Thrombogenic potential of transcatheter aortic valve implantation with trivial paravalvular leakage
Siegel, Rolland
2014-01-01
Background Significant paravalvular leakage after transcatheter aortic valve implantation (TAVI) correlates with increased morbidity and mortality, but adverse consequences of trivial paravalvular leakage have stimulated few investigations. Using a unique method distinctly different from other diagnostic approaches, we previously reported elevated backflow velocities of short duration (transients) in mechanical valve closure. In this study, similar transients were found in a transcatheter valve paravalvular leakage avatar. Methods Paravalvular leakage rate (zero to 58 mL/second) and aortic valve incompetence (volumetric back flow/forward flow; zero to 32%) were made adjustable using a mock transcatheter aortic valve device and tested in quasi-steady and pulsatile flow test systems. Projected dynamic valve area (PDVA) from the back illuminated mock transcatheter aortic valve device was measured and regional backflow velocities were derived by dividing volumetric flow rate by the PDVA over the open and closing valve phase and the total closed valve area derived from backflow leakage. Results Aortic incompetence from 1-32% generated negative backflow transients from 8 to 267 meters/second, a range not dissimilar to that measured in mechanical valves with zero paravalvular leakage. Optimal paravalvular leakage was identified; not too small generating high backflow transients, not too large considering volume overload and cardiac energy loss caused by defective valve behavior and fluid motion. Conclusions Thrombogenic potential of transcatheter aortic valves with trivial aortic incompetence and high magnitude regional backflow velocity transients was comparable to mechanical valves. This may have relevance to stroke rate, asymptomatic microembolic episodes and indications for anticoagulation therapy after transcatheter valve insertion. PMID:25333018
Analysis of Complex Valve and Feed Systems
NASA Technical Reports Server (NTRS)
Ahuja, Vineet; Hosangadi, Ashvin; Shipman, Jeremy; Cavallo, Peter; Dash, Sanford
2007-01-01
A numerical framework for analysis of complex valve systems supports testing of propulsive systems by simulating key valve and control system components in the test loop. In particular, it is designed to enhance the analysis capability in terms of identifying system transients and quantifying the valve response to these transients. This system has analysis capability for simulating valve motion in complex systems operating in diverse flow regimes ranging from compressible gases to cryogenic liquids. A key feature is the hybrid, unstructured framework with sub-models for grid movement and phase change including cryogenic cavitations. The multi-element unstructured framework offers improved predictions of valve performance characteristics under steady conditions for structurally complex valves such as pressure regulator valve. Unsteady simulations of valve motion using this computational approach have been carried out for various valves in operation at Stennis Space Center such as the split-body valve and the 10-in. (approx.25.4-cm) LOX (liquid oxygen) valve and the 4-in. (approx.10 cm) Y-pattern valve (liquid nitrogen). Such simulations make use of variable grid topologies, thereby permitting solution accuracy and resolving important flow physics in the seat region of the moving valve. An advantage to this software includes possible reduction in testing costs incurred due to disruptions relating to unexpected flow transients or functioning of valve/flow control systems. Prediction of the flow anomalies leading to system vibrations, flow resonance, and valve stall can help in valve scheduling and significantly reduce the need for activation tests. This framework has been evaluated for its ability to predict performance metrics like flow coefficient for cavitating venturis and valve coefficient curves, and could be a valuable tool in predicting and understanding anomalous behavior of system components at rocket propulsion testing and design sites.
HWANG, HUI-JEONG; YOON, KYUNG LIM; SOHN, IL SUK
2016-01-01
The present study reported the case of a 60-year-old female with patent ductus arteriosus (PDA) and a bicuspid aortic valve, who presented with transient severe left ventricular (LV) dysfunction following percutaneous closure of PDA, as identified by speckle tracking analysis. Transient LV dysfunction following PDA closure has previously been reported; however, severe LV dysfunction is rare. In the present case, the combination of a large PDA size, large amount of shunting, LV remodeling and bicuspid aortic valve may have induced serious deterioration of LV function following PDA closure. Furthermore, speckle-tracking echocardiography may be useful in the estimation of functional alterations in the myocardium of the LV following PDA closure. The observations detailed in the present study may improve the understanding of the pathophysiology and myocardial patterns of transient left ventricular dysfunction following PDA closure in adult humans. PMID:26998021
Hwang, Hui-Jeong; Yoon, Kyung Lim; Sohn, Il Suk
2016-03-01
The present study reported the case of a 60-year-old female with patent ductus arteriosus (PDA) and a bicuspid aortic valve, who presented with transient severe left ventricular (LV) dysfunction following percutaneous closure of PDA, as identified by speckle tracking analysis. Transient LV dysfunction following PDA closure has previously been reported; however, severe LV dysfunction is rare. In the present case, the combination of a large PDA size, large amount of shunting, LV remodeling and bicuspid aortic valve may have induced serious deterioration of LV function following PDA closure. Furthermore, speckle-tracking echocardiography may be useful in the estimation of functional alterations in the myocardium of the LV following PDA closure. The observations detailed in the present study may improve the understanding of the pathophysiology and myocardial patterns of transient left ventricular dysfunction following PDA closure in adult humans.
Espino, Daniel M; Shepherd, Duncan E T; Hukins, David W L
2014-01-01
A transient multi-physics model of the mitral heart valve has been developed, which allows simultaneous calculation of fluid flow and structural deformation. A recently developed contact method has been applied to enable simulation of systole (the stage when blood pressure is elevated within the heart to pump blood to the body). The geometry was simplified to represent the mitral valve within the heart walls in two dimensions. Only the mitral valve undergoes deformation. A moving arbitrary Lagrange-Euler mesh is used to allow true fluid-structure interaction (FSI). The FSI model requires blood flow to induce valve closure by inducing strains in the region of 10-20%. Model predictions were found to be consistent with existing literature and will undergo further development.
Numerical Modeling of Thermofluid Transients During Chilldown of Cryogenic Transfer Lines
NASA Technical Reports Server (NTRS)
Majumdar, Alok; Steadman, Todd
2003-01-01
The chilldown of fluid transfer lines is an important part of using cryogenic systems such as those found in both ground and space based applications. The chilldown process is a complex combination of both thermal and fluid transient phenomena. A cryogenic liquid flows through a transfer line that is initially at a much higher temperature than the cryogen. Transient heat transfer processes between the liquid and transfer line cause vaporization of the liquid, and this phase change can cause transient pressure and flow surges in the liquid. As the transfer line is cooled, these effects diminish until the liquid reaches a steady flow condition in the chilled transfer line. If these transient phenomena are not properly accounted for in the design process of a cryogenic system, it can lead to damage or failure of system components during operation. For such cases, analytical modeling is desirable for ensuring that a cryogenic system transfer line design is adequate for handling the effects of a chilldown process. The purpose of this paper is to present the results of a numerical model developed using Generalized Fluid System Simulation Program (GFSSP)'s new fluid transient capability in combination with its previously developed thermal transient capability to predict pressure and flow surge in cryogenic transfer lines during a chilldown process. An experiment performed by the National Bureau of Standards (NBS) in 1966 has been chosen as the baseline comparison case for this work. NBS s experimental set-up consisted of a 10.59 cubic foot supply dewar, an inlet valve, and a 200 foot long, in Outside Diameter (OD) vacuum jacketed copper transfer line that exhausted to atmosphere. Three different inlet valves, an in-port ball valve, a 1-in-port globe valve and a 1-in-port gate valve, were used in NBS's experiments. Experiments were performed using both liquid hydrogen and liquid nitrogen as the fluids. The proposed paper will include detailed comparisons of GFSSP's predictions with NBS's experimental results.
Variable reluctance proximity sensors for cryogenic valve position indication
NASA Technical Reports Server (NTRS)
Cloyd, R. A.
1982-01-01
A test was conducted to determine the performance of a variable reluctance proximity sensor system when installed in a space shuttle external tank vent/relief valve. The sensors were used as position indicators. The valve and sensors were cycled through a series of thermal transients; while the valve was being opened and closed pneumatically, the sensor's performance was being monitored. During these thermal transients, the vent valve was cooled ten times by liquid nitrogen and two times by liquid hydrogen. It was concluded that the sensors were acceptable replacements for the existing mechanical switches. However, the sensors need a mechanical override for the target similar to what is presently used with the mechanical switches. This override could insure contact between sensor and target and eliminate any problems of actuation gap growth caused by thermal gradients.
NASA Technical Reports Server (NTRS)
Mitchell, G. A.; Sanders, B. W.
1975-01-01
The throat of a Mach 2.5 inlet with a coldpipe termination was fitted with a stability-bypass system. System variations included several stability bypass entrance configurations. Poppet valves controlled the bypass airflow. The inlet stable airflow range achieved with each configuration was determined for both steady state conditions and internal pulse transients. Results are compared with those obtained without a stability bypass system. Transient results were also obtained for the inlet with a choke point at the diffuser exit and for the inlet with large and small stability bypass plenum volumes. Poppet valves at the stability bypass exit provided the inlet with a stable airflow range of 20 percent or greater at all static and transient conditions.
Waterhammer Transient Simulation and Model Anchoring for the Robotic Lunar Lander Propulsion System
NASA Technical Reports Server (NTRS)
Stein, William B.; Trinh, Huu P.; Reynolds, Michael E.; Sharp, David J.
2011-01-01
Waterhammer transients have the potential to adversely impact propulsion system design if not properly addressed. Waterhammer can potentially lead to system plumbing, and component damage. Multi-thruster propulsion systems also develop constructive/destructive wave interference which becomes difficult to predict without detailed models. Therefore, it is important to sufficiently characterize propulsion system waterhammer in order to develop a robust design with minimal impact to other systems. A risk reduction activity was performed at Marshall Space Flight Center to develop a tool for estimating waterhammer through the use of anchored simulation for the Robotic Lunar Lander (RLL) propulsion system design. Testing was performed to simulate waterhammer surges due to rapid valve closure and consisted of twenty-two series of waterhammer tests, resulting in more than 300 valve actuations. These tests were performed using different valve actuation schemes and three system pressures. Data from the valve characterization tests were used to anchor the models that employed MSCSoftware.EASY5 v.2010 to model transient fluid phenomena by using transient forms of mass and energy conservation. The anchoring process was performed by comparing initial model results to experimental data and then iterating the model input to match the simulation results with the experimental data. The models provide good correlation with experimental results, supporting the use of EASY5 as a tool to model fluid transients and provide a baseline for future RLL system modeling. This paper addresses tasks performed during the waterhammer risk reduction activity for the RLL propulsion system. The problem of waterhammer simulation anchoring as applied to the RLL system is discussed with results from the corresponding experimental valve tests. Important factors for waterhammer mitigation are discussed along with potential design impacts to the RLL propulsion system.
Luo, Ma-Ji; Chen, Guo-Hua; Ma, Yuan-Hao
2003-01-01
This paper presents a KIVA-3 code based numerical model for three-dimensional transient intake flow in the intake port-valve-cylinder system of internal combustion engine using body-fitted technique, which can be used in numerical study on internal combustion engine with vertical and inclined valves, and has higher calculation precision. A numerical simulation (on the intake process of a two-valve engine with a semi-sphere combustion chamber and a radial intake port) is provided for analysis of the velocity field and pressure field of different plane at different crank angles. The results revealed the formation of the tumble motion, the evolution of flow field parameters and the variation of tumble ratios as important information for the design of engine intake system.
NASA Technical Reports Server (NTRS)
Leggett, Nickolaus
1990-01-01
The ambient natural vacuum of space is proposed as a basis for electron valves. Each valve is an electron controlling structure similiar to a vacuum tube that is operated without a vacuum sustaining envelope. The natural vacuum electron valves discussed offer a viable substitute for solid state devices. The natural vacuum valve is highly resistant to ionizing radiation, system generated electromagnetic pulse, current transients, and direct exposure to space conditions.
Network Flow Simulation of Fluid Transients in Rocket Propulsion Systems
NASA Technical Reports Server (NTRS)
Bandyopadhyay, Alak; Hamill, Brian; Ramachandran, Narayanan; Majumdar, Alok
2011-01-01
Fluid transients, also known as water hammer, can have a significant impact on the design and operation of both spacecraft and launch vehicle propulsion systems. These transients often occur at system activation and shutdown. The pressure rise due to sudden opening and closing of valves of propulsion feed lines can cause serious damage during activation and shutdown of propulsion systems. During activation (valve opening) and shutdown (valve closing), pressure surges must be predicted accurately to ensure structural integrity of the propulsion system fluid network. In the current work, a network flow simulation software (Generalized Fluid System Simulation Program) based on Finite Volume Method has been used to predict the pressure surges in the feed line due to both valve closing and valve opening using two separate geometrical configurations. The valve opening pressure surge results are compared with experimental data available in the literature and the numerical results compared very well within reasonable accuracy (< 5%) for a wide range of inlet-to-initial pressure ratios. A Fast Fourier Transform is preformed on the pressure oscillations to predict the various modal frequencies of the pressure wave. The shutdown problem, i.e. valve closing problem, the simulation results are compared with the results of Method of Characteristics. Most rocket engines experience a longitudinal acceleration, known as "pogo" during the later stage of engine burn. In the shutdown example problem, an accumulator has been used in the feed system to demonstrate the "pogo" mitigation effects in the feed system of propellant. The simulation results using GFSSP compared very well with the results of Method of Characteristics.
NASA Astrophysics Data System (ADS)
Henclik, Sławomir
2018-03-01
The influence of dynamic fluid-structure interaction (FSI) onto the course of water hammer (WH) can be significant in non-rigid pipeline systems. The essence of this effect is the dynamic transfer of liquid energy to the pipeline structure and back, which is important for elastic structures and can be negligible for rigid ones. In the paper a special model of such behavior is analyzed. A straight pipeline with a steady flow, fixed to the floor with several rigid supports is assumed. The transient is generated by a quickly closed valve installed at the end of the pipeline. FSI effects are assumed to be present mainly at the valve which is fixed with a spring dash-pot attachment. Analysis of WH runs, especially transient pressure changes, for various stiffness and damping parameters of the spring dash-pot valve attachment is presented in the paper. The solutions are found analytically and numerically. Numerical results have been computed with the use of an own computer program developed on the basis of the four equation model of WH-FSI and the specific boundary conditions formulated at the valve. Analytical solutions have been found with the separation of variables method for slightly simplified assumptions. Damping at the dash-pot is taken into account within the numerical study. The influence of valve attachment parameters onto the WH courses was discovered and it was found the transient amplitudes can be reduced. Such a system, elastically attached shut-off valve in a pipeline or other, equivalent design can be a real solution applicable in practice.
NASA Technical Reports Server (NTRS)
Ferrera, J. D.
1972-01-01
The purpose of this report is to define and program the transient pneumatic flow equations necessary to determine, for a given set of conditions (geometry, pressures, temperatures, valve on time, etc.), the total nitrogen impulse and mass flow per pulse for the single pulsing of a Mariner type reaction control assembly valve. The rates of opening and closing of the valves are modeled, and electrical pulse durations from 20 to 100 ms are investigated. In developing the transient flow analysis, maximum use was made of the steady-state analysis. The impulse results are also compared to an equivalent square-wave impulse for both the Mariner Mars 1971 (MM'71) and Mariner Mars 1964 (MM'64) systems. It is demonstrated that, whereas in the MM'64 system, the actual impulse was as much as 56 percent higher than an assumed impulse (which is the product of the steady-state thrust and value on time i.e., the square wave), in the MM'71 system, these two values were in error in the same direction by only approximately 4 percent because of the larger nozzle areas and shorter valve stroke used.
NASA Astrophysics Data System (ADS)
Henclik, S.
2014-08-01
Transient flows in pipes (water hammer = WH) do appear in various situations and the accompanying pressure waves may involve serious perturbations in system functioning. To model these effects properly in the case of elastic pipe the dynamic fluid-structure interaction (FSI) should be taken into account. Fluid-structure couplings appear in various manners and the junction coupling is considered to be the strongest. This effect can be especially significant if the pipe can move as a whole body, which is possible when all its supports are not rigid. In the current paper a similar effect is numerically modelled. The pipe is fixed rigidly, but the valve at the end has a spring-dashpot mounting system, thus its motion is possible when WH is excited by the valve closuring. The boundary condition at the moving valve is modelled as a differential equation of motion. The valve hydraulic characteristics during closuring period are assumed by a time dependence of its loss factor. Preliminary numerical tests of that algorithm were done with an own computer program and it was found that the proper valve fixing system may produce significant lowering of WH pressures.
Reduced-order modellin for high-pressure transient flow of hydrogen-natural gas mixture
NASA Astrophysics Data System (ADS)
Agaie, Baba G.; Khan, Ilyas; Alshomrani, Ali Saleh; Alqahtani, Aisha M.
2017-05-01
In this paper the transient flow of hydrogen compressed-natural gas (HCNG) mixture which is also referred to as hydrogen-natural gas mixture in a pipeline is numerically computed using the reduced-order modelling technique. The study on transient conditions is important because the pipeline flows are normally in the unsteady state due to the sudden opening and closure of control valves, but most of the existing studies only analyse the flow in the steady-state conditions. The mathematical model consists in a set of non-linear conservation forms of partial differential equations. The objective of this paper is to improve the accuracy in the prediction of the HCNG transient flow parameters using the Reduced-Order Modelling (ROM). The ROM technique has been successfully used in single-gas and aerodynamic flow problems, the gas mixture has not been done using the ROM. The study is based on the velocity change created by the operation of the valves upstream and downstream the pipeline. Results on the flow characteristics, namely the pressure, density, celerity and mass flux are based on variations of the mixing ratio and valve reaction and actuation time; the ROM computational time cost advantage are also presented.
Simulations of Instabilities in Complex Valve and Feed Systems
NASA Technical Reports Server (NTRS)
Ahuja, Vineet; Hosangadi, Ashvin; Shipman, Jeremy; Cavallo, Peter A.
2006-01-01
CFD analyses are playing an increasingly important role in identifying and characterizing flow induced instabilities in rocket engine test facilities and flight systems. In this paper, we analyze instability mechanisms that range from turbulent pressure fluctuations due to vortex shedding in structurally complex valve systems to flow resonance in plug cavities to large scale pressure fluctuations due to collapse of cavitation induced vapor clouds. Furthermore, we discuss simulations of transient behavior related to valve motion that can serve as guidelines for valve scheduling. Such predictions of valve response to varying flow conditions is of crucial importance to engine operation and testing.
Transient Analysis of Pressurization and Pneumatic Subsystems of the X-34 Main Propulsion System
NASA Technical Reports Server (NTRS)
Hedayat, A.; Knight, K. C.; Chamption, R. H., Jr.; Kennedy, Jim W. (Technical Monitor)
2000-01-01
Transient models for the pressurization, vent/relief, and pneumatic subsystems of the X-34 Main Propulsion System are presented and simulation of their operation within prescribed requirements are provided. First, using ROCket Engine Transient Simulation (ROCETS) program, pressurization subsystem operation was simulated and helium requirements and the ullage thermodynamic condition within each propellant tank were calculated. Then, Overpressurization scenarios of propellant tanks and the response of vent/relief valves were evaluated using ROCETS simulation of simultaneous operation of the pressurization and vent/relief subsystems by incorporating the valves data into the model. Finally, the ROCETS simulation of in-flight operation of pneumatic subsystem predicted the overall helium consumption, Inter-Propellant Seal (IPS) purge flowrate and thermodynamic conditions, and Spin Start power.
Virtual solar field - An opportunity to optimize transient processes in line-focus CSP power plants
NASA Astrophysics Data System (ADS)
Noureldin, Kareem; Hirsch, Tobias; Pitz-Paal, Robert
2017-06-01
Optimizing solar field operation and control is a key factor to improve the competitiveness of line-focus solar thermal power plants. However, the risks of assessing new and innovative control strategies on operational power plants hinder such optimizations and result in applying more conservative control schemes. In this paper, we describe some applications for a whole solar field transient in-house simulation tool developed at the German Aerospace Centre (DLR), the Virtual Solar Field (VSF). The tool offers a virtual platform to simulate real solar fields while coupling the thermal and hydraulic conditions of the field with high computational efficiency. Using the tool, developers and operator can probe their control strategies and assess the potential benefits while avoiding the high risks and costs. In this paper, we study the benefits gained from controlling the loop valves and of using direct normal irradiance maps and forecasts for the field control. Loop valve control is interesting for many solar field operators since it provides a high degree of flexibility to the control of the solar field through regulating the flow rate in each loop. This improves the reaction to transient condition, such as passing clouds and field start-up in the morning. Nevertheless, due to the large number of loops and the sensitivity of the field control to the valve settings, this process needs to be automated and the effect of changing the setting of each valve on the whole field control needs to be taken into account. We used VSF to implement simple control algorithms to control the loop valves and to study the benefits that could be gained from using active loop valve control during transient conditions. Secondly, we study how using short-term highly spatially-resolved DNI forecasts provided by cloud cameras could improve the plant energy yield. Both cases show an improvement in the plant efficiency and outlet temperature stability. This paves the road for further investigations of new control strategies or for optimizations of the currently implemented ones.
Computational Modeling of Liquid and Gaseous Control Valves
NASA Technical Reports Server (NTRS)
Daines, Russell; Ahuja, Vineet; Hosangadi, Ashvin; Shipman, Jeremy; Moore, Arden; Sulyma, Peter
2005-01-01
In this paper computational modeling efforts undertaken at NASA Stennis Space Center in support of rocket engine component testing are discussed. Such analyses include structurally complex cryogenic liquid valves and gas valves operating at high pressures and flow rates. Basic modeling and initial successes are documented, and other issues that make valve modeling at SSC somewhat unique are also addressed. These include transient behavior, valve stall, and the determination of flow patterns in LOX valves. Hexahedral structured grids are used for valves that can be simplifies through the use of axisymmetric approximation. Hybrid unstructured methodology is used for structurally complex valves that have disparate length scales and complex flow paths that include strong swirl, local recirculation zones/secondary flow effects. Hexahedral (structured), unstructured, and hybrid meshes are compared for accuracy and computational efficiency. Accuracy is determined using verification and validation techniques.
NASA Technical Reports Server (NTRS)
Mitchell, G. A.; Sanders, B. W.
1974-01-01
The throat of a Mach 2.5 inlet with a coldpipe termination was fitted with a stability-bypass system. The inlet stable airflow range provided by various stability-bypass entrance configurations in alternate combination with several stability-bypass exit controls was determined for both steady-state conditions and internal transient pulses. Transient results were also obtained for the inlet with a choke point at the diffuser exit. Instart angles of attack were determined for the various stability-bypass entrance configurations. The response of the inlet-coldpipe system to internal and external oscillating disturbances was determined. Poppet valves at the stability-bypass exit provided an inlet stable airflow range of 28 percent or greater at all static and transient conditions.
An Analysis of the Impact of Valve Closure Time on the Course of Water Hammer
NASA Astrophysics Data System (ADS)
Kodura, Apoloniusz
2016-06-01
The knowledge of transient flow in pressure pipelines is very important for the designing and describing of pressure networks. The water hammer is the most common example of transient flow in pressure pipelines. During this phenomenon, the transformation of kinetic energy into pressure energy causes significant changes in pressure, which can lead to serious problems in the management of pressure networks. The phenomenon is very complex, and a large number of different factors influence its course. In the case of a water hammer caused by valve closing, the characteristic of gate closure is one of the most important factors. However, this factor is rarely investigated. In this paper, the results of physical experiments with water hammer in steel and PE pipelines are described and analyzed. For each water hammer, characteristics of pressure change and valve closing were recorded. The measurements were compared with the results of calculations perfomed by common methods used by engineers - Michaud's equation and Wood and Jones's method. The comparison revealed very significant differences between the results of calculations and the results of experiments. In addition, it was shown that, the characteristic of butterfly valve closure has a significant influence on water hammer, which should be taken into account in analyzing this phenomenon. Comparison of the results of experiments with the results of calculations? may lead to new, improved calculation methods and to new methods to describe transient flow.
NASA Technical Reports Server (NTRS)
Wichmann, Horst; Marquardt, Kaiser; Goforth, Alyssa
1993-01-01
Design of a direct-acting valve (DAV) for a primary thruster which is fully interchangeable with a thruster equipped with pilot-operated valves is described. The DAV is based on a bellows to isolate propellants form the actuator for maximum resistance to iron nitrate and other contamination and to select optimum materials for the actuator. It provides improved seal performance under all operating conditions and insensitivity to pressure transients. As compared with the existing pilot-operated valve, the DAV design is much simpler, consists of fewer parts, and will be lower in cost.
NASA Astrophysics Data System (ADS)
Zhang, Zh.
2016-11-01
Because of the limited value of the wave propagation speed in water the propagation of a pressure surge in transient flows can be tracked in the time series. This enables both the pressure head and the flow velocity in pipe flows to be determined as a function of both the coordinate along the pipe and the time. The propagation of the pressure surge includes both wave transmission and reflection. The latter occurs where the flow section is changed. The wave tracking method has been demonstrated as highly accurate and subsequently was applied to much more complex hydraulic systems, in which the pump is shut off and the spherical valve is simultaneously progressively closed. A combined four-quadrant characteristic of the pump and a spherical valve has been worked out, with which the computational procedure for the transient flow in the complex system could be significantly simplified. It has been demonstrated that not only the pressure surge in the hydraulic system but also the rotational speed of the pump could be satisfactorily computed. The computational algorithm has been demonstrated as quite simple, so that all calculations could be performed simply by means of the Microsoft Excel module.
Rodés-Cabau, Josep; Masson, Jean-Bernard; Welsh, Robert C; Garcia Del Blanco, Bruno; Pelletier, Marc; Webb, John G; Al-Qoofi, Faisal; Généreux, Philippe; Maluenda, Gabriel; Thoenes, Martin; Paradis, Jean-Michel; Chamandi, Chekrallah; Serra, Vicenç; Dumont, Eric; Côté, Mélanie
2017-07-10
The aim of this study was to compare aspirin plus clopidogrel with aspirin alone as antithrombotic treatment following transcatheter aortic valve replacement (TAVR) for the prevention of ischemic events, bleeding events, and death. Few data exist on the optimal antithrombotic therapy following TAVR. This was a randomized controlled trial comparing aspirin (80 to 100 mg/day) plus clopidogrel (75 mg/day) (dual antiplatelet therapy [DAPT]) versus aspirin alone (single-antiplatelet therapy [SAPT]) in patients undergoing TAVR with a balloon-expandable valve. The primary endpoint was the occurrence of death, myocardial infarction (MI), stroke or transient ischemic attack, or major or life-threatening bleeding (according to Valve Academic Research Consortium 2 definitions) within the 3 months following the procedure. The trial was prematurely stopped after the inclusion of 74% of the planned study population. A total of 222 patients were included, 111 allocated to DAPT and 111 to SAPT. The composite of death, MI, stroke or transient ischemic attack, or major or life-threatening bleeding tended to occur more frequently in the DAPT group (15.3% vs. 7.2%, p = 0.065). There were no differences between groups in the occurrence of death (DAPT, 6.3%; SAPT, 3.6%; p = 0.37), MI (DAPT, 3.6%; SAT, 0.9%; p = 0.18), or stroke or transient ischemic attack (DAPT, 2.7%; SAPT, 0.9%; p = 0.31) at 3 months. DAPT was associated with a higher rate of major or life-threatening bleeding events (10.8% vs. 3.6% in the SAPT group, p = 0.038). There were no differences between groups in valve hemodynamic status post-TAVR. This small trial showed that SAPT (vs. DAPT) tended to reduce the occurrence of major adverse events following TAVR. SAPT reduced the risk for major or life-threatening events while not increasing the risk for MI or stroke. Larger studies are needed to confirm these results. (Aspirin Versus Aspirin + Clopidogrel Following Transcatheter Aortic Valve Implantation: The ARTE Trial [ARTE], NCT01559298; Aspirin Versus Aspirin+Clopidogrel as Antithrombotic Treatment Following TAVI [ARTE], NCT02640794). Copyright © 2017 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Analysis of the STS-126 Flow Control Valve Structural-Acoustic Coupling Failure
NASA Technical Reports Server (NTRS)
Jones, Trevor M.; Larko, Jeffrey M.; McNelis, Mark E.
2010-01-01
During the Space Transportation System mission STS-126, one of the main engine's flow control valves incurred an unexpected failure. A section of the valve broke off during liftoff. It is theorized that an acoustic mode of the flowing fuel, coupled with a structural mode of the valve, causing a high cycle fatigue failure. This report documents the analysis efforts conducted in an attempt to verify this theory. Hand calculations, computational fluid dynamics, and finite element methods are all implemented and analyses are performed using steady-state methods in addition to transient analysis methods. The conclusion of the analyses is that there is a critical acoustic mode that aligns with a structural mode of the valve
Barnett, H. J.; Gent, M.; Sackett, D. L.; Taylor, D. W.
1980-01-01
Transient cerebral ischemia and stroke may occur despite previous occlusion of the artery supplying the appropriate part of the brain. After occlusion of the internal carotid artery, emboli may pass from the "stump" of the occluded artery to later produce transient cerebral ischemia or a stroke. Transient cerebral ischemia and stroke are due to a variety of conditions, some of which are strongly correlated with platelet thromboembolism, while others have little primary relation to thrombosis. The impact of this on therapeutic considerations is obvious. Thromboembolism related to myxomatous degeneration of the mitral valve (mitral valve prolapse) is a factor to consider in determining the cause of a stroke in a younger person. The prognosis for the different varieties of threatened stroke is incompletely known. When due to arteriosclerosis of the large cerebral arteries, transient cerebral ischemia and minor strokes pose a cumulative threat for a major stroke or death of 13% in the first year, 22% in the second year and 30% in the third year. Images FIG. 1 FIG. 2 FIG. 3 FIG. 4 PMID:6989464
NASA Technical Reports Server (NTRS)
Mazurkivich, Pete; Chandler, Frank; Grayson, Gary
2005-01-01
To meet the requirements for the 2nd Generation Reusable Launch Vehicle (RLV), a unique propulsion feed system concept was identified using crossfeed between the booster and orbiter stages that could reduce the Two-Stage-to-Orbit (TSTO) vehicle weight and development cost by approximately 25%. A Main Propulsion System (MPS) crossfeed water demonstration test program was configured to address all the activities required to reduce the risks for the MPS crossfeed system. A transient, one-dimensional system simulation was developed for the subscale crossfeed water flow tests. To ensure accurate representation of the crossfeed valve's dynamics in the system model, a high-fidelity, three-dimensional, computational fluid-dynamics (CFD) model was employed. The results from the CFD model were used to specify the valve's flow characteristics in the system simulation. This yielded a crossfeed system model that was anchored to the specific valve hardware and achieved good agreement with the measured test data. These results allowed the transient models to be correlated and validated and used for full scale mission predictions. The full scale model simulations indicate crossfeed is ' viable with the system pressure disturbances at the crossfeed transition being less than experienced by the propulsion system during engine start and shutdown transients.
Transient flow characteristics of a high speed rotary valve
NASA Astrophysics Data System (ADS)
Browning, Patrick H.
Pressing economic and environmental concerns related to the performance of fossil fuel burning internal combustion engines have revitalized research in more efficient, cleaner burning combustion methods such as homogeneous charge compression ignition (HCCI). Although many variations of such engines now exist, several limiting factors have restrained the full potential of HCCI. A new method patented by West Virginia University (WVU) called Compression Ignition by Air Injection (CIBAI) may help broaden the range of effective HCCI operation. The CIBAI process is ideally facilitated by operating two synchronized piston-cylinders mounted head-to-head with one of the cylinders filled with a homogeneous mixture of air and fuel and the other cylinder filled with air. A specialized valve called the cylinder connecting valve (CCV) separates the two cylinders, opens just before reaching top dead center (TDC), and allows the injection air into the charge to achieve autoignition. The CCV remains open during the entire power stroke such that upon ignition the rapid pressure rise in the charge cylinder forces mass flow back through the CCV into the air-only cylinder. The limited mass transfer between the cylinders through the CCV limits the theoretical auto ignition timing capabilities and thermal efficiency of the CIBAI cycle. Research has been performed to: (1) Experimentally measure the transient behavior of a potential CCV design during valve opening between two chambers maintained at constant pressure and again at constant volume; (2) Develop a modified theoretical CCV mass flow model based upon the measured cold flow valve performance that is capable of predicting the operating conditions required for successful mixture autoignition; (3) Make recommendations for future CCV designs to maximize CIBAI combustion range. Results indicate that the modified-ball CCV design offers suitable transient flow qualities required for application to the CIBAI concept. Mass injection events were experimentally mapped as a function of valve speed, inter-cylinder pressure ratios and volume ratios and the results were compared to compressible flow theoretical models. Specifically, the transient behavior suggested a short-lived loss-mode initiation closely resembled by shock tube theory followed by a quasi-steady flow regime resembling choked flow behavior. An empirical model was then employed to determine the useful range of the CCV design as applied to a four-stroke CIBAI engine cycle modeled using a 1-D quasi-steady numerical method, with particular emphasis on the cyclic timing of the CCV opening. Finally, a brief discussion of a high-temperature version of the CCV design is presented.
Application of several variable-valve-timing concepts to an LHR engine
NASA Technical Reports Server (NTRS)
Morel, T.; Keribar, R.; Sawlivala, M.; Hakim, N.
1987-01-01
The paper discusses advantages provided by electronically controlled hydraulically activated valves (ECVs) when applied to low heat rejection (LHR) engines. The ECV concept provides additional engine control flexibility by allowing for a variable valve timing as a function of speed and load, or for a given transient condition. The results of a study carried out to assess the benefits that this flexibility can offer to an LHR engine indicated that, when judged on the benefits to BSFC, volumetric efficiency, and peak firing pressure, ECVs would provide only modest benefits in comparison to conventional valve profiles. It is noted, however, that once installed on the engine, the ECVs would permit a whole range of certain more sophisticated variable valve timing strategies not otherwise possible, such as high compression cranking, engine braking, cylinder cutouts, and volumetric efficiency timing with engine speed.
Inflammatory Regulation of Valvular Remodeling: The Good(?), the Bad, and the Ugly
Mahler, Gretchen J.; Butcher, Jonathan T.
2011-01-01
Heart valve disease is unique in that it affects both the very young and very old, and does not discriminate by financial affluence, social stratus, or global location. Research over the past decade has transformed our understanding of heart valve cell biology, yet still more remains unclear regarding how these cells respond and adapt to their local microenvironment. Recent studies have identified inflammatory signaling at nearly every point in the life cycle of heart valves, yet its role at each stage is unclear. While the vast majority of evidence points to inflammation as mediating pathological valve remodeling and eventual destruction, some studies suggest inflammation may provide key signals guiding transient adaptive remodeling. Though the mechanisms are far from clear, inflammatory signaling may be a previously unrecognized ally in the quest for controlled rapid tissue remodeling, a key requirement for regenerative medicine approaches for heart valve disease. This paper summarizes the current state of knowledge regarding inflammatory mediation of heart valve remodeling and suggests key questions moving forward. PMID:21792386
NASA Technical Reports Server (NTRS)
Schultz, D. F.
1971-01-01
The steady state thermodynamic cycle balance of the single preburner staged combustion engine, coupled with dynamic transient analyses, dictated in detail the location and requirements for each valve defined in this volume. Valve configuration selections were influenced by overall engine and vehicle system weight and failure mode determinations. Modulating valve actuators are external to the valve and are line replaceable. Development and satisfactory demonstration of a high pressure dynamic shaft seal has made this configuration practical. Pneumatic motor driven actuators that use engine pumped hydrogen gas as the working fluid are used. The helium control system is proposed as a module containing a cluster of solenoid actuated valves. The separable couplings and flanges are designed to assure minimum leakage with minimum coupling weight. The deflection of the seal surface in the flange is defined by finite element analysis that has been confirmed with test data. The seal design proposed has passed preliminary pressure cycling and thermal cycling tests.
NASA Technical Reports Server (NTRS)
Cole, G. L.; Dustin, M. O.; Neiner, G. H.
1975-01-01
Results of a wind tunnel investigation are presented. The inlet was modified so that airflow can be removed through a porous cowl-bleed region in the vicinity of the throat. Bleed plenum exit flow area is controlled by relief type mechanical valves. Unlike valves in previous systems, these are made for use in a high Mach flight environment and include refinements so that the system could be tested on a NASA YF-12 aircraft. The valves were designed to provide their own reference pressure. The results show that the system can absorb internal-airflow-transients that are too fast for a conventional bypass door control system and that the two systems complement each other quite well. Increased tolerance to angle of attack and Mach number changes is indicated. The valves should provide sufficient time for the inlet control system to make geometry changes required to keep the inlet started.
NASA Technical Reports Server (NTRS)
Korte, John J.
1990-01-01
A numerical simulation of the actuation system for the propulsion control valve (PCV) of the NASA Langley Aircraft Landing Dynamics Facility was developed during the preliminary design of the PCV and used throughout the entire project. The simulation is based on a predictive model of the PCV which is used to evaluate and design the actuation system. The PCV controls a 1.7 million-pound thrust water jet used in propelling a 108,000-pound test carriage. The PCV can open and close in 0.300 second and deliver over 9,000 gallons of water per sec at pressures up to 3150 psi. The numerical simulation results are used to predict transient performance and valve opening characteristics, specify the hydraulic control system, define transient loadings on components, and evaluate failure modes. The mathematical model used for numerically simulating the mechanical fluid power system is described, and numerical results are demonstrated for a typical opening and closing cycle of the PCV. A summary is then given on how the model is used in the design process.
Unsteady Analyses of Valve Systems in Rocket Engine Testing Environments
NASA Technical Reports Server (NTRS)
Shipman, Jeremy; Hosangadi, Ashvin; Ahuja, Vineet
2004-01-01
This paper discusses simulation technology used to support the testing of rocket propulsion systems by performing high fidelity analyses of feed system components. A generalized multi-element framework has been used to perform simulations of control valve systems. This framework provides the flexibility to resolve the structural and functional complexities typically associated with valve-based high pressure feed systems that are difficult to deal with using traditional Computational Fluid Dynamics (CFD) methods. In order to validate this framework for control valve systems, results are presented for simulations of a cryogenic control valve at various plug settings and compared to both experimental data and simulation results obtained at NASA Stennis Space Center. A detailed unsteady analysis has also been performed for a pressure regulator type control valve used to support rocket engine and component testing at Stennis Space Center. The transient simulation captures the onset of a modal instability that has been observed in the operation of the valve. A discussion of the flow physics responsible for the instability and a prediction of the dominant modes associated with the fluctuations is presented.
Modeling and Optimization for Management of Intermittent Water Supply
NASA Astrophysics Data System (ADS)
Lieb, A. M.; Wilkening, J.; Rycroft, C.
2014-12-01
In many urban areas, piped water is supplied only intermittently, as valves direct water to different parts of the water distribution system at different times. The flow is transient, and may transition between free-surface and pressurized, resulting in complex dynamical features with important consequences for water suppliers and users. These consequences include degradation of distribution system components, compromised water quality, and inequitable water availability. The goal of this work is to model the important dynamics and identify operating conditions that mitigate certain negative effects of intermittent water supply. Specifically, we will look at controlling valve parameters occurring as boundary conditions in a network model of transient, transition flow through closed pipes. Gradient-based optimization will be used to find boundary values to minimize pressure gradients and ensure equitable water availability at system endpoints.
Optimal Dynamics of Intermittent Water Supply
NASA Astrophysics Data System (ADS)
Lieb, Anna; Wilkening, Jon; Rycroft, Chris
2014-11-01
In many urban areas of the developing world, piped water is supplied only intermittently, as valves direct water to different parts of the water distribution system at different times. The flow is transient, and may transition between free-surface and pressurized, resulting in complex dynamical features with important consequences for water suppliers and users. These consequences include degradation of distribution system components, compromised water quality, and inequitable water availability. The goal of this work is to model the important dynamics and identify operating conditions that mitigate certain negative effects of intermittent water supply. Specifically, we will look at valve parameters occurring as boundary conditions in a network model of transient, transition flow through closed pipes. Optimization will be used to find boundary values to minimize pressure gradients and ensure equitable water availability.
NASA Technical Reports Server (NTRS)
Sanders, B. W.
1980-01-01
The throat of a Mach 2.5 inlet that was attached to a turbojet engine was fitted with a poppet-valve-controlled stability bypass system that was designed to provide a large, stable airflow range. Propulsion system response and stability bypass performance were determined for several transient airflow disturbances, both internal and external. Internal airflow disturbances included reductions in overboard bypass airflow, power lever angle, and primary-nozzle area as well as compressor stall. For reference, data are also included for a conventional, fixed-exit bleed system. The poppet valves greatly increased inlet stability and had no adverse effects on propulsion system performance. Limited unstarted-inlet bleed performance data are presented.
Cardiac Calcified Amorphous Tumor of the Mitral Valve Presenting as Transient Ischemic Attack
Abbasi Teshnizi, Mohammad; Ghorbanzadeh, Atefeh; Zirak, Nahid; Manafi, Babak
2017-01-01
Cardiac calcified amorphous tumors (CATs) are an extremely rare nonneoplastic intracardiac masses. They have been reported in the literature in only a few cases. Thus, the incidence, pathogenesis, and best approach to the treatment are not certain. We report a case of CATs on the atrial surface of the anterior mitral valve leaflet in a 37-year-old female who was diagnosed by histopathological examination after surgical removal. PMID:28194283
Cardiac Calcified Amorphous Tumor of the Mitral Valve Presenting as Transient Ischemic Attack.
Abbasi Teshnizi, Mohammad; Ghorbanzadeh, Atefeh; Zirak, Nahid; Manafi, Babak; Moeinipour, Aliasghar
2017-01-01
Cardiac calcified amorphous tumors (CATs) are an extremely rare nonneoplastic intracardiac masses. They have been reported in the literature in only a few cases. Thus, the incidence, pathogenesis, and best approach to the treatment are not certain. We report a case of CATs on the atrial surface of the anterior mitral valve leaflet in a 37-year-old female who was diagnosed by histopathological examination after surgical removal.
Water-hammer pressure waves interaction at cross-section changes in series in viscoelastic pipes
NASA Astrophysics Data System (ADS)
Meniconi, S.; Brunone, B.; Ferrante, M.
2012-08-01
In view of scarcity of both experimental data and numerical models concerning transient behavior of cross-section area changes in pressurized liquid flow, the paper presents laboratory data and numerical simulation of the interaction of a surge wave with a partial blockage by a valve, a single pipe contraction or expansion and a series of pipe contraction/expansion in close proximity.With regard to a single change of cross-section area, laboratory data point out the completely different behavior with respect to one of the partially closed in-line valves with the same area ratio. In fact, for the former the pressure wave interaction is not regulated by the steady-state local head loss. With regard to partial blockages, transient tests have shown that the smaller the length, the more intense the overlapping of pressure waves due to the expansion and contraction in series.Numerically, the need for taking into account both the viscoelasticity and unsteady friction is demonstrated, since the classical water-hammer theory does not simulate the relevant damping of pressure peaks and gives rise to a time shifting between numerical and laboratory data. The transient behavior of a single local head loss has been checked by considering tests carried out in a system with a partially closed in-line valve. As a result, the reliability of the quasi steady-state approach for local head loss simulation has been demonstrated in viscoelastic pipes. The model parameters obtained on the basis of transients carried out in single pipe systems have then been used to simulate transients in the more complex pipe systems. These numerical experiments show the great importance of the length of the small-bore pipe with respect to one of the large-bore pipes. Precisely, until a gradually flow establishes in the small-bore pipe, the smaller such a length, the better the quality of the numerical simulation.
NASA Astrophysics Data System (ADS)
Canetta, D.; Capozza, A.; Iovino, G.
The transient response following pump trip-offs and start-ups was investigated in the sea water system of a nuclear power plant. Specific care was devoted to water column separation and cavity collapse phenomena. A computer program designed for analysis of complex hydraulic networks was used. It is found that dangerous overpressures can be avoided by the use of loop seals. The design of the vacuum breaker valves of the loop seals and the optimization of overall transient behavior is discussed.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Garbett, K; Mendler, O J; Gardner, G C
In PWR steam generator tube rupture (SGTR) faults, a direct pathway for the release of radioactive fission products can exist if there is a coincident stuck-open safety relief valve (SORV) or if the safety relief valve is cycled. In addition to the release of fission products from the bulk steam generator water by moisture carryover, there exists the possibility that some primary coolant may be released without having first mixed with the bulk water - a process called primary coolant bypassing. The MB-2 Phase II test program was designed specifically to identify the processes for droplet carryover during SGTR faultsmore » and to provide data of sufficient accuracy for use in developing physical models and computer codes to describe activity release. The test program consisted of sixteen separate tests designed to cover a range of steady-state and transient fault conditions. These included a full SGTR/SORV transient simulation, two SGTR overfill tests, ten steady-state SGTR tests at water levels ranging from very low levels in the bundle up to those when the dryer was flooded, and three moisture carryover tests without SGTR. In these tests the influence of break location and the effect of bypassing the dryer were also studied. In a final test the behavior with respect to aerosol particles in a dry steam generator, appropriate to a severe accident fault, was investigated.« less
Cavitating flow during water hammer using a generalized interface vaporous cavitation model
NASA Astrophysics Data System (ADS)
Sadafi, Mohamadhosein; Riasi, Alireza; Nourbakhsh, Seyed Ahmad
2012-10-01
In a transient flow simulation, column separation may occur when the calculated pressure head decreases to the saturated vapor pressure head in a computational grid. Abrupt valve closure or pump failure can result in a fast transient flow with column separation, potentially causing problems such as pipe failure, hydraulic equipment damage, cavitation or corrosion. This paper reports a numerical study of water hammer with column separation in a simple reservoir-pipeline-valve system and pumping station. The governing equations for two-phase transient flow in pipes are solved based on the method of characteristics (MOC) using a generalized interface vaporous cavitating model (GIVCM). The numerical results were compared with the experimental data for validation purposes, and the comparison indicated that the GIVCM describes the experimental results more accurately than the discrete vapor cavity model (DVCM). In particular, the GIVCM correlated better with the experimental data than the DVCM in terms of timing and pressure magnitude. The effects of geometric and hydraulic parameters on flow behavior in a pumping station with column separation were also investigated in this study.
Thrombolytic therapy for mitral valve thrombosis.
Lin, T K; Tsai, L M; Chen, J H; Yang, Y J
1997-05-01
A 44-year-old man with a St. Jude mitral valve was admitted because of progressive pulmonary edema. He was diagnosed with prosthetic heart valve thrombosis (PHVT) based on the findings of "muffled" prosthetic valve clicks. Doppler echocardiographic evidence of severe mitral stenosis and transesophageal echocardiographic evidence of limited mitral valve motility. Because the patient hesitated to undergo our recommended surgical treatment, he was immediately treated with intravenous recombinant tissue plasminogen activator (100 mg over 3 h) followed by heparinization. Two hours after the thrombolytic therapy, the prosthetic valve clicks became clearly audible and his congestive symptoms were dramatically improved. Follow-up echocardiography no longer-showed significant mitral valve obstruction. A transient cerebral ischemic attack occurred at the end of thrombolytic therapy but there were no neurologic sequalae. The patient, on warfarin therapy, was well at follow-up 8 months after discharge. Surgical intervention has long been the standard therapy for patients with PHVT. Our case experience suggests that thrombolytic therapy may be considered as an effective alternative to surgical intervention for selected patients with PHVT. In this report, we also review the current literature regarding the indications, effectiveness and safety of thrombolytic therapy in PHVT.
Parametric study on the performance of automotive MR shock absorbers
NASA Astrophysics Data System (ADS)
Gołdasz, J.; Dzierżek, S.
2016-09-01
The paper contains the results of a parametric study to explore the influence of various quantities on the performance range of semi-active automotive shock absorbers using the magnetorheological (MR) fluid under steady-state and transient excitations. The analysis was performed with simulated data and using a standard single-tube shock absorber configuration with a single-gap MR valve. Additionally, the impact of material variables and valves geometry was examined as the parameters were varied and its dynamic range studied.
CFD Analysis of the Anti-Surge Effects by Water Hammering
NASA Astrophysics Data System (ADS)
Kim, Tae-oh; Jeong, Hyo-min; Chung, Han-shik; Lee, Sin-il; Lee, Kwang-sung
2015-09-01
Water hammering occurs due to the surge effect that comes from operating the pump, sudden stop during the operating due to a blackout and rapid open and close of the valve. By the water hammering of the pipeline and the pump, the valve are damaged. In this paper, transient analysis is conducted by CFD (Computational Fluid Dynamics). The purpose of this paper is to provide the research data about the change of the pressure and flow in the pipe that caused by the water hammering.
NASA Astrophysics Data System (ADS)
Li, D. D.; Jiang, J.; Zhao, Z.; Yi, W. S.; Lan, G.
2013-12-01
We take a concrete pumping station as an example in this paper. Through the calculation of water hammer protection with a specific pumping station water supply project, and the analysis of the principle, mathematical models and boundary conditions of air vessel and over-pressure relief valve we show that the air vessel can protect the water conveyance system and reduce the transient pressure damage due to various causes. Over-pressure relief valve can effectively reduce the water hammer because the water column re-bridge suddenly stops the pump and prevents pipeline burst. The paper indicates that the combination set of air vessel and over-pressure relief valve can greatly reduce the quantity of the air valve and can eliminate the water hammer phenomenon in the pipeline system due to the vaporization and water column separation and re-bridge. The conclusion could provide a reference for the water hammer protection of long-distance water supply system.
Transient chaos and crisis phenomena in butterfly valves driven by solenoid actuators
NASA Astrophysics Data System (ADS)
Naseradinmousavi, Peiman; Nataraj, C.
2012-11-01
Chilled water systems used in the industry and on board ships are critical for safe and reliable operation. It is hence important to understand the fundamental physics of these systems. This paper focuses in particular on a critical part of the automation system, namely, actuators and valves that are used in so-called "smart valve" systems. The system is strongly nonlinear, and necessitates a nonlinear dynamic analysis to be able to predict all critical phenomena that affect effective operation and efficient design. The derived mathematical model includes electromagnetics, fluid mechanics, and mechanical dynamics. Nondimensionalization has been carried out in order to reduce the large number of parameters to a few critical independent sets to help carry out a broad parametric analysis. The system stability analysis is then carried out with the aid of the tools from nonlinear dynamic analysis. This reveals that the system is unstable in a certain region of the parameter space. The system is also shown to exhibit crisis and transient chaotic responses; this is characterized using Lyapunov exponents and power spectra. Knowledge and avoidance of these dangerous regimes is necessary for successful and safe operation.
Li, Wendy F; Pollard, Heidi; Karimi, Mohsen; Asnes, Jeremy D; Hellenbrand, William E; Shabanova, Veronika; Weismann, Constance G
2018-01-01
Trans-catheter (TC) pulmonary valve replacement (PVR) has become common practice for patients with right ventricular outflow tract obstruction (RVOTO) and/or pulmonic insufficiency (PI). Our aim was to compare PVR and right ventricular (RV) function of patients who received TC vs surgical PVR. Retrospective review of echocardiograms obtained at three time points: before, immediately after PVR, and most recent. Sixty-two patients (median age 19 years, median follow-up 25 months) following TC (N = 32) or surgical (N = 30) PVR at Yale-New Haven Hospital were included. Pulmonary valve and right ventricular function before, immediately after, and most recently after PVR. At baseline, the TC group had predominant RVOTO (74% vs 10%, P < .001), and moderate-severe PI was less common (61% vs 100%, P < .001). Immediate post-procedural PVR function was good throughout. At last follow-up, the TC group had preserved valve function, but the surgical group did not (moderate RVOTO: 6% vs 41%, P < .001; >mild PI: 0% vs 24%, P = .003). Patients younger than 17 years at surgical PVR had the highest risk of developing PVR dysfunction, while PVR function in follow-up was similar in adults. Looking at RV size and function, both groups had a decline in RV size following PVR. However, while RV function remained stable in the TC group, there was a transient postoperative decline in the surgical group. TC PVR in patients age <17 years is associated with better PVR function in follow-up compared to surgical valves. There was a transient decline in RV function following surgical but not TC PVR. TC PVR should therefore be the first choice in children who are considered for PVR, whenever possible. © 2017 Wiley Periodicals, Inc.
Role of vortices in cavitation formation in the flow across a mechanical heart valve.
Li, Chi-Pei; Lu, Po-Chien; Liu, Jia-Shing; Lo, Chi-Wen; Hwang, Ned H
2008-07-01
Cavitation occurs during mechanical heart valve closure when the local pressure drops below vapor pressure. The formation of stable gas bubbles may result in gaseous emboli, and secondarily cause transient ischemic attacks or strokes. It is noted that instantaneous valve closure, occluder rebound and high-speed leakage flow generate vortices that promote low-pressure regions in favor of stable bubble formation; however, to date no studies have been conducted for the quantitative measurement and analysis of these vortices. A Björk-Shiley Monostrut (BSM) monoleaflet valve was placed in the mitral position of a pulsatile mock circulatory loop. Particle image velocimetry (PIV) and pico coulomb (PCB) pressure measurements were applied. Flow field measurements were carried out at t = -5, -3, -1, -0.5, 0 (valve closure), 0.3, 0.5, 0.75, 1.19, 1.44, 1.69, 1.94, 2, 2.19, 2.54, 2.79, 3.04, 3.29, 3.54, 5 and 10 ms. The vortices were quantitatively analyzed using the Rankine vortex model. A single counter-clockwise vortex was The instantaneous formation of cavitation bubbles at mechanical heart valve (MHV) closure, which subsequently damage blood cells and valve integrity, is a well-known and widely studied phenomenon (1-4). Contributing factors seem to include the water-hammer, squeeze flow and Venturi effects, all of which are short-lived. Both, Dauzat et al. (5) and Sliwka et al. (6) have detected high-intensity transient signals (HITS) with transcranial Doppler ultrasound in the carotid and cerebral arteries of MHV recipients, while Deklunder (7) observed clinical occurrences of cerebral gas emboli that were not seen with bioprosthetic valves. These detected over the major orifice, while a pair of counter-rotating vortices was found over the minor orifice. Velocity profiles were consistent with Rankine vortices. The vortex strength and magnitude of the pressure drop peaked shortly after initial occluder-housing impact and rapidly decreased after 0.5 ms, indicating viscous dissipation, with a less significant contribution from the occluder rebound effect. The maximum pressure drop was on the order of magnitude of 40 mmHg. Detailed PIV measurements and quantitative analysis of the BSM mechanical heart valve revealed large-scale vortex formation immediately after valve closure. Of note, the vortices were typical of a Rankine vortex and the maximum pressure change at the vortex center was only 40 mmHg. These data support the conclusion that vortex formation alone cannot generate the magnitude of pressure drop required for cavitation bubble formation.
CONTEMPT/LT-028 Browns Ferry studies of an anticipated transient without scram
DOE Office of Scientific and Technical Information (OSTI.GOV)
Holcomb, E.E.
1983-01-01
The Browns Ferry Nuclear Plant containment response during the first 30 min of an anticipated transient without scram (ATWS) is the subject of this paper. Three cases, each initiated by a main steam isolation valve closure, are presented: the ATWS is mitigated by operator actions in the spirit of the General Electric Emergency Procedure Guidelines; the ATWS is managed by the plant automatic control systems; and the ATWS proceeds as in first case except that the drywell coolers are unavailable. Success of the standby liquid control system is assumed in the last two transients.
CardiaMed mechanical valve: mid-term results of a multicenter clinical trial.
Nazarov, Vladimir M; Zheleznev, Sergey I; Bogachev-Prokophiev, Alexandr V; Afanasyev, Alexandr V; Nemchenko, Eugene V; Jeltovskiy, Yuri V; Lavinyukov, Sergey O
2014-01-01
Prosthesis choice is a major concern in valvular surgery. A multicenter clinical trial was performed to assess the efficacy and safety of the CardiaMed prosthetic heart valve. The study enrolled 420 patients who underwent mitral (209) or aortic (211) valve replacement from 2003 to 2004 at 7 institutions in Russia, and who were followed up from 2006 to 2011. The mean age was 52.2 ± 10.2 years (range, 12-78 years), 47.4% were female, and 99.05% completed the study. The maximum observation term was 7.5 years (2188.5 patient-years); 1081.6 patient-years for aortic and 1106.9 patient-years for mitral valve replacement. The overall 7-year survival rate was 85.1% ± 3.7%; 86.1% ± 4.8% and 84.4% ± 5.4% for aortic and mitral valve replacement, respectively. The 7-year freedom from valve-related death was 93.9% ± 3.7% and 94.5% ± 3.2% for aortic and mitral valve replacement, respectively. When early mortality (<30 days) was excluded, these rates were 94.8% ± 3.1% and 93.8% ± 3.82%, respectively. Linearized valve-dependent complication rates were determined for structural valve failure (0%/patient-year overall), thrombosis (0.63%/patient-year, all for mitral valve replacement), thromboembolic complications including transient neurologic deficits (0.13%/patient-year overall, 0.5%/patient-year for aortic valve replacement, 0.8%/patient-year for mitral valve replacement), hemorrhagic bleeding (0.64%/patient-year overall, 0.55%/patient-year for aortic valve replacement, 0.09%/patient-year for mitral valve replacement), prosthetic endocarditis (0.28%/patient-year overall, 0.28%/patient-year for aortic valve replacement, 0%/patient-year for mitral valve replacement), and hemolysis (0%/patient-year overall). The CardiaMed mechanical heart valve prostheses meets world standards of safety and efficacy.
NASA Technical Reports Server (NTRS)
Atwell, Matthew J.; Melcher, John C.; Hurlbert, Eric A.; Morehead, Robert L.
2017-01-01
A liquid oxygen, liquid methane (LOX/LCH4) reaction control system (RCS) was tested at NASA Glenn Research Center's Plum Brook Station in the Spacecraft Propulsion Research Facility (B-2) under simulated altitude and thermal vacuum conditions. The RCS is a subsystem of the Integrated Cryogenic Propulsion Test Article (ICPTA) and was initially developed under Project Morpheus. Composed of two 28 lbf-thrust and two 7 lbf-thrust engines, the RCS is fed in parallel with the ICPTA main engine from four propellant tanks. 40 tests consisting of 1,010 individual thruster pulses were performed across 6 different test days. Major test objectives were focused on system dynamics, and included characterization of fluid transients, manifold priming, manifold thermal conditioning, thermodynamic vent system (TVS) performance, and main engine/RCS interaction. Peak surge pressures from valve opening and closing events were examined. It was determined that these events were impacted significantly by vapor cavity formation and collapse. In most cases the valve opening transient was more severe than the valve closing. Under thermal vacuum conditions it was shown that TVS operation is unnecessary to maintain liquid conditions at the thruster inlets. However, under higher heat leak environments the RCS can still be operated in a self-conditioning mode without overboard TVS venting, contingent upon the engines managing a range of potentially severe thermal transients. Lastly, during testing under cold thermal conditions the engines experienced significant ignition problems. Only after warming the thruster bodies with a gaseous nitrogen purge to an intermediate temperature was successful ignition demonstrated.
Micro-valve using induced-charge electrokinetic motion of Janus particle.
Daghighi, Yasaman; Li, Dongqing
2011-09-07
A new micro-valve using the electrokinetic motion of a Janus particle is introduced in this paper. A Janus particle with a conducting hemisphere and a non-conducting hemisphere is placed in a junction of several microchannels. Under an applied electric field, the induced-charge electrokinetic flow around the conducting side of the Janus particle forms vortices. The vortices push the particle moving forwards to block the entrance of a microchannel. By switching the direction of the applied electric field, the motion of the Janus particle can be changed to block different microchannels. This paper develops a theoretical model and conducts numerical simulations of the three-dimensional transient motion of the Janus particle. The results show that this Janus particle-based micro-valve is feasible for switching and controlling the flow rate in a microfluidic chip. This method is simple in comparison with other types of micro-valve methods. It is easy for fabrication, for operation control, and has a fast response time. To better understand the micro-valve functions, comparisons with a non-conducting particle and a fully conducting particle were made. Results proved that only a Janus particle can fulfill the requirements of such a micro-valve.
Optimization of valve opening process for the suppression of impulse exhaust noise
NASA Astrophysics Data System (ADS)
Li, Jingxiang; Zhao, Shengdun
2017-02-01
Impulse exhaust noise generated by the sudden impact of discharging flow of pneumatic systems has significant temporal characteristics including high sound pressure and rapid sound transient. The impulse noise exposures are more hazardous to hearing than the energy equivalent uniform noise exposures. This paper presents a novel approach to suppress the peak sound pressure as a major indicator of impulsiveness of the impulse exhaust noise by an optimization of the opening process of valve. Relationships between exhaust flow and impulse noise are described by thermodynamics and noise generating mechanism. Then an optimized approach by controlling the valve opening process is derived under a constraint of pre-setting exhaust time. A modified servo-direct-driven valve was designed and assembled in a typical pneumatic system for the verification experiments comparing with an original solenoid valve. Experimental results with groups of initial cylinder pressures and pre-setting exhaust times are shown to verify the effects of the proposed optimization. Some indicators of energy-equivalent and impulsiveness are introduced to discuss the effects of the noise suppressions. Relationship between noise reduction and exhaust time delay is also discussed.
Jugular veins in transient global amnesia: innocent bystanders.
Baracchini, Claudio; Tonello, Simone; Farina, Filippo; Viaro, Federica; Atzori, Matteo; Ballotta, Enzo; Manara, Renzo
2012-09-01
Transient global amnesia (TGA) has been associated with an increased prevalence of internal jugular valve insufficiency and many patients report Valsalva-associated maneuvers before TGA onset. These findings have led to the assumption of hemodynamic alterations in intracranial veins inducing focal hippocampal ischemia. We investigated this hypothesis in patients with TGA and control subjects. Seventy-five patients with TGA and 75 age- and sex-matched healthy subjects were enrolled into a cross-sectional study. Extracranial and transcranial high-resolution venous echo-color-Doppler sonography was performed blindly in all patients and control subjects. Blood flow direction and velocities were recorded at the internal jugular veins, basal veins of Rosenthal, and vein of Galen, both at rest and during Valsalva-associated maneuvers. Mean age of patients with TGA was 60.3±8.0 years (median, 60 years; range, 44-78 years); 44 (59%) were female (female/male ratio: 1.42). Internal jugular valve insufficiency (left, right, or bilateral) was found to be more frequent in patients with TGA than in control subjects: 53 (70.7%) versus 22 (29.3%; P<0.05). Blood flow velocities in the deep cerebral veins of patients with TGA did not differ from control subjects both at rest and during Valsalva-associated maneuvers. Intracranial venous reflux was neither observed in patients with TGA nor in control subjects despite unilateral or bilateral internal jugular valve insufficiency during prolonged and maximal Valsalva-associated maneuvers. This study, although confirming the association between TGA and internal jugular valve insufficiency, challenges the hypothesis that cerebral venous congestion plays a significant role in the pathogenesis of TGA.
Analytical Solution to the Pneumatic Transient Rod System at ACRR
DOE Office of Scientific and Technical Information (OSTI.GOV)
Fehr, Brandon Michael
2016-01-08
The ACRR pulse is pneumatically driven by nitrogen in a system of pipes, valves and hoses up to the connection of the pneumatic system and mechanical linkages of the transient rod (TR). The main components of the TR pneumatic system are the regulator, accumulator, solenoid valve and piston-cylinder assembly. The purpose of this analysis is to analyze the flow of nitrogen through the TR pneumatic system in order to develop a motion profile of the piston during the pulse and be able to predict the pressure distributions inside both the cylinder and accumulators. The predicted pressure distributions will be validatedmore » against pressure transducer data, while the motion profile will be compared to proximity switch data. By predicting the motion of the piston, pulse timing will be determined and provided to the engineers/operators for verification. The motion profile will provide an acceleration distribution to be used in Razorback to more accurately predict reactivity insertion into the system.« less
TRACE Model for Simulation of Anticipated Transients Without Scram in a BWR
DOE Office of Scientific and Technical Information (OSTI.GOV)
Cheng L. Y.; Baek J.; Cuadra,A.
2013-11-10
A TRACE model has been developed for using theTRACE/PARCS computational package [1, 2] to simulate anticipated transients without scram (ATWS) events in a boiling water reactor (BWR). The model represents a BWR/5 housed in a Mark II containment. The reactor and the balance of plant systems are modeled in sufficient detail to enable the evaluation of plant responses and theeffectiveness of automatic and operator actions tomitigate this beyond design basis accident.The TRACE model implements features thatfacilitate the simulation of ATWS events initiated by turbine trip and closure of the main steam isolation valves (MSIV). It also incorporates control logic tomore » initiate actions to mitigate the ATWS events, such as water levelcontrol, emergency depressurization, and injection of boron via the standby liquid control system (SLCS). Two different approaches have been used to model boron mixing in the lower plenum of the reactor vessel: modulate coolant flow in the lower plenum by a flow valve, and use control logic to modular.« less
NASA Astrophysics Data System (ADS)
Vinod, M. P.; Vijayamohanan, K.; Joshi, S. N.
Effect of sodium silicate and phosphoric acid additives on the kinetics of oxygen evolution on PbO 2 electrodes in sulfuric acid has been studied in gelled and flooded electrolytes with relevance to valve-regulated lead/acid batteries. A comparison of the open-circuit potential versus time transients, with and without these additives, indicates that the additives suppress self-discharge of the electrodes. Tafel polarization studies also suggest that the addition of phosphoric acid attenuates the rate of oxygen evolution reaction. These findings have been supported with cyclic voltammetric data.
Hypovolemia induced systolic anterior motion of the mitral valve in two dogs.
Hammes, K; Novo Matos, J; Baron Toaldo, M; Glaus, T
2016-12-01
Systolic anterior (septal) motion of the mitral valve (SAM) is a common secondary phenomenon in hypertrophic cardiomyopathy (HCM) in people and cats. In humans, it is increasingly recognized that SAM may be found in other cardiac and non-cardiac disease states. In small animal cardiology, SAM unassociated with HCM has been described in dogs with mitral valve dysplasia and right ventricular pressure overload. In this report, we describe two cases of dogs where transient SAM was caused by hypovolemia. When SAM was present both dogs showed pseudohypertrophy and tachycardia. Important factors in the genesis of SAM in this scenario are probably hypovolemia induced changes in left ventricular geometry affecting the orientation of the mitral valve apparatus combined with elevated catecholamine levels. SAM associated with increased wall thickness is not pathognomonic of HCM; this observation is of particular clinical importance when extrapolated to species where HCM is highly prevalent, e.g., cats. An echocardiographic diagnosis should always be evaluated together with full clinical assessment of history and physical examination. Copyright © 2016. Published by Elsevier B.V.
Water Hammer Simulations of Monomethylhydrazine Propellant
NASA Technical Reports Server (NTRS)
Burkhardt, Zachary; Ramachandran, N.; Majumdar, A.
2017-01-01
Fluid Transient analysis is important for the design of spacecraft propulsion system to ensure structural stability of the system in the event of sudden closing or opening of the valve. Generalized Fluid System Simulation Program (GFSSP), a general purpose flow network code developed at NASA/MSFC is capable of simulating pressure surge due to sudden opening or closing of valve when thermodynamic properties of real fluid are available for the entire range of simulation. Specifically GFSSP needs an accurate representation of pressure density relationship in order to predict pressure surge during a fluid transient. Unfortunately, the available thermodynamic property programs such as REFPROP, GASP or GASPAK do not provide the thermodynamic properties of Monomethylhydrazine(MMH). This work illustrates the process used for building a customized table of properties of state variables from available properties and speed of sound that is required by GFSSP for simulation. Good agreement was found between the simulations and measured data. This method can be adopted for modeling flow networks and systems with other fluids whose properties are not known in detail in order to obtain general technical insight.
Experimental Verification of Steel Pipe Collapse under Vacuum Pressure Conditions
NASA Astrophysics Data System (ADS)
Autrique, R.; Rodal, E.
2016-11-01
Steel pipes are used widely in hydroelectric systems and in pumping systems. Both systems are subject to hydraulic transient effects caused by changes in boundary conditions, such as sudden valve closures, pump failures, or accidents. Water column separation, and its associated vaporization pressure inside the pipe, can cause the collapse of thin walled steel pipes subject to atmospheric pressure, as happened during the well known Oigawa Power Plant accident in Japan, in 1950. The conditions under which thin walled pipes subject to external pressure can collapse have been studied mathematically since the second half of the XIX century, with classical authors Southwell and Von Mises obtaining definitive equations for long and short pipes in the second decade of the XX century, in which the fundamental variables are the diameter to thickness ratio D/t and the length to diameter ratio L/D. In this paper, the predicted critical D/t ratio for steel pipe collapse is verified experimentally, in a physical model able to reproduce hydraulic transients, generating vacuum pressures through rapid upstream valve closures.
Two-phase flow in the cooling circuit of a cryogenic rocket engine
NASA Astrophysics Data System (ADS)
Preclik, D.
1992-07-01
Transient two-phase flow was investigated for the hydrogen cooling circuit of the HM7 rocket engine. The nuclear reactor code ATHLET/THESEUS was adapted to cryogenics and applied to both principal and prototype experiments for validation and simulation purposes. The cooling circuit two-phase flow simulation focused on the hydrogen prechilling and pump transient phase prior to ignition. Both a single- and a multichannel model were designed and employed for a valve leakage flow, a nominal prechilling flow, and a prechilling with a subsequent pump-transient flow. The latter case was performed in order to evaluate the difference between a nominal and a delayed turbo-pump start-up. It was found that an extension of the nominal prechilling sequence in the order of 1 second is sufficient to finally provide for liquid injection conditions of hydrogen which, as commonly known, is undesirable for smooth ignition and engine starting transients.
NASA Technical Reports Server (NTRS)
Neiner, G. H.; Dustin, M. O.; Cole, G. L.
1979-01-01
A stability-bleed system was installed in a YF-12 flight inlet that was subjected to internal and external airflow disturbances in the NASA Lewis 10 by 10 foot supersonic wind tunnel. The purpose of the system is to allow higher inlet performance while maintaining a substantial tolerance (without unstart) to internal and external disturbances. At Mach numbers of 2.47 and 2.76, the inlet tolerance to decreases in diffuser-exit corrected airflow was increased by approximately 10 percent of the operating-point airflow. The stability-bleed system complemented the terminal-shock-control system of the inlet and did not show interaction problems. For disturbances which caused a combined decrease in Mach number and increase in angle of attack, the system with valves operative kept the inlet started 4 to 28 times longer than with the valves inoperative. Hence, the stability system provides additional time for the inlet control system to react and prevent unstart. This was observed for initial Mach numbers of 2.55 and 2.68. For slow increase in angle of attack at Mach 2.47 and 2.76, the system kept the inlet started beyond the steady-state unstart angle. However, the maximum transient angles of attack without unstart could not be determined because wind-tunnel mechanical-stop limits for angle of attack were reached.
Thermodynamic and fluid mechanic analysis of rapid pressurization in a dead-end tube
NASA Technical Reports Server (NTRS)
Leslie, Ian H.
1989-01-01
Three models have been applied to very rapid compression of oxygen in a dead-ended tube. Pressures as high as 41 MPa (6000 psi) leading to peak temperatures of 1400 K are predicted. These temperatures are well in excess of the autoignition temperature (750 K) of teflon, a frequently used material for lining hoses employed in oxygen service. These findings are in accord with experiments that have resulted in ignition and combustion of the teflon, leading to the combustion of the stainless steel braiding and catastrophic failure. The system analyzed was representative of a capped off-high-pressure oxygen line, which could be part of a larger system. Pressurization of the larger system would lead to compression in the dead-end line, and possible ignition of the teflon liner. The model consists of a large plenum containing oxygen at the desired pressure (500 to 6000 psi). The plenum is connected via a fast acting valve to a stainless steel tube 2 cm inside diameter. Opening times are on the order of 15 ms. Downstream of the valve is an orifice sized to increase filling times to around 100 ms. The total length from the valve to the dead-end is 150 cm. The distance from the valve to the orifice is 95 cm. The models describe the fluid mechanics and thermodynamics of the flow, and do not include any combustion phenomena. A purely thermodynamic model assumes filling to be complete upstream of the orifice before any gas passes through the orifice. This simplification is reasonable based on experiment and computer modeling. Results show that peak temperatures as high as 4800 K can result from recompression of the gas after expanding through the orifice. An approximate transient model without an orifice was developed assuming an isentropic compression process. An analytical solution was obtained. Results indicated that fill times can be considerably shorter than valve opening times. The third model was a finite difference, 1-D transient compressible flow model. Results from the code show the recompression effect but predict much lower peak temperatures than the thermodynamic model.
Thermodynamic and fluid mechanic analysis of rapid pressurization in a dead-end tube
NASA Astrophysics Data System (ADS)
Leslie, Ian H.
1989-12-01
Three models have been applied to very rapid compression of oxygen in a dead-ended tube. Pressures as high as 41 MPa (6000 psi) leading to peak temperatures of 1400 K are predicted. These temperatures are well in excess of the autoignition temperature (750 K) of teflon, a frequently used material for lining hoses employed in oxygen service. These findings are in accord with experiments that have resulted in ignition and combustion of the teflon, leading to the combustion of the stainless steel braiding and catastrophic failure. The system analyzed was representative of a capped off-high-pressure oxygen line, which could be part of a larger system. Pressurization of the larger system would lead to compression in the dead-end line, and possible ignition of the teflon liner. The model consists of a large plenum containing oxygen at the desired pressure (500 to 6000 psi). The plenum is connected via a fast acting valve to a stainless steel tube 2 cm inside diameter. Opening times are on the order of 15 ms. Downstream of the valve is an orifice sized to increase filling times to around 100 ms. The total length from the valve to the dead-end is 150 cm. The distance from the valve to the orifice is 95 cm. The models describe the fluid mechanics and thermodynamics of the flow, and do not include any combustion phenomena. A purely thermodynamic model assumes filling to be complete upstream of the orifice before any gas passes through the orifice. This simplification is reasonable based on experiment and computer modeling. Results show that peak temperatures as high as 4800 K can result from recompression of the gas after expanding through the orifice. An approximate transient model without an orifice was developed assuming an isentropic compression process. An analytical solution was obtained.
Posttest data analysis of FIST experimental TRAC-BD1/MOD1 power transient experiment
DOE Office of Scientific and Technical Information (OSTI.GOV)
Wheatley, P.D.; Wagner, K.C.
The FIST power transient test 6PMC2 was analyzed to further the understanding of the FIST facility and provide an assessment of TRAC-BD1/MOD1. FIST power transient 6PMC2 investigated the thermal-hydraulic response following inadvertent closure of the main steam isolation valve and the subsequent failure of the reactor to scram. Failure of the high pressure core spray system was also assumed, resulting in only the reactor core isolation cooling flow for inventory makeup during the transient. The experiment was a sensitivity study with relatively high core power and low makeup rates. This study provides one of the first opportunities to assess TRAC-BD1/MOD1more » under power transient and natural circulation conditions with data from a facility with prototypical BWR geometry. The power transient test was analyzed with emphasis on the following phenomena: (a) the system pressure response, (b) the natural circulation flows and rates, and (c) the heater rod cladding temperature response. Based on the results of this study, TRAC-BD1/MOD1 can be expected to calculate the thermal-hydraulic behavior of a BWR during a power transient.« less
Bartonella vinsonii subsp. berkhoffii endocarditis in a dog from Saskatchewan
Cockwill, Ken R.; Taylor, Susan M.; Philibert, Helene M.; Breitschwerdt, Edward B.; Maggi, Ricardo G.
2007-01-01
A dog referred for lameness was diagnosed with culture-negative endocarditis. Antibodies to Bartonella spp. were detected. Antibiotic treatment resulted in transient clinical improvement, but the dog developed cardiac failure and was euthanized. Bartonella vinsonii subsp. berkhoffii genotype IV was identified within the aortic heart valve lesions by PCR amplification and DNA sequencing. PMID:17824328
NASA Technical Reports Server (NTRS)
Crane, Harold L.
1961-01-01
With an electric analog computer, an investigation has been made of the effects of control frictions and preloads on the transient longitudinal response of a fighter airplane during abrupt small attitude corrections. The simulation included the airplane dynamics, powered control system, feel system, and a simple linearized pseudopilot. Control frictions at the stick pivot and at the servo valve as well as preloads of the stick and valve were considered individually and in combinations. It is believed that the results which are presented in the form of time histories and vector diagrams present a more detailed illustration of the effects of stray forces and compensating forces in the longitudinal control system than has previously been available. Consistent with the results of previous studies, the present results show that any of these four friction and preload forces caused some deterioration of the response. However, even a small amount of valve friction caused an oscillatory pitching response during which the phasing of the valve friction was such that it caused energy to be fed into the pitching oscillation of the air-plane. Of the other friction and preload forces which were considered, it was found that stick preload was close to 180 deg. out of phase with valve friction and thus could compensate in large measure for valve friction as long as the cycling of the stick encompassed the trim point. Either stick friction or valve preload provided a smaller stabilizing effect primarily through a reduction in the amplitude of the resultant force vector acting on the control system. Some data were obtained on the effects of friction when the damping or inertia of the control system or the pilot lag was varied.
NASA Technical Reports Server (NTRS)
Crane, Harold L
1957-01-01
With an electric analog computer, an investigation has been made of the effects of control frictions and preloads on the transient longitudinal response of a fighter airplane during abrupt small attitude corrections. The simulation included the airplane dynamics, powered control system, feel system, and a simple linearized pseudopilot. Control frictions at the stick pivot and at the servo valve as well as preloads of the stick and valve were considered individually and in combinations. It is believed that the results which are presented in the form of time histories and vector diagrams present a more detailed illustration of the effects of stray forces and compensating forces in the longitudinal control system than has previously been available. Consistent with the results of previous studies, the present results show that any of thesefour friction and preload forces caused some deterioration of the response. However, even a small amount of valve friction caused an oscillatory pitching response during which the phasing of the valve friction was such that it caused energy to be fed into the pitching oscillation of the airplane. Of the other friction and preload forces which were considered, it was found that stick preload was close to 180 degrees out of phase with valve friction and thus could compensate in large measure for valve friction as long as the cycling of the stick encompassed the trim point. Either stick friction or valve preload provided a smaller stabilizing effect primarily through a reduction in the amplitude of the resultant force vector acting on the control system. Some data were obtained on the effects of friction when the damping or inertia of the control system or the pilot lag was varied.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Dallman, R J; Gottula, R C; Holcomb, E E
1987-05-01
An analysis of five anticipated transients without scram (ATWS) was conducted at the Idaho National Engineering Laboratory (INEL). The five detailed deterministic simulations of postulated ATWS sequences were initiated from a main steamline isolation valve (MSIV) closure. The subject of the analysis was the Browns Ferry Nuclear Plant Unit 1, a boiling water reactor (BWR) of the BWR/4 product line with a Mark I containment. The simulations yielded insights to the possible consequences resulting from a MSIV closure ATWS. An evaluation of the effects of plant safety systems and operator actions on accident progression and mitigation is presented.
NASA Technical Reports Server (NTRS)
Burkhardt, Z.; Ramachandran, N.; Majumdar, A.
2017-01-01
Fluid Transient analysis is important for the design of spacecraft propulsion system to ensure structural stability of the system in the event of sudden closing or opening of the valve. Generalized Fluid System Simulation Program (GFSSP), a general purpose flow network code developed at NASA/MSFC is capable of simulating pressure surge due to sudden opening or closing of valve when thermodynamic properties of real fluid are available for the entire range of simulation. Specifically GFSSP needs an accurate representation of pressure-density relationship in order to predict pressure surge during a fluid transient. Unfortunately, the available thermodynamic property programs such as REFPROP, GASP or GASPAK does not provide the thermodynamic properties of Monomethylhydrazine (MMH). This paper will illustrate the process used for building a customized table of properties of state variables from available properties and speed of sound that is required by GFSSP for simulation. Good agreement was found between the simulations and measured data. This method can be adopted for modeling flow networks and systems with other fluids whose properties are not known in detail in order to obtain general technical insight. Rigorous code validation of this approach will be done and reported at a future date.
Vacuum system transient simulator and its application to TFTR
DOE Office of Scientific and Technical Information (OSTI.GOV)
Sredniawski, J.
The vacuum system transient simulator (VSTS) models transient gas transport throughout complex networks of ducts, valves, traps, vacuum pumps, and other related vacuum system components. VSTS is capable of treating gas models of up to 10 species, for all flow regimes from pure molecular to continuum. Viscous interactions between species are considered as well as non-uniform temperature of a system. Although this program was specifically developed for use on the Tokamak Fusion Test Reactor (TFTR) project at Princeton, it is a generalized tool capable of handling a broad range of vacuum system problems. During the TFTR engineering design phase, VSTSmore » has been used in many applications. Two applications selected for presentation are: torus vacuum pumping system performance between 400 Ci tritium pulses and tritium backstreaming to neutral beams during pulses.« less
Prediction of pressure and flow transients in a gaseous bipropellant reaction control rocket engine
NASA Technical Reports Server (NTRS)
Markowsky, J. J.; Mcmanus, H. N., Jr.
1974-01-01
An analytic model is developed to predict pressure and flow transients in a gaseous hydrogen-oxygen reaction control rocket engine feed system. The one-dimensional equations of momentum and continuity are reduced by the method of characteristics from partial derivatives to a set of total derivatives which describe the state properties along the feedline. System components, e.g., valves, manifolds, and injectors are represented by pseudo steady-state relations at discrete junctions in the system. Solutions were effected by a FORTRAN IV program on an IBM 360/65. The results indicate the relative effect of manifold volume, combustion lag time, feedline pressure fluctuations, propellant temperature, and feedline length on the chamber pressure transient. The analytical combustion model is verified by good correlation between predicted and observed chamber pressure transients. The developed model enables a rocket designer to vary the design parameters analytically to obtain stable combustion for a particular mode of operation which is prescribed by mission objectives.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Wheatley, P.D.; Wagner, K.C.
The FIST power transient test 6PMC2 was analyzed to further the understanding of the FIST facility and provide an assessment of TRAC-BD1/MOD1. FIST power transient 6PMC2 investigated the thermal-hydraulic response following inadvertent closure of the main steam isolation valve and the subsequent failure of the reactor to scram. Failure of the high pressure core spray system was also assumed, resulting on only the reactor core isolation cooling flow for inventory makeup during the transient. The experiment was a sensitivity study with relatively high core power and low makeup rates. This study provides one of the first opportunities to assess TRAC-BD1/MOD1more » under power transient and natural circulation conditions with data from a facility with prototypical BWR geometry. The power transient test was analyzed with emphasis on the following phenomena; (a) the system pressure response, (b) the natural circulation flows and rates, and (c) the heater rod cladding temperature response. Based on the results of this study, TRAC-BD1/MOD1 can be expected to calculate the thermal-hydraulic behavior of a BWR during a power transient.« less
Short-course thrombolysis as the first line of therapy for cardiac valve thrombosis.
Manteiga, R; Carlos Souto, J; Altès, A; Mateo, J; Arís, A; Dominguez, J M; Borrás, X; Carreras, F; Fontcuberta, J
1998-04-01
To retrospectively evaluate the clinical and echocardiographic criteria of thrombolytic therapy for mechanical heart valve thrombosis. Nineteen consecutive patients with 22 instances of prosthetic heart valve thrombosis (14 mitral, 2 aortic, 3 tricuspid, and 3 pulmonary) were treated with short-course thrombolytic therapy as first option of treatment in absence of contraindications. The thrombolytic therapy protocol consisted of streptokinase (1,500,000 IU in 90 minutes) (n = 18) in one (n = 7) or two (n = 11) cycles or recombinant tissue-type plasminogen activator (100 mg in 90 minutes) (n = 4). Overall success was seen in 82%, immediate complete success in 59%, and partial success in 23%. Six patients without total response to thrombolytic therapy underwent surgery, and pannus was observed in 83%. Six patients showed complications: allergy, stroke, transient ischemic attack, coronary embolism, minor bleeding, and one death. At diagnosis, 10 patients evidenced atrial thrombus by transesophageal echocardiography, 3 of whom experienced peripheral embolism during thrombolysis. Four episodes of rethrombosis were observed (16%). The survivorship was 84% with a mean follow-up of 42.6 months. A short-course of thrombolytic therapy may be considered first-line therapy for prosthetic heart valve thrombosis. The risk of peripheral embolism may be evaluated for the presence of atrial thrombus by transesophageal echocardiography at diagnosis.
Vacuum system design and tritium inventory for the TFTR charge exchange diagnostic
DOE Office of Scientific and Technical Information (OSTI.GOV)
Medley, S.S.
The charge exchange diagnostic for the TFTR is comprised of two analyzer systems which contain a total of twenty independent mass/energy analyzers and one diagnostic neutral beam tentatively rated at 80 keV, 15 A. The associated vacuum systems were analyzed using the Vacuum System Transient Simulator (VSTS) computer program which models the transient transport of multi-gas species through complex networks of ducts, valves, traps, vacuum pumps, and other related vacuum system components. In addition to providing improved design performance at reduced cost, the analysis yields estimates for the exchange of tritium from the torus to the diagnostic components and ofmore » the diagnostic working gases to the torus.« less
Transient Region Coverage in the Propulsion IVHM Technology Experiment
NASA Technical Reports Server (NTRS)
Balaban, Edward; Sweet, Adam; Bajwa, Anupa; Maul, William; Fulton, Chris; Chicatelli, amy
2004-01-01
Over the last several years researchers at NASA Glenn and Ames Research Centers have developed a real-time fault detection and isolation system for propulsion subsystems of future space vehicles. The Propulsion IVHM Technology Experiment (PITEX), as it is called follows the model-based diagnostic methodology and employs Livingstone, developed at NASA Ames, as its reasoning engine. The system has been tested on,flight-like hardware through a series of nominal and fault scenarios. These scenarios have been developed using a highly detailed simulation of the X-34 flight demonstrator main propulsion system and include realistic failures involving valves, regulators, microswitches, and sensors. This paper focuses on one of the recent research and development efforts under PITEX - to provide more complete transient region coverage. It describes the development of the transient monitors, the corresponding modeling methodology, and the interface software responsible for coordinating the flow of information between the quantitative monitors and the qualitative, discrete representation Livingstone.
NASA Technical Reports Server (NTRS)
Cady, E. C.
1977-01-01
A design analysis, is developed based on experimental data, to predict the effects of transient flow and pressure surges (caused either by valve or pump operation, or by boiling of liquids in warm lines) on the retention performance of screen acquisition systems. A survey of screen liquid acquisition system applications was performed to determine appropriate system environment and classification. A screen model was developed which assumed that the screen device was a uniformly distributed composite orthotropic structure, and which accounted for liquid inflow/outflow, gas ingestion quality, screen stress, and liquid spill. A series of 177 tests using 13 specimens (5 screen meshes, 4 screen device construction/backup methods, and 2 orientations) with three test fluids (isopropyl alcohol, Freon 114, and LH2) provided data which verified important features of the screen model and resulted in a design tool which could accurately predict the transient startup performance acquisition devices.
NASA Astrophysics Data System (ADS)
Ueki, Hironobu; Ishida, Masahiro; Sakaguchi, Daisaku
2005-06-01
In order to investigate the effect of transient needle opening on early stage of spray behavior, simultaneous measurements of velocity and size of droplet were conducted by a newly developed laser 2-focus velocimeter (L2F). The micro-scale probe of the L2F was consisted of two foci with a distance of 36 µm. The tested nozzle had a single hole with a diameter of 0.2 mm. The measurements of injection pressure, needle lift, and crank angle were synchronized with the spray measurement by the L2F at the position 10 mm downstream from the nozzle exit. It has been clearly shown that the velocity and size of droplet increase with needle valve opening and that the probability density distribution of droplet size can be fitted to the Nukiyama-Tanasawa distribution under the transient needle opening condition.
[Long-term outcomes of Ahmed glaucoma valve implantation for treating refractory glaucoma].
Xu, Yumei; Hong, Tao; Li, Wanming
2015-02-10
To explore the efficacies and complications of Ahmed glaucoma valve implantation for treating refractory glaucoma. A retrospective study of case series was conducted for 24 patients (26 eyes) with refractory glaucoma from February 2001 to July 2008 at our hospital. Ahmed glaucoma valve implantation was performed. Pre- and post-operative best spectacle-corrected visual acuity (BSCVA), intraocular pressure (IOP), number of medications and complications were recorded and analyzed. The follow-up period was 58-159 months. The post-operative values of IOP were 13.02+/-6.79, 11.43+/-5.24 and 18.56+/-6.43 mmHg at 1 day, 1 month and the last follow-up respectively. There were significant difference when compared with pre-operative IOP (37.59+/-10.76 mmHg, P < 0.01). And 65.38% of eyes maintained or gained ≥ 1 line of BSCVA. But there was no significant difference with pre-operative BSCVA (P = 0.110). Twenty eyes required anti-glaucoma drugs after glaucoma valve implantation and the average number of medication was 1.72+/-0.98. There was significant difference with the pre-operative medication number 2.7 ± 0.7 (P = 0.001). The surgical success rate was 73.1%. And the causes of failure were endophthalmitis, corneal endothelial decompensation, persistent conjunctival wound non-healing, glaucoma valve exposure and loss of light perception.Early postoperative complications were ocular hypotony, shallow anterior chamber, hyphema, transient high IOP and tube occlusion. And long-term complications included encapsulated cyst formation, tube exposure, corneal endothelial decompensation and endophthalmitis. Ahmed glaucoma valve implantation is efficacious for refractory glaucoma.However, clinicians should pay attention to the prevention and treatment of complications.
Numerical Hydraulic Study on Seawater Cooling System of Combined Cycle Power Plant
NASA Astrophysics Data System (ADS)
Kim, J. Y.; Park, S. M.; Kim, J. H.; Kim, S. W.
2010-06-01
As the rated flow and pressure increase in pumping facilities, a proper design against surges and severe cavitations in the pipeline system is required. Pressure surge due to start-up, shut-down process and operation failure causes the water hammer in upstream of the closing valve and the cavitational hammer in downstream of the valve. Typical cause of water hammer is the urgent closure of valves by breakdown of power supply and unexpected failure of pumps. The abrupt changes in the flow rate of the liquid results in high pressure surges in upstream of the valves, thus kinetic energy is transformed into potential energy which leads to the sudden increase of the pressure that is called as water hammer. Also, by the inertia, the liquid continues to flow downstream of the valve with initial speed. Accordingly, the pressure decreases and an expanding vapor bubble known as column separation are formed near the valve. In this research, the hydraulic study on the closed cooling water heat exchanger line, which is the one part of the power plant, is introduced. The whole power plant consists of 1,200 MW combined power plant and 220,000 m3/day desalination facility. Cooling water for the plant is supplied by sea water circulating system with a capacity of 29 m3/s. The primary focus is to verify the steady state hydraulic capacity of the system. The secondary is to quantify transient issues and solutions in the system. The circuit was modeled using a commercial software. The stable piping network was designed through the hydraulic studies using the simulation for the various scenarios.
Transient particle emission measurement with optical techniques
NASA Astrophysics Data System (ADS)
Bermúdez, Vicente; Luján, José M.; Serrano, José R.; Pla, Benjamín
2008-06-01
Particulate matter is responsible for some respiratory and cardiovascular diseases. In addition, it is one of the most important pollutants of high-speed direct injection (HSDI) passenger car engines. Current legislation requires particulate dilution tunnels for particulate matter measuring. However for development work, dilution tunnels are expensive and sometimes not useful since they are not able to quantify real-time particulate emissions during transient operation. In this study, the use of a continuous measurement opacimeter and a fast response HFID is proven to be a good alternative to obtain instantaneous particle mass emissions during transient operation (due to particulate matter consisting mainly of soot and SOF). Some methods and correlations available from literature, but developed for steady conditions, are evaluated during transient operation by comparing with mini-tunnel measurements during the entire MVEG-A transient cycle. A new correlation was also derived from this evaluation. Results for soot and SOF (obtained from the new correlation proposed) are compared with soot and SOF captured with particulate filters, which have been separated by means of an SOF extraction method. Finally, as an example of ECU design strategies using these sort of correlations, the EGR valve opening is optimized during transient operation. The optimization is performed while simultaneously taking into account instantaneous fuel consumption, particulate emissions (calculated with the proposed correlation) and other regulated engine pollutants.
Ebacher, G; Besner, M C; Clément, B; Prévost, M
2012-09-01
Intrusion events caused by transient low pressures may result in the contamination of a water distribution system (DS). This work aims at estimating the range of potential intrusion volumes that could result from a real downsurge event caused by a momentary pump shutdown. A model calibrated with transient low pressure recordings was used to simulate total intrusion volumes through leakage orifices and submerged air vacuum valves (AVVs). Four critical factors influencing intrusion volumes were varied: the external head of (untreated) water on leakage orifices, the external head of (untreated) water on submerged air vacuum valves, the leakage rate, and the diameter of AVVs' outlet orifice (represented by a multiplicative factor). Leakage orifices' head and AVVs' orifice head levels were assessed through fieldwork. Two sets of runs were generated as part of two statistically designed experiments. A first set of 81 runs was based on a complete factorial design in which each factor was varied over 3 levels. A second set of 40 runs was based on a latin hypercube design, better suited for experimental runs on a computer model. The simulations were conducted using commercially available transient analysis software. Responses, measured by total intrusion volumes, ranged from 10 to 366 L. A second degree polynomial was used to analyze the total intrusion volumes. Sensitivity analyses of both designs revealed that the relationship between the total intrusion volume and the four contributing factors is not monotonic, with the AVVs' orifice head being the most influential factor. When intrusion through both pathways occurs concurrently, interactions between the intrusion flows through leakage orifices and submerged AVVs influence intrusion volumes. When only intrusion through leakage orifices is considered, the total intrusion volume is more largely influenced by the leakage rate than by the leakage orifices' head. The latter mainly impacts the extent of the area affected by intrusion. Copyright © 2012 Elsevier Ltd. All rights reserved.
Investigation of erosion behavior in different pipe-fitting using Eulerian-Lagrangian approach
NASA Astrophysics Data System (ADS)
Kulkarni, Harshwardhan; Khadamkar, Hrushikesh; Mathpati, Channamallikarjun
2017-11-01
Erosion is a wear mechanism of piping system in which wall thinning occurs because of turbulent flow along with along with impact of solid particle on the pipe wall, because of this pipe ruptures causes costly repair of plant and personal injuries. In this study two way coupled Eulerian-Lagrangian approach is used to solve the liquid solid (water-ferrous suspension) flow in the different pipe fitting namely elbow, t-junction, reducer, orifice and 50% open gate valve. Simulations carried out using incomressible transient solver in OpenFOAM for different Reynolds's number (10k, 25k, 50k) and using WenYu drag model to find out possible higher erosion region in pipe fitting. Used transient solver is a hybrid in nature which is combination of Lagrangian library and pimpleFoam. Result obtained from simulation shows that exit region of elbow specially downstream of straight, extradose of the bend section more affected by erosion. Centrifugal force on solid particle at bend affect the erosion behavior. In case of t-junction erosion occurs below the locus of the projection of branch pipe on the wall. For the case of reducer, orifice and a gate valve reduction area as well as downstream is getting more affected by erosion because of increase in velocities.
A Numerical Investigation of the Startup Transient in a Wave Rotor
NASA Technical Reports Server (NTRS)
Paxson, Daniel E.
1996-01-01
The startup process is investigated for a hypothetical four-port wave rotor, envisioned as a topping cycle for a small gas turbine engine. The investigation is conducted numerically using a multi-passage, one-dimensional CFD-based wave rotor simulation in combination with lumped volume models for the combustor, exhaust valve plenum, and rotor center cavity components. The simulation is described and several startup transients are presented which illustrate potential difficulties for the specific cycle design investigated. In particular it is observed that, prior to combustor light-off, or just after, the flow through the combustor loop is reversed from the design direction. The phenomenon is demonstrated and several possible modifications techniques are discussed which avoid or overcome the problem.
Pretest analysis document for Test S-FS-7
DOE Office of Scientific and Technical Information (OSTI.GOV)
Hall, D.G.
This report documents the pretest calculations completed for Semiscale Test S-FS-7. This test will simulate a transient initiated by a 14.3% break in a steam generator bottom feedwater line downstream of the check valve. The initial conditions represent normal operating conditions for a C-E System 80 nuclear power plant. Predictions of transients resulting from feedwater line breaks in these plants have indicated that significant primary system overpressurization may occur. The results of a RELAP5/MOD2/CY21 code calculation indicate that the test objectives for Test S-FS-7 can be achieved. The primary system overpressurization will occur but pose no threat to personnel ormore » to plant integrity. 3 refs., 15 figs., 5 tabs.« less
Pretest analysis document for Test S-FS-11
DOE Office of Scientific and Technical Information (OSTI.GOV)
Hall, D.G.; Shaw, R.A.
This report documents the pretest calculations completed for Semiscale Test S-FS-11. This test will simulate a transient initiated by a 50% break in a steam generator bottom feedwater line downstream of the check valve. The initial conditions represents normal operating conditions for a C-E System 80 nuclear plant. Prediction of transients resulting from feedwater line breaks in these plants have indicated that significant primary system overpressurization may occur. The results of a RELAP5/MOD2/CY21 code calculation indicate that the test objectives for Test S-FS-11 can be achieved. The primary system overpressurization will occur but pose no threat to personnel or plantmore » integrity. 3 refs., 15 figs., 5 tabs.« less
Numerical model of spray combustion in a single cylinder diesel engine
NASA Astrophysics Data System (ADS)
Acampora, Luigi; Sequino, Luigi; Nigro, Giancarlo; Continillo, Gaetano; Vaglieco, Bianca Maria
2017-11-01
A numerical model is developed for predicting the pressure cycle from Intake Valve Closing (IVC) to the Exhaust Valve Opening (EVO) events. The model is based on a modified one-dimensional (1D) Musculus and Kattke spray model, coupled with a zero-dimensional (0D) non-adiabatic transient Fed-Batch reactor model. The 1D spray model provides an estimate of the fuel evaporation rate during the injection phenomenon, as a function of time. The 0D Fed-Batch reactor model describes combustion. The main goal of adopting a 0D (perfectly stirred) model is to use highly detailed reaction mechanisms for Diesel fuel combustion in air, while keeping the computational cost as low as possible. The proposed model is validated by comparing its predictions with experimental data of pressure obtained from an optical single cylinder Diesel engine.
NASA Astrophysics Data System (ADS)
Nezhad, Mohsen Motahari; Shojaeefard, Mohammad Hassan; Shahraki, Saeid
2016-02-01
In this study, the experiments aimed at analyzing thermally the exhaust valve in an air-cooled internal combustion engine and estimating the thermal contact conductance in fixed and periodic contacts. Due to the nature of internal combustion engines, the duration of contact between the valve and its seat is too short, and much time is needed to reach the quasi-steady state in the periodic contact between the exhaust valve and its seat. Using the methods of linear extrapolation and the inverse solution, the surface contact temperatures and the fixed and periodic thermal contact conductance were calculated. The results of linear extrapolation and inverse methods have similar trends, and based on the error analysis, they are accurate enough to estimate the thermal contact conductance. Moreover, due to the error analysis, a linear extrapolation method using inverse ratio is preferred. The effects of pressure, contact frequency, heat flux, and cooling air speed on thermal contact conductance have been investigated. The results show that by increasing the contact pressure the thermal contact conductance increases substantially. In addition, by increasing the engine speed the thermal contact conductance decreases. On the other hand, by boosting the air speed the thermal contact conductance increases, and by raising the heat flux the thermal contact conductance reduces. The average calculated error equals to 12.9 %.
Dynamic water behaviour due to one trapped air pocket in a laboratory pipeline apparatus
NASA Astrophysics Data System (ADS)
Bergant, A.; Karadžić, U.; Tijsseling, A.
2016-11-01
Trapped air pockets may cause severe operational problems in hydropower and water supply systems. A locally isolated air pocket creates distinct amplitude, shape and timing of pressure pulses. This paper investigates dynamic behaviour of a single trapped air pocket. The air pocket is incorporated as a boundary condition into the discrete gas cavity model (DGCM). DGCM allows small gas cavities to form at computational sections in the method of characteristics (MOC). The growth of the pocket and gas cavities is described by the water hammer compatibility equation(s), the continuity equation for the cavity volume, and the equation of state of an ideal gas. Isentropic behaviour is assumed for the trapped gas pocket and an isothermal bath for small gas cavities. Experimental investigations have been performed in a laboratory pipeline apparatus. The apparatus consists of an upstream end high-pressure tank, a horizontal steel pipeline (total length 55.37 m, inner diameter 18 mm), four valve units positioned along the pipeline including the end points, and a downstream end tank. A trapped air pocket is captured between two ball valves at the downstream end of the pipeline. The transient event is initiated by rapid opening of the upstream end valve; the downstream end valve stays closed during the event. Predicted and measured results for a few typical cases are compared and discussed.
Dulamea, Adriana O; Matei, Costel; Mindruta, Ioana; Ionescu, Virgil
2015-10-12
Based on a case report, the authors reviewed the data about involuntary emotional expression disorder (IEED). IEED includes the syndromes of pathological laughing and crying (PLC) and emotional lability (EL). PLC is a rare disorder of emotional expression characterized by relatively uncontrollable episodes of laughter and crying or both that do not have an apparent motivating stimulus. Authors report the case of a 59-year-old man who presented with recurrent episodes of PLC of approximately 2 min duration, consisting of accelerated breathing, emission of guttural, snoring sounds, frowning of the eyebrows, followed by laughter accompanied by motor restlessness of all four limbs. PLC episodes preceded left carotid transient ischemic attacks (TIA's) manifested by reversible aphasia and right hemiparesis. Electroencephalography performed during PLC episodes revealed no spike-wave activity. Brain magnetic resonance imaging showed lacunar infarcts in the left lenticulo-capsulo-thalamic area and multiple round lesions in the cortical-subcortical and in the deep white matter of frontal-parietal-occipital lobes bilaterally, with T2 hyperintensity, T1 isointensity and no diffusion changes. The episodes were interpreted as transient ischemic attacks although gelastic seizures could not be excluded. The etiological investigations revealed unstable plaques on the left carotid artery bulb and the aortic arch and a degenerative mitral valve stenosis. The patient was treated first with antiplatelet therapy and antiepileptic drugs but PLC stopped only after anticoagulation was started. During follow-up the patient continued to have left carotid and vertebrobasilar TIA's being on oral anticoagulation. The patient became asymptomatic only after mitral valve replacement was performed. This case illustrates the difficulty distinguishing between gelastic epilepsy and TIA's in cases of PLC episodes and discuss the neuroanatomic bases and pathophysiology of this rare condition.
Simulation of IST Turbomachinery Power-Neutral Tests with the ANL Plant Dynamics Code
DOE Office of Scientific and Technical Information (OSTI.GOV)
Moisseytsev, A.; Sienicki, J. J.
The validation of the Plant Dynamics Code (PDC) developed at Argonne National Laboratory (ANL) for the steady-state and transient analysis of supercritical carbon dioxide (sCO2) systems has been continued with new test data from the Naval Nuclear Laboratory (operated by Bechtel Marine Propulsion Corporation) Integrated System Test (IST). Although data from three runs were provided to ANL, only two of the data sets were analyzed and described in this report. The common feature of these tests is the power-neutral operation of the turbine-compressor shaft, where no external power through the alternator was provided during the tests. Instead, the shaft speedmore » was allowed to change dictated by the power balance between the turbine, the compressor, and the power losses in the shaft. The new test data turned out to be important for code validation for several reasons. First, the power-neutral operation of the shaft allows validation of the shaft dynamics equations in asynchronous mode, when the shaft is disconnected from the grid. Second, the shaft speed control with the compressor recirculation (CR) valve not only allows for testing the code control logic itself, but it also serves as a good test for validation of both the compressor surge control and the turbine bypass control actions, since the effect of the CR action on the loop conditions is similar for both of these controls. Third, the varying compressor-inlet temperature change test allows validation of the transient response of the precooler, a shell-and-tube heat exchanger. The first transient simulation of the compressor-inlet temperature variation Test 64661 showed a much slower calculated response of the precooler in the calculations than the test data. Further investigation revealed an error in calculating the heat exchanger tube mass for the PDC dynamic equations that resulted in a slower change in the tube wall temperature than measured. The transient calculations for both tests were done in two steps. The first step was done in the same fashion as the FY15 IST analysis, where the CR valve position and the turbine-compressor shaft speed were specified through the PDC input based on the test values. On the second step, the turbine-compressor shaft dynamics equations were invoked by specifying that the shaft is disconnected from the grid. In addition, the CR valve control was used to control the shaft speed, based on the turbine bypass control logic already implemented in the PDC. For the shaft power balance, the friction (windage) loss is calculated based on the shaft balance at the steady-state conditions and is assumed to be scaled to the third power of shaft speed in the transient. Both the steady-state and transient simulations of both tests showed good agreement with the test data. The only significant difference was the turbine performance, which was not predicted as well as it was in the previous IST simulation, resulting in the prediction of a somewhat different flow split between the two turbines. This flow split difference is believed to be the result of not addressing the recent turbine modifications in the model. In addition, the full simulation of the turbine-compressor speed variation Test 65261-P with shaft speed control showed greater a difference with the test data later in the transient than the other test. Further analysis of the results revealed that this difference is most likely due to scaling the shaft windage losses only with the shaft speed and ignoring its dependency on the fluid density in the shaft cavity. Based on the results of steady state and transient calculations of the Tests 64661 and 65216-P, several areas of future improvements for the PDC simulation of the IST are identified.« less
Role of vortices in growth of microbubbles at mitral mechanical heart valve closure.
Rambod, Edmond; Beizai, Masoud; Sahn, David J; Gharib, Morteza
2007-07-01
This study is aimed at refining our understanding of the role of vortex formation at mitral mechanical heart valve (MHV) closure and its association with the high intensity transient signals (HITS) seen in echocardiographic studies with MHV recipients. Previously reported numerical results described a twofold process leading to formation of gas-filled microbubbles in-vitro: (1) nucleation and (2) growth of micron size bubbles. The growth itself consists of two processes: (a) diffusion and (b) sudden pressure drop due to valve closure. The role of diffusion has already been shown to govern the initial growth of nuclei. Pressure drop at mitral MHV closure may be attributed to other phenomena such as squeezed flow, water hammer and primarily, vortex cavitation. Mathematical analysis of vortex formation at mitral MHV closure revealed that a closing velocity of approximately 12 m/s can induce a strong regurgitant vortex which in return can instigate a local pressure drop of about 0.9 atm. A 2D experimental model of regurgitant flows was used to substantiate the impact of vortices. At simulated flow and pressure conditions, a regurgitant vortex was observed to drastically enlarge micron size hydrogen bubbles at its core.
Birnbaum, J; Ulrich, S M; Schramm, R; Hagl, C; Lehner, A; Fischer, M; Haas, N A; Heineking, B
2017-02-01
In pediatric heart transplantation, the size of the donor organ is an important criterion for organ allocation. Oversized donor hearts are often accepted with good results, but some complications in relation to a high donor-recipient ratio have been described. Our patient was transplanted for progressive heart failure in dilated cardiomyopathy. The donor-to-recipient weight ratio was 3 (donor weight 65 kg, recipient weight 22 kg). The intra-operative echocardiography before chest closure showed excellent cardiac function, no tricuspid valve regurgitation, and a normal central venous pressure. After chest closure, central venous pressure increased substantially and echocardiography revealed a severe tricuspid insufficiency. As other reasons for right ventricular dysfunction, that is, myocardial ischemia, pulmonary hypertension, and rejection, were excluded, we assumed that the insufficiency was caused by an alteration of the right ventricular geometry. After 1 week, the valve insufficiency regressed to a minimal degree. In pediatric heart transplant patients with a high donor-to-recipient weight ratio, the outlined complication may occur. If other reasons for right ventricular heart failure can be ruled out, this entity is most likely caused by an acute and transient alteration of the right ventricular geometry that may disappear over time. © 2016 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.
1D-3D coupling for hydraulic system transient simulations
NASA Astrophysics Data System (ADS)
Wang, Chao; Nilsson, Håkan; Yang, Jiandong; Petit, Olivier
2017-01-01
This work describes a coupling between the 1D method of characteristics (MOC) and the 3D finite volume method of computational fluid dynamics (CFD). The coupling method is applied to compressible flow in hydraulic systems. The MOC code is implemented as a set of boundary conditions in the OpenFOAM open source CFD software. The coupling is realized by two linear equations originating from the characteristics equation and the Riemann constant equation, respectively. The coupling method is validated using three simple water hammer cases and several coupling configurations. The accuracy and robustness are investigated with respect to the mesh size ratio across the interface, and 3D flow features close to the interface. The method is finally applied to the transient flow caused by the closing and opening of a knife valve (gate) in a pipe, where the flow is driven by the difference in free surface elevation between two tanks. A small region surrounding the moving gate is resolved by CFD, using a dynamic mesh library, while the rest of the system is modeled by MOC. Minor losses are included in the 1D region, corresponding to the contraction of the flow from the upstream tank into the pipe, a separate stationary flow regulation valve, and a pipe bend. The results are validated with experimental data. A 1D solution is provided for comparison, using the static gate characteristics obtained from steady-state CFD simulations.
Simulation Analysis of Computer-Controlled pressurization for Mixture Ratio Control
NASA Technical Reports Server (NTRS)
Alexander, Leslie A.; Bishop-Behel, Karen; Benfield, Michael P. J.; Kelley, Anthony; Woodcock, Gordon R.
2005-01-01
A procedural code (C++) simulation was developed to investigate potentials for mixture ratio control of pressure-fed spacecraft rocket propulsion systems by measuring propellant flows, tank liquid quantities, or both, and using feedback from these measurements to adjust propellant tank pressures to set the correct operating mixture ratio for minimum propellant residuals. The pressurization system eliminated mechanical regulators in favor of a computer-controlled, servo- driven throttling valve. We found that a quasi-steady state simulation (pressure and flow transients in the pressurization systems resulting from changes in flow control valve position are ignored) is adequate for this purpose. Monte-Carlo methods are used to obtain simulated statistics on propellant depletion. Mixture ratio control algorithms based on proportional-integral-differential (PID) controller methods were developed. These algorithms actually set target tank pressures; the tank pressures are controlled by another PID controller. Simulation indicates this approach can provide reductions in residual propellants.
Predicted performance of an integrated modular engine system
NASA Technical Reports Server (NTRS)
Binder, Michael; Felder, James L.
1993-01-01
Space vehicle propulsion systems are traditionally comprised of a cluster of discrete engines, each with its own set of turbopumps, valves, and a thrust chamber. The Integrated Modular Engine (IME) concept proposes a vehicle propulsion system comprised of multiple turbopumps, valves, and thrust chambers which are all interconnected. The IME concept has potential advantages in fault-tolerance, weight, and operational efficiency compared with the traditional clustered engine configuration. The purpose of this study is to examine the steady-state performance of an IME system with various components removed to simulate fault conditions. An IME configuration for a hydrogen/oxygen expander cycle propulsion system with four sets of turbopumps and eight thrust chambers has been modeled using the Rocket Engine Transient Simulator (ROCETS) program. The nominal steady-state performance is simulated, as well as turbopump thrust chamber and duct failures. The impact of component failures on system performance is discussed in the context of the system's fault tolerant capabilities.
Rise and fall of NT-proBNP in aortic valve intervention.
Hultkvist, Henrik; Holm, Jonas; Svedjeholm, Rolf; Vánky, Farkas
2018-01-01
To describe the dynamics of N-terminal pro-B-type natriuretic peptide (NT-proBNP) from preoperative evaluation to 6-month follow-up in patients undergoing aortic valve intervention, and to evaluate NT-proBNP with regard to 1-year mortality. At preoperative evaluation, we prospectively included 462 patients accepted for aortic valve intervention. The median time to surgical aortic valve replacement (SAVR; n=336) or transcatheter aortic valve implantation (TAVI; n=126) was 4 months. NT-proBNP was measured at enrolment for preoperative evaluation, on the day of surgery, postoperatively on day 1, day 3 and at the 6-month follow-up. Subgroups of patients undergoing SAVR with aortic regurgitation and aortic stenosis with and without coronary artery bypass were also analysed. NT-proBNP remained stable in all subgroups during the preoperative waiting period, but displayed a substantial transient early postoperative increase with a peak on day 3 except in the TAVI group, which peaked on day 1. At the 6-month follow-up, NT-proBNP had decreased to or below the preoperative level in all groups. In the SAVR group, NT-proBNP preoperatively and on postoperative days 1 and 3 revealed significant discriminatory power with regard to 1-year mortality (area under the curve (AUC)=0.79, P=0.0001; AUC=0.71, P=0.03; and AUC=0.79, P=0.002, respectively). This was not found in the TAVI group, which had higher levels of NT-proBNP both preoperatively and at the 6-month follow-up compared with the SAVR group. The dynamic profile of NT-proBNP differed between patients undergoing TAVI and SAVR. NT-proBNP in the perioperative course was associated with increased risk of 1-year mortality in SAVR but not in TAVI.
Bade, Arun Shivajirao; Shaikh, Shakil Sattar Ahmed; Khemani, Hemant; Singh, Gurkirat; Bansal, Narender Omprakash
2018-01-01
Background Thrombosis is a complication of prosthetic valves on oral anticoagulants which is associated with significant morbidity and mortality. A re-operation carries a substantial risk, with mortality rate from 10% to 15% in selected series, which may be 2- or 3-folds higher in critically ill patients. This study conducted in a tertiary care cardiology unit aimed to evaluate the effectiveness and safety of thrombolytic therapy in stuck mitral bileaflet heart valves. Methods As a prospective observational study, clinical symptoms and fluoroscopy were the mainstay in diagnosis of stuck mitral valve. Gradient across the valve by transthoracic echocardiography was used to monitor the therapy every 6 h. Fall of mean gradient more than 50% was considered as successful thrombolysis. And final results were again checked by fluoroscopy with documentation of improved leaflet movement. Results Totally we studied 34 patients. Patients receiving thrombolytic therapy with streptokinase achieved an overall 91.2% freedom from a repeat operation or major complications, a large subcutaneous hematoma occurred in one ( 2.9%), reoperation required in two due to failure of treatment (5.9%), allergic reaction in one (2.9%), one patient developed transient neurologic dysfunction (2.9%) and one patient died during therapy due to refractory cardiogenic shock(2.9%). All patients including those with delayed presentation (> 14 days) and hemodynamically unstable patients had good results similar to those who presented within 14 days and hemodynamically stable. Mortality was higher in unstable patients and reoperation was higher with delayed presentation. Conclusions Thrombolysis with streptokinase is highly successful and safe therapy in hemodynamically stable as well as unstable patients, or those with early or delayed presentation with stuck bileaflet mitral valves, especially in centers where round the clock cardiothoracic surgery backup is not available. PMID:29904451
The sense of hearing in the Pacific oyster, Magallana gigas
Charifi, Mohcine; Sow, Mohamedou; Ciret, Pierre; Benomar, Soumaya
2017-01-01
There is an increasing concern that anthropogenic noise could have a significant impact on the marine environment, but there is still insufficient data for most invertebrates. What do they perceive? We investigated this question in oysters Magallana gigas (Crassostrea gigas) using pure tone exposures, accelerometer fixed on the oyster shell and hydrophone in the water column. Groups of 16 oysters were exposed to quantifiable waterborne sinusoidal sounds in the range of 10 Hz to 20 kHz at various acoustic energies. The experiment was conducted in running seawater using an experimental flume equipped with suspended loudspeakers. The sensitivity of the oysters was measured by recording their valve movements by high-frequency noninvasive valvometry. The tests were 3 min tone exposures including a 70 sec fade-in period. Three endpoints were analysed: the ratio of responding individuals in the group, the resulting changes of valve opening amplitude and the response latency. At high enough acoustic energy, oysters transiently closed their valves in response to frequencies in the range of 10 to <1000 Hz, with maximum sensitivity from 10 to 200 Hz. The minimum acoustic energy required to elicit a response was 0.02 m∙s-2 at 122 dBrms re 1 μPa for frequencies ranging from 10 to 80 Hz. As a partial valve closure cannot be differentiated from a nociceptive response, it is very likely that oysters detect sounds at lower acoustic energy. The mechanism involved in sound detection and the ecological consequences are discussed. PMID:29069092
Butchart, E G; Moreno de la Santa, P; Rooney, S J; Lewis, P A
1994-03-01
To determine the effect of risk factors and trigger factors on cerebrovascular events, 622 patients who survived mitral valve replacement between December 1979 and December 1992 were analyzed. Ninety-six patients suffered 139 nonhemorrhagic cerebrovascular events. Data were available on 138 events in 95 patients. There were 32 transient ischemic attacks (TIAs), 57 reversible ischemic neurological deficits (RINDs), and 49 strokes. Age, sex, atrial fibrillation, left atrial size, systemic hypertension, and abnormal body mass index did not discriminate between patients who suffered events and those who did not. In contrast, smoking status differed significantly between patients who suffered events and those who did not. Among current or recent ex-smokers, the risk of stroke or RIND was significantly higher than in non-smokers (p < < 0.001). The odds ratio of suffering any type of event in patients who smoked at any time postoperatively versus those who did not smoke was 2.9 (95% confidence interval: 1.8 to 4.6). Of 61 patients contacted directly, 30% recalled an infective episode immediately prior to their event. A diurnal and seasonal influence on events was also detected with peaks in the morning and in the winter months, respectively (both p < 0.001). It is concluded that there is persuasive evidence for the involvement of several nonprosthetic factors in the incidence of cerebrovascular events after mitral valve replacement. This has implications for patient management and for future analysis of prosthetic heart valve series.
Zhu, Weibin; Park, Jong M.; White, Michael J.; Nellis, Gregory F.; Gianchandani, Yogesh B.
2011-01-01
This article reports the evaluation of a Joule–Thomson (JT) cooling system that combines two custom micromachined components—a Si/glass-stack recuperative heat exchanger and a piezoelectrically actuated expansion microvalve. With the microvalve controlling the flow rate, this system can modulate cooling to accommodate varying refrigeration loads. The perforated plate Si/glass heat exchanger is fabricated with a stack of alternating silicon plates and Pyrex glass spacers. The microvalve utilizes a lead zirconate titanate actuator to push a Si micromachined valve seat against a glass plate, thus modulating the flow passing through the gap between the valve seat and the glass plate. The fabricated heat exchanger has a footprint of 1×1 cm2 and a length of 35 mm. The size of the micromachined piezoelectrically actuated valve is about 1×1×1 cm3. In JT cooling tests, the temperature of the system was successfully controlled by adjusting the input voltage of the microvalve. When the valve was fully opened (at an input voltage of −30 V), the system cooled down to a temperature as low as 254.5 K at 430 kPa pressure difference between inlet and outlet at steady state and 234 K at 710 kPa in a transient state. The system provided cooling powers of 75 mW at 255 K and 150 mW at 258 K. Parasitic heat loads at 255 K are estimated at approximately 700 mW. PMID:21552354
The sense of hearing in the Pacific oyster, Magallana gigas.
Charifi, Mohcine; Sow, Mohamedou; Ciret, Pierre; Benomar, Soumaya; Massabuau, Jean-Charles
2017-01-01
There is an increasing concern that anthropogenic noise could have a significant impact on the marine environment, but there is still insufficient data for most invertebrates. What do they perceive? We investigated this question in oysters Magallana gigas (Crassostrea gigas) using pure tone exposures, accelerometer fixed on the oyster shell and hydrophone in the water column. Groups of 16 oysters were exposed to quantifiable waterborne sinusoidal sounds in the range of 10 Hz to 20 kHz at various acoustic energies. The experiment was conducted in running seawater using an experimental flume equipped with suspended loudspeakers. The sensitivity of the oysters was measured by recording their valve movements by high-frequency noninvasive valvometry. The tests were 3 min tone exposures including a 70 sec fade-in period. Three endpoints were analysed: the ratio of responding individuals in the group, the resulting changes of valve opening amplitude and the response latency. At high enough acoustic energy, oysters transiently closed their valves in response to frequencies in the range of 10 to <1000 Hz, with maximum sensitivity from 10 to 200 Hz. The minimum acoustic energy required to elicit a response was 0.02 m∙s-2 at 122 dBrms re 1 μPa for frequencies ranging from 10 to 80 Hz. As a partial valve closure cannot be differentiated from a nociceptive response, it is very likely that oysters detect sounds at lower acoustic energy. The mechanism involved in sound detection and the ecological consequences are discussed.
Kusajima, Kunio; Fujita, Tomoyuki; Hata, Hiroki; Shimahara, Yusuke; Miura, Sayaka; Kobayashi, Junjiro
2016-01-01
OBJECTIVES Concomitant tricuspid valve surgery with mitral valve surgery is recommended for patients with severe functional tricuspid regurgitation (TR). However, the treatment for 2+ TR (mild TR) remains controversial. Here, we evaluated the long-term results of untreated 2+ TR in patients undergoing mitral valve surgery. METHODS We retrospectively reviewed the records of 96 patients with untreated 2+ TR among 885 patients who underwent mitral valve surgery from 2003 to 2010. Exclusion criteria were tricuspid valve surgery (TVS), emergency surgery, primary TR and pacemaker lead through the tricuspid valve. We assessed survival and freedom from heart failure. The freedom from 3+ (moderate) or 4+ (severe) TR was investigated by echocardiographic data at pre- and postoperative week 1, then at 1, 3, 5, 7 and 10 postoperative years, which were compared with those in patients who had 2+ TR preoperatively and underwent concomitant TVS in the same period (n = 47). RESULTS The mean follow-up was 7.1 ± 2.7 years. There was no 30-day mortality. The survival rate was 97.5% at 5 years and 87.5% at 10 years. The independent risk factors for mortality were age (OR 1.2, P = 0.03) and left ventricular ejection fraction (OR 0.9, P = 0.03). Untreated 2+ TR improved transiently within the first postoperative year (P < 0.001), but progressed again in the mid- to long term. Freedom from ≥3+ TR was 64.2% at 5 years and 46.7% at 10 years, which was significantly lower than that from ≥3+ TR in patients who underwent concomitant TVS (P = 0.006). The independent risk factors for TR progression (≥3 + TR) were age (OR 1.1, P = 0.005), atrial fibrillation (OR 2.2, P = 0.04) and tricuspid annular diameter (TAD) index (mm/m2; OR 1.1, P = 0.02). Receiver operating characteristic curves showed that the optimal TAD index cut-off value was 21.0 for long-term survival [area under the curve (AUC) = 0.72] and 21.2 for TR progression (AUC = 0.64). CONCLUSIONS Although untreated, 2+ TR significantly improved after mitral valve surgery, it then progressed again in the mid- to long term. Therefore, concomitant TVS should be considered in patients with 2+ TR who have dilated tricuspid annulus or atrial fibrillation, if feasible. PMID:26993477
NASA Technical Reports Server (NTRS)
1980-01-01
Detailed computer models of the engine were developed to predict both the steady state and transient operation of the engine system. Mechanical design layout drawings were prepared for the following components: thrust chamber and nozzle; extendible nozzle actuating mechanism and seal; LOX turbopump and boost pump; hydrogen turbopump and boost pump; and the propellant control valves. The necessary heat transfer, stress, fluid flow, dynamic, and performance analyses were performed to support the mechanical design.
1981-04-30
fluid temperature should exceed 145°F. The flow control module contains all the hydraulic circuit elements necessary for both the pressure line to and...are contained in three basic modules : 1) the hydraulic power supply, 2) a flow control module containing valving, accumulators and filters, and 3) the...hydraulic transient overpressures, is located in the flow control module , as are the high and low pressure filters. The load frame (MTS Systems Corp
DOE Office of Scientific and Technical Information (OSTI.GOV)
Dodson, R.J.; Feltus, M.A.
The low-temperature overpressurization protection system (LTOPS) is designed to protect the reactor pressure vessel (RPV) from brittle failure during startup and cooldown maneuvers in Westinghouse pressurized water reactors. For the Salem power plants, the power-operated relief valves (PORVs) mitigate pressure increases above a setpoint where an operational startup transient may put the RPV in the embrittlement fracture zone. The Title 10, Part 50, Code of Federal Regulations Appendix G limit, given by plant technical specifications, conservatively bounds the maximum pressure allowed during those transients where the RPV can suffer brittle fracture (usually below 350{degrees}F). The Appendix G limit is amore » pressure versus temperature curve that is more restrictive at lower RPV temperatures and allows for higher pressures as the temperature approaches the upper bounding fracture temperature.« less
Surge detection on an automotive turbocharger during transient phases
NASA Astrophysics Data System (ADS)
Deligant, M.; Danlos, A.; Podevin, P.; Clenci, A.; Guilain, S.
2017-10-01
The surge limit on automotive turbocharger needs to be avoided to prevent operations with pressure and mass flow oscillations. Mild surge is accompanied by noise which is disturbing. Deep surge can cause significant loss of engine power and severe drivability issues. It is necessary to know the stationary limit in order to match a turbocharger with an engine, ensuring enough surge margin. However, this choice does not guarantee surge free operation during transient functioning. In this paper, the surge onset of a compressor while closing a downstream valve is studied. Various tests have been carried out varying the closing time, the position of the initial operating point and the volume of the circuit. The inlet and outlet signals of physical parameters are analyzed with spectral and temporal methods in order to define the instant of the surge occurrence.
Pretest analysis document for Test S-FS-6
DOE Office of Scientific and Technical Information (OSTI.GOV)
Shaw, R.A.; Hall, D.G.
This report documents the pretest analyses completed for Semiscale Test S-FS-6. This test will simulate a transient initiated by a 100% break in a steam generator bottom feedwater line downstream of the check valve. The initial conditions represent normal operating conditions for a C-E System 80 nuclear power plant. Predictions of transients resulting from feedwater line breaks in these plants have indicated that significant primary system overpressurization may occur. The enclosed analyses include a RELAP5/MOD2/CY21 code calculation and preliminary results from a facility hot, integrated test which was conducted to near S-FS-6 specifications. The results of these analyses indicate thatmore » the test objectives for Test S-FS-6 can be achieved. The primary system overpressurization will pose no threat to personnel or plant integrity.« less
Intelligent Engine Systems: Adaptive Control
NASA Technical Reports Server (NTRS)
Gibson, Nathan
2008-01-01
We have studied the application of the baseline Model Predictive Control (MPC) algorithm to the control of main fuel flow rate (WF36), variable bleed valve (AE24) and variable stator vane (STP25) control of a simulated high-bypass turbofan engine. Using reference trajectories for thrust and turbine inlet temperature (T41) generated by a simulated new engine, we have examined MPC for tracking these two reference outputs while controlling a deteriorated engine. We have examined the results of MPC control for six different transients: two idle-to-takeoff transients at sea level static (SLS) conditions, one takeoff-to-idle transient at SLS, a Bode power command and reverse Bode power command at 20,000 ft/Mach 0.5, and a reverse Bode transient at 35,000 ft/Mach 0.84. For all cases, our primary focus was on the computational effort required by MPC for varying MPC update rates, control horizons, and prediction horizons. We have also considered the effects of these MPC parameters on the performance of the control, with special emphasis on the thrust tracking error, the peak T41, and the sizes of violations of the constraints on the problem, primarily the booster stall margin limit, which for most cases is the lone constraint that is violated with any frequency.
NASA Astrophysics Data System (ADS)
Vitkovskii, I. L.; Nikonov, S. P.; Ryasnyi, S. I.
2014-02-01
The subject of this paper is a transient caused by connection of a standby loop to three operating circulation pumps at the initial reactor heat rate equal to 70% of the rated value without preliminarily reducing it to 30% of the rated level as required by the safe operation regulations. Failure of the following normal operation systems is supposed: the first- and the second-type warning protection systems, all quick-acting reducing devices releasing steam into the auxiliary manifold, the electric heaters of the pressurizer, the pressurizer injection system, the primary cooling circuit fluid makeup/blow-through systems, and the blocking systems to shut down the main circulation pump after the level in the steam generator is exceeded. In addition, it is supposed that, under transient conditions, the valves of the turbine regulation system will be in the position in which they were at the moment of the initial event until generation of the signal for positive closing of the turbine stop valves. The first signal to actuate the reactor emergency protection system (EPS) is skipped. The failure of all quick-acting reducing devices releasing steam into the atmosphere is assumed. In addition to equipment failure, at the moment when the main circulation pump is connected, the operator erroneously puts in a new setting to maintain the power allowable for four pumps in operation-in the calculations it was taken equal to 104% of the rated level at most considering the accuracy of evaluating and maintaining the reactor heat rate-and the working group of the reactor protection and control system (P&CS) starts moving upward. On reaching the set power level, the automatic reactor power regulator stops operating and the P&CS elements remain in the position in which they are at the moment. Compliance with the design safety criteria for the adopted scenario of the transient is demonstrated.
NASA Astrophysics Data System (ADS)
Yi, Guodong; Li, Jin
2018-03-01
The master cylinder hydraulic system is the core component of the fineblanking press that seriously affects the machine performance. A key issue in the design of the master cylinder hydraulic system is dealing with the heavy shock loads in the fineblanking process. In this paper, an equivalent model of the master cylinder hydraulic system is established based on typical process parameters for practical fineblanking; then, the response characteristics of the master cylinder slider to the step changes in the load and control current are analyzed, and lastly, control strategies for the proportional valve are studied based on the impact of the control parameters on the kinetic stability of the slider. The results show that the kinetic stability of the slider is significantly affected by the step change of the control current, while it is slightly affected by the step change of the system load, which can be improved by adjusting the flow rate and opening time of the proportional valve.
Unusual cause of central aortic prosthetic regurgitation during transcatheter replacement.
López-Mínguez, José Ramón; Millán-Núñez, Victoria; González-Fernández, Reyes; Nogales-Asensio, Juan Manuel; Fuentes-Cañamero, María Eugenia; Merchán-Herrera, Antonio
2016-04-01
Transcatheter aortic valve replacement (TAVR) is an increasingly common procedure for the treatment of aortic stenosis in elderly patients with comorbidities that prevent the use of standard surgery. It has been shown that implantation without aortic regurgitation is related to lower mortality. Mild paravalvular regurgitation is inevitable in some cases due to calcification of the aortic annulus and its usually somewhat elliptical shape. Central regurgitation is less common, but has been associated with valve overdilatation in cases in which reduction of paravalvular regurgitation was attempted after the initial inflation. However, there are no reported cases of central prosthetic aortic regurgitation due to acute LV dysfunction. We report a case in which central aortic regurgitation occurred due to transient ventricular dysfunction secondary to occlusion of the right coronary artery by an embolus. The regurgitation disappeared after thrombus aspiration and normal ventricular function was immediately recovered. Copyright © 2016 Sociedade Portuguesa de Cardiologia. Published by Elsevier España. All rights reserved.
Kusajima, Kunio; Fujita, Tomoyuki; Hata, Hiroki; Shimahara, Yusuke; Miura, Sayaka; Kobayashi, Junjiro
2016-07-01
Concomitant tricuspid valve surgery with mitral valve surgery is recommended for patients with severe functional tricuspid regurgitation (TR). However, the treatment for 2+ TR (mild TR) remains controversial. Here, we evaluated the long-term results of untreated 2+ TR in patients undergoing mitral valve surgery. We retrospectively reviewed the records of 96 patients with untreated 2+ TR among 885 patients who underwent mitral valve surgery from 2003 to 2010. Exclusion criteria were tricuspid valve surgery (TVS), emergency surgery, primary TR and pacemaker lead through the tricuspid valve. We assessed survival and freedom from heart failure. The freedom from 3+ (moderate) or 4+ (severe) TR was investigated by echocardiographic data at pre- and postoperative week 1, then at 1, 3, 5, 7 and 10 postoperative years, which were compared with those in patients who had 2+ TR preoperatively and underwent concomitant TVS in the same period (n = 47). The mean follow-up was 7.1 ± 2.7 years. There was no 30-day mortality. The survival rate was 97.5% at 5 years and 87.5% at 10 years. The independent risk factors for mortality were age (OR 1.2, P = 0.03) and left ventricular ejection fraction (OR 0.9, P = 0.03). Untreated 2+ TR improved transiently within the first postoperative year (P < 0.001), but progressed again in the mid- to long term. Freedom from ≥3+ TR was 64.2% at 5 years and 46.7% at 10 years, which was significantly lower than that from ≥3+ TR in patients who underwent concomitant TVS (P = 0.006). The independent risk factors for TR progression (≥3 + TR) were age (OR 1.1, P = 0.005), atrial fibrillation (OR 2.2, P = 0.04) and tricuspid annular diameter (TAD) index (mm/m(2); OR 1.1, P = 0.02). Receiver operating characteristic curves showed that the optimal TAD index cut-off value was 21.0 for long-term survival [area under the curve (AUC) = 0.72] and 21.2 for TR progression (AUC = 0.64). Although untreated, 2+ TR significantly improved after mitral valve surgery, it then progressed again in the mid- to long term. Therefore, concomitant TVS should be considered in patients with 2+ TR who have dilated tricuspid annulus or atrial fibrillation, if feasible. © The Author 2016. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.
NASA Astrophysics Data System (ADS)
Saberimoghaddam, Ali; Bahri Rasht Abadi, Mohammad Mahdi
2018-01-01
Joule-Thomson cooling systems are commonly used in gas liquefaction. In small gas liquefiers, transient cool-down time is high. Selecting suitable conditions for cooling down process leads to decrease in time and cost. In the present work, transient thermal behavior of Joule-Thomson cooling system including counter current helically coiled tube in tube heat exchanger, expansion valve, and collector was studied using experimental tests and simulations. The experiments were performed using small gas liquefier and nitrogen gas as working fluid. The heat exchanger was thermally studied by experimental data obtained from a small gas liquefier. In addition, the simulations were performed using experimental data as variable boundary conditions. A comparison was done between presented and conventional methods. The effect of collector heat capacity and convection heat transfer coefficient inside the tubes on system performance was studied using temperature profiles along the heat exchanger.
ATWS analysis for Browns Ferry Nuclear Plant Unit 1
DOE Office of Scientific and Technical Information (OSTI.GOV)
Dallman, R.J.; Jouse, W.C.
1985-01-01
Analyses of postulated Anticipated Transients Without Scram (ATWS) were performed at the Idaho National Engineering Laboratory (INEL). The Browns Ferry Nuclear Plant Unit 1 (BFNP1) was selected as the subject of this work because of the cooperation of the Tennessee Valley Authority (TVA). The work is part of the Severe Accident Sequence Analysis (SASA) Program of the US Nuclear Regulatory Commission (NRC). A Main Steamline Isolation Valve (MSIV) closure served as the transient initiator for these analyses, which proceeded a complete failure to scram. Results from the analyses indicate that operator mitigative actions are required to prevent overpressurization of themore » primary containment. Uncertainties remain concerning the effectiveness of key mitigative actions. The effectiveness of level control as a power reduction procedure is limited. Power level resulting from level control only reduce the Pressure Suppression Pool (PSP) heatup rate from 6 to 4F/min.« less
Software Models Impact Stresses
NASA Technical Reports Server (NTRS)
Hanshaw, Timothy C.; Roy, Dipankar; Toyooka, Mark
1991-01-01
Generalized Impact Stress Software designed to assist engineers in predicting stresses caused by variety of impacts. Program straightforward, simple to implement on personal computers, "user friendly", and handles variety of boundary conditions applied to struck body being analyzed. Applications include mathematical modeling of motions and transient stresses of spacecraft, analysis of slamming of piston, of fast valve shutoffs, and play of rotating bearing assembly. Provides fast and inexpensive analytical tool for analysis of stresses and reduces dependency on expensive impact tests. Written in FORTRAN 77. Requires use of commercial software package PLOT88.
Normal and Abnormal Development of the Intrapericardial Arterial Trunks in Man and Mouse
Anderson, Robert H.; Chaudhry, Bill; Mohun, Timothy J.; Bamforth, Simon D.; Hoyland, Darren; Phillips, Helen M.; Webb, Sandra; Moorman, Antoon F.J.; Brown, Nigel A.; Henderson, Deborah J.
2014-01-01
Aims The definitive cardiac outflow channels have three components: the intrapericardial arterial trunks; the arterial roots with valves; and the ventricular outflow tracts. We studied the normal and abnormal development of the most distal of these, the arterial trunks, comparing findings in mouse and man. Methods and Results Using lineage tracing and three-dimensional visualization by episcopic reconstruction and scanning electron microscopy, we studied embryonic day 9.5 to 12.5 mouse hearts, clarifying the development of the outflow tracts distal to the primordia of the arterial valves. We characterize a transient aortopulmonary foramen, located between the leading edge of a protrusion from the dorsal wall of the aortic sac and the distal margins of the two outflow cushions. The foramen is closed by fusion of the protrusion, with its cap of neural crest cells, with the neural crest cell-filled cushions; the resulting structure then functioning transiently as an aortopulmonary septum. Only subsequent to this closure is it possible to recognize, more proximally, the previously described aortopulmonary septal complex. The adjacent walls of the intrapericardial trunks are derived from the protrusion and distal parts of the outflow cushions, while the lateral walls are formed from intrapericardial extensions of pharyngeal mesenchyme derived from the second heart field. Conclusions We provide, for the first time, objective evidence of the mechanisms of closure of an aortopulmonary foramen that exists distally between the lumens of the developing intrapericardial arterial trunks. Our findings provide insights into the formation of aortopulmonary windows and the variants of common arterial trunk. PMID:22499773
VO2 Off Transient Kinetics in Extreme Intensity Swimming.
Sousa, Ana; Figueiredo, Pedro; Keskinen, Kari L; Rodríguez, Ferran A; Machado, Leandro; Vilas-Boas, João P; Fernandes, Ricardo J
2011-01-01
Inconsistencies about dynamic asymmetry between the on- and off- transient responses in oxygen uptake are found in the literature. Therefore, the purpose of this study was to characterize the oxygen uptake off-transient kinetics during a maximal 200-m front crawl effort, as examining the degree to which the on/off regularity of the oxygen uptake kinetics response was preserved. Eight high level male swimmers performed a 200-m front crawl at maximal speed during which oxygen uptake was directly measured through breath-by-breath oxymetry (averaged every 5 s). This apparatus was connected to the swimmer by a low hydrodynamic resistance respiratory snorkel and valve system. The on- and off-transient phases were symmetrical in shape (mirror image) once they were adequately fitted by a single-exponential regression models, and no slow component for the oxygen uptake response was developed. Mean (± SD) peak oxygen uptake was 69.0 (± 6.3) mL·kg(-1)·min(-1), significantly correlated with time constant of the off- transient period (r = 0.76, p < 0.05) but not with any of the other oxygen off-transient kinetic parameters studied. A direct relationship between time constant of the off-transient period and mean swimming speed of the 200-m (r = 0.77, p < 0.05), and with the amplitude of the fast component of the effort period (r = 0.72, p < 0.05) were observed. The mean amplitude and time constant of the off-transient period values were significantly greater than the respective on- transient. In conclusion, although an asymmetry between the on- and off kinetic parameters was verified, both the 200-m effort and the respectively recovery period were better characterized by a single exponential regression model. Key pointsThe VO2 slow component was not observed in the recovery period of swimming extreme efforts;The on and off transient periods were better fitted by a single exponential function, and so, these effort and recovery periods of swimming extreme efforts are symmetrical;The rate of VO2 decline during the recovery period may be due to not only the magnitude of oxygen debt but also the VO2peak obtained during the effort period.
Best-estimate coupled RELAP/CONTAIN analysis of inadvertent BWR ADS valve opening transient
DOE Office of Scientific and Technical Information (OSTI.GOV)
Feltus, M.A.; Muftuoglu, A.K.
1993-01-01
Noncondensible gases may become dissolved in boiling water reactor (BWR) water-level instrumentation during normal operations. Any dissolved noncondensible gases inside these water columns may come out of solution during rapid depressurization events and displace water from the reference leg piping, resulting in a false high level. Significant errors in water-level indication are not expected to occur until the reactor pressure vessel (RPV) pressure has dropped below [approximately]450 psig. These water level errors may cause a delay or failure in emergency core cooling system (ECCS) actuation. The RPV water level is monitored using the pressure of a water column having amore » varying height (reactor water level) that is compared to the pressure of a water column maintained at a constant height (reference level). The reference legs have small-diameter pipes with varying lengths that provide a constant head of water and are located outside the drywell. The amount of noncondensible gases dissolved in each reference leg is very dependent on the amount of leakage from the reference leg and its geometry and interaction of the reactor coolant system with the containment, i.e., torus or suppression pool, and reactor building. If a rapid depressurization causes an erroneously high water level, preventing automatic ECCS actuation, it becomes important to determine if there would be other adequate indications for operator response. In the postulated inadvertent opening of all seven automatic depressurization system (ADS) valves, the ECCS signal on high drywell pressure would be circumvented because the ADS valves discharge directly into the suppression pool. A best-estimate analysis of such an inadvertent opening of all ADS valves would have to consider the thermal-hydraulic coupling between the pool, drywell, reactor building, and RPV.« less
Anticoagulation in pregnant women with mechanical heart valve prostheses
Meschengieser, S; Fondevila, C; Santarelli, M; Lazzari, M
1999-01-01
OBJECTIVE—To evaluate the outcome of pregnancy in women with mechanical heart valve prostheses in relation to the anticoagulant treatment used in the first trimester and the incidence of thrombotic and bleeding complications. METHODS—92 pregnancies in 59 women were followed between 1986 and 1997. In 31 pregnancies, oral anticoagulants were discontinued when pregnancy was diagnosed and subcutaneous heparin was started (12 500 U every 12 hours) adjusted to prolong the adjusted partial thromboplastin time to twice the control level. In the second trimester oral anticoagulants were resumed but changed to heparin again 15 days before the expected delivery date. In 61 pregnancies oral anticoagulants were continued during the first trimester. The same regimen of heparin was used for delivery. RESULTS—Abortion or fetal losses were similar (p = 0.5717) in women exposed to oral anticoagulants in the first trimester (13/61; 25%) compared with those who received adjusted subcutaneous heparin (6/31; 19%). Embolic episodes were more common (p = 0.0029) in women who received heparin (4.92%) compared with those on oral anticoagulants (0.33%). Embolic episodes were cerebral and transient. No valve thromboses were observed. No malformations appeared in the 71 newborns, except for one case of hydrocephalus. There were no maternal deaths secondary to thrombotic complications. The only death was the result of major bleeding after the delivery of a premature stillborn. CONCLUSIONS—Oral anticoagulants seem to be safer for the mother than adjusted subcutaneous heparin. Heparin does not offer a clear advantage over oral anticoagulation in the pregnancy outcome. Keywords: pregnancy; oral anticoagulants; heparin; prosthetic valves PMID:10377303
Statistical characteristics of mechanical heart valve cavitation in accelerated testing.
Wu, Changfu; Hwang, Ned H C; Lin, Yu-Kweng M
2004-07-01
Cavitation damage has been observed on mechanical heart valves (MHVs) undergoing accelerated testing. Cavitation itself can be modeled as a stochastic process, as it varies from beat to beat of the testing machine. This in-vitro study was undertaken to investigate the statistical characteristics of MHV cavitation. A 25-mm St. Jude Medical bileaflet MHV (SJM 25) was tested in an accelerated tester at various pulse rates, ranging from 300 to 1,000 bpm, with stepwise increments of 100 bpm. A miniature pressure transducer was placed near a leaflet tip on the inflow side of the valve, to monitor regional transient pressure fluctuations at instants of valve closure. The pressure trace associated with each beat was passed through a 70 kHz high-pass digital filter to extract the high-frequency oscillation (HFO) components resulting from the collapse of cavitation bubbles. Three intensity-related measures were calculated for each HFO burst: its time span; its local root-mean-square (LRMS) value; and the area enveloped by the absolute value of the HFO pressure trace and the time axis, referred to as cavitation impulse. These were treated as stochastic processes, of which the first-order probability density functions (PDFs) were estimated for each test rate. Both the LRMS value and cavitation impulse were log-normal distributed, and the time span was normal distributed. These distribution laws were consistent at different test rates. The present investigation was directed at understanding MHV cavitation as a stochastic process. The results provide a basis for establishing further the statistical relationship between cavitation intensity and time-evolving cavitation damage on MHV surfaces. These data are required to assess and compare the performance of MHVs of different designs.
Durães, André Rodrigues; de Souza Roriz, Pollianna; de Almeida Nunes, Bianca; Albuquerque, Felipe Pinho E; de Bulhões, Fábio Vieira; de Souza Fernandes, Andre Mauricio; Aras, Roque
2016-06-01
Dabigatran is a direct thrombin inhibitor shown to be an effective alternative to warfarin in patients with non-valvular atrial fibrillation (AF). We evaluated the use of dabigatran in patients with bioprosthetic mitral and/or aortic valve replacement and AF. We selected 34 and randomized 27 patients in a 1:1 ratio to receive dabigatran or warfarin. The primary endpoint was the presence of a new intracardiac thrombus at 90 days, by transesophageal echocardiogram (TEE). Secondary endpoints included the development of dense spontaneous echo contrast (SEC) and incidence of stroke (ischemic or hemorrhagic), myocardium infarction, valve thrombosis and peripheral embolic events. The trial was terminated prematurely because of low enrollment. There were 27 patients in total: 15 patients placed in the dabigatran group and 12 in the warfarin group. After 90 days, one patient (8.3 %) in the warfarin group and none in the dabigatran group had developed a new intracardiac thrombus. In the dabigatran group, two patients (13.3 %) developed dense SEC versus one patient (8.3 %) in the warfarin group. In the warfarin group, one patient (8.3 %) presented ischemic stroke, and none did in the dabigatran group. We observed no cases of hemorrhagic stroke, valve thrombosis, embolic events or myocardial infarction in either group throughout the study. However, one patient (6.7 %) in the dabigatran group had a fully recovered transient ischemic attack and one patient in the warfarin group died of heart failure. The use of dabigatran appears to be similar to warfarin in preventing the formation of intracardiac thrombus. Clinicaltrials.gov NCT01868243.
Advanced space engine powerhead breadboard assembly system study
NASA Technical Reports Server (NTRS)
Campbell, R. G.
1978-01-01
The objective of this study was to establish a preliminary design of a Powerhead Breadboard Assembly (PBA) for an 88 964-Newton (20,000-pound) thrust oxygen/hydrogen staged combustion cycle engine for use in orbital transfer vehicle propulsion. Existing turbopump, preburner, and thrust chamber components were integrated with interconnecting ducting, a heat exchanger, and a control system to complete the PBA design. Cycle studies were conducted to define starting transients and steady-state balances for the completed design. Specifications were developed for all valve applications and the conditions required for the control system integration with the facility for system test were defined.
NASA Technical Reports Server (NTRS)
Clark, E. C.
1975-01-01
Thruster valve assemblies (T/VA's) were subjected to the development test program for the combined JPL Low-Cost Standardized Spacecraft Equipment (LCSSE) and Mariner Jupiter/Saturn '77 spacecraft (MJS) programs. The development test program was designed to achieve the following program goals: (1) demonstrate T/VA design compliance with JPL Specifications, (2) to conduct a complete performance Cf map of the T/VA over the full operating range of environment, (3) demonstrate T/VA life capability and characteristics of life margin for steady-state limit cycle and momentum wheel desaturation duty cycles, (4) verification of structural design capability, and (5) generate a computerized performance model capable of predicting T/VA operation over pressures ranging from 420 to 70 psia, propellant temperatures ranging from 140 F to 40 F, pulse widths of 0.008 to steady-state operation with unlimited duty cycle capability, and finally predict the transient performance associated with reactor heatup during any given duty cycle, start temperature, feed pressure, and propellant temperature conditions.
NASA Technical Reports Server (NTRS)
Ryan, Harry M.; Coote, David J.; Ahuja, Vineet; Hosangadi, Ashvin
2006-01-01
Accurate modeling of liquid rocket engine test processes involves assessing critical fluid mechanic and heat and mass transfer mechanisms within a cryogenic environment, and accurately modeling fluid properties such as vapor pressure and liquid and gas densities as a function of pressure and temperature. The Engineering and Science Directorate at the NASA John C. Stennis Space Center has developed and implemented such analytic models and analysis processes that have been used over a broad range of thermodynamic systems and resulted in substantial improvements in rocket propulsion testing services. In this paper, we offer an overview of the analyses techniques used to simulate pressurization and propellant fluid systems associated with the test stands at the NASA John C. Stennis Space Center. More specifically, examples of the global performance (one-dimensional) of a propellant system are provided as predicted using the Rocket Propulsion Test Analysis (RPTA) model. Computational fluid dynamic (CFD) analyses utilizing multi-element, unstructured, moving grid capability of complex cryogenic feed ducts, transient valve operation, and pressurization and mixing in propellant tanks are provided as well.
ATWS at Browns Ferry Unit One - accident sequence analysis
DOE Office of Scientific and Technical Information (OSTI.GOV)
Harrington, R.M.; Hodge, S.A.
1984-07-01
This study describes the predicted response of Unit One at the Browns Ferry Nuclear Plant to a postulated complete failure to scram following a transient occurrence that has caused closure of all Main Steam Isolation Valves (MSIVs). This hypothetical event constitutes the most severe example of the type of accident classified as Anticipated Transient Without Scram (ATWS). Without the automatic control rod insertion provided by scram, the void coefficient of reactivity and the mechanisms by which voids are formed in the moderator/coolant play a dominant role in the progression of the accident. Actions taken by the operator greatly influence themore » quantity of voids in the coolant and the effect is analyzed in this report. The progression of the accident sequence under existing and under recommended procedures is discussed. For the extremely unlikely cases in which equipment failure and wrongful operator actions might lead to severe core damage, the sequence of emergency action levels and the associated timing of events are presented.« less
RAMONA-3B application to Browns Ferry ATWS
DOE Office of Scientific and Technical Information (OSTI.GOV)
Slovik, G.C.; Neymotin, L.Y.; Saha, P.
1985-01-01
The Anticipated Transient Without Scram (ATWS) is known to be a dominant accident sequence for possible core melt in a Boiling Water Reactor (BWR). A recent Probabilistic Risk Assessment (PRA) analysis for the Browns Ferry nuclear power plant indicates that ATWS is the second most dominant transient for core melt in BWR/4 with Mark I containment. The most dominant sequence being the failure of long term decay heat removal function of the Residual Heat Removal (RHR) system. Of all the various ATWS scenarios, the Main Steam Isolation Valve (MSIV) closure ATWS sequence was chosen for present analysis because of itsmore » relatively high frequency of occurrence and its challenge to the residual heat removal system and containment integrity. The objective of this paper is to discuss four MSIV closure ATWS calculations using the RAMONA-3B code. The paper is a summary of a report being prepared for the USNRC Severe Accident Sequence Analysis (SASA) program which should be referred to for details. 10 refs., 20 figs., 3 tabs.« less
Frequency Dynamics of the First Heart Sound
NASA Astrophysics Data System (ADS)
Wood, John Charles
Cardiac auscultation is a fundamental clinical tool but first heart sound origins and significance remain controversial. Previous clinical studies have implicated resonant vibrations of both the myocardium and the valves. Accordingly, the goals of this thesis were threefold, (1) to characterize the frequency dynamics of the first heart sound, (2) to determine the relative contribution of the myocardium and the valves in determining first heart sound frequency, and (3) to develop new tools for non-stationary signal analysis. A resonant origin for first heart sound generation was tested through two studies in an open-chest canine preparation. Heart sounds were recorded using ultralight acceleration transducers cemented directly to the epicardium. The first heart sound was observed to be non-stationary and multicomponent. The most dominant feature was a powerful, rapidly-rising frequency component that preceded mitral valve closure. Two broadband components were observed; the first coincided with mitral valve closure while the second significantly preceded aortic valve opening. The spatial frequency of left ventricular vibrations was both high and non-stationary which indicated that the left ventricle was not vibrating passively in response to intracardiac pressure fluctuations but suggested instead that the first heart sound is a propagating transient. In the second study, regional myocardial ischemia was induced by left coronary circumflex arterial occlusion. Acceleration transducers were placed on the ischemic and non-ischemic myocardium to determine whether ischemia produced local or global changes in first heart sound amplitude and frequency. The two zones exhibited disparate amplitude and frequency behavior indicating that the first heart sound is not a resonant phenomenon. To objectively quantify the presence and orientation of signal components, Radon transformation of the time -frequency plane was performed and found to have considerable potential for pattern classification. Radon transformation of the Wigner spectrum (Radon-Wigner transform) was derived to be equivalent to dechirping in the time and frequency domains. Based upon this representation, an analogy between time-frequency estimation and computed tomography was drawn. Cohen's class of time-frequency representations was subsequently shown to result from simple changes in reconstruction filtering parameters. Time-varying filtering, adaptive time-frequency transformation and linear signal synthesis were also performed from the Radon-Wigner representation.
Hosoba, Soh; Yamamoto, Masanori; Shioda, Kayoko; Sago, Mitsuru; Koyama, Yutaka; Shimura, Tetsuro; Kagase, Ai; Tada, Norio; Naganuma, Toru; Araki, Motoharu; Yamanaka, Futoshi; Shirai, Shinichi; Watanabe, Yusuke; Hayashida, Kentaro
2018-03-01
Favourable results have been reported for monitored anaesthesia care that includes local anaesthesia and conscious sedation [minimalist approach (MA)] for transfemoral transcatheter aortic valve replacement (TAVR). However, the efficacy of MA is still controversial in Japan. We describe our experience from a Japanese multicentre registry. Between October 2013 and April 2016, 1215 consecutive Japanese patients with symptomatic, severe aortic stenosis undergoing TAVR with self-expandable or balloon-expandable valves were prospectively included in the Optimized transCathEter vAlvular intervention-Transcatheter Aortic Valve Implantation (OCEAN-TAVI) registry. Of these patients, we retrospectively reviewed 921 consecutive patients who underwent elective transfemoral-TAVR. We evaluated the perioperative results of MA-TAVR and non-minimalist approach (NMA) TAVR using propensity score matching analysis. A total of 118 patients underwent MA-TAVR, and 802 patients underwent NMA-TAVR [median age 84 vs 85 years, P = 0.25; Society of Thoracic Surgeons (STS) score 7.6 vs 6.4, P = 0.01]. One hundred eighteen matched pairs were compared after propensity score matching. In-hospital mortality and stroke/transient ischaemic attack were not significantly different between the MA-TAVR and the NMA-TAVR groups (2.5% vs 0.8%, P = 0.3; 1.7% vs 0.8%, P = 0.6, respectively). Major or life-threatening bleeding and the transfusion rate were significantly lower in the MA-TAVR group (3.4% vs 17%, P = 0.003; 6.8% vs 29%, P = 0.0002, respectively). The total intensive care unit days and length of hospital stay were significantly lower in the MA-TAVR group (P ≤ 0.0002). MA-TAVR has similar results to NMA-TAVR in terms of mortality and stroke in this Japanese multicentre registry. Shorter procedure time and hospital stays were seen in the MA-TAVR group. MA-TAVR is as safe and effective as NMA-TAVR. © The Author 2017. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.
A transient fault-valve mechanism operating in upper crustal level, Sierras Pampeanas, Argentina
NASA Astrophysics Data System (ADS)
Japas, María Silvia; Urbina, Nilda Esther; Sruoga, Patricia; Garro, José Matías; Ibañes, Oscar
2016-11-01
Located in the Sierras Pampeanas (the broken-foreland of the Pampean flat slab segment in the southern Central Andes), the Cerro Tiporco volcanic field shows Neogene hydrothermal activity linked to migration of arc-magmatism into the foreland. Late Neogene deposits comprise epithermal vein systems emplaced in Precambrian-Early Palaeozoic igneous-metamorphic basement, Late Miocene sedimentary rocks and Early Pliocene volcaniclastic rocks. Mineralization consists of calcareous onyx, aragonite and calcite veins as well as travertine deposits. Onyx and aragonite occur as fill of low-displacement nearly vertical reverse-sinistral faults striking NW, and nearly horizontal dilatant fractures. The latter consist of load-removal induced fractures affecting the igneous-metamorphic rocks, as well as bedding planes in the Late Miocene sediments. The presence of veins recording multiple fracture episodes and crack-and-seal growth of veins suggests relatively low differential stress and supralithostatic fluid pressure, as well as cyclic changes in pore pressure and high mineral-deposition/fracture-opening ratio. These conditions support a mechanism of fault-valve behaviour during onyx and aragonite vein emplacement. The fault-valve mechanism involves fractures associated with impermeable barriers between environments with different fluid pressure. Faulting generated an appreciable directional permeability triggering fluid migration from the highest to the lowest pressure region, with subsequent deposition and sealing that started a new pressurization-faulting-sealing cycle. Late aragonite and calcite veins suggest a change in kinematics indicating the onset of tectonic-load conditions.
Jablonski, Ireneusz; Mroczka, Janusz
2010-01-01
The paper offers an enhancement of the classical interrupter technique algorithm dedicated to respiratory mechanics measurements. Idea consists in exploitation of information contained in postocclusional transient states during indirect measurement of parameter characteristics by model identification. It needs the adequacy of an inverse analogue to general behavior of the real system and a reliable algorithm of parameter estimation. The second one was a subject of reported works, which finally showed the potential of the approach to separation of airway and tissue response in a case of short-term excitation by interrupter valve operation. Investigations were conducted in a regime of forward-inverse computer experiment.
Modeling of Transient Flow Mixing of Streams Injected into a Mixing Chamber
NASA Technical Reports Server (NTRS)
Voytovych, Dmytro M.; Merkle, Charles L.; Lucht, Robert P.; Hulka, James R.; Jones, Gregg W.
2006-01-01
Ignition is recognized as one the critical drivers in the reliability of multiple-start rocket engines. Residual combustion products from previous engine operation can condense on valves and related structures thereby creating difficulties for subsequent starting procedures. Alternative ignition methods that require fewer valves can mitigate the valve reliability problem, but require improved understanding of the spatial and temporal propellant distribution in the pre-ignition chamber. Current design tools based mainly on one-dimensional analysis and empirical models cannot predict local details of the injection and ignition processes. The goal of this work is to evaluate the capability of the modern computational fluid dynamics (CFD) tools in predicting the transient flow mixing in pre-ignition environment by comparing the results with the experimental data. This study is a part of a program to improve analytical methods and methodologies to analyze reliability and durability of combustion devices. In the present paper we describe a series of detailed computational simulations of the unsteady mixing events as the cold propellants are first introduced into the chamber as a first step in providing this necessary environmental description. The present computational modeling represents a complement to parallel experimental simulations' and includes comparisons with experimental results from that effort. A large number of rocket engine ignition studies has been previously reported. Here we limit our discussion to the work discussed in Refs. 2, 3 and 4 which is both similar to and different from the present approach. The similarities arise from the fact that both efforts involve detailed experimental/computational simulations of the ignition problem. The differences arise from the underlying philosophy of the two endeavors. The approach in Refs. 2 to 4 is a classical ignition study in which the focus is on the response of a propellant mixture to an ignition source, with emphasis on the level of energy needed for ignition and the ensuing flame propagation issues. Our focus in the present paper is on identifying the unsteady mixing processes that provide the propellant mixture in which the ignition source is to be placed. In particular, we wish to characterize the spatial and temporal mixture distribution with a view toward identifying preferred spatial and temporal locations for the ignition source. As such, the present work is limited to cold flow (pre-ignition) conditions
Mayr, Benedikt; Firschke, Christian; Erlebach, Magdalena; Bleiziffer, Sabine; Krane, Markus; Joner, Michael; Herold, Ulf; Nöbauer, Christian; Lange, Rüdiger; Deutsch, Marcus-André
2018-02-26
Simultaneous surgical off-pump coronary revascularization and transcatheter aortic valve implantation (TAVI) as a hybrid procedure may be a therapeutic option for patients with a TAVI indication who are not suitable for percutaneous coronary intervention and for patients who have an indication for combined surgical aortic valve implantation and coronary artery bypass grafting but present with a porcelain aorta. Early outcomes of these patients are analysed in this study. From February 2011 to April 2017, hybrid TAVI/off-pump coronary artery bypass (OPCAB) was performed in 12 (60%) patients, hybrid TAVI/minimally invasive direct coronary artery bypass in 6 (30%) patients and staged TAVI/OPCAB in 2 (10%) patients. Endpoints of this study were 30-day mortality, device success and postoperative adverse events as defined by the updated Valve Academic Research Consortium (VARC-2). The median age at the time of surgery was 77 years [interquartile range (IQR), 70-81] with a median logistic EuroSCORE and Society of Thoracic Surgeons' Predicted Risk score of 16.1% (IQR, 9.3-28.1) and 3.9% (IQR, 2.2-5.6), respectively. The median Synergy between PCI with Taxus and Cardiac Surgery score was 16.5 (IQR, 9.8-22.8). TAVI implantation routes were transaortic in 9 (45%) patients, transapical and transfemoral in 5 (25%) patients each and transsubclavian in 1 (5%) patient. Complete myocardial revascularization was achieved in 75% of patients. Device success rate was 100%. Paravalvular aortic regurgitation did not exceed mild in any patient. Stroke/transient ischaemic attack, vascular complications and myocardial infarction were not observed. Re-exploration for bleeding was required in 1 (5%) patient. Thirty-day mortality was 0%. Hybrid OPCAB/MIDCAB and TAVI prove to be a safe and effective alternative treatment option in selected higher risk patients.
2014-01-01
Background Mechanical heart valve replacement has an inherent risk of thromboembolic events (TEs). Current guidelines recommend an international normalized ratio (INR) of at least 2.5 after mechanical mitral valve replacement (MVR). This study aimed to evaluate the effects of a low INR (2.0–2.5) on thromboembolic and bleeding complications in patients with mechanical MVR on warfarin therapy. Methods One hundred and thirty-five patients who underwent mechanical MVR were enrolled in this study. The end points of this study were defined as TEs (valve thrombosis, transient ischemic attack, stroke) and bleeding (all minor and major bleeding) complications. Patients were followed up for a mean of 39.6 months and the mean INR of the patients was calculated. After data collection, patients were divided into 3 groups according to their mean INR, as follows: group 1 (n = 34), INR <2.0; group 2 (n = 49), INR 2.0–2.5; and group 3 (n = 52), INR >2.5. Results A total of 22 events (10 [7.4%] thromboembolic and 12 [8.8%] bleeding events) occurred in the follow-up period. The mean INR was an independent risk factor for the development of TEs. Mean INR and neurological dysfunction were independent risk factors for the development of bleeding events. A statistically significant positive correlation was found between the log mean INR and all bleeding events, and a negative correlation was found between the log mean INR and all TEs. The total number of events was significantly lower in group 2 than in groups 1 and 3 (P = 0.036). Conclusions This study showed that a target INRs of 2.0–2.5 are acceptable for preventing TEs and safe in terms of bleeding complications in patients with mechanical MVR. PMID:24885719
Advanced Supercritical Carbon Dioxide Brayton Cycle Development
DOE Office of Scientific and Technical Information (OSTI.GOV)
Anderson, Mark; Sienicki, James; Moisseytsev, Anton
2015-10-21
Fluids operating in the supercritical state have promising characteristics for future high efficiency power cycles. In order to develop power cycles using supercritical fluids, it is necessary to understand the flow characteristics of fluids under both supercritical and two-phase conditions. In this study, a Computational Fluid Dynamic (CFD) methodology was developed for supercritical fluids flowing through complex geometries. A real fluid property module was implemented to provide properties for different supercritical fluids. However, in each simulation case, there is only one species of fluid. As a result, the fluid property module provides properties for either supercritical CO 2 (S-CO 2)more » or supercritical water (SCW). The Homogeneous Equilibrium Model (HEM) was employed to model the two-phase flow. HEM assumes two phases have same velocity, pressure, and temperature, making it only applicable for the dilute dispersed two-phase flow situation. Three example geometries, including orifices, labyrinth seals, and valves, were used to validate this methodology with experimental data. For the first geometry, S-CO 2 and SCW flowing through orifices were simulated and compared with experimental data. The maximum difference between the mass flow rate predictions and experimental measurements is less than 5%. This is a significant improvement as previous works can only guarantee 10% error. In this research, several efforts were made to help this improvement. First, an accurate real fluid module was used to provide properties. Second, the upstream condition was determined by pressure and density, which determines supercritical states more precise than using pressure and temperature. For the second geometry, the flow through labyrinth seals was studied. After a successful validation, parametric studies were performed to study geometric effects on the leakage rate. Based on these parametric studies, an optimum design strategy for the see-through labyrinth seals was proposed. A stepped labyrinth seal, which mimics the behavior of the labyrinth seal used in the Sandia National Laboratory (SNL) S-CO 2 Brayton cycle, was also tested in the experiment along with simulations performed. The rest of this study demonstrates the difference of valves' behavior under supercritical fluid and normal fluid conditions. A small-scale valve was tested in the experiment facility using S-CO 2. Different percentages of opening valves were tested, and the measured mass flow rate agreed with simulation predictions. Two transients from a real S-CO 2 Brayton cycle design provided the data for valve selection. The selected valve was studied using numerical simulation, as experimental data is not available.« less
DOE Office of Scientific and Technical Information (OSTI.GOV)
Arroyo, R.; Rebollo, L.
1993-06-01
This document presents the comparison between the simulation results and the plant measurements of a real event that took place in JOSE CABRERA nuclear power plant in August 30th, 1984. The event was originated by the total, continuous and inadverted opening of the pressurizer spray valve PCV-400A. JOSE CABRERA power plant is a single loop Westinghouse PWR belonging to UNION ELECTRICA FENOSA, S.A. (UNION FENOSA), an Spanish utility which participates in the International Code Assessment and Applications Program (ICAP) as a member of UNIDAD ELECTRICA, S.A. (UNESA). This is the second of its two contributions to the Program: the firstmore » one was an application case and this is an assessment one. The simulation has been performed using the RELAP5/MOD2 cycle 36.04 code, running on a CDC CYBER 180/830 computer under NOS 2.5 operating system. The main phenomena have been calculated correctly and some conclusions about the 3D characteristics of the condensation due to the spray and its simulation with a 1D tool have been got.« less
Barresi, Antonello A; Rasetto, Valeria; Marchisio, Daniele L
2018-05-15
This manuscript shows how computational models, mainly based on Computational Fluid Dynamics (CFD), can be used to simulate different parts of an industrial freeze-drying equipment and to properly design them; in particular, the freeze-dryer chamber and the duct connecting the chamber with the condenser, with the valves and vanes eventually present are analysed in this work. In Part 1, it will be shown how CFD can be employed to improve specific designs, to perform geometry optimization, to evaluate different design choices and how it is useful to evaluate the effect on product drying and batch variance. Such an approach allows an in-depth process understanding and assessment of the critical aspects of lyophilisation. This can be done by running either steady-state or transient simulations with imposed sublimation rates or with multi-scale approaches. This methodology will be demonstrated on freeze-drying equipment of different sizes, investigating the influence of the equipment geometry and shelf inter-distance. The effect of valve type (butterfly and mushroom) and shape on duct conductance and critical flow conditions will be instead investigated in Part 2. Copyright © 2018. Published by Elsevier B.V.
Overpressure and Fluid Diffusion Causing Non-hydrological Transient GNSS Displacements
NASA Astrophysics Data System (ADS)
Rossi, Giuliana; Fabris, Paolo; Zuliani, David
2017-11-01
In this work, global navigation satellite system (GNSS) observations from the northern tip of the Adria microplate are analysed to differentiate non-periodic (transient) tectonic signals from other deviations from the linear trends primarily due to hydrological loading effects. We tested a recently proposed hypothesis that a porosity wave generated by fault-valve mechanisms in a seismogenic fault in the Bovec basin (western Slovenia) propagated throughout the surrounding region. After excluding potential spatially correlated common-mode errors in the considered time series, we investigated the relationship between the GNSS observations and periodic hydrological loading variations. The tests demonstrated that subtracting the hydrological term was effective at the global scale and that the frequency band of the transient signal (1.5 < T < 3.5 years) was not correlated with hydrological effects at the local scale (within a few kilometres of the station). Next, the results of previous works are used to calculate the permeability values and pore-pressure state at the source of the transient signal. The permeability values for the four main rock formations in the region are consistent with independent observations for similar lithotypes. The ratio between the effective stress and lithostatic load for different vertical profiles in the Bovec area indicated a state of overpressure, with pore-pressure close to the value of the lithostatic load. Thus, our results help define a scenario in which the porosity wave could have originated. Indeed, the formation of the domains of interconnected fractures, such as during the formation of a porosity wave, increases the permeability values, thereby relieving overpressure and restoring a state of equilibrium.
Overpressure and Fluid Diffusion Causing Non-hydrological Transient GNSS Displacements
NASA Astrophysics Data System (ADS)
Rossi, Giuliana; Fabris, Paolo; Zuliani, David
2018-05-01
In this work, global navigation satellite system (GNSS) observations from the northern tip of the Adria microplate are analysed to differentiate non-periodic (transient) tectonic signals from other deviations from the linear trends primarily due to hydrological loading effects. We tested a recently proposed hypothesis that a porosity wave generated by fault-valve mechanisms in a seismogenic fault in the Bovec basin (western Slovenia) propagated throughout the surrounding region. After excluding potential spatially correlated common-mode errors in the considered time series, we investigated the relationship between the GNSS observations and periodic hydrological loading variations. The tests demonstrated that subtracting the hydrological term was effective at the global scale and that the frequency band of the transient signal (1.5 < T < 3.5 years) was not correlated with hydrological effects at the local scale (within a few kilometres of the station). Next, the results of previous works are used to calculate the permeability values and pore-pressure state at the source of the transient signal. The permeability values for the four main rock formations in the region are consistent with independent observations for similar lithotypes. The ratio between the effective stress and lithostatic load for different vertical profiles in the Bovec area indicated a state of overpressure, with pore-pressure close to the value of the lithostatic load. Thus, our results help define a scenario in which the porosity wave could have originated. Indeed, the formation of the domains of interconnected fractures, such as during the formation of a porosity wave, increases the permeability values, thereby relieving overpressure and restoring a state of equilibrium.
Pretest analysis document for Test S-NH-2
DOE Office of Scientific and Technical Information (OSTI.GOV)
Streit, J.E.; Owca, W.A.
This report documents the pretest analysis calculation completed with the RELAP5/MOD2/CY3601 code for Semiscale MOD-2C Test S-NH-2. The test will simulate the transient that results from the shear in a small diameter penetration of a cold leg, equivalent to 2.1% of the cold leg flow area. The high pressure injection system is assumed to be inoperative throughout the transient. The recovery procedure consists of latching open both steam generator atmospheric dump valves, supplying both steam generators with auxiliary feedwater system is assumed to be partially inoperative so the auxiliary feedwater flow is degraded. Recovery will be initiated upon a peakmore » cladding temperature of 811/sup 0/K (1000/sup 0/F). The test will be terminated when primary pressure has been reduced to the low pressure injection system setpoint of 1.38 MPa (200 psia). The calculated results indicate that the test objectives can be achieved and the proposed test scenario poses no threat to personnel or to plant integrity. 7 refs., 16 figs., 2 tabs.« less
Passive containment cooling system with drywell pressure regulation for boiling water reactor
Hill, Paul R.
1994-01-01
A boiling water reactor having a regulating valve for placing the wetwell in flow communication with an intake duct of the passive containment cooling system. This subsystem can be adjusted to maintain the drywell pressure at (or slightly below or above) wetwell pressure after the initial reactor blowdown transient is over. This addition to the PCCS design has the benefit of eliminating or minimizing steam leakage from the drywell to the wetwell in the longer-term post-LOCA time period and also minimizes the temperature difference between drywell and wetwell. This in turn reduces the rate of long-term pressure buildup of the containment, thereby extending the time to reach the design pressure limit.
Study of cryogenic propellant systems for loading the space shuttle
NASA Technical Reports Server (NTRS)
Voth, R. O.; Steward, W. G.; Hall, W. J.
1974-01-01
Computer programs were written to model the liquid oxygen loading system for the space shuttle. The programs allow selection of input data through graphic displays which schematically depict the part of the system being modeled. The computed output is also displayed in the form of graphs and printed messages. Any one of six computation options may be selected. The first four of these pertain to thermal stresses, pressure surges, cooldown times, flow rates and pressures during cooldown. Options five and six deal with possible water hammer effects due to closing of valves, steady flow and transient response to changes in operating conditions after cooldown. Procedures are given for operation of the graphic display unit and minicomputer.
Zhou, Nana; Yang, Chen; Tucker, David
2015-02-01
Thermal management in the fuel cell component of a direct fired solid oxide fuel cell gas turbine (SOFC/GT) hybrid power system can be improved by effective management and control of the cathode airflow. The disturbances of the cathode airflow were accomplished by diverting air around the fuel cell system through the manipulation of a hot-air bypass valve in open loop experiments, using a hardware-based simulation facility designed and built by the U.S. Department of Energy, National Energy Technology Laboratory (NETL). The dynamic responses of the fuel cell component and hardware component of the hybrid system were studied in this paper.
A NARX damper model for virtual tuning of automotive suspension systems with high-frequency loading
NASA Astrophysics Data System (ADS)
Alghafir, M. N.; Dunne, J. F.
2012-02-01
A computationally efficient NARX-type neural network model is developed to characterise highly nonlinear frequency-dependent thermally sensitive hydraulic dampers for use in the virtual tuning of passive suspension systems with high-frequency loading. Three input variables are chosen to account for high-frequency kinematics and temperature variations arising from continuous vehicle operation over non-smooth surfaces such as stone-covered streets, rough or off-road conditions. Two additional input variables are chosen to represent tuneable valve parameters. To assist in the development of the NARX model, a highly accurate but computationally excessive physical damper model [originally proposed by S. Duym and K. Reybrouck, Physical characterization of non-linear shock absorber dynamics, Eur. J. Mech. Eng. M 43(4) (1998), pp. 181-188] is extended to allow for high-frequency input kinematics. Experimental verification of this extended version uses measured damper data obtained from an industrial damper test machine under near-isothermal conditions for fixed valve settings, with input kinematics corresponding to harmonic and random road profiles. The extended model is then used only for simulating data for training and testing the NARX model with specified temperature profiles and different valve parameters, both in isolation and within quarter-car vehicle simulations. A heat generation and dissipation model is also developed and experimentally verified for use within the simulations. Virtual tuning using the quarter-car simulation model then exploits the NARX damper to achieve a compromise between ride and handling under transient thermal conditions with harmonic and random road profiles. For quarter-car simulations, the paper shows that a single tuneable NARX damper makes virtual tuning computationally very attractive.
Protasiewicz, Marcin; Gajek, Jacek; Mysiak, Andrzej
2013-01-01
We report a case of a 65-year-old female patient after replacement of aortic and mitral valve with mechanical prostheses and implantation of a pacemaker hospitalized in our clinic due to acute coronary syndrome complicated with cardiac arrest due to ventricular fibrillation. The electrocardiogram performed on admission showed signs of myocardial infarction with concomitant ventricular pacing. After successful resuscitation the coronary angiography was performed, which showed occlusion of the left circumflex artery (LCx) by thrombus. On the basis of intravascular ultrasound imaging the presence of vulnerable plaque, parietal thrombus and dissection of LCx were excluded. It suggested that occlusion of the LCx resulted from its embolism by left-sided heart thrombus due to subtherapeutic oral anticoagulation. In this case suboptimal anticoagulation was partially iatrogenic. Two weeks before the patient had been given vitamin K intravenously due to indeterminable international normalized ratio (INR) level, which caused transient resistance to oral anticoagulants. This case report illustrates tragic difficulties in the treatment with vitamin K antagonists, which concern as many as 2/3 of anticoagulated patients. These troubles contributed to the search for new, more efficient and safer anticoagulants. There are two classes of new oral anticoagulant drugs, which do not require monitoring of coagulation: direct thrombin inhibitors (e.g. dabigatran) and factor Xa inhibitors (e.g. rivaroxaban). In spite of their proven efficacy in the prevention of ischaemic stroke related to atrial fibrillation and prevention or treatment of deep vein thrombosis and pulmonary embolism, the use of new oral anticoagulants for the treatment of patients with mechanical valve prostheses needs further research. PMID:24570697
Protasiewicz, Marcin; Rojek, Aleksandra; Gajek, Jacek; Mysiak, Andrzej
2013-01-01
We report a case of a 65-year-old female patient after replacement of aortic and mitral valve with mechanical prostheses and implantation of a pacemaker hospitalized in our clinic due to acute coronary syndrome complicated with cardiac arrest due to ventricular fibrillation. The electrocardiogram performed on admission showed signs of myocardial infarction with concomitant ventricular pacing. After successful resuscitation the coronary angiography was performed, which showed occlusion of the left circumflex artery (LCx) by thrombus. On the basis of intravascular ultrasound imaging the presence of vulnerable plaque, parietal thrombus and dissection of LCx were excluded. It suggested that occlusion of the LCx resulted from its embolism by left-sided heart thrombus due to subtherapeutic oral anticoagulation. In this case suboptimal anticoagulation was partially iatrogenic. Two weeks before the patient had been given vitamin K intravenously due to indeterminable international normalized ratio (INR) level, which caused transient resistance to oral anticoagulants. This case report illustrates tragic difficulties in the treatment with vitamin K antagonists, which concern as many as 2/3 of anticoagulated patients. These troubles contributed to the search for new, more efficient and safer anticoagulants. There are two classes of new oral anticoagulant drugs, which do not require monitoring of coagulation: direct thrombin inhibitors (e.g. dabigatran) and factor Xa inhibitors (e.g. rivaroxaban). In spite of their proven efficacy in the prevention of ischaemic stroke related to atrial fibrillation and prevention or treatment of deep vein thrombosis and pulmonary embolism, the use of new oral anticoagulants for the treatment of patients with mechanical valve prostheses needs further research.
49 CFR 236.383 - Valve locks, valves, and valve magnets.
Code of Federal Regulations, 2010 CFR
2010-10-01
... 49 Transportation 4 2010-10-01 2010-10-01 false Valve locks, valves, and valve magnets. 236.383... Inspection and Tests § 236.383 Valve locks, valves, and valve magnets. Valve locks on valves of the non-cut-off type shall be tested at least once every three months, and valves and valve magnets shall be...
Spray Bar Zero-Gravity Vent System for On-Orbit Liquid Hydrogen Storage
NASA Technical Reports Server (NTRS)
Hastings, L. J.; Flachbart, R. H.; Martin, J. J.; Hedayat, A.; Fazah, M.; Lak, T.; Nguyen, H.; Bailey, J. W.
2003-01-01
During zero-gravity orbital cryogenic propulsion operations, a thermodynamic vent system (TVS) concept is expected to maintain tank pressure control without propellant resettling. In this case, a longitudinal spray bar mixer system, coupled with a Joule-Thompson (J-T) valve and heat exchanger, was evaluated in a series of TVS tests using the 18 cu m multipurpose hydrogen test bed. Tests performed at fill levels of 90, 50, and 25 percent, coupled with heat tank leaks of about 20 and 50 W, successfully demonstrated tank pressure control within a 7-kPa band. Based on limited testing, the presence of helium constrained the energy exchange between the gaseous and liquid hydrogen (LH2) during the mixing cycles. A transient analytical model, formulated to characterize TVS performance, was used to correlate the test data. During self-pressurization cycles following tank lockup, the model predicted faster pressure rise rates than were measured; however, once the system entered the cyclic self-pressurization/mixing/venting operational mode, the modeled and measured data were quite similar. During a special test at the 25-percent fill level, the J-T valve was allowed to remain open and successfully reduced the bulk LH2 saturation pressure from 133 to 70 kPa in 188 min.
NASA Astrophysics Data System (ADS)
Zhang, Xiaoxi; Cheng, Yongguang; Xia, Linsheng; Yang, Jiandong
2016-11-01
This paper reports the preliminary progress in the CFD simulation of the reverse water-hammer induced by the collapse of a draft-tube cavity in a model pump-turbine during the runaway process. Firstly, the Fluent customized 1D-3D coupling model for hydraulic transients and the Schnerr & Sauer cavitation model for cavity development are introduced. Then, the methods are validated by simulating the benchmark reverse water-hammer in a long pipe caused by a valve instant closure. The simulated head history at the valve agrees well with the measured data in literature. After that, the more complicated reverse water-hammer in the draft-tube of a runaway model pump-turbine, which is installed in a model pumped-storage power plant, is simulated. The dynamic processes of a vapor cavity, from generation, expansion, shrink to collapse, are shown. After the cavity collapsed, a sudden increase of pressure can be evidently observed. The process is featured by a locally expending and collapsing vapor cavity that is around the runner cone, which is different from the conventional recognition of violent water- column separation. This work reveals the possibility for simulating the reverse water-hammer phenomenon in turbines by 3D CFD.
Passive containment cooling system with drywell pressure regulation for boiling water reactor
Hill, P.R.
1994-12-27
A boiling water reactor is described having a regulating valve for placing the wetwell in flow communication with an intake duct of the passive containment cooling system. This subsystem can be adjusted to maintain the drywell pressure at (or slightly below or above) wetwell pressure after the initial reactor blowdown transient is over. This addition to the PCCS design has the benefit of eliminating or minimizing steam leakage from the drywell to the wetwell in the longer-term post-LOCA time period and also minimizes the temperature difference between drywell and wetwell. This in turn reduces the rate of long-term pressure buildup of the containment, thereby extending the time to reach the design pressure limit. 4 figures.
Aortic or Mitral Valve Replacement With the Biocor and Biocor Supra
2017-04-26
Aortic Valve Insufficiency; Aortic Valve Regurgitation; Aortic Valve Stenosis; Aortic Valve Incompetence; Mitral Valve Insufficiency; Mitral Valve Regurgitation; Mitral Valve Stenosis; Mitral Valve Incompetence
NASA Astrophysics Data System (ADS)
Yusop, Hanafi M.; Ghazali, M. F.; Yusof, M. F. M.; Remli, M. A. Pi; Kamarulzaman, M. H.
2017-10-01
In a recent study, the analysis of pressure transient signals could be seen as an accurate and low-cost method for leak and feature detection in water distribution systems. Transient phenomena occurs due to sudden changes in the fluid’s propagation in pipelines system caused by rapid pressure and flow fluctuation due to events such as closing and opening valves rapidly or through pump failure. In this paper, the feasibility of the Hilbert-Huang transform (HHT) method/technique in analysing the pressure transient signals in presented and discussed. HHT is a way to decompose a signal into intrinsic mode functions (IMF). However, the advantage of HHT is its difficulty in selecting the suitable IMF for the next data postprocessing method which is Hilbert Transform (HT). This paper reveals that utilizing the application of an integrated kurtosis-based algorithm for a z-filter technique (I-Kaz) to kurtosis ratio (I-Kaz-Kurtosis) allows/contributes to/leads to automatic selection of the IMF that should be used. This technique is demonstrated on a 57.90-meter medium high-density polyethylene (MDPE) pipe installed with a single artificial leak. The analysis results using the I-Kaz-kurtosis ratio revealed/confirmed that the method can be used as an automatic selection of the IMF although the noise level ratio of the signal is low. Therefore, the I-Kaz-kurtosis ratio method is recommended as a means to implement an automatic selection technique of the IMF for HHT analysis.
NASA Astrophysics Data System (ADS)
Yusop, Hanafi M.; Ghazali, M. F.; Yusof, M. F. M.; PiRemli, M. A.; Karollah, B.; Rusman
2017-10-01
Pressure transient signal occurred due to sudden changes in fluid propagation filled in pipelines system, which is caused by rapid pressure and flow fluctuation in a system, such as closing and opening valve rapidly. The application of Hilbert-Huang Transform (HHT) as the method to analyse the pressure transient signal utilised in this research. However, this method has the difficulty in selecting the suitable IMF for the further post-processing, which is Hilbert Transform (HT). This paper proposed the implementation of Integrated Kurtosis-based Algorithm for z-filter Technique (I-kaz) to kurtosis ratio (I-kaz-Kurtosis) for that allows automatic selection of intrinsic mode function (IMF) that’s should be used. This work demonstrated the synthetic pressure transient signal generates using transmission line modelling (TLM) in order to test the effectiveness of I-kaz as the autonomous selection of intrinsic mode function (IMF). A straight fluid network was designed using TLM fixing with higher resistance at some point act as a leak and connecting to the pipe feature (junction, pipefitting or blockage). The analysis results using I-kaz-kurtosis ratio revealed that the method can be utilised as an automatic selection of intrinsic mode function (IMF) although the noise level ratio of the signal is lower. I-kaz-kurtosis ratio is recommended and advised to be implemented as automatic selection of intrinsic mode function (IMF) through HHT analysis.
Wipper, Sabine; Lohrenz, Christina; Ahlbrecht, Oliver; Carpenter, Sebastian W; Tsilimparis, Nikolaos; Kersten, Jan Felix; Detter, Christian; Debus, Eike S; Kölbel, Tilo
2015-06-01
To compare the technical feasibility and hemodynamic alterations during antegrade transcardiac access routes vs conventional transfemoral access (TFA) for endovascular treatment of the ascending aorta in a porcine model. Antegrade transseptal access (TSA), transapical access (TAA), and TFA were used for implantation of custom-made endografts into the ascending aorta under fluoroscopy (6 pigs each). Hemodynamic parameters, myocardial and cerebral blood flow, and carotid artery blood flow were evaluated during baseline (T1), sheath advancement (T2), after sheath retraction (T3), and after endograft deployment (T4). Endograft deployment was feasible in all animals; all coronary arteries remained patent. Hemodynamic parameters were comparable in all 3 study groups during all measurements. During T2, transient hemodynamic alteration occurred in all groups, with transient severe valve insufficiency in TSA and TAA reflected by the higher pulmonary to mean arterial pressure ratio (p<0.05) as compared with TFA. Values stabilized again at T3 and remained stable until T4. The innominate artery was partially occluded in 4 (TSA), 3 (TAA), and 5 (TFA) animals. There was no deterioration of myocardial or cerebral perfusion during the procedures. Endograft deployment and fluoroscopy times during TAA were shorter than in TSA and TFA. TSA, TFA, and TAA to the ascending aorta are feasible for endograft delivery to the ascending aorta in a porcine model. Transient hemodynamic instability in TSA and TAA recovered to near preoperative values. TAA appeared technically easier. © The Author(s) 2015.
Nuclear radiation actuated valve
Christiansen, David W.; Schively, Dixon P.
1985-01-01
A nuclear radiation actuated valve for a nuclear reactor. The valve has a valve first part (such as a valve rod with piston) and a valve second part (such as a valve tube surrounding the valve rod, with the valve tube having side slots surrounding the piston). Both valve parts have known nuclear radiation swelling characteristics. The valve's first part is positioned to receive nuclear radiation from the nuclear reactor's fuel region. The valve's second part is positioned so that its nuclear radiation induced swelling is different from that of the valve's first part. The valve's second part also is positioned so that the valve's first and second parts create a valve orifice which changes in size due to the different nuclear radiation caused swelling of the valve's first part compared to the valve's second part. The valve may be used in a nuclear reactor's core coolant system.
Flex Fuel Optimized SI and HCCI Engine
DOE Office of Scientific and Technical Information (OSTI.GOV)
Zhu, Guoming; Schock, Harold; Yang, Xiaojian
The central objective of the proposed work is to demonstrate an HCCI (homogeneous charge compression ignition) capable SI (spark ignited) engine that is capable of fast and smooth mode transition between SI and HCCI combustion modes. The model-based control technique was used to develop and validate the proposed control strategy for the fast and smooth combustion mode transition based upon the developed control-oriented engine; and an HCCI capable SI engine was designed and constructed using production ready two-step valve-train with electrical variable valve timing actuating system. Finally, smooth combustion mode transition was demonstrated on a metal engine within eight enginemore » cycles. The Chrysler turbocharged 2.0L I4 direct injection engine was selected as the base engine for the project and the engine was modified to fit the two-step valve with electrical variable valve timing actuating system. To develop the model-based control strategy for stable HCCI combustion and smooth combustion mode transition between SI and HCCI combustion, a control-oriented real-time engine model was developed and implemented into the MSU HIL (hardware-in-the-loop) simulation environment. The developed model was used to study the engine actuating system requirement for the smooth and fast combustion mode transition and to develop the proposed mode transition control strategy. Finally, a single cylinder optical engine was designed and fabricated for studying the HCCI combustion characteristics. Optical engine combustion tests were conducted in both SI and HCCI combustion modes and the test results were used to calibrate the developed control-oriented engine model. Intensive GT-Power simulations were conducted to determine the optimal valve lift (high and low) and the cam phasing range. Delphi was selected to be the supplier for the two-step valve-train and Denso to be the electrical variable valve timing system supplier. A test bench was constructed to develop control strategies for the electrical variable valve timing (VVT) actuating system and satisfactory electrical VVT responses were obtained. Target engine control system was designed and fabricated at MSU for both single-cylinder optical and multi-cylinder metal engines. Finally, the developed control-oriented engine model was successfully implemented into the HIL simulation environment. The Chrysler 2.0L I4 DI engine was modified to fit the two-step vale with electrical variable valve timing actuating system. A used prototype engine was used as the base engine and the cylinder head was modified for the two-step valve with electrical VVT actuating system. Engine validation tests indicated that cylinder #3 has very high blow-by and it cannot be reduced with new pistons and rings. Due to the time constraint, it was decided to convert the four-cylinder engine into a single cylinder engine by blocking both intake and exhaust ports of the unused cylinders. The model-based combustion mode transition control algorithm was developed in the MSU HIL simulation environment and the Simulink based control strategy was implemented into the target engine controller. With both single-cylinder metal engine and control strategy ready, stable HCCI combustion was achived with COV of 2.1% Motoring tests were conducted to validate the actuator transient operations including valve lift, electrical variable valve timing, electronic throttle, multiple spark and injection controls. After the actuator operations were confirmed, 15-cycle smooth combustion mode transition from SI to HCCI combustion was achieved; and fast 8-cycle smooth combustion mode transition followed. With a fast electrical variable valve timing actuator, the number of engine cycles required for mode transition can be reduced down to five. It was also found that the combustion mode transition is sensitive to the charge air and engine coolant temperatures and regulating the corresponding temperatures to the target levels during the combustion mode transition is the key for a smooth combustion mode transition. As a summary, the proposed combustion mode transition strategy using the hybrid combustion mode that starts with the SI combustion and ends with the HCCI combustion was experimentally validated on a metal engine. The proposed model-based control approach made it possible to complete the SI-HCCI combustion mode transition within eight engine cycles utilizing the well controlled hybrid combustion mode. Without intensive control-oriented engine modeling and HIL simulation study of using the hybrid combustion mode during the mode transition, it would be impossible to validate the proposed combustion mode transition strategy in a very short period.« less
Design and characterization of a cough simulator.
Zhang, Bo; Zhu, Chao; Ji, Zhiming; Lin, Chao-Hsin
2017-02-23
Expiratory droplets from human coughing have always been considered as potential carriers of pathogens, responsible for respiratory infectious disease transmission. To study the transmission of disease by human coughing, a transient repeatable cough simulator has been designed and built. Cough droplets are generated by different mechanisms, such as the breaking of mucus, condensation and high-speed atomization from different depths of the respiratory tract. These mechanisms in coughing produce droplets of different sizes, represented by a bimodal distribution of 'fine' and 'coarse' droplets. A cough simulator is hence designed to generate transient sprays with such bimodal characteristics. It consists of a pressurized gas tank, a nebulizer and an ejector, connected in series, which are controlled by computerized solenoid valves. The bimodal droplet size distribution is characterized for the coarse droplets and fine droplets, by fibrous collection and laser diffraction, respectively. The measured size distributions of coarse and fine droplets are reasonably represented by the Rosin-Rammler and log-normal distributions in probability density function, which leads to a bimodal distribution. To assess the hydrodynamic consequences of coughing including droplet vaporization and polydispersion, a Lagrangian model of droplet trajectories is established, with its ambient flow field predetermined from a computational fluid dynamics simulation.
Pressurization, Pneumatic, and Vent Subsystems of the X-34 Main Propulsion System
NASA Technical Reports Server (NTRS)
Hedayat, A.; Steadman, T. E.; Brown, T. M.; Knight, K. C.; White, C. E., Jr.; Champion, R. H., Jr.
1998-01-01
In pressurization systems, regulators and orifices are use to control the flow of the pressurant. For the X-34 Main Propulsion System, three pressurization subsystem design configuration options were considered. In the first option, regulators were used while in the other options, orifices were considered. In each design option, the vent/relief system must be capable of relieving the pressurant flow without allowing the tank pressure to rise above proof, therefore, impacts on the propellant tank vent system were investigated and a trade study of the pressurization system was conducted. The analysis indicated that design option using regulators poses least risk. Then, a detailed transient thermal/fluid analysis of the recommended pressurization system was performed. Helium usage, thermodynamic conditions, and overpressurization of each propellant tank were evaluated. The pneumatic and purge subsystem is used for pneumatic valve actuation, Inter-Propellant Seal purges, Engine Spin Start, and engine purges at the required interface pressures, A transient analysis of the pneumatic and purge subsystem provided helium usage and flow rates to Inter-Propellant Seal and engine interfaces. Fill analysis of the helium bottles of pressurization and pneumatic subsystems during ground operation was performed. The required fill time and the stored
Materials-Enabled High-Efficiency (MEHE) Heavy-Duty Diesel Engines
DOE Office of Scientific and Technical Information (OSTI.GOV)
Kass, M.; Veliz, M.
2011-09-30
The purpose of this Cooperative Research and Development Agreement (CRADA) between UTBattelle, Inc. and Caterpillar, Inc. was to improve diesel engine efficiency by incorporating advanced materials to enable higher combustion pressures and temperatures necessary for improved combustion. The project scope also included novel materials for use in advanced components and designs associated with waste-heat recovery and other concepts for improved thermal efficiency. Caterpillar initially provided ORNL with a 2004 Tier 2 C15 ACERT diesel engine (designed for on-highway use) and two 600 hp motoring dynamometers. The first year of the CRADA effort was focused on establishing a heavy-duty experimental enginemore » research cell. First year activities included procuring, installing and commissioning the cell infrastructure. Infrastructure components consisted of intake air handling system, water tower, exhaust handling system, and cell air conditioning. Other necessary infrastructure items included the fuel delivery system and bottled gas handling to support the analytical instrumentation. The second year of the CRADA focused on commissioning the dynamometer system to enable engine experimentation. In addition to the requirements associated with the dynamometer controller, the electrical system needed a power factor correction system to maintain continuity with the electrical grid. During the second year the engine was instrumented and baseline operated to confirm performance and commission the dynamometer. The engine performance was mapped and modeled according to requirements provided by Caterpillar. This activity was further supported by a Work-for-Others project from Caterpillar to evaluate a proprietary modeling system. A second Work-for-Others activity was performed to evaluate a novel turbocharger design. This project was highly successful and may lead to new turbocharger designs for Caterpillar heavy-duty diesel engines. During the third (and final) year of the CRADA, a novel valve material was evaluated to assess high temperature performance and durability. A series of prototype valves, composed of a unique nickel-alloy was placed in the engine head. The engine was aggressively operated using a transient test cycle for 200 hours. The valve recession was periodically measured to determine valve performance. Upon completion of the test the valves were removed and returned to Caterpillar for additional assessment. Industrial in-kind support was available throughout the project period. Review of the status and research results were carried out on a regular basis (meetings and telecons) which included direction for future work activities. A significant portion of the industrial support was in the form of information exchange and technical consultation.« less
Conradi, Lenard; Kloth, Benjamin; Seiffert, Moritz; Schirmer, Johannes; Koschyk, Dietmar; Blankenberg, Stefan; Reichenspurner, Hermann; Diemert, Patrick; Treede, Hendrik
2014-12-01
Recently, the feasibility of valve-in-valve procedures using current first-generation transcatheter heart valves (THV) in cases of structural valve degeneration has been reported as an alternative to conventional open repeat valve replacement. By design, certain biological valve xenografts carry a high risk of coronary ostia occlusion due to lateral displacement of leaflets after valve-in-valve procedures. In the present report we aimed to prove feasibility and safety of transapical valve-in-valve implantation of the JenaValve THV in two cases of degenerated Mitroflow bioprostheses. We herein report two cases of successful transapical valve-in-valve procedures using a JenaValve THV implanted in Sorin Mitroflow bioprostheses for structural valve degeneration. Both patients were alive and in good clinical condition at 30 days from the procedure. However, increased transvalvular gradients were noted in both cases. Transcatheter valve-in-valve implantation of a JenaValve THV is a valid alternative for patients with degenerated Mitroflow bioprostheses of sufficient size and in the presence of short distances to the coronary ostia who are too ill for conventional repeat open heart surgery. Increased pressure gradients have to be expected and weighed against the disadvantages of other treatment options when planning such a procedure.
NASA Technical Reports Server (NTRS)
Wojciechowski, Bogdan V. (Inventor); Pegg, Robert J. (Inventor)
2003-01-01
A fast-acting valve includes an annular valve seat that defines an annular valve orifice between the edges of the annular valve seat, an annular valve plug sized to cover the valve orifice when the valve is closed, and a valve-plug holder for moving the annular valve plug on and off the annular valve seat. The use of an annular orifice reduces the characteristic distance between the edges of the valve seat. Rather than this distance being equal to the diameter of the orifice, as it is for a conventional circular orifice, the characteristic distance equals the distance between the inner and outer radii (for a circular annulus). The reduced characteristic distance greatly reduces the gap required between the annular valve plug and the annular valve seat for the valve to be fully open, thereby greatly reducing the required stroke and corresponding speed and acceleration of the annular valve plug. The use of a valve-plug holder that is under independent control to move the annular valve plug between its open and closed positions is important for achieving controllable fast operation of the valve.
Nuclear-radiation-actuated valve. [Patent application; for increasing coolant flow to blanket
Christiansen, D.W.; Schively, D.P.
1982-01-19
The present invention relates to a breeder reactor blanket fuel assembly coolant system valve which increases coolant flow to the blanket fuel assembly to minimize long-term temperature increases caused by fission of fissile fuel created from fertile fuel through operation of the breeder reactor. The valve has a valve first part (such as a valve rod with piston) and a valve second part (such as a valve tube surrounding the valve rod, with the valve tube having side slots surrounding the piston). Both valve parts have known nuclear radiation swelling characteristics. The valve's first part is positioned to receive nuclear radiation from the nuclear reactor's fuel region. The valve's second part is positioned so that its nuclear radiation induced swelling is different from that of the valve's first part. The valve's second part also is positioned so that the valve's first and second parts create a valve orifice which changes in size due to the different nuclear radiation caused swelling of the valve's first part compared to the valve's second part. The valve may be used in a nuclear reactor's core coolant system.
Slide valve apparatus for internal combustion engine
DOE Office of Scientific and Technical Information (OSTI.GOV)
Taylor, B.A.; McMahan, T.O.
This patent describes an internal combustion engine including a combustion cylinder having an opening at one end thereof, a piston mounted within the cylinder for coaxial reciprocable movement, a driven crankshaft, and a connecting rod connecting the crankshaft to the cylinder for linear reciprocable movement of the piston in response to the rotary movement of the crankshaft, a valve apparatus comprising: (a) a valve chamber extending longitudinally across and in fluid communication with the opening in the cylinder, (b) an intake valve plate having a longitudinal axis mounted within the valve chamber for slidable, reciprocable, longitudinal movement, (c) an exhaustmore » valve plate having a longitudinal axis mounted within the valve chamber alongside the intake valve plate for slidable, reciprocable, longitudinal movement and parallel to the longitudinal axis of the intake valve plate, (d) each of the valve plates having a plurality of longitudinally spaced valve ports therein, the valve ports comprising movable intake valve ports in the intake valve plate and movable exhaust valve ports in the exhaust valve plate, (e) the valve chamber comprising a planar wall on the opposite side of the valve plates from the cylinder opening and having a plurality of fixed valve ports therethrough. The fixed valve ports being equal in number and substantially equal in size and spacing as the movable intake and exhaust valve ports, whereby the movable intake valve ports are adapted to register with their corresponding fixed valve ports when the intake valve plate is in its intake operative position for opening fluid communication between the cylinder and the corresponding fixed valve ports.« less
Bioprosthetic Valve Fracture to Facilitate Transcatheter Valve-in-Valve Implantation.
Allen, Keith B; Chhatriwalla, Adnan K; Cohen, David J; Saxon, John T; Aggarwal, Sanjeev; Hart, Anthony; Baron, Suzanne; Davis, J Russell; Pak, Alex F; Dvir, Danny; Borkon, A Michael
2017-11-01
Valve-in-valve transcatheter aortic valve replacement is less effective in small surgical bioprostheses. We evaluated the feasibility of bioprosthetic valve fracture with a high-pressure balloon to facilitate valve-in-valve transcatheter aortic valve replacement. In vitro bench testing on aortic tissue valves was performed on 19-mm and 21-mm Mitroflow (Sorin, Milan, Italy), Magna and Magna Ease (Edwards Lifesciences, Irvine, CA), Trifecta and Biocor Epic (St. Jude Medical, Minneapolis, MN), and Hancock II and Mosaic (Medtronic, Minneapolis, MN). High-pressure balloons Tru Dilation, Atlas Gold, and Dorado (C.R. Bard, Murray Hill, NJ) were used to determine which valves could be fractured and at what pressure fracture occurred. Mitroflow, Magna, Magna Ease, Mosaic, and Biocor Epic surgical valves were successfully fractured using high-pressures balloon 1 mm larger than the labeled valve size whereas Trifecta and Hancock II surgical valves could not be fractured. Only the internal valve frame was fractured, and the sewing cuff was never disrupted. Manufacturer's rated burst pressures for balloons were exceeded, with fracture pressures ranging from 8 to 24 atmospheres depending on the surgical valve. Testing further demonstrated that fracture facilitated the expansion of previously constrained, underexpanded transcatheter valves (both balloon and self-expanding) to the manufacturer's recommended size. Bench testing demonstrates that the frame of most, but not all, bioprosthetic surgical aortic valves can be fractured using high-pressure balloons. The safety of bioprosthetic valve fracture to optimize valve-in-valve transcatheter aortic valve replacement in small surgical valves requires further clinical investigation. Copyright © 2017 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.
Advanced User Interface Capabilities for Application on Portable Computers
1992-02-01
0 060 iI 1 ........ ... r: switch_ e( pressure , switch ) r6 : pilot- valvecircuit(A, pilot valve) pilot_ valve_ circuit(B, pilot valve) r7...shutoff_valvecircwt(A, shutoff valve) shutoff_valve circuit(B, shutoff valve) r: pressure_ switch_ circuit(A, pressure switch ) pressure_ switch circuit(B... pressure switch ) r: indicator(A, pilot valve) indicator(B, pilot valve) indicator(A, shutoff valve) indicator(B, shutoff valve) indicator(A, pressure
Reul, Ross M.; Ramchandani, Mahesh K.; Reardon, Michael J.
2017-01-01
Surgical aortic valve replacement is the gold standard procedure to treat patients with severe, symptomatic aortic valve stenosis or insufficiency. Bioprosthetic valves are used for surgical aortic valve replacement with a much greater prevalence than mechanical valves. However, bioprosthetic valves may fail over time because of structural valve deterioration; this often requires intervention due to severe bioprosthetic valve stenosis or regurgitation or a combination of both. In select patients, transcatheter aortic valve replacement is an alternative to surgical aortic valve replacement. Transcatheter valve-in-valve (ViV) replacement is performed by implanting a transcatheter heart valve within a failing bioprosthetic valve. The transcatheter ViV operation is a less invasive procedure compared with reoperative surgical aortic valve replacement, but it has been associated with specific complications and requires extensive preoperative work-up and planning by the heart team. Data from experimental studies and analyses of results from clinical procedures have led to strategies to improve outcomes of these procedures. The type, size, and implant position of the transcatheter valve can be optimized for individual patients with knowledge of detailed dimensions of the surgical valve and radiographic and echocardiographic measurements of the patient's anatomy. Understanding the complexities of the ViV procedure can lead surgeons to make choices during the original surgical valve implantation that can make a future ViV operation more technically feasible years before it is required. PMID:29743998
Heart valve replacement with the Sorin tilting-disc prosthesis. A 10-year experience.
Milano, A; Bortolotti, U; Mazzucco, A; Mossuto, E; Testolin, L; Thiene, G; Gallucci, V
1992-02-01
From 1978 to 1988, 697 patients with a mean age of 48 +/- 11 years (range 5 to 75 years) received a Sorin tilting-disc prosthesis; 358 had had aortic valve replacement, 247 mitral valve replacement, and 92 mitral and aortic valve replacement. Operative mortality rates were 7.8%, 11.3%, and 10.8%, respectively, in the three groups. Cumulative duration of follow-up is 1650 patient-years for aortic valve replacement (maximum follow-up 11.4 years), 963 patient-years for mitral valve replacement (maximum follow-up 9.9 years) and 328 patient-years for mitral and aortic valve replacement (maximum follow-up 9.4 years). Actuarial survival at 9 years is 72% +/- 4% after mitral valve replacement, 70% +/- 3% after aortic valve replacement, and 50% +/- 12% after mitral and aortic valve replacement, and actuarial freedom from valve-related deaths is 97% +/- 2% after mitral valve replacement, 92% +/- 2% after aortic valve replacement, and 62% +/- 15% after mitral and aortic valve replacement. Thromboembolic events occurred in 21 patients with aortic valve replacement (1.3% +/- 0.2%/pt-yr), in 12 with mitral valve replacement (1.2% +/- 0.3% pt-yr), and in seven with mitral and aortic valve replacement (2.1% +/- 0.8%), with one case of prosthetic thrombosis in each group; actuarial freedom from thromboembolism at 9 years is 92% +/- 3% after mitral valve replacement, 91% +/- 3% after aortic valve replacement, and 74% +/- 16% after mitral and aortic valve replacement. Anticoagulant-related hemorrhage was observed in 15 patients after aortic valve replacement (0.9% +/- 0.2%/pt-yr), in 9 after mitral valve replacement (0.9% +/- 0.3%/pt-yr), and in 6 with mitral and aortic valve replacement (0.9% +/- 0.5%/pt-yr); actuarial freedom from this complication at 9 years is 94% +/- 2% after aortic valve replacement, 91% +/- 4% after mitral valve replacement, and 68% +/- 16% after mitral and aortic valve replacement. Actuarial freedom from reoperation at 9 years is 97% +/- 2% after mitral and aortic valve replacement, 92% +/- 4% after mitral valve replacement, and 89% +/- 3% after aortic valve replacement, with no cases of mechanical fracture. The Sorin valve has shown a satisfactory long-term overall performance, comparable with other mechanical prostheses, and an excellent durability that renders it a reliable heart valve substitute for the mitral and aortic positions.
Bergman, Ulf C.
1984-01-01
The safety valve contains a resilient gland to be held between a valve seat and a valve member and is secured to the valve member by a sleeve surrounding the end of the valve member adjacent to the valve seat. The sleeve is movable relative to the valve member through a limited axial distance and a gap exists between said valve member and said sleeve.
Valve assembly for internal combustion engine
DOE Office of Scientific and Technical Information (OSTI.GOV)
Wakeman, R.J.; Shea, S.F.
1989-09-05
This patent describes an improvement in a valve assembly for an internal combustion engine of the type including a valve having a valve stem, a valve guideway for mounting this valve for reciprocal strokes between opened and seated position, and spring means for biasing the valve into the seated position. The improvement comprising a valve spool of greater cross-sectional diameter as compared to the valve stem, and a valve spool guideway within which the valve spool is movable during the strokes of the valve, an upper surface of the valve spool and a portion of the spool guideway collectively establishingmore » a damper chamber which varies in volume during the valve strokes. a feed passage for introducing oil into the damper chamber, and a bleed passage for discharging oil from the damper chamber. The bleed passages each laterally opening into the valve spool guideway.« less
Kiffer, Micah S.; Tentarelli, Stephen Clyde
2016-02-09
Excess flow shutoff valve comprising a valve body, a valve plug, a partition, and an activation component where the valve plug, the partition, and activation component are disposed within the valve body. A suitable flow restriction is provided to create a pressure difference between the upstream end of the valve plug and the downstream end of the valve plug when fluid flows through the valve body. The pressure difference exceeds a target pressure difference needed to activate the activation component when fluid flow through the valve body is higher than a desired rate, and thereby closes the valve.
Method of Manufacturing Carbon Fiber Reinforced Carbon Composite Valves
NASA Technical Reports Server (NTRS)
Rivers, H. Kevin (Inventor); Ransone, Philip O. (Inventor); Northam, G. Burton (Inventor)
1998-01-01
A method for forming a carbon composite valve for internal combustion engines is discussed. The process includes the steps of braiding carbon fiber into a rope thereby forming a cylindrically shaped valve stem portion and continuing to braid said fiber while introducing into the braiding carbon fiber rope a carbon matrix plug having an outer surface in a net shape of a valve head thereby forming a valve head portion. The said carbon matrix plug acting as a mandrel over which said carbon fiber rope is braided, said carbon fiber rope and carbon matrix plug forming a valve head portion suitable for mating with a valve seat; cutting said braided carbon valve stem portion at one end to form a valve tip and cutting said braided carbon fiber after said valve head portion to form a valve face and thus provide a composite valve preform; and densifying said preform by embedding the braided carbon in a matrix of carbon to convert said valve stem portion to a valve stem and said valve head portion to a valve head thereby providing said composite valve.
Bradley, Scott M
2013-10-01
The contents of this article were presented in the session "Aortic insufficiency in the teenager" at the congenital parallel symposium of the 2013 Society of Thoracic Surgeons (STS) annual meeting. The accompanying articles detail the approaches of aortic valve repair and the Ross procedure.(1,2) The current article focuses on prosthetic valve replacement. For many young patients requiring aortic valve surgery, either aortic valve repair or a Ross procedure provides a good option. The advantages include avoidance of anticoagulation and potential for growth. In other patients, a prosthetic valve is an appropriate alternative. This article discusses the current state of knowledge regarding mechanical and bioprosthetic valve prostheses and their specific advantages relative to valve repair or a Ross procedure. In current practice, young patients requiring aortic valve surgery frequently undergo valve replacement with a prosthetic valve. In STS adult cardiac database, among patients ≤30 years of age undergoing aortic valve surgery, 34% had placement of a mechanical valve, 51% had placement of a bioprosthetic valve, 9% had aortic valve repair, and 2% had a Ross procedure. In the STS congenital database, among patients 12 to 30 years of age undergoing aortic valve surgery, 21% had placement of a mechanical valve, 18% had placement of a bioprosthetic valve, 30% had aortic valve repair, and 24% had a Ross procedure. In the future, the balance among these options may be altered by design improvements in prosthetic valves, alternatives to warfarin, the development of new patch materials for valve repair, and techniques to avoid Ross autograft failure.
Water hammer caused by closure of turbine safety spherical valves
NASA Astrophysics Data System (ADS)
Karadžić, U.; Bergant, A.; Vukoslavčević, P.
2010-08-01
This paper investigates water hammer effects caused by closure of spherical valves against the discharge. During the first phase of modernisation of Perućica high-head hydropower plant (HPP), Montenegro, safety spherical valves (inlet turbine valves) have been refurbished on the first two Pelton turbine units. The valve closure is controlled by the valve actuator (hydraulic servomotor). Because the torque acting on the valve body is dependent on flow conditions the valve closing time may vary significantly for different flow velocities (passive valve). For the passive valve the torques acting on the valve body should be considered in the valve model. The valve closing time results from numerical simulation. On the contrary, for the active valve the valve closing time is assumed prior to simulation. The spherical valve boundary condition is incorporated into the method of characteristics (MOC) algorithm. The staggered (diamond) grid in applying the MOC is used in this paper. The passive valve boundary condition is described by the water hammer equations, the valve equation that relates discharge to pressure head drop and the dynamic equation of the valve body motion (torque equation). The active valve boundary condition is described by the first two equations, respectively. Standard quasi-steady friction model is used for estimating friction losses in plant's tunnel and penstocks. Numerical results using both the active and the passive spherical valve models are compared with results of measurements. It has been found that the influence of flow conditions on the spherical valve closing time is minor for the cases considered. Computed and measured results agree reasonably well.
Garg, Aatish; Parashar, Akhil; Agarwal, Shikhar; Aksoy, Olcay; Hammadah, Muhammad; Poddar, Kanhaiya Lal; Puri, Rishi; Svensson, Lars G; Krishnaswamy, Amar; Tuzcu, E Murat; Kapadia, Samir R
2015-02-15
The SAPIEN-XT is a newer generation balloon-expandable valve created of cobalt chromium frame, as opposed to the stainless steel frame used in the older generation SAPIEN valve. We sought to determine if there was difference in acute recoil between the two valves. All patients who underwent transfemoral-transcatheter aortic valve replacement using the SAPIEN-XT valve at the Cleveland Clinic were included. Recoil was measured using biplane cine-angiographic image analysis of valve deployment. Acute recoil was defined as [(valve diameter at maximal balloon inflation) - (valve diameter after deflation)]/valve diameter at maximal balloon inflation (reported as percentage). Patients undergoing SAPIEN valve implantation were used as the comparison group. Among the 23 mm valves, the mean (standard deviation-SD) acute recoil was 2.77% (1.14) for the SAPIEN valve as compared to 3.75% (1.52) for the SAPIEN XT valve (P = 0.04). Among the 26 mm valves, the mean (SD) acute recoil was 2.85% (1.4) for the SAPIEN valve as compared to 4.32% (1.63) for the SAPIEN XT valve (P = 0.01). Multivariable linear regression analysis demonstrated significantly greater adjusted recoil in the SAPIEN XT valves as compared to the SAPIEN valves by 1.43% [(95% CI: 0.69-2.17), P < 0.001]. However, the residual peak gradient was less for SAPIEN XT compared to SAPIEN valves [18.86 mm Hg versus 23.53 mm Hg (P = 0.01)]. Additionally, no difference in paravalvular leak was noted between the two valve types (P = 0.78). The SAPIEN XT valves had significantly greater acute recoil after deployment compared to the SAPIEN valves. Implications of this difference in acute recoil on valve performance need to be investigated in future studies. © 2014 Wiley Periodicals, Inc.
Aydin, Unal; Gul, Mehmet; Aslan, Serkan; Akkaya, Emre; Yildirim, Aydin
2015-04-28
Transcatheter valve implantation is a novel interventional technique, which was developed as an alternative therapy for surgical aortic valve replacement in inoperable patients with severe aortic stenosis. Despite limited experience in using transcatheter valve implantation for mitral and aortic regurgitation, transapical transcatheter aortic valve implantation and valve-in-valve implantation for degenerated mitral valve bioprosthesis can be performed in high-risk patients who are not candidates for conventional replacement surgery. In this case, we present the simultaneous transcatheter valve implantation via transapical approach for both degenerated bioprosthetic mitral valve with severe regurgitation and pure severe aortic regurgitation.
Downhole surge valve for earth boring apparatus
DOE Office of Scientific and Technical Information (OSTI.GOV)
Lee, D.W.
1990-05-29
This patent describes a boring tool assembly having an underground percussion mole boring tool powered by a working fluid, the tool being driven through the earth by a rigid drill string pushed by a drilling frame, and a downhole valve assembly fixed between the downhole end of the drill string and the too, the improved downhole valve assembly. It comprises: a valve spool having an open first end, a closed second end and a peripheral sidewall, an axial bore extending partly through the valve spool from the open first end; a radial passage adjacent the closed second and of themore » valve spool, the radial passage extending radially from the valve spool axial bore through the valve spool peripheral sidewall; an axial groove in the peripheral sidewall of the valve spool; a valve body having a first end, a second end and a peripheral sidewall, an axial bore extending through the valve body, the valve spool extending through the valve body axial bore so that the second end of the valve body is adjacent the closed second end of the valve spool, the valve spool being axially moveable within the valve body axial bore; an axial slot; a free-floating key element; a valve housing; and seal means.« less
3T magnetic resonance imaging testing of externally programmable shunt valves
Zabramski, Joseph M.; Preul, Mark C.; Debbins, Josef; McCusker, Daniel J.
2012-01-01
Background: Exposure of externally programmable shunt-valves (EPS-valves) to magnetic resonance imaging (MRI) may lead to unexpected changes in shunt settings, or affect the ability to reprogram the valve. We undertook this study to examine the effect of exposure to a 3T MRI on a group of widely used EPS-valves. Methods: Evaluations were performed on first generation EPS-valves (those without a locking mechanism to prevent changes in shunt settings by external magnets other than the programmer) and second generation EPS-valves (those with a locking mechanisms). Fifteen new shunt-valves were divided into five groups of three identical valves each, and then exposed to a series of six simulated MRI scans. After each of the exposures, the valves were evaluated to determine if the valve settings had changed, and whether the valves could be reprogrammed. The study produced 18 evaluations for each line of shunt-valves. Results: Exposure of the first generation EPS-valves to a 3T magnetic field resulted in frequent changes in the valve settings; however, all valves retained their ability to be reprogrammed. Repeated exposure of the second generation EPS-valves has no effect on shunt valve settings, and all valves retained their ability to be interrogated and reprogrammed. Conclusions: Second generation EPS-valves with locking mechanisms can be safely exposed to repeated 3T MRI systems, without evidence that shunt settings will change. The exposure of the first generation EPS-valves to 3T MRI results in frequent changes in shunt settings that necessitate re-evaluation soon after MRI to avoid complications. PMID:22937481
40 CFR Appendix Viii to Part 85 - Vehicle and Engine Parameters and Specifications
Code of Federal Regulations, 2010 CFR
2010-07-01
... pressure. 3. Valves (intake and exhaust). a. Head diameter dimension. b. Valve lifter or actuator type and... diameter dimension. b. Valve lifter or actuator type and valve lash dimension. 5. Camshaft timing. a. Valve... dimension. b. Valve lifter or actuator type and valve lash dimension. 5. Camshaft timing. a. Valve opening...
40 CFR Appendix Viii to Part 85 - Vehicle and Engine Parameters and Specifications
Code of Federal Regulations, 2011 CFR
2011-07-01
... pressure. 3. Valves (intake and exhaust). a. Head diameter dimension. b. Valve lifter or actuator type and... diameter dimension. b. Valve lifter or actuator type and valve lash dimension. 5. Camshaft timing. a. Valve... dimension. b. Valve lifter or actuator type and valve lash dimension. 5. Camshaft timing. a. Valve opening...
Fast-acting valve and uses thereof
Meyer, J.A.
1980-05-16
A very fast acting valve capable of producing a very well-defined plug of gas suitable for filling a theta pinch vacuum vessel is given. The valve requires no springs, instead being stopped mainly by a nonlinear force. Thus, the valve is not subject to bouncing; and the ratio of the size of the valve housing to the size of the valve stem is smaller than it would be if springs were needed to stop the valve stem. Furthermore, the valve can be used for thousands of valve firings with no apparent valve damage.
Fast-acting valve and uses thereof
Meyer, James A.
1982-01-01
A very fast acting valve capable of producing a very well-defined plug of gas suitable for filling a theta pinch vacuum vessel is given. The valve requires no springs, instead being stopped mainly by a nonlinear force. Thus, the valve is not subject to bouncing; and the ratio of the size of the valve housing to the size of the valve stem is smaller than it would be if springs were needed to stop the valve stem. Furthermore, the valve can be used for thousands of valve firings with no apparent valve damage.
Brown, Stephen C; Cools, Bjorn; Gewillig, Marc
2016-09-01
Bioprosthetic valves degenerate over time. Transcatheter valve-in-valve procedures have become an attractive alternative to surgery. However, every valve increasingly diminishes the diameter of the valvar orifice. We report a 12-year-old female who had a previous transcatheter tricuspid valve-in-valve procedure; cracking the ring of a Carpentier Edwards Perimount valve by means of an ultrahigh pressure balloon allowed implantation of a further larger percutaneous valve. The advantage of this novel approach permits enlarging the inner valve diameter and may facilitate future interventions and prolong time to surgery. © 2016 Wiley Periodicals, Inc. © 2016 Wiley Periodicals, Inc.
All metal valve structure for gas systems
Baker, Ray W.; Pawlak, Donald A.; Ramey, Alford J.
1984-11-13
A valve assembly with a resilient metal seat member is disclosed for providing a gas-tight seal in a gas handling system. The valve assembly also includes a valve element for sealing against the valve seat member; and an actuating means for operating the valve element. The valve seat member is a one-piece stainless steel ring having a central valve port and peripheral mounting flange, and an annular corrugation in between. A groove between the first and second ridges serves as a flexure zone during operation of the valve member and thus provides the seating pressure between the inner ridge or valve seat and the valve element. The outer annular ridge has a diameter less than said valve element to limit the seating motion of the valve element, preventing non-elastic deformation of the seat member.
Hydraulic engine valve actuation system including independent feedback control
Marriott, Craig D
2013-06-04
A hydraulic valve actuation assembly may include a housing, a piston, a supply control valve, a closing control valve, and an opening control valve. The housing may define a first fluid chamber, a second fluid chamber, and a third fluid chamber. The piston may be axially secured to an engine valve and located within the first, second and third fluid chambers. The supply control valve may control a hydraulic fluid supply to the piston. The closing control valve may be located between the supply control valve and the second fluid chamber and may control fluid flow from the second fluid chamber to the supply control valve. The opening control valve may be located between the supply control valve and the second fluid chamber and may control fluid flow from the supply control valve to the second fluid chamber.
All-metal valve structure for gas systems
Baker, R.W.; Pawlak, D.A.; Ramey, A.J.
1982-06-10
A valve assembly with a resilient metal seat member is disclosed for providing a gas-tight seal in a gas handling system. The valve assembly also includes a valve element for sealing against the valve seat member; and an actuating means for operating the valve element. The valve seat member is a one-piece stainless steel ring having a central valve port and peripheral mounting flange, and an annular corrugation in between. A groove between the first and second ridges serves as a flexure zone during operation of the valve member and thus provides the seating pressure between the inner ridge or valve seat and the valve element. The outer annular ridge has a diameter less than said valve element to limit the seating motion of the valve element, preventing non-elastic deformation of the seat member.
Late complications in patients with Björk-Shiley and St. Jude Medical heart valve replacement.
Horstkotte, D; Körfer, R; Seipel, L; Bircks, W; Loogen, F
1983-09-01
Valve-related complications after Björk-Shiley mitral valve implantation (n = 475), aortic valve implantation (n = 424), or mitral-aortic valve implantation (n = 119) were compared with those after St. Jude Medical mitral valve replacement (n = 173), aortic valve replacement (n = 152), or mitral-aortic valve replacement (n = 69). All patients were placed on anticoagulant therapy with phenprocoumon early after operation. All patients had a comparable follow-up time of approximately 23 months, which showed that cumulative thromboembolic rates were significantly higher after St. Jude valve implantation than after Björk-Shiley valve implantation. Reoperations were necessary because of valve thrombosis (0.46%), perivalvular leakage (2.2%), or prosthetic valve endocarditis with perivalvular regurgitation (0.46%). One Björk-Shiley mitral valve prosthesis had to be replaced because of fracture of the outlet strut. Without significant intergroup differences, hemorrhage due to anticoagulant treatment was the most frequent complication. Thromboembolic complications were significantly more frequent after Björk-Shiley mitral, aortic, and double valve replacements than after St. Jude valve implantation. This may lead to consideration of changes in the prophylaxis of thrombus formations in the St. Jude valve, especially in aortic valve replacements, in patients with sinus rhythm.
Low energy high pressure miniature screw valve
Fischer, Gary J [Sandia Park, NM; Spletzer, Barry L [Albuquerque, NM
2006-12-12
A low energy high pressure screw valve having a valve body having an upper portion and a lower portion, said lower portion of said valve body defining an inlet flow passage and an outlet flow passage traversing said valve body to a valve seat, said upper portion of said valve body defining a cavity at said valve seat, a diaphragm restricting flow between said upper portion of said valve body and said lower portion, said diaphragm capable of engaging said valve seat to restrict fluid communication between said inlet passage and said outlet passage, a plunger within said cavity supporting said diaphragm, said plunger being capable of engaging said diaphragm with said valve seat at said inlet and outlet fluid passages, said plunger being in point contact with a drive screw having threads engaged with opposing threads within said upper portion of said valve body such engagement allowing motion of said drive screw within said valve body.
[Valvuloplasty with balloon catheter in biologic prosthesis. Reality or illusion].
Ledesma Velasco, M; Verdín Vázquez, R; Acosta Valdez, J L; Munayer Calderón, J; Salgado Escobar, J L; Arias Monroy, L; Flores Mendoza, J
1989-01-01
We performed catheter balloon valvuloplasty (CBV) on 8 stenotic operatively-excised bioprosthetic valves (2 Hancock and 6 Ionescu Shiley). Pathology of valves before CBV included degenerative changes: commissural fusion by mounds of calcific deposits (2 valves), fibrotic and focally calcified leaflets (7 valves) and stiff and thick valves (1 valve). Inflation of the balloon resulted in commissural splitting (2 valves), leaflet cracks and fractures (3 valves). Removal of the deflated balloon catheter was associated with debris dislodgement (3 valves). In one case the valve was unable to close with potential for acute regurgitation. Thus, CBV of bioprosthetic valves can split fused commissures by similar mechanisms as in native valves. CBV may fracture calcific deposits causing acute emboli. It can also disrupt the leaflets causing acute insufficiency. The findings suggest a limited role of CBV in the treatment of stenotic bioprosthetic valves in mitral and aortic position.
30. Engine controls and valve gear, looking aft on main ...
30. Engine controls and valve gear, looking aft on main (promenade) deck level. Threaded admission valve lift rods (two at immediate left of chronometer) permit adjustment of valve timing in lower and upper admission valves of cylinder (left rod controls lower valve, right rod upper valve). Valve rods are lifted by jaw-like "wipers" during operation. Exhaust valve lift rods and wipers are located to right of chronometer. Crank at extreme right drives valve wiper shaft when engaged to end of eccentric rod, shown under "Crank Indicator" dial. Pair of handles to immediate left of admission valve rods control condenser water valves; handles to right of exhaust valve rods control feedwater flow to boilers from pumps. Gauges indicate boiler pressure (left) and condenser vacuum (right); "Crank Indicator" on wall aids engineer in keeping engine crank off "dead-center" at stop so that engine may be easily restarted. - Steamboat TICONDEROGA, Shelburne Museum Route 7, Shelburne, Chittenden County, VT
Dukas, Jr., Stephen J.
1990-01-01
A bellows sealed plug valve includes a valve body having an inlet passage and an outlet passage, a valve chamber between the inlet and outlet passages. A valve plug has substantially the same shape as the valve chamber and is rotatably disposed therein. A shaft is movable linearly in response to a signal from a valve actuator. A bellows is sealingly disposed between the valve chamber and the valve actuator and means are located between the bellows and the valve plug for converting linear movement of the shaft connected to the valve actuator to rotational movement of the plug. Various means are disclosed including helical thread mechanism, clevis mechanism and rack and pinion mechanism, all for converting linear motion to rotational motion.
Mesofluidic two stage digital valve
Jansen, John F; Love, Lonnie J; Lind, Randall F; Richardson, Bradley S
2013-12-31
A mesofluidic scale digital valve system includes a first mesofluidic scale valve having a valve body including a bore, wherein the valve body is configured to cooperate with a solenoid disposed substantially adjacent to the valve body to translate a poppet carried within the bore. The mesofluidic scale digital valve system also includes a second mesofluidic scale valve disposed substantially perpendicular to the first mesofluidic scale valve. The mesofluidic scale digital valve system further includes a control element in communication with the solenoid, wherein the control element is configured to maintain the solenoid in an energized state for a fixed period of time to provide a desired flow rate through an orifice of the second mesofluidic valve.
Wang, Lv; Lu, Fang-Lin; Wang, Chong; Tan, Meng-Wei; Xu, Zhi-yun
2014-12-01
The Society of Thoracic Surgeons 2008 cardiac surgery risk models have been developed for heart valve surgery with and without coronary artery bypass grafting. The aim of our study was to evaluate the performance of Society of Thoracic Surgeons 2008 cardiac risk models in Chinese patients undergoing single valve surgery and the predicted mortality rates of those undergoing multiple valve surgery derived from the Society of Thoracic Surgeons 2008 risk models. A total of 12,170 patients underwent heart valve surgery from January 2008 to December 2011. Combined congenital heart surgery and aortal surgery cases were excluded. A relatively small number of valve surgery combinations were excluded. The final research population included the following isolated heart valve surgery types: aortic valve replacement, mitral valve replacement, and mitral valve repair. The following combined valve surgery types were included: mitral valve replacement plus tricuspid valve repair, mitral valve replacement plus aortic valve replacement, and mitral valve replacement plus aortic valve replacement and tricuspid valve repair. Evaluation was performed by using the Hosmer-Lemeshow test and C-statistics. Data from 9846 patients were analyzed. The Society of Thoracic Surgeons 2008 cardiac risk models showed reasonable discrimination and poor calibration (C-statistic, 0.712; P = .00006 in Hosmer-Lemeshow test). Society of Thoracic Surgeons 2008 models had better discrimination (C-statistic, 0.734) and calibration (P = .5805) in patients undergoing isolated valve surgery than in patients undergoing multiple valve surgery (C-statistic, 0.694; P = .00002 in Hosmer-Lemeshow test). Estimates derived from the Society of Thoracic Surgeons 2008 models exceeded the mortality rates of multiple valve surgery (observed/expected ratios of 1.44 for multiple valve surgery and 1.17 for single valve surgery). The Society of Thoracic Surgeons 2008 cardiac surgery risk models performed well when predicting the mortality for Chinese patients undergoing valve surgery. The Society of Thoracic Surgeons 2008 models were suitable for single valve surgery in a Chinese population; estimates of mortality for multiple valve surgery derived from the Society of Thoracic Surgeons 2008 models were less accurate. Copyright © 2014 The American Association for Thoracic Surgery. Published by Elsevier Inc. All rights reserved.
Murphy, David J; McEvoy, Sinead H; Iyengar, Sri; Feuchtner, Gudrun; Cury, Ricardo C; Roobottom, Carl; Baumueller, Stephan; Alkadhi, Hatem; Dodd, Jonathan D
2014-08-01
To assess the diagnostic accuracy of standard axial 64-slice chest CT compared to aortic valve image plane ECG-gated cardiac CT for bicuspid aortic valves. The standard axial chest CT scans of 20 patients with known bicuspid aortic valves were blindly, randomly analyzed for (i) the appearance of the valve cusps, (ii) the largest aortic sinus area, (iii) the longest aortic cusp length, (iv) the thickest aortic valve cusp and (v) valve calcification. A second blinded reader independently analyzed the appearance of the valve cusps. Forty-two age- and sex-matched patients with known tricuspid aortic valves were used as controls. Retrospectively ECG-gated cardiac CT multiphase reconstructions of the aortic valve were used as the gold-standard. Fourteen (21%) scans were scored as unevaluable (7 bicuspid, 7 tricuspid). Of the remainder, there were 13 evaluable bicuspid valves, ten of which showed an aortic valve line sign, while the remaining three showed a normal Mercedes-Benz appearance owing to fused valve cusps. The 35 evaluable tricuspid aortic valves all showed a normal Mercedes-Benz appearance (P=0.001). Kappa analysis=0.62 indicating good interobserver agreement for the aortic valve cusp appearance. Aortic sinus areas, aortic cusp lengths and aortic cusp thicknesses of ≥ 3.8 cm(2), 3.2 cm and 1.6mm respectively on standard axial chest CT best distinguished bicuspid from tricuspid aortic valves (P<0.0001 for all). Of evaluable scans, the sensitivity, specificity, positive and negative predictive values of standard axial chest CT in diagnosing bicuspid aortic valves was 77% (CI 0.54-1.0), 100%, 100% and 70% respectively. The aortic valve is evaluable in approximately 80% of standard chest 64-slice CT scans. Bicuspid aortic valves may be diagnosed on evaluable scans with good diagnostic accuracy. An aortic valve line sign, enlarged aortic sinuses and elongated, thickened valve cusps are specific CT features. Copyright © 2014 Elsevier Ireland Ltd. All rights reserved.
Ross operation in children: late results.
Elkins, R C; Lane, M M; McCue, C
2001-11-01
Although the Ross operation has become the accepted aortic valve replacement in children, the long-term fate of the pulmonary autograft valve remains unknown. To assess mid-term and late results of autograft valve durability, patient survival and valve-related morbidity, a retrospective review of patients (age range: 3 days to 17 years) having a Ross operation between November 1986 and May 2001 were reviewed. Medical records and patient contacts with all but two of 167 current survivors of 178 consecutive patients having an aortic valve replacement as a Ross operation have been completed during the past two years. The most recent echocardiographic evaluation was reviewed for autograft valve and homograft valve function. Operative mortality was 4.5% (8/178), with three late deaths (two were non-valve-related) for an actuarial survival of 92+/-3% at 12 years. Actuarial freedom from autograft valve degeneration (reoperation or severe insufficiency of autograft valve or valve-related death) was 90+/-4% at 12 years. Autograft valve degeneration was not affected by technique of insertion (141 root replacement, 37 intra-aortic), aortic valve morphology (157 bicuspid or unicuspid, 26 tricuspid), or age at operation. Autograft valve degeneration was worse in patients with a primary lesion of aortic insufficiency than in those with aortic stenosis (p = 0.03). Autograft valve reoperation was required in 12 patients, with autograft valve replacement in seven. Actuarial freedom from autograft replacement was 93+/-3% at 12 years. Homograft valve replacement was required in seven patients, with actuarial freedom from replacement of 90+/-4% at 12 years. Eight additional patients have homograft valve obstruction (gradient > or =50 mmHg), and seven have severe pulmonary insufficiency. Survival and freedom from aortic valve replacement are excellent in children. Homograft valve late function remains a concern, and efforts to improve homograft durability should be encouraged.
... valve . Learn about the different types of stenosis: Aortic stenosis Tricuspid stenosis Pulmonary stenosis Mitral stenosis Outlook for ... Disease "Innocent" Heart Murmur Problem: Valve Stenosis - Problem: Aortic Valve Stenosis - Problem: Mitral Valve Stenosis - Problem: Tricuspid Valve Stenosis - ...
... and replacing it with a man-made or biological valve. Biological valves are made from pig, cow, or human ... the valve. Man-made valves last longer than biological valves and usually don’t have to be ...
... working correctly. Most valve replacements involve the aortic Tricuspid valve and mitral valves. The aortic valve separates ... where it shouldn’t. This is called incompetence, insufficiency or regurgitation. • Prolapse — mitral valve flaps don’t ...
FLUID MECHANICS OF ARTIFICIAL HEART VALVES
Dasi, Lakshmi P; Simon, Helene A; Sucosky, Philippe; Yoganathan, Ajit P
2009-01-01
SUMMARY 1. Artificial heart valves have been in use for over five decades to replace diseased heart valves. Since the first heart valve replacement performed with a caged-ball valve, more than 50 valve designs have been developed, differing principally in valve geometry, number of leaflets and material. To date, all artificial heart valves are plagued with complications associated with haemolysis, coagulation for mechanical heart valves and leaflet tearing for tissue-based valve prosthesis. For mechanical heart valves, these complications are believed to be associated with non-physiological blood flow patterns. 2. In the present review, we provide a bird’s-eye view of fluid mechanics for the major artificial heart valve types and highlight how the engineering approach has shaped this rapidly diversifying area of research. 3. Mechanical heart valve designs have evolved significantly, with the most recent designs providing relatively superior haemodynamics with very low aerodynamic resistance. However, high shearing of blood cells and platelets still pose significant design challenges and patients must undergo life-long anticoagulation therapy. Bioprosthetic or tissue valves do not require anticoagulants due to their distinct similarity to the native valve geometry and haemodynamics, but many of these valves fail structurally within the first 10–15 years of implantation. 4. These shortcomings have directed present and future research in three main directions in attempts to design superior artificial valves: (i) engineering living tissue heart valves; (ii) development of advanced computational tools; and (iii) blood experiments to establish the link between flow and blood damage. PMID:19220329
Fluid mechanics of artificial heart valves.
Dasi, Lakshmi P; Simon, Helene A; Sucosky, Philippe; Yoganathan, Ajit P
2009-02-01
1. Artificial heart valves have been in use for over five decades to replace diseased heart valves. Since the first heart valve replacement performed with a caged-ball valve, more than 50 valve designs have been developed, differing principally in valve geometry, number of leaflets and material. To date, all artificial heart valves are plagued with complications associated with haemolysis, coagulation for mechanical heart valves and leaflet tearing for tissue-based valve prosthesis. For mechanical heart valves, these complications are believed to be associated with non-physiological blood flow patterns. 2. In the present review, we provide a bird's-eye view of fluid mechanics for the major artificial heart valve types and highlight how the engineering approach has shaped this rapidly diversifying area of research. 3. Mechanical heart valve designs have evolved significantly, with the most recent designs providing relatively superior haemodynamics with very low aerodynamic resistance. However, high shearing of blood cells and platelets still pose significant design challenges and patients must undergo life-long anticoagulation therapy. Bioprosthetic or tissue valves do not require anticoagulants due to their distinct similarity to the native valve geometry and haemodynamics, but many of these valves fail structurally within the first 10-15 years of implantation. 4. These shortcomings have directed present and future research in three main directions in attempts to design superior artificial valves: (i) engineering living tissue heart valves; (ii) development of advanced computational tools; and (iii) blood experiments to establish the link between flow and blood damage.
Fault Study of Valve Based on Test Analysis and Comparison
NASA Astrophysics Data System (ADS)
Cheng, Li; Yang, Wukui; Liang, Tao; Xu, Yu; Chen, Chao
2017-10-01
The valve of a certain type of small engine often has the fault phenomenon of abnormal vibration noise and can’t close under the specified pressure, which may cause the engine automatic stop because of valve incomplete close leading to fuel leakage during test and startup on the bench. By test study compared to imported valve with the same use function and test condition valve, and put forward the thinking of improving valve structure, compared no-improved valve to improved valve by adopting Fluent field simulation software. As a result, improved valve can restore close pressure of valve, restrain abnormal vibration noise phenomenon, and effectively compensate compression value of spring because of steel ball contacting position downward with valve casing.
Single leg separation prevalence among explanted Björk-Shiley prosthetic heart valves.
Blot, William J; Signorello, Lisa B; Cohen, Sarah S; Ibrahim, Michel A
2007-11-01
Björk-Shiley convexo-concave (BSCC) prosthetic heart valves are believed to have been implanted in over 86,000 patients worldwide. Limited data are available on the prevalence of single leg separations (SLS) of the valves' outlet struts, a potential precursor to complete valve fracture. Data maintained by the manufacturer, including results of examinations for SLS in explanted valves, were merged with available information on the characteristics of the valve. The prevalence of SLS in the examined valves was calculated according to valve angle, size, position, and study. Among 343 examined valves, the overall prevalence of SLS was 8.2%, but this varied significantly by valve size, being three-fold higher among 29+ mm valves than among smaller valves, with statistically non-significantly higher prevalences among mitral than aortic, and among 70 degrees than 60 degrees valves. By applying the size, position and angle-specific SLS prevalences to the worldwide valve distribution, it is estimated that SLS may be present in 6.8% (95% confidence limits 4.1-9.4%) of all BSCC valves. These findings suggest that SLS may affect between 820 and 1,880 of the almost 20,000 BSCC valves among surviving patients worldwide. Such estimates help frame the context for potential patient screenings, should imaging and acoustic techniques to detect SLS become available.
Naber, Christoph K; Ghanem, Alexander; Abizaid, Alexander A; Wolf, Alexander; Sinning, Jan-Malte; Werner, Nikos; Nickenig, Georg; Schmitz, Thomas; Grube, Eberhard
2012-05-15
We describe the first-in-human experience with a novel cerebral embolic protection device used during transcatheter aortic valve implantation (TAVI). One current challenge of TAVI is the reduction of procedural stroke. Procedural mobilisation of debris is a known source of cerebral embolisation. Mechanical protection by transient filtration of cerebral blood flow might reduce the embolic burden during TAVI. We aimed to evaluate the feasibility and safety of the Claret CE Pro™ cerebral protection device in patients undergoing TAVI. Patients scheduled for TAVI were prospectively enrolled at three centres. The Claret CE Pro™ (Claret Medical, Inc. Santa Rosa, CA, USA) cerebral protection device was placed via the right radial/brachial artery prior to TAVI and was removed after the procedure. The primary endpoint was technical success rate. Secondary endpoints encompassed procedural and 30-day stroke rates, as well as device-related complications. Deployment of the Claret CE Pro™ cerebral protection device was intended for use in 40 patients, 35 devices were implanted into the aortic arch. Technical success rate with delivery of the proximal and distal filter was 60% for the first generation device and 87% for the second-generation device. Delivery times for the first-generation device were 12.4±12.1 minutes and 4.4 ± 2.5 minutes for the second-generation device (p<0.05). The quantity of contrast used related to the Claret CE Pro System was 19.6 ± 3.8 ml. Captured debris was documented in at least 19 of 35 implanted devices (54.3%). No procedural transient ischaemic attacks, minor strokes or major strokes occurred. Thirty-day follow-up showed one minor stroke occurring 30 days after the procedure, and two major strokes both occurring well after the patient had completed TAVI. The use of the Claret CE Pro™ system is feasible and safe. Capture of debris in more than half of the patients provides evidence for the potential to reduce the procedural cerebral embolic burden utilising this dedicated filter system during TAVI.
Upton, Hubert Allen; Garcia, Pablo
1999-08-24
A check valve for use in a GDCS of a nuclear reactor and having a motor driven disk including a rotatable armature for rotating the check valve disk over its entire range of motion is described. In one embodiment, the check valve includes a valve body having a coolant flow channel extending therethrough. The coolant flow channel includes an inlet end and an outlet end. A valve body seat is located on an inner surface of the valve body. The check valve further includes a disk assembly, sometimes referred to as the motor driven disc, having a counterweight and a disk shaped valve. The disk valve includes a disk base having a seat for seating with the valve body seat. The disk assembly further includes a first hinge pin member which extends at least partially through the disk assembly and is engaged to the disk. The disk valve is rotatable relative to the first hinge pin member. The check valve also includes a motor having a stator frame with a stator bore therein. An armature is rotatably positioned within the stator bore and the armature is coupled to the disk valve to cause the disk valve to rotate about its full range of motion.
Upton, H.A.; Garcia, P.
1999-08-24
A check valve for use in a GDCS of a nuclear reactor and having a motor driven disk including a rotatable armature for rotating the check valve disk over its entire range of motion is described. In one embodiment, the check valve includes a valve body having a coolant flow channel extending therethrough. The coolant flow channel includes an inlet end and an outlet end. A valve body seat is located on an inner surface of the valve body. The check valve further includes a disk assembly, sometimes referred to as the motor driven disc, having a counterweight and a disk shaped valve. The disk valve includes a disk base having a seat for seating with the valve body seat. The disk assembly further includes a first hinge pin member which extends at least partially through the disk assembly and is engaged to the disk. The disk valve is rotatable relative to the first hinge pin member. The check valve also includes a motor having a stator frame with a stator bore therein. An armature is rotatably positioned within the stator bore and the armature is coupled to the disk valve to cause the disk valve to rotate about its full range of motion. 5 figs.
46 CFR 56.20-9 - Valve construction.
Code of Federal Regulations, 2011 CFR
2011-10-01
...) motion of the handwheel or operating lever as seen by one facing the end of the valve stem. Each gate, globe, and angle valve must generally be of the rising-stem type, preferably with the stem threads... nonrising-stem valve will be acceptable. Each nonrising-stem valve, lever-operated valve, or other valve...
46 CFR 56.20-9 - Valve construction.
Code of Federal Regulations, 2013 CFR
2013-10-01
...) motion of the handwheel or operating lever as seen by one facing the end of the valve stem. Each gate, globe, and angle valve must generally be of the rising-stem type, preferably with the stem threads... nonrising-stem valve will be acceptable. Each nonrising-stem valve, lever-operated valve, or other valve...
46 CFR 56.20-9 - Valve construction.
Code of Federal Regulations, 2012 CFR
2012-10-01
...) motion of the handwheel or operating lever as seen by one facing the end of the valve stem. Each gate, globe, and angle valve must generally be of the rising-stem type, preferably with the stem threads... nonrising-stem valve will be acceptable. Each nonrising-stem valve, lever-operated valve, or other valve...
46 CFR 56.20-9 - Valve construction.
Code of Federal Regulations, 2014 CFR
2014-10-01
...) motion of the handwheel or operating lever as seen by one facing the end of the valve stem. Each gate, globe, and angle valve must generally be of the rising-stem type, preferably with the stem threads... nonrising-stem valve will be acceptable. Each nonrising-stem valve, lever-operated valve, or other valve...
Microfluidic Valves Made From Polymerized Polyethylene Glycol Diacrylate
Rogers, Chad I.; Oxborrow, Joseph B.; Anderson, Ryan R.; Tsai, Long-Fang; Nordin, Gregory P.; Woolley, Adam T.
2013-01-01
Pneumatically actuated, non-elastomeric membrane valves fabricated from polymerized polyethylene glycol diacrylate (poly-PEGDA) have been characterized for temporal response, valve closure, and long-term durability. A ~100 ms valve opening time and a ~20 ms closure time offer valve operation as fast as 8 Hz with potential for further improvement. Comparison of circular and rectangular valve geometries indicates that the surface area for membrane interaction in the valve region is important for valve performance. After initial fabrication, the fluid pressure required to open a closed circular valve is ~50 kPa higher than the control pressure holding the valve closed. However, after ~1000 actuations to reconfigure polymer chains and increase elasticity in the membrane, the fluid pressure required to open a valve becomes the same as the control pressure holding the valve closed. After these initial conditioning actuations, poly-PEGDA valves show considerable robustness with no change in effective operation after 115,000 actuations. Such valves constructed from non-adsorptive poly-PEGDA could also find use as pumps, for application in small volume assays interfaced with biosensors or impedance detection, for example. PMID:24357897
Lotfi, Shahram; Becker, Michael; Moza, Ajay; Autschbach, Rüdiger; Marx, Nikolaus; Schröder, Jörg
2017-09-10
Transcatheter aortic valve implantation has become an accepted treatment modality for inoperable or high-risk surgical patients with symptomatic severe aortic stenosis. We report the case of a 70-year-old white man who was treated for severe symptomatic aortic regurgitation using transcatheter aortic valve implantation from the apical approach. Because of recurrent cardiac decompensation 4 weeks after implantation he underwent the implantation of a left ventricular assist device system. A year later echocardiography showed a severe transvalvular central insufficiency. Our heart team decided to choose a valve-in-valve approach while reducing the flow rate of left ventricular assist device to minimum and pacing with a frequency of 140 beats/minute. There was an excellent result and our patient is doing well with no relevant insufficiency of the aortic valve at 12-month follow-up. This is the first report about a successful treatment of a stenotic JenaValve using a CoreValve Evolut R; the use of a CoreValve Evolut R prosthesis may be an optimal option for valve-in-valve procedures.
Bennett, Charles L.; Sewall, Noel; Boroa, Carl
2014-08-19
An engine based on a reciprocating piston engine that extracts work from pressurized working fluid. The engine includes a harmonic oscillator inlet valve capable of oscillating at a resonant frequency for controlling the flow of working fluid into of the engine. In particular, the inlet valve includes an inlet valve head and a spring arranged together as a harmonic oscillator so that the inlet valve head is moveable from an unbiased equilibrium position to a biased closed position occluding an inlet. Upon releasing the inlet valve the inlet valve head undergoes a single oscillation past the equilibrium positio to a maximum open position and returns to a biased return position close to the closed position to choke the flow and produce a pressure drop across the inlet valve causing the inlet valve to close. Protrusions carried either by the inlet valve head or piston head are used to bump open the inlet valve from the closed position and initiate the single oscillation of the inlet valve head, and protrusions carried either by the outlet valve head or piston head are used to close the outlet valve ahead of the bump opening of the inlet valve.
Bosmans, Johan; Bleiziffer, Sabine; Gerckens, Ulrich; Wenaweser, Peter; Brecker, Stephen; Tamburino, Corrado; Linke, Axel
2015-07-21
Transcatheter aortic valve replacement (TAVR) enables treatment of high-risk patients with symptomatic aortic stenosis without open-heart surgery; however, the benefits are mitigated by the potential for neurological events. This study sought to determine the timing and causes of clinically relevant neurological events after self-expandable TAVR. We enrolled 1,015 patients, of whom 996 underwent TAVR with a self-expandable system at 44 TAVR-experienced centers in Europe, Colombia, and Israel. Neurological events were evaluated for 3 distinct time periods: periprocedural (0 to 1 days post TAVR); early (2 to 30 days); and late (31 to 730 days). In this real-world study, neurological events were first referred to the site neurologist and then reviewed by an independent neurologist. The overall stroke rate was 1.4% through the first day post-procedure, 3.0% at 30 days, and 5.6% at 2 years. There were no significant predictors of periprocedural stroke or stroke/transient ischemic attack (TIA) combined. Significant predictors of early stroke were acute kidney injury (p = 0.03), major vascular complication (p = 0.04), and female sex (p = 0.04). For stroke/TIA combined, prior atrial fibrillation (p = 0.03) and major vascular complication (p = 0.009) were predictive. Coronary artery bypass graft surgery was the only significant predictor of late stroke (p = 0.007) or late stroke/TIA (p = 0.06). Treatment of high-risk patients with aortic stenosis using a self-expandable system was associated with a low stroke rate at short- and long-term follow-up. Multivariable predictors of clinically relevant neurological events differed on the basis of the timing after TAVR. (CoreValve Advance International Post Market Study; NCT01074658). Copyright © 2015 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Husser, Oliver; Kessler, Thorsten; Burgdorf, Christof; Templin, Christian; Pellegrini, Costanza; Schneider, Simon; Kasel, Albert Markus; Kastrati, Adnan; Schunkert, Heribert; Hengstenberg, Christian
2016-02-01
Transcatheter aortic valve implantation is increasingly used in patients with aortic stenosis. Post-procedural intraventricular conduction abnormalities and permanent pacemaker implantations remain a serious concern. Recently, the Edwards SAPIEN 3 prosthesis has replaced the SAPIEN XT. We sought to determine the incidences of new-onset intraventricular conduction abnormalities and permanent pacemaker implantations by comparing the 2 devices. We analyzed the last consecutive 103 patients undergoing transcatheter aortic valve implantation with SAPIEN XT before SAPIEN 3 was used in the next 105 patients. To analyze permanent pacemaker implantations and new-onset intraventricular conduction abnormalities, patients with these conditions at baseline were excluded. Electrocardiograms were recorded at baseline, after the procedure, and before discharge. SAPIEN 3 was associated with higher device success (100% vs 92%; P=.005) and less paravalvular leakage (0% vs 7%; P<.001). The incidence of permanent pacemaker implantations was 12.6% (23 of 183) with no difference between the 2 groups (SAPIEN 3: 12.5% [12 of 96] vs SAPIEN XT: 12.6% [11 of 87]; P=.99). SAPIEN 3 was associated with a higher rate of new-onset intraventricular conduction abnormalities (49% vs 27%; P=.007) due to a higher rate of fascicular blocks (17% vs 5%; P=.021). There was no statistically significant difference in transient (29% [20 of 69] vs persistent 19% [12 of 64]; P=.168) left bundle branch blocks (28% [19 of 69] vs 17% [11 of 64]; P=.154) when SAPIEN 3 was compared with SAPIEN XT. We found a trend toward a higher rate of new-onset intraventricular conduction abnormalities with SAPIEN 3 compared with SAPIEN XT, although this did not result in a higher permanent pacemaker implantation rate. Copyright © 2015 Sociedad Española de Cardiología. Published by Elsevier España, S.L.U. All rights reserved.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Love, Lonnie J.; Mell, Ellen
2015-02-01
AeroValve s innovative pneumatic valve technology recycles compressed air through the valve body with each cycle of the valve, and was reported to reduce compressed air requirements by an average of 25% 30%.This technology collaboration project between ORNL and Aerovalve confirms the energy efficiency of valve performance. Measuring air consumption per work completed, the AeroValve was as much as 85% better than the commercial Festo valve.
Carbon Fiber Reinforced Carbon Composite Valve for an Internal Combustion Engine
NASA Technical Reports Server (NTRS)
Rivers, H. Kevin (Inventor); Ransone, Philip O. (Inventor); Northam, G. Burton (Inventor)
1999-01-01
A carbon fiber reinforced carbon composite valve for internal combustion engines and the like formed of continuous carbon fibers throughout the valve's stem and head is disclosed. The valve includes braided carbon fiber material over axially aligned unidirectional carbon fibers forming a valve stem; the braided and unidirectional carbon fibers being broomed out at one end of the valve stem forming the shape of the valve head; the valve-shaped structure being densified and rigidized with a matrix of carbon containing discontinuous carbon fibers: and the finished valve being treated to resist oxidation. Also disclosed is a carbon matrix plug containing continuous and discontinuous carbon fibers and forming a net-shape valve head acting as a mandrel over which the unidirectional and braided carbon fibers are formed according to textile processes. Also disclosed are various preform valves and processes for making finished and preform carbon fiber reinforced carbon composite valves.
Patient-specific pediatric silicone heart valve models based on 3D ultrasound
NASA Astrophysics Data System (ADS)
Ilina, Anna; Lasso, Andras; Jolley, Matthew A.; Wohler, Brittany; Nguyen, Alex; Scanlan, Adam; Baum, Zachary; McGowan, Frank; Fichtinger, Gabor
2017-03-01
PURPOSE: Patient-specific heart and valve models have shown promise as training and planning tools for heart surgery, but physically realistic valve models remain elusive. Available proprietary, simulation-focused heart valve models are generic adult mitral valves and do not allow for patient-specific modeling as may be needed for rare diseases such as congenitally abnormal valves. We propose creating silicone valve models from a 3D-printed plastic mold as a solution that can be adapted to any individual patient and heart valve at a fraction of the cost of direct 3D-printing using soft materials. METHODS: Leaflets of a pediatric mitral valve, a tricuspid valve in a patient with hypoplastic left heart syndrome, and a complete atrioventricular canal valve were segmented from ultrasound images. A custom software was developed to automatically generate molds for each valve based on the segmentation. These molds were 3D-printed and used to make silicone valve models. The models were designed with cylindrical rims of different sizes surrounding the leaflets, to show the outline of the valve and add rigidity. Pediatric cardiac surgeons practiced suturing on the models and evaluated them for use as surgical planning and training tools. RESULTS: Five out of six surgeons reported that the valve models would be very useful as training tools for cardiac surgery. In this first iteration of valve models, leaflets were felt to be unrealistically thick or stiff compared to real pediatric leaflets. A thin tube rim was preferred for valve flexibility. CONCLUSION: The valve models were well received and considered to be valuable and accessible tools for heart valve surgery training. Further improvements will be made based on surgeons' feedback.
Development of Long-Lifetime Pulsed Gas Valves for Pulsed Electric Thrusters
NASA Technical Reports Server (NTRS)
Burkhardt, Wendel M.; Crapuchettes, John M.; Addona, Brad M.; Polzin, Kurt A.
2015-01-01
The design and test results for two types of pulsed gas valves are presented. The valves, a piezo valve and a solenoid actuated valve, must have exceedingly long lifetime to support gas-fed pulsed electric thruster operation for missions of interest. The performance of both valves was tested, with both demonstrating the capability to throttle the gas flow rate while maintaining low leakage levels below 10(exp -3) sccs of He at the beginning of valve lifetime. The piezo valve varies the flow rate by changing the amount that the valve is open, which is a function of applied voltage. This valve demonstrated continuous throttlability from 0-10 mL/s, with opening and closing times of 100 microsecond or less. The solenoid actuated valve flow rate changes as a function of the inlet gas pressure, with demonstrated flow rates in these tests from 2.7-11 mL per second. The valve response time is slower than the piezo valve, opening in 1-2 ms and closing in several ms. The solenoid actuated valve was tested to one million cycles, with the valve performance remaining relatively unchanged throughout the test. Galling of the sliding plunger caused the valve to bind and fail just after one million cycles, but at this point in the test the valve sealing surface leak rate still appeared to be well below the maximum target leak rake of 1×10(exp -3) sccs of He.
NASA Astrophysics Data System (ADS)
Dalesandro, Andrew A.; Theilacker, Jay; Van Sciver, Steven
2012-06-01
Safe operation of superconducting radio frequency (SRF) cavities require design consideration of a sudden catastrophic loss of vacuum (SCLV) adjacent with liquid helium (LHe) vessels and subsequent dangers. An experiment is discussed to test the longitudinal effects of SCLV along the beam line of a string of scaled SRF cavities. Each scaled cavity includes one segment of beam tube within a LHe vessel containing 2 K saturated LHe, and a riser pipe connecting the LHe vessel to a common gas header. At the beam tube inlet is a fast acting solenoid valve to simulate SCLV and a high/low range orifice plate flow-meter to measure air influx to the cavity. The gas header exit also has an orifice plate flow-meter to measure helium venting the system at the relief pressure of 0.4 MPa. Each cavity is instrumented with Validyne pressure transducers and Cernox thermometers. The purpose of this experiment is to quantify the time required to spoil the beam vacuum and the effects of transient heat and mass transfer on the helium system. Heat transfer data is expected to reveal a longitudinal effect due to the geometry of the experiment. Details of the experimental design criteria and objectives are presented.
He, Guoxi; Liang, Yongtu; Li, Yansong; Wu, Mengyu; Sun, Liying; Xie, Cheng; Li, Feng
2017-06-15
The accidental leakage of long-distance pressurized oil pipelines is a major area of risk, capable of causing extensive damage to human health and environment. However, the complexity of the leaking process, with its complex boundary conditions, leads to difficulty in calculating the leakage volume. In this study, the leaking process is divided into 4 stages based on the strength of transient pressure. 3 models are established to calculate the leaking flowrate and volume. First, a negative pressure wave propagation attenuation model is applied to calculate the sizes of orifices. Second, a transient oil leaking model, consisting of continuity, momentum conservation, energy conservation and orifice flow equations, is built to calculate the leakage volume. Third, a steady-state oil leaking model is employed to calculate the leakage after valves and pumps shut down. Moreover, sensitive factors that affect the leak coefficient of orifices and volume are analyzed respectively to determine the most influential one. To validate the numerical simulation, two types of leakage test with different sizes of leakage holes were conducted from Sinopec product pipelines. More validations were carried out by applying commercial software to supplement the experimental insufficiency. Thus, the leaking process under different leaking conditions are described and analyzed. Copyright © 2017 Elsevier B.V. All rights reserved.
De Gaetano, Francesco; Serrani, Marta; Bagnoli, Paola; Brubert, Jacob; Stasiak, Joanna; Moggridge, Geoff D.; Costantino, Maria Laura
2016-01-01
Introduction Only mechanical and biological heart valve prostheses are currently commercially available. The former show longer durability but require anticoagulant therapy, the latter display better fluid dynamic behaviour but do not have adequate durability. New Polymeric Heart Valves (PHVs) could potentially combine the haemodynamic properties of biological valves with the durability of mechanical valves. This work presents a hydrodynamic evaluation of two groups of newly developed supra-annular tri-leaflet prosthetic heart valves made from styrenic block copolymers (SBC): Poli-Valves. Methods Two types of Poli-Valves made of SBC differing in polystyrene fraction content were tested under continuous and pulsatile flow conditions as prescribed by ISO 5840 Standard. An ad - hoc designed pulse duplicator allowed the valve prototypes to be tested at different flow rates and frequencies. Pressure and flow were recorded; pressure drops, effective orifice area (EOA), and regurgitant volume were computed to assess the valve’s behaviour. Results Both types Poli-Valves met the minimum requirements in terms of regurgitation and EOA as specified by ISO 5840 Standard. Results were compared with five mechanical heart valves (MHVs) and five tissue heart valves (THVs), currently available on the market. Conclusion Based on these results, polymeric heart valves based on styrenic block copolymers, as Poli-Valves are, can be considered as promising alternative for heart valve replacement in near future. PMID:26689146
De Gaetano, Francesco; Serrani, Marta; Bagnoli, Paola; Brubert, Jacob; Stasiak, Joanna; Moggridge, Geoff D; Costantino, Maria Laura
2015-11-01
Only mechanical and biological heart valve prostheses are currently commercially available. The former show longer durability but require anticoagulant therapy; the latter display better fluid dynamic behavior but do not have adequate durability. New Polymeric Heart Valves (PHVs) could potentially combine the hemodynamic properties of biological valves with the durability of mechanical valves. This work presents a hydrodynamic evaluation of 2 groups of newly developed supra-annular, trileaflet prosthetic heart valves made from styrenic block copolymers (SBC): Poli-Valves. 2 types of Poli-Valves made of SBC and differing in polystyrene fraction content were tested under continuous and pulsatile flow conditions as prescribed by ISO 5840 Standard. A pulse duplicator designed ad hoc allowed the valve prototypes to be tested at different flow rates and frequencies. Pressure and flow were recorded; pressure drops, effective orifice area (EOA), and regurgitant volume were computed to assess the behavior of the valve. Both types of Poli-Valves met the minimum requirements in terms of regurgitation and EOA as specified by the ISO 5840 Standard. Results were compared with 5 mechanical heart valves (MHVs) and 5 tissue heart valves (THVs), currently available on the market. Based on these results, PHVs based on styrenic block copolymers, as are Poli-Valves, can be considered a promising alternative for heart valve replacement in the near future.
NASA Astrophysics Data System (ADS)
Zhao, Jianhua; Zhou, Songlin; Lu, Xianghui; Gao, Dianrong
2015-09-01
The double flapper-nozzle servo valve is widely used to launch and guide the equipment. Due to the large instantaneous flow rate of servo valve working under specific operating conditions, the temperature of servo valve would reach 120°C and the valve core and valve sleeve deform in a short amount of time. So the control precision of servo valve significantly decreases and the clamping stagnation phenomenon of valve core appears. In order to solve the problem of degraded control accuracy and clamping stagnation of servo valve under large temperature difference circumstance, the numerical simulation of heat-fluid-solid coupling by using finite element method is done. The simulation result shows that zero position leakage of servo valve is basically impacted by oil temperature and change of fit clearance. The clamping stagnation is caused by warpage-deformation and fit clearance reduction of the valve core and valve sleeve. The distribution rules of the temperature and thermal-deformation of shell, valve core and valve sleeve and the pressure, velocity and temperature field of flow channel are also analyzed. Zero position leakage and electromagnet's current when valve core moves in full-stroke are tested using Electro-hydraulic Servo-valve Characteristic Test-bed of an aerospace sciences and technology corporation. The experimental results show that the change law of experimental current at different oil temperatures is roughly identical to simulation current. The current curve of the electromagnet is smooth when oil temperature is below 80°C, but the amplitude of current significantly increases and the hairy appears when oil temperature is above 80°C. The current becomes smooth again after the warped valve core and valve sleeve are reground. It indicates that clamping stagnation is caused by warpage-deformation and fit clearance reduction of valve core and valve sleeve. This paper simulates and tests the heat-fluid-solid coupling of double flapper-nozzle servo valve, and the obtained results provide the reference value for the design of double flapper-nozzle force feedback servo valve.
Fujita, Buntaro; Scholtz, Smita; Ensminger, Stephan
2016-04-01
Coronary obstruction during transcatheter aortic valve implantation is a potentially life-threatening complication. Most of the widely used transcatheter heart valves require a certain distance between the basal aortic annular plane and the origins of the coronary arteries. We report the case of a successful valve-in-valve procedure with an Edwards SAPIEN XT valve into a JenaValve as a bail-out procedure in a patient with a low originating left coronary artery and a heavily calcified aorta. © 2015 Wiley Periodicals, Inc.
Transcatheter Aortic Valve Replacement for Native Aortic Valve Regurgitation
Spina, Roberto; Anthony, Chris; Muller, David WM
2015-01-01
Transcatheter aortic valve replacement with either the balloon-expandable Edwards SAPIEN XT valve, or the self-expandable CoreValve prosthesis has become the established therapeutic modality for severe aortic valve stenosis in patients who are not deemed suitable for surgical intervention due to excessively high operative risk. Native aortic valve regurgitation, defined as primary aortic incompetence not associated with aortic stenosis or failed valve replacement, on the other hand, is still considered a relative contraindication for transcatheter aortic valve therapies, because of the absence of annular or leaflet calcification required for secure anchoring of the transcatheter heart valve. In addition, severe aortic regurgitation often coexists with aortic root or ascending aorta dilatation, the treatment of which mandates operative intervention. For these reasons, transcatheter aortic valve replacement has been only sporadically used to treat pure aortic incompetence, typically on a compassionate basis and in surgically inoperable patients. More recently, however, transcatheter aortic valve replacement for native aortic valve regurgitation has been trialled with newer-generation heart valves, with encouraging results, and new ancillary devices have emerged that are designed to stabilize the annulus–root complex. In this paper we review the clinical context, technical characteristics and outcomes associated with transcatheter treatment of native aortic valve regurgitation. PMID:29588674
Should patients with Björk-Shiley valves undergo prophylactic replacement?
Birkmeyer, J D; Marrin, C A; O'Connor, G T
1992-08-29
About 85,000 patients have undergone replacement of diseased heart valves with prosthetic Björk-Shiley convexo-concave (CC) valves. These valves are prone to fracture of the outlet strut, which leads to acute valve failure that is usually fatal. Should patients with these valves undergo prophylactic replacement to avoid fracture? The incidence of strut fracture varies between 0% and 1.5% per year, depending on valve opening angle (60 degrees or 70 degrees), diameter (less than 29 mm or greater than or equal to 29 mm), and location (aortic or mitral). Other factors include the patient's life expectancy and the expected morbidity and mortality associated with reoperation. We have used decision analysis to identify the patients most likely to benefit from prophylactic reoperation. The incidence of outlet strut fracture was estimated from the data of three large studies on CC valves, and stratified by opening angle, diameter, and location. A Markov decision analysis model was used to estimate life expectancy for patients undergoing prophylactic valve replacement and for those not undergoing reoperation. Prophylactic valve replacement does not benefit patients with CC valves that have low strut fracture risks (60 degrees aortic valves and less than 29 mm, 60 degrees mitral valves). For most patients with CC valves that have high strut fracture risks (greater than or equal to 29 mm, 70 degrees CC), prophylactic valve replacement increases life expectancy. However, elderly patients with such valves benefit from prophylactic reoperation only if the risk of operative mortality is low. Patient age and operative risk are most important in recommendations for patients with CC valves that have intermediate strut fracture risks (less than 29 mm, 70 degrees valves and greater than or equal to 29 mm, 60 degrees mitral valves). For all patients and their doctors facing the difficult decision on whether to replace CC valves, individual estimates of operative mortality risk that take account of both patient-specific and institution-specific factors are essential.
Cevasco, Marisa; Mick, Stephanie L; Kwon, Michael; Lee, Lawrence S; Chen, Edward P; Chen, Frederick Y
2013-05-01
Currently, there is no universal standard for sizing bioprosthetic aortic valves. Hence, a standardized comparison was performed to clarify this issue. Every size of four commercially available bioprosthetic aortic valves marketed in the United States (Biocor Supra; Mosaic Ultra; Magna Ease; Mitroflow) was obtained. Subsequently, custom sizers were created that were accurate to 0.0025 mm to represent aortic roots 18 mm through 32 mm, and these were used to measure the external diameter of each valve. Using the effective orifice area (EOA) and transvalvular pressure gradient (TPG) data submitted to the FDA, a comparison was made between the hemodynamic properties of valves with equivalent manufacturer stated sizes and valves with equivalent measured external diameters. Based on manufacturer size alone, the valves at first seemed to be hemodynamically different from each other, with Mitroflow valves appearing to be hemodynamically superior, having a large EOA and equivalent or superior TPG (p < 0.05). However, Mitroflow valves had a larger measured external diameter than the other valves of a given numerical manufacturer size. Valves with equivalent external diameters were then compared, regardless of the stated manufacturer sizes. For truly equivalently sized valves (i.e., by measured external diameter) there was no clear hemodynamic difference. There was no statistical difference in the EOAs between the Biocor Supra, Mosaic Ultra, and Mitroflow valves, and the Magna Ease valve had a statistically smaller EOA (p < 0.05). On comparing the mean TPG, the Biocor Supra and Mitroflow valves had statistically equivalent gradients to each other, as did the Mosaic Ultra and Magna Ease valves. When comparing valves of the same numerical manufacturer size, there appears to be a difference in hemodynamic performance across different manufacturers' valves according to FDA data. However, comparing equivalently measured valves eliminates the differences between valves produced by different manufacturers.
5. DIABLO DAM: DETAIL VIEW OF RELIEF VALVES AT ELEVATION ...
5. DIABLO DAM: DETAIL VIEW OF RELIEF VALVES AT ELEVATION 1044. VALVE IN FOREGROUND IS A BUTTERFLY VALVE SIX FEET IN DIAMETER; VALVE TO THE REAR IS A JOHNSON-TYPE NEEDLE VALVE BOTH VALVES WERE MANUFACTURED BY THE PELTON WATER WHEEL COMPANY, 1989. - Skagit Power Development, Diablo Dam, On Skagit River, 6.9 miles upstream from Newhalem, Newhalem, Whatcom County, WA
The Effect of Valve Cooling upon Maximum Permissible Engine Output as Limited by Knock
NASA Technical Reports Server (NTRS)
Munger, Maurice; Wilsted, H D; Mulcahy, B A
1942-01-01
A Wright GR-1820-G200 cylinder was tested over a wide range of fuel-air ratios at maximum permissible power output as limited by knock with three different degrees of valve cooling. The valves used were stock valves (solid inlet valve and hollow sodium-cooled exhaust valve), hollow valves with no coolant, and hollow valves with flowing water as a coolant. Curves showing the variation in maximum permissible values of inlet-air pressure, indicated mean effective pressure, cylinder charge, and indicated specific fuel consumption with change in fuel-air ratio and valve cooling are shown. The use of valves cooled by a stream of water passing through their hollow interiors permitted indicated mean effective pressures 10 percent higher than the mean effective pressures permissible with stock valves when the engine was operated with fuel-air ratios from 0.055 to 0.065. Operation of the engine with lean mixtures with uncooled hollow valves resulted in power output below the output obtained with the stock valves. The data show an increase in maximum permissible indicated mean effective pressure due to cooling the valves, which averages only 2.1 percent with fuel-air ratios from 0.075 to 0.105.
Reynolds Shear Stress for Textile Prosthetic Heart Valves in Relation to Fabric Design
Bark, David L.; Koupei, Atieh Yousefi; Forleo, Marcio; Vaesken, Antoine; Heim, Frederic; Dasi, Lakshmi P.
2016-01-01
The most widely implanted prosthetic heart valves are either mechanical or bioprosthetic. While the former suffers from thrombotic risks, the latter suffers from a lack of durability. Textile valves, alternatively, can be designed with durability and to exhibit hemodynamics similar to the native valve, lowering the risk for thrombosis. Deviations from native valve hemodynamics can result in an increased Reynolds Shear Stress (RSS), which has the potential to instigate hemolysis or shear-induced thrombosis. This study is aimed at characterizing flow in multiple textile valve designs with an aim of developing a low profile valve. Valves were created using a shaping process based on heating a textile membrane and placed within a left heart simulator. Turbulence and bulk hemodynamics were assessed through particle imaging velocimetry (PIV), along with flow and pressure measurements. Overall, RSS was reduced for low profile valves relative to high profile valves, but was otherwise similar among low profile valves. However, leakage was found in 3 of the 4 low profile valve designs driving the fabric design for low profile valves. Through textile design, low profile valves can be created with favorable hemodynamics. PMID:26919564
Benito, Natividad; Miró, José M.; de Lazzari, Elisa; Cabell, Christopher H; del Río, Ana; Altclas, Javier; Commerford, Patrick; Delahaye, Francois; Dragulescu, Stefan; Giamarellou, Helen; Habib, Gilbert; Kamarulzaman, Adeeba; Kumar, A. Sampath; Nacinovich, Francisco M.; Suter, Fredy; Tribouilloy, Christophe; Venugopal, K; Moreno, Asuncion; Fowler, Vance G.
2013-01-01
Background The clinical profile and outcome of nosocomial and non-nosocomial health care–associated native valve endocarditis are not well defined. Objective To describe the prevalence, clinical characteristics, and outcomes of nosocomial and non-nosocomial health care–associated native valve endocarditis. Design Prospective observational study. Setting 61 hospitals in 28 countries. Patients Patients with definite native valve endocarditis and no history of injection drug use who were enrolled in the International Collaboration on Endocarditis–Prospective Cohort Study from June 2000 to August 2005. Measurements Characteristics of nosocomial and non-nosocomial health care–associated native valve endocarditis cases were described and compared with those cases acquired in the community. Results Health care–associated native valve endocarditis was present in 557 (34%) of 1622 patients with native valve endocarditis and no history of injection drug use (nosocomial native valve endocarditis 303 patients [54%]; non-nosocomial health care–associated native valve endocarditis 254 patients [46%]). Staphylococcus aureus was the most common cause of health care-associated native valve endocarditis (nosocomial native valve endocarditis, 47%; non-nosocomial health care–associated native valve endocarditis, 42%; p=0.3), with a notable proportion of methicillin-resistant S. aureus (nosocomial native valve endocarditis, 57%; non-nosocomial health care–associated native valve endocarditis, 41%; p=0.014). Patients with health care–associated native valve endocarditis had lower rates of cardiac surgery (41% health care–associated native valve endocarditis vs 51% community-acquired native valve endocarditis, p<0.001) and higher in-hospital mortality rates than patients with community-acquired native valve endocarditis (25% health care–associated native valve endocarditis vs. 13% community-acquired native valve endocarditis vs., p<0.001). Multivariable analysis confirmed a higher mortality associated with health care–associated native valve endocarditis (incidence risk ratio=1.20 (CI 95%, 1.03–1.61). Limitations This study involves tertiary hospitals with cardiac surgery programs. The results may not be generalized to patient populations receiving care in other types of facility. Conclusions More than one-third of all cases of native valve endocarditis in non-drug users involve contact with health care. S. aureus is the leading cause of health care–associated native valve endocarditis. Non-nosocomial health care–associated native valve endocarditis is common, especially in the US. Patients with health care-associated and community-acquired native valve endocarditis differ in their presentation, microbiology, and outcome. By contrast, patients with nosocomial and non-nosocomial healthcare-associated endocarditis are similar. PMID:19414837
Thompson, J.L.; Upton, H.A.
1999-04-27
Breaker valve assemblies for a simplified boiling water nuclear reactor are described. The breaker valve assembly, in one form, includes a valve body and a breaker valve. The valve body includes an interior chamber, and an inlet passage extends from the chamber and through an inlet opening to facilitate transporting particles from outside of the valve body to the interior chamber. The breaker valve is positioned in the chamber and is configured to substantially seal the inlet opening. Particularly, the breaker valve includes a disk which is sized to cover the inlet opening. The disk is movably coupled to the valve body and is configured to move substantially concentrically with respect to the valve opening between a first position, where the disk completely covers the inlet opening, and a second position, where the disk does not completely cover the inlet opening. 1 fig.
Thompson, Jeffrey L.; Upton, Hubert Allen
1999-04-27
Breaker valve assemblies for a simplified boiling water nuclear reactor are described. The breaker valve assembly, in one form, includes a valve body and a breaker valve. The valve body includes an interior chamber, and an inlet passage extends from the chamber and through an inlet opening to facilitate transporting particles from outside of the valve body to the interior chamber. The breaker valve is positioned in the chamber and is configured to substantially seal the inlet opening. Particularly, the breaker valve includes a disk which is sized to cover the inlet opening. The disk is movably coupled to the valve body and is configured to move substantially concentrically with respect to the valve opening between a first position, where the disk completely covers the inlet opening, and a second position, where the disk does not completely cover the inlet opening.
Microfluidic valve array control system integrating a fluid demultiplexer circuit
NASA Astrophysics Data System (ADS)
Kawai, Kentaro; Arima, Kenta; Morita, Mizuho; Shoji, Shuichi
2015-06-01
This paper proposes an efficient control method for the large-scale integration of microvalves in microfluidic systems. The proposed method can control 2n individual microvalves with 2n + 2 control lines (where n is an integer). The on-chip valves are closed by applying pressure to a control line, similar to conventional pneumatic microvalves. Another control line closes gate valves between the control line to the on-chip valves and the on-chip valves themselves, to preserve the state of the on-chip valves. The remaining control lines select an activated gate valve. While the addressed gate valve is selected by the other control lines, the corresponding on-chip valve is actuated by applying input pressure to the control line to the on-chip valves. Using this method would substantially reduce the number of world-to-chip connectors and off-chip valve controllers. Experiments conducted using a fabricated 28 microvalve array device, comprising 256 individual on-chip valves controlled with 18 (2 × 8 + 2) control lines, yielded switching speeds for the selected on-chip valve under 90 ms.
Olivieri, Laura J.; Baba, Ridhwan Y.; Arai, Andrew E.; Bandettini, W. Patricia; Rosing, Douglas R.; Bakalov, Vladimir; Sachdev, Vandana; Bondy, Carolyn A.
2014-01-01
Background Congenital aortic valve fusion is associated with aortic dilation, aneurysm and rupture in girls and women with Turner syndrome (TS). Our objective was to characterize aortic valve structure in subjects with TS, and determine the prevalence of aortic dilation and valve dysfunction associated with different types of aortic valves. Methods and Results The aortic valve and thoracic aorta were characterized by cardiovascular magnetic resonance imaging in 208 subjects with TS in an IRB-approved natural history study. Echocardiography was used to measure peak velocities across the aortic valve, and the degree of aortic regurgitation. Four distinct valve morphologies were identified: tricuspid aortic valve (TAV) 64%(n=133), partially fused aortic valve (PF) 12%(n=25), bicuspid aortic valve (BAV) 23%(n=47), and unicuspid aortic valve (UAV) 1%(n=3). Age and body surface area (BSA) were similar in the 4 valve morphology groups. There was a significant trend, independent of age, towards larger BSA-indexed ascending aortic diameters (AADi) with increasing valve fusion. AADi were (mean +/− SD) 16.9 +/− 3.3 mm/m2, 18.3 +/− 3.3 mm/m2, and 19.8 +/− 3.9 mm/m2 (p<0.0001) for TAV, PF and BAV+UAV respectively. PF, BAV, and UAV were significantly associated with mild aortic regurgitation and elevated peak velocities across the aortic valve. Conclusions Aortic valve abnormalities in TS occur with a spectrum of severity, and are associated with aortic root dilation across age groups. Partial fusion of the aortic valve, traditionally regarded as an acquired valve problem, had an equal age distribution and was associated with an increased AADi. PMID:24084490
High pressure capillary micro-fluidic valve device and a method of fabricating same
Crocker, Robert W [Fremont, CA; Caton, Pamela F [Berkely, CA; Gerhardt, Geoff C [Milbury, MA
2007-04-17
A freeze-thaw valve and a method of micro-machining the freeze-thaw valve is provided and includes a valve housing, wherein the valve housing defines a housing cavity and includes a housing inlet, a housing vent, a capillary tubing inlet and a capillary tubing outlet. A valve body is provided, at least a portion of which is lithographically constructed, wherein the valve body includes a refrigerant inlet, a refrigerant outlet and an expansion chamber. The expansion chamber is disposed to communicate the refrigerant inlet with the refrigerant outlet and includes a restriction region having a flow restriction. Additionally, the valve body is disposed within the housing cavity to form an insulating channel between the valve housing and the valve body.
Shape Memory Actuated Normally Open Permanent Isolation Valve
NASA Technical Reports Server (NTRS)
Ramspacher, Daniel J. (Inventor); Bacha, Caitlin E. (Inventor)
2017-01-01
A valve assembly for an in-space propulsion system includes an inlet tube, an outlet tube, a valve body coupling the inlet tube to the outlet tube and defining a propellant flow path, a valve stem assembly disposed within the valve body, an actuator body coupled to the valve body, the valve stem assembly extending from an interior of the valve body to an interior of the actuator body, and an actuator assembly disposed within the actuator body and coupled to the valve stem assembly, the actuator assembly including a shape memory actuator member that when heated to a transition temperature is configured to enable the valve stem assembly to engage the outlet tube and seal the propellant flow path.
Chhatriwalla, Adnan K; Allen, Keith B; Saxon, John T; Cohen, David J; Aggarwal, Sanjeev; Hart, Anthony J; Baron, Suzanne J; Dvir, Danny; Borkon, A Michael
2017-07-01
Valve-in-valve (VIV) transcatheter aortic valve replacement (TAVR) may be less effective in small surgical valves because of patient/prosthesis mismatch. Bioprosthetic valve fracture (BVF) using a high-pressure balloon can be performed to facilitate VIV TAVR. We report data from 20 consecutive clinical cases in which BVF was successfully performed before or after VIV TAVR by inflation of a high-pressure balloon positioned across the valve ring during rapid ventricular pacing. Hemodynamic measurements and calculation of the valve effective orifice area were performed at baseline, immediately after VIV TAVR, and after BVF. BVF was successfully performed in 20 patients undergoing VIV TAVR with balloon-expandable (n=8) or self-expanding (n=12) transcatheter valves in Mitroflow, Carpentier-Edwards Perimount, Magna and Magna Ease, Biocor Epic and Biocor Epic Supra, and Mosaic surgical valves. Successful fracture was noted fluoroscopically when the waist of the balloon released and by a sudden drop in inflation pressure, often accompanied by an audible snap. BVF resulted in a reduction in the mean transvalvular gradient (from 20.5±7.4 to 6.7±3.7 mm Hg, P <0.001) and an increase in valve effective orifice area (from 1.0±0.4 to 1.8±0.6 cm 2 , P <0.001). No procedural complications were reported. BVF can be performed safely in small surgical valves to facilitate VIV TAVR with either balloon-expandable or self-expanding transcatheter valves and results in reduced residual transvalvular gradients and increased valve effective orifice area. © 2017 American Heart Association, Inc.
Pregnancy-induced remodeling of heart valves.
Pierlot, Caitlin M; Moeller, Andrew D; Lee, J Michael; Wells, Sarah M
2015-11-01
Recent studies have demonstrated remodeling of aortic and mitral valves leaflets under the volume loading and cardiac expansion of pregnancy. Those valves' leaflets enlarge with altered collagen fiber architecture, content, and cross-linking and biphasic changes (decreases, then increases) in extensibility during gestation. This study extends our analyses to right-sided valves, with additional compositional measurements for all valves. Valve leaflets were harvested from nonpregnant heifers and pregnant cows. Leaflet structure was characterized by leaflet dimensions, and ECM composition was determined using standard biochemical assays. Histological studies assessed changes in cellular and ECM components. Leaflet mechanical properties were assessed using equibiaxial mechanical testing. Collagen thermal stability and cross-linking were assessed using denaturation and hydrothermal isometric tension tests. Pulmonary and tricuspid leaflet areas increased during pregnancy by 35 and 55%, respectively. Leaflet thickness increased by 20% only in the pulmonary valve and largely in the fibrosa (30% thickening). Collagen crimp length was reduced in both the tricuspid (61%) and pulmonary (42%) valves, with loss of crimped area in the pulmonary valve. Thermomechanics showed decreased collagen thermal stability with surprisingly maintained cross-link maturity. The pulmonary leaflet exhibited the biphasic change in extensibility seen in left side valves, whereas the tricuspid leaflet mechanics remained largely unchanged throughout pregnancy. The tricuspid valve exhibits a remodeling response during pregnancy that is significantly diminished from the other three valves. All valves of the heart remodel in pregnancy in a manner distinct from cardiac pathology, with much similarity valve to valve, but with interesting valve-specific responses in the aortic and tricuspid valves. Copyright © 2015 the American Physiological Society.
Sarı, Cenk; Baştuğ, Serdal; Kasapkara, Hacı Ahmet; Durmaz, Tahir; Keleş, Telat; Akçay, Murat; Aslan, Abdullah Nabi; Bayram, Nihal Akar; Bozkurt, Engin
2015-01-01
Introduction Transcatheter aortic valve implantation for severe symptomatic aortic stenosis in patients with a previous mitral valve prosthesis is technically challenging, and pre-procedural comprehensive assessment of these patients before transcatheter aortic valve implantation is vital for an uncomplicated and successful procedure. Aim We want to share our experience with transcatheter aortic valve implantation in patients with a preexisting functional mitral valve prosthesis and describe a series of important technical and pre-procedural details. Material and methods At our center, 135 patients with symptomatic severe aortic stenosis were treated with transcatheter aortic valve implantation. Six of them with a preexisting mitral valve prosthesis received an Edwards SAPIEN XT valve through the transfemoral route. Results Transcatheter aortic valve implantation was performed successfully in all 6 patients without any deformation of the cobalt-chromium/steel stents of the aortic valve bioprosthesis. Also no distortion or malfunction in the mitral valve prosthesis was observed after the procedure. There were no complications during the hospitalization period. Post-procedural echocardiography revealed no or mild aortic paravalvular regurgitation and normal valve function in all the patients. In addition, serial echocardiographic examination demonstrated that both the stability and function of the aortic and mitral prosthetic valves were normal without any deterioration in the gradients and the degree of the regurgitation at long-term follow-ups. Conclusions Our experience confirms that transcatheter aortic valve implantation is technically feasible in patients with previous mitral valve replacement but comprehensive evaluation of patients by multimodal imaging techniques such as transesophageal echocardiography and multislice computed tomography is mandatory for a successful and safe procedure. PMID:26677380
Check valve installation in pilot operated relief valve prevents reverse pressurization
NASA Technical Reports Server (NTRS)
Oswalt, L.
1966-01-01
Two check valves prevent reverse flow through pilot-operated relief valves of differential area piston design. Title valves control pressure flow to ensure that the piston dome pressure is always at least as great as the main relief valve discharge pressure.
Gilchrist, I C; Cardella, J F; Fox, P S; Pae, W E; el-Ghamry Sabe, A A; Landis, J R; Localio, A R; Kunselman, A R; Hopper, K D
1997-02-01
Cineradiography can identify patients with single-leg fractured Björk-Shiley Convexo-Concave valves, although little is known about the sensitivity and specificity of this technique. We evaluated three normal and six (0 microm gap) single-leg fractured Björk-Shiley valves that were placed in a working phantom model. Valves were randomly imaged a total of 33 times and duplicated into a 120-valve series with a 1:9 ratio of abnormal/normal valves. Six reviewers independently graded each valve and demonstrated markedly different rates of identifying the fractured valves. Average sensitivity at the grade that clinically results in valve explanation was 47%. Among the normal valves, a correct identification was made 96% (range 91% to 99%) of the time. Present radiographic technology may have significant difficulty in identifying true single-leg fracture in Björk-Shiley valves with limb separations that are common among clinically explanted valves.
Mechanical valve replacement in congenital heart disease.
Fiane, A E; Lindberg, H L; Saatvedt, K; Svennevig, J L
1996-05-01
Mechanical valves are the prosthesis of choice in valve replacement in children. However, the problem of somatic growth leading to patient-valve mismatch remains present, and the appropriate anticoagulation regimen remains controversial. We present our experience of valve replacement in a young population over 20 years. Between 1972 and 1992, 48 patients (34 males and 14 females), mean age 11.2 years (range 0.4-27.4 years), underwent mechanical valve replacement at our institution. Aortic valve replacement was performed in 28 patients (58.3%), mitral valve replacement in 13 (27.1%), tricuspid valve replacement in six (12.5%) and pulmonary valve replacement in one patient (2.1%). The prostheses used were: St. Jude Medical (n = 2), Björk-Shiley (n = 14), Medtronic Hall (n = 16), Duromedics (n = 2) and CarboMedics (n = 14). Early mortality was 14.3%, 10.7% for aortic valve replacement and 30.8% for mitral valve replacement. Mean follow up for all patients was 8.3 years (range 0-22 years), with a total of 398 patient-years. Seven patients died during the follow up (17.1%). Survival after 10 years, including operative mortality, was 81% for aortic valve replacement, 33% for mitral valve replacement, 83% for tricuspid valve replacement and 100% for pulmonary valve replacement. All patients were anticoagulated with warfarin. In eight patients (16.7%) an antiplatelet drug (aspirin or dipyridamole) was added. Major events included paravalvular leak in six patients (1.5%/pty), valve thrombosis in five (mitral position in two, tricuspid in three) (1.3%/pty) and endocarditis in one patient (0.3%/pty). Minor thromboembolic events occurred in three patients (0.8%/pty) and minor hemorrhagic events in three (0.8%/pty). No patients developed hemolytic anemia and there was no case of structural failure. In our experience, mechanical prostheses in congenital heart disease were associated with significant morbidity and mortality, however long term survival after aortic valve replacement was good (81% at 10 years). Thromboembolic and hemorrhagic events were of minor significance. Atrio-ventricular valve replacement carried the highest risk of valve thrombosis and we now give warfarin and an antiplatelet drug to children undergoing mechanical valve implantation in this position.
Kogon, Brian; Mori, Makoto; Alsoufi, Bahaaldin; Kanter, Kirk; Oster, Matt
2015-06-01
Pulmonary valve disruption in patients with tetralogy of Fallot and congenital pulmonary stenosis often results in pulmonary insufficiency, right ventricular dilation, and tricuspid valve regurgitation. Management of functional tricuspid regurgitation at the time of subsequent pulmonary valve replacement remains controversial. Our aims were to (1) analyze tricuspid valve function after pulmonary valve replacement through midterm follow-up and (2) determine the benefits, if any, of concomitant tricuspid annuloplasty. Thirty-five patients with tetralogy of Fallot or congenital pulmonary stenosis were analyzed. All patients had been palliated in childhood by disrupting the pulmonary valve, and all patients had at least moderate tricuspid valve regurgitation at the time of subsequent pulmonary valve replacement. Preoperative and serial postoperative echocardiograms were analyzed. Pulmonary and tricuspid regurgitation, along with right ventricular dilation and dysfunction were scored as 0 (none), 1 (mild), 2 (moderate), and 3 (severe). Right ventricular volume and area were also calculated. Comparisons were made between patients who underwent pulmonary valve replacement alone and those who underwent concomitant tricuspid valve annuloplasty. At 1 month after pulmonary valve replacement, there were significant reductions in pulmonary valve regurgitation (mean 3 vs 0.39, p < 0.0001), tricuspid valve regurgitation (mean 2.33 vs 1.3, p < 0.0001), and in right ventricular dilation, volume, and area. There was no difference in the degree of tricuspid regurgitation 1 month postoperatively between patients who underwent concomitant tricuspid annuloplasty and those who underwent pulmonary valve replacement alone (mean 1.31 vs 1.29, p = 0.81). However, at latest follow-up (mean 7.0 ± 2.8 years), the degree of tricuspid regurgitation was significantly higher in the concomitant annuloplasty group (mean 1.87 vs 1.12, p = 0.005). In patients with at least moderate tricuspid valve regurgitation, significant improvement in tricuspid valve function and right ventricular size occurs in the first postoperative month after pulmonary valve replacement, irrespective of concomitant tricuspid valve annuloplasty. The tricuspid valve appears to function better over the midterm if annuloplasty is not performed. Copyright © 2015 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.
Shunt Devices for the Treatment of Adult Hydrocephalus: Recent Progress and Characteristics
MIYAKE, Hiroji
2016-01-01
Various types of shunt valves have been developed during the past 50 years, most of which can be classified into the following categories: (1) fixed differential pressure valves; (2) fixed differential pressure (DP) valves with an antisiphon mechanism; (3) programmable DP valves; (4) programmable DP valves with an antisiphon mechanism; and (5) programmable antisiphon valves. When considering the myriad of possible postoperative condition changes, such as the onset of accidental non-related diseases or trauma in adults, and changes in normal physiological development or anticipation of future shunt removal in children, it has become standard to use the programmable valve as a first choice for cerebrospinal fluid shunting. However, it is still unclear what type of shunt valve is suitable for each individual case. Based on the results of SINPHONI and more recently SINPHONI 2 trials, the programmable DP valve is recommended as the first line shunt valve. The programmable DP valve with an antisiphon mechanism is thought to be beneficial for tall, slender patients, who have a tendency for easily developing complications of overdrainage, however, this type of valve must be used cautiously in obese patients because of the increased risk of underdrainage. Although the current evidence is still insufficient, the programmable antisiphon valve, which costs the same as the programmable DP valve, is also thought to be the first line shunt valve. The quick reference table is applicable for most shunt valves, and for patients with either the ventriculoperitoneal or the lumboperitoneal shunt. PMID:27041631
Shunt Devices for the Treatment of Adult Hydrocephalus: Recent Progress and Characteristics.
Miyake, Hiroji
2016-05-15
Various types of shunt valves have been developed during the past 50 years, most of which can be classified into the following categories: (1) fixed differential pressure valves; (2) fixed differential pressure (DP) valves with an antisiphon mechanism; (3) programmable DP valves; (4) programmable DP valves with an antisiphon mechanism; and (5) programmable antisiphon valves. When considering the myriad of possible postoperative condition changes, such as the onset of accidental non-related diseases or trauma in adults, and changes in normal physiological development or anticipation of future shunt removal in children, it has become standard to use the programmable valve as a first choice for cerebrospinal fluid shunting. However, it is still unclear what type of shunt valve is suitable for each individual case. Based on the results of SINPHONI and more recently SINPHONI 2 trials, the programmable DP valve is recommended as the first line shunt valve. The programmable DP valve with an antisiphon mechanism is thought to be beneficial for tall, slender patients, who have a tendency for easily developing complications of overdrainage, however, this type of valve must be used cautiously in obese patients because of the increased risk of underdrainage. Although the current evidence is still insufficient, the programmable antisiphon valve, which costs the same as the programmable DP valve, is also thought to be the first line shunt valve. The quick reference table is applicable for most shunt valves, and for patients with either the ventriculoperitoneal or the lumboperitoneal shunt.
Observation of cavitation in a mechanical heart valve in a total artificial heart.
Lee, Hwansung; Tsukiya, Tomonori; Homma, Akihiko; Kamimura, Tadayuki; Takewa, Yoshiaki; Nishinaka, Tomohiro; Tatsumi, Eisuke; Taenaka, Yoshiyuki; Takano, Hisateru; Kitamura, Soichiro
2004-01-01
Recently, cavitation on the surface of mechanical heart valves has been studied as a cause of fractures occurring in implanted mechanical heart valves. The cause of cavitation in mechanical heart valves was investigated using the 25 mm Medtronic Hall valve and the 23 mm Omnicarbon valve. Closing of these valves in the mitral position was simulated in an electrohydraulic totally artificial heart. Tests were conducted under physiologic pressures at heart rates from 60 to 100 beats per minute with cardiac outputs from 4.8 to 7.7 L/min. The disk closing motion was measured by a laser displacement sensor. A high-speed video camera was used to observe the cavitation bubbles in the mechanical heart valves. The maximum closing velocity of the Omnicarbon valve was faster than that of the Medtronic Hall valve. In both valves, the closing velocity of the leaflet, used as the cavitation threshold, was approximately 1.3-1.5 m/s. In the case of the Medtronic Hall valve, cavitation bubbles were generated by the squeeze flow and by the effects of the venturi and the water hammer. With the Omnicarbon valve, the cavitation bubbles were generated by the squeeze flow and the water hammer. The mechanism leading to the development of cavitation bubbles depended on the valve closing velocity and the valve stop geometry. Most of the cavitation bubbles were observed around the valve stop and were generated by the squeeze flow.
NASA Technical Reports Server (NTRS)
Srinivasan, K. V.
1986-01-01
The design and development of a large diameter high pressure quick acting propulsion valve and valve actuator is described. The valve is the heart of a major test facility dedicated to conducting full scale performance tests of aircraft landing systems. The valve opens in less than 300 milliseconds releasing a 46-centimeter- (18-in.-) diameter water jet and closes in 300 milliseconds. The four main components of the valve, i.e., valve body, safety shutter, high speed shutter, and pneumatic-hydraulic actuator, are discussed. This valve is unique and may have other aerospace and industrial applications.
Flow diverter value and flow diversion method
NASA Technical Reports Server (NTRS)
Arline, S. B.; Carlson, R. L. (Inventor)
1964-01-01
A flow diverter valve applicable to any fluid flow system requiring rapid bleed or bypass is disclosed. Examples of application of the flow diverter valve to a liquid rocket and a turbojet aircraft engine are given. Features of the valve include: (1) an independent fluid source is used to activate the flow diverter valve toward its closed position during its initial stage of travel; (2) the flow diverter port area and size is unlimited and the valve travel is unlimited; and (3) the valve housing is fabricated such that the valve can be a one step valve, a two step valve, or include as many steps as are found desirable.
NASA Technical Reports Server (NTRS)
Srinivasan, K. V.
1986-01-01
This paper describes the design and development of a large diameter high pressure quick acting propulsion valve and valve actuator. The valve is the heart of a major test facility dedicated to conducting full scale performance tests of aircraft landing gear systems. The valve opens in less than 300 milliseconds releasing a 46 cm (18 in) diameter water jet and closes in 300 milliseconds. The four main components of the valve, i.e., valve body, safety shutter, high speed shutter, and pneumatic-hydraulic actuator, are discussed. This valve is unique and may have other aerospace and industrial applications.
DeWall, Kevin G.; Watkins, John C; Nitzel, Michael E.
2006-08-29
Apparatus for actuating a valve includes a support frame and at least one valve driving linkage arm, one end of which is rotatably connected to a valve stem of the valve and the other end of which is rotatably connected to a screw block. A motor connected to the frame is operatively connected to a motor driven shaft which is in threaded screw driving relationship with the screw block. The motor rotates the motor driven shaft which drives translational movement of the screw block which drives rotatable movement of the valve driving linkage arm which drives translational movement of the valve stem. The valve actuator may further include a sensory control element disposed in operative relationship with the valve stem, the sensory control element being adapted to provide control over the position of the valve stem by at least sensing the travel and/or position of the valve stem.
Fox, P R; Miller, M W; Liu, S K
1992-11-15
Mitral stenosis was diagnosed noninvasively by echocardiography and Doppler imaging in 2 Bull Terriers. Two-dimensional echocardiography revealed severe atrial and moderate left ventricular dilatation; severely reduced mitral valve opening excursion; doming of the cranial mitral valve leaflet into the left ventricle during diastole; thickened, nodular cranial mitral valve leaflets; and reduced mitral valve orifice. M-mode echocardiographic findings additionally indicated greatly diminished mitral valve E to F slope and abnormal caudal mitral valve leaflet motion. Color flow Doppler imaging revealed bright bursts of color with aliasing originating from the stenotic mitral valve orifice, extending into the left atrium during systole, and into the left atrium during diastole. Spectral Doppler recordings revealed transvalvular mitral valve gradients and prolonged pressure half-times. Necropsy performed on 1 dog revealed extremely thickened, nodular, and stiff mitral valves with short, thickened, and fused chordae tendineae. The diagnosis of mitral valve stenosis was easily facilitated with diagnostic ultrasonography.
Method, apparatus and system for controlling fluid flow
McMurtrey, Ryan D.; Ginosar, Daniel M.; Burch, Joesph V.
2007-10-30
A system, apparatus and method of controlling the flow of a fluid are provided. In accordance with one embodiment of the present invention, a flow control device includes a valve having a flow path defined therethrough and a valve seat in communication with the flow path with a valve stem disposed in the valve seat. The valve stem and valve seat are cooperatively configured to cause mutual relative linear displacement thereof in response to rotation of the valve stem. A gear member is coupled with the rotary stem and a linear positioning member includes a portion which complementarily engages the gear member. Upon displacement of the linear positioning member along a first axis, the gear member and rotary valve stem are rotated about a second axis and the valve stem and valve seat are mutually linearly displaced to alter the flow of fluid through the valve.
FLUID PRESSURE AND CAM OPERATED VACUUM VALVE
Batzer, T.H.
1963-11-26
An ultra-high vacuum valve that is bakable, reusable, and capable of being quickly opened and closed is described. A translationally movable valve gate having an annular ridge is adapted to contact an annular soft metal gasket disposed at the valve seat such that the soft metal gasket extends beyond the annular ridge on all sides. The valve gate is closed, by first laterally aligning the valve gate with the valve seat and then bringing the valve gate and valve seat into seating contact by the translational movement of a ramp-like wedging means that engages similar ramp-like stractures at the base of the valve gate to force the valve gate into essentially pressureless contact with the annular soft metal gasket. This gasket is then pressurized from beneath by a fluid thereby effecting a vacuura tight seal between the gasket and the ridge. (AEC)
Dysfunction of an On-X Heart Valve by Pannus.
Abad, Cipriano; Urso, Stefano; Gomez, Elsa; De la Vega, Maria
2016-09-01
A 68-year-old woman with a history of previous double-valve replacement with On-X mechanical heart valves presented with clinical, echocardiographic and cardiac catheterization signs of obstruction of the On-X tricuspid heart valve prosthesis. The patient was successfully reoperated, but at surgery the valve was seen to be invaded by an abnormal overgrowth of pannus that blocked one of the leaflets. A small amount of non-obstructive fresh thrombus was also observed. The valve was successfully replaced with a biological heart valve prosthesis. The patient was discharged home, and is doing well four months after the operation, when echocardiography demonstrated normal function in the tricuspid valve. The present case represents the first ever report of pannus formation and subsequent dysfunction in an On-X heart valve, and also the first case of tricuspid valve malfunction and obstruction using this type of heart valve substitute.
Zhuang, H D; Zhang, X D
2015-05-01
A fast valve based on the double-layer eddy-current repulsion mechanism has been developed on Experimental Advanced Superconducting Tokamak (EAST). In addition to a double-layer eddy-current coil, a preload system was added to improve the security of the valve, whereby the valve opens more quickly and the open-valve time becomes shorter, making it much safer than before. In this contribution, testing platforms, open-valve characteristics, and throughput of the fast valve are discussed. Tests revealed that by choosing appropriate parameters the valve opened within 0.15 ms, and open-valve times were no longer than 2 ms. By adjusting working parameter values, the maximum number of particles injected during this open-valve time was estimated at 7 × 10(22). The fast valve will become a useful tool to further explore disruption mitigation experiments on EAST in 2015.
How to Make a Heart Valve: From Embryonic Development to Bioengineering of Living Valve Substitutes
MacGrogan, Donal; Luxán, Guillermo; Driessen-Mol, Anita; Bouten, Carlijn; Baaijens, Frank; de la Pompa, José Luis
2014-01-01
Cardiac valve disease is a significant cause of ill health and death worldwide, and valve replacement remains one of the most common cardiac interventions in high-income economies. Despite major advances in surgical treatment, long-term therapy remains inadequate because none of the current valve substitutes have the potential for remodeling, regeneration, and growth of native structures. Valve development is coordinated by a complex interplay of signaling pathways and environmental cues that cause disease when perturbed. Cardiac valves develop from endocardial cushions that become populated by valve precursor mesenchyme formed by an epithelial–mesenchymal transition (EMT). The mesenchymal precursors, subsequently, undergo directed growth, characterized by cellular compartmentalization and layering of a structured extracellular matrix (ECM). Knowledge gained from research into the development of cardiac valves is driving exploration into valve biomechanics and tissue engineering directed at creating novel valve substitutes endowed with native form and function. PMID:25368013
Solenoid Valve With Self-Compensation
NASA Technical Reports Server (NTRS)
Woeller, Fritz H.; Matsumoto, Yutaka
1987-01-01
New solenoid-operated miniature shutoff valve provides self-compensation of differential pressure forces that cause jamming or insufficient valve closure as in single-seal valves. Dual-seal valve is bidirectional. Valve simultaneously seals both inlet and outlet tubes by pressing single disk of silicone rubber against ends of both.
NASA Technical Reports Server (NTRS)
Brown, Andrew M.; Mulder, Andrew
2017-01-01
NASA is developing a new launch vehicle, called the Space Launch System (SLS), which is intended on taking humans out of low earth orbit to destinations including the moon, asteroids, and Mars. The propulsion system for the core stage of this vehicle includes four RS-25 Liquid Hydrogen/Oxygen rocket engines. These engines are upgraded versions of the Space Shuttle Main Engines (SSME); the upgrades include higher power levels and affordability enhancements. As with any new vehicle, the Main Propulsion System (MPS), which include the feedlines and ancillary hardware connecting the engines to the fuel and oxidizer tanks, had to be redesigned (figure 1 - export clearance in progress), as the previous MPS for the SSME's was inherently part of the Space Shuttle System, which had a completely different overall configuration.
Optical, Electrical and Magnetic Studies of Pi-Conjugated Organic Semiconductor Systems
DOE Office of Scientific and Technical Information (OSTI.GOV)
Vardeny, Zeev Valentine
Over the duration of this grant our group has studied the transient and cw optical response of various π-conjugated polymers, oligomers, single crystals, fullerene molecules and blends of organic donor-acceptor molecules. We have been also involved in complementary experiments such as magneto-optical studies and spin-physics. We have advanced the field of photophysics of these materials by providing information on their excited state energies and primodal and long-lived photoexcitations such as singlet excitons, triplet excitons, polaron-pairs, excimers and exciplexes. We also fabricated various organic optoelectronic devices such as organic light emitting diodes (OLED), electrochemical cells, organic diodes, organic spin-valves (OSV), andmore » organic photovoltaic (OPV) solar cells. These devices benefited the society in terms of cheap and energy saving illumination, as well as harnessing the solar energy.« less
Scanlan, Adam B; Nguyen, Alex V; Ilina, Anna; Lasso, Andras; Cripe, Linnea; Jegatheeswaran, Anusha; Silvestro, Elizabeth; McGowan, Francis X; Mascio, Christopher E; Fuller, Stephanie; Spray, Thomas L; Cohen, Meryl S; Fichtinger, Gabor; Jolley, Matthew A
2018-03-01
Mastering the technical skills required to perform pediatric cardiac valve surgery is challenging in part due to limited opportunity for practice. Transformation of 3D echocardiographic (echo) images of congenitally abnormal heart valves to realistic physical models could allow patient-specific simulation of surgical valve repair. We compared materials, processes, and costs for 3D printing and molding of patient-specific models for visualization and surgical simulation of congenitally abnormal heart valves. Pediatric atrioventricular valves (mitral, tricuspid, and common atrioventricular valve) were modeled from transthoracic 3D echo images using semi-automated methods implemented as custom modules in 3D Slicer. Valve models were then both 3D printed in soft materials and molded in silicone using 3D printed "negative" molds. Using pre-defined assessment criteria, valve models were evaluated by congenital cardiac surgeons to determine suitability for simulation. Surgeon assessment indicated that the molded valves had superior material properties for the purposes of simulation compared to directly printed valves (p < 0.01). Patient-specific, 3D echo-derived molded valves are a step toward realistic simulation of complex valve repairs but require more time and labor to create than directly printed models. Patient-specific simulation of valve repair in children using such models may be useful for surgical training and simulation of complex congenital cases.
Scotten, Lawrence N; Siegel, Rolland
2015-08-01
Exploration for causes of prosthetic valve thrombogenicity has frequently focused on forward or post-closure flow detail. In prior laboratory studies, we uncovered high amplitude flow velocities of short duration close to valve closure implying potential for substantial shear stress with subsequent initiation of blood coagulation pathways. This may be relevant to widely accepted clinical disparity between mechanical and tissue valves vis-à-vis thrombogenicity. With a series of prototype bi-leaflet mechanical valves, we attempt reduction of closure related velocities with the objective of identifying a prototype valve with thrombogenic potential similar to our tissue valve control. This iterative design approach may find application in preclinical assessment of valves for anticoagulation independence. Tested valves included: prototype mechanical bi-leaflet BVs (n=56), controls (n=2) and patented early prototype mechanicals (n=2) from other investigators. Pulsatile and quasi-steady flow systems were used for testing. Projected dynamic valve area (PDVA) was measured using previously described novel technology. Flow velocity over the open and closing periods was determined by volumetric flow rate/PDVA. For the closed valve interval, use was made of data obtained from quasi-steady back pressure/flow tests. Performance was ranked by a proposed thrombogenicity potential index (TPI) relative to tissue and mechanical control valves. Optimization of the prototype valve designs lead to a 3-D printed model (BV3D). For the mitral/aortic site, BV3D has lower TPI (1.10/1.47) relative to the control mechanical valve (3.44/3.93) and similar to the control tissue valve (ideal TPI ≤1.0). Using unique technology, rapid prototyping and thrombogenicity ranking, optimization of experimental valves for reduced thrombogenic potential was expedited and simplified. Innovative mechanical valve configurations were identified that merit consideration for further development which may bring the anti-coagulation independent mechanical valve within reach.
Automatic Mesh Generation of Hybrid Mesh on Valves in Multiple Positions in Feedline Systems
NASA Technical Reports Server (NTRS)
Ross, Douglass H.; Ito, Yasushi; Dorothy, Fredric W.; Shih, Alan M.; Peugeot, John
2010-01-01
Fluid flow simulations through a valve often require evaluation of the valve in multiple opening positions. A mesh has to be generated for the valve for each position and compounding. The problem is the fact that the valve is typically part of a larger feedline system. In this paper, we propose to develop a system to create meshes for feedline systems with parametrically controlled valve openings. Herein we outline two approaches to generate the meshes for a valve in a feedline system at multiple positions. There are two issues that must be addressed. The first is the creation of the mesh on the valve for multiple positions. The second is the generation of the mesh for the total feedline system including the valve. For generation of the mesh on the valve, we will describe the use of topology matching and mesh generation parameter transfer. For generation of the total feedline system, we will describe two solutions that we have implemented. In both cases the valve is treated as a component in the feedline system. In the first method the geometry of the valve in the feedline system is replaced with a valve at a different opening position. Geometry is created to connect the valve to the feedline system. Then topology for the valve is created and the portion of the topology for the valve is topology matched to the standard valve in a different position. The mesh generation parameters are transferred and then the volume mesh for the whole feedline system is generated. The second method enables the user to generate the volume mesh on the valve in multiple open positions external to the feedline system, to insert it into the volume mesh of the feedline system, and to reduce the amount of computer time required for mesh generation because only two small volume meshes connecting the valve to the feedline mesh need to be updated.
Xu, Tong-yi; Zhang, Zhi-gang; Li, Xin; Han, Lin; Xu, Zhi-yun
2014-01-01
Background Since 2000, transcatheter pulmonary valve replacement has steadily advanced. However, the available prosthetic valves are restricted to bioprosthesis which have defects like poor durability. Polymeric heart valve is thought as a promising alternative to bioprosthesis. In this study, we introduced a novel polymeric transcatheter pulmonary valve and evaluated its feasibility and safety in sheep by a hybrid approach. Methods We designed a novel polymeric trileaflet transcatheter pulmonary valve with a balloon-expandable stent, and the valve leaflets were made of 0.1-mm expanded polytetrafluoroethylene (ePTFE) coated with phosphorylcholine. We chose glutaraldehyde-treated bovine pericardium valves as control. Pulmonary valve stents were implanted in situ by a hybrid transapical approach in 10 healthy sheep (8 for polymeric valve and 2 for bovine pericardium valve), weighing an average of 22.5±2.0 kg. Angiography and cardiac catheter examination were performed after implantation to assess immediate valvular functionality. After 4-week follow-up, angiography, echocardiography, computed tomography, and cardiac catheter examination were used to assess early valvular function. One randomly selected sheep with polymeric valve was euthanized and the explanted valved stent was analyzed macroscopically and microscopically. Findings Implantation was successful in 9 sheep. Angiography at implantation showed all 9 prosthetic valves demonstrated orthotopic position and normal functionality. All 9 sheep survived at 4-week follow-up. Four-week follow-up revealed no evidence of valve stent dislocation or deformation and normal valvular and cardiac functionality. The cardiac catheter examination showed the peak-peak transvalvular pressure gradient of the polymeric valves was 11.9±5.0 mmHg, while that of two bovine pericardium valves were 11 and 17 mmHg. Gross morphology demonstrated good opening and closure characteristics. No thrombus or calcification was seen macroscopically. Conclusions This design of the novel ePTFE transcatheter pulmonary valve is safe and effective to deploy in sheep by hybrid approach, and the early valvular functionality is good. PMID:24926892
Amstrup Funder, Jonas; Christian Danielsen, Carl; Baandrup, Ulrik; Martin Bibby, Bo; Carl Andelius, Ted; Toft Brøndum, Emil; Wang, Tobias; Michael Hasenkam, J
2017-01-01
Heart valves which exist naturally in an extreme-pressure system must have evolved in a way to resist the stresses of high pressure. Giraffes are interesting as they naturally have a blood pressure twice that of humans. Thus, knowledge regarding giraffe heart valves may aid in developing techniques to design improved pressure-resistant biological heart valves. Heart valves from 12 giraffes and 10 calves were explanted and subjected to either biomechanical or morphological examinations. Strips from the heart valves were subjected to cyclic loading tests, followed by failure tests. Thickness measurements and analyses of elastin and collagen content were also made. Valve specimens were stained with hematoxylin and eosin, elastic van Gieson stain, Masson's trichrome and Fraser-Lendrum stain, as well as immunohistochemical reactions for morphological examinations. The aortic valve was shown to be 70% (95% CI 42-103%) stronger in the giraffe than in its bovine counterpart (p <0.001). No significant difference was found between mitral or pulmonary valves. After normalization for collagen, no significant differences were found in strength between species. The giraffe aortic valve was found to be significantly stiffer than the bovine aortic valve (p <0.001), with no significant difference between mitral and pulmonary valves. On a dry weight basis, the aortic (10.9%), pulmonary (4.3%), and mitral valves (9.6%) of giraffes contained significantly more collagen than those of calves. The elastin contents of the pulmonary valves (2.5%) and aortic valves (1.5%) were also higher in giraffes. The greater strength of the giraffe aortic valve is most likely due to a compact collagen construction. Both, collagen and elastin contents were higher in giraffes than in calves, which would make giraffe valves more resistant to the high-pressure forces. However, collagen also stiffens and thickens the valves. The mitral leaflets showed similar (but mostly insignificant) trends in strength, stiffness, and collagen content.
Siegel, Rolland
2015-01-01
Background Exploration for causes of prosthetic valve thrombogenicity has frequently focused on forward or post-closure flow detail. In prior laboratory studies, we uncovered high amplitude flow velocities of short duration close to valve closure implying potential for substantial shear stress with subsequent initiation of blood coagulation pathways. This may be relevant to widely accepted clinical disparity between mechanical and tissue valves vis-à-vis thrombogenicity. With a series of prototype bi-leaflet mechanical valves, we attempt reduction of closure related velocities with the objective of identifying a prototype valve with thrombogenic potential similar to our tissue valve control. This iterative design approach may find application in preclinical assessment of valves for anticoagulation independence. Methods Tested valves included: prototype mechanical bi-leaflet BVs (n=56), controls (n=2) and patented early prototype mechanicals (n=2) from other investigators. Pulsatile and quasi-steady flow systems were used for testing. Projected dynamic valve area (PDVA) was measured using previously described novel technology. Flow velocity over the open and closing periods was determined by volumetric flow rate/PDVA. For the closed valve interval, use was made of data obtained from quasi-steady back pressure/flow tests. Performance was ranked by a proposed thrombogenicity potential index (TPI) relative to tissue and mechanical control valves. Results Optimization of the prototype valve designs lead to a 3-D printed model (BV3D). For the mitral/aortic site, BV3D has lower TPI (1.10/1.47) relative to the control mechanical valve (3.44/3.93) and similar to the control tissue valve (ideal TPI ≤1.0). Conclusions Using unique technology, rapid prototyping and thrombogenicity ranking, optimization of experimental valves for reduced thrombogenic potential was expedited and simplified. Innovative mechanical valve configurations were identified that merit consideration for further development which may bring the anti-coagulation independent mechanical valve within reach. PMID:26417581
Feasibility study of inlet shock stability system of YF-12
NASA Technical Reports Server (NTRS)
Blausey, G. C.; Coleman, D. M.; Harp, D. S.
1972-01-01
The feasibility of self actuating bleed valves as a shock stabilization system in the inlet of the YF-12 is considered for vortex valves, slide valves, and poppet valves. Analytical estimation of valve performance indicates that only the slide and poppet valves located in the inlet cowl can meet the desired steady state stabilizing flows, and of the two the poppet valve is substantially faster in response to dynamic disturbances. The poppet valve is, therefore, selected as the best shock stability system for the YF-12 inlet.
NASA Astrophysics Data System (ADS)
Hopfgartner, J.; Posch, S.; Zuber, B.; Almbauer, R.; Krischan, K.; Stangl, S.
2017-08-01
Reed valves are widely used in hermetic reciprocating compressors and are responsible for a large part of the thermodynamic losses. Especially, the suction valve, which is opened nearly during the whole suction stroke, has a big potential for improvement. Usually, suction valves are opened only by vacuum created by the moving piston and should be closed before the compression stroke starts to avoid a reversed mass-flow through the valve. Therefore, the valves are prestressed, which results on the other hand in a higher flow resistance. In this work, a suction valve is investigated, which is not closed by the preload of the valve but by an electromagnetic coil located in the suction muffler neck. Shortly before the piston reaches its bottom dead centre, voltage is applied to the coil and a magnetic force is generated which pulls the valve shut. Thereby, the flow resistance through the valve can be reduced by changing the preload on the reed valve because it is no longer needed to close the valve. The investigation of this adapted valve and the electromagnetic coil is firstly done by numerical simulations including fluid structure interactions of the reed valves of a reciprocating compressor and secondly by experiments made on a calorimeter test bench.
Neethling, W M; Papadimitriou, J M; Swarts, E; Hodge, A J
2000-06-01
Valve related factors and patient related factors are responsible for calcification of valvular bioprostheses. Recent studies showed different donor and recipient species have different influences on the total calcification rate of bioprostheses. This study was performed to evaluate and compare Kangaroo aortic valve leaflets with porcine aortic valve leaflets. Experimental design. Prospective study. Setting. Cardio-thoracic experimental research of a university department. Glutaraldehyde-fixed Kangaroo and porcine valve leaflets were evaluated in vitro according to valve geometry (internal diameter and leaflet thickness), morphology (light and electron microscopy) and tensile strength. In vivo evaluation consisted of implantation in a rat model for 8 weeks, Von Kossa stain for calcium and atomic absorption spectrophotometry for total extractable calcium content. Kangaroo valves indicated a smaller internal valve diameter as well as a thinner valve leaflet (p<0.01, ANOVA) at corresponding body weight, less proteoglycan spicules in the fibrosa, increased elasticity (p<0.05) and low calcification potential (p<0.01, confidence interval 95%). Kangaroo aortic valve leaflets have different valvular qualities compared to porcine valve tissue. Kangaroo valve leaflets are significantly superior to porcine valve leaflets as far as calcification is concerned. These results are encouraging and suggest further in vivo evaluation in a larger animal model before clinical application can be considered.
Xuan, Yue; Wang, Zhongjie; Liu, Raymond; Haraldsson, Henrik; Hope, Michael D; Saloner, David A; Guccione, Julius M; Ge, Liang; Tseng, Elaine
2018-03-08
Guidelines for repair of bicuspid aortic valve-associated ascending thoracic aortic aneurysms have been changing, most recently to the same criteria as tricuspid aortic valve-ascending thoracic aortic aneurysms. Rupture/dissection occurs when wall stress exceeds wall strength. Recent studies suggest similar strength of bicuspid aortic valve versus tricuspid aortic valve-ascending thoracic aortic aneurysms; thus, comparative wall stress may better predict dissection in bicuspid aortic valve versus tricuspid aortic valve-ascending thoracic aortic aneurysms. Our aim was to determine whether bicuspid aortic valve-ascending thoracic aortic aneurysms had higher wall stresses than their tricuspid aortic valve counterparts. Patients with bicuspid aortic valve- and tricuspid aortic valve-ascending thoracic aortic aneurysms (bicuspid aortic valve = 17, tricuspid aortic valve = 19) greater than 4.5 cm underwent electrocardiogram-gated computed tomography angiography. Patient-specific 3-dimensional geometry was reconstructed and loaded to systemic pressure after accounting for prestress geometry. Finite element analyses were performed using the LS-DYNA solver (LSTC Inc, Livermore, Calif) with user-defined fiber-embedded material model to determine ascending thoracic aortic aneurysm wall stress. Bicuspid aortic valve-ascending thoracic aortic aneurysms 99th-percentile longitudinal stresses were 280 kPa versus 242 kPa (P = .028) for tricuspid aortic valve-ascending thoracic aortic aneurysms in systole. These stresses did not correlate to diameter for bicuspid aortic valve-ascending thoracic aortic aneurysms (r = -0.004) but had better correlation to tricuspid aortic valve-ascending thoracic aortic aneurysms diameter (r = 0.677). Longitudinal stresses on sinotubular junction were significantly higher in bicuspid aortic valve-ascending thoracic aortic aneurysms than in tricuspid aortic valve-ascending thoracic aortic aneurysms (405 vs 329 kPa, P = .023). Bicuspid aortic valve-ascending thoracic aortic aneurysm 99th-percentile circumferential stresses were 548 kPa versus 462 kPa (P = .033) for tricuspid aortic valve-ascending thoracic aortic aneurysms, which also did not correlate to bicuspid aortic valve-ascending thoracic aortic aneurysm diameter (r = 0.007). Circumferential and longitudinal stresses were greater in bicuspid aortic valve- than tricuspid aortic valve-ascending thoracic aortic aneurysms and were more pronounced in the sinotubular junction. Peak wall stress did not correlate with bicuspid aortic valve-ascending thoracic aortic aneurysm diameter, suggesting diameter alone in this population may be a poor predictor of dissection risk. Our results highlight the need for patient-specific aneurysm wall stress analysis for accurate dissection risk prediction. Copyright © 2018 The American Association for Thoracic Surgery. Published by Elsevier Inc. All rights reserved.
Muneretto, Claudio; Alfieri, Ottavio; Cesana, Bruno Mario; Bisleri, Gianluigi; De Bonis, Michele; Di Bartolomeo, Roberto; Savini, Carlo; Folesani, Gianluca; Di Bacco, Lorenzo; Rambaldini, Manfredo; Maureira, Juan Pablo; Laborde, Francois; Tespili, Maurizio; Repossini, Alberto; Folliguet, Thierry
2015-12-01
We sought to investigate the clinical outcomes of patients with isolated severe aortic stenosis and an intermediate- to high-risk profile treated by means of conventional surgery (surgical aortic valve replacement), sutureless valve implantation, or transcatheter aortic valve replacement in a multicenter evaluation. Among 991 consecutive patients with isolated severe aortic stenosis and an intermediate- to high-risk profile (Society of Thoracic Surgeons score >4 and logistic European System for Cardiac Operative Risk Evaluation I >10), a propensity score analysis was performed on the basis of the therapeutic strategy: surgical aortic valve replacement (n = 204), sutureless valve implantation (n = 204), and transcatheter aortic valve replacement (n = 204). Primary end points were 30-day mortality and overall survival at 24-month follow-up; the secondary end point was survival free from a composite end point of major adverse cardiac events (defined as cardiac-related mortality, myocardial infarction, cerebrovascular accidents, and major hemorrhagic events) and periprosthetic regurgitation greater than 2. Thirty-day mortality was significantly higher in the transcatheter aortic valve replacement group (surgical aortic valve replacement = 3.4% vs sutureless = 5.8% vs transcatheter aortic valve replacement = 9.8%; P = .005). The incidence of postprocedural was 3.9% in asurgical aortic valve replacement vs 9.8% in sutureless vs 14.7% in transcatheter aortic valve replacement (P< .001) and peripheral vascular complications occurred in 0% of surgicalaortic valve replacement vs 0% of sutureless vs 9.8% transcatheter aortic valve replacement (P< .001). At 24-month follow-up, overall survival (surgical aortic valve replacement = 91.3% ± 2.4% vs sutureless = 94.9% ± 2.1% vs transcatheter aortic valve replacement = 79.5% ± 4.3%; P < .001) and survival free from the composite end point of major adverse cardiovascular events and periprosthetic regurgitation were significantly better in patients undergoing surgical aortic valve replacement and sutureless valve implantation than in patients undergoing transcatheter aortic valve replacement (surgical aortic valve replacement = 92.6% ± 2.3% vs sutureless = 96% ± 1.8% vs transcatheter aortic valve replacement = 77.1% ± 4.2%; P < .001). Multivariate Cox regression analysis identified transcatheter aortic valve replacement as an independent risk factor for overall mortality hazard ratio (hazard ratio, 2.5; confidence interval, 1.1-4.2; P = .018). The use of transcatheter aortic valve replacement in patients with an intermediate- to high-risk profile was associated with a significantly higher incidence of perioperative complications and decreased survival at short- and mid-term when compared with conventional surgery and sutureless valve implantation. Copyright © 2015 The American Association for Thoracic Surgery. Published by Elsevier Inc. All rights reserved.
Globe stability during simulated vitrectomy with valved and non-valved trocar cannulas
Abulon, Dina Joy; Charles, Martin; Charles, Daniel E
2015-01-01
Purpose To compare the effects of valved and non-valved cannulas on intraocular pressure (IOP), fluid leakage, and vitreous incarceration during simulated vitrectomy. Methods Three-port pars plana incisions were generated in six rubber eyes using 23-, 25-, and 27-gauge valved and non-valved trocar cannulas. The models were filled with air and IOP was measured. Similar procedures were followed for 36 acrylic eyes filled with saline solution. Vitreous incarceration was analyzed in eleven rabbit and twelve porcine cadaver eyes. Results In the air-filled model, IOP loss was 89%–94% when two non-valved cannulas were unoccupied versus 1%–5% when two valved cannulas were unoccupied. In the fluid-filled model, with non-valved cannulas, IOP dropped while fluid leaked from the open ports. With two open ports, the IOP dropped to 20%–30% of set infusion pressure, regardless of infusion pressure and IOP compensation. The IOP was maintained in valved cannulas when one or two ports were left open, regardless of IOP compensation settings. There was no or minimal fluid leakage through open ports at any infusion pressure. Direct microscopic analysis of rabbit eyes showed that vitreous incarceration was significantly greater with 23-gauge non-valved than valved cannulas (P<0.005), and endoscopy of porcine eyes showed that vitreous incarceration was significantly greater with 23-gauge (P<0.05) and 27-gauge (P<0.05) non-valved cannulas. External observation of rabbit eyes showed vitreous prolapse through non-valved, but not valved, cannulas. Conclusion Valved cannulas surpassed non-valved cannulas in maintaining IOP, preventing fluid leakage, and reducing vitreous incarceration during simulated vitrectomy. PMID:26445520
Balaras, Elias; Cha, K S; Griffith, Bartley P; Gammie, James S
2009-03-01
Aortic valve bypass surgery treats aortic valve stenosis with a valve-containing conduit that connects the left ventricular apex to the descending thoracic aorta. After aortic valve bypass, blood is ejected from the left ventricle via both the native stenotic aortic valve and the conduit. We performed computational modeling to determine the effects of aortic valve bypass on aortic and cerebral blood flow, as well as the effect of conduit size on relative blood flow through the conduit and the native valve. The interaction of blood flow with the vascular boundary was modeled using a hybrid Eurelian-Lagrangian formulation, where an unstructured Galerkin finite element method was coupled with an immersed boundary approach. Our model predicted native (stenotic) valve to conduit flow ratios of 45:55, 52:48, and 60:40 for conduits with diameters of 20, 16, and 10 mm, respectively. Mean gradients across the native aortic valve were calculated to be 12.5, 13.8, and 17.6 mm Hg, respectively. Post-aortic valve bypass cerebral blood flow was unchanged from preoperative aortic valve stenosis configurations and was constant across all conduit sizes. In all cases modeled, cerebral blood flow was completely supplied by blood ejected across the native aortic valve. An aortic valve bypass conduit as small as 10 mm results in excellent relief of left ventricular outflow tract obstruction in critical aortic valve stenosis. The presence of an aortic valve bypass conduit has no effect on cerebral blood flow. All blood flow to the brain occurs via antegrade flow across the native stenotic valve; this configuration may decrease the long-term risk of cerebral thromboembolism.
Valve thrombosis following transcatheter aortic valve implantation: a systematic review.
Córdoba-Soriano, Juan G; Puri, Rishi; Amat-Santos, Ignacio; Ribeiro, Henrique B; Abdul-Jawad Altisent, Omar; del Trigo, María; Paradis, Jean-Michel; Dumont, Eric; Urena, Marina; Rodés-Cabau, Josep
2015-03-01
Despite the rapid global uptake of transcatheter aortic valve implantation, valve trombosis has yet to be systematically evaluated in this field. The aim of this study was to determine the clinical characteristics, diagnostic criteria, and treatment outcomes of patients diagnosed with valve thrombosis following transcatheter aortic valve implantation through a systematic review of published data. Literature published between 2002 and 2012 on valve thrombosis as a complication of transcatheter aortic valve implantation was identified through a systematic electronic search. A total of 11 publications were identified, describing 16 patients (mean age, 80 [5] years, 65% men). All but 1 patient (94%) received a balloon-expandable valve. All patients received dual antiplatelet therapy immediately following the procedure and continued to take either mono- or dual antiplatelet therapy at the time of valve thrombosis diagnosis. Valve thrombosis was diagnosed at a median of 6 months post-procedure, with progressive dyspnea being the most common symptom. A significant increase in transvalvular gradient (from 10 [4] to 40 [12] mmHg) was the most common echocardiographic feature, in addition to leaflet thickening. Thrombus was not directly visualized with echocardiography. Three patients underwent valve explantation, and the remaining received warfarin, which effectively restored the mean transvalvular gradient to baseline within 2 months. Systemic embolism was not a feature of valve thrombosis post-transcatheter aortic valve implantation. Although a rare, yet likely under-reported complication of post-transcatheter aortic valve implantation, progressive dyspnea coupled with an increasing transvalvular gradient on echocardiography within the months following the intervention likely signifies valve thrombosis. While direct thrombus visualization appears difficult, prompt initiation of oral anticoagulation therapy effectively restores baseline valve function. Copyright © 2014 Sociedad Española de Cardiología. Published by Elsevier España, S.L.U. All rights reserved.
NASA Technical Reports Server (NTRS)
Burgess, Robert K.; Yakos, David; Walthall, Bryan
2012-01-01
This invention utilizes a new method of opening and closing a ball valve. Instead of rotating the ball with a perpendicular stem (as is the case with standard ball valves), the ball is rotated around a fixed axis by two guide pins. This innovation eliminates the leak point that is present in all standard ball valves due to the penetration of an actuation stem through the valve body. The VOST (Venturi Off-Set-Technology) valve has been developed for commercial applications. The standard version of the valve consists of an off-set venturi flow path through the valve. This path is split at the narrowest portion of the venturi, allowing the section upstream from the venturi to be rotated. As this rotation takes place, the venturi becomes restricted as one face rotates with respect to the other, eventually closing off the flow path. A spring-loaded seal made of resilient material is embedded in the upstream face of the valve, making a leak-proof seal between the faces; thus a valve is formed. The spring-loaded lip seal is the only seal that can provide a class six, or bubble-tight, seal against the opposite face of the valve. Tearing action of the seal by high-velocity gas on this early design required relocation of the seal to the downstream face of the valve. In the stemless embodiment of this valve, inner and outer magnetic cartridges are employed to transfer mechanical torque from the outside of the valve to the inside without the use of a stem. This eliminates the leak path caused by the valve stems in standard valves because the stems penetrate through the bodies of these valves.
All in the family: matrimonial mitral valve clicks.
Desser, K B; Bokhari, S I; Benchimol, A; Romney, D
1981-05-01
Mitral valve clicks with or without late systolic murmurs were detected in genetically unrelated marital partners of 5 families. The first family represented 2 successive nonconsanguineous marital unions with 3 generations of mitral valve clicks. The second family included 1 natural and 2 adopted children with clinical and echographic evidence of mitral valve prolapse. The third family was comprised of asymptomatic parents, both with nonejection clicks and mitral valve prolapse, whose daughter presented 3 years previously with syncope, palpitations, and combined mitral and tricuspid valve prolapse. The fourth family had 3 members with auscultatory and ultrasonic manifestations of billowing mitral valve, whereas the fourth member had "silent mitral valve prolapse." The fifth family represented a mother with auscultatory and echographic evidence of mitral valve prolapse; her 14-year-old daughter had both mitral and tricuspid valve prolapse, whereas the son had a bicuspid aortic valve. Both children were products of a prior marriage, and her husband has symptomatic mitral valve prolapse. We conclude that matrimonial mitral valve prolapse probably reflects the purported (6--10%) prevalence of this disorder in the general population. The consequences of such marital union on progeny is currently unclear and warrants future investigation.
Structural valve deterioration in a starr-edwards mitral caged-disk valve prosthesis.
Aoyagi, Shigeaki; Tayama, Kei-Ichiro; Okazaki, Teiji; Shintani, Yusuke; Kono, Michitaka; Wada, Kumiko; Kosuga, Ken-Ichi; Mori, Ryusuke; Tanaka, Hiroyuki
2013-01-01
The durability of the Starr-Edwards (SE) mitral caged-disk valve, model 6520, is not clearly known, and structural valve deterioration in the SE disk valve is very rare. Replacement of the SE mitral disk valve was performed in 7 patients 23-40 years after implantation. Macroscopic examination of the removed disk valves showed no structural abnormalities in 3 patients, in whom the disk valves were removed at <26 years after implantation. Localized disk wear was found at the sites where the disk abutted the struts of the cage, in disk valves excised >36 years after implantation in 4 patients. Disk fracture, a longitudinal split in the disk along its circumference at the site of incorporation of the titanium ring, was detected in the valves removed 36 and 40 years after implantation, respectively, and many cracks were also observed on the outflow aspect of the disk removed 40 years after implantation. Disk fracture and localized disk wear were found in the SE mitral disk valves implanted >36 years previously. The present results suggest that SE mitral caged-disk valves implanted >20 years previously should be carefully followed up, and that those implanted >30 years previously should be electively replaced with modern prosthetic valves
Fracturing mechanics before valve-in-valve therapy of small aortic bioprosthetic heart valves.
Johansen, Peter; Engholt, Henrik; Tang, Mariann; Nybo, Rasmus F; Rasmussen, Per D; Nielsen-Kudsk, Jens Erik
2017-10-13
Patients with degraded bioprosthetic heart valves (BHV) who are not candidates for valve replacement may benefit from transcatheter valve-in-valve (VIV) therapy. However, in smaller-sized surgical BHV the resultant orifice may become too narrow. To overcome this, the valve frame can be fractured by a high-pressure balloon prior to VIV. However, knowledge on fracture pressures and mechanics are prerequisites. The aim of this study was to identify the fracture pressures needed in BHV, and to describe the fracture mechanics. Commonly used BHV of small sizes were mounted on a high-pressure balloon situated in a biplane fluoroscopic system with a high-speed camera. The instant of fracture was captured along with the balloon pressure. The valves were inspected for material protrusion and later dissected for fracture zone investigation and description. The valves with a polymer frame fractured at a lower pressure (8-10 atm) than those with a metal stent (19-26 atm). None of the fractured valves had elements protruding. VIV procedures in small-sized BHV may be performed after prior fracture of the valve frame by high-pressure balloon dilatation. This study provides tentative guidelines for expected balloon sizes and pressures for valve fracturing.
Tao, Ende; Wan, Li; Wang, WenJun; Luo, YunLong; Zeng, JinFu; Wu, Xia
2017-01-01
Surgery remains the primary form of treatment for infective endocarditis (IE). However, it is not clear what type of prosthetic valve provides a better prognosis. We conducted a meta-analysis to compare the prognosis of infective endocarditis treated with biological valves to cases treated with mechanical valves. Pubmed, Embase and Cochrane databases were searched from January 1960 to November 2016.Randomized controlled trials, retrospective cohorts and prospective studies comparing outcomes between biological valve and mechanical valve management for infective endocarditis were analyzed. The Newcastle-Ottawa Scale(NOS) was used to evaluate the quality of the literature and extracted data, and Stata 12.0 software was used for the meta-analysis. A total of 11 publications were included; 10,754 cases were selected, involving 6776 cases of biological valves and 3,978 cases of mechanical valves. The all-cause mortality risk of the biological valve group was higher than that of the mechanical valve group (HR = 1.22, 95% CI 1.03 to 1.44, P = 0.023), as was early mortality (RR = 1.21, 95% CI 1.02 to 1.43, P = 0.033). The recurrence of endocarditis (HR = 1.75, 95% CI 1.26 to 2.42, P = 0.001), as well as the risk of reoperation (HR = 1.79, 95% CI 1.15 to 2.80, P = 0.010) were more likely to occur in the biological valve group. The incidence of postoperative embolism was less in the biological valve group than in the mechanical valve group, but this difference was not statistically significant (RR = 0.90, 95% CI 0.76 to 1.07, P = 0.245). For patients with prosthetic valve endocarditis (PVE), there was no significant difference in survival rates between the biological valve group and the mechanical valve group (HR = 0.91, 95% CI 0.68 to 1.21, P = 0.520). The results of our meta-analysis suggest that mechanical valves can provide a significantly better prognosis in patients with infective endocarditis. There were significant differences in the clinical features of patients receiving a biological valve compared to patients receiving a mechanical valve. A large, multicenter retrospective study included in our meta-analysis suggested that any mortality risk of the biological valve group was significant higher than that of the mechanical valve group. However, the risk was no different after risk was adjusted. So, we thought the reason for this result may be related to the characteristics of the patient rather than valve dysfunction. It is still necessary to future randomized studies to verify this conclusion.
Magnetically operated check valve
NASA Technical Reports Server (NTRS)
Morris, Brian G. (Inventor); Bozeman, Richard J., Jr. (Inventor)
1994-01-01
A magnetically operated check valve is disclosed. The valve is comprised of a valve body and a movable poppet disposed therein. A magnet attracts the poppet to hold the valve shut until the force of fluid flow through the valve overcomes the magnetic attraction and moves the poppet to an unseated, open position. The poppet and magnet are configured and disposed to trap a magnetically attracted particulate and prevent it from flowing to a valve seating region.
Magnetically operated check valve
NASA Astrophysics Data System (ADS)
Morris, Brian G.; Bozeman, Richard J., Jr.
1994-06-01
A magnetically operated check valve is disclosed. The valve is comprised of a valve body and a movable poppet disposed therein. A magnet attracts the poppet to hold the valve shut until the force of fluid flow through the valve overcomes the magnetic attraction and moves the poppet to an unseated, open position. The poppet and magnet are configured and disposed to trap a magnetically attracted particulate and prevent it from flowing to a valve seating region.
Heart valve surgery - series (image)
... heart valves are either natural (biologic) or artificial (mechanical). Natural valves are from human donors (cadavers), modified ... artificial valves will require anticoagulation. The advantage of mechanical valves is that they last longer-thus, the ...
Improvement of a Pneumatic Control Valve with Self-Holding Function
NASA Astrophysics Data System (ADS)
Dohta, Shujiro; Akagi, Tetsuya; Kobayashi, Wataru; Shimooka, So; Masago, Yusuke
2017-10-01
The purpose of this study is to develop a small-sized, lightweight and low-cost control valve with low energy consumption and to apply it to the assistive system. We have developed some control valves; a tiny on/off valve using a vibration motor, and an on/off valve with self-holding function. We have also proposed and tested the digital servo valve with self-holding function using permanent magnets and a small-sized servo motor. In this paper, in order to improve the valve, an analytical model of the digital servo valve is proposed. And the simulated results by using the analytical model and identified parameters were compared with the experimental results. Then, the improved digital servo valve was designed based on the calculated results and tested. As a result, we realized the digital servo valve that can control the flow rate more precisely while maintaining its volume and weight compared with the previous valve. As an application of the improved valve, a position control system of rubber artificial muscle was built and the position control was performed successfully.
Guggenheim, S. Frederic
1986-01-01
A multi-port fluid valve apparatus is used to control the flow of fluids through a plurality of valves and includes a web, which preferably is a stainless steel endless belt. The belt has an aperture therethrough and is progressed, under motor drive and control, so that its aperture is moved from one valve mechanism to another. Each of the valve mechanisms comprises a pair of valve blocks which are held in fluid-tight relationship against the belt. Each valve block consists of a block having a bore through which the fluid flows, a first seal surrounding the bore and a second seal surrounding the first seal, with the distance between the first and second seals being greater than the size of the belt aperture. In order to open a valve, the motor progresses the belt aperture to where it is aligned with the two bores of a pair of valve blocks, such alignment permitting a flow of the fluid through the valve. The valve is closed by movement of the belt aperture and its replacement, within the pair of valve blocks, by a solid portion of the belt.
Intraluminal valves: development, function and disease
Geng, Xin; Cha, Boksik; Mahamud, Md. Riaj
2017-01-01
ABSTRACT The circulatory system consists of the heart, blood vessels and lymphatic vessels, which function in parallel to provide nutrients and remove waste from the body. Vascular function depends on valves, which regulate unidirectional fluid flow against gravitational and pressure gradients. Severe valve disorders can cause mortality and some are associated with severe morbidity. Although cardiac valve defects can be treated by valve replacement surgery, no treatment is currently available for valve disorders of the veins and lymphatics. Thus, a better understanding of valves, their development and the progression of valve disease is warranted. In the past decade, molecules that are important for vascular function in humans have been identified, with mouse studies also providing new insights into valve formation and function. Intriguing similarities have recently emerged between the different types of valves concerning their molecular identity, architecture and development. Shear stress generated by fluid flow has also been shown to regulate endothelial cell identity in valves. Here, we review our current understanding of valve development with an emphasis on its mechanobiology and significance to human health, and highlight unanswered questions and translational opportunities. PMID:29125824
Mpakopoulou, Maria; Brotis, Alexandros G; Gatos, Haralampos; Paterakis, Konstantinos; Fountas, Kostas N
2012-01-01
The aim of this study was to present our 10-year experience with the use of fixed-pressure and programmable valves in the treatment of adult patients requiring cerebrospinal fluid (CSF) diversion. Patients (n = 159; 89 male and 70 female) suffering from hydrocephalus of various causes underwent CSF shunt implantation. Forty fixed-pressure and 119 programmable valves were initially implanted. The observed revision rate was 40% in patients with fixed-pressure valves. In 20% of these patients, a revision due to valve mechanism malfunction was undertaken, and the initial valve was replaced with a programmable one. The revision rate in the adjustable-pressure valve subgroup was 20%. The infection rate for the fixed-pressure and programmable valve subgroups were 3%, and 1.7%, respectively. Similarly, subdural fluid collections were noticed in 17% and 4% of patients with fixed-pressure valves and programmable valves, respectively. The revision and over-drainage rates were significantly lower when using programmable valves, and thus, this type of valve is preferred whenever CSF has to be diverted.
Statistical Performance Evaluation Of Soft Seat Pressure Relief Valves
DOE Office of Scientific and Technical Information (OSTI.GOV)
Harris, Stephen P.; Gross, Robert E.
2013-03-26
Risk-based inspection methods enable estimation of the probability of failure on demand for spring-operated pressure relief valves at the United States Department of Energy's Savannah River Site in Aiken, South Carolina. This paper presents a statistical performance evaluation of soft seat spring operated pressure relief valves. These pressure relief valves are typically smaller and of lower cost than hard seat (metal to metal) pressure relief valves and can provide substantial cost savings in fluid service applications (air, gas, liquid, and steam) providing that probability of failure on demand (the probability that the pressure relief valve fails to perform its intendedmore » safety function during a potentially dangerous over pressurization) is at least as good as that for hard seat valves. The research in this paper shows that the proportion of soft seat spring operated pressure relief valves failing is the same or less than that of hard seat valves, and that for failed valves, soft seat valves typically have failure ratios of proof test pressure to set pressure less than that of hard seat valves.« less
Engine including hydraulically actuated valvetrain and method of valve overlap control
Cowgill, Joel [White Lake, MI
2012-05-08
An exhaust valve control method may include displacing an exhaust valve in communication with the combustion chamber of an engine to an open position using a hydraulic exhaust valve actuation system and returning the exhaust valve to a closed position using the hydraulic exhaust valve actuation assembly. During closing, the exhaust valve may be displaced for a first duration from the open position to an intermediate closing position at a first velocity by operating the hydraulic exhaust valve actuation assembly in a first mode. The exhaust valve may be displaced for a second duration greater than the first duration from the intermediate closing position to a fully closed position at a second velocity at least eighty percent less than the first velocity by operating the hydraulic exhaust valve actuation assembly in a second mode.
Carolan, Michael Francis; Cooke, John Albert; Buzinski, Michael David
2010-04-27
A gas flow isolation device includes a gas flow isolation valve movable from an opened condition to a closed condition. The module isolation valve in one embodiment includes a rupture disk in flow communication with a flow of gas when the module isolation valve is in an opened condition. The rupture disk ruptures when a predetermined pressure differential occurs across it causing the isolation valve to close. In one embodiment the valve is mechanically linked to the rupture disk to maintain the valve in an opened condition when the rupture disk is intact, and which permits the valve to move into a closed condition when the rupture disk ruptures. In another embodiment a crushable member maintains the valve in an open condition, and the flow of gas passed the valve upon rupturing of the rupture disk compresses the crushable member to close the isolation valve.
Pressure control valve. [inflating flexible bladders
NASA Technical Reports Server (NTRS)
Lambson, K. H. (Inventor)
1980-01-01
A control valve is provided which is adapted to be connected between a pressure source, such as a vacuum pump, and a pressure vessel so as to control the pressure in the vessel. The valve comprises a housing having a longitudinal bore which is connected between the pump and vessel, and a transversely movable valve body which controls the air flow through an air inlet in the housing. The valve body includes cylindrical and conical shaped portions which cooperate with reciprocally shaped portions of the housing to provide flow control. A filter in the air inlet removes foreign matter from the air. The bottom end of the valve body is screwed into the valve housing control knob formed integrally with the valve body and controls translation of the valve body, and the opening and closing of the valve.
van Gorp, Maarten J; van der Graaf, Yolanda; de Mol, Bas A J M; Bakker, Chris J G; Witkamp, Theo D; Ramos, Lino M P; Mali, Willem P T M
2004-03-01
To assess the relationship between heart valve history and susceptibility artifacts at magnetic resonance (MR) imaging of the brain in patients with Björk-Shiley convexoconcave (BSCC) valves. MR images of the brain were obtained in 58 patients with prosthetic heart valves: 20 patients had BSCC valve replacements, and 38 had other types of heart valves. Two experienced neuroradiologists determined the presence or absence of susceptibility artifacts in a consensus reading. Artifacts were defined as characteristic black spots that were visible on T2*-weighted gradient-echo MR images. The statuses of the 20 explanted BSCC valves-specifically, whether they were intact or had an outlet strut fracture (OSF) or a single-leg fracture (SLF)-had been determined earlier. Number of artifacts seen at brain MR imaging was correlated with explanted valve status, and differences were analyzed with nonparametric statistical tests. Significantly more patients with BSCC valves (17 [85%] of 20 patients) than patients with other types of prosthetic valves (18 [47%] of 38 patients) had susceptibility artifacts at MR imaging (P =.005). BSCC valve OSFs were associated with a significantly higher number of artifacts than were intact BSCC valves (P =.01). No significant relationship between SLF and number of artifacts was observed. Susceptibility artifacts at brain MR imaging are not restricted to patients with BSCC valves. These artifacts can be seen on images obtained in patients with various other types of fractured and intact prosthetic heart valves. Copyright RSNA, 2004
Molnar, Adrian; Muresan, Ioan; Trifan, Catalin; Pop, Dana; Sacui, Diana
2015-01-01
The introduction of Duke's criteria and the improvement of imaging methods has lead to an earlier and a more accurate diagnosis of infectious endocarditis (IE). The options for the best therapeutic approach and the timing of surgery are still a matter of debate and require a close colaboration between the cardiologist, the infectionist and the cardiac surgeon. We undertook a retrospective, descriptive study, spanning over a period of five years (from January 1st, 2007 to December 31st, 2012), on 100 patients who underwent surgery for native valve infectious endocarditis in our unit. The patients' age varied between 13 and 77 years (with a mean of 54 years), of which 85 were males (85%). The main microorganisms responsible for IE were: Streptococcus Spp. (21 cases - 21%), Staphylococcus Spp. (15 cases - 15%), and Enterococcus Spp. (9 cases - 9%). The potential source of infection was identified in 26 patients (26%), with most cases being in the dental area (16 cases - 16%). The lesions caused by IE were situated in the left heart in 96 patients (96%), mostly on the aortic valve (50 cases - 50%). In most cases (82%) we found preexisting endocardial lesions which predisposed to the development of IE, most of them being degenerative valvular lesions (38 cases - 38%). We performed the following surgical procedures: surgery on a single valve - aortic valve replacement (40 cases), mitral valve replacement (19 cases), mitral valve repair (1 case), surgery on more than one valve - mitral and aortic valve replacement (20 cases), aortic and tricuspid valve replacement (1 case), aortic valve replacement with a mechanical valve associated with mitral valve repair (5 cases), aortic valve replacement with a biological valve associated with mitral valve repair (2 cases), and mitral valve replacement with a mechanical valve combined with De Vega procedure on the tricuspid valve (1 case). In 5 patients (5%) the bacteriological examination of valve pieces excised during surgery was positive. In 3 cases it matched the germ identified in the hemocultures, and in 2 cases it evidenced another bacterium. The overall mortality of 5% is well between the limits presented in literature, being higher (30%) in patients who required emergency surgery. For the patients who return into our clinic with prosthetic valve endocarditis, the mortality after surgery was even higher (50%).
Nielsen, Joseph; Tokuhiro, Akira; Hiromoto, Robert; ...
2015-11-13
Evaluation of the impacts of uncertainty and sensitivity in modeling presents a significant set of challenges in particular to high fidelity modeling. Computational costs and validation of models creates a need for cost effective decision making with regards to experiment design. Experiments designed to validate computation models can be used to reduce uncertainty in the physical model. In some cases, large uncertainty in a particular aspect of the model may or may not have a large impact on the final results. For example, modeling of a relief valve may result in large uncertainty, however, the actual effects on final peakmore » clad temperature in a reactor transient may be small and the large uncertainty with respect to valve modeling may be considered acceptable. Additionally, the ability to determine the adequacy of a model and the validation supporting it should be considered within a risk informed framework. Low fidelity modeling with large uncertainty may be considered adequate if the uncertainty is considered acceptable with respect to risk. In other words, models that are used to evaluate the probability of failure should be evaluated more rigorously with the intent of increasing safety margin. Probabilistic risk assessment (PRA) techniques have traditionally been used to identify accident conditions and transients. Traditional classical event tree methods utilize analysts’ knowledge and experience to identify the important timing of events in coordination with thermal-hydraulic modeling. These methods lack the capability to evaluate complex dynamic systems. In these systems, time and energy scales associated with transient events may vary as a function of transition times and energies to arrive at a different physical state. Dynamic PRA (DPRA) methods provide a more rigorous analysis of complex dynamic systems. Unfortunately DPRA methods introduce issues associated with combinatorial explosion of states. This study presents a methodology to address combinatorial explosion using a Branch-and-Bound algorithm applied to Dynamic Event Trees (DET), which utilize LENDIT (L – Length, E – Energy, N – Number, D – Distribution, I – Information, and T – Time) as well as a set theory to describe system, state, resource, and response (S2R2) sets to create bounding functions for the DET. The optimization of the DET in identifying high probability failure branches is extended to create a Phenomenological Identification and Ranking Table (PIRT) methodology to evaluate modeling parameters important to safety of those failure branches that have a high probability of failure. The PIRT can then be used as a tool to identify and evaluate the need for experimental validation of models that have the potential to reduce risk. Finally, in order to demonstrate this methodology, a Boiling Water Reactor (BWR) Station Blackout (SBO) case study is presented.« less
Tricuspid Valve Repair in Infancy Using Neochordae: Three-Dimensional Echocardiographic Imaging.
Martin, Billie-Jean; Khoo, Nee S; Smallhorn, Jeffrey; Aklabi, Mohammed Al
2017-11-01
Tricuspid regurgitation (TR) in infancy poses a surgical challenge. Both two- and three-dimensional echocardiography (3DE) can provide detailed information about the mechanism(s) of valve failure and insights into valve adaptation during follow-up. We report two patients who underwent tricuspid valve repair using Gore-Tex neochordae, repairs which were facilitated by and assessed with 3DE. Both infants had less than mild residual TR and no valve tethering at hospital discharge. Furthermore, follow-up 3DEs have helped to confirm valve competence, lack of tethering, and growth of the valve and valve apparatus.
Study of multiple cycles valves
NASA Technical Reports Server (NTRS)
Wichmann, H.
1973-01-01
A discussion is presented regarding valves which can be cycled repeatedly and are available from industry for application in the inlet system for the Pioneer Venus Probe mass spectrometer. Both solenoid type and latching type valves are considered. The study is divided into two principal areas: (1) preparation of a valve specification reflecting the requirements of the inlet system cyclic valves for the Pioneer Venus Probe mass spectrometer and the submittal of this specification to potential valve suppliers for their response and proposal; (2) preparation of a design layout of an optimum cyclic valve meeting all of the valve specification requirements.
Fast acting multiple element valve
Yang, Jefferson Y. S.; Wada, James M.
1991-01-01
A plurality of slide valve elements having plural axial-spaced annular parts and an internal slide are inserted into a bulkhead in a fluid conduit from a downstream side of the bulkhead, locked in place by a bayonet coupling and set screw, and project through the bulkhead into the upstream conduit. Pneumatic lines connecting the slide valve element actuator to pilot valves are brought out the throat of the valve element to the downstream side. Pilot valves are radially spaced around the exterior of the valve to permit the pneumatic lines to be made identical, thereby to minimize adverse timing tolerances in operation due to pressure variations. Ring manifolds surround the valve adjacent respective pilot valve arrangements to further reduce adverse timing tolerances due to pressure variations, the manifolds being directly connected to the respective pilot valves. Position sensors are provided the valve element slides to signal the precise time at which a slide reaches or passes through a particular point in its stroke to initiate a calibrated timing function.
Recellularization of decellularized heart valves: Progress toward the tissue-engineered heart valve
VeDepo, Mitchell C; Detamore, Michael S; Hopkins, Richard A; Converse, Gabriel L
2017-01-01
The tissue-engineered heart valve portends a new era in the field of valve replacement. Decellularized heart valves are of great interest as a scaffold for the tissue-engineered heart valve due to their naturally bioactive composition, clinical relevance as a stand-alone implant, and partial recellularization in vivo. However, a significant challenge remains in realizing the tissue-engineered heart valve: assuring consistent recellularization of the entire valve leaflets by phenotypically appropriate cells. Many creative strategies have pursued complete biological valve recellularization; however, identifying the optimal recellularization method, including in situ or in vitro recellularization and chemical and/or mechanical conditioning, has proven difficult. Furthermore, while many studies have focused on individual parameters for increasing valve interstitial recellularization, a general understanding of the interacting dynamics is likely necessary to achieve success. Therefore, the purpose of this review is to explore and compare the various processing strategies used for the decellularization and subsequent recellularization of tissue-engineered heart valves. PMID:28890780
NASA Astrophysics Data System (ADS)
Jong, Rudiyanto P.; Osman, Kahar; Adib, M. Azrul Hisham M.
2012-06-01
Mitral valve prolapse without proper monitoring might lead to a severe mitral valve failure which eventually leads to a sudden death. Additional information on the mitral valve leaflet condition against the backflow volume would be an added advantage to the medical practitioner for their decision on the patients' treatment. A study on two dimensional echocardiography images has been conducted and the correlations between the backflow volume of the mitral regurgitation and mitral valve leaflet Young modulus have been obtained. Echocardiogram images were analyzed on the aspect of backflow volume percentage and mitral valve leaflet dimensions on different rates of backflow volume. Young modulus values for the mitral valve leaflet were obtained by using the principle of elastic deflection and deformation on the mitral valve leaflet. The results show that the backflow volume increased with the decrease of the mitral valve leaflet Young modulus which also indicate the condition of the mitral valve leaflet approaching failure at high backflow volumes. Mitral valve leaflet Young modulus values obtained in this study agreed with the healthy mitral valve leaflet Young modulus from the literature. This is an initial overview of the trend on the prediction of the behaviour between the fluid and the structure of the blood and the mitral valve which is extendable to a larger system of prediction on the mitral valve leaflet condition based on the available echocardiogram images.
BORED AND ASSEMBLED GATE VALVES RECEIVING PROTECTIVE COATING IN THE ...
BORED AND ASSEMBLED GATE VALVES RECEIVING PROTECTIVE COATING IN THE VALVE PAINT BOOTH OF THE VALVE ASSEMBLY BUILDING. - Stockham Pipe & Fittings Company, Valve Assembly Building, 4000 Tenth Avenue North, Birmingham, Jefferson County, AL
Complementary role of cardiac CT in the assessment of aortic valve replacement dysfunction
Moss, Alastair J; Dweck, Marc R; Dreisbach, John G; Williams, Michelle C; Mak, Sze Mun; Cartlidge, Timothy; Nicol, Edward D; Morgan-Hughes, Gareth J
2016-01-01
Aortic valve replacement is the second most common cardiothoracic procedure in the UK. With an ageing population, there are an increasing number of patients with prosthetic valves that require follow-up. Imaging of prosthetic valves is challenging with conventional echocardiographic techniques making early detection of valve dysfunction or complications difficult. CT has recently emerged as a complementary approach offering excellent spatial resolution and the ability to identify a range of aortic valve replacement complications including structural valve dysfunction, thrombus development, pannus formation and prosthetic valve infective endocarditis. This review discusses each and how CT might be incorporated into a multimodal cardiovascular imaging pathway for the assessment of aortic valve replacements and in guiding clinical management. PMID:27843568
NASA Technical Reports Server (NTRS)
Jadaan, Osama
2001-01-01
Present capabilities of the NASA CARES/Life (Ceramic Analysis and Reliability Evaluation of Structures/Life) code include probabilistic life prediction of ceramic components subjected to fast fracture, slow crack growth (stress corrosion), and cyclic fatigue failure modes. Currently, this code has the capability to compute the time-dependent reliability of ceramic structures subjected to simple time-dependent loading. For example, in slow crack growth (SCG) type failure conditions CARES/Life can handle the cases of sustained and linearly increasing time-dependent loads, while for cyclic fatigue applications various types of repetitive constant amplitude loads can be accounted for. In real applications applied loads are rarely that simple, but rather vary with time in more complex ways such as, for example, engine start up, shut down, and dynamic and vibrational loads. In addition, when a given component is subjected to transient environmental and or thermal conditions, the material properties also vary with time. The objective of this paper is to demonstrate a methodology capable of predicting the time-dependent reliability of components subjected to transient thermomechanical loads that takes into account the change in material response with time. In this paper, the dominant delayed failure mechanism is assumed to be SCG. This capability has been added to the NASA CARES/Life (Ceramic Analysis and Reliability Evaluation of Structures/Life) code, which has also been modified to have the ability of interfacing with commercially available FEA codes executed for transient load histories. An example involving a ceramic exhaust valve subjected to combustion cycle loads is presented to demonstrate the viability of this methodology and the CARES/Life program.
Aortic valve repair leads to a low incidence of valve-related complications.
Aicher, Diana; Fries, Roland; Rodionycheva, Svetlana; Schmidt, Kathrin; Langer, Frank; Schäfers, Hans-Joachim
2010-01-01
Aortic valve replacement for aortic regurgitation (AR) has been established as a standard treatment but implies prosthesis-related complications. Aortic valve repair is an alternative approach, but its mid- to long-term results still need to be defined. Over a 12-year period, 640 patients underwent aortic valve repair for regurgitation of a unicuspid (n=21), bicuspid (n=205), tricuspid (n=411) or quadricuspid (n=3) aortic valve. The mechanism of regurgitation involved prolapse (n=469) or retraction (n=20) of the cusps, and dilatation of the root (n=323) or combined pathologies. Treatment consisted of cusp repair (n=529), root repair (n=323) or a combination of both (n=208). The patients were followed clinically and echocardiographically; follow-up was complete in 98.5% (cumulative follow-up: 3035 patient years). Hospital mortality was 3.4% in the total patient cohort and 0.8% for isolated aortic valve repair. The incidences of thrombo-embolism (0.2% per patient per year) and endocarditis (0.16%per patient per year) were low. Freedom from re-operation at 5 and 10 years was 88% and 81% in bicuspid and 97% and 93% in tricuspid aortic valves (p=0.0013). At re-operation, 13 out of 36 valves could be re-repaired. Freedom from valve replacement was 95% and 90% in bicuspid and 97% and 94% in tricuspid aortic valves (p=0.36). Freedom from all valve-related complications at 10 years was 88%. Reconstructive surgery of the aortic valve is feasible with low mortality in many individuals with aortic regurgitation. Freedom from valve-related complications after valve repair seems superior compared to available data on standard aortic valve replacement. Copyright 2009 European Association for Cardio-Thoracic Surgery. Published by Elsevier B.V. All rights reserved.
Force measuring valve assemblies, systems including such valve assemblies and related methods
DeWall, Kevin George [Pocatello, ID; Garcia, Humberto Enrique [Idaho Falls, ID; McKellar, Michael George [Idaho Falls, ID
2012-04-17
Methods of evaluating a fluid condition may include stroking a valve member and measuring a force acting on the valve member during the stroke. Methods of evaluating a fluid condition may include measuring a force acting on a valve member in the presence of fluid flow over a period of time and evaluating at least one of the frequency of changes in the measured force over the period of time and the magnitude of the changes in the measured force over the period of time to identify the presence of an anomaly in a fluid flow and, optionally, its estimated location. Methods of evaluating a valve condition may include directing a fluid flow through a valve while stroking a valve member, measuring a force acting on the valve member during the stroke, and comparing the measured force to a reference force. Valve assemblies and related systems are also disclosed.
Simulation of proportional control of hydraulic actuator using digital hydraulic valves
NASA Astrophysics Data System (ADS)
Raghuraman, D. R. S.; Senthil Kumar, S.; Kalaiarasan, G.
2017-11-01
Fluid power systems using oil hydraulics in earth moving and construction equipment have been using proportional and servo control valves for a long time to achieve precise and accurate position control backed by system performance. Such valves are having feedback control in them and exhibit good response, sensitivity and fine control of the actuators. Servo valves and proportional valves are possessing less hysteresis when compared to on-off type valves, but when the servo valve spools get stuck in one position, a high frequency called as jitter is employed to bring the spool back, whereas in on-off type valves it requires lesser technology to retract the spool. Hence on-off type valves are used in a technology known as digital valve technology, which caters to precise control on slow moving loads with fast switching times and with good flow and pressure control mimicking the performance of an equivalent “proportional valve” or “servo valve”.
Noninvasive valve monitor using alternating electromagnetic field
Eissenberg, David M.; Haynes, Howard D.; Casada, Donald A.
1993-01-01
One or more electrical coils are carefully located on the outside of a valve body. An alternating current passing through the coil(s) results in an alternating electromagnetic field being transmitted into the valve body and valve internals. The electromagnetic field varies in intensity and polarity in the valve. As the position of a valve internal part is changed, the electromagnetic field throughout the valve body and its internals is altered. A passive receiver coil carefully located on the outside of the valve body detects the intensity of the electromagnetic field at that location as an induced electrical voltage in the coil. With the change in position of the valve internal part, there is a corresponding change in the induced voltage as a result of the alteration in the alternating electromagnetic field at that location. Changes in the voltage provide an indication of the position and motion of valve internals.
Noninvasive valve monitor using alternating electromagnetic field
Eissenberg, D.M.; Haynes, H.D.; Casada, D.A.
1993-03-16
One or more electrical coils are carefully located on the outside of a valve body. An alternating current passing through the coil(s) results in an alternating electromagnetic field being transmitted into the valve body and valve internals. The electromagnetic field varies in intensity and polarity in the valve. As the position of a valve internal part is changed, the electromagnetic field throughout the valve body and its internals is altered. A passive receiver coil carefully located on the outside of the valve body detects the intensity of the electromagnetic field at that location as an induced electrical voltage in the coil. With the change in position of the valve internal part, there is a corresponding change in the induced voltage as a result of the alteration in the alternating electromagnetic field at that location. Changes in the voltage provide an indication of the position and motion of valve internals.
Mieres, Juan; Menéndez, Marcelo; Fernández-Pereira, Carlos; Rubio, Miguel; Rodríguez, Alfredo E.
2015-01-01
Transcatheter Aortic Valve Replacement (TAVR) is performed in patients who are poor surgical candidates. Many patients have inadequate femoral access, and alternative access sites have been used such as the transapical approach discussed in this paper. We present an elderly and fragile patient not suitable for surgery for unacceptable high risk, including poor ventricular function, previous myocardial infarction with percutaneous coronary intervention, pericardial effusion, and previous cardiac surgery with replacement of mechanical mitral valve. Transapical aortic valve replacement with a second-generation self-expanding JenaValve is performed. The JenaValve is a second-generation transapical TAVR valve consisting of a porcine root valve mounted on a low-profile nitinol stent. The valve is fully retrievable and repositionable. We discuss transapical access, implantation technique, and feasibility of valve implantation in this extremely high surgical risk patient. PMID:26346128
Mieres, Juan; Menéndez, Marcelo; Fernández-Pereira, Carlos; Rubio, Miguel; Rodríguez, Alfredo E
2015-01-01
Transcatheter Aortic Valve Replacement (TAVR) is performed in patients who are poor surgical candidates. Many patients have inadequate femoral access, and alternative access sites have been used such as the transapical approach discussed in this paper. We present an elderly and fragile patient not suitable for surgery for unacceptable high risk, including poor ventricular function, previous myocardial infarction with percutaneous coronary intervention, pericardial effusion, and previous cardiac surgery with replacement of mechanical mitral valve. Transapical aortic valve replacement with a second-generation self-expanding JenaValve is performed. The JenaValve is a second-generation transapical TAVR valve consisting of a porcine root valve mounted on a low-profile nitinol stent. The valve is fully retrievable and repositionable. We discuss transapical access, implantation technique, and feasibility of valve implantation in this extremely high surgical risk patient.
Reduced-impact sliding pressure control valve for pneumatic hammer drill
Polsky, Yarom [Oak Ridge, TN; Grubelich, Mark C [Albuquerque, NM; Vaughn, Mark R [Albuquerque, NM
2012-05-15
A method and means of minimizing the effect of elastic valve recoil in impact applications, such as percussive drilling, where sliding spool valves used inside the percussive device are subject to poor positioning control due to elastic recoil effects experienced when the valve impacts a stroke limiting surface. The improved valve design reduces the reflected velocity of the valve by using either an energy damping material, or a valve assembly with internal damping built-in, to dissipate the compression stress wave produced during impact.
System and method for controlling engine knock using electro-hydraulic valve actuation
Brennan, Daniel G
2013-12-10
A control system for an engine includes a knock control module and a valve control module. The knock control module adjusts a period that one or more of an intake valve and an exhaust valve of a cylinder are open based on engine knock corresponding to the cylinder. The valve control module, based on the adjusted period, controls the one or more of the intake valve and the exhaust valve using one or more hydraulic actuators.
Francel, P C; Stevens, F A; Tompkins, P; Pollay, M
2001-02-01
The proper functioning of shunt valves in vivo is dependent on many factors, including the valve itself, the anti-siphon device or ASD (if included), patency of inlet and outlet tubing, and location of the valve. One important, but sometimes overlooked, consideration in valve function is the valve location relative to the tip of the ventricular inlet catheter. As with any pressure measurement, the zero or reference position is an important concept. In the case of shunt valves, the position of the proximal inlet catheter tip is fixed and therefore serves as the reference point for all pressure measurements. This study was conducted to document the importance of this relationship for the pressure/flow characteristics of the shunt valve. We bench-tested differential pressure valves (with integral anti-gravity devices; AGDs) from three manufacturers. Valves were connected to an "infinite" reservoir, and the starting head pressure for each was determined from product inserts. The inlet catheter tip was fixed at this position, and the valve body was moved in relation to the inlet catheter tip. Outflow rates were determined gravimetrically for positions varying between 4 cm above and 8 cm below the inlet catheter tip. All differential pressure valves utilized in this study that contained AGDs showed significant increases in outflow rate as the valve body was moved incrementally below the level of the inlet catheter tip. To allow functioning as a zero-hydrostatic pressure differential pressure valve, the AGD and the inlet catheter tip should be aligned at the same horizontal level.
Prêtre, René; Rosser, Barbara; Mueller, Christoph; Kretschmar, Oliver; Dave, Hitendu
2012-09-01
The purpose of this study was to review our experience with recycling of the pulmonary valve in cases of chronic pulmonary insufficiency after a transannular patch procedure as part of a repair of tetralogy of Fallot. Eight patients in whom the technique was used were reviewed. Technically, the valve was reapproximated at the anterior commissure if the valve leaflet was sufficiently developed and of good tissue quality. Additional corrections were performed in 5 patients (resection of an infundibular aneurysm [5 patients], repair of the tricuspid valve [1 patient]). The valve was competent with no or trivial regurgitation in 5 patients and a small regurgitation in 3 patients. There was no significant transvalvular gradient in 5 patients with tricuspid valves and a small gradient in 3 patients with a bicuspid valves (<23 mm Hg). The valve function remained stable over the follow-up period (median time, 32 months). Recycling of the pulmonary valve is an interesting concept that could avoid the necessary reoperations linked with valves or valved prostheses. The repair must be carefully followed in bicuspid valves because of a reduction in the opening area. Valve leaflets of good quality should be preserved during the primary repair of tetralogy of Fallot and the transannular incision should be made across the anterior commissure if possible. These steps should allow a few patients to profit from a recycling of their valves in the future. Copyright © 2012 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.
Martins, Carlo de Oliveira; Demarchi, Lea; Ferreira, Frederico Moraes; Pomerantzeff, Pablo Maria Alberto; Brandao, Carlos; Sampaio, Roney Orismar; Spina, Guilherme Sobreira; Kalil, Jorge; Cunha-Neto, Edecio; Guilherme, Luiza
2017-01-01
Autoimmune inflammatory reactions leading to rheumatic fever (RF) and rheumatic heart disease (RHD) result from untreated Streptococcus pyogenes throat infections in individuals who exhibit genetic susceptibility. Immune effector mechanisms have been described that lead to heart tissue damage culminating in mitral and aortic valve dysfunctions. In myxomatous valve degeneration (MXD), the mitral valve is also damaged due to non-inflammatory mechanisms. Both diseases are characterized by structural valve disarray and a previous proteomic analysis of them has disclosed a distinct profile of matrix/structural proteins differentially expressed. Given their relevance in organizing valve tissue, we quantitatively evaluated the expression of vimentin, collagen VI, lumican, and vitronectin as well as performed immunohistochemical analysis of their distribution in valve tissue lesions of patients in both diseases. We identified abundant expression of two isoforms of vimentin (45 kDa, 42 kDa) with reduced expression of the full-size protein (54 kDa) in RHD valves. We also found increased vitronectin expression, reduced collagen VI expression and similar lumican expression between RHD and MXD valves. Immunohistochemical analysis indicated disrupted patterns of these proteins in myxomatous degeneration valves and disorganized distribution in rheumatic heart disease valves that correlated with clinical manifestations such as valve regurgitation or stenosis. Confocal microscopy analysis revealed a diverse pattern of distribution of collagen VI and lumican into RHD and MXD valves. Altogether, these results demonstrated distinct patterns of altered valve expression and tissue distribution/organization of structural/matrix proteins that play important pathophysiological roles in both valve diseases.
Tao, Ende; Wan, Li; Wang, WenJun; Luo, YunLong; Zeng, JinFu; Wu, Xia
2017-01-01
Objective Surgery remains the primary form of treatment for infective endocarditis (IE). However, it is not clear what type of prosthetic valve provides a better prognosis. We conducted a meta-analysis to compare the prognosis of infective endocarditis treated with biological valves to cases treated with mechanical valves. Methods Pubmed, Embase and Cochrane databases were searched from January 1960 to November 2016.Randomized controlled trials, retrospective cohorts and prospective studies comparing outcomes between biological valve and mechanical valve management for infective endocarditis were analyzed. The Newcastle-Ottawa Scale(NOS) was used to evaluate the quality of the literature and extracted data, and Stata 12.0 software was used for the meta-analysis. Results A total of 11 publications were included; 10,754 cases were selected, involving 6776 cases of biological valves and 3,978 cases of mechanical valves. The all-cause mortality risk of the biological valve group was higher than that of the mechanical valve group (HR = 1.22, 95% CI 1.03 to 1.44, P = 0.023), as was early mortality (RR = 1.21, 95% CI 1.02 to 1.43, P = 0.033). The recurrence of endocarditis (HR = 1.75, 95% CI 1.26 to 2.42, P = 0.001), as well as the risk of reoperation (HR = 1.79, 95% CI 1.15 to 2.80, P = 0.010) were more likely to occur in the biological valve group. The incidence of postoperative embolism was less in the biological valve group than in the mechanical valve group, but this difference was not statistically significant (RR = 0.90, 95% CI 0.76 to 1.07, P = 0.245). For patients with prosthetic valve endocarditis (PVE), there was no significant difference in survival rates between the biological valve group and the mechanical valve group (HR = 0.91, 95% CI 0.68 to 1.21, P = 0.520). Conclusion The results of our meta-analysis suggest that mechanical valves can provide a significantly better prognosis in patients with infective endocarditis. There were significant differences in the clinical features of patients receiving a biological valve compared to patients receiving a mechanical valve. A large, multicenter retrospective study included in our meta-analysis suggested that any mortality risk of the biological valve group was significant higher than that of the mechanical valve group. However, the risk was no different after risk was adjusted. So, we thought the reason for this result may be related to the characteristics of the patient rather than valve dysfunction. It is still necessary to future randomized studies to verify this conclusion. PMID:28407024
Fluid check valve has fail-safe feature
NASA Technical Reports Server (NTRS)
Gaul, L. C.
1965-01-01
Check valve ensures unidirectional fluid flow and, in case of failure, vents the downstream fluid to the atmosphere and gives a positive indication of malfunction. This dual valve consists of a master check valve and a fail-safe valve.
75 FR 10696 - Airworthiness Directives; Fokker Services B.V. Model F.28 Mark 0070 and 0100 Airplanes
Federal Register 2010, 2011, 2012, 2013, 2014
2010-03-09
... form on actuators P/N 9409122 installed on fuel crossfeed valves and fuel fire shut-off valves. Tests... fuel crossfeed valves and fuel fire shut-off valves. Tests revealed that the ice can prevent the... Tests for Fuel Crossfeed Valves and Fuel Fire Shut-Off Valves (g) For airplanes with an actuator having...
Method and system for measuring gate valve clearances and seating force
Casada, Donald A.; Haynes, Howard D.; Moyers, John C.; Stewart, Brian K.
1996-01-01
Valve clearances and seating force, as well as other valve operational parameters, are determined by measuring valve stem rotation during opening and closing operations of a translatable gate valve. The magnitude of the stem rotation, and the relative difference between the stem rotation on opening and closing provides valuable data on the valve internals in a non-intrusive manner.
Code of Federal Regulations, 2014 CFR
2014-07-01
... Population Emission Factors—Storage Wellheads, Gas Service Connector 0.01 Valve 0.1 Pressure Relief Valve 0.17 Open Ended Line 0.03 Population Emission Factors—Other Components, Gas Service Low Continuous... Bleed Pneumatic Device Vents 2 2.35 1 Valves include control valves, block valves and regulator valves...
Code of Federal Regulations, 2013 CFR
2013-07-01
... Population Emission Factors—Storage Wellheads, Gas Service Connector 0.01 Valve 0.1 Pressure Relief Valve 0.17 Open Ended Line 0.03 Population Emission Factors—Other Components, Gas Service Low Continuous... Bleed Pneumatic Device Vents 2 2.35 1 Valves include control valves, block valves and regulator valves...
Aortic valve replacement for papillary fibroelastoma.
Arikan, Ali Ahmet; Omay, Oğuz; Aydın, Fatih; Kanko, Muhip; Gür, Sibel; Derviş, Emir; Yılmaz, Cansu Eda; Müezzinoğlu, Bahar
2017-06-01
Surgery is indicated for symptomatic patients with papillary fibroelastomas (PFE) on the aortic valve. The valve is commonly spared during tumor excision. Rarely, aortic valve replacement (AVR) is needed. We present a case requiring AVR for an aortic valve PFE and review the literature to determine the risk factors for failure of aortic valve-sparing techniques in patients with PFE. © 2017 Wiley Periodicals, Inc.
Long life valve design concepts
NASA Technical Reports Server (NTRS)
Jones, J. R.; Hall, A. H., Jr.
1975-01-01
Valve concept evaluation, final candidate selection, design, manufacture, and demonstration testing of a pneumatically actuated 10-inch hybrid poppet butterfly shutoff valve are presented. Conclusions and recommendations regarding those valve characteristics and features which would serve to guide in the formulation of future valve procurements are discussed. The pertinent design goals were temperature range of plus 200 to minus 423 F, valve inlet pressure 35 psia, actuation pressure 750 psia, main seal leakage 3 x 0.00001 sccs at 35 psia valve inlet pressure, and a storage and operating life of 10 years. The valve was designed to be compatible with RP-1, propane, LH2, LO2, He, and N2.
Noninvasive valve monitor using constant magnetic and/or DC electromagnetic field
Casada, D.A.; Haynes, H.D.
1993-08-17
One or more sources of steady magnetic field are carefully located on the outside of a valve body. The constant magnetic field is transmitted into the valve body and valve internals. A magnetic field detector carefully located on the outside of the valve body detects the intensity of the magnetic field at its location. As the position of a valve internal part is changed, there is an alteration in the magnetic field in the valve, and a consequent change in the detected magnetic field. Changes in the detected signal provide an indication of the position and motion of the valve internals.
NASA Technical Reports Server (NTRS)
Addona, Brad; Eddleman, David
2015-01-01
A developmental Main Oxidizer Valve (MOV) was designed by NASA-MSFC using additive manufacturing processes. The MOV is a pneumatically actuated poppet valve to control the flow of liquid oxygen to an engine's injector. A compression spring is used to return the valve to the closed state when pneumatic pressure is removed from the valve. The valve internal parts are cylindrical in shape, which lends itself to traditional lathe and milling operations. However, the valve body represents a complicated shape and contains the majority of the mass of the valve. Additive manufacturing techniques were used to produce a part that optimized mass and allowed for design features not practical with traditional machining processes.
Noninvasive valve monitor using constant magnetic and/or DC electromagnetic field
Casada, Donald A.; Haynes, Howard D.
1993-01-01
One or more sources of steady magnetic field are carefully located on the outside of a valve body. The constant magnetic field is transmitted into the valve body and valve internals. A magnetic field detector carefully located on the outside of the valve body detects the intensity of the magnetic field at its location. As the position of a valve internal part is changed, there is an alteration in the magnetic field in the valve, and a consequent change in the detected magnetic field. Changes in the detected signal provide an indication of the position and motion of the valve internals.
Thermal and mechanical analysis of major components for the advanced adiabatic diesel engine
NASA Technical Reports Server (NTRS)
1983-01-01
The proposed design for the light duty diesel is an in-line four cylinder spark assisted diesel engine mounted transversely in the front of the vehicle. The engine has a one piece cylinder head, with one intake valve and one exhaust valve per cylinder. A flat topped piston is used with a cylindrical combustion chamber recessed into the cylinder head directly under the exhaust valve. A single ceramic insert is cast into the cylinder head to insulate both the combustion chamber and the exhaust port. A similar ceramic insert is cast into the head to insulate the intake port. A ceramic faceplate is pressed into the combustion face of the head to insulate the face of the head from hot combustion gas. The valve seats are machined directly into the ceramic faceplate for the intake valve and into the ceramic exhaust pot insert for the exhaust valve. Additional ceramic applications in the head are the use of ceramic valve guides and ceramic insulated valves. The ceramic valve guides are press fit into the head and are used for increased wear resistance. The ceramic insulated valves are conventional valves with the valve faces plasma spray coated with ceramic for insulation.
Outcome of bioprosthetic valve replacement in dogs with tricuspid valve dysplasia.
Bristow, P; Sargent, J; Luis Fuentes, V; Brockman, D
2017-04-01
To describe the short-term and long-term outcome in dogs with tricuspid valve dysplasia undergoing tricuspid valve replacement under cardiopulmonary bypass. Data were collected from the hospital records of all dogs that had undergone tricuspid valve replacement under cardiopulmonary bypass between 2006 and 2012. Dogs were considered candidates for tricuspid valve replacement if they had severe tricuspid valve regurgitation associated with clinical signs of cardiac compromise. Nine dogs of six different breeds were presented. Median age was 13 months (range 7 to 61 months), median weight 26·5 kg (range 9·7 to 59 kg). Eight bovine pericardial valves and one porcine aortic valve were used. One non-fatal intraoperative complication occurred. Complications during hospitalisation occurred in six dogs, four of which were fatal. Of the five dogs discharged, one presented dead due to haemothorax after minor trauma seven days later. The four remaining dogs survived a median of 533 days; all of these dogs received a bovine pericardial valve. Based on our results, tricuspid valve replacement with bovine or porcine prosthetic valves is associated with a high incidence of complications. © 2017 British Small Animal Veterinary Association.
NASA Technical Reports Server (NTRS)
Farner, Bruce
2013-01-01
A moveable valve for controlling flow of a pressurized working fluid was designed. This valve consists of a hollow, moveable floating piston pressed against a stationary solid seat, and can use the working fluid to seal the valve. This open/closed, novel valve is able to use metal-to-metal seats, without requiring seat sliding action; therefore there are no associated damaging effects. During use, existing standard high-pressure ball valve seats tend to become damaged during rotation of the ball. Additionally, forces acting on the ball and stem create large amounts of friction. The combination of these effects can lead to system failure. In an attempt to reduce damaging effects and seat failures, soft seats in the ball valve have been eliminated; however, the sliding action of the ball across the highly loaded seat still tends to scratch the seat, causing failure. Also, in order to operate, ball valves require the use of large actuators. Positioning the metal-to-metal seats requires more loading, which tends to increase the size of the required actuator, and can also lead to other failures in other areas such as the stem and bearing mechanisms, thus increasing cost and maintenance. This novel non-sliding seat surface valve allows metal-to-metal seats without the damaging effects that can lead to failure, and enables large seating forces without damaging the valve. Additionally, this valve design, even when used with large, high-pressure applications, does not require large conventional valve actuators and the valve stem itself is eliminated. Actuation is achieved with the use of a small, simple solenoid valve. This design also eliminates the need for many seals used with existing ball valve and globe valve designs, which commonly cause failure, too. This, coupled with the elimination of the valve stem and conventional valve actuator, improves valve reliability and seat life. Other mechanical liftoff seats have been designed; however, they have only resulted in increased cost, and incurred other reliability issues. With this novel design, the seat is lifted by simply removing the working fluid pressure that presses it against the seat and no external force is required. By eliminating variables associated with existing ball and globe configurations that can have damaging effects upon a valve, this novel design reduces downtime in rocket engine test schedules and maintenance costs.
Effect of the sinus of valsalva on the closing motion of bileaflet prosthetic heart valves.
Ohta, Y; Kikuta, Y; Shimooka, T; Mitamura, Y; Yuhta, T; Dohi, T
2000-04-01
Conventional bileaflet prosthetic mechanical heart valves close passively with backflow. Naturally, the valve has problems associated with closure, such as backflow, water hammer effect, and fracture of the leaflet. On the other hand, in the case of the natural aortic valve, the vortex flow in the sinus of Valsalva pushes the leaflet to close, and the valve starts the closing motion earlier than the prosthetic valve as the forward flow decelerates. This closing mechanism is thought to decrease backflow at valve closure. In this study, we propose a new bileaflet mechanical valve resembling a drawbridge in shape, and the prototype valve was designed so that the leaflet closes with the help of the vortex flow in the sinus. The test valve was made of aluminum alloy, and its closing motion was compared to that of the CarboMedics (CM) valve. Both valves were driven by a computer controlled hydraulic mock circulator and were photographed at 648 frames/s by a high speed charge-coupled device (CCD) camera. Each frame of the valve motion image was analyzed with a personal computer, and the opening angles were measured. The flow rate was set as 5.0 L/min. The system was pulsed with 70 bpm, and the systolic/diastolic ratio was 0.3. Glycerin water was used as the circulation fluid at room temperature, and polystyrene particles were used to visualize the streamline. The model of the sinus of Valsalva was made of transparent silicone rubber. As a result, high speed video analysis showed that the test valve started the closing motion 41 ms earlier than the CM valve, and streamline analysis showed that the test valve had a closing mechanism similar to the natural one with the effect of vortex flow. The structure of the test valve was thought to be effective for soft closure and could solve problems associated with closure.
Karimi, Ashkan; Pourafshar, Negiin; Dibu, George; Beaver, Thomas M; Bavry, Anthony A
2017-06-01
A 55-year-old male with a history of two prior cardiac surgeries presented with decompensated heart failure due to severe bioprosthetic aortic valve insufficiency. A third operation was viewed prohibitively high risk and valve-in-valve trans-catheter aortic valve replacement was considered. There were however several high-risk features and technically challenging aspects including low coronary ostia height, poor visualization of the aortic sinuses, and difficulty in identification of the coplanar view due to severe aortic insufficiency, and a highly mobile aortic valve mass. After meticulous peri-procedural planning, trans-catheter aortic valve replacement was carried out with a SAPIEN 3 balloon-expandable valve without any complication. Strategies undertaken to navigate the technically challenging aspects of the case are discussed.
Liquid Nitrogen (Oxygen Simulent) Thermodynamic Venting System Test Data Analysis
NASA Technical Reports Server (NTRS)
Hedayat, A.; Nelson, S. L.; Hastings, L. J.; Flachbart, R. H.; Tucker, S. P.
2005-01-01
In designing systems for the long-term storage of cryogens in low gravity space environments, one must consider the effects of thermal stratification on excessive tank pressure that will occur due to environmental heat leakage. During low gravity operations, a Thermodynamic Venting System (TVS) concept is expected to maintain tank pressure without propellant resettling. The TVS consists of a recirculation pump, Joule-Thomson (J-T) expansion valve, and a parallel flow concentric tube heat exchanger combined with a longitudinal spray bar. Using a small amount of liquid extracted by the pump and passing it though the J-T valve, then through the heat exchanger, the bulk liquid and ullage are cooled, resulting in lower tank pressure. A series of TVS tests were conducted at the Marshall Space Flight Center using liquid nitrogen as a liquid oxygen simulant. The tests were performed at fill levels of 90%, 50%, and 25% with gaseous nitrogen and helium pressurants, and with a tank pressure control band of 7 kPa. A transient one-dimensional model of the TVS is used to analyze the data. The code is comprised of four models for the heat exchanger, the spray manifold and injector tubes, the recirculation pump, and the tank. The TVS model predicted ullage pressure and temperature and bulk liquid saturation pressure and temperature are compared with data. Details of predictions and comparisons with test data regarding pressure rise and collapse rates will be presented in the final paper.
Turbulence downstream of subcoronary stentless and stented aortic valves.
Funder, Jonas Amstrup; Frost, Markus Winther; Wierup, Per; Klaaborg, Kaj-Erik; Hjortdal, Vibeke; Nygaard, Hans; Hasenkam, J Michael
2011-08-11
Regions of turbulence downstream of bioprosthetic heart valves may cause damage to blood components, vessel wall as well as to aortic valve leaflets. Stentless aortic heart valves are known to posses several hemodynamic benefits such as larger effective orifice areas, lower aortic transvalvular pressure difference and faster left ventricular mass regression compared with their stented counterpart. Whether this is reflected by diminished turbulence formation, remains to be shown. We implanted either stented pericardial valve prostheses (Mitroflow), stentless valve prostheses (Solo or Toronto SPV) in pigs or they preserved their native valves. Following surgery, blood velocity was measured in the cross sectional area downstream of the valves using 10MHz ultrasonic probes connected to a dedicated pulsed Doppler equipment. As a measure of turbulence, Reynolds normal stress (RNS) was calculated at two different blood pressures (baseline and 50% increase). We found no difference in maximum RNS measurements between any of the investigated valve groups. The native valve had significantly lower mean RNS values than the Mitroflow (p=0.004), Toronto SPV (p=0.008) and Solo valve (p=0.02). There were no statistically significant differences between the artificial valve groups (p=0.3). The mean RNS was significantly larger when increasing blood pressure (p=0.0006). We, thus, found no advantages for the stentless aortic valves compared with stented prosthesis in terms of lower maximum or mean RNS values. Native valves have a significantly lower mean RNS value than all investigated bioprostheses. Copyright © 2011 Elsevier Ltd. All rights reserved.
Whitey SCHe Ball Valves Provide Test Port Isolation
DOE Office of Scientific and Technical Information (OSTI.GOV)
MISKA, C.R.
2000-09-15
These valves are 1/4 inch ball valves fabricated of 316 stainless steel. Packing is TFE (standard). They are used as normally closed isolation valves for test ports in the SCHe System between the gage root valve and the pressure indicator.
MOLNAR, ADRIAN; MURESAN, IOAN; TRIFAN, CATALIN; POP, DANA; SACUI, DIANA
2015-01-01
Background and aims The introduction of Duke’s criteria and the improvement of imaging methods has lead to an earlier and a more accurate diagnosis of infectious endocarditis (IE). The options for the best therapeutic approach and the timing of surgery are still a matter of debate and require a close colaboration between the cardiologist, the infectionist and the cardiac surgeon. Methods We undertook a retrospective, descriptive study, spanning over a period of five years (from January 1st, 2007 to December 31st, 2012), on 100 patients who underwent surgery for native valve infectious endocarditis in our unit. Results The patients’ age varied between 13 and 77 years (with a mean of 54 years), of which 85 were males (85%). The main microorganisms responsible for IE were: Streptococcus Spp. (21 cases – 21%), Staphylococcus Spp. (15 cases – 15%), and Enterococcus Spp. (9 cases – 9%). The potential source of infection was identified in 26 patients (26%), with most cases being in the dental area (16 cases – 16%). The lesions caused by IE were situated in the left heart in 96 patients (96%), mostly on the aortic valve (50 cases – 50%). In most cases (82%) we found preexisting endocardial lesions which predisposed to the development of IE, most of them being degenerative valvular lesions (38 cases – 38%). We performed the following surgical procedures: surgery on a single valve - aortic valve replacement (40 cases), mitral valve replacement (19 cases), mitral valve repair (1 case), surgery on more than one valve – mitral and aortic valve replacement (20 cases), aortic and tricuspid valve replacement (1 case), aortic valve replacement with a mechanical valve associated with mitral valve repair (5 cases), aortic valve replacement with a biological valve associated with mitral valve repair (2 cases), and mitral valve replacement with a mechanical valve combined with De Vega procedure on the tricuspid valve (1 case). In 5 patients (5%) the bacteriological examination of valve pieces excised during surgery was positive. In 3 cases it matched the germ identified in the hemocultures, and in 2 cases it evidenced another bacterium. Conclusion The overall mortality of 5% is well between the limits presented in literature, being higher (30%) in patients who required emergency surgery. For the patients who return into our clinic with prosthetic valve endocarditis, the mortality after surgery was even higher (50%). PMID:26609267
Design and CFD analysis of intake port and exhaust port for a 4 valve cylinder head engine
NASA Astrophysics Data System (ADS)
Latheesh, V. M.; Parthasarathy, P.; Baskaran, V.; Karthikeyan, S.
2018-02-01
In cylinder air motion in a compression ignition engine effects mixing of air-fuel, quality of combustion and emission produced. The primary objective is to design and analyze intake and the exhaust port for a four valve cylinder head to meet higher emission norms for a given diesel engine with two valves. In this work, an existing cylinder head designed for two valves was redesigned with 4 valves. The modern trend also confirms this approach. This is being followed in the design and development of new generation engines to meet the stringent environment norms, competition in market and demand for more fuel-efficient engines. The swirl ratio and flow coefficient were measured for different valve lifts using STAR CCM+. CFD results were validated with the two-valve cylinder experimental results. After validation, a comparison between two-valve and four-valve cylinder head was done. The conversion of two valve cylinder head to 4 valves may not support modern high swirl generating port layout and requires a trade-off between many design parameters.
Failing stentless Bioprostheses in patients with carcinoid heart valve disease.
Schaefer, Andreas; Sill, Bjoern; Schoenebeck, Jeannette; Schneeberger, Yvonne; Reichenspurner, Hermann; Gulbins, Helmut
2015-03-27
Carcinoid tumor with consecutive endocardial fibroelastosis of the right heart, known as carcinoid heart valve disease (CHVD) or Hedinger's syndrome, is accompanied by combined right-sided valvular dysfunction with regurgitation and stenosis of the affected valves. Cardiac surgery with replacement of the tricuspid and/or pulmonary valve is an established therapeutic option for patients with Hedinger's syndrome. Little is known about the long term outcome and the choice of prosthesis for the pulmonal position is still a matter of debate. The authors report three cases of pulmonary valve replacement with stentless bioprostheses (Medtronic Freestyle, Medtronic PLC, Minneapolis, MN, USA) due to severe pulmonary valve degeneration in consequence of Hedinger's syndrome. All patients presented with re-stenosis of the pulmonal valve conduit at the height of the anastomoses in a premature fashion. Due to the increased risk for repeat surgical valve replacement, two patients were treated by transcatheter heart valves. We do not recommend the replacement of the pulmonary valve with stentless bioprostheses in patients with CHVD. These valves presented with an extreme premature degeneration and consecutive re-stenosis and heart failure.
Wong, Sui-To; Wen, Eleanor; Fong, Dawson
2013-08-01
Malfunction of a Codman Hakim programmable valve due to jamming of its programmable component may necessitate shunt revision. The authors report a method for programming jammed Codman Hakim programmable valves by using a Strata II magnet and additional neodymium magnets. The programming method was derived after studying a jammed valve in the laboratory that was explanted from an 10-year-old boy with a history of fourth ventricle ependymoma. Programming the explanted valve with a Codman programmer failed, but rotating a Strata II magnet above the valve resulted in rotation of the spiral cam in the valve. It was found that the Strata II magnet could be used to program the jammed valve by rotating the magnet 90° or multiples of 90° above the valve. The strength of the magnetic field of the Strata II magnet was able to be increased by putting neodymium magnets on it. The programming method was then successfully used in a patient with a jammed Codman Hakim programmable valve. After successful programming using this method, clinical and radiological follow-up of the patient was advised.
Aerosol penetration through respirator exhalation valves.
Bellin, P; Hinds, W C
1990-10-01
Exhalation valves are a critical component of industrial respirators. They are designed to permit minimal inward leakage of air contaminants during inhalation and provide low resistance during exhalation. Under normal conditions, penetration of aerosol through exhalation valves is minimal. The exhalation valve is, however, a vulnerable component of a respirator and under actual working conditions may become dirty or damaged to the point of causing significant leakage. Aerosol penetration was measured for normal exhalation valves and valves compromised by paint or fine copper wires on the valve seat. Penetration increased with increasing wire diameter. A wire 250 microns in diameter allowed greater than 1% penetration into the mask cavity. Dirt or paint accumulated on the exhalation valve allowed a similar level of penetration. Work rate had little effect on observed penetration. Penetration decreased significantly with increasing aerosol particle size. The amount of material on the valve or valve seat necessary for significant (greater than 0.5%) inward leakage in a half-mask respirator could be readily observed by careful inspection of the exhalation valve and its seat in good lighting conditions.
Method and system for measuring gate valve clearances and seating force
Casada, D.A.; Haynes, H.D.; Moyers, J.C.; Stewart, B.K.
1996-01-30
Valve clearances and seating force, as well as other valve operational parameters, are determined by measuring valve stem rotation during opening and closing operations of a translatable gate valve. The magnitude of the stem rotation, and the relative difference between the stem rotation on opening and closing provides valuable data on the valve internals in a non-intrusive manner. 8 figs.
Energy conservation with automatic flow control valves
DOE Office of Scientific and Technical Information (OSTI.GOV)
Phillips, D.
Automatic flow control valves are offered in a wide range of sizes starting at 1/2 in. with flow rates of 0.5 gpm and up. They are also provided with materials and end connections to meet virtually any fan-coil system requirement. Among these are copper sweat type valves; ductile iron threaded valves; male/female threaded brass valves; and combination flow control/ball valves with union ends.
NASA Technical Reports Server (NTRS)
Hesser, R. J.; Gershman, R.
1975-01-01
A valve opening-response problem encountered during development of a control valve for the Skylab thruster attitude control system (TACS) is described. The problem involved effects of dynamic interaction among valves in the quad-redundant valve package. Also described is a detailed computer simulation of the quad-valve package which was helpful in resolving the problem.
Alskaf, Ebraham; McConkey, Hannah; Laskar, Nabila; Kardos, Attila
2016-06-20
The Medtronic ATS Open Pivot mechanical valve has been successfully used in heart valve surgery for more than two decades. We present the case of a patient who, 19 years following a tricuspid valve replacement with an ATS prosthesis as part of a triple valve operation following infective endocarditis, developed severe tricuspid regurgitation due to pannus formation.
Minakata, Kenji; Tanaka, Shiro; Tamura, Nobushige; Yanagi, Shigeki; Ohkawa, Yohei; Okonogi, Shuichi; Kaneko, Tatsuo; Usui, Akihiko; Abe, Tomonobu; Shimamoto, Mitsuomi; Takahara, Yoshiharu; Yamanaka, Kazuo; Yaku, Hitoshi; Sakata, Ryuzo
2017-07-25
The aim of this study was to assess the long-term outcomes of aortic valve replacement (AVR) with either mechanical or bioprosthetic valves according to age at operation.Methods and Results:A total of 1,002 patients (527 mechanical valves and 475 bioprosthetic valves) undergoing first-time AVR were categorized according to age at operation: group Y, age <60 years; group M, age 60-69 years; and group O, age ≥70 years). Outcomes were compared on propensity score analysis (adjusted for 28 variables). Hazard ratio (HR) was calculated using the Cox regression model with adjustment for propensity score with bioprosthetic valve as a reference (HR=1). There were no significant differences in overall mortality between mechanical and bioprosthetic valves for all age groups. Valve-related mortality was significantly higher for mechanical valves in group O (HR, 2.53; P=0.02). Reoperation rate was significantly lower for mechanical valves in group Y (HR, 0.16; P<0.01) and group M (no events for mechanical valves). Although the rate of thromboembolic events was higher in mechanical valves in group Y (no events for tissue valves) and group M (HR, 9.05; P=0.03), there were no significant differences in bleeding events between all age groups. The type of prosthetic valve used in AVR does not significantly influence overall mortality.
Tricuspid valve and percutaneous approach: No longer the forgotten valve!
Bouleti, Claire; Juliard, Jean-Michel; Himbert, Dominique; Iung, Bernard; Brochet, Eric; Urena, Marina; Dilly, Marie-Pierre; Ou, Phalla; Nataf, Patrick; Vahanian, Alec
2016-01-01
Tricuspid valve disease is mainly represented by tricuspid regurgitation (TR), which is a predictor of poor outcome. TR is usually secondary, caused by right ventricle pressure or volume overload, the leading cause being left-sided heart valve diseases. Tricuspid surgery for severe TR is recommended during left valve surgery, and consists of either a valve replacement or, most often, a tricuspid repair with or without prosthetic annuloplasty. When TR persists or worsens after left valvular surgery, redo isolated tricuspid surgery is associated with high mortality. In addition, a sizeable proportion of patients present with tricuspid surgery deterioration over time, and need a reintervention, which is associated with high morbi-mortality rates. In this context, and given the recent major breakthrough in the percutaneous treatment of aortic and mitral valve diseases, the tricuspid valve appears an appealing challenge, although it raises specific issues. The first applications of transcatheter techniques for tricuspid valve disease were valve-in-valve and valve-in-ring implantation for degenerated bioprosthesis or ring annuloplasty. Some concerns remain regarding prosthesis sizing, rapid ventricular pacing and the best approach, but these procedures appear to be safe and effective. More recently, bicuspidization using a transcatheter approach for the treatment of native tricuspid valve has been published, in two patients. Finally, other devices are in preclinical development. Copyright © 2015 Elsevier Masson SAS. All rights reserved.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Quake, Stephen R; Marcus, Joshua S; Hansen, Carl L
2015-01-13
Sieve valves for use in microfluidic device are provided. The valves are useful for impeding the flow of particles, such as chromatography beads or cells, in a microfluidic channel while allowing liquid solution to pass through the valve. The valves find particular use in making microfluidic chromatography modules.
NASA Astrophysics Data System (ADS)
Yao, Che; Li, Tao; Zhang, Hong; Zhou, Yanming
2017-08-01
In this paper, the characters of two control valves used for ammonia injection in SCR system are discussed. The linear/quadratic character between pressure drop/outlet flow rate and valve opening/dynamic pressure inlet are investigated using computational fluid dynamic (CFD) and response surface analysis (RSA) methods. The results show that the linear character of brake valve is significantly better than butterfly valve, which means that the brake valve is more suitable for ammonia injection adjustment than the butterfly valve.
High-performance space shuttle auxiliary propellant valve system
NASA Technical Reports Server (NTRS)
Smith, G. M.
1973-01-01
Several potential valve closures for the space shuttle auxiliary propulsion system (SS/APS) were investigated analytically and experimentally in a modeling program. The most promising of these were analyzed and experimentally evaluated in a full-size functional valve test fixture of novel design. The engineering investigations conducted for both model and scale evaluations of the SS/APS valve closures and functional valve fixture are described. Preliminary designs, laboratory tests, and overall valve test fixture designs are presented, and a final recommended flightweight SS/APS valve design is presented.
Herndon, Charles; Brown, Roger A.
2002-01-01
An apparatus and process for removing a ball valve is provided. The ball valve removal tool provides a handle sliding along the length of a shaft. One end of the shaft is secured within an interior cavity of a ball valve while the opposite end of the shaft defines a stop member. By providing a manual sliding force to the handle, the handle impacts the stop member and transmits the force to the ball valve. The direction of the force is along the shaft of the removal tool and disengages the ball valve from the ball valve housing.
Yoneyama, Fumiya; Okamura, Toru; Harada, Yorikazu; Okita, Yutaka
2018-02-01
A 13-year-old male presented with neoaortic root dilatation and severe aortic valve regurgitation 13 years following an arterial switch operation. The valve cusps were unbalanced due to an enlarged non-coronary cusp. A valve-sparing reimplantation with a cusp plication was performed which resulted in a competent valve with trivial regurgitation. Thus, even in an unbalanced cusp, valve-sparing reimplantation can be used for neoaortic root dilatation and valve regurgitation after an arterial switch operation. © 2018 Wiley Periodicals, Inc.
Investigation of the effect of different carbon film thickness on the exhaust valve
NASA Astrophysics Data System (ADS)
Karamangil, M. I.; Avci, A.; Bilal, H.
2008-03-01
Valves working under different loads and temperatures are the mostly forced engine elements. In an internal combustion engine, pressures and temperatures affecting on the valves vary with fuel type and the combustion characteristics of the fuel. Consequently, valves are exposed to different dynamic and thermal stress. In this study, stress distributions and temperature profiles on exhaust valve are obtained depending on different carbon film thickness. It is concluded that heat losses and valve temperatures decrease and valve surfaces are exposed to less thermal shocks with increasing carbon film thickness.
A low power, on demand electrothermal valve for wireless drug delivery applications
Li, Po-Ying; Givrad, Tina K.; Sheybani, Roya; Holschneider, Daniel P.; Maarek, Jean-Michel I.
2014-01-01
We present a low power, on demand Parylene MEMS electrothermal valve. A novel Ω-shaped thermal resistive element requires low power (~mW) and enables rapid valve opening (~ms). Using both finite element analysis and valve opening experiments, a robust resistive element design for improved valve opening performance in water was obtained. In addition, a thermistor, as an inrush current limiter, was added into the valve circuit to provide variable current ramping. Wireless activation of the valve using RF inductive power transfer was demonstrated. PMID:20024057
Belov, Iu V; Stepanenko, A B; Gens, A P; Charchian, E R; Savichev, D D
2007-01-01
Simultaneous surgical interventions on the aorta and valvular system of the heart were performed in four patients presenting with aortic dissections and aneurysms conditioned by Marfan's syndrome. The following reconstructive operations were carried out: 1) prosthetic repair of the aortic valve and the ascending portion of the aorta by means of a valve-containing conduit with replantation of the openings of the coronary arteries into the side of the prosthesis according to the Benthall - De Bono technique, annuloplasty of the tricuspid valve according to the De Vega technique, valvuloplasty of the mitral valve by the Alferi technique; 2) grafting of the aortic valve and the ascending portion of the aorta by means of a valve-containing conduit with replantation of the openings of the coronary arteries according to the Kabrol's technique, plasty of the tricuspid valve by the De Vega technique; 3) prosthetic repair of the aortic arch with distal wedge-like excision of the membrane of the dissection and directing the blood flow along the both channels, plasty of the mitral valve, plasty of the aortic valve and the ascending portion of the aorta with a valve-containing conduit, accompanied by replantation of the openings of the coronary arteries into the side of the graft according to the Benthall - De Bono technique; (4) plasty of the mitral valve with a disk graft through the fibrous ring of the aortic valve, prosthetic repair of the aortic valve and the ascending portion of the aorta with a valve-containing conduit, accompanied by replantation of the openings of the coronary arteries into the side according to the Benthall-De Bono technique.
Boersma, Doeke; Vink, Aryan; Moll, Frans L; de Borst, Gert J
2017-06-01
To evaluate the SailValve, a new self-expanding deep venous valve concept based on a single polytetrafluoroethylene cusp floating up and down in the bloodstream like a sail, acting as a flow regulator and allowing minimal reflux to reduce thrombogenicity. Both iliac veins of 5 pigs were implanted with SailValve devices; the first animal was an acute pilot experiment to show the feasibility of accurately positioning the SailValve via a femoral access. The other 4 animals were followed for 2 weeks (n=2) or 4 weeks (n=2) under a chronic implantation protocol. Patency and valve function were evaluated directly in all animals using ascending and descending phlebography after device placement and at termination in the chronic implant animals. For reasons of clinical relevance, a regimen of clopidogrel and calcium carbasalate was administered. Histological analysis was performed according to a predefined protocol by an independent pathologist. Deployment was technically feasible in all 10 iliac veins, and all were patent directly after placement. No perioperative or postoperative complications occurred. Ascending phlebograms in the follow-up animals confirmed the patency of all valves after 2 or 4 weeks. Descending phlebograms showed full function in 5 of 8 valves. Limited reflux was seen in 1 valve (4-week group), and the function in the remaining 2 valves (2-week group) was insufficient because of malpositioning. No macroscopic thrombosis was noted on histology. Histology in the follow-up groups revealed a progressive inflammatory reaction to the valves. This animal study shows the potential of the SailValve concept with sufficient valve function after adequate positioning and no (thrombogenic) occlusions after short-term follow-up. Future research is essential to optimize valve material and long-term patency.
Chiam, Paul Toon-Lim; Ewe, See-Hooi; Soon, Jia-Lin; Ho, Kay-Woon; Sin, Yong-Koong; Tan, Swee-Yaw; Lim, Soo-Teik; Koh, Tian-Hai; Chua, Yeow-Leng
2016-07-01
Percutaneous transcatheter aortic valve implantation (TAVI) has become an established therapy for inoperable and high-surgical-risk patients with severe aortic stenosis. Although TAVI in patients with degenerated surgical aortic bioprostheses (i.e. valve-in-valve TAVI) is increasingly reported in Western studies, such data is lacking in Asian patients. We describe the initial experience of valve-in-valve TAVI in Asia. Eight patients who underwent valve-in-valve TAVI due to degenerated aortic bioprostheses were enrolled. The mechanism of bioprosthetic valve failure was stenotic, regurgitation or mixed. All procedures were performed via transfemoral arterial access, using the self-expanding CoreValve prosthesis or balloon-expandable SAPIEN XT prosthesis. The mean age of the patients was 71.6 ± 13.2 years and five were male. Mean duration to surgical bioprosthesis degeneration was 10.2 ± 4.1 years. Valve-in-valve TAVI was successfully performed in all patients. CoreValve and SAPIEN XT prostheses were used in six and two patients, respectively. There were no deaths, strokes or permanent pacemaker requirement at 30 days, with one noncardiac mortality at one year. All patients experienced New York Heart Association functional class improvement. Post-procedure mean pressure gradients were 20 ± 11 mmHg and 22 ± 8 mmHg at 30 days and one year, respectively. Residual aortic regurgitation (AR) of more than mild severity occurred in one patient at 30 days. At one year, only one patient had mild residual AR. In our experience of valve-in-valve TAVI, procedural success was achieved in all patients without adverse events at 30 days. Good clinical and haemodynamic outcomes were sustained at one year. Copyright © Singapore Medical Association.
Advanced technology for space shuttle auxiliary propellant valves
NASA Technical Reports Server (NTRS)
Wichmann, H.
1973-01-01
Valves for the gaseous hydrogen/gaseous oxygen shuttle auxiliary propulsion system are required to feature low leakage over a wide temperature range coupled with high cycle life, long term compatibility and minimum maintenance. In addition, those valves used as thruster shutoff valves must feature fast response characteristics to achieve small, repeatable minimum impulse bits. These valve technology problems are solved by developing unique valve components such as sealing closures, guidance devices, and actuation means and by demonstrating two prototype valve concepts. One of the prototype valves is cycled over one million cycles without exceeding a leakage rate of 27 scc's per hour at 450 psia helium inlet pressure throughout the cycling program.
Kim, Jong Hun; Kim, Tae Youn; Choi, Jong Bum; Kuh, Ja Hong
2017-01-01
Patients requiring redo cardiac surgery for diseased heart valves other than mitral valves may show increased pressure gradients and reduced valve areas of previously placed mechanical mitral valves due to subvalvular pannus formation. We treated four women who had mechanical mitral valves inserted greater than or equal to 20 years earlier and who presented with circular pannus that protruded into the lower margin of the valve ring but did not impede leaflet motion. Pannus removal improved the haemodynamic function of the mitral valve. © The Author 2016. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.
Apparatus for passive removal of subsurface contaminants and mass flow measurement
Jackson, Dennis G [Augusta, GA; Rossabi, Joseph [Aiken, SC; Riha, Brian D [Augusta, GA
2003-07-15
A system for improving the Baroball valve and a method for retrofitting an existing Baroball valve. This invention improves upon the Baroball valve by reshaping the interior chamber of the valve to form a flow meter measuring chamber. The Baroball valve sealing mechanism acts as a rotameter bob for determining mass flow rate through the Baroball valve. A method for retrofitting a Baroball valve includes providing static pressure ports and connecting a measuring device, to these ports, for measuring the pressure differential between the Baroball chamber and the well. A standard curve of nominal device measurements allows the mass flow rate to be determined through the retrofitted Baroball valve.
Apparatus for passive removal of subsurface contaminants and volume flow measurement
Jackson, Dennis G.; Rossabi, Joseph; Riha, Brian D.
2002-01-01
A system for improving the Baroball valve and a method for retrofitting an existing Baroball valve. This invention improves upon the Baroball valve by reshaping the interior chamber of the valve to form a flow meter measuring chamber. The Baroball valve sealing mechanism acts as a rotameter bob for determining volume flow rate through the Baroball valve. A method for retrofitting a Baroball valve includes providing static pressure ports and connecting a measuring device, to these ports, for measuring the pressure differential between the Baroball chamber and the well. A standard curve of nominal device measurements allows the volume flow rate to be determined through the retrofitted Baroball valve.
Talavlikar, P. H.; Walbaum, P. R.; Kitchin, A. H.
1973-01-01
Twelve patients undergoing aortic and 28 undergoing mitral valve replacement with autologous fascia lata valves were studied before and six months after surgery. One aortic and 10 mitral valves were found to be significantly incompetent. Of the incompetent mitral valves, two appeared to have perivalvular leaks. Six of the remainder were associated with abnormal ventricular filling patterns. Valve failure was much less common when the design was modified to provide a loose cusp structure; out of 12 such valves none was incompetent. Transvalvular gradients persist with fascial valves though they are lower than with most mechanical prostheses. Ventricular function was greatly improved in successful aortic replacement but remained impaired in the case of mitral replacement. Valve failure appeared to be associated with, or accelerated by, haemodynamic stress rather than due to inevitable degenerative pathological processes. PMID:4731108
The Melody® valve and Ensemble® delivery system for transcatheter pulmonary valve replacement
McElhinney, Doff B; Hennesen, Jill T
2013-01-01
The Melody® transcatheter pulmonary valve (TPV) is a percutaneous valve system designed for the treatment of obstruction and/or regurgitation of prosthetic conduits placed between the right ventricle and pulmonary arteries in patients with congenital heart disease. In 2000, Melody TPV became the first transcatheter valve implanted in a human; in 2006 it became the first transcatheter valve commercially available anywhere in the world; and in 2010 it was launched as the first commercially available transcatheter valve in the United States. In this review, we present the clinical background against which the Melody valve was developed and implemented, introduce the rationale for and challenges of transcatheter valve technology for this population, outline the history and technical details of its development and use, and summarize currently available data concerning the performance of the device. PMID:23834411
Application of Model-based Prognostics to a Pneumatic Valves Testbed
NASA Technical Reports Server (NTRS)
Daigle, Matthew; Kulkarni, Chetan S.; Gorospe, George
2014-01-01
Pneumatic-actuated valves play an important role in many applications, including cryogenic propellant loading for space operations. Model-based prognostics emphasizes the importance of a model that describes the nominal and faulty behavior of a system, and how faulty behavior progresses in time, causing the end of useful life of the system. We describe the construction of a testbed consisting of a pneumatic valve that allows the injection of faulty behavior and controllable fault progression. The valve opens discretely, and is controlled through a solenoid valve. Controllable leaks of pneumatic gas in the testbed are introduced through proportional valves, allowing the testing and validation of prognostics algorithms for pneumatic valves. A new valve prognostics approach is developed that estimates fault progression and predicts remaining life based only on valve timing measurements. Simulation experiments demonstrate and validate the approach.
Locking apparatus for gate valves
Fabyan, J.; Williams, C.W.
A locking apparatus for fluid operated valves having a piston connected to the valve actuator which moves in response to applied pressure within a cylinder housing having a cylinder head, a catch block is secured to the piston, and the cylinder head incorporates a catch pin. Pressure applied to the cylinder to open the valve moves the piston adjacent to the cylinder head where the catch pin automatically engages the catch block preventing further movement of the piston or premature closure of the valve. Application of pressure to the cylinder to close the valve, retracts the catch pin, allowing the valve to close. Included are one or more selector valves, for selecting pressure application to other apparatus depending on the gate valve position, open or closed, protecting such apparatus from damage due to premature closing caused by pressure loss or operational error.
Bioprosthetic Aortic Valve Durability: A Meta-Regression of Published Studies.
Wang, Mansen; Furnary, Anthony P; Li, Hsin-Fang; Grunkemeier, Gary L
2017-09-01
To compare structural valve deterioration (SVD) among bioprosthetic aortic valve types, a PubMed search found 54 papers containing SVD-free curves extending to at least 10 years. The curves were digitized and fit to Weibull distributions, and the mean times to valve failure (MTTF) were calculated. Twelve valve models were collapsed into four valve types: Medtronic (Medtronic, Minneapolis, MN) and Edwards (Edwards Lifesciences, Irvine, CA) porcine; and Sorin (Sorin Group [now LivaNova], London, United Kingdom) and Edwards pericardial. Meta-regression found MTTF was associated with the patient's age, publication year, SVD definition, and valve type. Sorin pericardial valves had significantly lower risk-adjusted MTTF (higher SVD risk), and there were no significant differences in MTTF among the other three valve types. Copyright © 2017 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.
NASA Technical Reports Server (NTRS)
Tervo, John N.
1992-01-01
Valve plate made to translate as well as rotate. Valve opened and closed by turning shaft and lever. Interactions among lever, spring, valve plate, and face seal cause plate to undergo combination of translation and rotation so valve plate clears seal during parts of opening and closing motions.
Fluid mechanics of heart valves.
Yoganathan, Ajit P; He, Zhaoming; Casey Jones, S
2004-01-01
Valvular heart disease is a life-threatening disease that afflicts millions of people worldwide and leads to approximately 250,000 valve repairs and/or replacements each year. Malfunction of a native valve impairs its efficient fluid mechanic/hemodynamic performance. Artificial heart valves have been used since 1960 to replace diseased native valves and have saved millions of lives. Unfortunately, despite four decades of use, these devices are less than ideal and lead to many complications. Many of these complications/problems are directly related to the fluid mechanics associated with the various mechanical and bioprosthetic valve designs. This review focuses on the state-of-the-art experimental and computational fluid mechanics of native and prosthetic heart valves in current clinical use. The fluid dynamic performance characteristics of caged-ball, tilting-disc, bileaflet mechanical valves and porcine and pericardial stented and nonstented bioprostheic valves are reviewed. Other issues related to heart valve performance, such as biomaterials, solid mechanics, tissue mechanics, and durability, are not addressed in this review.
NASA Astrophysics Data System (ADS)
Matsuoka, Ken; Esumi, Motoki; Ikeguchi, Ken Bryan; Kasahara, Jiro; Matsuo, Akiko; Funaki, Ikkoh
We developed a novel coaxial rotary valve for a multi-tube PDE. Since this single valve can supply three different gases (fuel, oxidizer and purge gas) into a combustor, the unification of the valve systems for three different gases is possible by using our newly designed valve. A PDRE system can be simple and lightweight by using this valve, and thus its thrust-weight ratio can be increased. We proposed the design of a multi-tube rotary-valved PDRE system by this rotary valve. Moreover, in preparation for a multi-tube rotary-valved PDRE, we carried out the multi-cycle operation experiment by the single-tube rotary-valved PDRE system. The combustion wave velocity was measured to confirm the operation of the PDRE system. Deflagration-to-detonation transition (DDT) was confirmed and DDT distance decreased under the condition of high operation frequency. In addition, a maximum operation frequency was 159 Hz.
Monovalve with integrated fuel injector and port control valve, and engine using same
Milam, David M.
2001-11-06
An engine includes an engine casing that defines a hollow piston cavity separated from an exhaust passage and an intake passage by a valve seat. A gas exchange valve member is positioned adjacent the valve seat and is moveable between an open position and a closed position. The gas exchange valve member also defines an opening that opens into the hollow piston cavity. A needle valve member is positioned in the gas exchange valve member adjacent a nozzle outlet and is moveable between an inject position and a blocked position. A port control valve member, which has a hydraulic surface, is mounted around the gas exchange valve member and moveable between an intake position and an exhaust position. A pilot valve is moveable between a first position at which the port control hydraulic surface is exposed to a source of high pressure fluid, and a second position at which the port control hydraulic surface is exposed to a source of low pressure fluid.
46 CFR 52.01-120 - Safety valves and safety relief valves (modifies PG-67 through PG-73).
Code of Federal Regulations, 2012 CFR
2012-10-01
... 46 Shipping 2 2012-10-01 2012-10-01 false Safety valves and safety relief valves (modifies PG-67 through PG-73). 52.01-120 Section 52.01-120 Shipping COAST GUARD, DEPARTMENT OF HOMELAND SECURITY (CONTINUED) MARINE ENGINEERING POWER BOILERS General Requirements § 52.01-120 Safety valves and safety relief valves (modifies PG-67 through PG-73). (a)...
46 CFR 52.01-120 - Safety valves and safety relief valves (modifies PG-67 through PG-73).
Code of Federal Regulations, 2010 CFR
2010-10-01
... 46 Shipping 2 2010-10-01 2010-10-01 false Safety valves and safety relief valves (modifies PG-67 through PG-73). 52.01-120 Section 52.01-120 Shipping COAST GUARD, DEPARTMENT OF HOMELAND SECURITY (CONTINUED) MARINE ENGINEERING POWER BOILERS General Requirements § 52.01-120 Safety valves and safety relief valves (modifies PG-67 through PG-73). (a)...
46 CFR 52.01-120 - Safety valves and safety relief valves (modifies PG-67 through PG-73).
Code of Federal Regulations, 2013 CFR
2013-10-01
... 46 Shipping 2 2013-10-01 2013-10-01 false Safety valves and safety relief valves (modifies PG-67 through PG-73). 52.01-120 Section 52.01-120 Shipping COAST GUARD, DEPARTMENT OF HOMELAND SECURITY (CONTINUED) MARINE ENGINEERING POWER BOILERS General Requirements § 52.01-120 Safety valves and safety relief valves (modifies PG-67 through PG-73). (a)...
46 CFR 52.01-120 - Safety valves and safety relief valves (modifies PG-67 through PG-73).
Code of Federal Regulations, 2011 CFR
2011-10-01
... 46 Shipping 2 2011-10-01 2011-10-01 false Safety valves and safety relief valves (modifies PG-67 through PG-73). 52.01-120 Section 52.01-120 Shipping COAST GUARD, DEPARTMENT OF HOMELAND SECURITY (CONTINUED) MARINE ENGINEERING POWER BOILERS General Requirements § 52.01-120 Safety valves and safety relief valves (modifies PG-67 through PG-73). (a)...
46 CFR 52.01-120 - Safety valves and safety relief valves (modifies PG-67 through PG-73).
Code of Federal Regulations, 2014 CFR
2014-10-01
... 46 Shipping 2 2014-10-01 2014-10-01 false Safety valves and safety relief valves (modifies PG-67 through PG-73). 52.01-120 Section 52.01-120 Shipping COAST GUARD, DEPARTMENT OF HOMELAND SECURITY (CONTINUED) MARINE ENGINEERING POWER BOILERS General Requirements § 52.01-120 Safety valves and safety relief valves (modifies PG-67 through PG-73). (a)...
[Tricuspid valve insufficiency: what should be done?].
von Segesser, L K; Stauffer, J C; Delabays, A; Chassot, P G
1998-12-01
Tricuspid regurgitation is relatively common. Due to the progress made in echocardiography, its diagnosis is in general made readily and in reliable fashion. Basically one has to distinguish between functional tricuspid valve regurgitation due to volume and/or pressure overload of the right ventricle with intact valve structures versus tricuspid valve regurgitation due to pathologic valve structures. The clear identification of the regurgitation mechanism is of prime importance for the treatment. Functional tricuspid valve regurgitation can often be improved by medical treatment of heart failure, and eventually a tricuspid valve plasty can solve the problem. However, the presence of pathologic tricuspid valve structures makes in general more specific plastic surgical procedures and even prosthetic valve replacements necessary. A typical example for a structural tricuspid valve regurgitation is the case of a traumatic papillary muscle rupture. Due to the sudden onset, this pathology is not well tolerated and requires in general surgical reinsertion of the papillary muscle. In contrast, tricuspid valve regurgitation resulting from chronic pulmonary embolism with pulmonary artery hypertension, can be improved by pulmonary artery thrombendarteriectomy and even completely cured with an additional tricuspid annuloplasty. However, tricuspid regurgitations due to terminal heart failure are not be addressed with surgery directed to tricuspid valve repair or replacement. Heart transplantation, dynamic cardiomyoplasty or mechanical circulatory support should be evaluated instead.
A review of design and modeling of magnetorheological valve
NASA Astrophysics Data System (ADS)
Abd Fatah, Abdul Yasser; Mazlan, Saiful Amri; Koga, Tsuyoshi; Zamzuri, Hairi; Zeinali, Mohammadjavad; Imaduddin, Fitrian
2015-01-01
Following recent rapid development of researches in utilizing Magnetorheological (MR) fluid, a smart material that can be magnetically controlled to change its apparent viscosity instantaneously, a lot of applications have been established to exploit the benefits and advantages of using the MR fluid. One of the most important applications for MR fluid in devices is the MR valve, where it uses the popular flow or valve mode among the available working modes for MR fluid. As such, MR valve is widely applied in a lot of hydraulic actuation and vibration reduction devices, among them are dampers, actuators and shock absorbers. This paper presents a review on MR valve, discusses on several design configurations and the mathematical modeling for the MR valve. Therefore, this review paper classifies the MR valve based on the coil configuration and geometrical arrangement of the valve, and focusing on four different mathematical models for MR valve: Bingham plastic, Herschel-Bulkley, bi-viscous and Herschel-Bulkley with pre-yield viscosity (HBPV) models for calculating yield stress and pressure drop in the MR valve. Design challenges and opportunities for application of MR fluid and MR valve are also highlighted in this review. Hopefully, this review paper can provide basic knowledge on design and modeling of MR valve, complementing other reviews on MR fluid, its applications and technologies.
Bouma, Hjalmar R; Mungroop, Hubert E; de Geus, A Fred; Huisman, Daniel D; Nijsten, Maarten W N; Mariani, Massimo A; Scheeren, Thomas W L; Burgerhof, Johannes G M; Henning, Robert H; Epema, Anne H
2018-03-01
Perioperative acute kidney injury (AKI) is an important predictor of long-term all-cause mortality after coronary artery bypass (CABG). However, the effect of AKI on long-term mortality after cardiac valve operations is hitherto undocumented. Perioperative renal injury and long-term all-cause mortality after valve operations were studied in a prospective cohort of patients undergoing solitary valve operations (n=2,806) or valve operations combined with CABG (n=1,260) with up to 18 years of follow-up. Postoperative serum creatinine increase was classified according to AKI 0-3. Patients undergoing solitary CABG (n=4,938) with cardiopulmonary bypass served as reference. In both valve and valve+CABG operations, postoperative renal injury of AKI stage 1 or higher was progressively associated with an increase in long-term mortality (HR 2.27, p<0.05 for valve; HR 1.65, p<0.05 for valve operations combined with CABG; HR 1.56, p<0.05 for CABG). Notably, the mortality risk increased already substantially at serum creatinine rises of 10-25%, i.e. far below the threshold for AKI stage 1 after valve operations (HR 1.39, p<0.05), but not after valve operations combined with CABG or CABG only. An increase in serum creatinine by more than 10%during the first week following valve operation is associated with an increased risk for long-term mortality following cardiac valve operation. Thus, AKI-classification clearly underestimates long-term mortality risk in patients undergoing valve operations. Copyright © 2018 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.
Valve replacement in children: a challenge for a whole life.
Henaine, Roland; Roubertie, François; Vergnat, Mathieu; Ninet, Jean
2012-10-01
Valvular pathology in infants and children poses numerous challenges to the paediatric cardiac surgeon. Without question, valvular repair is the goal of intervention because restoration of valvular anatomy and physiology using native tissue allows for growth and a potentially better long-term outcome. When reconstruction fails or is not feasible, valve replacement becomes inevitable. Which valve for which position is controversial. Homograft and bioprosthetic valves achieve superior haemodynamic results initially but at the cost of accelerated degeneration. Small patient size and the risk of thromboembolism limit the usefulness of mechanical valves, and somatic outgrowth is an universal problem with all available prostheses. The goal of this article is to address valve replacement options for all four valve positions within the paediatric population. We review current literature and our practice to support our preferences. To summarize, a multitude of opinions and surgical experiences exist. Today, the valve choices that seem without controversy are bioprosthetic replacement of the tricuspid valve and Ross or Ross-Konno procedures when necessary for the aortic valve. On the other hand, bioprostheses may be implanted when annular pulmonary diameter is adequate; if not or in case of right ventricular outflow tract discontinuity, it is better to use a pulmonary homograft with the Ross procedure. Otherwise, a valved conduit. Mitral valve replacement remains the most problematic; the mechanical prosthesis must be placed in the annular position, avoiding oversizing. Future advances with tissue-engineered heart valves for all positions and new anticoagulants may change the landscape for valve replacement in the paediatric population. Copyright © 2012 Elsevier Masson SAS. All rights reserved.
Moche, Jason A; Cohen, Justin C; Pearlman, Steven J
2013-07-01
The objective of this work was to explore the utility of axial computed tomography (CT) imaging to objectively define a narrow internal nasal valve, and compare those findings with clinical examination and patient complaint. Retrospective review from a single facial plastic surgery center. We reviewed 40 consecutive patients evaluated for either sinusitis or nasal airway obstruction for which a CT scan was obtained at a single radiology institution. Thirty-six complete office records were examined for the presence of clinical internal valve narrowing and complaints of nasal obstruction. In total, 72 internal nasal valves were analyzed using axial plane CT and measurements were compared to clinical findings and presence of airway obstruction. Measured valve areas for clinically normal internal nasal valves averaged 0.47 cm(2) vs 0.28 cm(2) for clinically narrow valves, a decrease of 40.4%. In unobstructed nasal airways the valve area averaged 0.51 cm(2) vs 0.38 cm(2) in obstructed airways, a difference of 25.5%. A radiographically measured valve area of <0.30 cm(2) suggests clinical narrowing with a sensitivity of 71.4%, specificity of 88.9%, positive predictive value of 62.5%, and negative predictive value of 92.3%. Using standard axial CT imaging we describe an objective method of radiographically evaluating the nasal valve, demonstrating strong correlation with physical examination and patient complaint. Additionally, radiographic valve areas can be used to screen for clinically narrow nasal valves with good sensitivity and specificity, providing a novel straightforward method for nasal valve assessment. © 2012 ARS-AAOA, LLC.
A new paradigm for obtaining marketing approval for pediatric-sized prosthetic heart valves.
Yoganathan, Ajit P; Fogel, Mark; Gamble, Susan; Morton, Michael; Schmidt, Paul; Secunda, Jeff; Vidmar, Sara; Del Nido, Pedro
2013-10-01
Congenital heart valve disease is one of the most common abnormalities in children. There are limited technological solutions available for treating children with congenital heart valve diseases. The aim of this study is to provide the details of the consensus reached in terms of pediatric definitions, design approach, in vitro testing, and clinical trials, which may be used as guidance for developing prosthetic heart valves for the pediatric indication. In stark contrast to the various designs of adult-sized replacement valves available in the market, there are no Food and Drug Administration (FDA)-approved prosthetic heart valves available for use in the pediatric population. There is a pressing need for FDA-approved pediatric valve devices in the United States. The pediatric patient population has been typically excluded from replacement heart valve trials for several reasons. In January 2010, heart valve manufacturers and pediatric clinicians collaborated with academicians and FDA staff in a workshop to suggest ways to successfully evaluate pediatric prosthetic valves and conduct pediatric clinical trials to provide acceptable heart valve replacement options for this patient population. Recommendations, derived from ISO 5840:2005 and the 2010 FDA Draft Replacement Heart Valve Guidance, are provided for hydrodynamic, durability, and fatigue testing. The article specifically addresses in vitro and premarket and postmarket approval clinical studies that should be considered by a heart valve manufacturer for obtaining regulatory approval of pediatric sizes of prosthetic heart valve designs that are already approved for adult clinical use. Copyright © 2013 The American Association for Thoracic Surgery. Published by Mosby, Inc. All rights reserved.
Zhang, Ronghui; You, Feng; Lv, Zhihan; He, Zhaocheng; Wang, Haiwei; Huang, Ling
2016-12-12
In order to prevent the backward flow of piezoelectric pumps, this paper presents a single-active-chamber piezoelectric membrane pump with multiple passive check valves. Under the condition of a fixed total number of passive check valves, by means of changing the inlet valves and outlet valves' configuration, the pumping characteristics in terms of flow rate and backpressure are experimentally investigated. Like the maximum flow rate and backpressure, the testing results show that the optimal frequencies are significantly affected by changes in the number inlet valves and outlet valves. The variation ratios of the maximum flow rate and the maximum backpressure are up to 66% and less than 20%, respectively. Furthermore, the piezoelectric pump generally demonstrates very similar flow rate and backpressure characteristics when the number of inlet valves in one kind of configuration is the same as that of outlet valves in another configuration. The comparison indicates that the backflow from the pumping chamber to inlet is basically the same as the backflow from the outlet to the pumping chamber. No matter whether the number of inlet valves or the number of outlet valves is increased, the backflow can be effectively reduced. In addition, the backpressure fluctuation can be significantly suppressed with an increase of either inlet valves or outlet valves. It also means that the pump can prevent the backflow more effectively at the cost of power consumption. The pump is very suitable for conditions where more accurate flow rates are needed and wear and fatigue of check valves often occur.
Yoon, Hakyoung; Kim, Jaehwan; Nahm, Sang-Soep; Eom, Kidong
2017-07-11
Congenital pulmonary valve stenosis and patent ductus arteriosus are common congenital heart defects in dogs. However, concurrence of atypical pulmonary valve stenosis and patent ductus arteriosus is uncommon. This report describes the anatomic, histopathologic, and echocardiographic features in a dog with concomitant pulmonary valve stenosis and patent ductus arteriosus with atypical pulmonary valve dysplasia that included a fibrous band of tissue. A 1.5-year-old intact female Chihuahua dog weighing 3.3 kg presented with a continuous grade VI cardiac murmur, poor exercise tolerance, and an intermittent cough. Echocardiography indicated pulmonary valve stenosis, a thickened dysplastic valve without annular hypoplasia, and a type IIA patent ductus arteriosus. The pulmonary valve was thick line-shaped in systole and dome-shaped towards the right ventricular outflow tract in diastole. The dog suffered a fatal cardiac arrest during an attempted balloon pulmonary valvuloplasty. Necropsy revealed pulmonary valve dysplasia, commissural fusion, and incomplete opening and closing of the pulmonary valve because of a fibrous band of tissue causing adhesion between the right ventricular outflow tract and the dysplastic intermediate cusp of the valve. A fibrous band of tissue between the right ventricular outflow track and the pulmonary valve should be considered as a cause of pulmonary valve stenosis. Pulmonary valve stenosis and patent ductus arteriosus can have conflicting effects on diastolic and systolic dysfunction, respectively. Therefore, beta-blockers should always be used carefully, particularly in patients with a heart defect where there is concern about left ventricular systolic function.
In Vitro Evaluation of a Novel Hemodynamically Optimized Trileaflet Polymeric Prosthetic Heart Valve
Claiborne, Thomas E.; Sheriff, Jawaad; Kuetting, Maximilian; Steinseifer, Ulrich; Slepian, Marvin J.; Bluestein, Danny
2013-01-01
Calcific aortic valve disease is the most common and life threatening form of valvular heart disease, characterized by stenosis and regurgitation, which is currently treated at the symptomatic end-stages via open-heart surgical replacement of the diseased valve with, typically, either a xenograft tissue valve or a pyrolytic carbon mechanical heart valve. These options offer the clinician a choice between structural valve deterioration and chronic anticoagulant therapy, respectively, effectively replacing one disease with another. Polymeric prosthetic heart valves (PHV) offer the promise of reducing or eliminating these complications, and they may be better suited for the new transcatheter aortic valve replacement (TAVR) procedure, which currently utilizes tissue valves. New evidence indicates that the latter may incur damage during implantation. Polymer PHVs may also be incorporated into pulsatile circulatory support devices such as total artificial heart and ventricular assist devices that currently employ mechanical PHVs. Development of polymer PHVs, however, has been slow due to the lack of sufficiently durable and biocompatible polymers. We have designed a new trileaflet polymer PHV for surgical implantation employing a novel polymer—xSIBS—that offers superior bio-stability and durability. The design of this polymer PHV was optimized for reduced stresses, improved hemodynamic performance, and reduced thrombogenicity using our device thrombogenicity emulation (DTE) methodology, the results of which have been published separately. Here we present our new design, prototype fabrication methods, hydrodynamics performance testing, and platelet activation measurements performed in the optimized valve prototype and compare it to the performance of a gold standard tissue valve. The hydrodynamic performance of the two valves was comparable in all measures, with a certain advantage to our valve during regurgitation. There was no significant difference between the platelet activation rates of our polymer valve and the tissue valve, indicating that similar to the latter, its recipients may not require anticoagulation. This work proves the feasibility of our optimized polymer PHV design and brings polymeric valves closer to clinical viability. PMID:23445066
O'Sullivan, Katie E; Gough, Aideen; Segurado, Ricardo; Barry, Mitchel; Sugrue, Declan; Hurley, John
2014-05-01
Paravalvular regurgitation (PVR) following transcatheter aortic valve implantation (TAVI) is associated with poor survival. The two main valve delivery systems used to date differ significantly in both structure and deployment technique. The primary objective of this study was to perform a systematic review and meta-analysis of studies identifying PVR in patients post-TAVI using Medtronic CoreValve (MCV) and Edward Sapien (ES) valves in order to identify whether a significant difference exists between valve types. The secondary objective was to identify additional factors predisposing to PVR to provide an overview of the other associated considerations. A systematic review and meta-analysis of the current literature to identify PVR rate in patients with MCV and ES valves was performed. We also sought to examine other factors predisposing to PVR. A total of 5910 patients were identified from 9 studies. PVR rates for MCV and ES were analysed. MCV was associated with a higher PVR rate of 15.75% [95% confidence interval (CI) 12.48-19.32] compared with ES 3.93% [95% CI 1.05-8.38]. We separately reviewed predisposing factors associated with PVR. A formal comparison of the MCV and ES valve leakage rates by mixed-effects meta-regression with a fixed-effect moderator variable for valve type (MCV or ES) suggested a statistically significant difference in leakage rate between the two valve types (P = 0.0002). Unfavourable anatomical and pathological factors as well as valve choice have an impact on rates of PVR. Additionally, certain anatomical features dictate valve choice. A direct comparison of all the predisposing factors at this time is not possible and will require prospective multivariate analysis. There is, however, a significant difference in the PVR rates between valves based on the published observational data available to date. The ES valve associated with a lower incidence of PVR overall; therefore, we conclude that valve choice is indeed a significant determinant of PVR post-TAVI.
Triple valve surgery: a 25-year experience.
Yilmaz, Mustafa; Ozkan, Murat; Böke, Erkmen
2004-09-01
Surgical treatment of rheumatic valvular disease still constitutes a significant number of cardiac operations in developing countries. Despite improvements in myocardial protection and cardiopulmonary bypass techniques, triple valve operations (aortic, mitral and tricuspid valves) are still challenging because of longer duration of cardiopulmonary bypass and higher degree of myocardial decompensation. This study was instituted in order to assess results of triple valve surgery. Between 1977 and 2002, 34 patients underwent triple valve surgery in our clinic by the same surgeon (EB). Eleven patients underwent triple valve replacement (32.4%) and 23 underwent tricuspid valve annuloplasty with aortic and mitral valve replacements (67.6%). There was no significant difference between the two groups of patients who underwent triple valve replacement and aortic and mitral valve replacement with tricuspid valve annuloplasty. There were 4 hospital deaths (11.8%) occurring within 30 days. The duration of follow-up for 30 survivors ranged from 6 to 202 months (mean 97 months). The actuarial survival rates were 85%, 72%, and 48% at 5, 10, and 15 years respectively. Actuarial freedom from reoperation rates at 5, 10, and 15 years was 86.3%, 71.9%, and 51.2%, respectively. Freedom from cerebral thromboembolism and anticoagulation-related hemorrhage rates, expressed in actuarial terms was 75.9% and 62.9% at 5 and 10 years. Major cerebral complications occurred in 10 of the 30 patients. We prefer replacing, if repairing is not possible, the tricuspid valve, with a bileaflet mechanical prosthesis in a patient with valve replacement of the left heart who will be anticoagulated in order to avoid unfavorable properties of bioprosthesis like degeneration and of old generation mechanical prosthesis like thrombosis and poor hemodynamic function. In recent years, results of triple valve surgery either with tricuspid valve conservation or valve replacement in suitable cases have become encouraging with improvements in surgical techniques and myocardial preservation methods.
Trans-apical aortic valve implantation in patients with severe calcification of the ascending aorta.
Buz, Semih; Pasic, Miralem; Unbehaun, Axel; Drews, Thorsten; Dreysse, Stephan; Kukucka, Marian; Mladenow, Alexander; Hetzer, Roland
2011-08-01
In patients with calcification of the ascending aorta, postoperative stroke and mortality rates remain high after conventional aortic valve replacement, but the results of trans-apical aortic valve implantation in these patients are not known. We evaluate the outcome of trans-apical aortic valve implantation in patients with severely calcified ascending aorta in a single center with expanded procedural experience. Between April 2008 and July 2010, 258 patients underwent trans-apical aortic valve implantation using Edwards Sapien valve. By computed tomography (CT) scan, we identified 46 (18%) patients with severe calcification of the ascending aorta (16 with porcelain aorta and 30 with severe, but not complete, calcification). Of 46 patients (mean age 77 ± 10 years, range 63-90 years; EuroSCORE (European System for Cardiac Operative Risk Evaluation) 45 ± 22%; STS (Society of Thoracic Surgeons) score 23 ± 13) with calcified aorta, 15 received 23-mm valves and 31 patients 26-mm valves. Primary valve implantation was successful in 44 patients and a second valve was implanted (valve-in-valve) in two. Six patients underwent concomitant interventions (three elective percutaneous coronary intervention (PCI), one off-pump coronary artery bypass (OPCAB), one tricuspid valve reconstruction, and one left-ventricular (LV) aneurysmectomy). The final procedural results showed valve incompetence (trace or grade 1) in 17 (37%) patients and paravalvular leak in 15 (32.6%) (trace in 10 and grade 1 in five). There was no 30-day mortality. Postoperatively, cranial CT showed new cerebral ischemia areas in three patients (6.2%), but only one patient (2.1%) experienced postoperative neurological deficit (temporary aphasia). Survival at 6 and 12 months was 88% and 85.2%, respectively. Trans-apical aortic valve implantation can be performed safely in patients with aortic valve stenosis and severe calcification of the ascending aorta. Copyright © 2010 European Association for Cardio-Thoracic Surgery. Published by Elsevier B.V. All rights reserved.
Sritharan, Deepa; Fathi, Parinaz; Weaver, Jason D; Retta, Stephen M; Wu, Changfu; Duraiswamy, Nandini
2018-06-12
After implantation of a transcatheter bioprosthetic heart valve its original circular circumference may become distorted, which can lead to changes in leaflet coaptation and leaflets that are stretched or sagging. This may lead to early structural deterioration of the valve as seen in some explanted transcatheter heart valves. Our in vitro study evaluates the effect of leaflet deformations seen in elliptical configurations on the damage patterns of the leaflets, with circular valve deformation as the control. Bovine pericardial tissue heart valves were subjected to accelerated wear testing under both circular (N = 2) and elliptical (N = 4) configurations. The elliptical configurations were created by placing the valve inside custom-made elliptical holders, which caused the leaflets to sag or stretch. The hydrodynamic performance of the valves was monitored and high resolution images were acquired to evaluate leaflet damage patterns over time. In the elliptically deformed valves, sagging leaflets experienced more damage from wear compared to stretched leaflets; the undistorted leaflets of the circular valves experienced the least leaflet damage. Free-edge thinning and tearing were the primary modes of damage in the sagging leaflets. Belly region thinning was seen in the undistorted and stretched leaflets. Leaflet and fabric tears at the commissures were seen in all valve configurations. Free-edge tearing and commissure tears were the leading cause of valve hydrodynamic incompetence. Our study shows that mechanical wear affects heart valve pericardial leaflets differently based on whether they are undistorted, stretched, or sagging in a valve configuration. Sagging leaflets are more likely to be subjected to free-edge tear than stretched or undistorted leaflets. Reducing leaflet stress at the free edge of non-circular valve configurations should be an important factor to consider in the design and/or deployment of transcatheter bioprosthetic heart valves to improve their long-term performance.
14 CFR 23.995 - Fuel valves and controls.
Code of Federal Regulations, 2010 CFR
2010-01-01
... AIRWORTHINESS STANDARDS: NORMAL, UTILITY, ACROBATIC, AND COMMUTER CATEGORY AIRPLANES Powerplant Fuel System... valve rapidly after it has been closed. (c) Each valve and fuel system control must be supported so that... lines connected to the valve. (d) Each valve and fuel system control must be installed so that gravity...
14 CFR 23.995 - Fuel valves and controls.
Code of Federal Regulations, 2012 CFR
2012-01-01
... AIRWORTHINESS STANDARDS: NORMAL, UTILITY, ACROBATIC, AND COMMUTER CATEGORY AIRPLANES Powerplant Fuel System... valve rapidly after it has been closed. (c) Each valve and fuel system control must be supported so that... lines connected to the valve. (d) Each valve and fuel system control must be installed so that gravity...
14 CFR 23.995 - Fuel valves and controls.
Code of Federal Regulations, 2014 CFR
2014-01-01
... AIRWORTHINESS STANDARDS: NORMAL, UTILITY, ACROBATIC, AND COMMUTER CATEGORY AIRPLANES Powerplant Fuel System... valve rapidly after it has been closed. (c) Each valve and fuel system control must be supported so that... lines connected to the valve. (d) Each valve and fuel system control must be installed so that gravity...
14 CFR 23.995 - Fuel valves and controls.
Code of Federal Regulations, 2013 CFR
2013-01-01
... AIRWORTHINESS STANDARDS: NORMAL, UTILITY, ACROBATIC, AND COMMUTER CATEGORY AIRPLANES Powerplant Fuel System... valve rapidly after it has been closed. (c) Each valve and fuel system control must be supported so that... lines connected to the valve. (d) Each valve and fuel system control must be installed so that gravity...
14 CFR 23.995 - Fuel valves and controls.
Code of Federal Regulations, 2011 CFR
2011-01-01
... AIRWORTHINESS STANDARDS: NORMAL, UTILITY, ACROBATIC, AND COMMUTER CATEGORY AIRPLANES Powerplant Fuel System... valve rapidly after it has been closed. (c) Each valve and fuel system control must be supported so that... lines connected to the valve. (d) Each valve and fuel system control must be installed so that gravity...
NASA Technical Reports Server (NTRS)
Macglashan, W. F., Jr.
1980-01-01
Zero-leakage valve has fluid-sealing diaphragm support and flat sievelike sealing surface. Diaphragm-support valve is easy to fabricate and requires minimum maintenance. Potential applications include isolation valve for waste systems and remote air-actuated valve. Device is also useful in controlling flow of liquid fluorine and corrosive fluids at high pressures.
Liquid rocket valve components
NASA Technical Reports Server (NTRS)
1973-01-01
A monograph on valves for use with liquid rocket propellant engines is presented. The configurations of the various types of valves are described and illustrated. Design criteria and recommended practices for the various valves are explained. Tables of data are included to show the chief features of valve components in use on operational vehicles.
Hardee, Harry C.
1991-01-01
A rotary pneumatic valve which is thrust balanced and the pneumatic pressure developed produces only radial loads on the valve cylinder producing negligible resistance and thus minimal torque on the bearings of the valve. The valve is multiplexed such that at least two complete switching cycles occur for each revolution of the cylinder spindle.
DOE Office of Scientific and Technical Information (OSTI.GOV)
McKnight, Timothy E.; Johnson, Anthony; Moise, Kenneth J.
Valve implant systems positionable within a flow passage, the systems having an inlet, an outlet, and a remotely activatable valve between the inlet and outlet, with the valves being operable to provide intermittent occlusion of the flow path. A remote field is applied to provide thermal or magnetic activation of the valves.
Transapical JenaValve in a patient with mechanical mitral valve prosthesis.
O' Sullivan, Katie E; Casserly, Ivan; Hurley, John
2015-04-01
We report the first case of transcatheter aortic valve replacement implantation using JenaValve™ in a patient with mechanical mitral valve prosthesis. We believe that the design features of this valve may be particularly suited for use in this setting. © 2014 Wiley Periodicals, Inc.
46 CFR 56.20-15 - Valves employing resilient material.
Code of Federal Regulations, 2010 CFR
2010-10-01
....20-15 Shipping COAST GUARD, DEPARTMENT OF HOMELAND SECURITY (CONTINUED) MARINE ENGINEERING PIPING SYSTEMS AND APPURTENANCES Valves § 56.20-15 Valves employing resilient material. (a) A valve in which the... piping system manifolds; (ii) Isolation valves in cross-connects between two piping systems, at least one...
Magnetically operated check valve
NASA Astrophysics Data System (ADS)
Morris, Brian G.; Bozeman, Richard J., Jr.
1993-03-01
A magnetically operated check valve is disclosed having, in one aspect, a valve body and a movable poppet disposed therein. A magnet attracts the poppet to hold the valve shut until the force of fluid flow through the valve overcomes the magnetic attraction and moves the poppet to an unseated, open position. The poppet and magnet are configured and disposed to trap a magnetically attracted particulate and prevent it from flowing to a valve seating region.
Development of a Calibration Rig for a Large Multi-Component Rotor Balance
2000-05-01
valve pressure reducer pressure manifold pressure switch pressure transducer pressure relief valve pressure gage off on control valve pressure switch on...Each of the four manifolds has been equipped with a pressure switch , a pressure transducer, a pressure gage, and a pressure relief valve. If the...valve. A pressure switch is installed between the servo valve and the actuator. This pressure switch is used as a diagnostic indicator by the
Line Fluid Actuated Valve Development Program. [for application on the space shuttle
NASA Technical Reports Server (NTRS)
Lynch, R. A.
1975-01-01
The feasibility of a line-fluid actuated valve design for potential application as a propellant-control valve on the space shuttle was examined. Design and analysis studies of two prototype valve units were conducted and demonstrated performance is reported. It was shown that the line-fluid actuated valve concept offers distinct weight and electrical advantages over alternate valve concepts. Summaries of projected performance and design goals are also included.
Valve system incorporating single failure protection logic
Ryan, Rodger; Timmerman, Walter J. H.
1980-01-01
A valve system incorporating single failure protective logic. The system consists of a valve combination or composite valve which allows actuation or de-actuation of a device such as a hydraulic cylinder or other mechanism, integral with or separate from the valve assembly, by means of three independent input signals combined in a function commonly known as two-out-of-three logic. Using the input signals as independent and redundant actuation/de-actuation signals, a single signal failure, or failure of the corresponding valve or valve set, will neither prevent the desired action, nor cause the undesired action of the mechanism.
System and method for controlling hydraulic pressure in electro-hydraulic valve actuation systems
Brennan, Daniel G; Marriott, Craig D; Cowgill, Joel; Wiles, Matthew A; Patton, Kenneth James
2014-09-23
A control system for an engine includes a first lift control module and a second lift control module. The first lift control module increases lift of M valves of the engine to a predetermined valve lift during a period before disabling or re-enabling N valves of the engine. The second lift control module decreases the lift of the M valves to a desired valve lift during a period after enabling or re-enabling the N valves of the engine, wherein N and M are integers greater than or equal to one.
NASA Technical Reports Server (NTRS)
Robbins, John S.
1988-01-01
Unit removed with minimal disturbance. Valve inlet and outlet ports adjacent to each other on same side of valve body. Ports inserted into special manifold on fluid line. Valve body attached to manifold by four bolts or, alternatively, by toggle clamps. Electromechanical actuator moves in direction parallel to fluid line to open and close valve. When necessary to clean valve, removed simply by opening bolts or toggle clamps. No need to move or separate ports of fluid line. Valve useful where disturbance of fluid line detrimental or where fast maintenance essential - in oil and chemical industries, automotive vehicles, aircraft, and powerplants.
Programmable valve shunts: are they really better?
Kataria, Rashim; Kumar, Vimal; Mehta, Veer Singh
2012-01-01
Programmable valve shunts allows selection of opening pressure of shunt valve. In the presented article, a unique complication pertaining to programmable shunts has been discussed. A 5-year-old boy who had tectal plate low grade glioma with obstructive hydrocephalus was managed with Codman programmable ventriculoperitoneal shunt. There was a spontaneous change in the opening pressure of the shunt valve leading to shunt malfunction. Routinely used household appliances produce a magnetic field strong enough to cause change in the setting of shunt valve pressure and may lead to valve malfunction. Other causes of programmable valve malfunction also discussed.
[Pannus Formation Two Years after Bioprosthetic Aortic Valve Implantation;Report of a Case].
Ono, Kimiyo; Kuroda, Hiroaki
2015-08-01
We report a case of early deterioration of the bioprosthetic aortic valve 23 months postoperatively. A 77-year-old man who had undergone aortic valve replacement with a 23-mm Epic valve( St. Jude Medical [SJM])presented to us after a syncopal episode. Echocardiography revealed severe aortic stenosis, and redo aortic valve replacement with a 21-mm SJM mechanical valve was performed. All 3 cusps of the tissue valve were thickened by fibrous pannus overgrowth. Neither calcification nor invasion of inflammatory cells was observed. The cause of pannus formation at such an early stage after implantation remains unknown.
Ozkan, Mehmet; Astarcioglu, Mehmet Ali; Karakoyun, Suleyman; Balkanay, Mehmet
2012-02-01
Obstruction to a prosthetic cardiac valve is a well-recognized complication of cardiac valve replacement. Malfunction of the mobile component of a prosthetic valve to open or close correctly may occur in consequence of intrinsic or extrinsic causes (thrombus, vegetation, entrapment of left ventricular myocardium, suture entanglement, and pannus formation) that may result prosthetic valve stenosis and/or insufficiency. In the case we report a 48-year-old female with valve dysfunction occurred early after surgery, as one valve leaflet was only able to partially open due to suture entrapment. © 2011, Wiley Periodicals, Inc.
Dual motion valve with single motion input
NASA Technical Reports Server (NTRS)
Belew, Robert (Inventor)
1987-01-01
A dual motion valve includes two dual motion valve assemblies with a rotary input which allows the benefits of applying both rotary and axial motion to a rotary sealing element with a plurality of ports. The motion of the rotary sealing element during actuation provides axial engagement of the rotary sealing element with a stationary valve plate which also has ports. Fluid passages are created through the valve when the ports of the rotary sealing element are aligned with the ports of the stationary valve plate. Alignment is achieved through rotation of the rotary sealing element with respect to the stationary valve plate. The fluid passages provide direct paths which minimize fluid turbulence created in the fluid as it passes through the valve.
Apparatus comprising magnetically actuated valves and uses thereof
Edwards, Thayne L.; Harper, Jason C.
2016-07-12
The present invention, in part, relates to an apparatus having a single-use, normally-closed fluidic valve that is initially maintained in the closed position by a valve element bonded to an adhesive coating. The valve is opened using a magnetic force. The valve element includes a magnetic material or metal. In some examples, the valve is opened by bringing a magnet in proximity to the valve element to provide a magnetic force that delaminates the valve element from the adhesive coating. In particular, the apparatus can be useful for on-chip amplification and/or detection of various targets, including biological targets and any amplifiable targets. Such apparatuses and methods are useful for in-field or real-time detection of targets, especially in limited resource settings.
Mixed mode fuel injector with individually moveable needle valve members
Stewart, Chris; Chockley, Scott A.; Ibrahim, Daniel R.; Lawrence, Keith; Tomaseki, Jay; Azam, Junru H.; Tian, Steven Ye; Shafer, Scott F.
2004-08-03
A fuel injector includes a homogenous charge nozzle outlet set and a conventional nozzle outlet set controlled respectively, by first and second needle valve members. One of the needle valve members moves to an open position while the other needle valve member remains stationary for a homogeneous charge injection event. The former needle valve member stays stationary while the other needle valve member moves to an open position for a conventional injection event. One of the needle valve members is at least partially positioned in the other needle valve member. Thus, the injector can perform homogeneous charge injection events, conventional injection events, or even a mixed mode having both types of injection events in a single engine cycle.
Sorbent-Based Atmosphere Revitalization System
NASA Technical Reports Server (NTRS)
Knox, James C (Inventor); Miller, Lee A. (Inventor)
2017-01-01
The present invention is a sorbent-based atmosphere revitalization (SBAR) system using treatment beds each having a bed housing, primary and secondary moisture adsorbent layers, and a primary carbon dioxide adsorbent layer. Each bed includes a redirecting plenum between moisture adsorbent layers, inlet and outlet ports connected to inlet and outlet valves, respectively, and bypass ports connected to the redirecting plenums. The SBAR system also includes at least one bypass valve connected to the bypass ports. An inlet channel connects inlet valves to an atmosphere source. An outlet channel connects the bypass valve and outlet valves to the atmosphere source. A vacuum channel connects inlet valves, the bypass valve and outlet valves to a vacuum source. In use, one bed treats air from the atmosphere source while another bed undergoes regeneration. During regeneration, the inlet, bypass, and outlet valves sequentially open to the vacuum source, removing accumulated moisture and carbon dioxide.
Special functions of valve organs of blood-sucking female mosquitoes
NASA Astrophysics Data System (ADS)
Kim, Boheum; Lee, Sangjoon
2010-11-01
Food-feeding insects usually have valve organs to regulate the sucking flow effectively. Female mosquitoes sucking lots of blood instantaneously have a unique valve system between two pumping organs located in their head. The valve system seems to prevent reverse flow and to grind granule particles such as red blood cells. To understand the functional characteristics of this valve organ in detail, the volumetric flow rate passing through the valves and their interaction with the two-pumps need to be investigated. However, it is very difficult to observe the dynamic behaviors of pumping organs and valve system. In this study, the dynamic motions of valve organs of blood-sucking female mosquitoes were observed under in vivo condition using synchrotron X-ray micro imaging technique. X-ray micro computed tomography was also employed to examine the three-dimensional internal structure of the blood pumping system including valve organs.
SLM Produced Hermetically Sealed Isolation Valve
NASA Technical Reports Server (NTRS)
Richard, James
2014-01-01
Marshall Space Flight Center (MSFC) has developed a valve concept to replace traditional pyrotechnic-driven isolation valves. This paper will describe the valve design and development process. The valve design uses a stem/wedge to support a disk inside the valve. That disk hermetically seals the pressurized fluids. A release mechanism holds the stem/wedge and a large spring in place. When required to open, a solenoid is energized and pulls the release mechanism allowing the spring to pull the stem/wedge away from the disk. Now the disk is unsupported and the pressure ruptures the disk allowing flow to the outlet of the valve. This paper will provide details of this design, describe the development testing, and show the results from the valve level tests performed. Also, a trade study is presented to show the advantages of this design to a conventional pyrotechnic-based valve.
Buch, W S; Pipkin, R D; Hancock, W D; Fogarty, T J
1975-11-01
From March 1971 through April 1975, one hundred twenty patients underwent mitral valve replacement with a Hancock "stabilized glutaraldehyde process" porcine aortic xenograft. A simultaneous canine experimental series was also carried out. In the clinical series, the early mortality was 8.3%. Actuarial analyses of all patients predicts survival at two years of 81.0% and at four years of 70.0%. The predicted survival for patients without coronary disease or prior prosthetic valve replacement is 87.5% at two years and 77.5% at four years. There were four thromboembolic episodes, a rate of 2.4% per patient-year. None were fatal. No valve failure were noted. Histologic examination and shrink temperature analysis of recovered valves show excellent tissue preservation at 40 months. The data indicate that the Hancock valve is durable, enjoys a low incidence of thromboembolism, and may be the valve of choice for mitral valve replacement.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Moore, Wayne R.
A control valve includes a first conduit having a first inlet and a first outlet and defining a first passage; a second conduit having a second inlet and a second outlet and defining a second passage, the second conduit extending into the first passage such that the second inlet is located within the first passage; and a valve plate disposed pivotably within the first passage, the valve plate defining a valve plate surface. Pivoting of the valve plate within the first passage varies flow from the first inlet to the first outlet and the valve plate is pivotal between amore » first position and a second position such that in the first position the valve plate substantially prevents fluid communication between the first passage and the second passage and such that in the second position the valve plate permits fluid communication between the first passage and the second passage.« less
Cells for tissue engineering of cardiac valves.
Jana, Soumen; Tranquillo, Robert T; Lerman, Amir
2016-10-01
Heart valve tissue engineering is a promising alternative to prostheses for the replacement of diseased or damaged heart valves, because tissue-engineered valves have the ability to remodel, regenerate and grow. To engineer heart valves, cells are harvested, seeded onto or into a three-dimensional (3D) matrix platform to generate a tissue-engineered construct in vitro, and then implanted into a patient's body. Successful engineering of heart valves requires a thorough understanding of the different types of cells that can be used to obtain the essential phenotypes that are expressed in native heart valves. This article reviews different cell types that have been used in heart valve engineering, cell sources for harvesting, phenotypic expression in constructs and suitability in heart valve tissue engineering. Natural and synthetic biomaterials that have been applied as scaffold systems or cell-delivery platforms are discussed with each cell type. Copyright © 2015 John Wiley & Sons, Ltd. Copyright © 2015 John Wiley & Sons, Ltd.
Transhepatic implant of a trimmed Melody™ valved stent in tricuspid position in a 1-year-old infant.
Cools, Bjorn; Rega, Filip; Gewillig, Marc
2017-02-15
Percutaneous valved stent implantation is precluded in small infants because large delivery sheaths and large devices. We describe a procedure in a 1-year-old boy in whom a 19 mm Epic™ valve in tricuspid position had become dysfunctional. As the internal diameter of the prosthetic valve was about 16 mm, the only available valve was the Melody™ valved stent. Technical modifications were required to address issues like venous access, the bulky delivery system, and the length of the valved stent. The Melody™ valved stent was surgically trimmed and mounted on a 16 mm Tyshak balloon, access was provided transhepatically through a short 18 Fr sheath. After deployment, the intrahepatic route was sealed with two vascular plugs (8 and 10 mm) in tandem. The procedure was uncomplicated with perfect valve function 18 months after implant. © 2016 Wiley Periodicals, Inc. © 2016 Wiley Periodicals, Inc.
SLM Produced Hermetically Sealed Isolation Valve
NASA Technical Reports Server (NTRS)
Richard, James A.
2014-01-01
Marshall Space Flight Center (MSFC) has developed a valve concept to replace traditional pyrotechnic driven isolation valves. This paper will describe the valve design and development process. The valve design uses a stem/wedge to support a disk inside the valve. That disk hermetically seals the pressurized fluids. A release mechanism holds the stem/wedge and a large spring in place. When required to open, a solenoid is energized and pulls the release mechanism allowing the spring to pull the stem/wedge away from the disk. Now the disk is unsupported and the pressure ruptures the disk allowing flow to the outlet of the valve. This paper will provide details of this design, describe the development testing, and show the results from the valve level tests performed. Also, a trade study is presented to show the advantages of this design to a conventional pyrotechnic based valve.
In Vitro Hydrodynamic Assessment of a New Transcatheter Heart Valve Concept (the TRISKELE).
Rahmani, Benyamin; Tzamtzis, Spyros; Sheridan, Rose; Mullen, Michael J; Yap, John; Seifalian, Alexander M; Burriesci, Gaetano
2017-04-01
This study presents the in vitro hydrodynamic assessment of the TRISKELE, a new system suitable for transcatheter aortic valve implantation (TAVI), aiming to mitigate the procedural challenges experienced with current technologies. The TRISKELE valve comprises three polymeric leaflet and an adaptive sealing cuff, supported by a novel fully retrievable self-expanding nitinol wire frame. Valve prototypes were manufactured in three sizes of 23, 26, and 29 mm by automated dip-coating of a biostable polymer, and tested in a hydrodynamic bench setup in mock aortic roots of 21, 23, 25, and 27 mm annulus, and compared to two reference valves suitable for equivalent implantation ranges: Edwards SAPIEN XT and Medtronic CoreValve. The TRISKELE valves demonstrated a global hydrodynamic performance comparable or superior to the controls with significant reduction in paravalvular leakage. The TRISKELE valve exhibits enhanced anchoring and improved sealing. The valve is currently under preclinical investigation.
Static Flow Characteristics of a Mass Flow Injecting Valve
NASA Technical Reports Server (NTRS)
Mattern, Duane; Paxson, Dan
1995-01-01
A sleeve valve is under development for ground-based forced response testing of air compression systems. This valve will be used to inject air and to impart momentum to the flow inside the first stage of a multi-stage compressor. The valve was designed to deliver a maximum mass flow of 0.22 lbm/s (0.1 kg/s) with a maximum valve throat area of 0.12 sq. in (80 sq. mm), a 100 psid (689 KPA) pressure difference across the valve and a 68 F, (20 C) air supply. It was assumed that the valve mass flow rate would be proportional to the valve orifice area. A static flow calibration revealed a nonlinear valve orifice area to mass flow relationship which limits the maximum flow rate that the valve can deliver. This nonlinearity was found to be caused by multiple choking points in the flow path. A simple model was used to explain this nonlinearity and the model was compared to the static flow calibration data. Only steady flow data is presented here. In this report, the static flow characteristics of a proportionally controlled sleeve valve are modelled and validated against experimental data.
Odelin, Gaëlle; Faure, Emilie; Coulpier, Fanny; Di Bonito, Maria; Bajolle, Fanny; Studer, Michèle; Avierinos, Jean-François; Charnay, Patrick; Topilko, Piotr; Zaffran, Stéphane
2018-01-03
Although cardiac neural crest cells are required at early stages of arterial valve development, their contribution during valvular leaflet maturation remains poorly understood. Here, we show in mouse that neural crest cells from pre-otic and post-otic regions make distinct contributions to the arterial valve leaflets. Genetic fate-mapping analysis of Krox20-expressing neural crest cells shows a large contribution to the borders and the interleaflet triangles of the arterial valves. Loss of Krox20 function results in hyperplastic aortic valve and partially penetrant bicuspid aortic valve formation. Similar defects are observed in neural crest Krox20 -deficient embryos. Genetic lineage tracing in Krox20 -/- mutant mice shows that endothelial-derived cells are normal, whereas neural crest-derived cells are abnormally increased in number and misplaced in the valve leaflets. In contrast, genetic ablation of Krox20 -expressing cells is not sufficient to cause an aortic valve defect, suggesting that adjacent cells can compensate this depletion. Our findings demonstrate a crucial role for Krox20 in arterial valve development and reveal that an excess of neural crest cells may be associated with bicuspid aortic valve. © 2018. Published by The Company of Biologists Ltd.
Study of the Pressure and Velocity Across the Aortic Valve
NASA Astrophysics Data System (ADS)
Kyung, Seo Young; Chung, Erica Soyun; Lee, Joo Hee; Kyung, Hayoung; Choi, Si Young
Biomechanics of the heart, requiring an extensive understanding of the complexity of the heart, have become the interests of many biomedical engineers in cardiology today. In order to study aortic valve disease, engineers have focused on the data obtained through bio-fluid flow analysis. To further this study, physical and computational analysis on the biomechanical determinants of blood flow in the stenosed aortic valve have been examined. These observations, along with the principles of cardiovascular physiology, confirm that when blood flows through the valve opening, pressure gradient across the valve is produced as a result of stenosis of the aortic valve. The aortic valve gradient is used to interpret the increase and decrease on each side of the defective valve. To compute different pressure gradients across the aortic valve, this paper analyzes Aortic Valve Areas (AVA) using simulations based on the continuity equation and Gorlin equation. The data obtained from such analysis consist of patients in the AS category that display mild Aortic Valve Velocity (AVV) and pressure gradient. Such correlation results in the construction of a dependent relationship between severe AS causing LV systolic dysfunction and the transaortic velocity.
Recently patented transcatheter aortic valves in clinical trials.
Neragi-Miandoab, Siyamek; Skripochnik, Edvard; Salemi, Arash; Girardi, Leonard
2013-12-01
The most widely used heart valve worldwide is the Edwards Sapien, which currently has 60% of the worldwide transcatheter aortic valve implantation (TAVI) market. The CoreValve is next in line in popularity, encompassing 35% of the worldwide TAVI market. Although these two valves dominate the TAVI market, a number of newer transcatheter valves have been introduced and others are in early clinical evaluation. The new valves are designed to reduce catheter delivery diameter, improve ease of positioning and sealing, and facilitate repositioning or removal. The most recent transcatheter valves for transapical use include Acurate TA (Symetis), Engager (Medtronic), and JenaValve the Portico (St Jude), Sadra Lotus Medical (Boston Scientific), and the Direct Flow Medical. These new inventions may introduce more effective treatment options for high-risk patients with severe aortic stenosis. Improvements in transcatheter valves and the developing variability among them may allow for more tailored approaches with respect to patient's anatomy, while giving operators the opportunity to choose devices they feel more comfortable with. Moreover, introducing new devices to the market will create a competitive environment among producers that will reduce high prices and expand availability. The present review article includes a discussion of recent patents related to Transcatheter Aortic Valves.
Thermostatic Valves Containing Silicone-Oil Actuators
NASA Technical Reports Server (NTRS)
Bhandari, Pradeep; Birur, Gajanana C.; Bame, David P.; Karlmann, Paul B.; Prina, Mauro; Young, William; Fisher, Richard
2009-01-01
Flow-splitting and flow-mixing thermally actuated spool valves have been developed for controlling flows of a heat-transfer fluid in a temperature-regulation system aboard the Mars Science Laboratory (MSL) rover. Valves like these could also be useful in terrestrial temperature-regulation systems, including automobile air-conditioning systems and general refrigeration systems. These valves are required to provide smoother actuation over a wider temperature range than the flow-splitting, thermally actuated spool valves used in the Mars Explorer Rover (MER). Also, whereas the MER valves are unstable (tending to oscillate) in certain transition temperature ranges, these valves are required not to oscillate. The MER valves are actuated by thermal expansion of a wax against spring-loaded piston rods (as in common automotive thermostats). The MSL valves contain similar actuators that utilize thermal expansion of a silicone oil, because silicone-oil actuators were found to afford greater and more nearly linear displacements, needed for smoother actuation, over the required wider temperature range. The MSL valves also feature improved spool designs that reflect greater understanding of fluid dynamics, consideration of pressure drops in valves, and a requirement for balancing of pressures in different flow branches.
Development of Overflow-Prevention Valve with Trigger Mechanism.
NASA Astrophysics Data System (ADS)
Ishino, Yuji; Mizuno, Takeshi; Takasaki, Masaya
2016-09-01
A new overflow-prevention valve for combustible fluid is developed which uses a trigger mechanism. Loading arms for combustible fluid are used for transferring oil from a tanker to tanks and vice versa. The loading arm has a valve for preventing overflow. Overflow- prevention valves cannot use any electric component to avoid combustion. Therefore, the valve must be constructed only by mechanical parts. The conventional overflow-prevention valve uses fluid and pneumatic forces. It consists of a sensor probe, a cylinder, a main valve for shutting off the fluid and a locking mechanism for holding an open state of the main valve. The proposed overflow-prevention valve uses the pressure due to the height difference between the fluid level of the tank and the sensor probe. However, the force of the cylinder produced by the pressure is too small to release the locking mechanism. Therefore, a trigger mechanism is introduced between the cylinder and the locking mechanism. The trigger mechanism produces sufficient force to release the locking mechanism and close the main valve when the height of fluid exceeds a threshold value. A trigger mechanism is designed and fabricated. The operation necessary for closing the main valve is conformed experimentally.
Blot, William J; Ibrahim, Michel A; Ivey, Tom D; Acheson, Donald E; Brookmeyer, Ron; Weyman, Arthur; Defauw, Joseph; Smith, J Kermit; Harrison, Donald
2005-05-31
The first Björk-Shiley convexoconcave (BSCC) prosthetic heart valves were implanted in 1978. The 25th anniversary provided a stimulus to summarize the research data relevant to BSCC valve fracture, patient management, and current clinical options. Published and unpublished data on the risks of BSCC valve fracture and replacement were compiled, and strategies for identifying candidates for prophylactic valve reoperation were summarized. By December 2003, outlet strut fractures (OSFs), often with fatal outcomes, had been reported in 633 BSCC valves (0.7% of 86,000 valves implanted). Fractures still continue to occur, but average rates of OSFs in 60 degrees valves are now <0.1% per year. OSF risk varies markedly by valve characteristics, especially valve angle and size, with weaker effects associated with other manufacturing variables. OSF risks are mildly lower among women than men but decline sharply with advancing age. The risks of valve replacement typically greatly exceed those of OSF. By comparing individualized estimated risks of OSF versus valve replacement, guidelines have been developed to identify the small percentage of BSCC patients (mostly younger men) who would be expected to have a gain in life expectancy should reoperative surgery be performed. Twenty-five years after the initial BSCC valve implants, fractures continue to occur. Continued monitoring of BSCC patients is needed to track and quantify risks and enable periodic updating of guidelines for patients and their physicians.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Sochman, Jan; Peregrin, Jan H.; Rocek, Miloslav
Purpose. To evaluate the feasibility of one-step implantation of a new type of stent-based mechanical aortic disc valve prosthesis (MADVP) above and across the native aortic valve and its short-term function in swine with both functional and dysfunctional native valves. Methods. The MADVP consisted of a folding disc valve made of silicone elastomer attached to either a nitinol Z-stent (Z model) or a nitinol cross-braided stent (SX model). Implantation of 10 MADVPs (6 Z and 4 SX models) was attempted in 10 swine: 4 (2 Z and 2 SX models) with a functional native valve and 6 (4 Z andmore » 2 SX models) with aortic regurgitation induced either by intentional valve injury or by MADVP placement across the native valve. MADVP function was observed for up to 3 hr after implantation. Results. MADVP implantation was successful in 9 swine. One animal died of induced massive regurgitation prior to implantation. Four MADVPs implanted above functioning native valves exhibited good function. In 5 swine with regurgitation, MADVP implantation corrected the induced native valve dysfunction and the device's continuous good function was observed in 4 animals. One MADVP (SX model) placed across native valve gradually migrated into the left ventricle. Conclusion. The tested MADVP can be implanted above and across the native valve in a one-step procedure and can replace the function of the regurgitating native valve. Further technical development and testing are warranted, preferably with a manufactured MADVP.« less
Valve stem and packing assembly
Wordin, J.J.
1991-09-03
A valve stem and packing assembly is provided in which a rotatable valve stem includes a first tractrix surface for sliding contact with a stem packing and also includes a second tractrix surface for sliding contact with a bonnet. Force is applied by means of a spring, gland flange, and gland on the stem packing so the stem packing seals to the valve stem and bonnet. This configuration serves to create and maintain a reliable seal between the stem packing and the valve stem. The bonnet includes a second complementary tractrix surface for contacting the second sliding tractrix surface, the combination serving as a journal bearing for the entire valve stem and packing assembly. The journal bearing so configured is known as a Schiele's pivot. The Schiele's pivot also serves to maintain proper alignment of the valve stem with respect to the bonnet. Vertical wear between the surfaces of the Schiele's pivot is uniform at all points of contact between the second sliding tractrix surface and the second complementary tractrix surface of a bonnet. The valve stem is connected to a valve plug by means of a slip joint. The valve is opened and closed by rotating the valve stem. The slip joint compensates for wear on the Schiele's pivot and on the valve plug. A ledge is provided on the valve bonnet for the retaining nut to bear against. The ledge prevents over tightening of the retaining nut and the resulting excessive friction between stem and stem packing. 2 figures.
Valve stem and packing assembly
Wordin, John J.
1991-01-01
A valve stem and packing assembly is provided in which a rotatable valve stem includes a first tractrix surface for sliding contact with a stem packing and also includes a second tractrix surface for sliding contact with a bonnet. Force is applied by means of a spring, gland flange, and gland on the stem packing so the stem packing seals to the valve stem and bonnet. This configuration serves to create and maintain a reliable seal between the stem packing and the valve stem. The bonnet includes a second complementary tractrix surface for contacting the second sliding tractrix surface, the combination serving as a journal bearing for the entire valve stem and packing assembly. The journal bearing so configured is known as a Schiele's pivot. The Schiele's pivot also serves to maintain proper alignment of the valve stem with respect to the bonnet. Vertical wear between the surfaces of the Schiele's pivot is uniform at all points of contact between the second sliding tractrix surface and the second complementary tractrix surface of a bonnet. The valve stem is connected to a valve plug by means of a slip joint. The valve is opened and closed by rotating the valve stem. The slip joint compensates for wear on the Schiele's pivot and on the valve plug. A ledge is provided on the valve bonnet for the retaining nut to bear against. The ledge prevents overtightening of the retaining nut and the resulting excessive friction between stem and stem packing.
Fixation and mounting of porcine aortic valves for use in mock circuits.
Schlöglhofer, Thomas; Aigner, Philipp; Stoiber, Martin; Schima, Heinrich
2013-10-01
Investigations of the circulatory system in vitro use mock circuits that require valves to mimic the cardiac situation. Whereas mechanical valves increase water hammer effects due to inherent stiffness and do not allow the use of pressure lines or catheters, bioprosthetic valves are expensive and of limited durability in test fluids. Therefore, we developed a cheap, fast, alternative method to mount valves obtained from the slaughterhouse in mock circuits. Porcine aortic roots were obtained from the abattoir and used either in native condition or after fixation. Fixation was performed at a constant retrograde pressure to ensure closed valve position. Fixation time was 4 h in a 0.5%-glutaraldehyde phosphate buffer. The fixed valves were molded into a modular mock circulation connector using a fast curing silicone. Valve functionality was evaluated in a pulsatile setting (cardiac output = 4.7 l/min, heart rate = 80 beats/min) and compared before and after fixation. Leaflet motion was recorded with a high-speed camera and valve insufficiency was quantified by leakage flow under steady pressure application (80 mmHg). Under physiological conditions the aortic valves showed almost equal leaflet motion in native and fixed conditions. However, the leaflets of the native valves showed lower stiffness and more fluttering during systole than the fixed specimens. Under retrograde pressure, fresh and fixed valves showed small leakage flows of <30 ml/min. The new mounting and fixation procedure is a fast method to fabricate low cost biologic valves for the use in mock circuits.
García-Bengochea, José; Sierra, Juan; González-Juanatey, José R; Rubio, José; Vega, Marino; Fernández, Angel L; Sánchez, Daniel
2006-05-01
The study aim was to evaluate the durability of the new Mitroflow 12A pericardial bioprosthesis and to assess ventricular mass regression after aortic valve replacement (AVR) in patients with small valves. A total of 326 Mitroflow 12A pericardial bioprostheses was implanted without any other associated valve procedure. The mean patient age was 73.5 +/- 6.1 years; 252 patients (78.0%) were in NYHA class III/IV. Small valves (19 and 21 mm) were implanted in 212 patients (65.6%). The total follow up period was 837.1 patient-years (pt-yr). A subset of 61 patients with preoperative stenosis was selected and submitted to conventional echo-Doppler assessment at a mean period of 11.1 months after surgery. Hospital mortality was 8.6%. At eight years of follow up, survival was 57.1%. Freedom from structural valve deterioration (SVD) was 86.5% per pt-yr. Mean gradients were significantly reduced postoperatively for each valve size (to 18 +/- 8 mmHg for 19-mm valves and 12 +/- 4 mmHg for 25-mm valves). The effective orifice area (EOA) was also increased significantly for all valve sizes (to 1.1 +/- 0.1 cm2 for 19-mm valves and 1.8 +/- 0.2 cm2 for 25-mm valves). The left ventricular mass index (LVMI) decreased significantly, from 177 +/- 29 to 136 +/- 22 g/m2 for the 19-mm valve, and from 200 +/- 42 to 132 +/- 22 g/m2 for the 25-mm valve. The EOA index (EOAI) showed mismatch for the 19- and 21-mm valves (0.74 and 0.82 cm2/m2, respectively). The diagrammatic calculation between LVMI and relative wall thickness after surgery showed that 29.7% of patients achieved a normal pattern of remodeling (including 19- and 21-mm valves), despite various degrees of mismatch. The new Mitroflow 12A pericardial bioprosthesis showed an absence of mechanical failure after an eight-year follow up, with a satisfactory rate of SVD. Significant reductions in LVMI and improved ventricular geometry were observed, despite the small valve sizes implanted.
Structural deterioration of the Freestyle aortic valve: mode of presentation and mechanisms.
Mohammadi, Siamak; Baillot, Richard; Voisine, Pierre; Mathieu, Patrick; Dagenais, François
2006-08-01
Structural valve deterioration is the major cause of bioprosthetic valve failure. Because of the unique design features and anti-calcification treatment of the Freestyle (Medtronic Inc, Minneapolis, Minn) stentless bioprosthesis, development of structural valve deterioration may differ in comparison with other bioprosthetic valves. This study evaluates the mechanisms and clinical presentation of structural valve deterioration in the Freestyle stentless bioprosthesis. Between January 1993 and August 2005, 608 patients underwent aortic valve replacement with a Freestyle stentless bioprosthesis. The implantation technique was subcoronary in 475 patients and a root replacement in 133 patients. Mean overall follow-up was 5.6 +/- 3.4 years. Follow-up was complete in all patients. Clinical and echocardiographic follow-ups were conducted prospectively. Freedom from structural valve deterioration was 95.8% at 10 years. Twelve patients showed evidence of structural valve deterioration and underwent reoperation for aortic regurgitation (n = 10) or aortic stenosis (n = 2). The mean age of patients with structural valve deterioration was significantly lower than patients without structural valve deterioration (62.6 +/- 8.2 years vs 68.6 +/- 8.3 years, P = .02). The median time between implantation and explantation was 8.7 years (range: 1.9-13.3 years). Eleven structural valve deteriorations occurred after subcoronary implantation, and 1 structural valve deterioration occurred after root implantation (P = .4). The mechanisms of structural valve deterioration were leaflet tears in 10 patients (6 in the left coronary cusp and 4 in the right coronary cusp), severe valve calcification in 1 patient, and cusp fibrosis in 1 patient. The interval between onset of symptoms and reoperation was acute or subacute in 10 patients. At 10 years, the Freestyle stentless bioprosthesis shows excellent freedom from structural valve deterioration. Structural valve deterioration in the Freestyle stentless bioprosthesis relates to leaflet tear with minimal calcification in the majority of cases. Because of the fast onset of symptoms with leaflet tear, patients with a Freestyle stentless bioprosthesis should be informed of the preferential mode of failure and time-frame of symptoms.
RELAP5 posttest calculation of IAEA-SPE-4
DOE Office of Scientific and Technical Information (OSTI.GOV)
Petelin, S.; Mavko, B.; Parzer, I.
The International Atomic Energy Agency`s Fourth Standard Problem Exercise (IAEA-SPE-4) was performed at the PMK-2 facility. The PMK-2 facility is designed to study processes following small- and medium-size breaks in the primary system and natural circulation in VVER-440 plants. The IAEA-SPE-4 experiment represents a cold-leg side small break, similar to the IAEA-SPE-2, with the exception of the high-pressure safety injection being unavailable, and the secondary side bleed and feed initiation. The break valve was located at the dead end of a vertical downcomer, which in fact simulates a break in the reactor vessel itself, and should be unlikely to happenmore » in a real nuclear power plant (NPP). Three different RELAP5 code versions were used for the transient simulation in order to assess the calculations with test results.« less
Quantitative Global Heat Transfer in a Mach-6 Quiet Tunnel
NASA Technical Reports Server (NTRS)
Sullivan, John P.; Schneider, Steven P.; Liu, Tianshu; Rubal, Justin; Ward, Chris; Dussling, Joseph; Rice, Cody; Foley, Ryan; Cai, Zeimin; Wang, Bo;
2012-01-01
This project developed quantitative methods for obtaining heat transfer from temperature sensitive paint (TSP) measurements in the Mach-6 quiet tunnel at Purdue, which is a Ludwieg tube with a downstream valve, moderately-short flow duration and low levels of heat transfer. Previous difficulties with inferring heat transfer from TSP in the Mach-6 quiet tunnel were traced to (1) the large transient heat transfer that occurs during the unusually long tunnel startup and shutdown, (2) the non-uniform thickness of the insulating coating, (3) inconsistencies and imperfections in the painting process and (4) the low levels of heat transfer observed on slender models at typical stagnation temperatures near 430K. Repeated measurements were conducted on 7 degree-half-angle sharp circular cones at zero angle of attack in order to evaluate the techniques, isolate the problems and identify solutions. An attempt at developing a two-color TSP method is also summarized.
Balloon Ascent: 3-D Simulation Tool for the Ascent and Float of High-Altitude Balloons
NASA Technical Reports Server (NTRS)
Farley, Rodger E.
2005-01-01
The BalloonAscent balloon flight simulation code represents a from-scratch development using Visual Basic 5 as the software platform. The simulation code is a transient analysis of balloon flight, predicting the skin and gas temperatures along with the 3-D position and velocity in a time and spatially varying environment. There are manual and automated controls for gas valving and the dropping of ballast. Also, there are many handy calculators, such as appropriate free lift, and steady-state thermal solutions with temperature gradients. The strength of this simulation model over others in the past is that the infrared environment is deterministic rather than guessed at. The ground temperature is specified along with the emissivity, which creates a ground level IR environment that is then partially absorbed as it travels upward through the atmosphere to the altitude of the balloon.
Control means for a gas turbine engine
NASA Technical Reports Server (NTRS)
Beitler, R. S.; Sellers, F. J.; Bennett, G. W. (Inventor)
1982-01-01
A means is provided for developing a signal representative of the actual compressor casing temperature, a second signal representative of compressor inlet gas temperature, and a third signal representative of compressor speed. Another means is provided for receiving the gas temperature and compressor speed signals and developing a schedule output signal which is a representative of a reference casing temperature at which a predetermined compressor blade stabilized clearance is provided. A means is also provided for comparing the actual compressor casing temperature signal and the reference casing temperature signal and developing a clearance control system representative of the difference. The clearance control signal is coupled to a control valve which controls a flow of air to the compressor casing to control the clearance between the compressor blades and the compressor casing. The clearance control signal can be modified to accommodate transient characteristics. Other embodiments are disclosed.
Common carotid artery dissection in the course of acute aortic dissection De Bakey type I.
Zwierzyńska, Elzbieta; Bec, Leszek; Sklinda, Katarzyna; Walecki, Jerzy; Garlicki, Mirosław; Pniewski, Jarosław
2007-01-01
Extracranial artery dissection is a rare finding. It may be the result of acute aortic dissection. We present the case of a 50-year-old patient admitted to Department of Neurology CSK MSWiA with signs of transient ischaemic attack. Both computed tomography (CT) and magnetic resonance imaging revealed no ischaemic changes within the central nervous system. The ultrasonography revealed dissection of the right common carotid artery, brachiocephalic trunk and aorta. Acute aortic dissection De Bakey type I was confirmed on CT scans of the chest and abdomen. Emergency cardiosurgery was performed because of acute aortic valve insufficiency, threatening infarction in the territory of the right coronary artery and heart tamponade. During the surgery the false aortic lumen was closed and the ascending graft was implanted. Short- and long-term outcome was good. Neither neurological symptoms nor cardiovascular insufficiency was observed within 9 months of follow-up.
Thermodynamic performance testing of the orbiter flash evaporator system
NASA Technical Reports Server (NTRS)
Jaax, J. R.; Melgares, M. A.; Frahm, J. P.
1980-01-01
System level testing of the space shuttle orbiter's development flash evaporator system (FES) was performed in a thermal vacuum chamber capable of simulating ambient ascent, orbital, and entry temperature and pressure profiles. The test article included the evaporator assembly, high load and topping exhaust duct and nozzle assemblies, and feedwater supply assembly. Steady state and transient heat load, water pressure/temperature and ambient pressure/temperature profiles were imposed by especially designed supporting test hardware. Testing in 1978 verified evaporator and duct heater thermal design, determined FES performance boundaries, and assessed topping evaporator plume characteristics. Testing in 1979 combined the FES with the other systems in the orbiter active thermal control subsystem (ATCS). The FES met or exceeded all nominal and contingency performance requirements during operation with the integrated ATCS. During both tests stability problems were encountered during steady state operations which resulted in subsequent design changes to the water spray nozzle and valve plate assemblies.
Durability of hand-sewn valves in the right ventricular outlet.
Nunn, Graham R; Bennetts, Jayme; Onikul, Ella
2008-08-01
The objective was to compare the medium- and long-term outcomes for pericardial monocusp valves, polytetrafluoroethylene (Gore-Tex, WL Gore and Associates Inc, Flagstaff, Ariz) 0.1-mm monocusp valves, and bileaflet 0.l-mm polytetrafluoroethylene valves and their efficiency in the right ventricular outlet. We reviewed all hand-sewn right ventricular outlet valves created by the author (Graham R. Nunn) in the setting of repaired tetralogy of Fallot or equivalent right ventricular outlet pathology when the native pulmonary valve could not be preserved. The valves were assessed by serial transthoracic echocardiography and more recently by magnetic resonance imaging angiography for late valve function. The bileaflet polytetrafluoroethylene valves were constructed in a standardized fashion from a semicircle of 0.1-mm polytetrafluoroethylene (the radius of which equaled the length of the outflow tract incision) that gave a lengthened free edge to the leaflets, central fixation of the free edge posteriorly just proximal to the branch pulmonary arteries, and generous augmentation of the outflow tract with polytetrafluoroethylene patch-plasty. The bileaflet configuration shortens the closing time against the posterior wall, and the leaflets are forced to maintain their configuration without prolapse into the right ventricular outlet. The valve can be generously oversized in young children to try to avoid the need for replacement. A total of 54 patients met the selection criteria--22 patients received fresh autologous pericardial monocusps, 7 patients received polytetrafluoroethylene (0.1-mm) monocusps, and 25 patients received bileaflet polytetrafluoroethylene (0.1-mm) outlet valves. The pericardial valves have the longest follow-up, and all valves developed free pulmonary incompetence. Polytetrafluoroethylene monocusps had reliable competence early after surgery but progressed to pulmonary incompetence. The bileaflet polytetrafluoroethylene (0.1-mm) valves have remained competent with regurgitant fractions of only 5% to 30% (magnetic resonance imaging angiography), and this has remained stable with time. The maximum follow-up for these valves is 5 years. No stenosis or peripheral emboli have been recognized, and no valves have been replaced to date. Hand-sewn bileaflet polytetrafluoroethylene valves in the right ventricular outlet can reliably provide competence and maintain function in the medium term. Their shape and size allow placement in young children with a reasonable expectation that they will remain competent with growth of the native annulus and not require replacement. Their durability is superior to the pericardial and polytetrafluoroethylene monocusp valves in this series.
Li, Mingwen; Xiao, Yingbin; Chen, Daozhong; Liu, Liming; Ma, Liming; Wang, Pingfan; Jia, Kui; Yang, Kai; Chen, Lin
2016-05-18
Star GK valves were widely used in China, and we studied the clinical follow-up results of patients with Star GK valve implants for more than one year. Clinical data were collected from those patients who had Star GK valve implants for over one year. Patients were divided into three groups: (1) AVR group: received aortic valve replacement surgery. Based on the valve model this group was further sub-divided into two groups: 21A group, and 23A group; (2) MVR group: received mitral valve replacement surgery. Based on the valve model this group was further sub-divided into three groups: 25M group, 27M group, and 29M group; (3) DVR group: received combined replacement surgeries including AVR + MVR. According to postoperative follow-up time these patients were divided into two groups: 1-year group and 3-year group. Follow-up data were collected by telephone, outpatient visits, or correspondence. Clinical data were aggregated by professional data scientists to conduct independent analyses. 959 patients were included in the study following Star GK valve implant. Follow-up after 1 year found that thrombosis occurred in 4 cases, hemorrhage in 15 cases, left heart failure in 13 cases, paravalvular leakage in 5 cases, and death due to cardiac causes in 2 cases. The long-term efficacy of Star GK valve implants was satisfactory with low incidence of valve-related complications, and following Star GK valve implant, valve and blood were highly compatible and blood component damage was minor. Very low incidence rate of thrombosis was observed following Star GK valve implant, however, attention should be paid to adjust the anticoagulation intensity.
Sharma, Ashutosh; Cote, Anita T; Hosking, Martin C K; Harris, Kevin C
2017-07-24
The aim of this study was to systematically evaluate the incidence of infective endocarditis (IE) in right ventricle-to-pulmonary artery conduits and valves, comparing bovine jugular vein (BJV) valves with all others. Recent evidence suggests that the incidence of IE is higher in patients with congenital heart disease who have undergone implantation of BJV valves in the pulmonary position compared with other valves. Systematic searches of published research were conducted using electronic databases (MEDLINE, Embase, and CINAHL) and citations cross-referenced current to April 2016. Included studies met the following criteria: patients had undergone right ventricle-to-pulmonary artery conduit or percutaneous pulmonary valve implantation, and investigators reported on the type of conduit or valve implanted, method of intervention (surgery or catheter based), IE incidence, and follow-up time. Fifty studies (Levels of Evidence: 2 to 4) were identified involving 7,063 patients. The median cumulative incidence of IE was higher for BJV compared with other valves (5.4% vs. 1.2%; p < 0.0001) during a median follow-up period of 24.0 and 35.5 months, respectively (p = 0.03). For patients with BJV valves, the incidence of IE was not different between surgical and catheter-based valve implantation (p = 0.83). There was a higher incidence of endocarditis with BJV valves than other types of right ventricle-to-pulmonary artery conduits. There was no difference in the incidence of endocarditis between catheter-based bovine valves and surgically implanted bovine valves, suggesting that the substrate for future infection is related to the tissue rather than the method of implantation. Copyright © 2017 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Modification and performance evaluation of a mono-valve engine
NASA Astrophysics Data System (ADS)
Behrens, Justin W.
A four-stroke engine utilizing one tappet valve for both the intake and exhaust gas exchange processes has been built and evaluated. The engine operates under its own power, but has a reduced power capacity than the conventional 2-valve engine. The reduction in power is traced to higher than expected amounts of exhaust gases flowing back into the intake system. Design changes to the cylinder head will fix the back flow problems, but the future capacity of mono-valve engine technology cannot be estimated. The back flow of exhaust gases increases the exhaust gas recirculation (EGR) rate and deteriorates combustion. Intake pressure data shows the mono-valve engine requires an advanced intake valve closing (IVC) time to prevent back flow of charge air. A single actuation camshaft with advanced IVC was tested in the mono-valve engine, and was found to improve exhaust scavenging at TDC and nearly eliminated all charge air back flow at IVC. The optimum IVC timing is shown to be approximately 30 crank angle degrees after BDC. The mono-valve cylinder head utilizes a rotary valve positioned above the tappet valve. The open spaces inside the rotary valveand between the rotary valve and tappet valve represent a common volume that needs to be reduced in order to reduce the base EGR rate. Multiple rotary valve configurations were tested, and the size of the common volume was found to have no effect on back flow but a direct effect on the EGR rate and engine performance. The position of the rotary valve with respect to crank angle has a direct effect on the scavenging process. Optimum scavenging occurs when the intake port is opened just after TDC.
Cost-effectiveness of homograft heart valve replacement surgery: an introductory study.
Yaghoubi, Mohsen; Aghayan, Hamid Reza; Arjmand, Babak; Emami-Razavi, Seyed Hassan
2011-05-01
The clinical effectiveness of heart valve replacement surgery has been well documented. Mechanical and homograft valves are used routinely for replacement of damaged heart valves. Homograft valves are produced in our country but we import the mechanical valves. To our knowledge the cost-effectiveness of homograft valve has not been assessed. The objective of the present study was to compare the cost-effectiveness of homograft valve replacement with mechanical valve replacement surgery. Our samples were selected from 200 patients that underwent homograft and mechanical heart valve replacement surgery in Imam-Khomeini hospital (2000-2005). In each group we enrolled 30 patients. Quality of life was measured using the SF-36 questionnaire and utility was measured in quality-adjusted life years (QALYs). For each group we calculated the price of heart valve and hospitalization charges. Finally the cost-effectiveness of each treatment modalities were summarized as costs per QALYs gained. Forty male and twenty female participated in the study. The mean score of quality of life was 66.06 (SD = 9.22) in homograft group and 57.85 (SD = 11.30) in mechanical group (P < 0.05). The mean QALYs gained in homograft group was 0.67 more than mechanical group. The incremental cost-effectiveness ratio (ICER) revealed a cost savings of 1,067 US$ for each QALY gained in homograft group. Despite limitation of this introductory study, we concluded that homograft valve replacement was more effective and less expensive than mechanical valve. These findings can encourage healthcare managers and policy makers to support the production of homograft valves and allocate more recourse for developing such activities.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Bond, N.D.
1982-05-01
This report covers the development of the various letdown valves used for the two-stage high pressure and temperature coal slurry letdown system as used at the H-Coal Pilot Plant. The period covered in this report was from the prestart-up oil circulation through Coal Runs No. 1 - No. 7. The valves covered are the Willis, which was used exclusively from Coal Runs No. 1 - No. 5, the Cameron and the Kieley and Mueller. The LV-202B Kieley and Mueller and LV-204B Cameron valves again showed little valve wear during short Coal Run No. 7, which demonstrates that the full potentialmore » of these valve designs has not been achieved yet. The problem with the Kieley and Mueller plug freezing will be looked at further, with addition of grease ports and a possible new designed plug shaft and stem guide being made for the valve. The Willis valves developed the same body leaks around the bonnet areas that occurred during Coal Run No. 6. This will be looked at before Coal Run No. 8, but no further trim development is planned. To summarize the progress of the LV-202 and LV-204 valves, the Willis was developed to last about 100 hours, which is the expected life for this valve design in our coal liquefaction process; whereas, the Cameron and Kieley and Mueller valves have lasted for days with good results. The Cameron and Kieley and Mueller valves still have not reached their full potential in plant operation, and, along with the new Masoneilan Sasol, Masoneilan Prototype, Hammel Dahl and Paul valves, future progress in Coal Run No. 8 for the high pressure and temperature letdown valves is anticipated.« less
Oborny, Michael C.; Paul, Phillip H.; Hencken, Kenneth R.; Frye-Mason, Gregory C.; Manginell, Ronald P.
2001-01-01
A valve for controlling fluid flows. This valve, which includes both an actuation device and a valve body provides: the ability to incorporate both the actuation device and valve into a unitary structure that can be placed onto a microchip, the ability to generate higher actuation pressures and thus control higher fluid pressures than conventional microvalves, and a device that draws only microwatts of power. An electrokinetic pump that converts electric potential to hydraulic force is used to operate, or actuate, the valve.
[Tricuspid valve regurgitation : Indications and operative techniques].
Lange, R; Piazza, N; Günther, T
2017-11-01
Functional tricuspid valve (TV) regurgitation secondary to left heart disease (e.g. mitral insufficiency and stenosis) is observed in 75% of the patients with TV regurgitation and is thus the most common etiology; therefore, the majority of patients who require TV surgery, undergo concomitant mitral and/or aortic valve surgery. Uncorrected moderate and severe TV regurgitation may persist or even worsen after mitral valve surgery, leading to progressive heart failure and death. Patients with moderate to severe TV regurgitation show a 3-year survival rate of 40%. Surgery is indicated in patients with severe TV regurgitation undergoing left-sided valve surgery and in patients with severe isolated primary regurgitation without severe right ventricular (RV) dysfunction. For patients requiring mitral valve surgery, tricuspid valve annuloplasty should be considered even in the absence of significant regurgitation, when severe annular dilatation (≥40 mm or >21 mm/m 2 ) is present. Functional TV regurgitation is primarily treated with valve reconstruction which carries a lower perioperative risk than valve replacement. Valve replacement is rarely required. Tricuspid valve repair with ring annuloplasty is associated with better survival and a lower reoperation rate than suture annuloplasty. Long-term results are not available. The severity of the heart insufficiency and comorbidities (e.g. renal failure and liver dysfunction) are the essential determinants of operative mortality and long-term survival. Tricuspid valve reoperations are rarely necessary and associated with a considerable mortality.
Linke, Axel; Walther, Thomas; Schuler, Gerhard
2010-03-01
Treatment of aortic stenosis remains challenging in older individuals, as their perioperative mortality for open heart surgery is increased due to comorbidities. Transcatheter aortic valve implantation using the CoreValve ReValving System (Medtronic, Minneapolis, USA) and the Edwards SAPIEN transcatheter heart valve (THV; Edwards Lifescience, Irvine, California, USA) represents an alternative to conventional valve replacement in elderly patients that have a high risk for conventional surgery. This article summarizes the evidence-base from recent clinical trials. The early results of these landmark studies suggest that transcatheter aortic valve implantation with either one of the prosthesis is feasible, safe, improves hemodynamics and, therefore, might be an alternative to conventional aortic valve replacement in very high-risk patients. However, all of the available transcatheter heart valves have certain disadvantages, limiting their use in daily clinical practice. The process of decision making, which valve to use and which access route to choose is illustrated in this article through clinical case scenarios. Additionally, the lessons learned thus far from the European perspective and the potential impact on the future use in the US are discussed. Despite of the progress in this field, we are still lacking an optimal transcatheter heart valve. Once it is available, we can take the plunge to compare transcatheter valve implantation with convention surgery in severe aortic stenosis!
Park, Byungjoon; Sung, Kiick; Park, Pyo Won
2018-01-25
This study aimed to evaluate the safety and feasibility of transaortic mitral pannus removal (TMPR).Methods and Results:Between 2004 and 2016, 34 patients (median age, 57 years; 30 women) with rheumatic disease underwent pannus removal on the ventricular side of a mechanical mitral valve through the aortic valve during reoperation. The median time interval from the previous surgery was 14 years. TMPR was performed after removal of the mechanical aortic valve (n=21) or diseased native aortic valve (n=11). TMPR was performed in 2 patients through a normal aortic valve. The mitral transprosthetic mean pressure gradient (TMPG) was ≥5 mmHg in 11 patients, including 3 with prosthetic valve malfunction. Prophylactic TMPR was performed in 23 patients. There were no early deaths. Concomitant operations included 22 tricuspid valve surgeries (13 replacements, 15 repairs) and 32 aortic valve replacements (24 repeats, 8 primary). The mean gradient in patients who had mitral TMPG ≥5 mmHg was significantly decreased from 6.46±1.1 to 4.37±1.17 mmHg at discharge (P<0.001). No mechanical valve malfunction was apparent on last echocardiography. TMPR is a safe and effective procedure for patients with malfunction or stenosis of a mechanical mitral valve and may be considered an alternative approach in patients with pannus overgrowth in such valves.
Hybrid textile heart valve prosthesis: preliminary in vitro evaluation.
Vaesken, Antoine; Pidancier, Christian; Chakfe, Nabil; Heim, Frederic
2016-09-22
Transcatheter aortic valve implantation (TAVI) is nowadays a popular alternative technique to surgical valve replacement for critical patients. Biological valve tissue has been used in these devices for over a decade now with over 100,000 implantations. However, material degradations due to crimping for catheter insertion purpose have been reported, and with only 6-year follow-up, no information is available about the long-term durability of biological tissue. Moreover, expensive biological tissue harvesting and chemical treatment procedures tend to promote the development of synthetic valve leaflet materials. Textile polyester (PET) material is characterized by outstanding folding and strength properties combined with proven biocompatibility and could therefore be considered as a candidate to replace biological valve leaflets in TAVI devices. Nevertheless, the material should be preferentially partly elastic in order to limit water hammer effects at valve closing time and prevent exaggerated stress from occurring into the stent and the valve. The purpose of the present work is to study in vitro the mechanical as well as the hydrodynamic behavior of a hybrid elastic textile valve device combining non-deformable PET yarn and elastic polyurethane (PU) yarn. The hybrid valve properties are compared with those of a non-elastic textile valve. Testing results show improved hydrodynamic properties with the elastic construction. However, under fatigue conditions, the interaction between PU and PET yarns tends to limit the valve durability.
DOE Office of Scientific and Technical Information (OSTI.GOV)
Medley, S.S.
The application of charge exchange analyzers for the measurement of ion temperature in fusion plasma experiments requires a direct connection between the diagnostic and plasma-discharge vacuum chambers. Differential pumping of the gas load from the diagnostic stripping cell operated at > or approx. = 10/sup -3/ Torr is required to maintain the analyzer chamber at a pressure of < or approx. = 10/sup -6/ Torr. The migration of gases between the diagnostic and plasma vacuum chambers must be minimized. In particular, introduction of the analyzer stripping cell gas into the plasma chamber having a base pressure of < or approx.more » = 10/sup -8/ Torr must be suppressed. The charge exchange diagnostic for the Tokamak Fusion Test Reactor (TFTR) is comprised of two analyzer systems designed to contain a total of 18 independent mass/energy analyzers and one diagnostic neutral beam rated at 80 keV, 15 A. The associated arrays of multiple, interconnected vacuum systems were analyzed using the Vacuum System Transient Simulator (Vsts) computer program which models the transient transport of multigas species through complex networks of ducts, valves, traps, vacuum pumps, and other related vacuum system components. In addition to providing improved design performance at reduced costs, the analysis yields estimates for the exchange of tritium from the torus to the diagnostic components and of the diagnostic working gases to the torus.« less
Leak detection in medium density polyethylene (MDPE) pipe using pressure transient method
NASA Astrophysics Data System (ADS)
Amin, M. M.; Ghazali, M. F.; PiRemli, M. A.; Hamat, A. M. A.; Adnan, N. F.
2015-12-01
Water is an essential part of commodity for a daily life usage for an average person, from personal uses such as residential or commercial consumers to industries utilization. This study emphasizes on detection of leaking in medium density polyethylene (MDPE) pipe using pressure transient method. This type of pipe is used to analyze the position of the leakage in the pipeline by using Ensemble Empirical Mode Decomposition Method (EEMD) with signal masking. Water hammer would induce an impulse throughout the pipeline that caused the system turns into a surge of water wave. Thus, solenoid valve is used to create a water hammer through the pipelines. The data from the pressure sensor is collected using DASYLab software. The data analysis of the pressure signal will be decomposed into a series of wave composition using EEMD signal masking method in matrix laboratory (MATLAB) software. The series of decomposition of signals is then carefully selected which reflected intrinsic mode function (IMF). These IMFs will be displayed by using a mathematical algorithm, known as Hilbert transform (HT) spectrum. The IMF signal was analysed to capture the differences. The analyzed data is compared with the actual measurement of the leakage in term of percentage error. The error recorded is below than 1% and it is proved that this method highly reliable and accurate for leak detection.
Code of Federal Regulations, 2011 CFR
2011-01-01
... 14 Aeronautics and Space 3 2011-01-01 2011-01-01 false Oil valves. 125.137 Section 125.137....137 Oil valves. (a) Each oil valve must— (1) Comply with § 125.155; (2) Have positive stops or... the valve. (b) The closing of an oil shutoff means must not prevent feathering the propeller, unless...
Code of Federal Regulations, 2010 CFR
2010-01-01
... 14 Aeronautics and Space 3 2010-01-01 2010-01-01 false Oil valves. 125.137 Section 125.137....137 Oil valves. (a) Each oil valve must— (1) Comply with § 125.155; (2) Have positive stops or... the valve. (b) The closing of an oil shutoff means must not prevent feathering the propeller, unless...
21 CFR 870.3935 - Prosthetic heart valve holder.
Code of Federal Regulations, 2010 CFR
2010-04-01
... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Prosthetic heart valve holder. 870.3935 Section... heart valve holder. (a) Identification. A prosthetic heart valve holder is a device used to hold a replacement heart valve while it is being sutured into place. (b) Classification. Class I. The device is...
49 CFR 192.193 - Valve installation in plastic pipe.
Code of Federal Regulations, 2011 CFR
2011-10-01
... 49 Transportation 3 2011-10-01 2011-10-01 false Valve installation in plastic pipe. 192.193... Components § 192.193 Valve installation in plastic pipe. Each valve installed in plastic pipe must be designed so as to protect the plastic material against excessive torsional or shearing loads when the valve...
Code of Federal Regulations, 2012 CFR
2012-01-01
... 14 Aeronautics and Space 3 2012-01-01 2012-01-01 false Oil valves. 125.137 Section 125.137....137 Oil valves. (a) Each oil valve must— (1) Comply with § 125.155; (2) Have positive stops or... the valve. (b) The closing of an oil shutoff means must not prevent feathering the propeller, unless...
Code of Federal Regulations, 2014 CFR
2014-01-01
... 14 Aeronautics and Space 3 2014-01-01 2014-01-01 false Oil valves. 125.137 Section 125.137....137 Oil valves. (a) Each oil valve must— (1) Comply with § 125.155; (2) Have positive stops or... the valve. (b) The closing of an oil shutoff means must not prevent feathering the propeller, unless...
Code of Federal Regulations, 2013 CFR
2013-01-01
... 14 Aeronautics and Space 3 2013-01-01 2013-01-01 false Oil valves. 125.137 Section 125.137....137 Oil valves. (a) Each oil valve must— (1) Comply with § 125.155; (2) Have positive stops or... the valve. (b) The closing of an oil shutoff means must not prevent feathering the propeller, unless...
21 CFR 870.3935 - Prosthetic heart valve holder.
Code of Federal Regulations, 2013 CFR
2013-04-01
... 21 Food and Drugs 8 2013-04-01 2013-04-01 false Prosthetic heart valve holder. 870.3935 Section... heart valve holder. (a) Identification. A prosthetic heart valve holder is a device used to hold a replacement heart valve while it is being sutured into place. (b) Classification. Class I. The device is...
21 CFR 870.3935 - Prosthetic heart valve holder.
Code of Federal Regulations, 2012 CFR
2012-04-01
... 21 Food and Drugs 8 2012-04-01 2012-04-01 false Prosthetic heart valve holder. 870.3935 Section... heart valve holder. (a) Identification. A prosthetic heart valve holder is a device used to hold a replacement heart valve while it is being sutured into place. (b) Classification. Class I. The device is...
21 CFR 870.3935 - Prosthetic heart valve holder.
Code of Federal Regulations, 2014 CFR
2014-04-01
... 21 Food and Drugs 8 2014-04-01 2014-04-01 false Prosthetic heart valve holder. 870.3935 Section... heart valve holder. (a) Identification. A prosthetic heart valve holder is a device used to hold a replacement heart valve while it is being sutured into place. (b) Classification. Class I. The device is...
21 CFR 870.3935 - Prosthetic heart valve holder.
Code of Federal Regulations, 2011 CFR
2011-04-01
... 21 Food and Drugs 8 2011-04-01 2011-04-01 false Prosthetic heart valve holder. 870.3935 Section... heart valve holder. (a) Identification. A prosthetic heart valve holder is a device used to hold a replacement heart valve while it is being sutured into place. (b) Classification. Class I. The device is...
49 CFR 192.193 - Valve installation in plastic pipe.
Code of Federal Regulations, 2010 CFR
2010-10-01
... 49 Transportation 3 2010-10-01 2010-10-01 false Valve installation in plastic pipe. 192.193... Components § 192.193 Valve installation in plastic pipe. Each valve installed in plastic pipe must be designed so as to protect the plastic material against excessive torsional or shearing loads when the valve...
49 CFR 192.193 - Valve installation in plastic pipe.
Code of Federal Regulations, 2014 CFR
2014-10-01
... 49 Transportation 3 2014-10-01 2014-10-01 false Valve installation in plastic pipe. 192.193... Components § 192.193 Valve installation in plastic pipe. Each valve installed in plastic pipe must be designed so as to protect the plastic material against excessive torsional or shearing loads when the valve...
49 CFR 192.193 - Valve installation in plastic pipe.
Code of Federal Regulations, 2012 CFR
2012-10-01
... 49 Transportation 3 2012-10-01 2012-10-01 false Valve installation in plastic pipe. 192.193... Components § 192.193 Valve installation in plastic pipe. Each valve installed in plastic pipe must be designed so as to protect the plastic material against excessive torsional or shearing loads when the valve...
49 CFR 192.193 - Valve installation in plastic pipe.
Code of Federal Regulations, 2013 CFR
2013-10-01
... 49 Transportation 3 2013-10-01 2013-10-01 false Valve installation in plastic pipe. 192.193... Components § 192.193 Valve installation in plastic pipe. Each valve installed in plastic pipe must be designed so as to protect the plastic material against excessive torsional or shearing loads when the valve...
Loger, K; Engel, A; Haupt, J; Lima de Miranda, R; Lutter, G; Quandt, E
2016-03-01
Heart valves are constantly exposed to high dynamic loading and are prone to degeneration. Therefore, it is a challenge to develop a durable heart valve substitute. A promising approach in heart valve engineering is the development of hybrid scaffolds which are composed of a mechanically strong inorganic mesh enclosed by valvular tissue. In order to engineer an efficient, durable and very thin heart valve for transcatheter implantations, we developed a fabrication process for microstructured heart valve leaflets made from a nickel-titanium (NiTi) thin film shape memory alloy. To examine the capability of microstructured NiTi thin film as a matrix scaffold for tissue engineered hybrid heart valves, leaflets were successfully seeded with smooth muscle cells (SMCs). In vitro pulsatile hydrodynamic testing of the NiTi thin film valve leaflets demonstrated that the SMC layer significantly improved the diastolic sufficiency of the microstructured leaflets, without affecting the systolic efficiency. Compared to an established porcine reference valve model, magnetron sputtered NiTi thin film material demonstrated its suitability for hybrid tissue engineered heart valves.
NASA Astrophysics Data System (ADS)
Kaiser, Alexander
2016-11-01
The mitral valve is one of four valves in the human heart. The valve opens to allow oxygenated blood from the lungs to fill the left ventricle, and closes when the ventricle contracts to prevent backflow. The valve is composed of two fibrous leaflets which hang from a ring. These leaflets are supported like a parachute by a system of strings called chordae tendineae. In this talk, I will describe a new computational model of the mitral valve. To generate geometry, general information comes from classical anatomy texts and the author's dissection of porcine hearts. An MRI image of a human heart is used to locate the tips of the papillary muscles, which anchor the chordae tendineae, in relation to the mitral ring. The initial configurations of the valve leaflets and chordae tendineae are found by solving solving an equilibrium elasticity problem. The valve is then simulated in fluid (blood) using the immersed boundary method over multiple heart cycles in a model valve tester. We aim to identify features and mechanisms that influence or control valve function. Support from National Science Foundation, Graduate Research Fellowship Program, Grant DGE 1342536.
Research on digital system design of nuclear power valve
NASA Astrophysics Data System (ADS)
Zhang, Xiaolong; Li, Yuan; Wang, Tao; Dai, Ye
2018-04-01
With the progress of China's nuclear power industry, nuclear power plant valve products is in a period of rapid development, high performance, low cost, short cycle of design requirements for nuclear power valve is proposed, so there is an urgent need for advanced digital design method and integrated design platform to provide technical support. Especially in the background of the nuclear power plant leakage in Japan, it is more practical to improve the design capability and product performance of the nuclear power valve. The finite element numerical analysis is a common and effective method for the development of nuclear power valves. Nuclear power valve has high safety, complexity of valve chamber and nonlinearity of seal joint surface. Therefore, it is urgent to establish accurate prediction models for earthquake prediction and seal failure to meet engineering accuracy and calculation conditions. In this paper, a general method of finite element modeling for nuclear power valve assembly and key components is presented, aiming at revealing the characteristics and rules of finite element modeling of nuclear power valves, and putting forward aprecision control strategy for finite element models for nuclear power valve characteristics analysis.
Adjustable shunt valve-induced magnetic resonance imaging artifact: a comparative study.
Toma, Ahmed K; Tarnaris, Andrew; Grieve, Joan P; Watkins, Laurence D; Kitchen, Neil D
2010-07-01
In this paper, the authors' goal was to compare the artifact induced by implanted (in vivo) adjustable shunt valves in spin echo, diffusion weighted (DW), and gradient echo MR imaging pulse sequences. The MR images obtained in 8 patients with proGAV and 6 patients with Strata II adjustable shunt valves were assessed for artifact areas in different planes as well as the total volume for different pulse sequences. Artifacts induced by the Strata II valve were significantly larger than those induced by proGAV valve in spin echo MR imaging pulse sequence (29,761 vs 2450 mm(3) on T2-weighted fast spin echo, p = 0.003) and DW images (100,138 vs 38,955 mm(3), p = 0.025). Artifacts were more marked on DW MR images than on spin echo pulse sequence for both valve types. Adjustable valve-induced artifacts can conceal brain pathology on MR images. This should influence the choice of valve implantation site and the type of valve used. The effect of artifacts on DW images should be highlighted pending the development of less MR imaging artifact-inducing adjustable shunt valves.
Mvondo, Charles Mve; Pugliese, Marta; Giamberti, Alessandro; Chelo, David; Kuate, Liliane Mfeukeu; Boombhi, Jerome; Dailor, Ellen Marie
2016-01-01
Rheumatic valve disease, a consequence of acute rheumatic fever, remains endemic in developing countries in the sub-Saharan region where it is the leading cause of heart failure and cardiovascular death, involving predominantly a young population. The involvement of the mitral valve is pathognomonic and mitral surgery has become the lone therapeutic option for the majority of these patients. However, controversies exist on the choice between valve repair or prosthetic valve replacement. Although the advantages of mitral valve repair over prosthetic valve replacement in degenerative mitral disease are well established, this has not been the case for rheumatic lesions, where the use of prosthetic valves, specifically mechanical devices, even in poorly compliant populations remains very common. These patients deserve more accurate evaluation in the choice of the surgical strategy which strongly impacts the post-operative outcomes. This report discusses the factors supporting mitral repair surgery in rheumatic disease, according to the patients' characteristics and the effectiveness of the current repair techniques compared to prosthetic valve replacement in developing countries.
NASA Technical Reports Server (NTRS)
Butt, Adam; Popp, Christopher G.; Holt, Kimberly A.; Pitts, Hank M.
2010-01-01
The Ares I launch vehicle is the selected design, chosen to return humans to the moon, Mars, and beyond. It is configured in two inline stages: the First Stage is a Space Shuttle derived five-segment Solid Rocket Booster and the Upper Stage is powered by a Saturn V derived J-2X engine. During launch, roll control for the First Stage (FS) is handled by a dedicated Roll Control System (RoCS) located on the connecting Interstage. That system will provide the Ares I with the ability to counteract induced roll torque while any induced yaw or pitch moments are handled by vectoring of the booster nozzle. This paper provides an overview of NASA s Ares I FS RoCS cold flow development test program including detailed test objectives, types of tests run to meet those objectives, an overview of the results, and applicable lessons learned. The test article was built and tested at the NASA Marshall Space Flight Center in Huntsville, AL. The FS RoCS System Development Test Article (SDTA) is a full scale, flight representative water flow test article whose primary objective was to obtain fluid system performance data to evaluate integrated system level performance characteristics and verify analytical models. Development testing and model correlation was deemed necessary as there is little historical precedent for similar large flow, pulsing systems such as the FS RoCS. The cold flow development test program consisted of flight-similar tanks, pressure regulators, and thruster valves, as well as plumbing simulating flight geometries, combined with other facility grade components and structure. Orifices downstream of the thruster valves were used to simulate the pressure drop through the thrusters. Additional primary objectives of this test program were to: evaluate system surge pressure (waterhammer) characteristics due to thruster valve operation over a range of mission duty cycles at various feed system pressures, evaluate temperature transients and heat transfer in the pressurization system, including regulator blowdown and propellant ullage performance, measure system pressure drops for comparison to analysis of tubing and components, and validate system activation and re-activation procedures for the helium pressurant system. Secondary objectives included: validating system processes for loading, unloading, and purging, validating procedures and system response for multiple failure scenarios, including relief valve operation, and evaluating system performance for contingency scenarios. The test results of the cold flow development test program are essential in validating the performance and interaction of the Roll Control System and anchoring analysis tools and results to a Critical Design Review level of fidelity.
Hydraulically actuated gas exchange valve assembly and engine using same
Carroll, Thomas S.; Taylor, Gregory O.
2002-09-03
An engine comprises a housing that defines a hollow piston cavity that is separated from a gas passage by a valve seat. The housing further defines a biasing hydraulic cavity and a control hydraulic cavity. A gas valve member is also included in the engine and is movable relative to the valve seat between an open position at which the hollow piston cavity is open to the gas passage and a closed position in which the hollow piston cavity is blocked from the gas passage. The gas valve member includes a ring mounted on a valve piece and a retainer positioned between the ring and the valve piece. A closing hydraulic surface is included on the gas valve member and is exposed to liquid pressure in the biasing hydraulic cavity.
Jung, Joonho; Lee, Cheol Joo; Lim, Sang-Hyun; Choi, Ho; Park, Soo-Jin
2013-01-01
A 51-year-old male was admitted to the hospital with complaints of fever and hemoptysis. After evaluation of the fever focus, he was diagnosed with pulmonary valve infective endocarditis. Thus pulmonary valve replacement and antibiotics therapy were performed and discharged. He was brought to the emergency unit presenting with a high fever (>39℃) and general weakness 6 months after the initial operation. The echocardiography revealed prosthetic pulmonary valve endocarditis. Therefore, redo-pulmonary valve replacement using valved conduit was performed in the Rastelli fashion because of the risk of pulmonary arterial wall injury and recurrent endocarditis from the remnant inflammatory tissue. We report here on the successful surgical treatment of prosthetic pulmonary valve endocarditis with an alternative surgical method. PMID:23772409
Remote manual operator for space station intermodule ventilation valve
NASA Technical Reports Server (NTRS)
Guyaux, James R.
1996-01-01
The Remote Manual Operator (RMO) is a mechanism used for manual operation of the Space Station Intermodule Ventilation (IMV) valve and for visual indication of valve position. The IMV is a butterfly-type valve, located in the ventilation or air circulation ducts of the Space Station, and is used to interconnect or isolate the various compartments. The IMV valve is normally operated by an electric motor-driven actuator under computer or astronaut control, but it can also be operated manually with the RMO. The IMV valve RMO consists of a handle with a deployment linkage, a gear-driven flexible shaft, and a linkage to disengage the electric motor actuator during manual operation. It also provides visual indication of valve position. The IMV valve RMO is currently being prepared for qualification testing.
Hsi, David H; Ryan, Gerald F; Taft, Janice; Arnone, Thomas J
2003-01-01
An 81-year-old woman was evaluated for prosthetic mitral valve function. She had received a Harken disk mitral valve 29 years earlier due to severe mitral valve disease. This particular valve prosthesis is known for premature disk edge wear and erosion. The patients 2-dimensional Doppler echocardiogram showed the distinctive appearance of a disk mitral valve prosthesis. Color Doppler in diastole showed a unique crown appearance, with initial flow acceleration around the disk followed by convergence to laminar flow in the left ventricle. Cineradiographic imaging revealed normal valve function and minimal disk erosion. We believe this to be the longest reported follow-up of a surviving patient with a rare Harken disk valve. We present images with unique echocardiographic and cineangiographic features.
A 29-Year-Old Harken Disk Mitral Valve
Hsi, David H.; Ryan, Gerald F.; Taft, Janice; Arnone, Thomas J.
2003-01-01
An 81-year-old woman was evaluated for prosthetic mitral valve function. She had received a Harken disk mitral valve 29 years earlier due to severe mitral valve disease. This particular valve prosthesis is known for premature disk edge wear and erosion. The patient's 2-dimensional Doppler echocardiogram showed the distinctive appearance of a disk mitral valve prosthesis. Color Doppler in diastole showed a unique crown appearance, with initial flow acceleration around the disk followed by convergence to laminar flow in the left ventricle. Cineradiographic imaging revealed normal valve function and minimal disk erosion. We believe this to be the longest reported follow-up of a surviving patient with a rare Harken disk valve. We present images with unique echocardiographic and cineangiographic features. (Tex Heart Inst J 2003;30:319–21) PMID:14677746
Hurwitz, Seth Eric; Waxman, Daniel; Hecht, Susan
2009-09-01
Pannus formation and valve thrombus can cause prosthetic valve failure. The authors report the case of a 50-year-old woman who presented to the emergency room with decompensated heart failure secondary to mechanical valve dysfunction. On two-dimensional and transesophageal echocardiography, the patient had severe aortic stenosis and regurgitation. A thrombus seen on the valve was felt to be the etiology of her prosthetic valve failure. She underwent emergent cardiac surgery for aortic valve replacement. Pathology revealed that although a small thrombus was present, extensive pannus was the underlying mechanism of valve dysfunction. Differentiation between pannus and thrombus may have important clinical implications, but this case illustrates that distinguishing between these entities by echocardiographic and clinical criteria may not be possible.
Comparison of Outcomes of Resident-performed Ahmed Valve Implantation vs Trabeculectomy
Kammerdiener, Leah L; Wannamaker, Kendall W; Fan, Jie; Sharpe, Elizabeth D
2016-01-01
ABSTRACT Aims: To compare outcomes of resident-performed Ahmed valve surgery vs trabeculectomy in a Veteran Affairs medical facility. Materials and methods: A retrospective cohort of 103 eyes in 91 patients receiving Ahmed valve (valve) or trabeculectomy (trab) performed at a Veterans Administration Medical Center by residents in their third year of training. The primary outcomes included intraocular pressure (IOP), treatment failure, and complications over 1 year. Results: Of 103 eyes, 44 received valve and 59 received trab. Primary open-angle glaucoma was primary diagnosis more often in trab, while neovascular glaucoma predominated in the valve group (p < 0.001). Preoperative mean IOP was 35.1 ± 11.8 and 24.5 ± 7.1 mm Hg for valve and trabeculectomy respectively (p < 0.001), but at 1 year the IOP difference between groups was not statistically significant (p = 0.064). Overall, 11 (25.0%) and 11 (18.6%) eyes met any criteria for failure for valve and trab respectively. At 1 year, 22.5% of valves had IOP > 21 mm Hg vs only 4.3% of trab (p = 0.02). Complications were infrequent. There were no intraoperative complications for valve, whereas five for trab. Most common immediate complication for valve was hyphema. Both groups had low rates of choroidal effusions and reoperation. Conclusion: Ahmed valve implantation and trabeculectomy produce significant reductions in IOP when performed by residents-in-training. Valves tend to be used more frequently in patients with secondary glaucoma. Although complication profiles differ between procedures, both are safe and well tolerated when performed by resident physicians. Clinical significance: This study provides support for evidence-based patient counseling that supervised, resident-performed Ahmed valve implantation and trabeculectomy are indeed safe and effective. How to cite this article: Sharpe RA, Kammerdiener LL, Wannamaker KW, Fan J, Sharpe ED. Comparison of Outcomes of Resident-performed Ahmed Valve Implantation vs Trabeculectomy. J Curr Glaucoma Pract 2016;10(2):60-67. PMID:27536049
FLUID-STRUCTURE INTERACTION MODELS OF THE MITRAL VALVE: FUNCTION IN NORMAL AND PATHOLOGIC STATES
DOE Office of Scientific and Technical Information (OSTI.GOV)
Kunzelman, K. S.; Einstein, Daniel R.; Cochran, R. P.
2007-08-29
Successful mitral valve repair is dependent upon a full understanding of normal and abnormal mitral valve anatomy and function. Computational analysis is one such method that can be applied to simulate mitral valve function in order to analyze the roles of individual components, and evaluate proposed surgical repair. We developed the first three-dimensional, finite element (FE) computer model of the mitral valve including leaflets and chordae tendineae, however, one critical aspect that has been missing until the last few years was the evaluation of fluid flow, as coupled to the function of the mitral valve structure. We present here ourmore » latest results for normal function and specific pathologic changes using a fluid-structure interaction (FSI) model. Normal valve function was first assessed, followed by pathologic material changes in collagen fiber volume fraction, fiber stiffness, fiber splay, and isotropic stiffness. Leaflet and chordal stress and strain, and papillary muscle force was determined. In addition, transmitral flow, time to leaflet closure, and heart valve sound were assessed. Model predictions in the normal state agreed well with a wide range of available in-vivo and in-vitro data. Further, pathologic material changes that preserved the anisotropy of the valve leaflets were found to preserve valve function. By contrast, material changes that altered the anisotropy of the valve were found to profoundly alter valve function. The addition of blood flow and an experimentally driven microstructural description of mitral tissue represent significant advances in computational studies of the mitral valve, which allow further insight to be gained. This work is another building block in the foundation of a computational framework to aid in the refinement and development of a truly noninvasive diagnostic evaluation of the mitral valve. Ultimately, it represents the basis for simulation of surgical repair of pathologic valves in a clinical and educational setting.« less
NASA Astrophysics Data System (ADS)
Jung, Wooseok; Barrett, Matthew; Brooks, Carla; Rivera, Andrew; Birdsell, Dawn N.; Wagner, David M.; Zenhausern, Frederic
2015-12-01
We present a new elastomeric valve for integrated nucleic acid analysis by capillary electrophoresis. The valve functions include metering to capture a designated volume of biological sample into a polymerase chain reaction (PCR) chamber, sealing to preserve the sample during PCR cycling, and transfer of the PCR-products and on-chip formamide post-processing for the analysis of DNA fragments by capillary gel electrophoresis. This new valve differs from prior art polydimethylsiloxane (PDMS) valves in that the valve is not actuated externally by air-pressure or vacuum so that it simplifies a DNA analysis system by eliminating the need for an air-pressure or vacuum source, and off-cartridge solenoid valves, control circuit boards and software. Instead, the new valve is actuated by a thermal cycling peltier assembly integrated within the hardware instrument that tightly comes in contact with a microfluidic cartridge for thermal activation during PCR, so that it spontaneously closes the valve without an additional actuator system. The valve has bumps in the designated locations so that it has a self-alignment that does not require precise alignment of a valve actuator. Moreover, the thickness of the new valve is around 600 μm with an additional bump height of 400 μm so that it is easy to handle and very feasible to fabricate by injection molding compared to other PDMS valves whose thicknesses are around 30-100 μm. The new valve provided over 95% of metering performance in filling the fixed volume of the PCR chamber, preserved over 97% of the sample volume during PCR, and showed very comparable capillary electrophoresis peak heights to the benchtop assay tube controls with very consistent transfer volume of the PCR-product and on-chip formamide. The new valve can perform a core function for integrated nucleic acid analysis by capillary electrophoresis.
Fazal, Iftikhar A; Alfakih, Khaled; Wilcox, Robert G; Walsh, John T
2009-05-01
Aortic stenosis (AS) is the most common indication for valve surgery. Recent data suggested an increased risk of cerebral emboli when the aortic valve is crossed to obtain 'pull-back' gradient. We conducted a large questionnaire based study to evaluate current practice in the assessment of aortic valve gradient amongst cardiologists and the preferences of cardiac surgeons in the UK. E-mail questionnaires were sent to 645 (72%) UK consultant cardiologists and to 198 (92%) UK consultant cardiac surgeons. 232 cardiologists and 52 cardiac surgeons responded. 53% of cardiologists routinely attempt to cross the valve in moderate AS while only 23% do so in severe AS. 38% of cardiologists in the age group '50+ years' cross the valve in severe AS compared to 13% in the age group '30-40 years'. Common reasons given for crossing a stenosed valve included 'to verify the echocardiographic gradient' (85%) and 'maintaining skill' (24%). 64% of cardiologists have changed their views on the necessity of crossing the valve in the last ten years. Although the majority appreciate the increased risk of crossing the valve only 18% of patients are consented differently if crossing the valve is planned. 26% of cardiac surgeons prefer the valve to be crossed to provide information on 'pull-back' gradient, 32% for LV function assessment and 14% to confirm MV competence. 92% would accept echocardiographic data alone if both the gradient and aortic valve area were available and considered correct. Our survey found that the practice of crossing the aortic valve has changed in the last 10 years and that younger consultant cardiologists are less likely to cross the aortic valve. Increasing confidence in echocardiographic data and potential complications of crossing the valve are implicated. 92% of cardiac surgeons do not require the valve to be crossed if the echo data is considered accurate.
Sardari Nia, Peyman; Heuts, Samuel; Daemen, Jean; Luyten, Peter; Vainer, Jindrich; Hoorntje, Jan; Cheriex, Emile; Maessen, Jos
2017-02-01
Mitral valve repair performed by an experienced surgeon is superior to mitral valve replacement for degenerative mitral valve disease; however, many surgeons are still deterred from adapting this procedure because of a steep learning curve. Simulation-based training and planning could improve the surgical performance and reduce the learning curve. The aim of this study was to develop a patient-specific simulation for mitral valve repair and provide a proof of concept of personalized medicine in a patient prospectively planned for mitral valve surgery. A 65-year old male with severe symptomatic mitral valve regurgitation was referred to our mitral valve heart team. On the basis of three-dimensional (3D) transoesophageal echocardiography and computed tomography, 3D reconstructions of the patient's anatomy were constructed. By navigating through these reconstructions, the repair options and surgical access were chosen (minimally invasive repair). Using rapid prototyping and negative mould fabrication, we developed a process to cast a patient-specific mitral valve silicone replica for preoperative repair in a high-fidelity simulator. Mitral valve and negative mould were printed in systole to capture the pathology when the valve closes. A patient-specific mitral valve silicone replica was casted and mounted in the simulator. All repair techniques could be performed in the simulator to choose the best repair strategy. As the valve was printed in systole, no special testing other than adjusting the coaptation area was required. Subsequently, the patient was operated, mitral valve pathology was validated and repair was successfully done as in the simulation. The patient-specific simulation and planning could be applied for surgical training, starting the (minimally invasive) mitral valve repair programme, planning of complex cases and the evaluation of new interventional techniques. The personalized medicine could be a possible pathway towards enhancing reproducibility, patient's safety and effectiveness of a complex surgical procedure. © The Author 2016. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.
NASA Astrophysics Data System (ADS)
Dong, Dai; Li, Xiaoning
2015-03-01
High-pressure solenoid valve with high flow rate and high speed is a key component in an underwater driving system. However, traditional single spool pilot operated valve cannot meet the demands of both high flow rate and high speed simultaneously. A new structure for a high pressure solenoid valve is needed to meet the demand of the underwater driving system. A novel parallel-spool pilot operated high-pressure solenoid valve is proposed to overcome the drawback of the current single spool design. Mathematical models of the opening process and flow rate of the valve are established. Opening response time of the valve is subdivided into 4 parts to analyze the properties of the opening response. Corresponding formulas to solve 4 parts of the response time are derived. Key factors that influence the opening response time are analyzed. According to the mathematical model of the valve, a simulation of the opening process is carried out by MATLAB. Parameters are chosen based on theoretical analysis to design the test prototype of the new type of valve. Opening response time of the designed valve is tested by verifying response of the current in the coil and displacement of the main valve spool. The experimental results are in agreement with the simulated results, therefore the validity of the theoretical analysis is verified. Experimental opening response time of the valve is 48.3 ms at working pressure of 10 MPa. The flow capacity test shows that the largest effective area is 126 mm2 and the largest air flow rate is 2320 L/s. According to the result of the load driving test, the valve can meet the demands of the driving system. The proposed valve with parallel spools provides a new method for the design of a high-pressure valve with fast response and large flow rate.
Seiffert, Moritz; Bader, Ralf; Kappert, Utz; Rastan, Ardawan; Krapf, Stephan; Bleiziffer, Sabine; Hofmann, Steffen; Arnold, Martin; Kallenbach, Klaus; Conradi, Lenard; Schlingloff, Friederike; Wilbring, Manuel; Schäfer, Ulrich; Diemert, Patrick; Treede, Hendrik
2014-10-01
This analysis reports on the initial German multicenter experience with the JenaValve (JenaValve Technology GmbH, Munich, Germany) transcatheter heart valve for the treatment of pure aortic regurgitation. Experience with transcatheter aortic valve implantation (TAVI) for severe aortic regurgitation is limited due to the risk of insufficient anchoring of the valve stent within the noncalcified aortic annulus. Transapical TAVI with a JenaValve for the treatment of severe aortic regurgitation was performed in 31 patients (age 73.8 ± 9.1 years) in 9 German centers. All patients were considered high risk for surgery (logistic EuroSCORE [European System for Cardiac Operative Risk Evaluation] 23.6 ± 14.5%) according to a local heart team consensus. Procedural results and clinical outcomes up to 6 months were analyzed. Implantation was successful in 30 of 31 cases (aortic annulus diameter 24.7 ± 1.5 mm); transcatheter heart valve dislodgement necessitated valve-in-valve implantation in 1 patient. Post-procedural aortic regurgitation was none/trace in 28 of 31 and mild in 3 of 31 patients. During follow-up, 2 patients underwent valvular reinterventions (surgical aortic valve replacement for endocarditis, valve-in-valve implantation for increasing paravalvular regurgitation). All-cause mortality was 12.9% and 19.3% at 30 days and 6 months, respectively. In the remaining patients, a significant improvement in New York Heart Association class was observed and persisted up to 6 months after TAVI. Aortic regurgitation remains a challenging pathology for TAVI. After initial demonstration of feasibility, this multicenter study revealed the JenaValve transcatheter heart valve as a reasonable option in this subset of patients. However, a significant early noncardiac mortality related to the high-risk population emphasizes the need for careful patient selection. Copyright © 2014 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
NASA Technical Reports Server (NTRS)
Perkins, Gerald S. (Inventor)
1979-01-01
A check valve which closes more rapidly to prevent wearing of the valve seat and of the valve member that seals thereagainst, including a solenoid or other actuator that aids the normal spring to quickly close the valve at approximately the time when downpath fluid flow would stop, the actuator then being deenergized. The control circuit that operates the actuator can include a pair of pressure sensors sensing pressure both upstream and downstream from the valve seat. Where the valve is utilized to control flow to or from a piston pump, energization of the actuator can be controlled by sensing when the pump piston reaches its extreme of travel.
Treede, Hendrik; Rastan, Ardawan; Ferrari, Markus; Ensminger, Stephan; Figulla, Hans-Reiner; Mohr, Friedrich-Wilhelm
2012-09-01
The JenaValve is a next-generation TAVI device which consists of a well-proven porcine root valve mounted on a low-profile nitinol stent. Feeler guided positioning and clip fixation on the diseased leaflets allow for anatomically correct implantation of the device without rapid pacing. Safety and efficacy of transapical aortic valve implantation using the JenaValve were evaluated in a multicentre prospective study that showed good short and midterm results. The valve was CE-mark released in Europe in September 2011. A post-market registry ensures on-going and prospective data collection in "real-world" patients. The transfemoral JenaValve delivery system will be evaluated in a first-in-man study in the near future.
Quantification and comparison of the mechanical properties of four human cardiac valves.
Pham, Thuy; Sulejmani, Fatiesa; Shin, Erica; Wang, Di; Sun, Wei
2017-05-01
Although having the same ability to permit unidirectional flow within the heart, the four main valves-the mitral valve (MV), aortic (AV), tricuspid (TV) and pulmonary (PV) valves-experience different loading conditions; thus, they exhibit different structural integrity from one another. Most research on heart valve mechanics have been conducted mainly on MV and AV or an individual valve, but none quantify and compare the mechanical and structural properties among the four valves from the same aged patient population whose death was unrelated to cardiovascular disease. A total of 114 valve leaflet samples were excised from 12 human cadavers whose death was unrelated to cardiovascular disease (70.1±3.7years old). Tissue mechanical and structural properties were characterized by planar biaxial mechanical testing and histological methods. The experimental data were then fitted with a Fung-type constitutive model. The four valves differed substantially in thickness, degree of anisotropy, and stiffness. The leaflets of the left heart (the AV leaflets and the anterior mitral leaflets, AML) were significantly stiffer and less compliant than their counterparts in the right heart. TV leaflets were the most extensible and isotropic, while AML and AV leaflets were the least extensible and the most anisotropic. Age plays a significant role in the reduction of leaflet stiffness and extensibility with nearly straightened collagen fibers observed in the leaflet samples from elderly groups (65years and older). Results from 114 human leaflet samples not only provided a baseline quantification of the mechanical properties of aged human cardiac valves, but also offered a better understanding of the age-dependent differences among the four valves. It is hoped that the experimental data collected and the associated constitutive models in this study can facilitate future studies of valve diseases, treatments and the development of interventional devices. Most research on heart valve mechanics have been conducted mainly on mitral and aortic valves or an individual valve, but none quantify and compare the mechanical and structural properties among the four valves from the same relatively healthy elderly patient population. In this study, the mechanical and microstructural properties of 114 leaflets of aortic, mitral, pulmonary and tricuspid valves from 12 human cadaver hearts were mechanically tested, analyzed and compared. Our results not only provided a baseline quantification of the mechanical properties of aged human valves, but a age range between patients (51-87years) also offers a better understanding of the age-dependent differences among the four valves. It is hoped that the obtained experimental data and associated constitutive parameters can facilitate studies of valve diseases, treatments and the development of interventional devices. Copyright © 2017 Acta Materialia Inc. Published by Elsevier Ltd. All rights reserved.
Effect of prior aortic valve intervention on results of the Ross operation.
Sakaguchi, Hidehito; Elkins, Ronald C; Lane, Mary M; McCue, Carolyn
2003-07-01
Patient-related factors, aortic insufficiency, bicuspid aortic valve, aortic annulus dilatation, ascending aortic dilatation or aneurysm, and aortic valve endocarditis have been suggested as affecting the results of the Ross operation. The study aim was to assess the impact of prior aortic valve intervention on early and late results of a Ross operation. A total of 399 patients who underwent surgery between August 1986 and September 2000 were reviewed retrospectively. The patients were grouped as: no prior aortic valve intervention (NOAVI, n = 219); prior aortic valvuloplasty (AVP, n = 106); prior balloon aortic valvuloplasty (AVB, n = 40); and prior aortic valve replacement (AVR, n = 34). Details of operative and late mortality, autograft valve function, and homograft valve function were analyzed. Operative mortality was higher for AVB (10%; three deaths in neonates) than the other groups (from 2.3% to 5.9%) (p = 0.084). Freedom from autograft valve degeneration, defined as severe autograft valve insufficiency, non-endocarditis autograft valve reoperation or valve-related death, ranged from 93 +/- 3% for AVP to 76 +/- 8% for NOAVI at 10 years (p = 0.43). Freedom from homograft reoperation in the pulmonary position was 100% for AVB at six years, and 99 +/- 1% for AVP, 82 +/- 8% for NOAVI, and 70 +/- 13% for AVR at 10 years (p = 0.0026). There appears to be no significant difference between patients with and without prior aortic valve surgery, with respect to operative mortality or late autograft function. However, patients with prior AVR appear to have a significantly higher homograft reoperation rate after a Ross operation, the reasons for which are uncertain.
Yamaki, F; Nakano, K; Endo, M; Hashimoto, A; Koyanagi, H
1994-02-01
Although a high incidence of strut fracture of the Björk-Shiley convexo-concave (C-C) valve prosthesis has been reported, it is still controversial whether the prosthesis functioning normally should be replaced electively. To clarify our policy for this issue, we reviewed 28 patients who had undergone mitral valve replacement with a C-C valve prosthesis, and evaluated long-term results according to STS guidelines. The incidence of valve-related complications expressed as % patient-year were; structural deterioration 0.30, nonstructural dysfunction 0.30, thromboembolism 1.20. There were no thrombosed valves, anticoagulant-related hemorrhage or prosthetic valve endocarditis. The actuarial free rate after valve replacement with the C-C at 10 years, constructed by the Kaplan-Meier method, were compared with those with St. Jude Medical (SJM) valve prosthesis. (1) Actuarial survival: 77.9 vs 87.3 (NS), (2) Reoperation-free: 91.6 vs 98.0 (p < 0.05), (3) thromboembolism-free: 83.7 vs 83.9 (NS), (4) Event-free (hospital death+reoperation+valve-related complications: 71.2 vs 77.7 (NS). The long-term results with the C-C were compatible to those with a SJM valve prosthesis. However, if a strut fracture occurs, it is very difficult to save the patient's life. The hospital mortality of re-replacement of valve prosthesis during the last 10 years, at our institute, was 2.6%, which is almost equal to the cumulative risk of strut fracture of the C-C valve after 10 years (3.0%).(ABSTRACT TRUNCATED AT 250 WORDS)
Hydraulic Evaluation of Culvert Valves at Eisenhower and Snell Locks, St. Lawrence Seaway
2015-06-01
ER D C/ CH L TR -1 5- 7 Hydraulic Evaluation of Culvert Valves at Eisenhower and Snell Locks, St. Lawrence Seaway Co as ta l a nd H...client/default. ERDC/CHL TR-15-7 June 2015 Hydraulic Evaluation of Culvert Valves at Eisenhower and Snell Locks, St. Lawrence Seaway...filling valve well of the Snell Lock’s south-wall culvert . The new vertical-frame valve operated at a slower rate than the double-skin-plate valve
Van Dyke, William J.
1992-01-01
A fast valve is disclosed that can close on the order of 7 milliseconds. It is closed by the force of a compressed air spring with the moving parts of the valve designed to be of very light weight and the valve gate being of wedge shaped with O-ring sealed faces to provide sealing contact without metal to metal contact. The combination of the O-ring seal and an air cushion create a soft final movement of the valve closure to prevent the fast air acting valve from having a harsh closing.
TTK Chitra tilting disc heart valve model TC2: An assessment of fatigue life and durability.
Subhash, N N; Rajeev, Adathala; Sujesh, Sreedharan; Muraleedharan, C V
2017-08-01
Average age group of heart valve replacement in India and most of the Third World countries is below 30 years. Hence, the valve for such patients need to be designed to have a service life of 50 years or more which corresponds to 2000 million cycles of operation. The purpose of this study was to assess the structural performance of the TTK Chitra tilting disc heart valve model TC2 and thereby address its durability. The TC2 model tilting disc heart valves were assessed to evaluate the risks connected with potential structural failure modes. To be more specific, the studies covered the finite element analysis-based fatigue life prediction and accelerated durability testing of the tilting disc heart valves for nine different valve sizes. First, finite element analysis-based fatigue life prediction showed that all nine valve sizes were in the infinite life region. Second, accelerated durability test showed that all nine valve sizes remained functional for 400 million cycles under experimental conditions. The study ensures the continued function of TC2 model tilting disc heart valves over duration in excess of 50 years. The results imply that the TC2 model valve designs are structurally safe, reliable and durable.
Mitral valve prolapse and hyperthyroidism: effect of patient selection.
Zullo, M A; Devereux, R B; Kramer-Fox, R; Lutas, E M; Brown, W T
1985-11-01
Patients with mitral valve prolapse and hyperthyroidism have common symptoms; the most outstanding symptom is palpitation. To determine whether or not common symptoms contributed to the reported association of these conditions, we evaluated 220 patients with symptomatic mitral valve prolapse and 216 first-degree relatives in 72 families; 65 relatives with mitral valve prolapse and 151 relatives without mitral valve prolapse, all greater than or equal to 16 years of age. Thirty subjects, aged 49 +/- 13 years (p less than 0.025 vs entire study group), had thyroid disease (23 subjects had definite thyroid disease, seven subjects had probable); 27 of 30 subjects with thyroid disease (90%) were female (p less than 0.005). The age- and sex-adjusted prevalence of hyperthyroidism was significantly higher in probands with mitral valve prolapse than in family members without mitral valve prolapse (3.5% vs 0%, p = 0.03), while an intermediate prevalence of hyperthyroidism (2.2%) was observed in family members with mitral valve prolapse. Thus, the prevalence of hyperthyroidism is increased among symptomatic patients with mitral valve prolapse as compared to family members without mitral valve prolapse, but the prevalence of thyroid conditions is similar among family members with or without this condition. These findings are explained by the effect of common symptoms on clinical detection of both mitral valve prolapse and hyperthyroidism.
... called anticoagulants) for the rest of their lives. Biological valves are made from animal tissue (called a ... for valve replacement (called an autograft). Patients with biological valves usually do not need to take blood- ...
46 CFR 78.17-5 - Valves and closing appliances.
Code of Federal Regulations, 2010 CFR
2010-10-01
... 46 Shipping 3 2010-10-01 2010-10-01 false Valves and closing appliances. 78.17-5 Section 78.17-5..., Drills, and Inspections § 78.17-5 Valves and closing appliances. (a) It shall be the duty of the master to see that all valves, including cross connecting valves where fitted, and other appliances such as...
40 CFR 61.242-6 - Standards: Open-ended valves or lines.
Code of Federal Regulations, 2010 CFR
2010-07-01
... for Equipment Leaks (Fugitive Emission Sources) § 61.242-6 Standards: Open-ended valves or lines. (a)(1) Each open-ended valve or line shall be equipped with a cap, blind flange, plug, or a second valve... 40 Protection of Environment 8 2010-07-01 2010-07-01 false Standards: Open-ended valves or lines...
Spool-Valve Pressure-Difference Regulator
NASA Technical Reports Server (NTRS)
Grasso, A. P.
1983-01-01
Valves maintain preset pressure difference between gas flows. Two spool valves connected by shaft move back and forth in response to changes in pressure in oxygen and hydrogen chambers. Spool-valve assembly acts to restore pressures to preset difference. By eliminating diaphragms, pressure exerted directly on external end of spool valve; however, forces and therefore sensitivity of assembly are reduced.
Mitral valve surgery - minimally invasive
... flow. Your valve has developed an infection (infectious endocarditis). You have severe mitral valve prolapse that is ... function. Damage to your heart valve from infection (endocarditis). A minimally invasive procedure has many benefits. There ...
Code of Federal Regulations, 2010 CFR
2010-10-01
... PIPELINE Design Requirements § 195.116 Valves. Each valve installed in a pipeline system must comply with the following: (a) The valve must be of a sound engineering design. (b) Materials subject to the...
Mechanism of valvular regurgitation.
Khoo, Nee S; Smallhorn, Jeffery F
2011-10-01
Despite improvements in surgical techniques, valvular regurgitation results in major morbidity in children with heart disease. Functional anatomy, mechanisms of valve closure and adaptation to changing hemodynamic stress in normal mitral and tricuspid valves are complex and only partially understood. As well, pathology of atrioventricular valve regurgitation is further complicated by congenital valve abnormalities involving leaflet tissue, supporting chordal apparatus and displaced papillary muscles. This review provides a current understanding of the mechanisms that result in atrioventricular valve failure. Mitral valve leaflets have contractile elements, in addition to atrial muscle modulation of leaflet tension. When placed under mechanical tethering stress, the mitral valve adapts by leaflet expansion, which increases coaptation surface reserve and chordal thickening. Both pediatric and adult studies are increasingly reporting on the importance of subvalvar apparatus function in maintaining valve competency. The maintenance of efficient valve function is accomplished by a complex series of events involving atrial and annular contraction, annular deformation, active leaflet tension, chordal transmission of papillary muscle contractions and ventricular contraction.