Sample records for ventilation flow rate

  1. Evaluation of a turbine flow meter (Ventilometer Mark 2) in the measurement of ventilation.

    PubMed

    Cooper, C B; Harris, N D; Howard, P

    1990-01-01

    We have evaluated a turbine flow meter (Ventilometer Mark 2, PK Morgan, Kent, UK) at low flow rates and levels of ventilation which are likely to be encountered during exercise in patients with chronic respiratory disease. Pulsatile flows were generated from a volume-cycled mechanical ventilator, the flow wave-form was modified by damping to simulate a human breathing pattern. Comparative measurements of ventilation were made whilst varying tidal volume (VT) from 0.22 to 1.131 and respiratory rate (fR) from 10 to 35 min-1. At lower levels of ventilation the instrument tended to underread especially with increasing fR. The calibration factor must be adjusted to match the level of ventilation if the measurement errors are to be within 5%.

  2. Fuselage ventilation due to wind flow about a postcrash aircraft

    NASA Technical Reports Server (NTRS)

    Stuart, J. W.

    1980-01-01

    Postcrash aircraft fuselage fire development, dependent on the internal and external fluid dynamics is discussed. The natural ventilation rate, a major factor in the internal flow patterns and fire development is reviewed. The flow about the fuselage as affected by the wind and external fire is studied. An analysis was performend which estimated the rates of ventilation produced by the wind for a limited idealized environmental configuration. The simulation utilizes the empirical pressure coefficient distribution of an infinite circular cylinder near a wall with its boundary later flow to represent the atmospheric boundary layer. The resulting maximum ventilation rate for two door size openings, with varying circumferential location in a common 10 mph wind was an order of magnitude greater than the forced ventilation specified in full scale fire testing. The parameter discussed are: (1) fuselage size and shape, (2) fuselage orientation and proximity to the ground, (3) fuselage-openings size and location, (4) wind speed and direction, and (5) induced flow of the external fire plume is recommended. The fire testing should be conducted to a maximum ventilation rate at least an order of magnitude greater than the inflight air conditioning rates.

  3. Impact of High-flow Nasal Cannula Therapy in Quality Improvement and Clinical Outcomes in a Non-invasive Ventilation Device-free Pediatric Intensive Care Unit.

    PubMed

    Can, Fulva Kamit; Anil, Ayse Berna; Anil, Murat; Zengin, Neslihan; Bal, Alkan; Bicilioglu, Yuksel; Gokalp, Gamze; Durak, Fatih; Ince, Gulberat

    2017-10-15

    To analyze the change in quality indicators due to the use of high-flow nasal cannula therapy as a non-invasive ventilation method in children with respiratory distress/failure in a non-invasive ventilation device-free pediatric intensive care unit. Retrospective chart review of children with respiratory distress/failure admitted 1 year before (period before high-flow nasal cannula therapy) and 1 year after (period after high-flow nasal cannula therapy) the introduction of high-flow nasal cannula therapy. We compared quality indicators as rate of mechanical ventilation, total duration of mechanical ventilation, rate of re-intubation, pediatric intensive care unit length of stay, and mortality rate between these periods. Between November 2012 and November 2014, 272 patients: 141 before and 131 after high-flow nasal cannula therapy were reviewed (median age was 20.5 mo). Of the patients in the severe respiratory distress/failure subgroup, the rate of intubation was significantly lower in period after than in period before high-flow nasal cannula therapy group (58.1% vs. 76.1%; P <0.05). The median pediatric intensive care unit length of stay was significantly shorter in patients who did not require mechanical ventilation in the period after than in the period before high-flow nasal cannula therapy group (3d vs. 4d; P<0,05). Implementation of high-flow nasal cannula therapy in pediatric intensive care unit significantly improves the quality of therapy and its outcomes.

  4. Proportional mechanical ventilation through PWM driven on/off solenoid valve.

    PubMed

    Sardellitti, I; Cecchini, S; Silvestri, S; Caldwell, D G

    2010-01-01

    Proportional strategies for artificial ventilation are the most recent form of synchronized partial ventilatory assistance and intra-breath control techniques available in clinical practice. Currently, the majority of commercial ventilators allowing proportional ventilation uses proportional valves to generate the flow rate pattern. This paper proposes on-off solenoid valves for proportional ventilation given their small size, low cost and short switching time, useful for supplying high frequency ventilation. A new system based on a novel fast switching driver circuit combined with on/off solenoid valve is developed. The average short response time typical of onoff solenoid valves was further reduced through the driving circuit for the implementation of PWM control. Experimental trials were conducted for identifying the dynamic response of the PWM driven on/off valve and for verifying its effectiveness in generating variable-shaped ventilatory flow rate patterns. The system was able to smoothly follow the reference flow rate patterns also changing in time intervals as short as 20 ms, achieving a flow rate resolution up to 1 L/min and repeatability in the order of 0.5 L/min. Preliminary results showed the feasibility of developing a stand alone portable device able to generate both proportional and high frequency ventilation by only using on-off solenoid valves.

  5. Significant Improvements in Pyranometer Nighttime Offsets Using High-Flow DC Ventilation

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Michalsky, Joseph J.; Kutchenreiter, Mark; Long, Charles N.

    Ventilators are used to keep the domes of pyranometers clean and dry, but they affect the nighttime offset as well. This paper examines different ventilation strategies. For the several commercial single-black-detector pyranometers with ventilators examined here, high flow rate (50 CFM and higher), 12 VDC fans lower the offsets, lower the scatter, and improve the predictability of the offsets during the night compared with lower flow rate 35 CFM, 120 VAC fans operated in the same ventilator housings. Black-and-white pyranometers sometimes show improvement with DC ventilation, but in some cases DC ventilation makes the offsets slightly worse. Since the offsetsmore » for these black-and-white pyranometers are always small, usually no more than 1 Wm -2, whether AC or DC ventilated, changing their ventilation to higher CFM DC ventilation is not imperative. Future work should include all major manufacturers of pyranometers and unventilated, as well as, ventilated pyranometers. Lastly, an important outcome of future research will be to clarify under what circumstances nighttime data can be used to predict daytime offsets.« less

  6. Significant Improvements in Pyranometer Nighttime Offsets Using High-Flow DC Ventilation

    DOE PAGES

    Michalsky, Joseph J.; Kutchenreiter, Mark; Long, Charles N.

    2017-06-20

    Ventilators are used to keep the domes of pyranometers clean and dry, but they affect the nighttime offset as well. This paper examines different ventilation strategies. For the several commercial single-black-detector pyranometers with ventilators examined here, high flow rate (50 CFM and higher), 12 VDC fans lower the offsets, lower the scatter, and improve the predictability of the offsets during the night compared with lower flow rate 35 CFM, 120 VAC fans operated in the same ventilator housings. Black-and-white pyranometers sometimes show improvement with DC ventilation, but in some cases DC ventilation makes the offsets slightly worse. Since the offsetsmore » for these black-and-white pyranometers are always small, usually no more than 1 Wm -2, whether AC or DC ventilated, changing their ventilation to higher CFM DC ventilation is not imperative. Future work should include all major manufacturers of pyranometers and unventilated, as well as, ventilated pyranometers. Lastly, an important outcome of future research will be to clarify under what circumstances nighttime data can be used to predict daytime offsets.« less

  7. Barotrauma and microvascular injury in lungs of nonadult rabbits: effect of ventilation pattern.

    PubMed

    Peevy, K J; Hernandez, L A; Moise, A A; Parker, J C

    1990-06-01

    To study the pulmonary microvascular injury produced by ventilation barotrauma, the isolated perfused lungs of 4 to 6-wk-old New Zealand white rabbits were ventilated by one of the following methods: peak inspiratory pressure (PIP) 23 cm H2O, gas flow rate 1.1 L/min (group 1); PIP 27 cm H2O, gas flow rate 6.9 L/min (group 2); PIP 50 cm H2O, gas flow rate 1.9 L/min (group 3); or PIP 53 cm H2O, gas flow rate 8.3 L/min (group 4). Microvascular permeability was assessed using the capillary filtration coefficient (Kfc) before and 5, 30, and 60 min after a 15-min period of ventilation. Baseline Kfc was not significantly different between groups. A significant increase over the baseline Kfc was noted at 60 min in group 2 and in all postventilation Kfc values in groups 3 and 4 (p less than .05). Group 1 Kfc values did not change significantly after ventilation. At all post-ventilation times, values for Kfc were significantly greater in groups 3 and 4 than in group 1 (p less than .05). Group 4 Kfc values were significantly greater than those in group 2 at 5 and 30 min postventilation. These data indicate that high PIP, and to a lesser extent, high gas flow rates cause microvascular injury in the compliant nonadult lung and suggest that the combination of high PIP and high gas flow rates are the most threatening to microvascular integrity.

  8. An advanced expiratory circuit for the recovery of perfluorocarbon liquid from non-saturated perfluorocarbon vapour during partial liquid ventilation: an experimental model

    PubMed Central

    Dunster, Kimble R; Davies, Mark W; Fraser, John F

    2006-01-01

    Background The loss of perfluorocarbon (PFC) vapour in the expired gases during partial liquid ventilation should be minimized both to prevent perfluorocarbon vapour entering the atmosphere and to re-use the recovered PFC liquid. Using a substantially modified design of our previously described condenser, we aimed to determine how much perfluorocarbon liquid could be recovered from gases containing PFC and water vapour, at concentrations found during partial liquid ventilation, and to determine if the amount recovered differed with background flow rate (at flow rates suitable for use in neonates). Methods The expiratory line of a standard ventilator circuit set-up was mimicked, with the addition of two condensers. Perfluorocarbon (30 mL of FC-77) and water vapour, at concentrations found during partial liquid ventilation, were passed through the circuit at a number of flow rates and the percentage recovery of the liquids measured. Results From 14.2 mL (47%) to 27.3 mL (91%) of the infused 30 mL of FC-77 was recovered at the flow rates studied. Significantly higher FC-77 recovery was obtained at lower flow rates (ANOVA with Bonferroni's multiple comparison test, p < 0.0001). As a percentage of the theoretical maximum recovery, 64 to 95% of the FC-77 was recovered. Statistically significantly less FC-77 was recovered at 5 Lmin-1 (ANOVA with Bonferroni's multiple comparison test, p < 0.0001). Amounts of perfluorocarbon vapour recovered were 47%, 50%, 81% and 91% at flow rates of 10, 5, 2 and 1 Lmin-1, respectively. Conclusion Using two condensers in series 47% to 91% of perfluorocarbon liquid can be recovered, from gases containing perfluorocarbon and water vapour, at concentrations found during partial liquid ventilation. PMID:16457722

  9. Significant Improvements in Pyranometer Nighttime Offsets Using High-Flow DC Ventilation

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Kutchenreiter, Mark; Michalski, J.J.; Long, C.N.

    2017-05-22

    Accurate solar radiation measurements using pyranometers are required to understand radiative impacts on the Earth's energy budget, solar energy production, and to validate radiative transfer models. Ventilators of pyranometers, which are used to keep the domes clean and dry, also affect instrument thermal offset accuracy. This poster presents a high-level overview of the ventilators for single-black-detector pyranometers and black-and-white pyranometers. For single-black-detector pyranometers with ventilators, high-flow-rate (50-CFM and higher), 12-V DC fans lower the offsets, lower the scatter, and improve the predictability of nighttime offsets compared to lower-flow-rate (35-CFM), 120-V AC fans operated in the same type of environmental setup.more » Black-and-white pyranometers, which are used to measure diffuse horizontal irradiance, sometimes show minor improvement with DC fan ventilation, but their offsets are always small, usually no more than 1 W/m2, whether AC- or DC-ventilated.« less

  10. CFD Approach To Investigate The Flow Characteristics In Bi-Directional Ventilated Disc Brake

    NASA Astrophysics Data System (ADS)

    Munisamy, Kannan M.; Yusoff, Mohd. Zamri; Shuaib, Norshah Hafeez; Thangaraju, Savithry K.

    2010-06-01

    This paper presents experimental and Computational Fluids Dynamics (CFD) investigations of the flow in ventilated brake discs. Development of an experiment rig with basic measuring devices are detailed out and following a validation study, the possible improvement in the brake cooling can be further analyzed using CFD analysis. The mass flow rate is determined from basic flow measurement technique following that the conventional bi-directional passenger car is simulated using commercial CFD software FLUENT™. The CFD simulation is used to investigate the flow characteristics in between blade flow of the bi-directional ventilated disc brake.

  11. Decline of the performance of a portable axial-flow fan due to the friction and duct bending loss of a connected flexible duct.

    PubMed

    Ojima, Jun

    2017-03-28

    In a job site, a portable fan is often used to ventilate a confined space. When a portable fan is applied to such a space, the actual ventilation flow rate must be accurately estimated in advance because the safety level of contaminant and oxygen concentrations in the space will determine the ventilation requirements. When a portable fan is used with a flexible duct, the actual flow rate of the fan decreases due to the friction and duct bending loss of the duct. Intending to show the decline of a fan performance, the author conducted laboratory experiments and reported the quantitative effect of the friction and duct bending loss of a flexible duct to the flow rate of a portable fan. Four commercial portable fans of different specifications were procured for the experiments, and the decline of the performance of each portable fan due to the friction loss etc. of a connected flexible duct was investigated by measuring actual flow rate. The flow rate showed an obvious decrease from the rated flow rate when a flexible duct was connected. Connection of a straight polyester flexible duct and a straight aluminum flexible duct reduced the flow rates to 81.2 - 52.9% and less than 50%, respectively. The flow rate decreased with an increase of the bend angle of the flexible duct. It is recommended that flow rate check of a portable fan should be diligently carried out in every job site.

  12. Increased ventilation by fish leads to a higher risk of parasitism.

    PubMed

    Mikheev, Victor N; Pasternak, Anna F; Valtonen, E Tellervo; Taskinen, Jouni

    2014-06-23

    Fish are common intermediate hosts of trematode cercariae and their gills can potentially serve as important sites of penetration by these larval stages. We experimentally tested the hypothesis that volume of ventilation flow across the gills contributes to acquisition of these parasites by fish. We manipulated the intensity of ventilation by using different oxygen concentrations. Juvenile Oncorhynchus mykiss were individually exposed for 10 minutes to a standard dose of Diplostomum pseudospathaceum cercariae at three levels of oxygen concentration, 30, 60 and 90%. Ventilation amplitude (measured as a distance between left and right operculum), operculum beat rate, and the number of cercariae established in the eyes of fish were recorded. Fish reacted to low oxygen concentration with wider expansion of opercula (but not with increasing beat rate), leading to an increase in ventilation volume. As expected, the intensity of infection increased with decreasing oxygen saturation-probably due to a higher exposure to cercariae caused by increased ventilation under low oxygen concentrations. The number of cercariae acquired by an individual fish was positively correlated with ventilation amplitude and with ventilation volume, but not with operculum beat rate. However, even though the infection rate increased under these circumstances, the proportion of larval trematodes successfully establishing in fish eyes decreased with increasing ventilation volume, suggesting that the high flow velocity, although increasing host exposure to cercarial parasites, may interfere with the ability of these parasites to penetrate their hosts. There was no difference in the behaviour of trematode cercariae exposed to low and high oxygen concentrations. A reduction in oxygen saturation resulted in an increase in ventilation volume across the gills and in doing so an increase in the exposure of fish to cercariae. A significant correlation between ventilation volume and parasitism represents the first experimental evidence that this physiological mechanism generates variation in transmission of parasites to fish hosts. Other factors that modify ventilation flow, e.g. physiological or social stressors, are expected to produce similar effects on the transmission success of the parasites penetrating fish hosts using the gills.

  13. Increased ventilation by fish leads to a higher risk of parasitism

    PubMed Central

    2014-01-01

    Background Fish are common intermediate hosts of trematode cercariae and their gills can potentially serve as important sites of penetration by these larval stages. We experimentally tested the hypothesis that volume of ventilation flow across the gills contributes to acquisition of these parasites by fish. We manipulated the intensity of ventilation by using different oxygen concentrations. Methods Juvenile Oncorhynchus mykiss were individually exposed for 10 minutes to a standard dose of Diplostomum pseudospathaceum cercariae at three levels of oxygen concentration, 30, 60 and 90%. Ventilation amplitude (measured as a distance between left and right operculum), operculum beat rate, and the number of cercariae established in the eyes of fish were recorded. Results Fish reacted to low oxygen concentration with wider expansion of opercula (but not with increasing beat rate), leading to an increase in ventilation volume. As expected, the intensity of infection increased with decreasing oxygen saturation—probably due to a higher exposure to cercariae caused by increased ventilation under low oxygen concentrations. The number of cercariae acquired by an individual fish was positively correlated with ventilation amplitude and with ventilation volume, but not with operculum beat rate. However, even though the infection rate increased under these circumstances, the proportion of larval trematodes successfully establishing in fish eyes decreased with increasing ventilation volume, suggesting that the high flow velocity, although increasing host exposure to cercarial parasites, may interfere with the ability of these parasites to penetrate their hosts. There was no difference in the behaviour of trematode cercariae exposed to low and high oxygen concentrations. Conclusion A reduction in oxygen saturation resulted in an increase in ventilation volume across the gills and in doing so an increase in the exposure of fish to cercariae. A significant correlation between ventilation volume and parasitism represents the first experimental evidence that this physiological mechanism generates variation in transmission of parasites to fish hosts. Other factors that modify ventilation flow, e.g. physiological or social stressors, are expected to produce similar effects on the transmission success of the parasites penetrating fish hosts using the gills. PMID:24954703

  14. Ventilation through a small-bore catheter: optimizing expiratory ventilation assistance.

    PubMed

    Hamaekers, A E W; Borg, P A J; Götz, T; Enk, D

    2011-03-01

    Emergency ventilation through a small-bore transtracheal catheter can be lifesaving in a 'cannot intubate, cannot ventilate' situation. Ejectors, capable of creating suction by the Bernoulli principle, have been proposed to facilitate expiration through small-bore catheters. In this bench study, we compared a novel, purpose-built ventilation ejector (DE 5) with a previously proposed, modified industrial ejector (SBP 07). The generated insufflation pressures, suction pressures in static and dynamic situations, and also suction capacities and entrainment ratios of the SBP 07 and the DE 5 were determined. The DE 5 was also tested in a lung simulator with a simulated complete upper airway obstruction. Inspiratory and expiratory times through a transtracheal catheter were measured at various flow rates and achievable minute volumes were calculated. In a static situation, the SBP 07 showed a more negative pressure build-up compared with the DE 5. However, in a dynamic situation, the DE 5 generated a more negative pressure, resulting in a higher suction capacity. Employment of the DE 5 at a flow rate of 18 litre min(-1) allowed a minute volume through the transtracheal catheter of up to 8.27 litre min(-1) at a compliance of 100 ml cm H(2)O(-1). The efficiency of the DE 5 depended on the flow rate of the driving gas and the compliance of the lung simulator. In laboratory tests, the DE 5 is an optimized ventilation ejector suitable for applying expiratory ventilation assistance. Further research may confirm the clinical applicability as a portable emergency ventilator for use with small-bore catheters.

  15. Ventilation and infiltration in high-rise apartment buildings

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Diamond, R.C.; Feustel, H.E.; Dickerhoff, D.J.

    1996-03-01

    Air flow, air leakage measurements and numerical simulations were made on a 13-story apartment building to characterize the ventilation rates for the individual apartments. Parametric simulations were performed for specific conditions, e.g., height, orientation, outside temperature and wind speed. Our analysis of the air flow simulations suggest that the ventilation to the individual units varies considerably. With the mechanical ventilation system disabled and no wind, units at the lower level of the building have adequate ventilation only on days with high temperature differences, while units on higher floors have no ventilation at all. Units facing the windward side will bemore » over-ventilated when the building experiences wind directions between west and north. At the same time, leeward apartments did not experience any fresh air-because, in these cases, air flows enter the apartments from the corridor and exit through the exhaust shafts and the cracks in the facade. Even with the mechanical ventilation system operating, we found wide variation in the air flows to the individual apartments. In addition to the specific case presented here, these findings have more general implications for energy retrofits and health and comfort of occupants in high-rise apartment buildings.« less

  16. A novel fiber-optic measurement system for the evaluation of performances of neonatal pulmonary ventilators

    NASA Astrophysics Data System (ADS)

    Battista, L.; Scorza, A.; Botta, F.; Sciuto, S. A.

    2016-02-01

    Published standards for the performance evaluation of pulmonary ventilators are mainly directed to manufacturers rather than to end-users and often considered inadequate or not comprehensive. In order to contribute to overcome the problems above, a novel measurement system was proposed and tested with waveforms of mechanical ventilation by means of experimental trials carried out with infant ventilators typically used in neonatal intensive care units: the main quantities of mechanical ventilation in newborns are monitored, i.e. air flow rate, differential pressure and volume from infant ventilator are measured by means of two novel fiber-optic sensors (OFSs) developed and characterized by the authors, while temperature and relative humidity of air mass are obtained by two commercial transducers. The proposed fiber-optic sensors (flow sensor Q-OFS, pressure sensor P-OFS) showed measurement ranges of air flow and pressure typically encountered in neonatal mechanical ventilation, i.e. the air flow rate Q ranged from 3 l min-1 to 18 l min-1 (inspiratory) and from  -3 l min-1 to  -18 l min-1 (expiratory), the differential pressure ΔP ranged from  -15 cmH2O to 15 cmH2O. In each experimental trial carried out with different settings of the ventilator, outputs of the OFSs are compared with data from two reference sensors (reference flow sensor RF, reference pressure sensor RP) and results are found consistent: flow rate Q showed a maximum error between Q-OFS and RF up to 13 percent, with an output ratio Q RF/Q OFS of not more than 1.06  ±  0.09 (least square estimation, 95 percent confidence level, R 2 between 0.9822 and 0.9931). On the other hand the maximum error between P-OFS and RP on differential pressure ΔP was lower than 10 percent, with an output ratio ΔP RP/ΔP OFS between 0.977  ±  0.022 and 1.0  ±  0.8 (least square estimation, 95 percent confidence level, R 2 between 0.9864 and 0.9876). Despite the possible improvements, results were encouraging and suggested the proposed measurement system can be considered suitable for performances evaluation of neonatal ventilators and useful for both end-users and manufacturers.

  17. Technical aspects and clinical implications of high frequency jet ventilation with a solenoid valve.

    PubMed

    Carlon, G C; Miodownik, S; Ray, C; Kahn, R C

    1981-01-01

    High frequency jet ventilation (HFJV) is an incompletely studied technique of mechanical respiratory support. The authors have built a ventilator based on a solenoid valve, that allows independent selection of respiratory rate and inspiratory/expiratory ratio. The ventilator can be synchronized to the heart rate. Humidification is provided by warm saline dripped in front of the injector nozzle, so that the jet stream itself acts as a nebulizer. Tube diameter, length, and deformability are fundamental determinants of inspiratory flow rate and wave form. Cannula kinking and inadequate humidification were the most significant sources of complications.

  18. 40 CFR 98.323 - Calculating GHG emissions.

    Code of Federal Regulations, 2014 CFR

    2014-07-01

    ... systems (metric tons CH4). CH4V = Quarterly CH4 liberated from each ventilation monitoring point (metric... vent holes are collected, you must calculate the quarterly CH4 liberated from the ventilation system... CH4 liberated from a ventilation monitoring point (metric tons CH4). V = Volumetric flow rate for the...

  19. The influence of broiler activity, growth rate, and litter on carbon dioxide balances for the determination of ventilation flow rates in broiler production.

    PubMed

    Calvet, S; Estellés, F; Cambra-López, M; Torres, A G; Van den Weghe, H F A

    2011-11-01

    Carbon dioxide balances are useful in determining ventilation rates in livestock buildings. These balances need an accurate estimation of the CO(2) produced by animals and their litter to determine the ventilation flows. To estimate the daily variation in ventilation flow, it is necessary to precisely know the daily variation pattern of CO(2) production, which mainly depends on animal activity. The objective of this study was to explore the applicability of CO(2) balances for determining ventilation flows in broiler buildings. More specifically, this work aimed to quantify the amount of CO(2) produced by the litter, as well as the amount of CO(2) produced by the broilers, as a function of productive parameters, and to analyze the influence of broiler activity on CO(2) emissions. Gas concentrations and ventilation flows were simultaneously measured in 3 trials, with 1 under experimental conditions and the other 2 in a commercial broiler farm. In the experimental assay, broiler activity was also determined. At the end of the experimental trial, on the day after the removal of the broilers, the litter accounted for 20% of the total CO(2) produced, and the broilers produced 3.71 L/h of CO(2) per kg of metabolic weight. On the commercial farm, CO(2) production was the same for the 2 cycles (2.60 L/h per kg of metabolic weight, P > 0.05). However, substantial differences were found between CO(2) and broiler activity patterns after changes in light status. A regression model was used to explain these differences (R(2) = 0.52). Carbon dioxide increased with bird activity, being on average 3.02 L/h per kg of metabolic weight for inactive birds and 4.73 L/h per kg of metabolic weight when bird activity was highest. Overall, CO(2) balances are robust tools for determining the daily average ventilation flows in broiler farms. These balances could also be applied at more frequent intervals, but in this case, particular care is necessary after light status changes because of discrepancy between animal activity and CO(2) production.

  20. Control of airborne infectious diseases in ventilated spaces

    PubMed Central

    Nielsen, Peter V.

    2009-01-01

    We protect ourselves from airborne cross-infection in the indoor environment by supplying fresh air to a room by natural or mechanical ventilation. The air is distributed in the room according to different principles: mixing ventilation, displacement ventilation, etc. A large amount of air is supplied to the room to ensure a dilution of airborne infection. Analyses of the flow in the room show that there are a number of parameters that play an important role in minimizing airborne cross-infection. The air flow rate to the room must be high, and the air distribution pattern can be designed to have high ventilation effectiveness. Furthermore, personalized ventilation may reduce the risk of cross-infection, and in some cases, it can also reduce the source of infection. Personalized ventilation can especially be used in hospital wards, aircraft cabins and, in general, where people are in fixed positions. PMID:19740921

  1. Effect of accessory ostia on maxillary sinus ventilation: a computational fluid dynamics (CFD) study.

    PubMed

    Zhu, Jian Hua; Lee, Heow Pueh; Lim, Kian Meng; Gordon, Bruce R; Wang, De Yun

    2012-08-15

    We evaluated, by CFD simulation, effects of accessory ostium (AO) on maxillary sinus ventilation. A three-dimensional nasal model was constructed from an adult CT scan with two left maxillary AOs (sinus I) and one right AO (sinus II), then compared to an identical control model with all AOs sealed (sinuses III and IV). Transient simulations of quiet inspiration and expiration at 15 L/min, and nasal blow at 48 L/min, were calculated for both models using low-Reynolds-number turbulent analysis. At low flows, ventilation rates in sinuses with AOs (I ≈ 0.46 L/min, II ≈ 0.54 L/min), were both more than a magnitude higher than sinuses without AOs (II I ≈ 0.019 L/min, IV ≈ 0.020 L/min). Absence of AO almost completely prevented sinus ventilation. Increased ventilation of sinuses with AOs is complex. Under high flow conditions mimicking nose blowing, in sinuses II, III, and IV, the sinus flow rate increased. In contrast, the airflow direction through sinus I reversed between inspiration and expiration, while it remained almost constant throughout the respiration cycle in sinus II. CFD simulation demonstrated that AOs markedly increase maxillary sinus airflow rates and alter sinus air circulation patterns. Whether these airflow changes impact maxillary sinus physiology or pathophysiology is unknown. Copyright © 2012 Elsevier B.V. All rights reserved.

  2. Inside anesthesia breathing circuits: time to reach a set sevoflurane concentration in toddlers and newborns: simulation using a test lung.

    PubMed

    Kern, Delphine; Larcher, Claire; Basset, Bertrand; Alacoque, Xavier; Fesseau, Rose; Samii, Kamran; Minville, Vincent; Fourcade, Olivier

    2012-08-01

    We measured the time it takes to reach the desired inspired anesthetic concentration using the Primus (Drägerwerk, AG, Lübeck, Germany) and the Avance (GE Datex-Ohmeda, Munich, Germany) anesthesia machines with toddler and newborn ventilation settings. The time to reach 95% of inspired target sevoflurane concentration was measured during wash-in from 0 to 6 vol% sevoflurane and during wash-out from 6 to 0 vol% with fresh gas flows equal to 1 and 2 times the minute ventilation. The Avance was faster than the Primus (65 seconds [95% confidence interval (CI): 55 to 78] vs 310 seconds [95% CI: 261 to 359]) at 1.5 L/min fresh gas flow, tidal volume of 50 mL, and 30 breaths/min. Times were shorter by the same magnitude at higher fresh gas flows and higher minute ventilation rates. The effect of doubling fresh gas flow was variable and less than expected. The Primus is slower during newborn than toddler ventilation, whereas the Avance's response time was the same for newborn and toddler ventilation. Our data confirm that the time to reach the target-inspired anesthetic concentration depends on breathing circuit volume, fresh gas flow, and minute ventilation.

  3. Regenerative Blower for EVA Suit Ventilation Fan

    NASA Technical Reports Server (NTRS)

    Izenson, Michael G.; Chen, Weibo; Paul, Heather L.

    2010-01-01

    Portable life support systems in future space suits will include a ventilation subsystem driven by a dedicated fan. This ventilation fan must meet challenging requirements for pressure rise, flow rate, efficiency, size, safety, and reliability. This paper describes research and development that showed the feasibility of a regenerative blower that is uniquely suited to meet these requirements. We proved feasibility through component tests, blower tests, and design analysis. Based on the requirements for the Constellation Space Suit Element (CSSE) Portable Life Support System (PLSS) ventilation fan, we designed the critical elements of the blower. We measured the effects of key design parameters on blower performance using separate effects tests, and used the results of these tests to design a regenerative blower that will meet the ventilation fan requirements. We assembled a proof-of-concept blower and measured its performance at sub-atmospheric pressures that simulate a PLSS ventilation loop environment. Head/flow performance and maximum efficiency point data were used to specify the design and operating conditions for the ventilation fan. We identified materials for the blower that will enhance safety for operation in a lunar environment, and produced a solid model that illustrates the final design. The proof-of-concept blower produced the flow rate and pressure rise needed for the CSSE ventilation subsystem while running at 5400 rpm, consuming only 9 W of electric power using a non-optimized, commercial motor and controller and inefficient bearings. Scaling the test results to a complete design shows that a lightweight, compact, reliable, and low power regenerative blower can meet the performance requirements for future space suit life support systems.

  4. Evaluation of design ventilation requirements for enclosed parking facilities

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ayari, A.; Krarti, M.

    2000-07-01

    This paper proposes a new design approach to determine the ventilation requirements for enclosed parking garages. The design approach accounts for various factors that affect the indoor air quality within a parking facility, including the average CO emission rate, the average travel time, the number of cars, and the acceptable CO level within the parking garage. This paper first describes the results of a parametric analysis based on the design method that was developed. Then the design method is presented to explain how the ventilation flow rate can be determined for any enclosed parking facility. Finally, some suggestions are proposedmore » to save fan energy for ventilating parking garages using demand ventilation control strategies.« less

  5. Analysis of the velocity distribution in different types of ventilation system ducts

    NASA Astrophysics Data System (ADS)

    Peszyński, Kazimierz; Olszewski, Lukasz; Smyk, Emil; Perczyński, Daniel

    2018-06-01

    The paper presents the results obtained during the preliminary studies of circular and rectangular ducts before testing the properties elements (elbows, tees, etc.)of rectangular with rounded corners ducts. The fundamental problem of the studies was to determine the flow rate in the ventilation duct. Due to the size of the channel it was decided to determine the flow rate based on the integration of flow velocity over the considered cross-section. This method requires knowledge of the velocity distribution in the cross section. Approximation of the measured actual profile by the classic and modified Prandtl power-law velocity profile was analysed.

  6. City ventilation of Hong Kong at no-wind conditions

    NASA Astrophysics Data System (ADS)

    Yang, Lina; Li, Yuguo

    We hypothesize that city ventilation due to both thermally-driven mountain slope flows and building surface flows is important in removing ambient airborne pollutants in the high-rise dense city Hong Kong at no-wind conditions. Both spatial and temporal urban surface temperature profiles are an important boundary condition for studying city ventilation by thermal buoyancy. Field measurements were carried out to investigate the diurnal thermal behavior of urban surfaces (mountain slopes, and building exterior walls and roofs) in Hong Kong by using the infrared thermography. The maximum urban surface temperature was measured in the early noon hours (14:00-15:00 h) and the minimum temperature was observed just before sunrise (5:00 h). The vertical surface temperature of the building exterior wall was found to increase with height at daytime and the opposite occurred at nighttime. The solar radiation and the physical properties of the various urban surfaces were found to be important factors affecting the surface thermal behaviors. The temperature difference between the measured maximum and minimum surface temperatures of the four selected exterior walls can be at the highest of 16.7 °C in the early afternoon hours (15:00 h). Based on the measured surface temperatures, the ventilation rate due to thermal buoyancy-induced wall surface flows of buildings and mountain slope winds were estimated through an integral analysis of the natural convection flow over a flat surface. At no-wind conditions, the total air change rate by the building wall flows (2-4 ACH) was found to be 2-4 times greater than that by the slope flows due to mountain surface (1 ACH) due to larger building exterior surface areas and temperature differences with surrounding air. The results provide useful insights into the ventilation of a high-rise dense city at no-wind conditions.

  7. Field evaluation of ventilation system performance in enclosed parking garages

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Ayari, A.M.; Grot, D.A.; Krarti, M.

    2000-07-01

    This paper summarizes the results of a field study to determine the ventilation requirements and the contaminant levels in existing enclosed parking garages. The testing was conducted in seven parking garages with different sizes, traffic flow patterns, vehicle types, and locations. In particular, the study compares the actual ventilation rates measured using the tracer gas technique with the ventilation requirements of ANSI/ASHRAE Standard 62-1989. In addition, the field test evaluated the effectiveness of the existing ventilation systems in maintaining acceptable contaminant levels within enclosed parking garages.

  8. Flow measurement in mechanical ventilation: a review.

    PubMed

    Schena, Emiliano; Massaroni, Carlo; Saccomandi, Paola; Cecchini, Stefano

    2015-03-01

    Accurate monitoring of flow rate and volume exchanges is essential to minimize ventilator-induced lung injury. Mechanical ventilators employ flowmeters to estimate the amount of gases delivered to patients and use the flow signal as a feedback to adjust the desired amount of gas to be delivered. Since flowmeters play a crucial role in this field, they are required to fulfill strict criteria in terms of dynamic and static characteristics. Therefore, mechanical ventilators are equipped with only the following kinds of flowmeters: linear pneumotachographs, fixed and variable orifice meters, hot wire anemometers, and ultrasonic flowmeters. This paper provides an overview of these sensors. Their working principles are described together with their relevant advantages and disadvantages. Furthermore, the most promising emerging approaches for flowmeters design (i.e., fiber optic technology and three dimensional micro-fabrication) are briefly reviewed showing their potential for this application. Copyright © 2015 IPEM. Published by Elsevier Ltd. All rights reserved.

  9. Simulation of erosion by a particulate airflow through a ventilator

    NASA Astrophysics Data System (ADS)

    Ghenaiet, A.

    2015-08-01

    Particulate flows are a serious problem in air ventilation systems, leading to erosion of rotor blades and aerodynamic performance degradation. This paper presents the numerical results of sand particle trajectories and erosion patterns in an axial ventilator and the subsequent blade deterioration. The flow field was solved separately by using the code CFX- TASCflow. The Lagrangian approach for the solid particles tracking implemented in our inhouse code considers particle and eddy interaction, particle size distribution, particle rebounds and near walls effects. The assessment of erosion wear is based on the impact frequency and local values of erosion rate. Particle trajectories and erosion simulation revealed distinctive zones of impacts with high rates of erosion mainly on the blade pressure side, whereas the suction side is eroded around the leading edge.

  10. Enhanced Photoacoustic Gas Analyser Response Time and Impact on Accuracy at Fast Ventilation Rates during Multiple Breath Washout

    PubMed Central

    Horsley, Alex; Macleod, Kenneth; Gupta, Ruchi; Goddard, Nick; Bell, Nicholas

    2014-01-01

    Background The Innocor device contains a highly sensitive photoacoustic gas analyser that has been used to perform multiple breath washout (MBW) measurements using very low concentrations of the tracer gas SF6. Use in smaller subjects has been restricted by the requirement for a gas analyser response time of <100 ms, in order to ensure accurate estimation of lung volumes at rapid ventilation rates. Methods A series of previously reported and novel enhancements were made to the gas analyser to produce a clinically practical system with a reduced response time. An enhanced lung model system, capable of delivering highly accurate ventilation rates and volumes, was used to assess in vitro accuracy of functional residual capacity (FRC) volume calculation and the effects of flow and gas signal alignment on this. Results 10–90% rise time was reduced from 154 to 88 ms. In an adult/child lung model, accuracy of volume calculation was −0.9 to 2.9% for all measurements, including those with ventilation rate of 30/min and FRC of 0.5 L; for the un-enhanced system, accuracy deteriorated at higher ventilation rates and smaller FRC. In a separate smaller lung model (ventilation rate 60/min, FRC 250 ml, tidal volume 100 ml), mean accuracy of FRC measurement for the enhanced system was minus 0.95% (range −3.8 to 2.0%). Error sensitivity to flow and gas signal alignment was increased by ventilation rate, smaller FRC and slower analyser response time. Conclusion The Innocor analyser can be enhanced to reliably generate highly accurate FRC measurements down at volumes as low as those simulating infant lung settings. Signal alignment is a critical factor. With these enhancements, the Innocor analyser exceeds key technical component recommendations for MBW apparatus. PMID:24892522

  11. The dependency of expiratory airway collapse on pump system and flow rate in liquid ventilated rabbits.

    PubMed

    Meinhardt, J P; Ashton, B A; Annich, G M; Quintel, M; Hirschl, R B

    2003-05-30

    To evaluate the influence of pump system and flow pattern on expiratory airway collapse (EAC) in total perfluorocarbon ventilation. - Prospective, controlled, randomized animal trial for determination of (1) post-mortem changes by repeated expiration procedures (EP) with a constant flow piston pump (PP) before and after sacrifice (n = 8 rabbits), (2) differences between pump systems by subjecting animals to both PP and roller pump (RP) circuits for expiration (n = 16 rabbits). EP were performed using a servo-controlled shut-off at airway pressures < 25 cm H subset 2O randomly with either pump at different flows. - Expired volumes before and after sacrifice were not significantly different. PP and RP revealed identical mean flows, while significantly more liquid was drained using PP (p<0.05). Increasing differences towards higher flow rates indicated profound flow pulsatility in RP. - (1) post-mortem changes in expired volumes are not significant, (2) EAC is related to flow rate and pump system; (3) relationship between expiratory flow rate and drainable liquid volume is linear inverse; (4) PP provides higher drainage than RP. - Expiratory airway collapse is related to flow rate and pump system, post mortem changes in expirable volumes are not significant. Relationship between expiratory flow rate and drainable liquid volume is linear inverse, piston pump expiration provides higher drainage volumes than roller pump expiration.

  12. Evaluating humidity recovery efficiency of currently available heat and moisture exchangers: a respiratory system model study.

    PubMed

    Lucato, Jeanette Janaina Jaber; Adams, Alexander Bernard; Souza, Rogério; Torquato, Jamili Anbar; Carvalho, Carlos Roberto Ribeiro; Marini, John J

    2009-01-01

    To evaluate and compare the efficiency of humidification in available heat and moisture exchanger models under conditions of varying tidal volume, respiratory rate, and flow rate. Inspired gases are routinely preconditioned by heat and moisture exchangers to provide a heat and water content similar to that provided normally by the nose and upper airways. The absolute humidity of air retrieved from and returned to the ventilated patient is an important measurable outcome of the heat and moisture exchangers' humidifying performance. Eight different heat and moisture exchangers were studied using a respiratory system analog. The system included a heated chamber (acrylic glass, maintained at 37 degrees C), a preserved swine lung, a hygrometer, circuitry and a ventilator. Humidity and temperature levels were measured using eight distinct interposed heat and moisture exchangers given different tidal volumes, respiratory frequencies and flow-rate conditions. Recovery of absolute humidity (%RAH) was calculated for each setting. Increasing tidal volumes led to a reduction in %RAH for all heat and moisture exchangers while no significant effect was demonstrated in the context of varying respiratory rate or inspiratory flow. Our data indicate that heat and moisture exchangers are more efficient when used with low tidal volume ventilation. The roles of flow and respiratory rate were of lesser importance, suggesting that their adjustment has a less significant effect on the performance of heat and moisture exchangers.

  13. Gas pre-warming for improving performances of heated humidifiers in neonatal ventilation.

    PubMed

    Schena, E; De Paolis, E; Silvestri, S

    2011-01-01

    Adequate temperature and humidification of gas delivered must be performed during long term neonatal ventilation to avoid potential adverse health effects. Literature shows that performances of heated humidifiers are, at least in some cases, quite poor. In this study, a novel approach to gas conditioning, consisting of gas warming upstream the humidification chamber, is presented. Gas pre-warming, in combination with a control strategy based on a mathematical model taking into account a number of parameters, allows to significantly improve the heated humidifier performances. The theoretical model has been validated and experimental trials have been carried out in the whole volumetric flow-rate (Q) range of neonatal ventilation (lower than 10 L · min(-1)). Experimental results (temperature values ranging from 36 °C to 38 °C and relative humidity values from 90 % to 98 % in the whole range of Q) show values very close to the ideal thermo-hygrometric conditions. The proposed solution allows to avoid vapor condensation at low flow rates and decrease of relative humidity at high flow rates.

  14. Bias flow rate and ventilation efficiency during adult high-frequency oscillatory ventilation: a lung model study.

    PubMed

    Nagano, Osamu; Yumoto, Tetsuya; Nishimatsu, Atsunori; Kanazawa, Shunsuke; Fujita, Takahisa; Asaba, Sunao; Yamanouchi, Hideo

    2018-04-19

    Bias flow (BF) is essential to maintain mean airway pressure (MAP) and to washout carbon dioxide (CO 2 ) from the oscillator circuit during high-frequency oscillatory ventilation (HFOV). If the BF rate is inadequate, substantial CO 2 rebreathing could occur and ventilation efficiency could worsen. With lower ventilation efficiency, the required stroke volume (SV) would increase in order to obtain the same alveolar ventilation with constant frequency. The aim of this study was to assess the effect of BF rate on ventilation efficiency during adult HFOV. The R100 oscillator (Metran, Japan) was connected to an original lung model internally equipped with a simulated bronchial tree. The actual SV was measured with a flow sensor placed at the Y-piece. Carbon dioxide (CO 2 ) was continuously insufflated into the lung model ([Formula: see text]CO 2 ), and the partial pressure of CO 2 (PCO 2 ) in the lung model was monitored. Alveolar ventilation ([Formula: see text]A) was estimated as [Formula: see text]CO 2 divided by the stabilized value of PCO 2 . [Formula: see text]A was evaluated by setting SV from 80 to 180 mL (10 mL increments, n = 5) at a frequency of 8 Hz, a MAP of 25 cmH 2 O, and a BF of 10, 20, 30, and 40 L/min (study 1). Ventilation efficiency was calculated as [Formula: see text]A divided by the actual minute volume. The experiment was also performed with an actual SV of 80, 100, and 120 mL and a BF from 10 to 60 L/min (10 L/min increments: study 2). Study 1: With the same setting SV, the [Formula: see text]A with a BF of 20 L/min or more was significantly higher than that with a BF of 10 L/min. Study 2: With the same actual SV, the [Formula: see text]A and the ventilation efficiency with a BF of 30 L/min or more were significantly higher than those with a BF of 10 or 20 L/min. Increasing BF up to 30 L/min or more improved ventilation efficiency in the R100 oscillator.

  15. Constant-flow ventilation in canine experimental pulmonary emphysema.

    PubMed

    Hachenberg, T; Wendt, M; Meyer, J; Struckmeier, O; Lawin, P

    1989-07-01

    The efficacy of constant-flow ventilation (CFV) was investigated in eight mongrel dogs before (control-phase) and after development of papain-induced panlobular emphysema (PLE-phase). For CFV, heated, humidified and oxygen-enriched air was continuously delivered via two catheters positioned within each mainstem bronchus at flow rates (V) of 0.33, 0.5 and 0.66 l/s. Data obtained during intermittent positive pressure ventilation (IPPV) served as reference. In the control-phase, Pao2 was lower (P less than or equal to 0.05) and alveolo-arterial O2 difference (P(A-a)O2) was higher (P less than or equal to 0.01) during CFV at all flow rates when compared with IPPV. This may be due to inhomogeneities of intrapulmonary gas distribution and increased ventilation-perfusion (VA/Q) mismatching. Paco2 and V showed a hyperbolic relationship; constant normocapnia (5.3 kPa) was achieved at 0.48 +/- 0.21 l/s (V53). Development of PLE resulted in an increase of functional residual capacity (FRC), residual volume (RV) and static compliance (Cstat) (P less than or equal to 0.05). PaO2 had decreased and P(A-a)O2 had increased (P less than or equal to 0.05), indicating moderate pulmonary dysfunction. Oxygenation during CFV was not significantly different in the PLE-phase when compared with the control-phase. Paco2 and V showed a hyperbolic relationship and V5.3 was even lower than in the control-group (0.42 +/- 0.13 l/s). In dogs with emphysematous lungs CFV maintains sufficient gas exchange. This may be due to preferential ventilation of basal lung units, thereby counterbalancing the effects of impaired lung morphometry and increased airtrapping. Conventional mechanical ventilation is more effective in terms of oxygenation and CO2-elimination.

  16. Reductions in dead space ventilation with nasal high flow depend on physiological dead space volume: metabolic hood measurements during sleep in patients with COPD and controls.

    PubMed

    Biselli, Paolo; Fricke, Kathrin; Grote, Ludger; Braun, Andrew T; Kirkness, Jason; Smith, Philip; Schwartz, Alan; Schneider, Hartmut

    2018-05-01

    Nasal high flow (NHF) reduces minute ventilation and ventilatory loads during sleep but the mechanisms are not clear. We hypothesised NHF reduces ventilation in proportion to physiological but not anatomical dead space.11 subjects (five controls and six chronic obstructive pulmonary disease (COPD) patients) underwent polysomnography with transcutaneous carbon dioxide (CO 2 ) monitoring under a metabolic hood. During stable non-rapid eye movement stage 2 sleep, subjects received NHF (20 L·min -1 ) intermittently for periods of 5-10 min. We measured CO 2 production and calculated dead space ventilation.Controls and COPD patients responded similarly to NHF. NHF reduced minute ventilation (from 5.6±0.4 to 4.8±0.4 L·min -1 ; p<0.05) and tidal volume (from 0.34±0.03 to 0.3±0.03 L; p<0.05) without a change in energy expenditure, transcutaneous CO 2 or alveolar ventilation. There was a significant decrease in dead space ventilation (from 2.5±0.4 to 1.6±0.4 L·min -1 ; p<0.05), but not in respiratory rate. The reduction in dead space ventilation correlated with baseline physiological dead space fraction (r 2 =0.36; p<0.05), but not with respiratory rate or anatomical dead space volume.During sleep, NHF decreases minute ventilation due to an overall reduction in dead space ventilation in proportion to the extent of baseline physiological dead space fraction. Copyright ©ERS 2018.

  17. Computational Aerodynamic Simulations of a Spacecraft Cabin Ventilation Fan Design

    NASA Technical Reports Server (NTRS)

    Tweedt, Daniel L.

    2010-01-01

    Quieter working environments for astronauts are needed if future long-duration space exploration missions are to be safe and productive. Ventilation and payload cooling fans are known to be dominant sources of noise, with the International Space Station being a good case in point. To address this issue cost effectively, early attention to fan design, selection, and installation has been recommended, leading to an effort by NASA to examine the potential for small-fan noise reduction by improving fan aerodynamic design. As a preliminary part of that effort, the aerodynamics of a cabin ventilation fan designed by Hamilton Sundstrand has been simulated using computational fluid dynamics codes, and the computed solutions analyzed to quantify various aspects of the fan aerodynamics and performance. Four simulations were performed at the design rotational speed: two at the design flow rate and two at off-design flow rates. Following a brief discussion of the computational codes, various aerodynamic- and performance-related quantities derived from the computed flow fields are presented along with relevant flow field details. The results show that the computed fan performance is in generally good agreement with stated design goals.

  18. Effect of cabin ventilation rate on ultrafine particle exposure inside automobiles.

    PubMed

    Knibbs, Luke D; de Dear, Richard J; Morawska, Lidia

    2010-05-01

    We alternately measured on-road and in-vehicle ultrafine (<100 nm) particle (UFP) concentration for 5 passenger vehicles that comprised an age range of 18 years. A range of cabin ventilation settings were assessed during 301 trips through a 4 km road tunnel in Sydney, Australia. Outdoor air flow (ventilation) rates under these settings were quantified on open roads using tracer gas techniques. Significant variability in tunnel trip average median in-cabin/on-road (I/O) UFP ratios was observed (0.08 to approximately 1.0). Based on data spanning all test automobiles and ventilation settings, a positive linear relationship was found between outdoor air flow rate and I/O ratio, with the former accounting for a substantial proportion of variation in the latter (R(2) = 0.81). UFP concentrations recorded in-cabin during tunnel travel were significantly higher than those reported by comparable studies performed on open roadways. A simple mathematical model afforded the ability to predict tunnel trip average in-cabin UFP concentrations with good accuracy. Our data indicate that under certain conditions, in-cabin UFP exposures incurred during tunnel travel may contribute significantly to daily exposure. The UFP exposure of automobile occupants appears strongly related to their choice of ventilation setting and vehicle.

  19. Demand controlled ventilating systems: Sensor market survey. Energy conservation in buildings and community systems programme, annex 18, December 1991

    NASA Astrophysics Data System (ADS)

    Raatschen, W.; Sjoegren, M.

    The subject of indoor and outdoor air quality has generated a great deal of attention in many countries. Areas of concern include outgassing of building materials as well as occupant-generated pollutants such as carbon dioxide, moisture, and odors. Progress has also been made towards addressing issues relating to the air tightness of the building envelope. Indoor air quality studies indicate that better control of supply flow rates as well as the air distribution pattern within buildings are necessary. One method of maintaining good indoor air quality without extensive energy consumption is to control the ventilation rate according to the needs and demands of the occupants, or to preserve the building envelope. This is accomplished through the use of demand controlled ventilating (DCV) systems. The specific objective of Annex 18 is to develop guidelines for demand controlled ventilating systems based on state of the art analyses, case studies on ventilation effectiveness, and proposed ventilation rates for different users in domestic, office, and school buildings.

  20. Simulated Altitude Investigation of Stewart-Warner Model 906-B Combustion Heater

    NASA Technical Reports Server (NTRS)

    Ebersbach, Frederick R.; Cervenka, Adolph J.

    1947-01-01

    An investigation has been conducted to determine thermal and pressure-drop performance and the operational characteristics of a Stewart-Warner model 906-B combustion heater. The performance tests covered a range of ventilating-air flows from 500 to 3185 pounds per hour, combustion-air pressure drops from 5 to 35 inches of water, and pressure altitudes from sea level to 41,000 feet. The operational characteristics investigated were the combustion-air flows for sustained combustion and for consistent ignition covering fuel-air ratios ranging from 0.033 to 0.10 and pressure altitudes from sea level to 45,000 feet. Rated heat output of 50,000 Btu per hour was obtained at pressure altitudes up to 27,000 feet for ventilating-air flows greater than 800 pounds per hour; rated output was not obtained at ventilating-air flow below 800 pounds per hour at any altitude. The maximum heater efficiency was found to be 60.7 percent at a fuel-air ratio of 0.050, a sea-level pressure altitude, a ventilating-air temperature of 0 F, combustion-air temperature of 14 F, a ventilating-air flow of 690 pounds per hour, and a combustion-air flow of 72.7 pounds per hour. The minimum combustion-air flow for sustained combustion at a pressure altitude of 25,000 feet was about 9 pounds per hour for fuel-air ratios between 0.037 and 0.099 and at a pressure altitude of 45,000 feet increased to 18 pounds per hour at a fuel-air ratio of 0.099 and 55 pounds per hour at a fuel-air ratio of 0.036. Combustion could be sustained at combustion-air flows above values of practical interest. The maximum flow was limited, however, by excessively high exhaust-gas temperature or high pressure drop. Both maximum and minimum combustion-air flows for consistent ignition decrease with increasing pressure altitude and the two curves intersect at a pressure altitude of approximately 25,000 feet and a combustion-air flow of approximately 28 pounds per hour.

  1. Simple, Inexpensive Model Spirometer for Understanding Ventilation Volumes

    ERIC Educational Resources Information Center

    Giuliodori, Mauricio J.; DiCarlo, Stephen E.

    2004-01-01

    Spirometers are useful for enhancing students' understanding of normal lung volumes, capacities, and flow rates. Spirometers are also excellent for understanding how lung diseases alter ventilation volumes. However, spirometers are expensive, complex, and not appropriate for programs with limited space and budgets. Therefore, we developed a…

  2. A multiscale MDCT image-based breathing lung model with time-varying regional ventilation

    PubMed Central

    Yin, Youbing; Choi, Jiwoong; Hoffman, Eric A.; Tawhai, Merryn H.; Lin, Ching-Long

    2012-01-01

    A novel algorithm is presented that links local structural variables (regional ventilation and deforming central airways) to global function (total lung volume) in the lung over three imaged lung volumes, to derive a breathing lung model for computational fluid dynamics simulation. The algorithm constitutes the core of an integrative, image-based computational framework for subject-specific simulation of the breathing lung. For the first time, the algorithm is applied to three multi-detector row computed tomography (MDCT) volumetric lung images of the same individual. A key technique in linking global and local variables over multiple images is an in-house mass-preserving image registration method. Throughout breathing cycles, cubic interpolation is employed to ensure C1 continuity in constructing time-varying regional ventilation at the whole lung level, flow rate fractions exiting the terminal airways, and airway deformation. The imaged exit airway flow rate fractions are derived from regional ventilation with the aid of a three-dimensional (3D) and one-dimensional (1D) coupled airway tree that connects the airways to the alveolar tissue. An in-house parallel large-eddy simulation (LES) technique is adopted to capture turbulent-transitional-laminar flows in both normal and deep breathing conditions. The results obtained by the proposed algorithm when using three lung volume images are compared with those using only one or two volume images. The three-volume-based lung model produces physiologically-consistent time-varying pressure and ventilation distribution. The one-volume-based lung model under-predicts pressure drop and yields un-physiological lobar ventilation. The two-volume-based model can account for airway deformation and non-uniform regional ventilation to some extent, but does not capture the non-linear features of the lung. PMID:23794749

  3. Investigation into the behaviors of ventilated supercavities in unsteady flow

    NASA Astrophysics Data System (ADS)

    Shao, Siyao; Wu, Yue; Haynes, Joseph; Arndt, Roger E. A.; Hong, Jiarong

    2018-05-01

    A systematic investigation of ventilated supercavitation behaviors in an unsteady flow is conducted using a high-speed water tunnel at the Saint Anthony Falls Laboratory. The cavity is generated with a forward facing model under varying ventilation rates and cavitator sizes. The unsteady flow is produced by a gust generator consisting of two hydrofoils flapping in unison with a varying angle of attack (AoA) and frequency (fg). The current experiment reveals five distinct cavity states, namely, the stable state, wavy state, pulsating state I, pulsating state II, and collapsing state, based on the variation of cavity geometry and pressure signatures inside the cavity. The distribution of cavity states over a broad range of unsteady conditions is summarized in a cavity state map. It shows that the transition of the supercavity from the stable state to pulsating and collapsing states is primarily induced by increasing AoA while the transition to the wavy state triggers largely by increasing fg. Remarkably, the state map over the non-dimensionalized half wavelength and wave amplitude of the perturbation indicates that the supercavity loses its stability and transitions to pulsating or collapsing states when the level of its distortion induced by the flow unsteadiness exceeds the cavity dimension under a steady condition. The state maps under different ventilation rates and cavitator sizes yield similar distribution but show that the occurrence of the cavity collapse can be suppressed with increasing ventilation coefficient or cavitator size. Such knowledge can be integrated into designing control strategies for the supercavitating devices operating under different unsteady conditions.

  4. Experimental investigation into the interaction between the human body and room airflow and its effect on thermal comfort under stratum ventilation.

    PubMed

    Cheng, Y; Lin, Z

    2016-04-01

    Room occupants' comfort and health are affected by the airflow. Nevertheless, they themselves also play an important role in indoor air distribution. This study investigated the interaction between the human body and room airflow under stratum ventilation. Simplified thermal manikin was employed to effectively resemble the human body as a flow obstacle and/or free convective heat source. Unheated and heated manikins were designed to fully evaluate the impact of the manikin at various airflow rates. Additionally, subjective human tests were conducted to evaluate thermal comfort for the occupants in two rows. The findings show that the manikin formed a local blockage effect, but the supply airflow could flow over it. With the body heat from the manikin, the air jet penetrated farther compared with that for the unheated manikin. The temperature downstream of the manikin was also higher because of the convective effect. Elevating the supply airflow rate from 7 to 15 air changes per hour varied the downstream airflow pattern dramatically, from an uprising flow induced by body heat to a jet-dominated flow. Subjective assessments indicated that stratum ventilation provided thermal comfort for the occupants in both rows. Therefore, stratum ventilation could be applied in rooms with occupants in multiple rows. © 2015 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

  5. Evaluating Humidity Recovery Efficiency of Currently Available Heat and Moisture Exchangers: A Respiratory System Model Study

    PubMed Central

    Lucato, Jeanette Janaina Jaber; Adams, Alexander Bernard; Souza, Rogério; Torquato, Jamili Anbar; Carvalho, Carlos Roberto Ribeiro; Marini, John J

    2009-01-01

    OBJECTIVES: To evaluate and compare the efficiency of humidification in available heat and moisture exchanger models under conditions of varying tidal volume, respiratory rate, and flow rate. INTRODUCTION: Inspired gases are routinely preconditioned by heat and moisture exchangers to provide a heat and water content similar to that provided normally by the nose and upper airways. The absolute humidity of air retrieved from and returned to the ventilated patient is an important measurable outcome of the heat and moisture exchangers’ humidifying performance. METHODS: Eight different heat and moisture exchangers were studied using a respiratory system analog. The system included a heated chamber (acrylic glass, maintained at 37°C), a preserved swine lung, a hygrometer, circuitry and a ventilator. Humidity and temperature levels were measured using eight distinct interposed heat and moisture exchangers given different tidal volumes, respiratory frequencies and flow-rate conditions. Recovery of absolute humidity (%RAH) was calculated for each setting. RESULTS: Increasing tidal volumes led to a reduction in %RAH for all heat and moisture exchangers while no significant effect was demonstrated in the context of varying respiratory rate or inspiratory flow. CONCLUSIONS: Our data indicate that heat and moisture exchangers are more efficient when used with low tidal volume ventilation. The roles of flow and respiratory rate were of lesser importance, suggesting that their adjustment has a less significant effect on the performance of heat and moisture exchangers. PMID:19578664

  6. Actual performance of mechanical ventilators in ICU: a multicentric quality control study.

    PubMed

    Govoni, Leonardo; Dellaca', Raffaele L; Peñuelas, Oscar; Bellani, Giacomo; Artigas, Antonio; Ferrer, Miquel; Navajas, Daniel; Pedotti, Antonio; Farré, Ramon

    2012-01-01

    Even if the performance of a given ventilator has been evaluated in the laboratory under very well controlled conditions, inappropriate maintenance and lack of long-term stability and accuracy of the ventilator sensors may lead to ventilation errors in actual clinical practice. The aim of this study was to evaluate the actual performances of ventilators during clinical routines. A resistance (7.69 cmH(2)O/L/s) - elastance (100 mL/cmH(2)O) test lung equipped with pressure, flow, and oxygen concentration sensors was connected to the Y-piece of all the mechanical ventilators available for patients in four intensive care units (ICUs; n = 66). Ventilators were set to volume-controlled ventilation with tidal volume = 600 mL, respiratory rate = 20 breaths/minute, positive end-expiratory pressure (PEEP) = 8 cmH(2)O, and oxygen fraction = 0.5. The signals from the sensors were recorded to compute the ventilation parameters. The average ± standard deviation and range (min-max) of the ventilatory parameters were the following: inspired tidal volume = 607 ± 36 (530-723) mL, expired tidal volume = 608 ± 36 (530-728) mL, peak pressure = 20.8 ± 2.3 (17.2-25.9) cmH(2)O, respiratory rate = 20.09 ± 0.35 (19.5-21.6) breaths/minute, PEEP = 8.43 ± 0.57 (7.26-10.8) cmH(2)O, oxygen fraction = 0.49 ± 0.014 (0.41-0.53). The more error-prone parameters were the ones related to the measure of flow. In several cases, the actual delivered mechanical ventilation was considerably different from the set one, suggesting the need for improving quality control procedures for these machines.

  7. High Resolution Time Series Cave Ventilation Processes and the Effects on Cave Air Chemistry and Drip Waters: Speleoclimatology and Proxy Calibration

    NASA Astrophysics Data System (ADS)

    Kowalczk, A. J.; Froelich, P. N.; Gaffka, C.; Tremaine, D.

    2008-12-01

    Continuous high resolution (sub-hourly), long-term (Nov 2007-present) monitoring of cave air chemistry (Temperature, Relative Humidity, Barometric Pressure, Radon-222, CO2, Air flow, Wind speed and direction) in a shallow subtropical cave (Hollow Ridge) in N Florida reveals two major ventilation mechanisms: 1) ventilation driven by winds across the cave entrances, and 2) ventilation driven by density differences between atmospheric and cave air. The degree and type of ventilation strongly influence the 222Rn and CO2 of cave air, which in turn affects the timing and extent of calcite deposition in speleothems. The degree of ventilation is estimated using a cave air CO2-δ13CO2 Keeling Plot, or a simple radon deficiency model. Results show cave air has an atmospheric component ranging from 10-90%. During fall and winter, average CO2 (700 ppmv) and 222Rn (50-100 dpm/L) are lower than in spring and summer (CO2 = 1200 ppmv; 222Rn = 1000 dpm/L) due to increased winter ventilation. Decreased ventilation during the summer allows CO2 and 222Rn levels to rise. Winter daily ventilation is primarily a function of density gradients between cave air and atmospheric air, while summer daily ventilation is primarily a function of late morning NW-NE winds above the cave. Stable isotope analyses of drip water (fracture drip and pore flow drip) and aquifer water from Hollow Ridge agree with previous isotope studies of drip water at Florida Caverns State Park, 2 km to the NE. During summer, isotopic composition of pore flow drip water (δ18O -3.8 to -4.0 per mil; δD -17.3 to -20.2 per mil VSMOW) and aquifer water (δ18O -4.0 per mil; δD -18.0 to -21.1 per mil) are similar to average annual weighted isotopic composition of precipitation (δ18O -3.6 per mil) while fracture drip waters (δ18O -3 to -3.4 per mil; δD -11.9 to -14.3 per mil) likely reflect the isotopic composition of individual precipitation events. Pore flow drip waters δ18O are weakly correlated with drip rates (enriched δ18O during periods of higher drip rates) but show no correlation to precipitation amount. Knowledge of the type of drip flow is important when considering stalagmites for paleoclimate studies. A significant decrease in drip rate was observed from June (1034 drips/hour) through August 2008 (34 drips/hour). Higher water demands during summer months with increased evapotranspiration may be responsible for this decrease. A semi-diurnal drip rate cycle, negatively correlated with barometric pressure, is also observed throughout the period. This strong negative correlation is hypothesized to be controlled by atmospheric tidal oscillations. Observations into the fall and winter seasons should reveal seasonality, if any, and if there is an evapotranspiration effect present in the water cycle. High resolution studies of cave air chemistry and ventilation processes will enhance knowledge of the timing, extent, and isotopic and chemical composition of calcite deposition. When combined with drip water and precipitation isotope analyses, these studies will improve the understanding and interpretation of high- resolution (sub-annual) speleothem paleoclimate records.

  8. Sensor-based demand controlled ventilation

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    De Almeida, A.T.; Fisk, W.J.

    In most buildings, occupancy and indoor pollutant emission rates vary with time. With sensor-based demand-controlled ventilation (SBDCV), the rate of ventilation (i.e., rate of outside air supply) also varies with time to compensate for the changes in pollutant generation. In other words, SBDCV involves the application of sensing, feedback and control to modulate ventilation. Compared to ventilation without feedback, SBDCV offers two potential advantages: (1) better control of indoor pollutant concentrations; and (2) lower energy use and peak energy demand. SBDCV has the potential to improve indoor air quality by increasing the rate of ventilation when indoor pollutant generation ratesmore » are high and occupants are present. SBDCV can also save energy by decreasing the rate of ventilation when indoor pollutant generation rates are low or occupants are absent. After providing background information on indoor air quality and ventilation, this report provides a relatively comprehensive discussion of SBDCV. Topics covered in the report include basic principles of SBDCV, sensor technologies, technologies for controlling air flow rates, case studies of SBDCV, application of SBDCV to laboratory buildings, and research needs. SBDCV appears to be an increasingly attractive technology option. Based on the review of literature and theoretical considerations, the application of SBDCV has the potential to be cost-effective in applications with the following characteristics: (a) a single or small number of dominant pollutants, so that ventilation sufficient to control the concentration of the dominant pollutants provides effective control of all other pollutants; (b) large buildings or rooms with unpredictable temporally variable occupancy or pollutant emission; and (c) climates with high heating or cooling loads or locations with expensive energy.« less

  9. Delivery of tidal volume from four anaesthesia ventilators during volume-controlled ventilation: a bench study.

    PubMed

    Wallon, G; Bonnet, A; Guérin, C

    2013-06-01

    Tidal volume (V(T)) must be accurately delivered by anaesthesia ventilators in the volume-controlled ventilation mode in order for lung protective ventilation to be effective. However, the impact of fresh gas flow (FGF) and lung mechanics on delivery of V(T) by the newest anaesthesia ventilators has not been reported. We measured delivered V(T) (V(TI)) from four anaesthesia ventilators (Aisys™, Flow-i™, Primus™, and Zeus™) on a pneumatic test lung set with three combinations of lung compliance (C, ml cm H2O(-1)) and resistance (R, cm H2O litre(-1) s(-2)): C60R5, C30R5, C60R20. For each CR, three FGF rates (0.5, 3, 10 litre min(-1)) were investigated at three set V(T)s (300, 500, 800 ml) and two values of PEEP (0 and 10 cm H2O). The volume error = [(V(TI) - V(Tset))/V(Tset)] ×100 was computed in body temperature and pressure-saturated conditions and compared using analysis of variance. For each CR and each set V(T), the absolute value of the volume error significantly declined from Aisys™ to Flow-i™, Zeus™, and Primus™. For C60R5, these values were 12.5% for Aisys™, 5% for Flow-i™ and Zeus™, and 0% for Primus™. With an increase in FGF, absolute values of the volume error increased only for Aisys™ and Zeus™. However, in C30R5, the volume error was minimal at mid-FGF for Aisys™. The results were similar at PEEP 10 cm H2O. Under experimental conditions, the volume error differed significantly between the four new anaesthesia ventilators tested and was influenced by FGF, although this effect may not be clinically relevant.

  10. Flow-field characteristics of high-temperature annular buoyant jets and their development laws influenced by ventilation system.

    PubMed

    Wang, Yi; Huang, Yanqiu; Liu, Jiaping; Wang, Hai; Liu, Qiuhan

    2013-01-01

    The flow-field characteristics of high-temperature annular buoyant jets as well as the development laws influenced by ventilation system were studied using numerical methods to eliminate the pollutants effectively in this paper. The development laws of high-temperature annular buoyant jets were analyzed and compared with previous studies, including radial velocity distribution, axial velocity and temperature decay, reattachment position, cross-section diameter, volumetric flow rate, and velocity field characteristics with different pressures at the exhaust hood inlet. The results showed that when the ratio of outer diameter to inner diameter of the annulus was smaller than 5/2, the flow-field characteristics had significant difference compared to circular buoyant jets with the same outer diameter. For similar diameter ratios, reattachment in this paper occurred further downstream in contrast to previous study. Besides, the development laws of volumetric flow rate and cross-section diameter were given with different initial parameters. In addition, through analyzing air distribution characteristics under the coupling effect of high-temperature annular buoyant jets and ventilation system, it could be found that the position where maximum axial velocity occurred was changing gradually when the pressure at the exhaust hood inlet changed from 0 Pa to -5 Pa.

  11. Flow-Field Characteristics of High-Temperature Annular Buoyant Jets and Their Development Laws Influenced by Ventilation System

    PubMed Central

    Liu, Jiaping; Wang, Hai; Liu, Qiuhan

    2013-01-01

    The flow-field characteristics of high-temperature annular buoyant jets as well as the development laws influenced by ventilation system were studied using numerical methods to eliminate the pollutants effectively in this paper. The development laws of high-temperature annular buoyant jets were analyzed and compared with previous studies, including radial velocity distribution, axial velocity and temperature decay, reattachment position, cross-section diameter, volumetric flow rate, and velocity field characteristics with different pressures at the exhaust hood inlet. The results showed that when the ratio of outer diameter to inner diameter of the annulus was smaller than 5/2, the flow-field characteristics had significant difference compared to circular buoyant jets with the same outer diameter. For similar diameter ratios, reattachment in this paper occurred further downstream in contrast to previous study. Besides, the development laws of volumetric flow rate and cross-section diameter were given with different initial parameters. In addition, through analyzing air distribution characteristics under the coupling effect of high-temperature annular buoyant jets and ventilation system, it could be found that the position where maximum axial velocity occurred was changing gradually when the pressure at the exhaust hood inlet changed from 0 Pa to −5 Pa. PMID:24000278

  12. High-Flow Nasal Cannula Oxygenation in Immunocompromised Patients With Acute Hypoxemic Respiratory Failure: A Groupe de Recherche Respiratoire en Réanimation Onco-Hématologique Study.

    PubMed

    Lemiale, Virginie; Resche-Rigon, Matthieu; Mokart, Djamel; Pène, Frédéric; Argaud, Laurent; Mayaux, Julien; Guitton, Christophe; Rabbat, Antoine; Girault, Christophe; Kouatchet, Achille; Vincent, François; Bruneel, Fabrice; Nyunga, Martine; Seguin, Amélie; Klouche, Kada; Colin, Gwenahel; Kontar, Loay; Perez, Pierre; Meert, Anne-Pascale; Benoit, Dominique D; Papazian, Laurent; Demoule, Alexandre; Chevret, Sylvie; Azoulay, Elie

    2017-03-01

    In immunocompromised patients with acute respiratory failure, invasive mechanical ventilation remains associated with high mortality. Choosing the adequate oxygenation strategy is of the utmost importance in that setting. High-flow nasal oxygen has recently shown survival benefits in unselected patients with acute respiratory failure. The objective was to assess outcomes of immunocompromised patients with hypoxemic acute respiratory failure treated with high-flow nasal oxygen. We performed a post hoc analysis of a randomized controlled trial of noninvasive ventilation in critically ill immunocompromised patients with hypoxemic acute respiratory failure. Twenty-nine ICUs in France and Belgium. Critically ill immunocompromised patients with hypoxemic acute respiratory failure. A propensity score-based approach was used to assess the impact of high-flow nasal oxygen compared with standard oxygen on day 28 mortality. Among 374 patients included in the study, 353 met inclusion criteria. Underlying disease included mostly malignancies (n = 296; 84%). Acute respiratory failure etiologies were mostly pneumonia (n = 157; 44.4%) or opportunistic infection (n = 76; 21.5%). Noninvasive ventilation was administered to 180 patients (51%). Invasive mechanical ventilation was ultimately needed in 142 patients (40.2%). Day 28 mortality was 22.6% (80 deaths). Throughout the ICU stay, 127 patients (36%) received high-flow nasal oxygen whereas 226 patients received standard oxygen. Ninety patients in each group (high-flow nasal oxygen or standard oxygen) were matched according to the propensity score, including 91 of 180 (51%) who received noninvasive ventilation. High-flow nasal oxygen was neither associated with a lower intubation rate (hazard ratio, 0.42; 95% CI, 0.11-1.61; p = 0.2) nor day 28 mortality (hazard ratio, 0.80; 95% CI, 0.45-1.42; p = 0.45). In immunocompromised patients with hypoxemic acute respiratory failure, high-flow nasal oxygen when compared with standard oxygen did not reduce intubation or survival rates. However, these results could be due to low statistical power or unknown confounders associated with the subgroup analysis. A randomized trial is needed.

  13. Numerical analyses of ventilated cavitation over a 2-D NACA0015 hydrofoil using two turbulence modeling methods

    NASA Astrophysics Data System (ADS)

    Yang, Dan-dan; Yu, An; Ji, Bin; Zhou, Jia-jian; Luo, Xian-wu

    2018-04-01

    The present paper studies the ventilated cavitation over a NACA0015 hydrofoil by numerical methods. The corresponding cavity evolutions are obtained at three ventilation rates by using the level set method. To depict the complicated turbulent flow structure, the filter-based density corrected model (FBDCM) and the modified partially-averaged Navier-Stokes (MPANS) model are applied in the present numerical analyses. It is indicated that the predicted results of the cavitation shedding dynamics by both turbulence models agree fairly well with the experimental data. It is also noted that the shedding frequency and the super cavity length predicted by the MPANS method are closer to the experiment data as compared to that predicted by the FBDCM model. The simulation results show that in the ventilated cavitation, the vapor cavity and the air cavity have the same shedding frequency. As the ventilated rate increases, the vapor cavity is depressed rapidly. The cavitation-vortex interaction in the ventilated cavitation is studied based on the vorticity transport equation (VTE) and the Lagrangian coherent structure (LCS). Those results demonstrate that the vortex dilatation and baroclinic torque terms are highly dependent on the evolution of the cavitation. In addition, from the LCSs and the tracer particles in the flow field, one may see the process from the attached cavity to the cloud cavity.

  14. Passive Cooling of Body Armor

    NASA Astrophysics Data System (ADS)

    Holtz, Ronald; Matic, Peter; Mott, David

    2013-03-01

    Warfighter performance can be adversely affected by heat load and weight of equipment. Current tactical vest designs are good insulators and lack ventilation, thus do not provide effective management of metabolic heat generated. NRL has undertaken a systematic study of tactical vest thermal management, leading to physics-based strategies that provide improved cooling without undesirable consequences such as added weight, added electrical power requirements, or compromised protection. The approach is based on evaporative cooling of sweat produced by the wearer of the vest, in an air flow provided by ambient wind or ambulatory motion of the wearer. Using an approach including thermodynamic analysis, computational fluid dynamics modeling, air flow measurements of model ventilated vest architectures, and studies of the influence of fabric aerodynamic drag characteristics, materials and geometry were identified that optimize passive cooling of tactical vests. Specific architectural features of the vest design allow for optimal ventilation patterns, and selection of fabrics for vest construction optimize evaporation rates while reducing air flow resistance. Cooling rates consistent with the theoretical and modeling predictions were verified experimentally for 3D mockups.

  15. Effectiveness of in-room air filtration and dilution ventilation for tuberculosis infection control.

    PubMed

    Miller-Leiden, S; Lobascio, C; Nazaroff, W W; Macher, J M

    1996-09-01

    Tuberculosis (TB) is a public health problem that may pose substantial risks to health care workers and others. TB infection occurs by inhalation of airborne bacteria emitted by persons with active disease. We experimentally evaluated the effectiveness of in-room air filtration systems, specifically portable air filters (PAFs) and ceiling-mounted air filters (CMAFs), in conjunction with dilution ventilation, for controlling TB exposure in high-risk settings. For each experiment, a test aerosol was continuously generated and released into a full-sized room. With the in-room air filter and room ventilation system operating, time-averaged airborne particle concentrations were measured at several points. The effectiveness of in-room air filtration plus ventilation was determined by comparing particle concentrations with and without device operation. The four PAFs and three CMAFs we evaluated reduced room-average particle concentrations, typically by 30% to 90%, relative to a baseline scenario with two air-changes per hour of ventilation (outside air) only. Increasing the rate of air flow recirculating through the filter and/or air flow from the ventilation did not always increase effectiveness. Concentrations were generally higher near the emission source than elsewhere in the room. Both the air flow configuration of the filter and its placement within the room were important, influencing room air flow patterns and the spatial distribution of concentrations. Air filters containing efficient, but non-high efficiency particulate air (HEPA) filter media were as effective as air filters containing HEPA filter media.

  16. Effectiveness of In-Room Air Filtration and Dilution Ventilation for Tuberculosis Infection Control.

    PubMed

    Miller-Leiden, S; Lohascio, C; Nazaroff, W W; Macher, J M

    1996-09-01

    Tuberculosis (TB) is a public health problem that may pose substantial risks to health care workers and others. TB infection occurs by inhalation of airborne bacteria emitted by persons with active disease. We experimentally evaluated the effectiveness of in-room air filtration systems, specifically portable air filters (PAFs) and ceiling-mounted air filters (CMAFs), in conjunction with dilution ventilation, for controlling TB exposure in high-risk settings. For each experiment, a test aerosol was continuously generated and released into a full-sized room. With the in-room air filter and room ventilation system operating, time-averaged airborne particle concentrations were measured at several points. The effectiveness of in-room air filtration plus ventilation was determined by comparing particle concentrations with and without device operation. The four PAFs and three CMAFs we evaluated reduced room-average particle concentrations, typically by 30% to 90%, relative to a baseline scenario with two air-changes per hour of ventilation (outside air) only. Increasing the rate of air flow recirculating through the filter and/or air flow from the ventilation did not always increase effectiveness. Concentrations were generally higher near the emission source than elsewhere in the room. Both the air flow configuration of the filter and its placement within the room were important, influencing room air flow patterns and the spatial distribution of concentrations. Air filters containing efficient, but non-high efficiency particulate air (HEPA) filter media were as effective as air filters containing HEPA filter media.

  17. Low-Flow Extracorporeal Carbon Dioxide Removal Using the Hemolung Respiratory Dialysis System® to Facilitate Lung-Protective Mechanical Ventilation in Acute Respiratory Distress Syndrome.

    PubMed

    Akkanti, Bindu; Rajagopal, Keshava; Patel, Kirti P; Aravind, Sangeeta; Nunez-Centanu, Emmanuel; Hussain, Rahat; Shabari, Farshad Raissi; Hofstetter, Wayne L; Vaporciyan, Ara A; Banjac, Igor S; Kar, Biswajit; Gregoric, Igor D; Loyalka, Pranav

    2017-06-01

    Extracorporeal carbon dioxide removal (ECCO 2 R) permits reductions in alveolar ventilation requirements that the lungs would otherwise have to provide. This concept was applied to a case of hypercapnia refractory to high-level invasive mechanical ventilator support. We present a case of an 18-year-old man who developed post-pneumonectomy acute respiratory distress syndrome (ARDS) after resection of a mediastinal germ cell tumor involving the left lung hilum. Hypercapnia and hypoxemia persisted despite ventilator support even at traumatic levels. ECCO 2 R using a miniaturized system was instituted and provided effective carbon dioxide elimination. This facilitated establishment of lung-protective ventilator settings and lung function recovery. Extracorporeal lung support increasingly is being applied to treat ARDS. However, conventional extracorporeal membrane oxygenation (ECMO) generally involves using large cannulae capable of carrying high flow rates. A subset of patients with ARDS has mixed hypercapnia and hypoxemia despite high-level ventilator support. In the absence of profound hypoxemia, ECCO 2 R may be used to reduce ventilator support requirements to lung-protective levels, while avoiding risks associated with conventional ECMO.

  18. Utilizing Diffusion Theory to predict carbon dioxide concentration in an indoor environment

    NASA Astrophysics Data System (ADS)

    Kramer, Andrew R.

    This research details a new method of relating sources of carbon dioxide to carbon dioxide concentration in a room operating in a reduced ventilation mode by utilizing Diffusion Theory. The theoretical basis of this research involved solving Fick's Second Law of Diffusion in spherical coordinates for a source of carbon dioxide flowing at a constant rate and located in the center of an impermeable spherical boundary. The solution was developed using a Laplace Transformation. A spherical diffusion test chamber was constructed and used to validate and benchmark the developed theory. The method was benchmarked by using Dispersion Coefficients for large carbon dioxide flow rates due to diffusion induced convection. The theoretical model was adapted to model a room operating with restricted ventilation in the presence of a known, constant source of carbon dioxide. The room was modeled as a sphere of volume equal to the room and utilized a Dispersion Coefficient that is consistent with published values. The developed Diffusion Model successfully predicted the spatial concentration of carbon dioxide in a room operating in a reduced ventilation mode in the presence of a source of carbon dioxide. The flow rates of carbon dioxide that were used in the room are comparable to the average flow rate of carbon dioxide from a person during quiet breathing, also known as the Tidal Breathing. This indicates the Diffusion Model developed from this research has the potential to correlate carbon dioxide concentration with static occupancy levels which can lead to energy savings through a reduction in air exchange rates when low occupancy is detected.

  19. Effect of ventilation velocity on hexavalent chromium and isocyanate exposures in aircraft paint spraying.

    PubMed

    Bennett, James; Marlow, David; Nourian, Fariba; Breay, James; Feng, Amy; Methner, Mark

    2018-03-01

    Exposure control system performance was evaluated during aircraft paint spraying at a military facility. Computational fluid dynamics (CFD) modeling, tracer gas testing, and exposure monitoring examined contaminant exposure vs. crossflow ventilation velocity. CFD modeling using the RNG k-ϵ turbulence model showed exposures to simulated methyl isobutyl ketone of 294 and 83.6 ppm, as a spatial average of five worker locations, for velocities of 0.508 and 0.381 m/s (100 and 75 fpm), respectively. In tracer gas experiments, observed supply/exhaust velocities of 0.706/0.503 m/s (136/99 fpm) were termed full-flow, and reduced velocities were termed 3/4-flow and half-flow. Half-flow showed higher tracer gas concentrations than 3/4-flow, which had the lowest time-averaged concentration, with difference in log means significant at the 95% confidence level. Half-flow compared to full-flow and 3/4-flow compared to full-flow showed no statistically significant difference. CFD modeling using these ventilation conditions agreed closely with the tracer results for the full-flow and 3/4-flow comparison, yet not for the 3/4-flow and half-flow comparison. Full-flow conditions at the painting facility produced a velocity of 0.528 m/s (104 fpm) midway between supply and exhaust locations, with the supply rate of 94.4 m 3 /s (200,000 cfm) exceeding the exhaust rate of 68.7 m 3 /s (146,000 cfm). Ventilation modifications to correct this imbalance created a midhangar velocity of 0.406 m/s (80.0 fpm). Personal exposure monitoring for two worker groups-sprayers and sprayer helpers ("hosemen")-compared process duration means for the two velocities. Hexavalent chromium (Cr[VI]) exposures were 500 vs. 360 µg/m 3 for sprayers and 120 vs. 170 µg/m 3 for hosemen, for 0.528 m/s (104 fpm) and 0.406 m/s (80.0 fpm), respectively. Hexamethylene diisocyanate (HDI) monomer means were 32.2 vs. 13.3 µg/m 3 for sprayers and 3.99 vs. 8.42 µg/m 3 for hosemen. Crossflow velocities affected exposures inconsistently, and local work zone velocities were much lower. Aircraft painting contaminant control is accomplished better with the unidirectional crossflow ventilation presented here than with other observed configurations. Exposure limit exceedances for this ideal condition reinforce continued use of personal protective equipment.

  20. Cost-Effectiveness Analysis of Nasal Continuous Positive Airway Pressure Versus Nasal High Flow Therapy as Primary Support for Infants Born Preterm.

    PubMed

    Huang, Li; Roberts, Calum T; Manley, Brett J; Owen, Louise S; Davis, Peter G; Dalziel, Kim M

    2018-05-01

    To compare the cost-effectiveness of 2 common "noninvasive" modes of respiratory support for infants born preterm. An economic evaluation was conducted as a component of a multicenter, randomized control trial from 2013 to 2015 enrolling infants born preterm at ≥28 weeks of gestation with respiratory distress, <24 hours old, who had not previously received endotracheal intubation and mechanical ventilation or surfactant. The economic evaluation was conducted from a healthcare sector perspective and the time horizon was from birth until death or first discharge. The cost-effectiveness of continuous positive airway pressure (CPAP) vs high-flow with "rescue" CPAP backup and high-flow without rescue CPAP backup (as sole primary support) were analyzed by using the hospital cost of inpatient stay in a tertiary center and the rates of endotracheal intubation and mechanical ventilation during admission. Hospital inpatient cost records for 435 infants enrolled in all Australian centers were obtained. With "rescue" CPAP backup, an incremental cost-effectiveness ratio was estimated of A$179 000 (US$123 000) per ventilation avoided if CPAP was used compared with high flow. Without rescue CPAP backup, cost per ventilation avoided was A$7000 (US$4800) if CPAP was used compared with high flow. As sole primary support, CPAP is highly likely to be cost-effective compared with high flow. Neonatal units choosing to use only one device should apply CPAP as primary respiratory support. Compared with high-flow with rescue CPAP backup, CPAP is unlikely to be cost-effective if willingness to pay per ventilation avoided is less than A$179 000 (US$123 000). Copyright © 2018 Elsevier Inc. All rights reserved.

  1. Respiratory mechanics by least squares fitting in mechanically ventilated patients: application on flow-limited COPD patients.

    PubMed

    Volta, Carlo A; Marangoni, Elisabetta; Alvisi, Valentina; Capuzzo, Maurizia; Ragazzi, Riccardo; Pavanelli, Lina; Alvisi, Raffaele

    2002-01-01

    Although computerized methods of analyzing respiratory system mechanics such as the least squares fitting method have been used in various patient populations, no conclusive data are available in patients with chronic obstructive pulmonary disease (COPD), probably because they may develop expiratory flow limitation (EFL). This suggests that respiratory mechanics be determined only during inspiration. Eight-bed multidisciplinary ICU of a teaching hospital. Eight non-flow-limited postvascular surgery patients and eight flow-limited COPD patients. Patients were sedated, paralyzed for diagnostic purposes, and ventilated in volume control ventilation with constant inspiratory flow rate. Data on resistance, compliance, and dynamic intrinsic positive end-expiratory pressure (PEEPi,dyn) obtained by applying the least squares fitting method during inspiration, expiration, and the overall breathing cycle were compared with those obtained by the traditional method (constant flow, end-inspiratory occlusion method). Our results indicate that (a) the presence of EFL markedly decreases the precision of resistance and compliance values measured by the LSF method, (b) the determination of respiratory variables during inspiration allows the calculation of respiratory mechanics in flow limited COPD patients, and (c) the LSF method is able to detect the presence of PEEPi,dyn if only inspiratory data are used.

  2. Work of breathing during lung-protective ventilation in patients with acute lung injury and acute respiratory distress syndrome: a comparison between volume and pressure-regulated breathing modes.

    PubMed

    Kallet, Richard H; Campbell, Andre R; Dicker, Rochelle A; Katz, Jeffrey A; Mackersie, Robert C

    2005-12-01

    Pressure-control ventilation (PCV) and pressure-regulated volume-control (PRVC) ventilation are used during lung-protective ventilation because the high, variable, peak inspiratory flow rate (V (I)) may reduce patient work of breathing (WOB) more than the fixed V (I) of volume-control ventilation (VCV). Patient-triggered breaths during PCV and PRVC may result in excessive tidal volume (V(T)) delivery unless the inspiratory pressure is reduced, which in turn may decrease the peak V (I). We tested whether PCV and PRVC reduce WOB better than VCV with a high, fixed peak V (I) (75 L/min) while also maintaining a low V(T) target. Fourteen nonconsecutive patients with acute lung injury or acute respiratory distress syndrome were studied prospectively, using a random presentation of ventilator modes in a crossover, repeated-measures design. A target V(T) of 6.4 + 0.5 mL/kg was set during VCV and PRVC. During PCV the inspiratory pressure was set to achieve the same V(T). WOB and other variables were measured with a pulmonary mechanics monitor (Bicore CP-100). There was a nonsignificant trend toward higher WOB (in J/L) during PCV (1.27 + 0.58 J/L) and PRVC (1.35 + 0.60 J/L), compared to VCV (1.09 + 0.59 J/L). While mean V(T) was not statistically different between modes, in 40% of patients, V(T) markedly exceeded the lung-protective ventilation target during PRVC and PCV. During lung-protective ventilation, PCV and PRVC offer no advantage in reducing WOB, compared to VCV with a high flow rate, and in some patients did not allow control of V(T) to be as precise.

  3. Pretest Predictions for Ventilation Tests

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Y. Sun; H. Yang; H.N. Kalia

    The objective of this calculation is to predict the temperatures of the ventilating air, waste package surface, concrete pipe walls, and insulation that will be developed during the ventilation tests involving various test conditions. The results will be used as input to the following three areas: (1) Decisions regarding testing set-up and performance. (2) Assessing how best to scale the test phenomena measured. (3) Validating numerical approach for modeling continuous ventilation. The scope of the calculation is to identify the physical mechanisms and parameters related to thermal response in the ventilation tests, and develop and describe numerical methods that canmore » be used to calculate the effects of continuous ventilation. Sensitivity studies to assess the impact of variation of linear power densities (linear heat loads) and ventilation air flow rates are included. The calculation is limited to thermal effect only.« less

  4. Visualizing dissolved oxygen transport for liquid ventilation in an in vitro model of the human airways

    NASA Astrophysics Data System (ADS)

    Janke, T.; Bauer, K.

    2017-04-01

    Up until to now, the measurement of dissolved oxygen concentrations during liquid ventilation is limited to the determination of averaged concentrations of the liquid entering or leaving the body. The work presented in this paper aims to extend the possible measurement techniques in the research of liquid ventilation. Therefore optical measurements of the dissolved oxygen concentration, using a luminescent sensor dye, are performed. The preparation of a suitable sensor liquid, based on the metal complex Dichlorotris(1,10)-(phenanthroline)ruthenium(II), is presented. A transparent simplified human lung geometry is used for conducting the experiments. Inspiratory as well as expiratory flow at three different constant flow rates is investigated, covering the flow regimes \\text{Re}=83 -333 and \\text{Pe}=33 300 -133 000. The applied measurement technique is capable to reveal distinctive concentration patterns during inspiration and expiration caused by the laminar flow characteristics. Allowing a sufficiently long flow duration, local concentration inhomogeneities disappear and an exponential rise and decay of the mean values can be observed for inspiration and expiration.

  5. Inhaled Epoprostenol Through Noninvasive Routes of Ventilator Support Systems.

    PubMed

    Ammar, Mahmoud A; Sasidhar, Madhu; Lam, Simon W

    2018-06-01

    The administration of inhaled epoprostenol (iEPO) through noninvasive routes of ventilator support systems has never been previously evaluated. Describe the use of iEPO when administered through noninvasive routes of ventilator support systems. Critically ill patients admitted to the intensive care unit who received iEPO through noninvasive routes were analyzed. Improvements in respiratory status and hemodynamic parameters were evaluated. Safety end points assessed included hypotension, rebound hypoxemia, significant bleeding, and thrombocytopenia. A total of 36 patients received iEPO through noninvasive routes: high-flow oxygen therapy through nasal cannula, n = 29 (81%) and noninvasive positive-pressure ventilation, n = 7 (19%). Sixteen patients had improvement in their respiratory status: mean decrease in fraction of inspired oxygen (FiO 2 ), 20% ± 13%; mean increase in partial pressure of arterial oxygen to FiO 2 (PaO 2 /FiO 2 ) ratio, 60 ± 50 mm Hg; and mean decrease in HFNC oxygen flow rate, 6 ± 3 liters per minute (LPM). Eight patients had declines in their respiratory status (mean increase in FiO 2 , 30% ± 20%; mean decrease in PaO 2 /FiO 2 ratio, 38 ± 20 mm Hg; and mean increase in HFNC oxygen flow rate, 15 ± 10 LPM), and 12 patients had no change in their respiratory status. Conclusion and Relevance: This represents the first evaluation of the administration of iEPO through noninvasive routes of ventilator support systems and demonstrates that in critically ill patients, iEPO could be administered through a noninvasive route. Further evaluation is needed to determine the extent of benefit with this route of administration.

  6. Experimental and numerical investigation of hydro power generator ventilation

    NASA Astrophysics Data System (ADS)

    Jamshidi, H.; Nilsson, H.; Chernoray, V.

    2014-03-01

    Improvements in ventilation and cooling offer means to run hydro power generators at higher power output and at varying operating conditions. The electromagnetic, frictional and windage losses generate heat. The heat is removed by an air flow that is driven by fans and/or the rotor itself. The air flow goes through ventilation channels in the stator, to limit the electrical insulation temperatures. The temperature should be kept limited and uniform in both time and space, avoiding thermal stresses and hot-spots. For that purpose it is important that the flow of cooling air is distributed uniformly, and that flow separation and recirculation are minimized. Improvements of the air flow properties also lead to an improvement of the overall efficiency of the machine. A significant part of the windage losses occurs at the entrance of the stator ventilation channels, where the air flow turns abruptly from tangential to radial. The present work focuses exclusively on the air flow inside a generator model, and in particular on the flow inside the stator channels. The generator model design of the present work is based on a real generator that was previously studied. The model is manufactured taking into consideration the needs of both the experimental and numerical methodologies. Computational Fluid Dynamics (CFD) results have been used in the process of designing the experimental setup. The rotor and stator are manufactured using rapid-prototyping and plexi-glass, yielding a high geometrical accuracy, and optical experimental access. A special inlet section is designed for accurate air flow rate and inlet velocity profile measurements. The experimental measurements include Particle Image Velocimetry (PIV) and total pressure measurements inside the generator. The CFD simulations are performed based on the OpenFOAM CFD toolbox, and the steady-state frozen rotor approach. Specific studies are performed, on the effect of adding "pick-up" to spacers, and the effects of the inlet fan blades on the flow rate through the model. The CFD results capture the experimental flow details to a reasonable level of accuracy.

  7. Non-invasive ventilation: evaluation of CO2 washout by intentional leaking in three recent oronasal masks. A pilot study.

    PubMed

    Medrinal, C; Prieur, G; Contal, O; Villiot-Danger, E; Doyle, C; Reychler, G; Quieffin, J

    2015-05-01

    Many types of interfaces with intentional leaks exist for Non Invasive Ventilation. The purpose of intentional leaks is to remove CO2 from the interface, however the calibration does not allow a sufficiently large flow and rebreathing of CO2 can occur. The aim of this study was to compare the CO2 rinsing capacities of three new generation oronasal masks with intentional leaks (A: Quattro®, [Resmed]; B: Amara® [Respironics]; C: Forma® [Fisher&Paykel]) in healthy subjects. Seventeen healthy volunteers were included in this prospective cross-sectional, randomized, double-blinded trial. Each subject underwent ventilation with a home ventilator (IPAP: 14 cmH2O; EPAP: 4 cmH2O) with each mask consecutively. Transcutaneous capnography (PtcCO2) recordings were carried out throughout the trial and ventilator data (tidal volume, respiratory rate, minute ventilation and unintentional leaks) were also analyzed. Mask comfort was assessed using a visual analog scale (0 to 10). The results showed no differences in PtcCO2 between masks (P=0.82). There were no significant differences in respiratory parameters (tidal volume, P=0.79; respiratory rate, P=0.65; minute ventilation, P=0.12) between masks. The rate of unintentional leaks were significantly lower for Mask A (P=0.016). Subjects rated Mask A and Mask C as more comfortable than Mask B (P=0.041). There was no effect of mask on PtcCO2 in healthy subjects. The mask with the highest comfort rating had not the lowest rate of unintentional leaks.

  8. Exploration Mission Particulate Matter Filtration Technology Performance Testing in a Simulated Spacecraft Cabin Ventilation System

    NASA Technical Reports Server (NTRS)

    Agui, Juan H.; Vijayakumar, R.; Perry, Jay L.; Frederick, Kenneth R.; Mccormick, Robert M.

    2017-01-01

    Human deep space exploration missions will require advances in long-life, low maintenance airborne particulate matter filtration technology. As one of the National Aeronautics and Space Administrations (NASA) developments in this area, a prototype of a new regenerable, multi-stage particulate matter filtration technology was tested in an International Space Station (ISS) module simulation facility. As previously reported, the key features of the filter system include inertial and media filtration with regeneration and in-place media replacement techniques. The testing facility can simulate aspects of the cabin environment aboard the ISS and contains flight-like cabin ventilation system components. The filtration technology test article was installed at the inlet of the central ventilation system duct and instrumented to provide performance data under nominal flow conditions. In-place regeneration operations were also evaluated. The real-time data included pressure drop across the filter stages, process air flow rate, ambient pressure, humidity and temperature. In addition, two video cameras positioned at the filtration technology test articles inlet and outlet were used to capture the mechanical performance of the filter media indexing operation under varying air flow rates. Recent test results are presented and future design recommendations are discussed.

  9. Hypoxic pulmonary vasoconstriction does not affect hydrostatic pulmonary edema formation.

    PubMed

    Cheney, F W; Bishop, M J; Eisenstein, B L; Artman, L D

    1987-02-01

    We studied the effects of regional hypoxic pulmonary vasoconstriction (HPV) on lobar flow diversion in the presence of hydrostatic pulmonary edema. Ten anesthetized dogs with the left lower lobe (LLL) suspended in a net for continuous weighing were ventilated with a bronchial divider so the LLL could be ventilated with either 100% O2 or a hypoxic gas mixture (90% N2-5% CO2-5% O2). A balloon was inflated in the left atrium until hydrostatic pulmonary edema occurred, as evidenced by a continuous increase in LLL weight. Left lower lobe flow (QLLL) was measured by electromagnetic flow meter and cardiac output (QT) by thermal dilution. At a left atrial pressure of 30 +/- 5 mmHg, ventilation of the LLL with the hypoxic gas mixture caused QLLL/QT to decrease from 17 +/- 4 to 11 +/- 3% (P less than 0.05), pulmonary arterial pressure to increase from 35 +/- 5 to 37 +/- 6 mmHg (P less than 0.05), and no significant change in rate of LLL weight gain. Gravimetric confirmation of our results was provided by experiments in four animals where the LLL was ventilated with an hypoxic gas mixture for 2 h while the right lung was ventilated with 100% O2. In these animals there was no difference in bloodless lung water between the LLL and right lower lobe. We conclude that in the presence of left atrial pressures high enough to cause hydrostatic pulmonary edema, HPV causes significant flow diversion from an hypoxic lobe but the decrease in flow does not affect edema formation.

  10. Predictors of need for noninvasive ventilation during respiratory tract infections in medically stable, non-ventilated subjects with amyotrophic lateral sclerosis.

    PubMed

    Sancho, Jesus; Servera, Emilio; Bañuls, Pilar; Marin, Julio

    2015-04-01

    Acute lower respiratory infections can impair muscle strength in patients with amyotrophic lateral sclerosis (ALS). When associated with an increase in load on the respiratory system, this situation may precipitate hypercapnic respiratory failure in non-ventilated patients with ALS. The aim of this study was to determine whether a clinical or functional parameter can predict the need for noninvasive ventilation (NIV) during an acute respiratory infection for medically stable, non-ventilated patients with ALS. This was a prospective study involving all non-ventilated subjects with ALS admitted due to an acute respiratory infection to a respiratory care unit from a tertiary hospital. Thirty-two non-ventilated subjects with ALS were admitted to our respiratory care unit due to an acute respiratory infection: 60.72 ± 10.54 y, 13 males, 23 with spinal onset, FVC of 1.58 ± 0.83 L, FVC of 56.21 ± 23.15% of predicted, peak cough flow of 3.41 ± 1.77 L/s, maximum insufflation capacity of 1.87 ± 0.94 L, revised Amyotrophic Lateral Sclerosis Functional Rating Scale score of 22.80 ± 8.83, and Norris bulbar score of 23.48 ± 12.14. Fifteen subjects required NIV during the episode. Logistic regression analysis showed that the only predictors of need for NIV were percent-of-predicted FVC (odds ratio of 1.06, 95% CI 1.01-1.11, P = .02) and peak cough flow (odds ratio of 2.57, 95% CI 1.18-5.59, P = .02). In medically stable, non-ventilated patients with ALS, measurement of percent-of-predicted FVC and peak cough flow can predict the need for NIV during an acute lower respiratory tract infection. Copyright © 2015 by Daedalus Enterprises.

  11. Record rates of pressurized gas-flow in the great horsetail, Equisetum telmateia. Were Carboniferous Calamites similarly aerated?

    PubMed

    Armstrong, Jean; Armstrong, William

    2009-01-01

    Significant pressurized (convective) ventilation has been demonstrated in some flowering wetland plants, for example water-lilies and reeds, but not previously in nonflowering plants. Here we investigated convective flows in the great horsetail, Equisetum telmateia, and the possibility that convections aerated the massive rhizomes of the Calamites, extinct giant horsetails of the Carboniferous. Convection in E. telmateia was examined in relation to induction sites, anatomical pathways, relative humidity (RH), external wind-speed, diurnal effects, rhizome resistance and pressure-gradients. A mathematical model, incorporating Calamite aeration anatomy, was applied in assessing potentials for convective aeration. Individual shoots of E. telmateia generated extremely high rates of humidity-induced convection: < or = 120 cm(3) min(-1) (internal wind-velocity: 10 cm s(-1)) with rates proportional to branch numbers and 1/RH. Flows passed through branches, stem and rhizome via low-resistance lacunae (vallecular canals) and vented via stubble. Stomata supported internal pressures up to 800 Pa. Anatomically, E. telmateia resembles the Calamites and modelling predicted possible flows of 70 l min(-1) per Calamite tree. This is the first demonstration of significant convective flow in a nonflowering species, indicating that plant ventilation by a type of 'molecular gas-pump' may date back 350 million yr or more. Stomatal form and low-resistance pathways may facilitate high flow rates.

  12. In-flight cabin smoke control.

    PubMed

    Eklund, T I

    1996-12-31

    Fatal accidents originating from in-flight cabin fires comprise only about 1% of all fatal accidents in the civil jet transport fleet. Nevertheless, the impossibility of escape during flight accentuates the hazards resulting from low visibility and toxic gases. Control of combustion products in an aircraft cabin is affected by several characteristics that make the aircraft cabin environment unique. The aircraft fuselage is pressurized in flight and has an air distribution system which provides ventilation jets from the ceiling level air inlets running along the cabin length. A fixed quantity of ventilation air is metered into the cabin and air discharge is handled primarily by pressure controlling outflow valves in the rear lower part of the fuselage. Earlier airplane flight tests on cabin smoke control used generators producing minimally buoyant smoke products that moved with and served as a telltales for overall cabin ventilation flows. Analytical studies were done with localized smoke production to predict the percent of cabin length that would remain smoke-free during continuous generation. Development of a buoyant smoke generator allowed simulation of a fire plume with controllable simulated temperature and heat release rates. Tests on a Boeing 757, modified to allow smoke venting out through the top of the cabin, showed that the buoyant smoke front moved at 0.46m/s (1.5ft/sec) with and 0.27m/sec (0.9ft/sec) against, the axial ventilation airflow. Flight tests in a modified Boeing 727 showed that a ceiling level counterflow of about 0.55m/sec (1.8ft/sec) was required to arrest the forward movement of buoyant smoke. A design goal of 0.61m/s (2ft/sec) axial cabin flow would require a flow rate of 99m3/min (3500ft3/min) in a furnished Boeing 757. The current maximum fresh air cabin ventilation flow is 78m3/min (2756 ft3/min). Experimental results indicate that buoyancy effects cause smoke movement behaviour that is not predicted by traditional design analyses and flight test methodologies. Augmenting available ventilation for smoke control remains a design and safety challenge.

  13. Gravity predominates over ventilatory pattern in the prevention of ventilator-associated pneumonia.

    PubMed

    Li Bassi, Gianluigi; Marti, Joan Daniel; Saucedo, Lina; Rigol, Montserrat; Roca, Ignasi; Cabanas, Maria; Muñoz, Laura; Ranzani, Otavio Tavares; Giunta, Valeria; Luque, Nestor; Esperatti, Mariano; Gabarrus, Albert; Fernandez, Laia; Rinaudo, Mariano; Ferrer, Miguel; Ramirez, Jose; Vila, Jordi; Torres, Antoni

    2014-09-01

    In the semirecumbent position, gravity-dependent dissemination of pathogens has been implicated in the pathogenesis of ventilator-associated pneumonia. We compared the preventive effects of a ventilatory strategy, aimed at decreasing pulmonary aspiration and enhancing mucus clearance versus the Trendelenburg position. Prospective randomized animal study. Animal research facility, University of Barcelona, Spain. Twenty-four Large White-Landrace pigs. Pigs were intubated and on mechanical ventilation for 72 hours. Following surgical preparation, pigs were randomized to be positioned: 1) in semirecumbent/prone position, ventilated with a duty cycle (TITTOT) of 0.33 and without positive end-expiratory pressure (control); 2) as in the control group, positive end-expiratory pressure of 5 cm H2O and TITTOT to achieve a mean expiratory-inspiratory flow bias of 10 L/min (treatment); 3) in Trendelenburg/prone position and ventilated as in the control group (Trendelenburg). Following randomization, Pseudomonas aeruginosa was instilled into the oropharynx. Mucus clearance rate was measured through fluoroscopic tracking of tracheal markers. Microspheres were instilled into the subglottic trachea to assess pulmonary aspiration. Ventilator-associated pneumonia was confirmed by histological/microbiological studies. The mean expiratory-inspiratory flow in the treatment, control, and Trendelenburg groups were 10.7 ± 1.7, 1.8 ± 3.7 and 4.3 ± 2.8 L/min, respectively (p < 0.001). Mucus clearance rate was 11.3 ± 9.9 mm/min in the Trendelenburg group versus 0.1 ± 1.0 in the control and 0.2 ± 1.0 in the treatment groups (p = 0.002). In the control group, we recovered 1.35% ± 1.24% of the instilled microspheres per gram of tracheal secretions, whereas 0.22% ± 0.25% and 0.97% ± 1.44% were recovered in the treatment and Trendelenburg groups, respectively (p = 0.031). Ventilator-associated pneumonia developed in 66.67%, 85.71%, and 0% of the animals in the control, treatment, and Trendelenburg groups (p < 0.001). The Trendelenburg position predominates over expiratory flow bias and positive end-expiratory pressure in the prevention of gravity-dependent translocation of oropharyngeal pathogens and development of ventilator-associated pneumonia. These findings further substantiate the primary role of gravity in the pathogenesis of ventilator-associated pneumonia.

  14. [Spontaneous ventilation in positive expiratory pressure in cardiogenic pulmonary edema. Prospective study].

    PubMed

    Bouquin, V; L'Her, E; Moriconi, M; Jobic, Y; Maheu, B; Guillo, P; Paris, A; Pennec, P Y; Boles, J M; Blanc, J J

    1998-10-01

    New equipment facilitating the use of spontaneous ventilation with positive expiratory pressure (PEP) has become available in France since January 1996. This technique was applied in 38 patients with severe cardiogenic pulmonary oedema and persistent respiratory distress despite high flow classical oxygen therapy and standard treatment. After 1 hour of ventilation with a flow of 220 l/min of 100% oxygen with an average PEP of 7.7 cm H20, a significant improvement of clinical (heart and respiratory rate) and biological parameters (arterial gases) was observed. There were no side effects. Four patients died during the hospital period and only 1 was intubated. Spontaneous ventilation with PEP is a simple technique for coronary care units and, compared with conventional oxygen therapy, it rapidly improves arterial oxygenation, reduces respiratory work and improves conditions of cardiac load. Acute severe cardiogenic pulmonary oedema seems to be an indication of choice, especially in the elderly, where it may help avoid an often controversial intubation.

  15. Effect of hypoxic breathing on cutaneous temperature recovery in man

    NASA Astrophysics Data System (ADS)

    Fahim, Mohammad

    1992-03-01

    Effect of hypoxia (12% O2) on skin temperature recovery was studied on healthy young men. Forty male volunteers free of any respiratory disorder were randomly selected to participate in the study. Skin temperature, peripheral blood flow, heart rate and end expiratory PO2 and PCO2 were measured. During hyoxic ventilation the peripheral blood flow was reduced and a corresponding drop in skin temperature occurred. This was partly due to hyperventilation associated with hypoxic ventilation. The recovery of skin temperature after cooling the hand for 2 min in cold water (10 12° C) took 5.5±0.1 min during normal air breathing; during hypoxic ventilation even after 9.1±0.3 min when the skin temperature recovery curve plateaued, the skin temperature remained about 2° C below control. The results of the present investigation indicate that hypoxia interferes with the normal functioning of the thermoregulatory mechanism in man. Hyperventilation associated with hypoxic ventilation is also partly responsible for incomplete recovery of skin temperature.

  16. Effect of stratified inequality of blood flow on gas exchange in liquid-filled lungs.

    NASA Technical Reports Server (NTRS)

    West, J. B.; Maloney, J. E.; Castle, B. L.

    1972-01-01

    This investigation set out to answer two questions: (1) are the distal alveoli in the terminal lung units less well perfused than the proximal alveoli, i.e., is there stratification of blood flow; and (2) if so, does this enhance gas exchange in the presence of stratified inequality of ventilation. Excised dog lungs were ventilated with saline and perfused with blood. Following single inspirations of xenon 133 in saline and various periods of breath holding, the expired xenon concentration against volume was measured and it confirmed marked stratified inequality of ventilation under these conditions. By measuring the rate of depletion of xenon from alveoli during a period of blood flow, we showed that the alveoli which emptied at the end of expiration had 16% less blood flow than those exhaling earlier. However, by measuring the xenon concentration in pulmonary venous blood, we found that about 10% less tracer was transferred from the alveoli into the blood when the inspired xenon was stratified within the respiratory zone. Thus while stratification of blood flow was confirmed, it was shown to impair rather than enhance the efficiency of gas transfer.

  17. Natural ventilation of buildings: opposing wind and buoyancy

    NASA Astrophysics Data System (ADS)

    Linden, Paul; Hunt, Gary

    1998-11-01

    The use of natural ventilation in buildings is an attractive way to reduce energy usage thereby reducing costs and CO2 emissions. Generally, it is necessary to remove excess heat from a building and the designer can use the buoyancy forces associated with the above ambient temperatures within the building to drive a flow - 'stack' ventilation. The most efficient mode is displacement ventilation where warm air accumulates near the top of the building and flows out through upper level vents and cooler air flows in at lower levels. Ventilation will also be driven between these lower and upper openings by the wind. We report on laboratory modeling and theory which investigates the effects of an opposing wind on stack ventilation driven by a constant source of heat within a space under displacement ventilation. We show that there is a critical wind speed, expressed in dimensionless terms as a critical Froude number, above which displacement ventilation is replaced by (less efficient) mixing ventilation with reversed flow. Below this critical speed, displacement ventilation, in which the interior has a two-layer stratification, is maintained. The criterion for the change in ventilation mode is derived from general considerations of mixing efficiencies in stratified flows. We conclude that even when wind effects might appear to be dominant, the inhibition of mixing by the stable stratification within the space ensures that stack ventilation can operate over a wide range of apparently adverse conditions.

  18. [The aspiratory resistance and filtration penetration of N95 filtering-facepiece respirators used widely in China].

    PubMed

    Wang, Xinyan; Shi, Tingming; Lu, Wei; Qin, Shaoxian; Liu, Yuewei; Tao, Ying; Zhang, Hongge; Chen, Weihong

    2015-01-01

    The objective of this study was to investigate the aspiratory resistance, filtration penetration and their influence factors of N95 filtering-facepiece respirators used widely in China. The total of 6 brands and 21 models of N95 filtering-facepiece respirators which are certified and big sales on the market. The aspiratory resistance and filtration efficiency filter penetration were measured while air pump ran from 10 L/min to 100 L/min using differential pressure gauge and the PortaCount, respectively. The filtration penetrations for 2 of the 21 models were lower than 95%, and the qualified rate for all models was 90.47%. The filtration penetrations gradually decreased when ventilation flow of air pump increased. The negative correlation was observed between filtration penetration and ventilation flow (r(2) = 0.711, P < 0.05). The resistances of all 21 models of N95 respirators met the requirements of the national standard. The aspiratory resistance started to elevate with the increasing of ventilation flow, and a positive correlation between both (r(2) = 0.878, P < 0.05). Significant differences of filtration penetration and aspiratory resistance were observed among between different brands (P < 0.05) although no differences of filtration penetration existed among different models of one brand (P > 0.05). But the differences of the aspiratory resistance among different models of one brand were statistically significant (P < 0.05). The aspiratory resistances of all N95 filtering-facepiece respirators used in this study met the requirements of the national standard. And the qualified ratio of filtration penetration of all models was higher than 90%. The influencing factors of aspiratory resistance included materials, size and ventilation flow. And influencing factors for filtration penetration were materials and ventilation flow.

  19. Are we fully utilizing the functionalities of modern operating room ventilators?

    PubMed

    Liu, Shujie; Kacmarek, Robert M; Oto, Jun

    2017-12-01

    The modern operating room ventilators have become very sophisticated and many of their features are comparable with those of an ICU ventilator. To fully utilize the functionality of modern operating room ventilators, it is important for clinicians to understand in depth the working principle of these ventilators and their functionalities. Piston ventilators have the advantages of delivering accurate tidal volume and certain flow compensation functions. Turbine ventilators have great ability of flow compensation. Ventilation modes are mainly volume-based or pressure-based. Pressure-based ventilation modes provide better leak compensation than volume-based. The integration of advanced flow generation systems and ventilation modes of the modern operating room ventilators enables clinicians to provide both invasive and noninvasive ventilation in perioperative settings. Ventilator waveforms can be used for intraoperative neuromonitoring during cervical spine surgery. The increase in number of new features of modern operating room ventilators clearly creates the opportunity for clinicians to optimize ventilatory care. However, improving the quality of ventilator care relies on a complete understanding and correct use of these new features. VIDEO ABSTRACT: http://links.lww.com/COAN/A47.

  20. Methane emissions and airflow patterns on a longwall face: Potential influences from longwall gob permeability distributions on a bleederless longwall panel.

    PubMed

    Schatzel, S J; Krog, R B; Dougherty, H

    2017-01-01

    Longwall face ventilation is an important component of the overall coal mine ventilation system. Increased production rates due to higher-capacity mining equipment tend to also increase methane emission rates from the coal face, which must be diluted by the face ventilation. Increases in panel length, with some mines exceeding 6,100 m (20,000 ft), and panel width provide additional challenges to face ventilation designs. To assess the effectiveness of current face ventilation practices at a study site, a face monitoring study with continuous monitoring of methane concentrations and automated recording of longwall shearer activity was combined with a tracer gas test on a longwall face. The study was conducted at a U.S. longwall mine operating in a thick, bituminous coal seam and using a U-type, bleederless ventilation system. Multiple gob gas ventholes were located near the longwall face. These boreholes had some unusual design concepts, including a system of manifolds to modify borehole vacuum and flow and completion depths close to the horizon of the mined coalbed that enabled direct communication with the mine atmosphere. The mine operator also had the capacity to inject nitrogen into the longwall gob, which occurred during the monitoring study. The results show that emission rates on the longwall face showed a very limited increase in methane concentrations from headgate to tailgate despite the occurrence of methane delays during monitoring. Average face air velocities were 3.03 m/s (596 fpm) at shield 57 and 2.20 m/s (433 fpm) at shield 165. The time required for the sulfur hexafluoride (SF 6 ) peak to occur at each monitoring location has been interpreted as being representative of the movement of the tracer slug. The rate of movement of the slug was much slower in reaching the first monitoring location at shield 57 compared with the other face locations. This lower rate of movement, compared with the main face ventilation, is thought to be the product of a flow path within and behind the shields that is moving in the general direction of the headgate to the tailgate. Barometric pressure variations were pronounced over the course of the study and varied on a diurnal basis.

  1. Evaluation of cough peak expiratory flow as a predictor of successful mechanical ventilation discontinuation: a narrative review of the literature.

    PubMed

    Jiang, Chuan; Esquinas, Antonio; Mina, Bushra

    2017-01-01

    A crucial step in the transition from mechanical ventilation to extubation is the successful performance of a spontaneous breathing trial (SBT). The American College of Chest Physicians (ACCP) Guidelines recommend removal of the endotracheal tube upon successful completion of a SBT. However, this does not guarantee successful extubation as there remains a risk of re-intubation. Guidelines have outlined ventilator liberation protocols, selected use of non-invasive ventilation on extubation, early mobilization, and dynamic ventilator metrics to prevent and better predict extubation failure. However, a significant percentage of patients still fail mechanical ventilation discontinuation. A common reason for re-intubation is having a weak cough strength, which reflects the inability to protect the airway. Evaluation of cough strength via objective measures using peak expiratory flow rate is a non-invasive and easily reproducible assessment which can predict extubation failure. We conducted a narrative review of the literature regarding use of cough strength as a predictive index for extubation failure risk. Results of our review show that cough strength, quantified objectively with a cough peak expiratory flow measurement (CPEF), is strongly associated with extubation success. Furthermore, various cutoff thresholds have been identified and can provide reasonable diagnostic accuracy and predictive power for extubation failure. These results demonstrate that measurement of the CPEF can be a useful tool to predict extubation failure in patients on MV who have passed a SBT. In addition, the data suggest that this diagnostic modality may reduce ICU length of stay, ICU expenditures, and morbidity and mortality.

  2. Building America Case Study: Sealed Crawl Spaces with Integrated Whole-House Ventilation in a Cold Climate, Ithaca, New York

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    "9One method of code-compliance for crawlspaces is to seal and insulate the crawlspace, rather than venting to the outdoors. However, codes require mechanical ventilation; either via conditioned supply air from the HVAC system, or a continuous exhaust ventilation strategy. As the CARB's building partner, Ithaca Neighborhood Housing Services, intended to use the unvented crawlspace in a recent development, CARB was interested in investigating a hybrid ventilation method that includes the exhaust air from the crawlspace as a portion of an ASHRAE 62.2 compliant whole-house ventilation strategy. This hybrid ventilation method was evaluated through a series of long-term monitoring tests thatmore » observed temperature, humidity, and pressure conditions through the home and crawlspace. Additionally, CARB worked with NREL to perform multi-point tracer gas testing on six separate ventilation strategies - varying portions of 62.2 required flow supplied by the crawlspace fan and an upstairs bathroom fan. The intent of the tracer gas testing was to identify effective Reciprocal Age of Air (RAoA), which is equivalent to the air change rate in well-mixed zones, for each strategy while characterizing localized infiltration rates in several areas of the home.« less

  3. Purging of working atmospheres inside freight containers.

    PubMed

    Braconnier, Robert; Keller, François-Xavier

    2015-06-01

    This article focuses on prevention of possible exposure to chemical agents, when opening, entering, and stripping freight containers. The container purging process is investigated using tracer gas measurements and numerical airflow simulations. Three different container ventilation conditions are studied, namely natural, mixed mode, and forced ventilation. The tests conducted allow purging time variations to be quantified in relation to various factors such as container size, degree of filling, or type of load. Natural ventilation performance characteristics prove to be highly variable, depending on environmental conditions. Use of a mechanically supplied or extracted airflow under mixed mode and forced ventilation conditions enables purging to be significantly accelerated. Under mixed mode ventilation, extracting air from the end of the container furthest from the door ensures quicker purging than supplying fresh air to this area. Under forced ventilation, purging rate is proportional to the applied ventilation flow. Moreover, purging rate depends mainly on the location at which air is introduced: the most favourable position being above the container loading level. Many of the results obtained during this study can be generalized to other cases of purging air in a confined space by general ventilation, e.g. the significance of air inlet positioning or the advantage of generating high air velocities to maximize stirring within the volume. © The Author 2015. Published by Oxford University Press on behalf of the British Occupational Hygiene Society.

  4. Numerical simulation and comparison of two ventilation methods for a restaurant - displacement vs mixed flow ventilation

    NASA Astrophysics Data System (ADS)

    Chitaru, George; Berville, Charles; Dogeanu, Angel

    2018-02-01

    This paper presents a comparison between a displacement ventilation method and a mixed flow ventilation method using computational fluid dynamics (CFD) approach. The paper analyses different aspects of the two systems, like the draft effect in certain areas, the air temperatureand velocity distribution in the occupied zone. The results highlighted that the displacement ventilation system presents an advantage for the current scenario, due to the increased buoyancy driven flows caused by the interior heat sources. For the displacement ventilation case the draft effect was less prone to appear in the occupied zone but the high heat emissions from the interior sources have increased the temperature gradient in the occupied zone. Both systems have been studied in similar conditions, concentrating only on the flow patterns for each case.

  5. Laboratory and Physical Modelling of Building Ventilation Flows

    NASA Astrophysics Data System (ADS)

    Hunt, Gary

    2001-11-01

    Heating and ventilating buildings accounts for a significant fraction of the total energy budget of cities and an immediate challenge in building physics is for the design of sustainable, low-energy buildings. Natural ventilation provides a low-energy solution as it harness the buoyancy force associated with temperature differences between the internal and external environment, and the wind to drive a ventilating flow. Modern naturally-ventilated buildings use innovative design solutions, e.g. glazed atria and solar chimneys, to enhance the ventilation and demand for these and other designs has far outstripped our understanding of the fluid mechanics within these buildings. Developing an understanding of the thermal stratification and movement of air provides a considerable challenge as the flows involve interactions between stratification and turbulence and often in complex geometries. An approach that has provided significant new insight into these flows and which has led to the development of design guidelines for architects and ventilation engineers is laboratory modelling at small-scale in water tanks combined with physical modelling. Density differences to drive the flow in simplified plexiglass models of rooms or buildings are provided by fresh and salt water solutions, and wind flow is represented by a mean flow in a flume tank. In tandom with the experiments, theoretical models that capture the essential physics of these flows have been developed in order to generalise the experimental results to a wide range of typical building geometries and operating conditions. This paper describes the application and outcomes of these modelling techniques to the study of a variety of natural ventilation flows in buildings.

  6. Hollow Fiber Ground Evaporator Unit Testing

    NASA Technical Reports Server (NTRS)

    Bue, Grant; Trevino, Luis; Tsioulos, Gus

    2010-01-01

    A candidate technology for 1-atmosphere suited heat rejection was developed and tested at NASA Johnson Space Center. The concept is to use a collection of microporous hydrophobic tubes potted between inlet and outlet headers with water as coolant. A pump provides flow between headers through the tubes which are subjected to fan driven cross flow of relatively dry air. The forced ventilation would sweep out the water vapor from the evaporation of the coolant rejecting heat from the coolant stream. The hollow fibers are obtained commercially (X50-215 Celgard) which are arranged in a sheet containing 5 fibers per linear inch. Two engineering development units were produced that vary the fold direction of the fiber sheets relative to the ventilation. These units were tested at inlet water temperatures ranging from 20 deg C to 30 deg C, coolant flow rates ranging from 10 to 90 kg/hr, and at three fan speeds. These results were used to size a system that could reject heat at a rate of 340 W.

  7. Evaluation of carbon dioxide rebreathing during exercise assisted by noninvasive ventilation with plateau exhalation valve

    PubMed Central

    Ou, Yong-er; Lin, Zhi-min; Hua, Dong-ming; Jiang, Ying; Huo, Ya-ting; Luo, Qun; Chen, Rong-Chang

    2017-01-01

    Noninvasive ventilation with a plateau exhalation valve (PEV) is often used as an adjunct to exercise to achieve a physiologic training effect in severe chronic obstructive pulmonary disease (COPD) patients. However, during exercise, with the increase of exhalation flow and respiratory rate and limited capability of PEV to exhale gases out of the circuit, it is still unknown whether CO2 rebreathing occurs in COPD patients ventilated during exercise assisted by single-limb circuit with a PEV. A maximal symptom-limited cycle exercise test was performed while ventilated on pressure support (inspiratory:expiratory pressure 14:4 cmH2O) in 18 male patients with stable severe COPD (mean ± standard deviation, forced expiratory volume in 1 s: 29.5%±6.9% predicted). At rest and during exercise, breathing pattern, mean expiratory flow, mean expiratory flow of PEV, and the mean inspiratory fraction of CO2 (tidal fractional concentration of inspired CO2 [FiCO2]) reinsufflated from the circuit was measured for each breath. In comparison with rest, with the significant increase of mean expiratory flow (0.39±0.15 vs 0.82±0.27 L/s), fractional concentration of end-tidal CO2 (2.6%±0.7% vs 5.5%±0.6%), and the significant decrease of mean expiratory flow of PEV (0.41±0.02 vs 0.39±0.03 L/s), tidal FiCO2 significantly increased at peak exercise (0.48%±0.19% vs 1.8%±0.6%) in patients with stable severe COPD. The inflection point of obvious CO2 rebreathing was 0.67±0.09 L/s (95% confidence interval 0.60–0.73 L/s). Ventilated by a single-limb tubing with PEV caused CO2 rebreathing to COPD patients during exercise. Patients with mean expiratory flow >0.60–0.73 L/s may be predisposed to a higher risk of CO2 rebreathing. PMID:28144134

  8. Experimental evaluation of the Skylab orbital workshop ventilation system concept

    NASA Technical Reports Server (NTRS)

    Allums, S. L.; Hastings, L. J.; Ralston, J. T.

    1972-01-01

    Extensive testing was conducted to evaluate the Orbital Workshop ventilation concept. Component tests were utilized to determine the relationship between operating characteristics at 1 and 0.34 atm. System tests were conducted at 1 atm within the Orbital Workshop full-scale mockup to assess delivered volumetric flow rate and compartment air velocities. Component tests with the Anemostat circular diffusers (plenum- and duct-mounted) demonstrated that the diffuser produced essentially equivalent airflow patterns and velocities in 1- and 0.34-atm environments. The tests also showed that the pressure drop across the diffuser could be scaled from 1 to 0.34 atm using the atmosphere pressure ratio. Fan tests indicated that the performance of a multiple, parallel-mounted fan cluster could be predicted by summing the single-fan flow rates at a given delta P.

  9. Development of an Outdoor Temperature-Based Control Algorithm for Residential Mechanical Ventilation Control

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Less, Brennan; Walker, Iain; Tang, Yihuan

    2014-06-01

    Smart ventilation systems use controls to ventilate more during those periods that provide either an energy or IAQ advantage (or both) and less during periods that provide a dis advantage. Using detailed building simulations, this study addresses one of the simplest and lowest cost types of smart controllers —outdoor temperature- based control. If the outdoor temperature falls below a certain cut- off, the fan is simply turned off. T he main principle of smart ventilation used in this study is to shift ventilation from time periods with large indoor -outdoor temperature differences, to periods where these differences are smaller, andmore » their energy impacts are expected to be less. Energy and IAQ performance are assessed relative to a base case of a continuously operated ventilation fan sized to comply with ASHRAE 62.2-2013 whole house ventilation requirements. In order to satisfy 62.2-2013, annual pollutant exposure must be equivalent between the temperature controlled and continuous fan cases. This requires ventilation to be greater than 62.2 requirements when the ventilation system operates. This is achieved by increasing the mechanical ventilation system air flow rates.« less

  10. Stereoscopic particle image velocimetry investigations of the mixed convection exchange flow through a horizontal vent

    NASA Astrophysics Data System (ADS)

    Varrall, Kevin; Pretrel, Hugues; Vaux, Samuel; Vauquelin, Olivier

    2017-10-01

    The exchange flow through a horizontal vent linking two compartments (one above the other) is studied experimentally. This exchange is here governed by both the buoyant natural effect due to the temperature difference of the fluids in both compartments, and the effect of a (forced) mechanical ventilation applied in the lower compartment. Such a configuration leads to uni- or bi-directional flows through the vent. In the experiments, buoyancy is induced in the lower compartment thanks to an electrical resistor. The forced ventilation is applied in exhaust or supply modes and three different values of the vent area. To estimate both velocity fields and flow rates at the vent, measurements are realized at thermal steady state, flush the vent in the upper compartment using stereoscopic particle image velocimetry (SPIV), which is original for this kind of flow. The SPIV measurements allows the area occupied by both upward and downward flows to be determined.

  11. Characteristics of coal mine ventilation air flows.

    PubMed

    Su, Shi; Chen, Hongwei; Teakle, Philip; Xue, Sheng

    2008-01-01

    Coal mine methane (CMM) is not only a greenhouse gas but also a wasted energy resource if not utilised. Underground coal mining is by far the most important source of fugitive methane emissions, and approximately 70% of all coal mining related methane is emitted to the atmosphere through mine ventilation air. Therefore, research and development on mine methane mitigation and utilisation now focuses on methane emitted from underground coal mines, in particular ventilation air methane (VAM) capture and utilisation. To date, most work has focused on the oxidation of very low concentration methane. These processes may be classified based on their combustion kinetic mechanisms into thermal oxidation and catalytic oxidation. VAM mitigation/utilisation technologies are generally divided into two basic categories: ancillary uses and principal uses. However, it is possible that the characteristics of ventilation air flows, for example the variations in methane concentration and the presence of certain compounds, which have not been reported so far, could make some potential VAM mitigation and utilisation technologies unfeasible if they cannot cope with the characteristics of mine site ventilation air flows. Therefore, it is important to understand the characteristics of mine ventilation air flows. Moreover, dust, hydrogen sulphide, sulphur dioxide, and other possible compounds emitted through mine ventilation air into the atmosphere are also pollutants. Therefore, this paper presents mine-site experimental results on the characteristics of mine ventilation air flows, including methane concentration and its variations, dust loadings, particle size, mineral matter of the dust, and other compounds in the ventilation air flows. The paper also discusses possible correlations between ventilation air characteristics and underground mining activities.

  12. Simultaneous measurement of ventilation using tracer gas techniques and VOC concentrations in homes, garages and vehicles.

    PubMed

    Batterman, Stuart; Jia, Chunrong; Hatzivasilis, Gina; Godwin, Chris

    2006-02-01

    Air exchange rates and interzonal flows are critical ventilation parameters that affect thermal comfort, air migration, and contaminant exposure in buildings and other environments. This paper presents the development of an updated approach to measure these parameters using perfluorocarbon tracer (PFT) gases, the constant injection rate method, and adsorbent-based sampling of PFT concentrations. The design of miniature PFT sources using hexafluorotoluene and octafluorobenzene tracers, and the development and validation of an analytical GC/MS method for these tracers are described. We show that simultaneous deployment of sources and passive samplers, which is logistically advantageous, will not cause significant errors over multiday measurement periods in building, or over shorter periods in rapidly ventilated spaces like vehicle cabins. Measurement of the tracers over periods of hours to a week may be accomplished using active or passive samplers, and low method detection limits (<0.025 microg m(-3)) and high precisions (<10%) are easily achieved. The method obtains the effective air exchange rate (AER), which is relevant to characterizing long-term exposures, especially when ventilation rates are time-varying. In addition to measuring the PFT tracers, concentrations of other volatile organic compounds (VOCs) are simultaneously determined. Pilot tests in three environments (residence, garage, and vehicle cabin) demonstrate the utility of the method. The 4 day effective AER in the house was 0.20 h(-1), the 4 day AER in the attached garage was 0.80 h(-1), and 16% of the ventilation in the house migrated from the garage. The 5 h AER in a vehicle traveling at 100 km h(-1) under a low-to-medium vent condition was 92 h(-1), and this represents the highest speed test found in the literature. The method is attractive in that it simultaneously determines AERs, interzonal flows, and VOC concentrations over long and representative test periods. These measurements are practical, cost-effective, and helpful in indoor air quality and other investigations.

  13. Emptying patterns of the lung studied by multiple-breath N2 washout

    NASA Technical Reports Server (NTRS)

    Lewis, S. M.

    1978-01-01

    Changes in the nitrogen concentration seen during the single-breath nitrogen washout reflect changes in relative flow (ventilation) from units with differing ventilation/volume ratios. The multiple-breath washout provides sufficient data on ventilation for units with varying ventilation/volume ratios to be plotted as a function of the volume expired. Flow from the dead space may also be determined. In young normals the emptying patterns are narrow and unimodal throughout the alveolar plateau with little or no flow from the dead space at the end of the breath. Older normals show more flow from the dead space, particularly toward the end of the breath, and some show a high ventilation/volume ratio mode early in the breath. Patients with obstructive lung disease have a high flow from the dead space which is present throughout the breath. A well ventilated mode at the end of the breath is seen in some obstructed subjects. Patients with cystic fibrosis showed a poorly ventilated mode appearing at the end of the breath as well as a very high dead space.

  14. Flow currents and ventilation in Langstroth beehives due to brood thermoregulation efforts of honeybees.

    PubMed

    Sudarsan, Rangarajan; Thompson, Cody; Kevan, Peter G; Eberl, Hermann J

    2012-02-21

    Beekeepers universally agree that ensuring sufficient ventilation is vital for sustaining a thriving, healthy honeybee colony. Despite this fact, surprisingly little is known about the ventilation and flow patterns in bee hives. We take a first step towards developing a model-based approach that uses computational fluid dynamics to simulate natural ventilation flow inside a standard Langstroth beehive. A 3-D model of a Langstroth beehive with one brood chamber and one honey super was constructed and inside it the honeybee colony was distributed among different clusters each occupying the different bee-spaces between frames in the brood chamber. For the purpose of modeling, each honeybee cluster was treated as an air-saturated porous medium with constant porosity. Heat and mass transfer interactions of the honeybees with the air, the outcome of metabolism, were captured in the porous medium model as source and sink terms appearing in the governing equations of fluid dynamics. The temperature of the brood that results from the thermoregulation efforts of the colony is applied as a boundary condition for the governing equations. The governing equations for heat, mass transport and fluid flow were solved using Fluent(©), a commercially available CFD program. The results from the simulations indicate that (a) both heat and mass transfer resulting from honeybee metabolism play a vital role in determining the structure of the flow inside the beehive and mass transfer cannot be neglected, (b) at low ambient temperatures, the nonuniform temperature profile on comb surfaces that results from brood incubation enhances flow through the honeybee cluster which removes much of the carbon-dioxide produced by the cluster resulting in lower carbon-dioxide concentration next to the brood, (c) increasing ambient (outside) air temperature causes ventilation flow rate to drop resulting in weaker flow inside the beehive. Flow visualization indicates that at low ambient air temperatures the flow inside the beehive has an interesting 3-D structure with the presence of large recirculating vortices occupying the space between honey super frames above the honeybee clusters in the brood chamber and the structure and strength of the flow inside and around the honeybee clusters changes as we increase the ambient air temperature outside the beehive. Copyright © 2011 Elsevier Ltd. All rights reserved.

  15. Cardiorespiratory interactions and blood flow generation during cardiac arrest and other states of low blood flow.

    PubMed

    Sigurdsson, Gardar; Yannopoulos, Demetris; McKnite, Scott H; Lurie, Keith G

    2003-06-01

    Recent advances in cardiopulmonary resuscitation have shed light on the importance of cardiorespiratory interactions during shock and cardiac arrest. This review focuses on recently published studies that evaluate factors that determine preload during chest compression, methods that can augment preload, and the detrimental effects of hyperventilation and interrupting chest compressions. Refilling of the ventricles, so-called ventricular preload, is diminished during cardiovascular collapse and resuscitation from cardiac arrest. In light of the potential detrimental effects and challenges of large-volume fluid resuscitations, other methods have increasing importance. During cardiac arrest, active decompression of the chest and impedance of inspiratory airflow during the recoil of the chest work by increasing negative intrathoracic pressure and, hence, increase refilling of the ventricles and increase cardiac preload, with improvement in survival. Conversely, increased frequency of ventilation has detrimental effects on coronary perfusion pressure and survival rates in cardiac arrest and severe shock. Prolonged interruption of chest compressions for delivering single-rescuer ventilation or analyzing rhythm before shock delivery is associated with decreased survival rate. Cardiorespiratory interactions are of profound importance in states of cardiovascular collapse in which increased negative intrathoracic pressure during decompression of the chest has a favorable effect and increased intrathoracic pressure with ventilation has a detrimental effect on survival rate.

  16. Influence of drive and timing mechanisms on breathing pattern and ventilation during mental task performance.

    PubMed

    Wientjes, C J; Grossman, P; Gaillard, A W

    1998-09-01

    Assessment of multiple respiratory measures may provide insight into how behavioral demands affect the breathing pattern. This is illustrated by data from a study among 44 subjects, in which tidal volume, respiration rate, minute ventilation and indices of central drive and timing mechanisms were assessed via inductive plethysmography, in addition to end-tidal PCO2. After a baseline, three conditions of a memory comparison task were presented. The first two conditions differed only with regard to the presence or absence of feedback of performance (NFB and FB). In the third 'all-or-nothing' (AON) condition, subjects only received a monetary bonus, if their performance exceeded that of the previous two conditions. Minute ventilation increased from baseline to all task conditions, and from NFB and FB to AON. Respiration rate increased in all task conditions, but there were no differences between task conditions. Tidal volume decreased during NFB, but was equal to baseline during FB and AON. Of the respiratory control indices, inspiratory flow rate covaried much more closely with minute ventilation than duty cycle. The task performance induced a minor degree of hyperventilation. The discussion focusses on how behavioral demands affect respiratory control processes to produce alterations in breathing pattern and ventilation.

  17. A Study on The Development of Local Exhaust Ventilation System (LEV’s) for Installation of Laser Cutting Machine

    NASA Astrophysics Data System (ADS)

    Harun, S. I.; Idris, S. R. A.; Tamar Jaya, N.

    2017-09-01

    Local exhaust ventilation (LEV) is an engineering system frequently used in the workplace to protect operators from hazardous substances. The objective of this project is design and fabricate the ventilation system as installation for chamber room of laser cutting machine and to stimulate the air flow inside chamber room of laser cutting machine with the ventilation system that designed. LEV’s fabricated with rated voltage D.C 10.8V and 1.5 ampere. Its capacity 600 ml, continuously use limit approximately 12-15 minute, overall length LEV’s fabricated is 966 mm with net weight 0.88 kg and maximum airflow is 1.3 meter cubic per minute. Stimulate the air flow inside chamber room of laser cutting machine with the ventilation system that designed and fabricated overall result get 2 main gas vapor which air and carbon dioxide. For air gas which experimented by using anemometer, general duct velocity that produce is same with other gas produce, carbon dioxide which 5 m/s until 10 m/s. Overall result for 5 m/s and 10 m/s as minimum and maximum duct velocity produce for both air and carbon dioxide. The air gas flow velocity that captured by LEV’s fabricated, 3.998 m/s average velocity captured from 5 m/s duct velocity which it efficiency of 79.960% and 7.667 m/s average velocity captured from 10 m/s duct velocity with efficiency of 76.665%. For carbon dioxide gas flow velocity that captured by LEV’s fabricated, 3.674 m/s average velocity captured from 5 m/s duct velocity which it efficiency of 73.480% and 8.255 m/s average velocity captured from 10 m/s duct velocity with efficiency of 82.545%.

  18. VOC Emission Reduction Study at the Hill Air Force Base Building 515 Painting Facility

    DTIC Science & Technology

    1990-09-01

    occurs during painting. A system for decreasing the flow to a downstream VOC emission control device can be designed that takes advantage of this...paint application process. A flow-reducing ventilation system that takes advantage of this operating characteristic can be designed in which the...flow from the second duct is vented to a VOC emission control device. The advantage of this system is that the flow rate to a VOC emission contro

  19. Neopuff T-piece resuscitator mask ventilation: Does mask leak vary with different peak inspiratory pressures in a manikin model?

    PubMed

    Maheshwari, Rajesh; Tracy, Mark; Hinder, Murray; Wright, Audrey

    2017-08-01

    The aim of this study was to compare mask leak with three different peak inspiratory pressure (PIP) settings during T-piece resuscitator (TPR; Neopuff) mask ventilation on a neonatal manikin model. Participants were neonatal unit staff members. They were instructed to provide mask ventilation with a TPR with three PIP settings (20, 30, 40 cm H 2 O) chosen in a random order. Each episode was for 2 min with 2-min rest period. Flow rate and positive end-expiratory pressure (PEEP) were kept constant. Airway pressure, inspiratory and expiratory tidal volumes, mask leak, respiratory rate and inspiratory time were recorded. Repeated measures analysis of variance was used for statistical analysis. A total of 12 749 inflations delivered by 40 participants were analysed. There were no statistically significant differences (P > 0.05) in the mask leak with the three PIP settings. No statistically significant differences were seen in respiratory rate and inspiratory time with the three PIP settings. There was a significant rise in PEEP as the PIP increased. Failure to achieve the desired PIP was observed especially at the higher settings. In a neonatal manikin model, the mask leak does not vary as a function of the PIP when the flow rate is constant. With a fixed rate and inspiratory time, there seems to be a rise in PEEP with increasing PIP. © 2017 Paediatrics and Child Health Division (The Royal Australasian College of Physicians).

  20. Data on the natural ventilation performance of windcatcher with anti-short-circuit device (ASCD).

    PubMed

    Nejat, Payam; Calautit, John Kaiser; Majid, Muhd Zaimi Abd; Hughes, Ben Richard; Jomehzadeh, Fatemeh

    2016-12-01

    This article presents the datasets which were the results of the study explained in the research paper 'Anti-short-circuit device: a new solution for short-circuiting in windcatcher and improvement of natural ventilation performance' (P. Nejat, J.K. Calautit, M.Z. Abd. Majid, B.R. Hughes, F. Jomehzadeh, 2016) [1] which introduces a new technique to reduce or prevent short-circuiting in a two-sided windcatcher and also lowers the indoor CO2 concentration and improve the ventilation distribution. Here, we provide details of the numerical modeling set-up and data collection method to facilitate reproducibility. The datasets includes indoor airflow, ventilation rates and CO2 concentration data at several points in the flow field. The CAD geometry of the windcatcher models are also included.

  1. [High flow nasal cannula oxygen therapy in the treatment of acute bronchiolitis in neonates].

    PubMed

    Bermúdez Barrezueta, Lorena; García Carbonell, Nuria; López Montes, Jorge; Gómez Zafra, Rafael; Marín Reina, Purificación; Herrmannova, Jana; Casero Soriano, Javier

    2017-01-01

    To determine whether the availability of heated humidified high-flow nasal cannula (HFNC) therapy was associated with a decrease in need for mechanical ventilation in neonates hospitalised with acute bronchiolitis. A combined retrospective and prospective (ambispective) cohort study was performed in a type II-B Neonatal Unit, including hospitalised neonates with acute bronchiolitis after the introduction of HFNC (HFNC-period; October 2011-April 2015). They were compared with a historical cohort prior to the availability of this technique (pre-HFNC; January 2008-May 2011). The need for mechanical ventilation between the two study groups was analysed. Clinical parameters and technique-related complications were evaluated in neonates treated with HFNC. A total of 112 neonates were included, 56 after the introduction of HFNC and 56 from the period before the introduction of HFNC. None of patients in the HFNC-period required intubation, compared with 3.6% of the patients in the pre-HFNC group. The availability of HFNC resulted in a significant decrease in the need for non-invasive mechanical ventilation (30.4% vs 10.7%; P=.01), with a relative risk (RR) of .353 (95% CI; .150-.829), an absolute risk reduction (ARR) of 19.6% (95% CI; 5.13 - 34.2), yielding a NNT of 5. In the HFNC-period, 22 patients received high flow therapy, and 22.7% (95% CI; 7.8 to 45.4) required non-invasive ventilation. Treatment with HFNC was associated with a significant decrease in heart rate (P=.03), respiratory rate (P=.01), and an improvement in the Wood-Downes Férres score (P=.00). No adverse effects were observed. The availability of HFNC reduces the need for non-invasive mechanical ventilation, allowing a safe and effective medical management of neonates with acute bronchiolitis. Copyright © 2015 Asociación Española de Pediatría. Publicado por Elsevier España, S.L.U. All rights reserved.

  2. A regulator for pressure-controlled total-liquid ventilation.

    PubMed

    Robert, Raymond; Micheau, Philippe; Avoine, Olivier; Beaudry, Benoit; Beaulieu, Alexandre; Walti, Hervé

    2010-09-01

    Total-liquid ventilation (TLV) is an innovative experimental method of mechanical-assisted ventilation in which lungs are totally filled and then ventilated with a tidal volume of perfluorochemical liquid by using a dedicated liquid ventilator. Such a novel medical device must resemble other conventional ventilators: it must be able to conduct controlled-pressure ventilation. The objective was to design a robust controller to perform pressure-regulated expiratory flow and to implement it on our latest liquid-ventilator prototype (Inolivent-4). Numerical simulations, in vitro experiments, and in vivo experiments in five healthy term newborn lambs have demonstrated that it was efficient to generate expiratory flows while avoiding collapses. Moreover, the in vivo results have demonstrated that our liquid ventilator can maintain adequate gas exchange, normal acid-base equilibrium, and achieve greater minute ventilation, better oxygenation and CO2 extraction, while nearing flow limits. Hence, it is our suggestion to perform pressure-controlled ventilation during expiration with minute ventilation equal or superior to 140 mL x min(-1) x kg(-1) in order to ensure PaCO2 below 55 mmHg. From a clinician's point of view, pressure-controlled ventilation greatly simplifies the use of the liquid ventilator, which will certainly facilitate its introduction in intensive care units for clinical applications.

  3. Factors contributing to the failure of Humidified High-Flow Nasal Cannulae.

    PubMed

    Teoh, Sophia; Clyde, Elizabeth; Dassios, Theodore; Greenough, Anne

    2018-05-24

    The use of humidified high-flow nasal cannulae (HHFNC) as an alternative mode of non-invasive ventilation (NIV) in neonates has become widespread. A survey of UK neonatal units showed the proportion using HHFNC had increased from 56% in 2012 to 87% in 2015 (1). A recently reported Cochrane Review (2) comparing the use of HHFNC against other NIV modes of ventilation immediately after birth or following extubation showed no significant difference in the rates of bronchopulmonary dysplasia (BPD) or death and no significant difference in the rates of treatment failure/reintubation. Benefits cited include a significantly reduced risk of nasal trauma as compared to continuous positive airway pressure (CPAP). Furthermore, both medical staff and parents (3) were found to prefer HHFNC to CPAP. Identification of infants in whom use of HHFNC as either a primary or step-down mode of respiratory support may be inappropriate might further reduce the failure rate of HHFNC. This article is protected by copyright. All rights reserved. This article is protected by copyright. All rights reserved.

  4. The Fluid Mechanics of Natural Ventilation

    NASA Astrophysics Data System (ADS)

    Linden, P. F.

    1999-01-01

    Natural ventilation of buildings is the flow generated by temperature differences and by the wind. The governing feature of this flow is the exchange between an interior space and the external ambient. Although the wind may often appear to be the dominant driving mechanism, in many circumstances temperature variations play a controlling feature on the ventilation since the directional buoyancy force has a large influence on the flow patterns within the space and on the nature of the exchange with the outside. Two forms of ventilation are discussed: mixing ventilation, in which the interior is at an approximately uniform temperature, and displacement ventilation, where there is strong internal stratification. The dynamics of these buoyancy-driven flows are considered, and the effects of wind on them are examined. The aim behind this work is to give designers rules and intuition on how air moves within a building; the research reveals a fascinating branch of fluid mechanics.

  5. The effect of flow limitation on the cardiorespiratory response to arousal from sleep under controlled conditions of chemostimulation in healthy older adults.

    PubMed

    Goff, Elizabeth A; Nicholas, Christian L; Kleiman, Jan; Spear, Owen; Morrell, Mary J; Trinder, John

    2012-12-01

    The influence of flow limitation on the magnitude of the cardiorespiratory response to arousal from sleep is of interest in older people, because they experience considerable flow limitation and frequent arousals from sleep. We studied older flow-limiting subjects, testing the hypothesis that the cardiorespiratory activation response would be larger when arousal occurred during flow limitation, compared to no flow limitation, and chemical stimuli were controlled. In 11 older adults [mean ± standard deviation (SD) age: 68 ± 5 years] ventilation was stabilized using continuous positive airway pressure, and flow limitation was induced by dialling down the pressure. Partial pressure of end-tidal carbon dioxide (PetCO(2)) was maintained by titration of the inspired CO(2) and hyperoxia was maintained using 40% O(2) balanced with nitrogen. Flow limitation at the time of arousal did not augment cardiovascular activation response (heart rate P = 0.7; systolic blood pressure P = 0.6; diastolic blood pressure P = 0.3), whereas ventilation was greater following arousals during flow limitation compared to no flow limitation (P < 0.001). The pre-post-arousal differences in ventilation reflected significant pre-arousal suppression (due to flow limitation) plus post-arousal activation. In summary, the cardiovascular response to arousal from sleep is not influenced by flow limitation at the time of arousal, when chemical stimuli are controlled in older adults. This finding may contribute to the decreased cardiovascular burden associated with sleep-disordered breathing reported in older adults, although our data do not exclude the possibility that flow limitation in the presence of mild hypoxic hypercapnia could increase the cardiovascular response to arousal. © 2012 European Sleep Research Society.

  6. Evaluation of the effects of dorsal versus lateral recumbency on the cardiopulmonary system during anesthesia with isoflurane in red-tailed hawks (Buteo jamaicensis).

    PubMed

    Hawkins, Michelle G; Malka, Shachar; Pascoe, Peter J; Solano, Adrian M; Kass, Philip H; Ohmura, Hajime; Jones, James H

    2013-01-01

    To evaluate the effects of dorsal versus lateral recumbency on the cardiopulmonary system during isoflurane anesthesia in red-tailed hawks (Buteo jamaicensis). 6 adult 1.1- to 1.6-kg red-tailed hawks. A randomized, crossover study was used to evaluate changes in respiratory rate, tidal volume, minute ventilation, heart rate, mean arterial and indirect blood pressures, and end-tidal Pco(2) measured every 5 minutes plus Paco(2) and Pao(2) and arterial pH measured every 15 minutes throughout a 75-minute study period. Respiratory rate was higher, tidal volume lower, and minute ventilation not different in lateral versus dorsal recumbency. Position did not affect heart rate, mean arterial blood pressure, or indirect blood pressure, although heart rate decreased during the anesthetic period. Birds hypoventilated in both positions and Paco(2) differed with time and position × time interaction. The Petco(2) position × time interaction was significant and Petco(2) was a mean of 7 Torr higher than Paco(2). The Paco(2) in dorsal recumbency was a mean of 32 Torr higher than in lateral recumbency. Birds in both positions developed respiratory acidosis. Differences in tidal volume with similar minute ventilation suggested red-tailed hawks in dorsal recumbency might have lower dead space ventilation. Despite similar minute ventilation in both positions, birds in dorsal recumbency hypoventilated more yet maintained higher Pao(2), suggesting parabronchial ventilatory or pulmonary blood flow distribution changes with position. The results refute the hypothesis that dorsal recumbency compromises ventilation and O(2) transport more than lateral recumbency in red-tailed hawks.

  7. Extubation success in premature infants with respiratory distress syndrome treated with bi-level nasal continuous positive airway pressure versus nasal intermittent positive pressure ventilation.

    PubMed

    Thomas, Patricia E; LeFlore, Judy

    2013-01-01

    Infants born prematurely with respiratory distress syndrome are at high risk for complications from mechanical ventilation. Strategies are needed to minimize their days on the ventilator. The purpose of this study was to compare extubation success rates in infants treated with 2 different types of continuous positive airway pressure devices. A retrospective cohort study design was used. Data were retrieved from electronic medical records for patients in a large, metropolitan, level III neonatal intensive care unit. A sample of 194 premature infants with respiratory distress syndrome was selected, 124 of whom were treated with nasal intermittent positive pressure ventilation and 70 with bi-level variable flow nasal continuous positive airway pressure (bi-level nasal continuous positive airway pressure). Infants in both groups had high extubation success rates (79% of nasal intermittent positive pressure ventilation group and 77% of bi-level nasal continuous positive airway pressure group). Although infants in the bi-level nasal continuous positive airway pressure group were extubated sooner, there was no difference in duration of oxygen therapy between the 2 groups. Promoting early extubation and extubation success is a vital strategy to reduce complications of mechanical ventilation that adversely affect premature infants with respiratory distress syndrome.

  8. In China, Students in Crowded Dormitories with a Low Ventilation Rate Have More Common Colds: Evidence for Airborne Transmission

    PubMed Central

    Sun, Yuexia; Wang, Zhigang; Zhang, Yufeng; Sundell, Jan

    2011-01-01

    Objective To test whether the incidence of common colds among college students in China is associated with ventilation rates and crowdedness in dormitories. Methods In Phase I of the study, a cross-sectional study, 3712 students living in 1569 dorm rooms in 13 buildings responded to a questionnaire about incidence and duration of common colds in the previous 12 months. In Phase II, air temperature, relative humidity and CO2 concentration were measured for 24 hours in 238 dorm rooms in 13 buildings, during both summer and winter. Out-to indoor air flow rates at night were calculated based on measured CO2 concentrations. Results In Phase I, 10% of college students reported an incidence of more than 6 common colds in the previous 12 months, and 15% reported that each infection usually lasted for more than 2 weeks. Students in 6-person dorm rooms were about 2 times as likely to have an incidence of common colds ≥6 times per year and a duration ≥2 weeks, compared to students in 3-person rooms. In Phase II, 90% of the measured dorm rooms had an out-to indoor air flow rate less than the Chinese standard of 8.3 L/s per person during the heating season. There was a dose-response relationship between out-to indoor air flow rate per person in dorm rooms and the proportion of occupants with annual common cold infections ≥6 times. A mean ventilation rate of 5 L/(s•person) in dorm buildings was associated with 5% of self reported common cold ≥6 times, compared to 35% at 1 L/(s•person). Conclusion Crowded dormitories with low out-to indoor airflow rates are associated with more respiratory infections among college students. PMID:22110607

  9. Generation rate of carbon monoxide from CO2 arc welding.

    PubMed

    Ojima, Jun

    2013-01-01

    CO poisoning has been a serious industrial hazard in Japanese workplaces. Although incomplete combustion is the major cause of CO generation, there is a risk of CO poisoning during some welding operations. The aim of the present study was to evaluate the generation rate of CO from CO2 arc welding under controlled laboratory conditions and estimate the ventilation requirements for the prevention of CO poisoning. Bead on plate welding was carried out with an automatic welding robot on a rolled steel base metal under several conditions. The concentration of emitted CO from the welding was measured by a real-time CO monitor in a well-ventilated laboratory that was free from ambient CO contamination. The generation rate of CO was obtained from the three measurements-the flow rate of the welding exhaust gas, CO concentration in the exhaust gas and the arcing time. Then the ventilation requirement to prevent CO poisoning was calculated. The generation rate of CO was found to be 386-883 ml/min with a solid wire and 331-1,293 ml/min with a flux cored wire respectively. It was found that the CO concentration in a room would be maintained theoretically below the OSHA PEL (50 ppm) providing the ventilation rate in the room was 6.6-25.9 m3/min. The actual ventilation requirement was then estimated to be 6.6-259 m3/min considering incomplete mixing. In order to prevent CO poisoning, some countermeasures against gaseous emission as well as welding fumes should be taken eagerly.

  10. An air flow sensor for neonatal mechanical ventilation applications based on a novel fiber-optic sensing technique.

    PubMed

    Battista, L; Sciuto, S A; Scorza, A

    2013-03-01

    In this work, a simple and low-cost air flow sensor, based on a novel fiber-optic sensing technique has been developed for monitoring air flows rates supplied by a neonatal ventilator to support infants in intensive care units. The device is based on a fiber optic sensing technique allowing (a) the immunity to light intensity variations independent by measurand and (b) the reduction of typical shortcomings affecting all biomedical fields (electromagnetic interference and patient electrical safety). The sensing principle is based on the measurement of transversal displacement of an emitting fiber-optic cantilever due to action of air flow acting on it; the fiber tip displacement is measured by means of a photodiode linear array, placed in front of the entrance face of the emitting optical fiber in order to detect its light intensity profile. As the measurement system is based on a detection of the illumination pattern, and not on an intensity modulation technique, it results less sensitive to light intensity fluctuation independent by measurand than intensity-based sensors. The considered technique is here adopted in order to develop two different configurations for an air flow sensor suitable for the measurement of air flow rates typically occurring during mechanical ventilation of newborns: a mono-directional and a bi-directional transducer have been proposed. A mathematical model for the air flow sensor is here proposed and a static calibration of two different arrangements has been performed: a measurement range up to 3.00 × 10(-4) m(3)∕s (18.0 l∕min) for the mono-directional sensor and a measurement range of ±3.00 × 10(-4) m(3)∕s (±18.0 l∕min) for the bi-directional sensor are experimentally evaluated, according to the air flow rates normally encountered during tidal breathing of infants with a mass lower than 10 kg. Experimental data of static calibration result in accordance with the proposed theoretical model: for the mono-directional configuration, the coefficient of determination r(2) is equal to 0.997; for the bi-directional configuration, the coefficient of determination r(2) is equal to 0.990 for positive flows (inspiration) and 0.988 for negative flows (expiration). Measurement uncertainty δQ of air flow rate has been evaluated by means of the propagation of distributions and the percentage error in the arrangement of bi-directional sensor ranges from a minimum of about 0.5% at -18.0 l∕min to a maximum of about 9% at -12.0 l∕min.

  11. An air flow sensor for neonatal mechanical ventilation applications based on a novel fiber-optic sensing technique

    NASA Astrophysics Data System (ADS)

    Battista, L.; Sciuto, S. A.; Scorza, A.

    2013-03-01

    In this work, a simple and low-cost air flow sensor, based on a novel fiber-optic sensing technique has been developed for monitoring air flows rates supplied by a neonatal ventilator to support infants in intensive care units. The device is based on a fiber optic sensing technique allowing (a) the immunity to light intensity variations independent by measurand and (b) the reduction of typical shortcomings affecting all biomedical fields (electromagnetic interference and patient electrical safety). The sensing principle is based on the measurement of transversal displacement of an emitting fiber-optic cantilever due to action of air flow acting on it; the fiber tip displacement is measured by means of a photodiode linear array, placed in front of the entrance face of the emitting optical fiber in order to detect its light intensity profile. As the measurement system is based on a detection of the illumination pattern, and not on an intensity modulation technique, it results less sensitive to light intensity fluctuation independent by measurand than intensity-based sensors. The considered technique is here adopted in order to develop two different configurations for an air flow sensor suitable for the measurement of air flow rates typically occurring during mechanical ventilation of newborns: a mono-directional and a bi-directional transducer have been proposed. A mathematical model for the air flow sensor is here proposed and a static calibration of two different arrangements has been performed: a measurement range up to 3.00 × 10-4 m3/s (18.0 l/min) for the mono-directional sensor and a measurement range of ±3.00 × 10-4 m3/s (±18.0 l/min) for the bi-directional sensor are experimentally evaluated, according to the air flow rates normally encountered during tidal breathing of infants with a mass lower than 10 kg. Experimental data of static calibration result in accordance with the proposed theoretical model: for the mono-directional configuration, the coefficient of determination r2 is equal to 0.997; for the bi-directional configuration, the coefficient of determination r2 is equal to 0.990 for positive flows (inspiration) and 0.988 for negative flows (expiration). Measurement uncertainty δQ of air flow rate has been evaluated by means of the propagation of distributions and the percentage error in the arrangement of bi-directional sensor ranges from a minimum of about 0.5% at -18.0 l/min to a maximum of about 9% at -12.0 l/min.

  12. Feasibility and safety of low-flow extracorporeal carbon dioxide removal to facilitate ultra-protective ventilation in patients with moderate acute respiratory distress sindrome.

    PubMed

    Fanelli, Vito; Ranieri, Marco V; Mancebo, Jordi; Moerer, Onnen; Quintel, Michael; Morley, Scott; Moran, Indalecio; Parrilla, Francisco; Costamagna, Andrea; Gaudiosi, Marco; Combes, Alain

    2016-02-10

    Mechanical ventilation with a tidal volume (VT) of 6 mL/kg/predicted body weight (PBW), to maintain plateau pressure (Pplat) lower than 30 cmH2O, does not completely avoid the risk of ventilator induced lung injury (VILI). The aim of this study was to evaluate safety and feasibility of a ventilation strategy consisting of very low VT combined with extracorporeal carbon dioxide removal (ECCO2R). In fifteen patients with moderate ARDS, VT was reduced from baseline to 4 mL/kg PBW while PEEP was increased to target a plateau pressure--(Pplat) between 23 and 25 cmH2O. Low-flow ECCO2R was initiated when respiratory acidosis developed (pH < 7.25, PaCO2 > 60 mmHg). Ventilation parameters (VT, respiratory rate, PEEP), respiratory compliance (CRS), driving pressure (DeltaP = VT/CRS), arterial blood gases, and ECCO2R system operational characteristics were collected during the period of ultra-protective ventilation. Patients were weaned from ECCO2R when PaO2/FiO2 was higher than 200 and could tolerate conventional ventilation settings. Complications, mortality at day 28, need for prone positioning and extracorporeal membrane oxygenation, and data on weaning from both MV and ECCO2R were also collected. During the 2 h run in phase, VT reduction from baseline (6.2 mL/kg PBW) to approximately 4 mL/kg PBW caused respiratory acidosis (pH < 7.25) in all fifteen patients. At steady state, ECCO2R with an average blood flow of 435 mL/min and sweep gas flow of 10 L/min was effective at correcting pH and PaCO2 to within 10 % of baseline values. PEEP values tended to increase at VT of 4 mL/kg from 12.2 to 14.5 cmH2O, but this change was not statistically significant. Driving pressure was significantly reduced during the first two days compared to baseline (from 13.9 to 11.6 cmH2O; p < 0.05) and there were no significant differences in the values of respiratory system compliance. Rescue therapies for life threatening hypoxemia such as prone position and ECMO were necessary in four and two patients, respectively. Only two study-related adverse events were observed (intravascular hemolysis and femoral catheter kinking). The low-flow ECCO2R system safely facilitates a low volume, low pressure ultra-protective mechanical ventilation strategy in patients with moderate ARDS.

  13. Cardiovascular, cerebrovascular, and respiratory changes induced by different types of music in musicians and non-musicians: the importance of silence.

    PubMed

    Bernardi, L; Porta, C; Sleight, P

    2006-04-01

    To assess the potential clinical use, particularly in modulating stress, of changes in the cardiovascular and respiratory systems induced by music, specifically tempo, rhythm, melodic structure, pause, individual preference, habituation, order effect of presentation, and previous musical training. Measurement of cardiovascular and respiratory variables while patients listened to music. University research laboratory for the study of cardiorespiratory autonomic function. 12 practising musicians and 12 age matched controls. After a five minute baseline, presentation in random order of six different music styles (first for a two minute, then for a four minute track), with a randomly inserted two minute pause, in either sequence. Breathing rate, ventilation, carbon dioxide, RR interval, blood pressure, mid-cerebral artery flow velocity, and baroreflex. Ventilation, blood pressure, and heart rate increased and mid-cerebral artery flow velocity and baroreflex decreased with faster tempi and simpler rhythmic structures compared with baseline. No habituation effect was seen. The pause reduced heart rate, blood pressure, and minute ventilation, even below baseline. An order effect independent of style was evident for mid-cerebral artery flow velocity, indicating a progressive reduction with exposure to music, independent of style. Musicians had greater respiratory sensitivity to the music tempo than did non-musicians. Music induces an arousal effect, predominantly related to the tempo. Slow or meditative music can induce a relaxing effect; relaxation is particularly evident during a pause. Music, especially in trained subjects, may first concentrate attention during faster rhythms, then induce relaxation during pauses or slower rhythms.

  14. Experimental system for the control of surgically induced infections

    NASA Technical Reports Server (NTRS)

    Tevebaugh, M. D.

    1971-01-01

    The development tests to be performed on the experimental system are described in detail. The test equipment, conditions, and procedures are given. The portable clean room tests include assembly, collapsability, portability, and storage; laminar flow rate; static pressure; air flow pattern; and electrostatic buildup. The other tests are on the ventilation system, human factors evaluation, electrical subsystem, and material compatibility.

  15. Quantifying air distribution, ventilation effectiveness and airborne pollutant transport in an aircraft cabin mockup

    NASA Astrophysics Data System (ADS)

    Wang, Aijun

    The health, safety and comfort of passengers during flight inspired this research into cabin air quality, which is closely related to its airflow distribution, ventilation effectiveness and airborne pollutant transport. The experimental facility is a full-scale aircraft cabin mockup. A volumetric particle tracking velocimetry (VPTV) technique was enhanced by incorporating a self-developed streak recognition algorithm. Two stable recirculation regions, the reverse flows above the seats and the main air jets from the air supply inlets formed the complicated airflow patterns inside the cabin mockup. The primary air flow was parallel to the passenger rows. The small velocity component in the direction of the cabin depth caused less net air exchange between the passenger rows than that parallel to the passenger rows. Different total air supply rate changed the developing behaviors of the main air jets, leading to different local air distribution patterns. Two indices, Local mean age of air and ventilation effectiveness factor (VEF), were measured at five levels of air supply rate and two levels of heating load. Local mean age of air decreased linearly with an increase in the air supply rate, while the VEF remained consistent when the air supply rate varied. The thermal buoyancy force from the thermal plume generated the upside plume flow, opposite to the main jet flow above the boundary seats and thus lowered the local net air exchange. The airborne transport dynamics depends on the distance between the source and the receptors, the relative location of pollutant source, and air supply rate. Exposure risk was significantly reduced with increased distance between source and receptors. Another possible way to decrease the exposure risk was to position the release source close to the exhaust outlets. Increasing the air supply rate could be an effective solution under some emergency situations. The large volume of data regarding the three-dimensional air velocities was visualized in the CAVE virtual environment. ShadowLight, a virtual reality application was used to import and navigate the velocity vectors through the virtual airspace. A real world demonstration and an active interaction with the three-dimensional air velocity data have been established.

  16. The Evolution of Unidirectional Pulmonary Airflow.

    PubMed

    Farmer, C G

    2015-07-01

    Conventional wisdom holds that the avian respiratory system is unique because air flows in the same direction through most of the gas-exchange tubules during both phases of ventilation. However, recent studies showing that unidirectional airflow also exists in crocodilians and lizards raise questions about the true phylogenetic distribution of unidirectional airflow, the selective drivers of the trait, the date of origin, and the functional consequences of this phenomenon. These discoveries suggest unidirectional flow was present in the common diapsid ancestor and are inconsistent with the traditional paradigm that unidirectional flow is an adaptation for supporting high rates of gas exchange. Instead, these discoveries suggest it may serve functions such as decreasing the work of breathing, decreasing evaporative respiratory water loss, reducing rates of heat loss, and facilitating crypsis. The divergence in the design of the respiratory system between unidirectionally ventilated lungs and tidally ventilated lungs, such as those found in mammals, is very old, with a minimum date for the divergence in the Permian Period. From this foundation, the avian and mammalian lineages evolved very different respiratory systems. I suggest the difference in design is due to the same selective pressure, expanded aerobic capacity, acting under different environmental conditions. High levels of atmospheric oxygen of the Permian Period relaxed selection for a thin blood-gas barrier and may have resulted in the homogeneous, broncho-alveolar design, whereas the reduced oxygen of the Mesozoic selected for a heterogeneous lung with an extremely thin blood-gas barrier. These differences in lung design may explain the puzzling pattern of ecomorphological diversification of Mesozoic mammals: all were small animals that did not occupy niches requiring a great aerobic capacity. The broncho-alveolar lung and the hypoxia of the Mesozoic may have restricted these mammals from exploiting niches of large body size, where cursorial locomotion can be advantageous, as well as other niches requiring great aerobic capacities, such as those using flapping flight. Furthermore, hypoxia may have exerted positive selection for a parasagittal posture, the diaphragm, and reduced erythrocyte size, innovations that enabled increased rates of ventilation and more rapid rates of diffusion in the lung. ©2015 Int. Union Physiol. Sci./Am. Physiol. Soc.

  17. Demonstration of Split-Flow Ventilation and Recirculation as Flow- Reduction Methods in an Air Force Paint Spray Booth. Volume 1

    DTIC Science & Technology

    1994-07-27

    of the split-flow and recirculation modifications in typical Air Force painting operations; itwas a proof-of- concept study only. It is recognized...recirculating ventilation. 4 To Implement this flow-reduction concept , it must first be established that recirculation does not cause an accumulation of toxic...ventilation concept . The concentration gradient is determined by height and direction of paint application. If the concentration in the top portion is

  18. Temperature effects on metabolic rate and cardiorespiratory physiology of the spiny rock lobster (Jasus edwardsii) during rest, emersion and recovery.

    PubMed

    Forgan, Leonard G; Tuckey, Nicholas P L; Cook, Denham G; Jerrett, Alistair R

    2014-05-01

    Although spiny rock lobster (Jasus edwardsii) is a wholly sub-littoral species, they show a considerable ability to survive prolonged emersion, a fact exploited during the commercial export of this species. Yet, despite this remarkable hardiness, basic information on how this species responds physiologically to emersion is somewhat lacking. Using flow-through respirometry and electrophysiological techniques, we identified that J. edwardsii undergoes marked physiological changes during rest, emersion and recovery over a broad range of temperatures (3.7-17.8 °C). Under resting conditions, routine metabolic rates (RMR) were 22.57 ± 2.39, 9.69 ± 0.55 and 8.09 ± 0.27 mL O2 h(-1), average heart rates (Hr) were 54.72 ± 4.46, 37.68 ± 2.86 and 29.67 ± 0.59 BPM, and ventilation frequencies were 83.71 ± 5.86, 45.34 ± 2.91 and 41.62 ± 0.65 BPM at 15.0, 7.5 and 3.7 °C, respectively. Notably, the surgical implantation of electrodes elevated RMR compared with non-surgical treatments. In surgery and non-surgery groups, Q 10 was calculated to be ca. 3.0. Upon emersion, rate of oxygen consumption and Hr decreased below resting rates in a temperature-dependent manner, but, along with rate of CO2 production, increased steadily during 24-h emersion. Ventilation frequencies upon emersion showed a contrasting response and increased significantly above resting rates. When returned to flow-through sea water for recovery, elevated respiration rates provided clear evidence of an O2 debt, and near-complete recovery was observed after 17 h at both 15.0 and 7.5 °C, but close to no debt was recovered at 3.7 °C. In addition, J. edwardsii was observed to undergo marked diurnal and periodic ventilation cycles, characterised by synchronous changes in RMR, Hr and ventilation frequency.

  19. Interactive simulation system for artificial ventilation on the internet: virtual ventilator.

    PubMed

    Takeuchi, Akihiro; Abe, Tadashi; Hirose, Minoru; Kamioka, Koichi; Hamada, Atsushi; Ikeda, Noriaki

    2004-12-01

    To develop an interactive simulation system "virtual ventilator" that demonstrates the dynamics of pressure and flow in the respiratory system under the combination of spontaneous breathing, ventilation modes, and ventilator options. The simulation system was designed to be used by unexperienced health care professionals as a self-training tool. The system consists of a simulation controller and three modules: respiratory, spontaneous breath, and ventilator. The respiratory module models the respiratory system by three resistances representing the main airway, the right and left lungs, and two compliances also representing the right and left lungs. The spontaneous breath module generates inspiratory negative pressure produced by a patient. The ventilator module generates driving force of pressure or flow according to the combination of the ventilation mode and options. These forces are given to the respiratory module through the simulation controller. The simulation system was developed using HTML, VBScript (3000 lines, 100 kB) and ActiveX control (120 kB), and runs on Internet Explorer (5.5 or higher). The spontaneous breath is defined by a frequency, amplitude and inspiratory patterns in the spontaneous breath module. The user can construct a ventilation mode by setting a control variable, phase variables (trigger, limit, and cycle), and options. Available ventilation modes are: controlled mechanical ventilation (CMV), continuous positive airway pressure, synchronized intermittent mandatory ventilation (SIMV), pressure support ventilation (PSV), SIMV + PSV, pressure-controlled ventilation (PCV), pressure-regulated volume control (PRVC), proportional assisted ventilation, mandatory minute ventilation (MMV), bilevel positive airway pressure (BiPAP). The simulation system demonstrates in a graph and animation the airway pressure, flow, and volume of the respiratory system during mechanical ventilation both with and without spontaneous breathing. We developed a web application that demonstrated the respiratory mechanics and the basic theory of ventilation mode.

  20. Ventilation Inception and Washout, Scaling, and Effects on Hydrodynamic Performance of a Surface Piercing Strut

    NASA Astrophysics Data System (ADS)

    Harwood, Casey; Young, Yin Lu; Ceccio, Steven

    2014-11-01

    High-lift devices that operate at or near a fluid free surface (such as surface-piercing or shallowly-submerged propellers and hydrofoils) are prone to a multiphase flow phenomenon called ventilation, wherein non-condensable gas is entrained in the low-pressure flow, forming a cavity around the body and dramatically altering the global hydrodynamic forces. Experiments are being conducted at the University of Michigan's towing tank using a canonical surface-piercing strut to investigate atmospheric ventilation. The goals of the work are (i) to gain an understanding of the dominant physics in fully wetted, partially ventilated, and fully ventilated flow regimes, (ii) to quantify the effects of governing dimensionless parameters on the transition between flow regimes, and (iii) to develop scaling relations for the transition between flow regimes. Using theoretical arguments and flow visualization techniques, new criteria are developed for classifying flow regimes and transition mechanisms. Unsteady transition mechanisms are described and mapped as functions of the governing non-dimensional parameters. A theoretical scaling relationship is developed for ventilation washout, which is shown to adequately capture the experimentally-observed washout boundary. This material is based upon work supported by the National Science Foundation Graduate Student Research Fellowship under Grant No. DGE 1256260. Support also comes from the Naval Engineering Education Center (Award No. N65540-10-C-003).

  1. Daytime Mouthpiece for Continuous Noninvasive Ventilation in Individuals With Amyotrophic Lateral Sclerosis.

    PubMed

    Bédard, Marie-Eve; McKim, Douglas A

    2016-10-01

    Noninvasive ventilation (NIV) is commonly used to provide ventilatory support for individuals with amyotrophic lateral sclerosis (ALS). Once 24-h ventilation is required, the decision between invasive tracheostomy ventilation and palliation is often faced. This study describes the use and outcomes of daytime mouthpiece ventilation added to nighttime mask ventilation for continuous NIV in subjects with ALS as an effective alternative. This was a retrospective study of 39 subjects with ALS using daytime mouthpiece ventilation over a 17-y period. Thirty-one subjects were successful with mouthpiece ventilation, 2 were excluded, 2 stopped because of lack of motivation, and 4 with bulbar subscores of the Revised Amyotrophic Lateral Sclerosis Functional Rating Scale (b-ALSFRS-R) between 0 and 3 physically failed to use it consistently. No subject in the successful group had a b-ALSFRS-R score of <6. Thirty of the successful subjects were able to generate a maximum insufflation capacity - vital capacity difference with lung volume recruitment. The median (range) survival to tracheostomy or death from initiation of nocturnal NIV and mouthpiece ventilation were 648 (176-2,188) and 286 (41-1,769) d, respectively. Peak cough flow with lung-volume recruitment >180 L/min at initiation of mouthpiece ventilation was associated with a longer survival (637 ± 468 vs 240 ± 158 d (P = .01). Mouthpiece ventilation provides effective ventilation and prolonged survival for individuals with ALS requiring full-time ventilatory support and maintaining adequate bulbar function. Copyright © 2016 by Daedalus Enterprises.

  2. PAINT SPRAY BOOTH MODIFICATIONS FOR RECIRCULATION VENTILATION

    EPA Science Inventory

    The control of emissions from spray painting operations has historically been cost prohibitive, due to the high exhaust flow rates coupled with low volatile organic compound (VOC) and hazardous air pollutant (HAP) Concentrations. Past studies, conducted by the U.S. EPA and U.S. ...

  3. The fluid mechanics of natural ventilation

    NASA Astrophysics Data System (ADS)

    Linden, Paul

    1999-11-01

    Natural ventilation of buildings is the flow generated by temperature differences and by the wind. Modern buildings have extreme designs with large, tall open plan spaces and large cooling requirements. Natural ventilation offers a means of cooling these buildings and providing good indoor air quality. The essential feature of ventilation is an exchange between an interior space and the external ambient. Recent work shows that in many circumstances temperature variations play a controlling feature on the ventilation since the directional buoyancy force has a large influence on the flow patterns within the space and on the nature of the exchange with the outside. Two forms of buoyancy-driven ventilation are discussed: mixing ventilation in which the interior is at approximately uniform temperature and displacement ventilation where there is strong internal stratification. The dynamics of these flows are considered and the effects of wind on them are examined both experimentally and theoretically. The aim behind this work is to give designers rules and intuition on how air moves within a building and the research shows a fascinating branch of fluid mechanics.

  4. Influence of ventilation structure on air flow distribution of large turbo-generator

    NASA Astrophysics Data System (ADS)

    Zhang, Liying; Ding, Shuye; Zhao, Zhijun; Yang, Jingmo

    2018-04-01

    For the 350 MW air - cooled turbo—generator, the rotor body is ventilated by sub -slots and 94 radial ventilation ducts and the end adopts arc segment and the straight section to acquire the wind. The stator is ventilated with five inlets and eight outlet air branches. In order to analyze the cooling effect of different ventilation schemes, a global physical model including the stator, rotor, casing and fan is established, and the assumptions and boundary conditions of the solution domain are given. the finite volume method is used to solve the problem, and the air flow distribution characteristics of each part of the motor under different ventilation schemes are obtained. The results show that the baffle at the end of the rotor can eliminate the eddy current at the end of the rotor, and make the flow distribution of cooling air more uniform and reasonable. The conclusions can provide reference for the design of motor ventilation structure.

  5. Oxygen delivery using neonatal self-inflating bags without reservoirs.

    PubMed

    Sugiura, Takahiro; Urushibata, Rei; Komatsu, Kenji; Shioda, Tsutomu; Ota, Tatsuki; Sato, Megumi; Okubo, Yumiko; Fukuoka, Tetsuya; Hosono, Shigeharu; Tamura, Masanori

    2017-02-01

    Guidelines recommend avoiding excessive oxygen during neonatal resuscitation. Recent studies have suggested that oxygen titration can be achieved using a self-inflating bag, but data on the effectiveness of resuscitators used in neonatal ventilation are scarce, The aim of this study was therefore to determine the amount of oxygen delivered using several brands of neonatal self-inflating resuscitation bags without reservoirs under different conditions with regard to oxygen flow rate, ventilation rate (VR), peak inspiratory pressure (PIP) range, and test lung compliance. Oxygen concentration was measured under a variety of conditions. Combinations of oxygen flow rate (10, 5.0, 3.0 and 1.0 L/min), VR (40, 60 inflations/min), PIP range (20-25 cmH 2 O, 35-40 cmH 2 O), and test lung compliance (0.6, 1.0, 3.0, and 5.0 mL/cmH 2 O) were examined using six kinds of self-inflating bag. Delivered oxygen concentration varied widely (30.1-96.7%) and had a significant positive correlation with gas flow rate in all of the bags. Delivered oxygen concentration was also negatively correlated with PIP in all of the bags and with VR in some of them. Test lung compliance did not affect delivered oxygen concentration. The use of neonatal resuscitation self-inflating bags without reservoirs resulted in different delivered oxygen concentrations depending on gas flow rate, VR, PIP, and manufacturer, but not on lung compliance. This suggests that targeted oxygen concentrations could be delivered, even in lungs with decreased compliance, during resuscitation. © 2016 Japan Pediatric Society.

  6. Protective garment ventilation system

    NASA Technical Reports Server (NTRS)

    Lang, R. (Inventor)

    1970-01-01

    A method and apparatus for ventilating a protective garment, space suit system, and/or pressure suits to maintain a comfortable and nontoxic atmosphere within is described. The direction of flow of a ventilating and purging gas in portions of the garment may be reversed in order to compensate for changes in environment and activity of the wearer. The entire flow of the ventilating gas can also be directed first to the helmet associated with the garment.

  7. Increase in pulmonary blood flow at birth: role of oxygen and lung aeration.

    PubMed

    Lang, Justin A R; Pearson, James T; Binder-Heschl, Corinna; Wallace, Megan J; Siew, Melissa L; Kitchen, Marcus J; te Pas, Arjan B; Fouras, Andreas; Lewis, Robert A; Polglase, Graeme R; Shirai, Mikiyasu; Hooper, Stuart B

    2016-03-01

    Lung aeration stimulates the increase in pulmonary blood flow (PBF) at birth, but the spatial relationships between PBF and lung aeration and the role of increased oxygenation remain unclear. Using simultaneous phase-contrast X-ray imaging and angiography, we have investigated the separate roles of lung aeration and increased oxygenation in PBF changes at birth using near-term (30 days of gestation) rabbit kits (n = 18). Rabbits were imaged before ventilation, then the right lung was ventilated with 100% nitrogen (N2), air or 100% O2 (oxygen), before all kits were switched to ventilation in air, followed by ventilation of both lungs using air. Unilateral ventilation of the right lung with 100% N2 significantly increased heart rate (from 69.4 ± 4.9 to 93.0 ± 15.0 bpm), the diameters of both left and right pulmonary axial arteries, number of visible vessels in both left and right lungs, relative PBF index in both pulmonary arteries, and reduced bolus transit time for both left and right axial arteries (from 1.34 ± 0.39 and 1.81 ± 0.43 s to 0.52 ± 0.17 and 0.89 ± 0.21 s in the left and right axial arteries, respectively). Similar changes were observed with 100% oxygen, but increases in visible vessel number and vessel diameter of the axial arteries were greater in the ventilated right lung during unilateral ventilation. These findings confirm that PBF increase at birth is not spatially related to lung aeration and that the increase in PBF to unventilated regions is unrelated to oxygenation, although oxygen can potentiate this increase. © 2015 The Authors. The Journal of Physiology © 2015 The Physiological Society.

  8. An air flow sensor for neonatal mechanical ventilation applications based on a novel fiber-optic sensing technique

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Battista, L.; Sciuto, S. A.; Scorza, A.

    2013-03-15

    In this work, a simple and low-cost air flow sensor, based on a novel fiber-optic sensing technique has been developed for monitoring air flows rates supplied by a neonatal ventilator to support infants in intensive care units. The device is based on a fiber optic sensing technique allowing (a) the immunity to light intensity variations independent by measurand and (b) the reduction of typical shortcomings affecting all biomedical fields (electromagnetic interference and patient electrical safety). The sensing principle is based on the measurement of transversal displacement of an emitting fiber-optic cantilever due to action of air flow acting on it;more » the fiber tip displacement is measured by means of a photodiode linear array, placed in front of the entrance face of the emitting optical fiber in order to detect its light intensity profile. As the measurement system is based on a detection of the illumination pattern, and not on an intensity modulation technique, it results less sensitive to light intensity fluctuation independent by measurand than intensity-based sensors. The considered technique is here adopted in order to develop two different configurations for an air flow sensor suitable for the measurement of air flow rates typically occurring during mechanical ventilation of newborns: a mono-directional and a bi-directional transducer have been proposed. A mathematical model for the air flow sensor is here proposed and a static calibration of two different arrangements has been performed: a measurement range up to 3.00 Multiplication-Sign 10{sup -4} m{sup 3}/s (18.0 l/min) for the mono-directional sensor and a measurement range of {+-}3.00 Multiplication-Sign 10{sup -4} m{sup 3}/s ({+-}18.0 l/min) for the bi-directional sensor are experimentally evaluated, according to the air flow rates normally encountered during tidal breathing of infants with a mass lower than 10 kg. Experimental data of static calibration result in accordance with the proposed theoretical model: for the mono-directional configuration, the coefficient of determination r{sup 2} is equal to 0.997; for the bi-directional configuration, the coefficient of determination r{sup 2} is equal to 0.990 for positive flows (inspiration) and 0.988 for negative flows (expiration). Measurement uncertainty {delta}Q of air flow rate has been evaluated by means of the propagation of distributions and the percentage error in the arrangement of bi-directional sensor ranges from a minimum of about 0.5% at -18.0 l/min to a maximum of about 9% at -12.0 l/min.« less

  9. PADDLEFISH BUCCAL FLOW VELOCITY DURING RAM SUSPENSION FEEDING AND RAM VENTILATION

    PubMed

    Cech; Cheer

    1994-01-01

    A micro-thermistor probe was inserted into the buccal cavity of freely swimming paddlefish to measure flow velocity during ram ventilation, ram suspension feeding and prey processing. Swimming speed was measured from videotapes recorded simultaneously with the buccal flow velocity measurements. Both swimming velocity and buccal flow velocity were significantly higher during suspension feeding than during ram ventilation. As the paddlefish shifted from ventilation to feeding, buccal flow velocity increased to approximately 60 % of the swimming velocity. During prey processing, buccal flow velocity was significantly higher than the swimming velocity, indicating that prey processing involves the generation of suction. The Reynolds number (Re) for flow at the level of the paddlefish gill rakers during feeding is about 30, an order of magnitude lower than the Re calculated previously for pump suspension-feeding blackfish. These data, combined with data available from the literature, indicate that the gill rakers of ram suspension-feeding teleost fishes may operate at a substantially lower Re than the rakers of pump suspension feeders.

  10. Characterization of natural ventilation in wastewater collection systems.

    PubMed

    Ward, Matthew; Corsi, Richard; Morton, Robert; Knapp, Tom; Apgar, Dirk; Quigley, Chris; Easter, Chris; Witherspoon, Jay; Pramanik, Amit; Parker, Wayne

    2011-03-01

    The purpose of the study was to characterize natural ventilation in full-scale gravity collection system components while measuring other parameters related to ventilation. Experiments were completed at four different locations in the wastewater collection systems of Los Angeles County Sanitation Districts, Los Angeles, California, and the King County Wastewater Treatment District, Seattle, Washington. The subject components were concrete gravity pipes ranging in diameter from 0.8 to 2.4 m (33 to 96 in.). Air velocity was measured in each pipe using a carbon-monoxide pulse tracer method. Air velocity was measured entering or exiting the components at vents using a standpipe and hotwire anemometer arrangement. Ambient wind speed, temperature, and relative humidity; headspace temperature and relative humidity; and wastewater flow and temperature were measured. The field experiments resulted in a large database of measured ventilation and related parameters characterizing ventilation in full-scale gravity sewers. Measured ventilation rates ranged from 23 to 840 L/s. The experimental data was used to evaluate existing ventilation models. Three models that were based upon empirical extrapolation, computational fluid dynamics, and thermodynamics, respectively, were evaluated based on predictive accuracy compared to the measured data. Strengths and weaknesses in each model were found and these observations were used to propose a concept for an improved ventilation model.

  11. Impact of particle emissions of new laser printers on modeled office room

    NASA Astrophysics Data System (ADS)

    Koivisto, Antti J.; Hussein, Tareq; Niemelä, Raimo; Tuomi, Timo; Hämeri, Kaarle

    2010-06-01

    In this study, we present how an indoor aerosol model can be used to characterize particle emitter and predict influence of the source on indoor air quality. Particle size-resolved emission rates were quantified and the source's influence on indoor air quality was estimated by using office model simulations. We measured particle emissions from three modern laser printers in a flow-through chamber. Measured parameters were used as input parameters for an indoor aerosol model, which we then used to quantify the particle emission rates. The same indoor aerosol model was used to simulate the effect of the particle emission source inside an office model. The office model consists of a mechanically ventilated empty room and the particle source. The aerosol from the ventilation air was a filtered urban background aerosol. The effect of the ventilation rate was studied using three different ventilation ratios 1, 2 and 3 h -1. According to the model, peak emission rates of the printers exceeded 7.0 × 10 8 s -1 (2.5 × 10 12 h -1), and emitted mainly ultrafine particles (diameter less than 100 nm). The office model simulation results indicate that a print job increases ultrafine particle concentration to a maximum of 2.6 × 10 5 cm -3. Printer-emitted particles increased 6-h averaged particle concentration over eleven times compared to the background particle concentration.

  12. A randomized controlled trial of post-extubation bubble continuous positive airway pressure versus Infant Flow Driver continuous positive airway pressure in preterm infants with respiratory distress syndrome.

    PubMed

    Gupta, Samir; Sinha, Sunil K; Tin, Win; Donn, Steven M

    2009-05-01

    To compare the efficacy and safety of bubble continuous positive airway pressure (CPAP) and Infant Flow Driver (IFD) CPAP for the post-extubation management of preterm infants with respiratory distress syndrome (RDS). A total of 140 preterm infants at 24 to 29 weeks' gestation or with a birth weight of 600 to 1500 g who were ventilated at birth for RDS were randomized to receive either IFD CPAP (a variable-flow device) or bubble CPAP (a continuous-flow device). A standardized protocol was used for extubation and CPAP. No crossover was allowed. The primary outcome was successful extubation maintained for at least 72 hours. Secondary outcomes included successful extubation maintained for 7 days, total duration of CPAP support, chronic lung disease, and complications of prematurity. Seventy-one infants were randomized to bubble CPAP, and 69 were randomized to IFD CPAP. Mean gestational age and birth weight were similar in the 2 groups, as were the proportions of infants who achieved successful extubation for 72 hours and for 7 days. However, the median duration of CPAP support was 50% shorter in the infants on bubble CPAP. Moreover, in the subset of infants who were ventilated for less than 14 days, the infants on bubble CPAP had a significantly lower extubation failure rate. There was no difference in the incidence of chronic lung disease or other complications between the 2 study groups. Bubble CPAP is as effective as IFD CPAP in the post-extubation management of infants with RDS; however, in infants ventilated for < or = 14 days, bubble CPAP is associated with a significantly higher rate of successful extubation. Bubble CPAP also is associated with a significantly reduced duration of CPAP support.

  13. Numerical modeling of water spray suppression of conveyor belt fires in a large-scale tunnel.

    PubMed

    Yuan, Liming; Smith, Alex C

    2015-05-01

    Conveyor belt fires in an underground mine pose a serious life threat to miners. Water sprinkler systems are usually used to extinguish underground conveyor belt fires, but because of the complex interaction between conveyor belt fires and mine ventilation airflow, more effective engineering designs are needed for the installation of water sprinkler systems. A computational fluid dynamics (CFD) model was developed to simulate the interaction between the ventilation airflow, the belt flame spread, and the water spray system in a mine entry. The CFD model was calibrated using test results from a large-scale conveyor belt fire suppression experiment. Simulations were conducted using the calibrated CFD model to investigate the effects of sprinkler location, water flow rate, and sprinkler activation temperature on the suppression of conveyor belt fires. The sprinkler location and the activation temperature were found to have a major effect on the suppression of the belt fire, while the water flow rate had a minor effect.

  14. Numerical modeling of water spray suppression of conveyor belt fires in a large-scale tunnel

    PubMed Central

    Yuan, Liming; Smith, Alex C.

    2015-01-01

    Conveyor belt fires in an underground mine pose a serious life threat to miners. Water sprinkler systems are usually used to extinguish underground conveyor belt fires, but because of the complex interaction between conveyor belt fires and mine ventilation airflow, more effective engineering designs are needed for the installation of water sprinkler systems. A computational fluid dynamics (CFD) model was developed to simulate the interaction between the ventilation airflow, the belt flame spread, and the water spray system in a mine entry. The CFD model was calibrated using test results from a large-scale conveyor belt fire suppression experiment. Simulations were conducted using the calibrated CFD model to investigate the effects of sprinkler location, water flow rate, and sprinkler activation temperature on the suppression of conveyor belt fires. The sprinkler location and the activation temperature were found to have a major effect on the suppression of the belt fire, while the water flow rate had a minor effect. PMID:26190905

  15. Pulmonary blood flow distribution in sheep: effects of anesthesia, mechanical ventilation, and change in posture

    NASA Technical Reports Server (NTRS)

    Walther, S. M.; Domino, K. B.; Glenny, R. W.; Hlastala, M. P.

    1997-01-01

    BACKGROUND: Recent studies providing high-resolution images of pulmonary perfusion have questioned the classical zone model of pulmonary perfusion. Hence the present work was undertaken to provide detailed maps of regional pulmonary perfusion to examine the influence of anesthesia, mechanical ventilation, and posture. METHODS: Pulmonary perfusion was analyzed with intravenous fluorescent microspheres (15 microm) in six sheep studied in four conditions: prone and awake, prone with pentobarbital-anesthesia and breathing spontaneously, prone with anesthesia and mechanical ventilation, and supine with anesthesia and mechanical ventilation. Lungs were air dried at total lung capacity and sectioned into approximately 1,100 pieces (about 2 cm3) per animal. The pieces were weighed and assigned spatial coordinates. Fluorescence was read on a spectrophotometer, and signals were corrected for piece weight and normalized to mean flow. Pulmonary blood flow heterogeneity was assessed using the coefficient of variation of flow data. RESULTS: Pentobarbital anesthesia and mechanical ventilation did not influence perfusion heterogeneity, but heterogeneity increased when the animals were in the supine posture (P < 0.01). Gravitational flow gradients were absent in the prone position but present in the supine (P < 0.001 compared with zero). Pulmonary perfusion was distributed with a hilar-to-peripheral gradient in animals breathing spontaneously (P < 0.05). CONCLUSIONS: The influence of pentobarbital anesthesia and mechanical ventilation on pulmonary perfusion heterogeneity is small compared with the effect of changes in posture. Analysis of flow gradients indicate that gravity plays a small role in determining pulmonary blood flow distribution.

  16. Investigation of induced recirculation during planned ventilation system maintenance

    PubMed Central

    Pritchard, C.J.; Scott, D.F.; Noll, J.D.; Voss, B.; Leonis, D.

    2015-01-01

    The Office of Mine Safety and Health Research (OMSHR) investigated ways to increase mine airflow to underground metal/nonmetal (M/NM) mine working areas to improve miners’ health and safety. One of those areas is controlled recirculation. Because the quantity of mine air often cannot be increased, reusing part of the ventilating air can be an effective alternative, if implemented properly, until the capacity of the present system is improved. The additional airflow can be used to provide effective dilution of contaminants and higher flow velocities in the underground mine environment. Most applications of controlled recirculation involve taking a portion of the return air and passing it back into the intake to increase the air volume delivered to the desired work areas. OMSHR investigated a Nevada gold mine where shaft rehabilitation was in progress and one of the two main fans was shut down to allow reduced air velocity for safe shaft work. Underground booster fan operating pressures were kept constant to maintain airflow to work areas, inducing controlled recirculation in one work zone. Investigation into system behavior and the effects of recirculation on the working area during times of reduced primary ventilation system airflow would provide additional information on implementation of controlled recirculation into the system and how these events affect M/NM ventilation systems. The National Institute for Occupational Safety and Health monitored the ventilation district when both main fans were operating and another scenario with one of the units turned off for maintenance. Airflow and contaminants were measured to determine the exposure effects of induced recirculation on miner health. Surveys showed that 19% controlled recirculation created no change in the overall district airflow distribution and a small reduction in district fresh air intake. Total dust levels increased only modestly and respirable dust levels were also low. Diesel particulate matter (DPM) levels showed a high increase in district intake mass flow, but minor increases in exposure levels related to the recirculation percentage. Utilization of DPM mass flow rates allows input into ventilation modeling programs to better understand and plan for ventilation changes and district recirculation effects on miners’ health. PMID:26190862

  17. The specific effect of metronome guidance on the quality of one-person cardiopulmonary resuscitation and rescuer fatigue.

    PubMed

    Chung, Tae Nyoung; Kim, Sun Wook; You, Je Sung; Cho, Young Soon; Chung, Sung Phil; Park, Incheol; Kim, Seung Ho

    2012-12-01

    Metronome guidance is a simple and economic feedback method of guiding cardiopulmonary resuscitation (CPR). It has been proven for its usefulness in regulating the rate of chest compression and ventilation, but it is not yet clear how metronome use may affect compression depth or rescuer fatigue. The aim of this study was to assess the specific effect that metronome guidance has on the quality of CPR and rescuer fatigue. One-person CPRs were performed by senior medical students on Resusci Anne® manikins (Laerdal, Stavanger, Norway) with personal-computer skill-reporting systems. Half of the students performed CPR with metronome guidance and the other half without. CPR performance data, duration, and before-after trial differences in mean arterial pressure (MAP) and heart rate (HR) were compared between groups. Average compression depth (ACD) of the first five cycles, compression rate, no-flow fraction, and ventilation count were significantly lower in the metronome group (p=0.028, < 0.001, 0.001, and 0.041, respectively). Total CPR duration, total work (ACD × total compression count), and the before-after trial differences of the MAP and HR did not differ between the two groups. Metronome guidance is associated with lower chest compression depth of the first five cycles, while shortening the no-flow fraction and the ventilation count in a simulated one-person CPR model. Metronome guidance does not have an obvious effect of intensifying rescuer fatigue. Copyright © 2012 Elsevier Inc. All rights reserved.

  18. Operating Room Environment Control. Part A: a Valve Cannister System for Anesthetic Gas Adsorption. Part B: a State-of-the-art Survey of Laminar Flow Operating Rooms. Part C: Three Laminar Flow Experiments

    NASA Technical Reports Server (NTRS)

    Meyer, J. S.; Kosovich, J.

    1973-01-01

    An anesthetic gas flow pop-off valve canister is described that is airtight and permits the patient to breath freely. Once its release mechanism is activated, the exhaust gases are collected at a hose adapter and passed through activated coal for adsorption. A survey of laminar air flow clean rooms is presented and the installation of laminar cross flow air systems in operating rooms is recommended. Laminar flow ventilation experiments determine drying period evaporation rates for chicken intestines, sponges, and sections of pig stomach.

  19. System Design Verification for Closed Loop Control of Oxygenation With Concentrator Integration.

    PubMed

    Gangidine, Matthew M; Blakeman, Thomas C; Branson, Richard D; Johannigman, Jay A

    2016-05-01

    Addition of an oxygen concentrator into a control loop furthers previous work in autonomous control of oxygenation. Software integrates concentrator and ventilator function from a single control point, ensuring maximum efficiency by placing a pulse of oxygen at the beginning of the breath. We sought to verify this system. In a test lung, fraction of inspired oxygen (FIO2) levels and additional data were monitored. Tests were run across a range of clinically relevant ventilator settings in volume control mode, for both continuous flow and pulse dose flow oxygenation. Results showed the oxygen concentrator could maintain maximum pulse output (192 mL) up to 16 breaths per minute. Functionality was verified across ranges of tidal volumes and respiratory rates, with and without positive end-expiratory pressure, in continuous flow and pulse dose modes. For a representative test at respiratory rate 16 breaths per minute, tidal volume 550 mL, without positive end-expiratory pressure, pulse dose oxygenation delivered peak FIO2 of 76.83 ± 1.41%, and continuous flow 47.81 ± 0.08%; pulse dose flow provided a higher FIO2 at all tested setting combinations compared to continuous flow (p < 0.001). These tests verify a system that provides closed loop control of oxygenation while integrating time-coordinated pulse-doses from an oxygen concentrator. This allows the most efficient use of resources in austere environments. Reprint & Copyright © 2016 Association of Military Surgeons of the U.S.

  20. Influence of respiratory muscle work on VO(2) and leg blood flow during submaximal exercise.

    PubMed

    Wetter, T J; Harms, C A; Nelson, W B; Pegelow, D F; Dempsey, J A

    1999-08-01

    The work of breathing (W(b)) normally incurred during maximal exercise not only requires substantial cardiac output and O(2) consumption (VO(2)) but also causes vasoconstriction in locomotor muscles and compromises leg blood flow (Q(leg)). We wondered whether the W(b) normally incurred during submaximal exercise would also reduce Q(leg). Therefore, we investigated the effects of changing the W(b) on Q(leg) via thermodilution in 10 healthy trained male cyclists [maximal VO(2) (VO(2 max)) = 59 +/- 9 ml. kg(-1). min(-1)] during repeated bouts of cycle exercise at work rates corresponding to 50 and 75% of VO(2 max). Inspiratory muscle work was 1) reduced 40 +/- 6% via a proportional-assist ventilator, 2) not manipulated (control), or 3) increased 61 +/- 8% by addition of inspiratory resistive loads. Increasing the W(b) during submaximal exercise caused VO(2) to increase; decreasing the W(b) was associated with lower VO(2) (DeltaVO(2) = 0.12 and 0.21 l/min at 50 and 75% of VO(2 max), respectively, for approximately 100% change in W(b)). There were no significant changes in leg vascular resistance (LVR), norepinephrine spillover, arterial pressure, or Q(leg) when W(b) was reduced or increased. Why are LVR, norepinephrine spillover, and Q(leg) influenced by the W(b) at maximal but not submaximal exercise? We postulate that at submaximal work rates and ventilation rates the normal W(b) required makes insufficient demands for VO(2) and cardiac output to require any cardiovascular adjustment and is too small to activate sympathetic vasoconstrictor efferent output. Furthermore, even a 50-70% increase in W(b) during submaximal exercise, as might be encountered in conditions where ventilation rates and/or inspiratory flow resistive forces are higher than normal, also does not elicit changes in LVR or Q(leg).

  1. Interactive effects of mechano- and chemo-receptor inputs on cardiorespiratory outputs in the toad.

    PubMed

    Wang, T; Taylor, E W; Reid, S G; Milsom, W K

    2004-04-20

    Arterial blood pressure (P(b)), pulmocutaneous blood flow (Q(pc)), heart rate (f(H)), and fictive ventilation (motor activity in the Vth cranial nerve, V(int)), were recorded from decerebrated, paralysed toads receiving unidirectional ventilation with experimental gas mixtures over a range of lung inflation. At the onset of spontaneous bouts of fictive ventilation, (Q(pc)) and P(b) increased immediately, often with changes in heart rate, implying central cardiorespiratory interactions. Inflation of the lungs with different gas mixtures revealed that the effect of hypercarbia on V(int) was reduced by lung inflation and that feedback from pulmonary stretch receptors may summate with central feedforward control of f(H) and (Q(pc)) in an interactive fashion. The results of bolus injections of cyanide into the carotid or the pulmonary circulations suggest there are oxygen sensitive receptors in both circuits that affect the cardiovascular system directly and respiratory activity by complex central interactions with inputs from central chemoreceptors and pulmonary stretch receptors.

  2. Heat performance resulting from combined effects of radiation and mixed convection in a rectangular cavity ventilated by injection or suction

    NASA Astrophysics Data System (ADS)

    Ezzaraa, K.; Bahlaoui, A.; Arroub, I.; Raji, A.; Hasnaoui, M.; Naïmi, M.

    2018-05-01

    In this work, we investigated numerically heat transfer by mixed convection coupled to thermal radiation in a vented rectangular enclosure uniformly heated from below with a constant heat flux. The fresh fluid is admitted into the cavity by injection or suction, by means of two openings located on the lower part of both right and left vertical sides. Another opening is placed on the middle of the top wall to ensure the ventilation. Air, a radiatively transparent medium, is considered to be the cooling fluid. The inner surfaces, in contact with the fluid, are assumed to be gray, diffuse emitters and reflectors of radiation with identical emissivities. The effects of some pertinent parameters such as the Reynolds number, 300 ≤ Re ≤ 5000, and the emissivity of the walls, 0 ≤ ɛ ≤ 0.85, on flow and temperature patterns as well as on the heat transfer rate within the enclosure are presented for the two ventilation modes (injection and suction). The results indicate that the flow and thermal structures are affected by the thermal radiation for the two modes of imposed flow. However, the suction mode is found to be more favorable to the heat transfer in comparison with the injection one.

  3. Moisture transfer through the membrane of a cross-flow energy recovery ventilator: Measurement and simple data-driven modeling

    Treesearch

    CR Boardman; Samuel V. Glass

    2015-01-01

    The moisture transfer effectiveness (or latent effectiveness) of a cross-flow, membrane based energy recovery ventilator is measured and modeled. Analysis of in situ measurements for a full year shows that energy recovery ventilator latent effectiveness increases with increasing average relative humidity and surprisingly increases with decreasing average temperature. A...

  4. Feasibility and safety of low-flow extracorporeal CO2 removal managed with a renal replacement platform to enhance lung-protective ventilation of patients with mild-to-moderate ARDS.

    PubMed

    Schmidt, Matthieu; Jaber, Samir; Zogheib, Elie; Godet, Thomas; Capellier, Gilles; Combes, Alain

    2018-05-10

    Extracorporeal carbon-dioxide removal (ECCO 2 R) might allow ultraprotective mechanical ventilation with lower tidal volume (VT) (< 6 ml/kg predicted body weight), plateau pressure (P plat ) (< 30 cmH 2 O), and driving pressure to limit ventilator-induced lung injury. This study was undertaken to assess the feasibility and safety of ECCO 2 R managed with a renal replacement therapy (RRT) platform to enable very low tidal volume ventilation of patients with mild-to-moderate acute respiratory distress syndrome (ARDS). Twenty patients with mild (n = 8) or moderate (n = 12) ARDS were included. VT was gradually lowered from 6 to 5, 4.5, and 4 ml/kg, and PEEP adjusted to reach 23 ≤ P plat  ≤ 25 cmH 2 O. Standalone ECCO 2 R (no hemofilter associated with the RRT platform) was initiated when arterial PaCO 2 increased by > 20% from its initial value. Ventilation parameters (VT, respiratory rate, PEEP), respiratory system compliance, P plat and driving pressure, arterial blood gases, and ECCO 2 R-system operational characteristics were collected during at least 24 h of very low tidal volume ventilation. Complications, day-28 mortality, need for adjuvant therapies, and data on weaning off ECCO 2 R and mechanical ventilation were also recorded. While VT was reduced from 6 to 4 ml/kg and P plat kept < 25 cmH 2 O, PEEP was significantly increased from 13.4 ± 3.6 cmH 2 O at baseline to 15.0 ± 3.4 cmH 2 O, and the driving pressure was significantly reduced from 13.0 ± 4.8 to 7.9 ± 3.2 cmH 2 O (both p < 0.05). The PaO 2 /FiO 2 ratio and respiratory-system compliance were not modified after VT reduction. Mild respiratory acidosis occurred, with mean PaCO 2 increasing from 43 ± 8 to 53 ± 9 mmHg and mean pH decreasing from 7.39 ± 0.1 to 7.32 ± 0.10 from baseline to 4 ml/kg VT, while the respiratory rate was not altered. Mean extracorporeal blood flow, sweep-gas flow, and CO 2 removal were 421 ± 40 ml/min, 10 ± 0.3 L/min, and 51 ± 26 ml/min, respectively. Mean treatment duration was 31 ± 22 h. Day-28 mortality was 15%. A low-flow ECCO 2 R device managed with an RRT platform easily and safely enabled very low tidal volume ventilation with moderate increase in PaCO 2 in patients with mild-to-moderate ARDS. ClinicalTrials.gov, NCT02606240. Registered on 17 November 2015.

  5. Ventrain: an ejector ventilator for emergency use.

    PubMed

    Hamaekers, A E W; Borg, P A J; Enk, D

    2012-06-01

    A small, flow-regulated, manually operated ventilator designed for ventilation through a narrow-bore transtracheal catheter (TTC) has become available (Ventrain, Dolphys Medical BV, Eindhoven, The Netherlands). It is driven by a predetermined flow of oxygen from a high-pressure source and facilitates expiration by suction. The aim of this bench study was to test the efficacy of this new ventilator. The driving pressure, generated insufflation, and suction pressures and also the suction capacity of the Ventrain were measured at different oxygen flows. The minute volume achieved in an artificial lung through a TTC with an inner diameter (ID) of 2 mm was determined at different settings. Oxygen flows of 6-15 litre min(-1) resulted in driving pressures of 0.5-2.3 bar. Insufflation pressures, measured proximal to the TTC, ranged from 23 to 138 cm H(2)O. The maximal subatmospheric pressure build-up was -217 cm H(2)O. The suction capacity increased to a maximum of 12.4 litre min(-1) at an oxygen flow of 15 litre min(-1). At this flow, the achievable minute volume through the TTC ranged from 5.9 to 7.1 litres depending on the compliance of the artificial lung. The results of this bench study suggest that the Ventrain is capable of achieving a normal minute volume for an average adult through a 2 mm ID TTC. Further in vivo studies are needed to determine the value of the Ventrain as a portable emergency ventilator in a 'cannot intubate, cannot ventilate' situation.

  6. A new system for continuous and remote monitoring of patients receiving home mechanical ventilation

    NASA Astrophysics Data System (ADS)

    Battista, L.

    2016-09-01

    Home mechanical ventilation is the treatment of patients with respiratory failure or insufficiency by means of a mechanical ventilator at a patient's home. In order to allow remote patient monitoring, several tele-monitoring systems have been introduced in the last few years. However, most of them usually do not allow real-time services, as they have their own proprietary communication protocol implemented and some ventilation parameters are not always measured. Moreover, they monitor only some breaths during the whole day, despite the fact that a patient's respiratory state may change continuously during the day. In order to reduce the above drawbacks, this work reports the development of a novel remote monitoring system for long-term, home-based ventilation therapy; the proposed system allows for continuous monitoring of the main physical quantities involved during home-care ventilation (e.g., differential pressure, volume, and air flow rate) and is developed in order to allow observations of different remote therapy units located in different places of a city, region, or country. The developed remote patient monitoring system is able to detect various clinical events (e.g., events of tube disconnection and sleep apnea events) and has been successfully tested by means of experimental tests carried out with pulmonary ventilators typically used to support sick patients.

  7. A new system for continuous and remote monitoring of patients receiving home mechanical ventilation.

    PubMed

    Battista, L

    2016-09-01

    Home mechanical ventilation is the treatment of patients with respiratory failure or insufficiency by means of a mechanical ventilator at a patient's home. In order to allow remote patient monitoring, several tele-monitoring systems have been introduced in the last few years. However, most of them usually do not allow real-time services, as they have their own proprietary communication protocol implemented and some ventilation parameters are not always measured. Moreover, they monitor only some breaths during the whole day, despite the fact that a patient's respiratory state may change continuously during the day. In order to reduce the above drawbacks, this work reports the development of a novel remote monitoring system for long-term, home-based ventilation therapy; the proposed system allows for continuous monitoring of the main physical quantities involved during home-care ventilation (e.g., differential pressure, volume, and air flow rate) and is developed in order to allow observations of different remote therapy units located in different places of a city, region, or country. The developed remote patient monitoring system is able to detect various clinical events (e.g., events of tube disconnection and sleep apnea events) and has been successfully tested by means of experimental tests carried out with pulmonary ventilators typically used to support sick patients.

  8. U.S. Navy Unmanned Test Methods and Performance Limits for Underwater Breathing Apparatus

    DTIC Science & Technology

    2015-06-01

    in liters per minute (STPD) ?̇?2 Metabolic oxygen consumption in liters per minute (STPD) max Maximum flow rate ?̇? Ventilation , first time...8-2 8.1.3 Oxygen consumption ... OXYGEN CONSUMPTION SIMULATION ........................................................................... 9-1 CHAPTER 9. NEDU TM 15-01 vii 9-1

  9. On the mound of Macrotermes michaelseni as an organ of respiratory gas exchange.

    PubMed

    Turner, J S

    2001-01-01

    Patterns and rates of air movements in the mounds and nests of Macrotermes michaelseni were studied using tracer methods. Wind is a significant source of energy for powering nest ventilation, despite the mound being a completely enclosed structure. Nests are ventilated by a tidal movement of air driven by temporal variation in wind speed and wind direction. Density gradients sufficiently steep to drive bulk flow by natural convection will be rare. However, metabolism-induced buoyant forces may interact with wind energy in a way that promotes homeostasis of the mound atmosphere.

  10. Eszopiclone and Dexmedetomidine Depress Ventilation in Obese Rats with Features of Metabolic Syndrome

    PubMed Central

    Filbey, William A.; Sanford, David T.; Baghdoyan, Helen A.; Koch, Lauren G.; Britton, Steven L.; Lydic, Ralph

    2014-01-01

    Study Objectives: Obesity alters the therapeutic window of sedative/hypnotic drugs and increases the probability of respiratory complications. The current experiments used an established rodent model of obesity to test the hypothesis that the sedative/hypnotic drugs eszopiclone and dexmedetomidine alter ventilation differentially in obese rats compared with lean/fit rats. Design: This study used a within-groups/between-groups experimental design. Setting: University of Michigan. Participants: Experiments were conducted using lean/fit rats (n = 21) and obese rats (n = 21) that have features of metabolic syndrome. Interventions: Breathing was measured with whole-body plethysmography after systemic administration of vehicle (control), the nonbenzodiazepine, benzodiazepine site agonist eszopiclone, or the alpha-2 adrenergic receptor agonist dexmedetomidine. Measurements and Results: Data were analyzed using two-way analysis of variance and appropriate post hoc comparisons. At baseline, the obese/metabolic syndrome rats had increased respiratory rates (21.6%), lower tidal volumes/body weight (-24.1%), and no differences in minute ventilation compared to lean/fit rats. In the obese rats, respiratory rate was decreased by dexmedetomidine (-29%), but not eszopiclone. In the lean and the obese rats, eszopiclone decreased tidal volume (-12%). Both sedative/hypnotic drugs caused a greater decrease in minute ventilation in the obese (-26.3%) than lean (-18%) rats. Inspiratory flow rate (VT / TI) of the obese rats was decreased by dexmedetomidine (-10.6%) and eszopiclone (-18%). Duty cycle (TI / TTOT) in both rat lines was decreased by dexmedetomidine (-16.5%) but not by eszopiclone. Conclusions: Dexmedetomidine, in contrast to eszopiclone, decreased minute ventilation in the obese/metabolic syndrome rats by depressing both duty cycle and inspiratory flow rate. The results show for the first time that the obese phenotype differentially modulates the respiratory effects of eszopiclone and dexmedetomidine. These differences in breathing are consistent with previously documented differences in sleep between lean/fit and obese rats. These findings also encourage future studies of obese/metabolic syndrome rats that quantify the effect of sedative/hypnotic drugs on respiratory mechanics as well as hypoxic and hypercapnic ventilatory responses. Continued findings of favorable homology between obese humans and rodents will support the interpretation that these obese rats offer a unique animal model for mechanistic studies. Citation: Filbey WA, Sanford DT, Baghdoyan HA, Koch LG, Britton SL, Lydic R. Eszopiclone and dexmedetomidine depress ventilation in obese rats with features of metabolic syndrome. SLEEP 2014;37(5):871-880. PMID:24790265

  11. Physiology in Medicine: Understanding dynamic alveolar physiology to minimize ventilator-induced lung injury.

    PubMed

    Nieman, Gary F; Satalin, Josh; Kollisch-Singule, Michaela; Andrews, Penny; Aiash, Hani; Habashi, Nader M; Gatto, Louis A

    2017-06-01

    Acute respiratory distress syndrome (ARDS) remains a serious clinical problem with the main treatment being supportive in the form of mechanical ventilation. However, mechanical ventilation can be a double-edged sword: if set improperly, it can exacerbate the tissue damage caused by ARDS; this is known as ventilator-induced lung injury (VILI). To minimize VILI, we must understand the pathophysiologic mechanisms of tissue damage at the alveolar level. In this Physiology in Medicine paper, the dynamic physiology of alveolar inflation and deflation during mechanical ventilation will be reviewed. In addition, the pathophysiologic mechanisms of VILI will be reviewed, and this knowledge will be used to suggest an optimal mechanical breath profile (MB P : all airway pressures, volumes, flows, rates, and the duration that they are applied at both inspiration and expiration) necessary to minimize VILI. Our review suggests that the current protective ventilation strategy, known as the "open lung strategy," would be the optimal lung-protective approach. However, the viscoelastic behavior of dynamic alveolar inflation and deflation has not yet been incorporated into protective mechanical ventilation strategies. Using our knowledge of dynamic alveolar mechanics (i.e., the dynamic change in alveolar and alveolar duct size and shape during tidal ventilation) to modify the MB P so as to minimize VILI will reduce the morbidity and mortality associated with ARDS. Copyright © 2017 the American Physiological Society.

  12. A numerical study on the influence of slope and curvature on smoke flow in special section tunnel with natural ventilation

    NASA Astrophysics Data System (ADS)

    Wang, Wenzhou; Zhou, Xianping; Liu, Zhigang; Liu, Ya; Liu, Wanfu; Hong, Li

    2017-09-01

    In this study, a special section tunnel model was established by using FDS (Fire Dynamics Simulator). The influences of lope and curvature on smoke flow under natural ventilation have been studied. The results showed that under the condition of natural ventilation, the slope has some influences on the smoke flow in special section tunnel. The smoke spreading speed is accelerated along the upstream direction and decrease along the downstream direction due to buoyancy effect of slope. The steeper the tunnel, the more obvious the buoyancy effect. The curvature has little effect on the flow of flue gas.

  13. Impact of Room Ventilation Rates on Mouse Cage Ventilation and Microenvironment.

    PubMed

    Reeb, Carolyn K.; Jones, Robert B.; Bearg, David W.; Bedigian, Hendrick; Paigen, Beverly

    1997-01-01

    To assess the impact of room ventilation on animal cage microenvironment, intracage ventilation rate, temperature, humidity, and concentrations of carbon dioxide and ammonia were monitored in nonpressurized, bonnet-topped mouse cages. Cages on the top, middle, and bottom rows of a mouse rack were monitored at room ventilation rates of 0, 5, 10, and 20 air changes/h (ACH). Ventilation inside the animal cage increased somewhat from 12.8 to 18.9 ACH as room ventilation rate in- creased from 0 to 20 ACH, but the differences were not statistically significant, and most of the increase occurred in cages in the top row nearest to the fresh air supply. Cages containing mice had ventilation rate between 10 and 15 ACH even when room ventilation was reduced to 0 ACH; this ventilation is a result of the thermal heat load of the mice. After 6 days of soiled bedding, intracage ammonia concentration was c 3 ppm at all room ventilation rates and was not affected by increasing room ventilation. Temperature inside cages did not change with increasing ventilation. Humidity inside cages significantly decreased with increasing ventilation, from 55% relative humidity at 5 ACH to 36% relative humidity at 20 ACH. Carbon dioxide concentration decreased from 2,500 ppm to 1,900 ppm when ventilation rate increased from 5 ACH to 10 ACH, but no further significant decrease was observed at 20 ACH. In conclusion, increasing the room ventilation rate higher than 5 ACH did not result in significant improvements in the cage microenvironment.

  14. A comparison between tracer gas and aerosol particles distribution indoors: The impact of ventilation rate, interaction of airflows, and presence of objects.

    PubMed

    Bivolarova, M; Ondráček, J; Melikov, A; Ždímal, V

    2017-11-01

    The study investigated the separate and combined effects of ventilation rate, free convection flow produced by a thermal manikin, and the presence of objects on the distribution of tracer gas and particles in indoor air. The concentration of aerosol particles and tracer gas was measured in a test room with mixing ventilation. Three layouts were arranged: an empty room, an office room with an occupant sitting in front of a table, and a single-bed hospital room. The room occupant was simulated by a thermal manikin. Monodisperse particles of three sizes (0.07, 0.7, and 3.5 μm) and nitrous oxide tracer gas were generated simultaneously at the same location in the room. The particles and gas concentrations were measured in the bulk room air, in the breathing zone of the manikin, and in the exhaust air. Within the breathing zone of the sitting occupant, the tracer gas emerged as reliable predictor for the exposure to all different-sized test particles. A change in the ventilation rate did not affect the difference in concentration distribution between tracer gas and larger particle sizes. Increasing the room surface area did not influence the similarity in the dispersion of the aerosol particles and the tracer gas. © 2017 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

  15. FLOW-i ventilator performance in the presence of a circle system leak.

    PubMed

    Lucangelo, Umberto; Ajčević, Miloš; Accardo, Agostino; Borelli, Massimo; Peratoner, Alberto; Comuzzi, Lucia; Zin, Walter A

    2017-04-01

    Recently, the FLOW-i anaesthesia ventilator was developed based on the SERVO-i intensive care ventilator. The aim of this study was to test the FLOW-i's tidal volume delivery in the presence of a leak in the breathing circuit. We ventilated a test lung model in volume-, pressure-, and pressure-regulated volume-controlled modes (VC, PC, and PRVC, respectively) with a FLOW-i. First, the circuit remained airtight and the ventilator was tested with fresh gas flows of 6, 1, and 0.3 L/min in VC, PC, and PRVC modes and facing 4 combinations of different resistive and elastic loads. Second, a fixed leak in the breathing circuit was introduced and the measurements repeated. In the airtight system, FLOW-i maintained tidal volume (VT) and circuit pressure at approximately the set values, independently of respiratory mode, load, or fresh gas flow. In the leaking circuit, set VT = 500 mL, FLOW-i delivered higher VTs in PC (about 460 mL) than in VC and PRVC, where VTs were substantially less than 500 mL. Interestingly, VT did not differ appreciably from 6 to 0.3 L/min of fresh air flow among the 3 ventilatory modes. In the absence of leakage, peak inspiratory pressures were similar, while they were 35-45 % smaller in PRVC and VC than in PC mode in the presence of leaks. In conclusion, FLOW-i maintained VT (down to fresh gas flows of 0.3 L/min) to 90 % of its preset value in PC mode, which was 4-5 times greater than in VC or PRVC modes.

  16. Measurement of lung volume in mechanically ventilated monkeys with an ultrasonic flow meter and the nitrogen washout method.

    PubMed

    Schibler, Andreas; Hammer, Jürg; Isler, Ruedi; Buess, Christian; Newth, Christopher J L

    2004-01-01

    Measurement of functional residual capacity (FRC) during mechanical ventilation is important to standardise respiratory system compliance and adjust the ventilator settings to optimise lung recruitment. In the present study we compared three methods to measure FRC. The bias flow nitrogen washout technique (FRC(N2MC)), the multiple breath nitrogen washout (FRC(MBNW)) and the multiple breath sulphur-hexafluoride washout using the molar mass signal of an ultrasonic flow meter (FRC(MBSF6)) were compared in six adult monkeys after endotracheal intubation and during spontaneous breathing and mechanical ventilation at three different positive end-expiratory pressure (PEEP) levels of 0, 5 and 10 cmH2O. Animal research laboratory. We found good agreement between all three methods and they all accurately measured changes in FRC when PEEP was increased. The coefficients of variance of the three measurement techniques were in the same range (1.3-9.2%). The measurement of the tracer gas concentration with the molar mass signal of the ultrasonic flow meter provides a good and simple alternative to respiratory mass spectrometer for FRC measurements in ventilated subjects.

  17. Are there benefits or harm from pressure targeting during lung-protective ventilation?

    PubMed

    MacIntyre, Neil R; Sessler, Curtis N

    2010-02-01

    Mechanically, breath design is usually either flow/volume-targeted or pressure-targeted. Both approaches can effectively provide lung-protective ventilation, but they prioritize different ventilation parameters, so their responses to changing respiratory-system mechanics and patient effort are different. These different response behaviors have advantages and disadvantages that can be important in specific circumstances. Flow/volume targeting guarantees a set minute ventilation but sometimes may be difficult to synchronize with patient effort, and it will not limit inspiratory pressure. In contrast, pressure targeting, with its variable flow, may be easier to synchronize and will limit inspiratory pressure, but it provides no control over delivered volume. Skilled clinicians can maximize benefits and minimize problems with either flow/volume targeting or pressure targeting. Indeed, as is often the case in managing complex life-support devices, it is operator expertise rather than the device design features that most impacts patient outcomes.

  18. 30 CFR 57.22213 - Air flow (III mines).

    Code of Federal Regulations, 2010 CFR

    2010-07-01

    ... 30 Mineral Resources 1 2010-07-01 2010-07-01 false Air flow (III mines). 57.22213 Section 57.22213... Methane in Metal and Nonmetal Mines Ventilation § 57.22213 Air flow (III mines). The quantity of air coursed through the last open crosscut in pairs or sets of entries, or through other ventilation openings...

  19. Mathematics of Ventilator-induced Lung Injury.

    PubMed

    Rahaman, Ubaidur

    2017-08-01

    Ventilator-induced lung injury (VILI) results from mechanical disruption of blood-gas barrier and consequent edema and releases of inflammatory mediators. A transpulmonary pressure (P L ) of 17 cmH 2 O increases baby lung volume to its anatomical limit, predisposing to VILI. Viscoelastic property of lung makes pulmonary mechanics time dependent so that stress (P L ) increases with respiratory rate. Alveolar inhomogeneity in acute respiratory distress syndrome acts as a stress riser, multiplying global stress at regional level experienced by baby lung. Limitation of stress (P L ) rather than strain (tidal volume [V T ]) is the safe strategy of mechanical ventilation to prevent VILI. Driving pressure is the noninvasive surrogate of lung strain, but its relations to P L is dependent on the chest wall compliance. Determinants of lung stress (V T , driving pressure, positive end-expiratory pressure, and inspiratory flow) can be quantified in terms of mechanical power, and a safe threshold can be determined, which can be used in decision-making between safe mechanical ventilation and extracorporeal lung support.

  20. Preventing Airborne Disease Transmission: Review of Methods for Ventilation Design in Health Care Facilities

    PubMed Central

    Aliabadi, Amir A.; Rogak, Steven N.; Bartlett, Karen H.; Green, Sheldon I.

    2011-01-01

    Health care facility ventilation design greatly affects disease transmission by aerosols. The desire to control infection in hospitals and at the same time to reduce their carbon footprint motivates the use of unconventional solutions for building design and associated control measures. This paper considers indoor sources and types of infectious aerosols, and pathogen viability and infectivity behaviors in response to environmental conditions. Aerosol dispersion, heat and mass transfer, deposition in the respiratory tract, and infection mechanisms are discussed, with an emphasis on experimental and modeling approaches. Key building design parameters are described that include types of ventilation systems (mixing, displacement, natural and hybrid), air exchange rate, temperature and relative humidity, air flow distribution structure, occupancy, engineered disinfection of air (filtration and UV radiation), and architectural programming (source and activity management) for health care facilities. The paper describes major findings and suggests future research needs in methods for ventilation design of health care facilities to prevent airborne infection risk. PMID:22162813

  1. Constructing a generalized network design model to study air distribution in ventilation networks in subway with a single-track tunnel

    NASA Astrophysics Data System (ADS)

    Lugin, IV

    2018-03-01

    In focus are the features of construction of the generalized design model for the network method to study air distribution in ventilation system in subway with the single-track tunnel. The generalizations, assumptions and simplifications included in the model are specified. The air distribution is calculated with regard to the influence of topology and air resistances of the ventilation network sections. The author studies two variants of the subway line: half-open and closed with dead end on the both sides. It is found that the total air exchange at a subway station depends on the station location within the line. The operating mode of fans remains unaltered in this case. The article shows that elimination of air leakage in the station ventilation room allows an increase in the air flow rate by 7–8% at the same energy consumption by fans. The influence of the stop of a train in the tunnel on the air distribution is illustrated.

  2. The development of rotary drum dryer for palm fruit sterilization

    NASA Astrophysics Data System (ADS)

    Hanifarianty, S.; Legwiriyakul, A.; Alimalbari, A.; Nuntadusit, C.; Theppaya, T.; Wae-Hayee, M.

    2018-01-01

    The aim of this research was to design and develop a rotary drum dryer for palm fruit sterilization. In this article, the results of the effect of ventilation hole number on the reduction of moisture content in palm fruit were presented. The experimental set up was a drum dryer which has 57.5 cm in a diameter and 90 cm in a length (the size was similar to 200-littre steel drum container). A driving gear and a gear motor rotated the drum dryer. The ventilation hole were drilled on the lateral side of the drum. The diameter of ventilation hole was 10 mm, and the number of ventilation hole were 18, 36 and 72 hole (each side was 9, 18 and 36 hole, respectively). In the experiment, the palm fruit was dried by using LPG to burn and heat the bottom of the drum. The flow rate of LPG was controlled to keep the temperature inside the drum steadily at 120°C.

  3. Integration of a photocatalytic multi-tube reactor for indoor air purification in HVAC systems: a feasibility study.

    PubMed

    van Walsem, Jeroen; Roegiers, Jelle; Modde, Bart; Lenaerts, Silvia; Denys, Siegfried

    2018-04-24

    This work is focused on an in-depth experimental characterization of multi-tube reactors for indoor air purification integrated in ventilation systems. Glass tubes were selected as an excellent photocatalyst substrate to meet the challenging requirements of the operating conditions in a ventilation system in which high flow rates are typical. Glass tubes show a low-pressure drop which reduces the energy demand of the ventilator, and additionally, they provide a large exposed surface area to allow interaction between indoor air contaminants and the photocatalyst. Furthermore, the performance of a range of P25-loaded sol-gel coatings was investigated, based on their adhesion properties and photocatalytic activities. Moreover, the UV light transmission and photocatalytic reactor performance under various operating conditions were studied. These results provide vital insights for the further development and scaling up of multi-tube reactors in ventilation systems which can provide a better comfort, improved air quality in indoor environments, and reduced human exposure to harmful pollutants.

  4. Preventing airborne disease transmission: review of methods for ventilation design in health care facilities.

    PubMed

    Aliabadi, Amir A; Rogak, Steven N; Bartlett, Karen H; Green, Sheldon I

    2011-01-01

    Health care facility ventilation design greatly affects disease transmission by aerosols. The desire to control infection in hospitals and at the same time to reduce their carbon footprint motivates the use of unconventional solutions for building design and associated control measures. This paper considers indoor sources and types of infectious aerosols, and pathogen viability and infectivity behaviors in response to environmental conditions. Aerosol dispersion, heat and mass transfer, deposition in the respiratory tract, and infection mechanisms are discussed, with an emphasis on experimental and modeling approaches. Key building design parameters are described that include types of ventilation systems (mixing, displacement, natural and hybrid), air exchange rate, temperature and relative humidity, air flow distribution structure, occupancy, engineered disinfection of air (filtration and UV radiation), and architectural programming (source and activity management) for health care facilities. The paper describes major findings and suggests future research needs in methods for ventilation design of health care facilities to prevent airborne infection risk.

  5. Application of 1 D Finite Element Method in Combination with Laminar Solution Method for Pipe Network Analysis

    NASA Astrophysics Data System (ADS)

    Dudar, O. I.; Dudar, E. S.

    2017-11-01

    The features of application of the 1D dimensional finite element method (FEM) in combination with the laminar solutions method (LSM) for the calculation of underground ventilating networks are considered. In this case the processes of heat and mass transfer change the properties of a fluid (binary vapour-air mix). Under the action of gravitational forces it leads to such phenomena as natural draft, local circulation, etc. The FEM relations considering the action of gravity, the mass conservation law, the dependence of vapour-air mix properties on the thermodynamic parameters are derived so that it allows one to model the mentioned phenomena. The analogy of the elastic and plastic rod deformation processes to the processes of laminar and turbulent flow in a pipe is described. Owing to this analogy, the guaranteed convergence of the elastic solutions method for the materials of plastic type means the guaranteed convergence of the LSM for any regime of a turbulent flow in a rough pipe. By means of numerical experiments the convergence rate of the FEM - LSM is investigated. This convergence rate appeared much higher than the convergence rate of the Cross - Andriyashev method. Data of other authors on the convergence rate comparison for the finite element method, the Newton method and the method of gradient are provided. These data allow one to conclude that the FEM in combination with the LSM is one of the most effective methods of calculation of hydraulic and ventilating networks. The FEM - LSM has been used for creation of the research application programme package “MineClimate” allowing to calculate the microclimate parameters in the underground ventilating networks.

  6. On the calculation of air flow rates to ventilate closed-type stations in subway with the double-track tunnel

    NASA Astrophysics Data System (ADS)

    Kiyanitsa, LA

    2018-03-01

    Metro is not only the most promising kind of public transport but also an important part of infrastructure in a modern city. As a place where large groups of people gather, subway is to ensure the required air exchange at the passenger platforms of the stations. The air flow rate for airing the stations is also determined based on the required temperature, humidity and MAC of gases. The present study estimates the required air flow rate at the passenger platform of the closed-type subway station with the double-track tunnel given the standard air temperature, humidity and gas concentration, as well as based on the condition of the specified air flow feed and air changes per hour. The article proposes the scheme of air recirculation from the double-track tunnel to the station.

  7. Effect of body position and ventilation on umbilical artery and venous blood flows during delayed umbilical cord clamping in preterm lambs.

    PubMed

    Hooper, Stuart B; Crossley, Kelly J; Zahra, Valerie A; van Vonderen, Jeroen; Moxham, Alison; Gill, Andrew W; Kluckow, Martin; Te Pas, Arjan B; Wallace, Euan M; Polglase, Graeme R

    2017-07-01

    While delayed umbilical cord clamping (UCC) is thought to facilitate placental to infant blood transfusion, the physiological factors regulating flow in the umbilical arteries and veins during delayed UCC is unknown. We investigated the effects of gravity, by changing fetal height relative to the placenta, and ventilation on umbilical blood flows and the cardiovascular transition during delayed UCC at birth. Catheters and flow probes were implanted into preterm lambs (128 days) prior to delivery to measure pulmonary, carotid, umbilical artery (UaBF) and umbilical venous (UvBF) blood flows. Lambs were placed either 10 cm below or 10 cm above the ewe. Ventilation commenced 2-3 min before UCC and continued for 30 min after UCC. Gravity reduced umbilical and cerebral flows when lambs were placed below the midline, but the reduction in UaBF and UvBF was similar. Ventilation during delayed UCC reduced UvBF and UaBF by similar amounts, irrespective of the lamb's position, such that flows into and out of the placenta remained balanced. The effects of ventilation on umbilical flows were much greater than the effects of gravity, but no net placental to lamb blood transfusion could be detected under any condition. Cardiovascular parameters, cerebral oxygen kinetics and final blood volumes were similar in both groups 5 min after UCC. Gravity caused small transient effects on umbilical and cerebral flow, but given changes were similar in umbilical arteries and veins, no net placental transfusion was detected. Ventilation during delayed UCC has a markedly greater influence on cardiovascular function in the newborn. Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://www.bmj.com/company/products-services/rights-and-licensing/.

  8. Characteristics of rain penetration through a gravity ventilator used for natural ventilation.

    PubMed

    Kim, Taehyeung; Lee, Dong Ho; Ahn, Kwangseog; Ha, Hyunchul; Park, Heechang; Piao, Cheng Xu; Li, Xiaoyu; Seo, Jeoungyoon

    2008-01-01

    Gravity ventilators rely simply on air buoyancy to extract air and are widely used to exhaust air contaminants and heat from workplaces using minimal energy. They are designed to maximize the exhaust flow rate, but the rain penetration sometimes causes malfunctioning. In this study, the characteristics of rain penetration through a ventilator were examined as a preliminary study to develop a ventilator with the maximum exhaust capacity while minimizing rain penetration. A model ventilator was built and exposed to artificial rain and wind. The paths, intensities and amounts of penetration through the ventilator were observed and measured in qualitative and quantitative fashions. In the first phase, the pathways and intensities of rain penetration were visually observed. In the second phase, the amounts of rain penetration were quantitatively measured under the different configurations of ventilator components that were installed based on the information obtained in the first-phase experiment. The effects of wind speed, grill direction, rain drainage width, outer wall height, neck height and leaning angle of the outer wall from the vertical position were analyzed. Wind speed significantly affected rain penetration. Under the low crosswind conditions, the rain penetration intensities were under the limit of detection. Under the high crosswind conditions, grill direction and neck height were the most significant factors in reducing rain penetration. The installation of rain drainage was also important in reducing rain penetration. The experimental results suggest that, with proper configurations of its components, a gravity ventilator can be used for natural ventilation without significant rain penetration problems.

  9. Regional gas transport in the heterogeneous lung during oscillatory ventilation

    PubMed Central

    Herrmann, Jacob; Tawhai, Merryn H.

    2016-01-01

    Regional ventilation in the injured lung is heterogeneous and frequency dependent, making it difficult to predict how an oscillatory flow waveform at a specified frequency will be distributed throughout the periphery. To predict the impact of mechanical heterogeneity on regional ventilation distribution and gas transport, we developed a computational model of distributed gas flow and CO2 elimination during oscillatory ventilation from 0.1 to 30 Hz. The model consists of a three-dimensional airway network of a canine lung, with heterogeneous parenchymal tissues to mimic effects of gravity and injury. Model CO2 elimination during single frequency oscillation was validated against previously published experimental data (Venegas JG, Hales CA, Strieder DJ, J Appl Physiol 60: 1025–1030, 1986). Simulations of gas transport demonstrated a critical transition in flow distribution at the resonant frequency, where the reactive components of mechanical impedance due to airway inertia and parenchymal elastance were equal. For frequencies above resonance, the distribution of ventilation became spatially clustered and frequency dependent. These results highlight the importance of oscillatory frequency in managing the regional distribution of ventilation and gas exchange in the heterogeneous lung. PMID:27763872

  10. Continuous High-Frequency Oscillation Therapy in Invasively Ventilated Pediatric Subjects in the Critical Care Setting.

    PubMed

    Morgan, Stephen; Hornik, Christoph P; Patel, Niyati; Williford, Walter L; Turner, David A; Cheifetz, Ira M

    2016-11-01

    Continuous high-frequency oscillation (CHFO) creates a pressure gradient in the small airways that accelerates expiratory flow. The intended use of CHFO therapy is to facilitate secretion removal and treat atelectasis. Our objective was to assess the feasibility, safety, and efficacy of CHFO in the mechanically ventilated pediatric population. After institutional review board approval, we retrospectively reviewed medical records of mechanically ventilated children treated with CHFO (the MetaNeb system) at our institution from July 1, 2007 through August 31, 2012. Patients supported with extracorporeal membrane oxygenation were excluded. We evaluated changes in ventilator settings in subjects with ventilator data documented within 6 h pre- and post-treatment. We evaluated arterial blood gas (ABG) results for individual treatments, comparing ABG results within 8 h pre-therapy to ABG results within 3 h post-treatment. Oxygen index and P aO 2 /F IO 2 were calculated. Demographic data, blood pressure, heart rate, and development of new air leak while being treated with CHFO were recorded. Pre- and post-CHFO measurements were compared using Wilcoxon signed-rank testing. Our cohort included 59 invasively ventilated subjects. Median age was 2 y (range 1 month to 19 y), and median weight was 14 kg (2-81 kg). We evaluated data on 528 total treatments (range per subject 1-39 treatments). Peak inspiratory pressure significantly decreased with CHFO, whereas other parameters, including P aCO 2 and breathing frequency, remained stable. There was no significant change in systolic blood pressure, diastolic blood pressure, or heart rate following treatment with CHFO. One subject (2%) developed a clinically insignificant pneumothorax during CHFO. CHFO is feasible and seems safe in our cohort of mechanically ventilated pediatric subjects. The rate of pneumothorax was consistent with that seen in similar pediatric ICU populations. These preliminary results suggest that CHFO may be beneficial by improving lung compliance in pediatric subjects with secretion-induced atelectasis. Prospective clinical studies are needed to further evaluate the clinical efficacy and safety of CHFO in children receiving invasive mechanical ventilation. Copyright © 2016 by Daedalus Enterprises.

  11. 30 CFR 57.22211 - Air flow (I-A mines).

    Code of Federal Regulations, 2010 CFR

    2010-07-01

    ... 30 Mineral Resources 1 2010-07-01 2010-07-01 false Air flow (I-A mines). 57.22211 Section 57.22211... Methane in Metal and Nonmetal Mines Ventilation § 57.22211 Air flow (I-A mines). The average air velocity in the last open crosscut in pairs or sets of developing entries, or through other ventilation...

  12. Key considerations on nebulization of antimicrobial agents to mechanically ventilated patients.

    PubMed

    Rello, J; Rouby, J J; Sole-Lleonart, C; Chastre, J; Blot, S; Luyt, C E; Riera, J; Vos, M C; Monsel, A; Dhanani, J; Roberts, J A

    2017-09-01

    Nebulized antibiotics have an established role in patients with cystic fibrosis or bronchiectasis. Their potential benefit to treat respiratory infections in mechanically ventilated patients is receiving increasing interest. In this consensus statement of the European Society of Clinical Microbiology and Infectious Diseases, the body of evidence of the therapeutic utility of aerosolized antibiotics in mechanically ventilated patients was reviewed and resulted in the following recommendations: Vibrating-mesh nebulizers should be preferred to jet or ultrasonic nebulizers. To decrease turbulence and limit circuit and tracheobronchial deposition, we recommend: (a) the use of specifically designed respiratory circuits avoiding sharp angles and characterized by smooth inner surfaces, (b) the use of specific ventilator settings during nebulization including use of a volume controlled mode using constant inspiratory flow, tidal volume 8 mL/kg, respiratory frequency 12 to 15 bpm, inspiratory:expiratory ratio 50%, inspiratory pause 20% and positive end-expiratory pressure 5 to 10 cm H 2 O and (c) the administration of a short-acting sedative agent if coordination between the patient and the ventilator is not obtained, to avoid patient's flow triggering and episodes of peak decelerating inspiratory flow. A filter should be inserted on the expiratory limb to protect the ventilator flow device and changed between each nebulization to avoid expiratory flow obstruction. A heat and moisture exchanger and/or conventional heated humidifier should be stopped during the nebulization period to avoid a massive loss of aerosolized particles through trapping and condensation. If these technical requirements are not followed, there is a high risk of treatment failure and adverse events in mechanically ventilated patients receiving nebulized antibiotics for pneumonia. Copyright © 2017 European Society of Clinical Microbiology and Infectious Diseases. Published by Elsevier Ltd. All rights reserved.

  13. Validity of thermally-driven small-scale ventilated filling box models

    NASA Astrophysics Data System (ADS)

    Partridge, Jamie L.; Linden, P. F.

    2013-11-01

    The majority of previous work studying building ventilation flows at laboratory scale have used saline plumes in water. The production of buoyancy forces using salinity variations in water allows dynamic similarity between the small-scale models and the full-scale flows. However, in some situations, such as including the effects of non-adiabatic boundaries, the use of a thermal plume is desirable. The efficacy of using temperature differences to produce buoyancy-driven flows representing natural ventilation of a building in a small-scale model is examined here, with comparison between previous theoretical and new, heat-based, experiments.

  14. Ventilation of an hydrofoil wake

    NASA Astrophysics Data System (ADS)

    Arndt, Roger; Lee, Seung Jae; Monson, Garrett

    2013-11-01

    Ventilation physics plays a role in a variety of important engineering applications. For example, hydroturbine ventilation is used for control of vibration and cavitation erosion and more recently for improving the dissolved oxygen content of the flow through the turbine. The latter technology has been the focus of an ongoing study involving the ventilation of an hydrofoil wake to determine the velocity and size distribution of bubbles in a bubbly wake. This was carried out by utilizing particle shadow velocimetry (PSV). This technique is a non-scattering approach that relies on direct in-line volume illumination by a pulsed source such as a light-emitting diode (LED). The data are compared with previous studies of ventilated flow. The theoretical results of Hinze suggest that a scaling relationship is possible that can lead to developing appropriate design parameters for a ventilation system. Sponsored by ONR and DOE.

  15. VA/Q distribution during heavy exercise and recovery in humans: implications for pulmonary edema

    NASA Technical Reports Server (NTRS)

    Schaffartzik, W.; Poole, D. C.; Derion, T.; Tsukimoto, K.; Hogan, M. C.; Arcos, J. P.; Bebout, D. E.; Wagner, P. D.

    1992-01-01

    Ventilation-perfusion (VA/Q) inequality has been shown to increase with exercise. Potential mechanisms for this increase include nonuniform pulmonary vasoconstriction, ventilatory time constant inequality, reduced large airway gas mixing, and development of interstitial pulmonary edema. We hypothesized that persistence of VA/Q mismatch after ventilation and cardiac output subside during recovery would be consistent with edema; however, rapid resolution would suggest mechanisms related to changes in ventilation and blood flow per se. Thirteen healthy males performed near-maximal cycle ergometry at an inspiratory PO2 of 91 Torr (because hypoxia accentuates VA/Q mismatch on exercise). Cardiorespiratory variables and inert gas elimination patterns were measured at rest, during exercise, and between 2 and 30 min of recovery. Two profiles of VA/Q distribution behavior emerged during heavy exercise: in group 1 an increase in VA/Q mismatch (log SDQ of 0.35 +/- 0.02 at rest and 0.44 +/- 0.02 at exercise; P less than 0.05, n = 7) and in group 2 no change in VA/Q mismatch (n = 6). There were no differences in anthropometric data, work rate, O2 uptake, or ventilation during heavy exercise between groups. Group 1 demonstrated significantly greater VA/Q inequality, lower vital capacity, and higher forced expiratory flow at 25-75% of forced vital capacity for the first 20 min during recovery than group 2. Cardiac index was higher in group 1 both during heavy exercise and 4 and 6 min postexercise. However, both ventilation and cardiac output returned toward baseline values more rapidly than did VA/Q relationships. Arterial pH was lower in group 1 during exercise and recovery. We conclude that greater VA/Q inequality in group 1 and its persistence during recovery are consistent with the hypothesis that edema occurs and contributes to the increase in VA/Q inequality during exercise. This is supported by observation of greater blood flows and acidosis and, presumably therefore, higher pulmonary vascular pressures in such subjects.

  16. Variable tidal volumes improve lung protective ventilation strategies in experimental lung injury.

    PubMed

    Spieth, Peter M; Carvalho, Alysson R; Pelosi, Paolo; Hoehn, Catharina; Meissner, Christoph; Kasper, Michael; Hübler, Matthias; von Neindorff, Matthias; Dassow, Constanze; Barrenschee, Martina; Uhlig, Stefan; Koch, Thea; de Abreu, Marcelo Gama

    2009-04-15

    Noisy ventilation with variable Vt may improve respiratory function in acute lung injury. To determine the impact of noisy ventilation on respiratory function and its biological effects on lung parenchyma compared with conventional protective mechanical ventilation strategies. In a porcine surfactant depletion model of lung injury, we randomly combined noisy ventilation with the ARDS Network protocol or the open lung approach (n = 9 per group). Respiratory mechanics, gas exchange, and distribution of pulmonary blood flow were measured at intervals over a 6-hour period. Postmortem, lung tissue was analyzed to determine histological damage, mechanical stress, and inflammation. We found that, at comparable minute ventilation, noisy ventilation (1) improved arterial oxygenation and reduced mean inspiratory peak airway pressure and elastance of the respiratory system compared with the ARDS Network protocol and the open lung approach, (2) redistributed pulmonary blood flow to caudal zones compared with the ARDS Network protocol and to peripheral ones compared with the open lung approach, (3) reduced histological damage in comparison to both protective ventilation strategies, and (4) did not increase lung inflammation or mechanical stress. Noisy ventilation with variable Vt and fixed respiratory frequency improves respiratory function and reduces histological damage compared with standard protective ventilation strategies.

  17. No effect of artificial gravity on lung function with exercise training during head-down bed rest

    NASA Astrophysics Data System (ADS)

    Su, Longxiang; Guo, Yinghua; Wang, Yajuan; Wang, Delong; Liu, Changting

    2016-04-01

    The aim of this study is to explore the effectiveness of microgravity simulated by head-down bed rest (HDBR) and artificial gravity (AG) with exercise on lung function. Twenty-four volunteers were randomly divided into control and exercise countermeasure (CM) groups for 96 h of 6° HDBR. Comparisons of pulse rate, pulse oxygen saturation (SpO2) and lung function were made between these two groups at 0, 24, 48, 72, 96 h. Compared with the sitting position, inspiratory capacity and respiratory reserve volume were significantly higher than before HDBR (0° position) (P < 0.05). Vital capacity, expiratory reserve volume, forced vital capacity, forced expiratory volume in 1 s, forced inspiratory vital capacity, forced inspiratory volume in 1 s, forced expiratory flow at 25, 50, and 75%, maximal mid-expiratory flow and peak expiratory flow were all significantly lower than those before HDBR (P < 0.05). Neither control nor CM groups showed significant differences in pulse rate, SpO2, pulmonary volume and pulmonary ventilation function over the HDBR observation time. Postural changes can lead to variation in lung volume and ventilation function, but a HDBR model induced no changes in pulmonary function and therefore should not be used to study AG countermeasures.

  18. The Use of Stepper Motor-Controlled Proportional Valve for Fio2 Calculation in the Ventilator and its Control with Fuzzy Logic.

    PubMed

    Gölcük, Adem; Güler, İnan

    2017-01-01

    This article proposes the employment of a proportional valve that can calculate the amount of oxygen in the air to be given to patient in accordance with the amount of FiO 2 which is set from the control menu of the ventilation device. To actualize this, a stepper motor-controlled proportional valve was used. Two counts of valves were employed in order to control the gases with 2 bar pressure that came from both the oxygen and medical air tanks. Oxygen and medical air manometers alongside the pressure regulators were utilized to perform this task. It is a fuzzy-logic-based controller which calculates at what rate the proportional valves will be opened and closed for FiO 2 calculation. Fluidity and pressure of air given by the ventilation device were tested with a FlowMeter while the oxygen level was tested using the electronic lung model. The obtained results from the study revealed that stepper motor controlled proportional valve could be safely used in ventilation devices. In this article, it was indicated that fluidity and pressure control could be carried out with just two counts of proportional valve, which could be done with many solenoid valves, so this reduces the cost of ventilator, electrical power consumed by the ventilator, and the dimension of ventilator.

  19. International Space Station USOS Crew Quarters Ventilation and Acoustic Design Implementation

    NASA Technical Reports Server (NTRS)

    Broyan, James Lee, Jr.

    2009-01-01

    The International Space Station (ISS) United States Operational Segment (USOS) has four permanent rack sized ISS Crew Quarters (CQ) providing a private crewmember space. The CQ uses Node 2 cabin air for ventilation/thermal cooling, as opposed to conditioned ducted air from the ISS Temperature Humidity Control System or the ISS fluid cooling loop connections. Consequently, CQ can only increase the air flow rate to reduce the temperature delta between the cabin and the CQ interior. However, increasing airflow causes increased acoustic noise so efficient airflow distribution is an important design parameter. The CQ utilized a two fan push-pull configuration to ensure fresh air at the crewmember s head position and reduce acoustic exposure. The CQ interior needs to be below Noise Curve 40 (NC-40). The CQ ventilation ducts are open to the significantly louder Node 2 cabin aisle way which required significantly acoustic mitigation controls. The design implementation of the CQ ventilation system and acoustic mitigation are very inter-related and require consideration of crew comfort balanced with use of interior habitable volume, accommodation of fan failures, and possible crew uses that impact ventilation and acoustic performance. This paper illustrates the types of model analysis, assumptions, vehicle interactions, and trade-offs required for CQ ventilation and acoustics. Additionally, on-orbit ventilation system performance and initial crew feedback is presented. This approach is applicable to any private enclosed space that the crew will occupy.

  20. Experimental system, and its evaluation for the control of surgically inducted infections

    NASA Technical Reports Server (NTRS)

    Tevebaugh, M. D.; Nelson, J. P.

    1972-01-01

    The effect is reported to design, fabricate, test and evaluate a prototype experimental system for the control of surgically induced infections. The purpose is to provide the cleanest possible environment within a hospital surgery room and eliminate contamination sources that could cause infections during surgery. The system design is described. The system provides for a portable laminar flow clean room, a full bubble helmet system with associated communications and ventilation subsystems for operating room personnel, and surgical gowns that minimize the migration of bacteria. The development test results consisting of portability, laminar flowrate, air flow pattern, electrostatic buildup, noise level, ventilation, human factors, electrical and material compatibility tests are summarized. The conclusions are that the experimental system is effective in reducing the airborne and wound contamination although the helmets and gowns may not be a significant part of this reduction. Definitive conclusions with regard to the infection rate cannot be made at this time.

  1. Unloading work of breathing during high-frequency oscillatory ventilation: a bench study

    PubMed Central

    van Heerde, Marc; Roubik, Karel; Kopelent, Vitek; Plötz, Frans B; Markhorst, Dick G

    2006-01-01

    Introduction With the 3100B high-frequency oscillatory ventilator (SensorMedics, Yorba Linda, CA, USA), patients' spontaneous breathing efforts result in a high level of imposed work of breathing (WOB). Therefore, spontaneous breathing often has to be suppressed during high-frequency oscillatory ventilation (HFOV). A demand-flow system was designed to reduce imposed WOB. Methods An external gas flow controller (demand-flow system) accommodates the ventilator fresh gas flow during spontaneous breathing simulation. A control algorithm detects breathing effort and regulates the demand-flow valve. The effectiveness of this system has been evaluated in a bench test. The Campbell diagram and pressure time product (PTP) are used to quantify the imposed workload. Results Using the demand-flow system, imposed WOB is considerably reduced. The demand-flow system reduces inspiratory imposed WOB by 30% to 56% and inspiratory imposed PTP by 38% to 59% compared to continuous fresh gas flow. Expiratory imposed WOB was decreased as well by 12% to 49%. In simulations of shallow to normal breathing for an adult, imposed WOB is 0.5 J l-1 at maximum. Fluctuations in mean airway pressure on account of spontaneous breathing are markedly reduced. Conclusion The use of the demand-flow system during HFOV results in a reduction of both imposed WOB and fluctuation in mean airway pressure. The level of imposed WOB was reduced to the physiological range of WOB. Potentially, this makes maintenance of spontaneous breathing during HFOV possible and easier in a clinical setting. Early initiation of HFOV seems more possible with this system and the possibility of weaning of patients directly on a high-frequency oscillatory ventilator is not excluded either. PMID:16848915

  2. REDUCTION IN INSPIRATORY FLOW ATTENUATES IL-8 RELEASE AND MAPK ACTIVATION OF LUNG OVERSTRETCH

    EPA Science Inventory

    Lung overstretch involves mechanical factors, including large tidal volumes (VT), which induce inflammatory responses. The current authors hypothesised that inspiratory flow contributes to ventilator-induced inflammation. Buffer-perfused rabbit lungs were ventilated for 2 h with ...

  3. Circulatory and respiratory effects of methoxyflurane in dogs: comparison of halothane.

    PubMed

    Steffey, E P; Farver, T B; Woliner, M J

    1984-12-01

    Circulatory and respiratory effects of 3 alveolar concentrations (representing 1.0, 1.5, and 2.0 times the minimal alveolar concentration, MAC) of methoxyflurane in O2 were compared with similar MAC multiples of halothane in O2. Eight adult mixed breed dogs that were healthy and nonmedicated were studied in cross-over fashion with both agents during conditions of controlled ventilation (CV; PaCO2 averaged 34 to 38 mm of Hg) and spontaneous ventilation (SV). When ventilation was controlled, methoxyflurane similar to halothane caused dose-related cardiovascular depression. Except for a greater heart rate and lesser stroke volume with methoxyflurane, little difference was noticed between the anesthetics at equivalent doses during CV. There was less dose-related circulatory depression during SV with both agents but particularly with methoxyflurane. During SV, PaCO2 increased progressively with increases in alveolar concentrations of methoxyflurane and halothane. Methoxyflurane caused significantly greater (P less than 0.05) hypoventilation than halothane only at 2.0 MAC. Except for a greater respiratory gas flow and inspiratory-expiratory gas flow ratio and a lesser inspiratory-expiratory time ratio with methoxyflurane, there was no anesthetic- or dose-response effect on respiratory variables.

  4. Development and application of a double-piston configured, total-liquid ventilatory support device.

    PubMed

    Meinhardt, J P; Quintel, M; Hirschl, R B

    2000-05-01

    Perfluorocarbon liquid ventilation has been shown to enhance pulmonary mechanics and gas exchange in the setting of respiratory failure. To optimize the total liquid ventilation process, we developed a volume-limited, time-cycled liquid ventilatory support, consisting of an electrically actuated, microprocessor-controlled, double-cylinder, piston pump with two separate limbs for active inspiration and expiration. Prospective, controlled, animal laboratory study, involving sequential application of conventional gas ventilation, partial ventilation (PLV), and total liquid ventilation (TLV). Research facility at a university medical center. A total of 12 normal adult New Zealand rabbits weighing 3.25+/-0.1 kg. Anesthestized rabbits were supported with gas ventilation for 30 mins (respiratory rate, 20 cycles/min; peak inspiratory pressure, 15 cm H2O; end-expiratory pressure, 5 cm H2O), then PLV was established with perflubron (12 mL/kg). After 15 mins, TLV was instituted (tidal volume, 18 mL/kg; respiratory rate, 7 cycles/min; inspiratory/expiratory ratio, 1:2 cycles/min). After 4 hrs of TLV, PLV was re-established. Of 12 animals, nine survived the 4-hr TLV period. During TLV, mean values +/- SEM were as follows: PaO2, 363+/-30 torr; PaCO2, 39+/-1.5 torr; pH, 7.39+/-0.01; static peak inspiratory pressure, 13.2+/-0.2 cm H2O; static endexpiratory pressure, 5.5+/-0.1 cm H2O. No significant changes were observed. When compared with gas ventilation and PLV, significant increases occurred in mean arterial pressure (62.4+/-3.5 torr vs. 74.0+/-1.2 torr) and central venous pressure (5.6+/-0.7 cm H2O vs. 7.8+/-0.2 cm H2O) (p < .05). Total liquid ventilation can be performed successfully utilizing piston pumps with active expiration. Considering the enhanced flow profiles, this device configuration provides advantages over others.

  5. Metabolically Derived Human Ventilation Rates: A Revised ...

    EPA Pesticide Factsheets

    EPA announced the availability of the final report, Metabolically Derived Human Ventilation Rates: A Revised Approach Based Upon Oxygen Consumption Rates. This report provides a revised approach for calculating an individual's ventilation rate directly from their oxygen consumption rate. This approach will be used to update the ventilation rate information in the Exposure Factors Handbook, which serve as a resources for exposure assessors for calculating inhalation and other exposures. In this report, EPA presents a revised approach in which ventilation rate is calculated directly from an individual's oxygen consumption rate.

  6. Percutaneous dilational tracheostomy (PDT) and prevention of blood aspiration with superimposed high-frequency jet ventilation (SHFJV) using the tracheotomy-endoscope (TED): results of numerical and experimental simulations.

    PubMed

    Nowak, Andreas; Langebach, Robin; Klemm, Eckart; Heller, Winfried

    2012-04-01

    We describe an innovative computer-based method for the analysis of gas flow using a modified airway management technique to perform percutaneous dilatational tracheotomy (PDT) with a rigid tracheotomy endoscope (TED). A test lung was connected via an artificial trachea with the tracheotomy endoscope and ventilated using superimposed high-frequency jet ventilation. Red packed cells were instilled during the puncture phase of a simulated percutaneous tracheotomy in a trachea model and migration of the red packed cells during breathing was continuously measured. Simultaneously, the calculation of the gas-flow within the endoscope was numerically simulated. In the experimental study, no backflow of blood occurred during the use of superimposed high-frequency jet ventilation (SHFJV) from the trachea into the endoscope nor did any transportation of blood into the lower respiratory tract occur. In parallel, the numerical simulations of the openings of TED show almost positive volume flows. Under the conditions investigated there is no risk of blood aspiration during PDT using the TED and simultaneous ventilation with SHFJV. In addition, no risk of impairment of endoscopic visibility exists through a backflow of blood into the TED. The method of numerical simulation offers excellent insight into the fluid flow even under highly transient conditions like jet ventilation.

  7. Modeling ventilation time in forage tower silos.

    PubMed

    Bahloul, A; Chavez, M; Reggio, M; Roberge, B; Goyer, N

    2012-10-01

    The fermentation process in forage tower silos produces a significant amount of gases, which can easily reach dangerous concentrations and constitute a hazard for silo operators. To maintain a non-toxic environment, silo ventilation is applied. Literature reviews show that the fermentation gases reach high concentrations in the headspace of a silo and flow down the silo from the chute door to the feed room. In this article, a detailed parametric analysis of forced ventilation scenarios built via numerical simulation was performed. The methodology is based on the solution of the Navier-Stokes equations, coupled with transport equations for the gas concentrations. Validation was achieved by comparing the numerical results with experimental data obtained from a scale model silo using the tracer gas testing method for O2 and CO2 concentrations. Good agreement was found between the experimental and numerical results. The set of numerical simulations made it possible to establish a simple analytical model to predict the minimum time required to ventilate a silo to make it safe to enter. This ventilation time takes into account the headspace above the forage, the airflow rate, and the initial concentrations of O2 and CO2. The final analytical model was validated with available results from the literature.

  8. Multifamily Individual Heating and Ventilation Systems, Lawrence, Massachusetts (Fact Sheet)

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Not Available

    The conversion of an older Massachusetts building into condominiums illustrates a safe, durable, and cost-effective solution for heating and ventilation systems that can potentially benefit millions of multifamily buildings. Merrimack Valley Habitat for Humanity (MVHfH) partnered with U.S. Department of Energy Building America team Building Science Corporation (BSC) to provide high performance affordable housing for 10 families in the retrofit of an existing mass masonry building (a former convent). The original ventilation design for the project was provided by a local engineer and consisted of a single large heat recovery ventilator (HRV) located in a mechanical room in the basementmore » with a centralized duct system providing supply air to the main living space and exhausting stale air from the single bathroom in each apartment. This design was deemed to be far too costly to install and operate for several reasons: the large central HRV was oversized and the specified flows to each apartment were much higher than the ASHRAE 62.2 rate; an extensive system of ductwork, smoke and fire dampers, and duct chases were specified; ductwork required a significant area of dropped ceilings; and the system lacked individual ventilation control in the apartments« less

  9. [The numerical simulation of the internal flow field inside the pressure generator of a continuous positive airway pressure ventilator].

    PubMed

    Cheng, Yunzhang; Zhu, Lihua; Zhang, Weiguo; Wu, Wenquan

    2011-12-01

    The problem of noise in ventilator has always been an important topic to study in the development of the ventilator. A great number of data are showing that there are still large gaps of research and application levels in noise control of the ventilator between China and some more advanced foreign countries. In this study, with cooperation of the Shanghai Medical Equipment Limited Liability Company, we used the computational fluid dynamics (CFD), software FLUENT, adopted the standard k-epsilon turbulence model and the SIMPLE algorithm to simulate the inner flow field of the continuous positive airway pressure (CPAP) ventilator's pressure generator. After a detailed analysis, we figured out that there are several deficiencies in this ventilator, like local reflow in volute, uneven velocity distribution and local negative pressure in inlet of the impeller, which easily lead to noise and affect the ventilator's performances. So, it needs to be improved to a certain extent.

  10. Effects of tidal volume on work of breathing during lung-protective ventilation in patients with acute lung injury and acute respiratory distress syndrome.

    PubMed

    Kallet, Richard H; Campbell, Andre R; Dicker, Rochelle A; Katz, Jeffrey A; Mackersie, Robert C

    2006-01-01

    To assess the effects of step-changes in tidal volume on work of breathing during lung-protective ventilation in patients with acute lung injury (ALI) or the acute respiratory distress syndrome (ARDS). Prospective, nonconsecutive patients with ALI/ARDS. Adult surgical, trauma, and medical intensive care units at a major inner-city, university-affiliated hospital. Ten patients with ALI/ARDS managed clinically with lung-protective ventilation. Five patients were ventilated at a progressively smaller tidal volume in 1 mL/kg steps between 8 and 5 mL/kg; five other patients were ventilated at a progressively larger tidal volume from 5 to 8 mL/kg. The volume mode was used with a flow rate of 75 L/min. Minute ventilation was maintained constant at each tidal volume setting. Afterward, patients were placed on continuous positive airway pressure for 1-2 mins to measure their spontaneous tidal volume. Work of breathing and other variables were measured with a pulmonary mechanics monitor (Bicore CP-100). Work of breathing progressively increased (0.86 +/- 0.32, 1.05 +/- 0.40, 1.22 +/- 0.36, and 1.57 +/- 0.43 J/L) at a tidal volume of 8, 7, 6, and 5 mL/kg, respectively. In nine of ten patients there was a strong negative correlation between work of breathing and the ventilator-to-patient tidal volume difference (R = -.75 to -.998). : The ventilator-delivered tidal volume exerts an independent influence on work of breathing during lung-protective ventilation in patients with ALI/ARDS. Patient work of breathing is inversely related to the difference between the ventilator-delivered tidal volume and patient-generated tidal volume during a brief trial of unassisted breathing.

  11. Numerical simulation of gas distribution in goaf under Y ventilation mode

    NASA Astrophysics Data System (ADS)

    Li, Shengzhou; Liu, Jun

    2018-04-01

    Taking the Y type ventilation of the working face as the research object, diffusion equation is introduced to simulate the diffusion characteristics of gas, using Navier-Stokes equation and Brinkman equation to simulate the gas flow in working face and goaf, the physical model of gas flow in coal mining face was established. With numerical simulation software COMSOL multiphysics methods, gas distribution in goaf under Y ventilation mode is simulated and gas distribution of the working face, the upper corner and goaf is analysised. The results show that the Y type ventilation system can effectively improve the corner gas accumulation and overrun problem.

  12. Regional gas transport in the heterogeneous lung during oscillatory ventilation.

    PubMed

    Herrmann, Jacob; Tawhai, Merryn H; Kaczka, David W

    2016-12-01

    Regional ventilation in the injured lung is heterogeneous and frequency dependent, making it difficult to predict how an oscillatory flow waveform at a specified frequency will be distributed throughout the periphery. To predict the impact of mechanical heterogeneity on regional ventilation distribution and gas transport, we developed a computational model of distributed gas flow and CO 2 elimination during oscillatory ventilation from 0.1 to 30 Hz. The model consists of a three-dimensional airway network of a canine lung, with heterogeneous parenchymal tissues to mimic effects of gravity and injury. Model CO 2 elimination during single frequency oscillation was validated against previously published experimental data (Venegas JG, Hales CA, Strieder DJ, J Appl Physiol 60: 1025-1030, 1986). Simulations of gas transport demonstrated a critical transition in flow distribution at the resonant frequency, where the reactive components of mechanical impedance due to airway inertia and parenchymal elastance were equal. For frequencies above resonance, the distribution of ventilation became spatially clustered and frequency dependent. These results highlight the importance of oscillatory frequency in managing the regional distribution of ventilation and gas exchange in the heterogeneous lung. Copyright © 2016 the American Physiological Society.

  13. Buoyancy and Pressure Driven Flow of Hot Gases in Vertical Shafts with Natural and Forced Ventilation

    NASA Astrophysics Data System (ADS)

    Tamm, Gunnar; Jaluria, Yogesh

    2003-11-01

    An experimental investigation has been carried out on the buoyancy and pressure induced flow of hot gases in vertical shafts, in order to simulate the propagation of combustion products in elevator shafts due to fire in multilevel buildings. Various geometrical configurations are studied, with regard to natural and forced ventilation imposed at the top or bottom of the vertical shaft. The aspect ratio is taken at a fixed value of 6 and the inflow conditions for the hot gases, at a vent near the bottom, are varied in terms of the Reynolds and Grashof numbers. Temperature measurements within the shaft allow a detailed study of the steady state thermal fields, from which optimal means for smoke alleviation in high-rise building fires may be developed. Flow visualization is also used to study the flow characteristics. The results obtained indicate a wall plume as the primary transport mechanism. Flow recirculation dominates at high Grashof number flows, while increased Reynolds numbers gives rise to greater mixing in the shaft. The development and stability of the flow and its effect on the spread of smoke and hot gases are assessed for the different shaft configurations and inlet conditions. It is found that the fastest smoke removal and lowest shaft temperatures occur for a configuration with natural ventilation at the top and forced ventilation up from the shaft bottom. It is also shown that forced ventilation can be used to arrest smoke spread, as well as to dilute the effects of the fire.

  14. Gas exchange and intrapulmonary distribution of ventilation during continuous-flow ventilation

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Vettermann, J.; Brusasco, V.; Rehder, K.

    1988-05-01

    In 12 anesthetized paralyzed dogs, pulmonary gas exchange and intrapulmonary inspired gas distribution were compared between continuous-flow ventilation (CFV) and conventional mechanical ventilation (CMV). Nine dogs were studied while they were lying supine, and three dogs were studied while they were lying prone. A single-lumen catheter for tracheal insufflation and a double-lumen catheter for bilateral endobronchial insufflation (inspired O2 fraction = 0.4; inspired minute ventilation = 1.7 +/- 0.3 (SD) 1.kg-1.min-1) were evaluated. Intrapulmonary gas distribution was assessed from regional 133Xe clearances. In dogs lying supine, CO2 elimination was more efficient with endobronchial insufflation than with tracheal insufflation, but themore » alveolar-arterial O2 partial pressure difference was larger during CFV than during CMV, regardless of the type of insufflation. By contrast, endobronchial insufflation maintained both arterial PCO2 and alveolar-arterial O2 partial pressure difference at significantly lower levels in dogs lying prone than in dogs lying supine. In dogs lying supine, the dependent lung was preferentially ventilated during CMV but not during CFV. In dogs lying prone, gas distribution was uniform with both modes of ventilation. The alveolar-arterial O2 partial pressure difference during CFV in dogs lying supine was negatively correlated with the reduced ventilation of the dependent lung, which suggests that increased ventilation-perfusion mismatching was responsible for the increase in alveolar-arterial O2 partial pressure difference. The more efficient oxygenation during CFV in dogs lying prone suggests a more efficient matching of ventilation to perfusion, presumably because the distribution of blood flow is also nearly uniform.« less

  15. A polyurethane cuffed endotracheal tube is associated with decreased rates of ventilator-associated pneumonia.

    PubMed

    Miller, Melissa A; Arndt, Jennifer L; Konkle, Mark A; Chenoweth, Carol E; Iwashyna, Theodore J; Flaherty, Kevin R; Hyzy, Robert C

    2011-06-01

    The aim of this study was to determine whether the use of a polyurethane-cuffed endotracheal tube would result in a decrease in ventilator-associated pneumonia rate. We replaced conventional endotracheal tube with a polyurethane-cuff endotracheal tube (Microcuff, Kimberly-Clark Corporation, Rosewell, Ga) in all adult mechanically ventilated patients throughout our large academic hospital from July 2007 to June 2008. We retrospectively compared the rates of ventilator-associated pneumonia before, during, and after the intervention year by interrupted time-series analysis. Ventilator-associated pneumonia rates decreased from 5.3 per 1000 ventilator days before the use of the polyurethane-cuffed endotracheal tube to 2.8 per 1000 ventilator days during the intervention year (P = .0138). During the first 3 months after return to conventional tubes, the rate of ventilator-associated pneumonia was 3.5/1000 ventilator days. Use of the polyurethane-cuffed endotracheal tube was associated with an incidence risk ratio of ventilator-associated pneumonia of 0.572 (95% confidence interval, 0.340-0.963). In statistical regression analysis controlling for other possible alterations in the hospital environment, as measured by rate of tracheostomy-ventilator-associated pneumonia, the incidence risk ratio of ventilator-associated pneumonia in patients intubated with polyurethane-cuffed endotracheal tube was 0.565 (P = .032; 95% confidence interval, 0.335-0.953). Use of a polyurethane-cuffed endotracheal tube was associated with a significant decrease in the rate of ventilator-associated pneumonia in our study. Copyright © 2011 Elsevier Inc. All rights reserved.

  16. Winter ventilation rates at primary schools: comparison between Portugal and Finland.

    PubMed

    Canha, N; Almeida, S M; Freitas, M C; Täubel, M; Hänninen, O

    2013-01-01

    This study focused on examination of ventilation rates in classrooms with two different types of ventilation systems: natural and mechanical. Carbon dioxide (CO2) measurements were conducted in primary schools of Portugal characterized by natural ventilation and compared to Finland where mechanical ventilation is the norm. The winter period was selected since this season exerts a great influence in naturally ventilated classrooms, where opening of windows and doors occurs due to outdoor atmospheric conditions. The ventilation rates were calculated by monitoring CO2 concentrations generated by the occupants (used as a tracer gas) and application of the buildup phase method. A comparison between both countries' results was conducted with respect to ventilation rates and how these levels corresponded to national regulatory standards. Finnish primary schools (n = 2) registered a mean ventilation rate of 13.3 L/s per person, which is higher than the recommended ventilation standards. However, the Finnish classroom that presented the lowest ventilation rate (7.2 L/s per person) displayed short-term CO2 levels above 1200 ppm, which is the threshold limit value (TLV) recommended by national guidelines. The Portuguese classrooms (n = 2) showed low ventilation rates with mean values of 2.4 L/s per person, which is markedly lower than the minimum recommended value of 7 L/s per person as defined by ASHRAE and 20% less than the REHVA minimum of 3 L/s per person. Carbon dioxide levels of 1000 ppm, close to the TLV of 1200 ppm, were also reached in both Portuguese classrooms studied. The situation in Portugal indicates a potentially serious indoor air quality problem and strengthens the need for intervention to improve ventilation rates in naturally ventilated classrooms.

  17. Unsteady two dimensional airloads acting on oscillating thin airfoils in subsonic ventilated wind tunnels

    NASA Technical Reports Server (NTRS)

    Fromme, J.; Golberg, M.

    1978-01-01

    The numerical calculation of unsteady two dimensional airloads which act upon thin airfoils in subsonic ventilated wind tunnels was studied. Neglecting certain quadrature errors, Bland's collocation method is rigorously proved to converge to the mathematically exact solution of Bland's integral equation, and a three way equivalence was established between collocation, Galerkin's method and least squares whenever the collocation points are chosen to be the nodes of the quadrature rule used for Galerkin's method. A computer program displayed convergence with respect to the number of pressure basis functions employed, and agreement with known special cases was demonstrated. Results are obtained for the combined effects of wind tunnel wall ventilation and wind tunnel depth to airfoil chord ratio, and for acoustic resonance between the airfoil and wind tunnel walls. A boundary condition is proposed for permeable walls through which mass flow rate is proportional to pressure jump.

  18. Mechanical breath profile of airway pressure release ventilation: the effect on alveolar recruitment and microstrain in acute lung injury.

    PubMed

    Kollisch-Singule, Michaela; Emr, Bryanna; Smith, Bradford; Roy, Shreyas; Jain, Sumeet; Satalin, Joshua; Snyder, Kathy; Andrews, Penny; Habashi, Nader; Bates, Jason; Marx, William; Nieman, Gary; Gatto, Louis A

    2014-11-01

    Improper mechanical ventilation settings can exacerbate acute lung injury by causing a secondary ventilator-induced lung injury. It is therefore important to establish the mechanism by which the ventilator induces lung injury to develop protective ventilation strategies. It has been postulated that the mechanism of ventilator-induced lung injury is the result of heterogeneous, elevated strain on the pulmonary parenchyma. Acute lung injury has been associated with increases in whole-lung macrostrain, which is correlated with increased pathology. However, the effect of mechanical ventilation on alveolar microstrain remains unknown. To examine whether the mechanical breath profile of airway pressure release ventilation (APRV), consisting of a prolonged pressure-time profile and brief expiratory release phase, reduces microstrain. In a randomized, nonblinded laboratory animal study, rats were randomized into a controlled mandatory ventilation group (n = 3) and an APRV group (n = 3). Lung injury was induced by polysorbate lavage. A thoracotomy was performed and an in vivo microscope was placed on the lungs to measure alveolar mechanics. In the controlled mandatory ventilation group, multiple levels of positive end-expiratory pressure (PEEP; 5, 10, 16, 20, and 24 cm H2O) were tested. In the APRV group, decreasing durations of expiratory release (time at low pressure [T(low)]) were tested. The T(low) was set to achieve ratios of termination of peak expiratory flow rate (T-PEFR) to peak expiratory flow rate (PEFR) of 10%, 25%, 50%, and 75% (the smaller this ratio is [ie, 10%], the more time the lung is exposed to low pressure during the release phase, which decreases end-expiratory lung volume and potentiates derecruitment). Alveolar perimeters were measured at peak inspiration and end expiration using digital image analysis, and strain was calculated by normalizing the change in alveolar perimeter length to the original length. Macrostrain was measured by volume displacement. Higher PEEP (16-24 cm H2O) and a brief T(low) (APRV T-PEFR to PEFR ratio of 75%) reduced microstrain. Microstrain was minimized with an APRV T-PEFR to PEFR ratio of 75% (mean [SEM], 0.05 [0.03]) and PEEP of 16 cm H2O (mean [SEM], 0.09 [0.08]), but an APRV T-PEFR to PEFR ratio of 75% also promoted alveolar recruitment compared with PEEP of 16 cm H2O (mean [SEM] total inspiratory area, 52.0% [2.9%] vs 29.4% [4.3%], respectively; P < .05). Whole-lung strain was correlated with alveolar microstrain in tested settings (P < .05) except PEEP of 16 cm H2O (P > .05). Increased positive-end expiratory pressure and reduced time at low pressure (decreased T(low)) reduced alveolar microstrain. Reduced microstrain and improved alveolar recruitment using an APRV T-PEFR to PEFR ratio of 75% may be the mechanism of lung protection seen in previous clinical and animal studies.

  19. IMPACT OF VENTILATION FREQUENCY AND PARENCHYMAL STIFFNESS ON FLOW AND PRESSURE DISTRIBUTION IN A CANINE LUNG MODEL

    PubMed Central

    Amini, Reza; Kaczka, David W.

    2013-01-01

    To determine the impact of ventilation frequency, lung volume, and parenchymal stiffness on ventilation distribution, we developed an anatomically-based computational model of the canine lung. Each lobe of the model consists of an asymmetric branching airway network subtended by terminal, viscoelastic acinar units. The model allows for empiric dependencies of airway segment dimensions and parenchymal stiffness on transpulmonary pressure. We simulated the effects of lung volume and parenchymal recoil on global lung impedance and ventilation distribution from 0.1 to 100 Hz, with mean transpulmonary pressures from 5 to 25 cmH2O. With increasing lung volume, the distribution of acinar flows narrowed and became more synchronous for frequencies below resonance. At higher frequencies, large variations in acinar flow were observed. Maximum acinar flow occurred at first antiresonance frequency, where lung impedance achieved a local maximum. The distribution of acinar pressures became very heterogeneous and amplified relative to tracheal pressure at the resonant frequency. These data demonstrate the important interaction between frequency and lung tissue stiffness on the distribution of acinar flows and pressures. These simulations provide useful information for the optimization of frequency, lung volume, and mean airway pressure during conventional ventilation or high frequency oscillation (HFOV). Moreover our model indicates that an optimal HFOV bandwidth exists between the resonant and antiresonant frequencies, for which interregional gas mixing is maximized. PMID:23872936

  20. Tracheal gas insufflation combined with high-frequency oscillatory ventilation.

    PubMed

    Dolan, S; Derdak, S; Solomon, D; Farmer, C; Johanningman, J; Gelineau, J; Smith, R B

    1996-03-01

    To determine the efficacy of tracheal insufflation delivered by two different catheter designs on CO2 elimination when used in conjunction with high-frequency oscillatory ventilation. A nonrandomized before and after trial. Each animal served as his own control. Ten mongrel dogs weighing 20.9 +/- 1.9 kg. Four animals were assigned to a normal lung group and six animals underwent lung injury by large volume saline lavage. Permissive hypercapnia was allowed to occur by selecting oscillator settings that would lead to alveolar hypoventilation. Proximal mean airway pressure was kept constant. Tracheal gas was insufflated at 1 cm above the carina for 30 min periods at gas flows of 5 to 15 L/min. Carinal pressure, hemodynamic parameters (cardiac output, mean arterial pressure, pulmonary artery occlusion pressure), and gas exchange parameters (PaCO2, PaO2, PaO2/FIO2, shunt fraction, D O2) were measured. For the normal dogs, at catheter flow of 15 L/min; the forward thrust catheter increased carinal pressure and Pao2/FIO2 BY 30% (p<.003) and 105% (p<.005), respectively. The forward thrust catheter reduced Paco2 by 40% (p<.04). The reverse thrust catheter increased PaO2/FIO2 by 102% (p<.001) and decreased pressure and PaCO2 by 44% (p<.001) and 34% (p<.003), respectively. For the injured dogs, at catheter flow rate of 15 L/min, the forward thrust catheter increased carinal pressure, PaO2, and PaO2/FIO2 by 6% (p<.001), 23% (p<.001), and 24% (p<.02), respectively. The forward thrust catheter reduced PaCO2 by 29% (p<.002). The reverse thrust catheter increased PaO2 and PaO2/FIO2 both by 11% (p<.02) and reduced carinal pressure and PaCO2 by 23% (p<.001) and 18% (p<.002), respectively. Tracheal gas insufflation is capable of improving oxygenation and ventilation in acute lung injury when combined with high-frequency oscillatory ventilation. The addition of this second gas flow at the level of the carina raises or lowers distal airway pressure, the magnitude of which is dependent on the direction and rate of gas flow. The beneficial effects of tracheal gas insufflation may be tempered by the long-term effects of altering distal airway pressure; lowering distal airway pressure may lead to atelectasis, whereas raising distal airway pressure may lead to an auto-positive end-expiratory pressure (auto-PEEP) effect.

  1. The Potential of Bee-Generated Carbon Dioxide for Control of Varroa Mite (Mesostigmata: Varroidae) in Indoor Overwintering Honey bee (Hymenoptera: Apidae) Colonies.

    PubMed

    Bahreini, Rassol; Currie, Robert W

    2015-10-01

    The objective of this study was to manipulate ventilation rate to characterize interactions between stocks of honey bees (Apis mellifera L.) and ventilation setting on varroa mite (Varroa destructor Anderson and Trueman) mortality in honey bee colonies kept indoors over winter. The first experiment used colonies established from stock selected locally for wintering performance under exposure to varroa (n = 6) and unselected bees (n = 6) to assess mite and bee mortality and levels of carbon dioxide (CO2) and oxygen (O2) in the bee cluster when kept under a simulated winter condition at 5°C. The second experiment, used colonies from selected bees (n = 10) and unselected bees (n = 12) that were exposed to either standard ventilation (14.4 liter/min per hive) or restricted ventilation (0.24 liter/min per hive, in a Plexiglas ventilation chamber) during a 16-d treatment period to assess the influence of restricted air flow on winter mortality rates of varroa mites and honey bees. Experiment 2 was repeated in early, mid-, and late winter. The first experiment showed that under unrestricted ventilation with CO2 concentrations averaging <2% there was no correlation between CO2 and varroa mite mortality when colonies were placed under low temperature. CO2 was negatively correlated with O2 in the bee cluster in both experiments. When ventilation was restricted, mean CO2 level (3.82 ± 0.31%, range 0.43-8.44%) increased by 200% relative to standard ventilation (1.29 ± 0.31%; range 0.09-5.26%) within the 16-d treatment period. The overall mite mortality rates and the reduction in mean abundance of varroa mite over time was greater under restricted ventilation (37 ± 4.2%) than under standard ventilation (23 ± 4.2%) but not affected by stock of bees during the treatment period. Selected bees showed overall greater mite mortality relative to unselected bees in both experiments. Restricting ventilation increased mite mortality, but did not affect worker bee mortality relative to that for colonies under standard ventilation. Restricted ventilation did not affect the overall level of Nosema compared with the control. However, there was an interaction between stock, season, and time of the trial. Unselected stock showed an increase in Nosema over time in the late winter trial that did not occur in the selected stock. In conclusion, these findings suggested that restricted ventilation has potential to suppress varroa mite in overwintering honey bee colonies via a low-cost and environmentally friendly measure. © The Authors 2015. Published by Oxford University Press on behalf of Entomological Society of America. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  2. Flow dynamics in pediatric rigid bronchoscopes using computer-aided design modeling software.

    PubMed

    Barneck, Mitchell D; Webb, J Taylor; Robinson, Ryan E; Grimmer, J Fredrik

    2016-08-01

    Observed complications during rigid bronchoscopy, including hypercarbia and hypoxemia, prompted us to assess how well rigid bronchoscopes serve as an airway device. We performed computer-aided design flow analysis of pediatric rigid bronchoscopes to gain insight into flow dynamics. We made accurate three-dimensional computer models of pediatric rigid bronchoscopes and endotracheal tubes. SOLIDWORKS (Dassault Systemes, Vélizy-Villacoublay, France) flow analysis software was used to analyze fluid dynamics during pressure-controlled and volume-controlled ventilation. Flow analysis was performed on rigid bronchoscopes and similar outer diameter endotracheal tubes comparing resistance, flow, and turbulence during two ventilation modalities and in common surgical scenarios. Increased turbulent flow was observed in bronchoscopes compared to more laminar flow in endotracheal tubes of similar outer diameter. Flow analysis displayed higher resistances in all pediatric bronchoscope sizes except one (3.0 bronchoscope) compared to similar-sized endotracheal tubes. Loss of adequate ventilation was observed if the bronchoscope was not assembled correctly or if increased peak inspiratory pressures were needed. Anesthesia flow to the patient was reduced by 63% during telescope insertion. Flow analysis illustrates increased turbulent flow and increased airflow resistance in all but one size of pediatric bronchoscopes compared to endotracheal tubes. This increased turbulence and resistance, along with the unanticipated gas distal exit pattern, may contribute to the documented hypercarbia and hypoxemia during procedures. These findings may explain why hypoxemia and hypercarbia are commonly observed during rigid bronchoscopy, especially when positive pressure ventilation is needed. NA Laryngoscope, 126:1940-1945, 2016. © 2015 The American Laryngological, Rhinological and Otological Society, Inc.

  3. Evaluation of resistance in 8 different heat-and-moisture exchangers: effects of saturation and flow rate/profile.

    PubMed

    Lucato, Jeanette Janaina Jaber; Tucci, Mauro Roberto; Schettino, Guilherme Paula Pinto; Adams, Alexander B; Fu, Carolina; Forti, Germano; de Carvalho, Carlos Roberto Ribeiro; de Souza, Rogério

    2005-05-01

    When endotracheal intubation is required during ventilatory support, the physiologic mechanisms of heating and humidifying the inspired air related to the upper airways are bypassed. The task of conditioning the air can be partially accomplished by heat-and-moisture exchangers (HMEs). To evaluate and compare with respect to imposed resistance, different types/models of HME: (1) dry versus saturated, (2) changing inspiratory flow rates. Eight different HMEs were studied using a lung model system. The study was conducted initially by simulating spontaneous breathing, followed by connecting the system directly to a mechanical ventilator to provide pressure-support ventilation. None of the encountered values of resistance (0.5\\N3.6 cm H(2)O/L/s) exceeded the limits stipulated by the previously described international standard for HMEs (International Standards Organization Draft International Standard 9360-2) (not to exceed 5.0 cm H(2)O with a flow of 1.0 L/s, even when saturated). The hygroscopic HME had less resistance than other types, independent of the precondition status (dry or saturated) or the respiratory mode. The hygroscopic HME also had a lesser increase in resistance when saturated. The resistance of the HME was little affected by increases in flow, but saturation did increase resistance in the hydrophobic and hygroscopic/hydrophobic HME to levels that could be important at some clinical conditions. Resistance was little affected by saturation in hygroscopic models, when compared to the hydrophobic or hygroscopic/hydrophobic HME. Changes in inspiratory flow did not cause relevant alterations in resistance.

  4. Personal and demographic factors and change of subjective indoor air quality reported by school children in relation to exposure at Swedish schools: a 2-year longitudinal study.

    PubMed

    Wang, Juan; Smedje, Greta; Nordquist, Tobias; Norbäck, Dan

    2015-03-01

    This paper studies changes in subjective indoor air quality (SIAQ) among school children and relates these data to repeated exposure measurements during a two-year follow-up period. Data on SIAQ and demographic information were gathered by a questionnaire sent to 1476 primary and secondary school pupils in 39 randomly selected schools at baseline and after two years (follow-up). Exposure measurements were applied after questionnaire data were collected at baseline and follow-up in approximately 100 classrooms. The arithmetic mean values for baseline and follow-up were: for indoor air temperature 23.6°C and 21.8°C and for outdoor air flow rate 5.4 L/s and 7.9L/s. Older children, those with atopy at baseline, and those in larger schools reported impaired SIAQ during follow-up. Installation of new ventilation systems, higher personal air flow rate and air exchange rate, and better illumination were associated with improved SIAQ. Higher CO2 levels were associated with impaired SIAQ. In conclusion, sufficient ventilation and illumination in classrooms are essential for the perception of good indoor air quality. Copyright © 2014 Elsevier B.V. All rights reserved.

  5. Bacterial dispersion in relation to operating room clothing.

    PubMed Central

    Whyte, W.; Vesley, D.; Hodgson, R.

    1976-01-01

    The effect of operating clothing on the dispersal of bacterial particles from the wearers was studied in a dispersal chamber. A comparison was made of six gowns as well as four types of trousers. The gowns were of three basic types, namely a conventional cotton type, disposable types made of non-woven fabric and those of the total-body exhaust system (Charnley type). The dispersal chamber could simulate conditions as expected both in down-flow unidirectional ultra-clean systems and in a conventional turbulent plenum-ventilated system. It was found that the disposable gowns would reduce the dispersal rate by about 30% in the simulated conventionally ventilated system and about 65% in the laminar flow system. The total-body exhaust system (Charnley) would reduce the count by 10-fold in the conventional ventilated system and by 66-fold in the laminar-flow system. The poor performance of the gowns in conventionally ventilated systems was caused by the dispersal of bacterial particles from underneath the gown (about 80%). This was not reduced by the disposable gown and only partially by the Charnley type. This small drop would be further decreased in a conventionally ventilated operating-room as only scrubbed staff would wear the gown. In order to overcome this poor performance in conventionally ventilated operating-rooms impervious trousers would be required. Four types were studied and it was demonstrated that those made either from Ventile or non-woven fabric would reduce the bacterial dispersion fourfold. As these tests had been carried out in an artificial environment checks were carried out in the unidirectional-flow operating-room during total-hip arthroplasty. This was done by comparing conventional cotton gowns with non-woven gowns and total-body exhaust gowns. The results showed good correlation between the operating room and the chamber with the non-woven fabric gown but the total-body exhaust system did not perform as well in the operating room (12-fold compared to 66-fold) the difference being possibly due to the contribution from the patient. However, as this comparison was that which would be most open to influence from other variables confidence could be placed on the chamber test results. Values were also obtained for the total number of bacterial particles dispersed by persons during a standard exercise wearing different clothing. This count was dependent on the clothing worn but a median count of between 1000 and 1500 bacterial particles/min. would be expected when conventional clothing was worn, with a range of between 300 and 19,000. This count could be reduced to about 100/min. if a total-body exhaust suit was worn (range 30-400). PMID:778258

  6. Effect of using a laryngeal tube on the no-flow time in a simulated, single-rescuer, basic life support setting with inexperienced users.

    PubMed

    Meyer, O; Bucher, M; Schröder, J

    2016-03-01

    The laryngeal tube (LT) is a recommended alternative to endotracheal intubation during advanced life support (ALS). Its insertion is relatively simple; therefore, it may also serve as an alternative to bag mask ventilation (BMV) for untrained personnel performing basic life support (BLS). Data support the influence of LT on the no-flow time (NFT) compared with BMV during ALS in manikin studies. We performed a manikin study to investigate the effect of using the LT for ventilation instead of BMV on the NFT during BLS in a prospective, randomized, single-rescuer study. All 209 participants were trained in BMV, but were inexperienced in using LT; each participant performed BLS during a 4-min time period. No significant difference in total NFT (LT: mean 81.1 ± 22.7 s; BMV: mean 83.2 ± 13.1 s, p = 0.414) was found; however, significant differences in the later periods of the scenario were identified. While ventilating with the LT, the proportion of chest compressions increased significantly from 67.2 to 73.2%, whereas the proportion of chest compressions increased only marginally when performing BMV. The quality of the chest compressions and the associated ventilation rate did not differ significantly. The mean tidal volume and mean minute volume were significantly lower when performing BMV. The NFT was significantly shorter in the later periods in a single-rescuer, cardiac arrest scenario when using an LT without previous training compared with BMV with previous training. A possible explanation for this result may be the complexity and workload of alternating tasks (e.g., time loss when reclining the head and positioning the mask for each ventilation during BMV).

  7. Using Optical Tweezers to Study Cell Mechanics during Airway Reopening

    NASA Astrophysics Data System (ADS)

    Yalcin, Huseyin; Wang, Jing; Ghadiali, Samir; Ou-Yang, H. Daniel

    2006-03-01

    Patients suffering from the acute respiratory distress syndrome (ARDS) must be mechanically ventilated in order to survive. However, these ventilation protocols may generate injurious hydrodynamic stresses especially during low tidal volume (VT) ventilation when the flow of micron-sized air bubbles displace the surrounding liquid. In-vitro studies in our lab revealed that microbubble flows can severally damage lung epithelial cells (EC). The degree of injury was elevated for sub-confluent monolayers in small channel heights. Under these conditions, the micromechanics of individual EC may influence the degree of cellular injury. To investigate the role of cell mechanics, we used an oscillating Optical Tweezers (OT) technique to measure the intrinsic mechanical properties of EC before and after the flow of microbubbles. Knowledge of how the EC's micromechanical properties influence cell viability may lead to the development of novel treatment therapies that enhance the EC's ability to withstand injurious hydrodynamic stresses during ventilation treatment.

  8. The effect of low ventilation rate with elevated bioeffluent concentration on work performance, perceived indoor air quality, and health symptoms.

    PubMed

    Maula, H; Hongisto, V; Naatula, V; Haapakangas, A; Koskela, H

    2017-11-01

    The aim of this laboratory experiment was to study the effects of ventilation rate, and related changes in air quality, predominantly bioeffluents, on work performance, perceived indoor air quality, and health symptoms in a typical conditions of modern open-plan office with low material and equipment emissions. In Condition A, outdoor air flow rate of 28.2 l/s person (CO 2 level 540 ppm) was applied and in Condition B, outdoor air flow rate was 2.3 l/s person (CO 2 level 2260 ppm). CO 2 concentration level was used as an indicator of bioeffluents. Performance was measured with seven different tasks which measure different cognitive processes. Thirty-six subjects participated in the experiment. The exposure time was 4 hours. Condition B had a weak negative effect on performance only in the information retrieval tasks. Condition B increased slightly subjective workload and perceived fatigue. No effects on health symptoms were found. The intensity of symptoms was low in both conditions. The experimental condition had an effect on perceived air quality and observed odor intensity only in the beginning of the session. Although the room temperature was controlled in both conditions, the heat was perceived to impair the performance more in Condition B. © 2017 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

  9. Steady-state balance model to calculate the indoor climate of livestock buildings, demonstrated for finishing pigs

    NASA Astrophysics Data System (ADS)

    Schauberger, G.; Piringer, M.; Petz, E.

    The indoor climate of livestock buildings is of importance for the well-being and health of animals and their production performance (daily weight gain, milk yield etc). By using a steady-state model for the sensible and latent heat fluxes and the CO2 and odour mass flows, the indoor climate of mechanically ventilated livestock buildings can be calculated. These equations depend on the livestock (number of animals and how they are kept), the insulation of the building and the characteristics of the ventilation system (ventilation rate). Since the model can only be applied to animal houses where the ventilation systems are mechanically controlled (this is the case for a majority of finishing pig units), the calculations were done for an example of a finishing pig unit with 1000 animal places. The model presented used 30 min values of the outdoor parameters temperature and humidity, collected over a 2-year period, as input. The projected environment inside the livestock building was compared with recommended values. The duration of condensation on the inside surfaces was also calculated.

  10. Flow analysis of airborne particles in a hospital operating room

    NASA Astrophysics Data System (ADS)

    Faeghi, Shiva; Lennerts, Kunibert

    2016-06-01

    Preventing airborne infections during a surgery has been always an important issue to deliver effective and high quality medical care to the patient. One of the important sources of infection is particles that are distributed through airborne routes. Factors influencing infection rates caused by airborne particles, among others, are efficient ventilation and the arrangement of surgical facilities inside the operating room. The paper studies the ventilation airflow pattern in an operating room in a hospital located in Tehran, Iran, and seeks to find the efficient configurations with respect to the ventilation system and layout of facilities. This study uses computational fluid dynamics (CFD) and investigates the effects of different inflow velocities for inlets, two pressurization scenarios (equal and excess pressure) and two arrangements of surgical facilities in room while the door is completely open. The results show that system does not perform adequately when the door is open in the operating room under the current conditions, and excess pressure adjustments should be employed to achieve efficient results. The findings of this research can be discussed in the context of design and controlling of the ventilation facilities of operating rooms.

  11. Effect of external PEEP in patients under controlled mechanical ventilation with an auto-PEEP of 5 cmH2O or higher.

    PubMed

    Natalini, Giuseppe; Tuzzo, Daniele; Rosano, Antonio; Testa, Marco; Grazioli, Michele; Pennestrì, Vincenzo; Amodeo, Guido; Berruto, Francesco; Fiorillo, Marialinda; Peratoner, Alberto; Tinnirello, Andrea; Filippini, Matteo; Marsilia, Paolo F; Minelli, Cosetta; Bernardini, Achille

    2016-12-01

    In some patients with auto-positive end-expiratory pressure (auto-PEEP), application of PEEP lower than auto-PEEP maintains a constant total PEEP, therefore reducing the inspiratory threshold load without detrimental cardiovascular or respiratory effects. We refer to these patients as "complete PEEP-absorbers." Conversely, adverse effects of PEEP application could occur in patients with auto-PEEP when the total PEEP rises as a consequence. From a pathophysiological perspective, all subjects with flow limitation are expected to be "complete PEEP-absorbers," whereas PEEP should increase total PEEP in all other patients. This study aimed to empirically assess the extent to which flow limitation alone explains a "complete PEEP-absorber" behavior (i.e., absence of further hyperinflation with PEEP), and to identify other factors associated with it. One hundred patients with auto-PEEP of at least 5 cmH2O at zero end-expiratory pressure (ZEEP) during controlled mechanical ventilation were enrolled. Total PEEP (i.e., end-expiratory plateau pressure) was measured both at ZEEP and after applied PEEP equal to 80 % of auto-PEEP measured at ZEEP. All measurements were repeated three times, and the average value was used for analysis. Forty-seven percent of the patients suffered from chronic pulmonary disease and 52 % from acute pulmonary disease; 61 % showed flow limitation at ZEEP, assessed by manual compression of the abdomen. The mean total PEEP was 7 ± 2 cmH2O at ZEEP and 9 ± 2 cmH2O after the application of PEEP (p < 0.001). Thirty-three percent of the patients were "complete PEEP-absorbers." Multiple logistic regression was used to predict the behavior of "complete PEEP-absorber." The best model included a respiratory rate lower than 20 breaths/min and the presence of flow limitation. The predictive ability of the model was excellent, with an overoptimism-corrected area under the receiver operating characteristics curve of 0.89 (95 % CI 0.80-0.97). Expiratory flow limitation was associated with both high and complete "PEEP-absorber" behavior, but setting a relatively high respiratory rate on the ventilator can prevent from observing complete "PEEP-absorption." Therefore, the effect of PEEP application in patients with auto-PEEP can be accurately predicted at the bedside by measuring the respiratory rate and observing the flow-volume loop during manual compression of the abdomen.

  12. The respiratory system during resuscitation: a review of the history, risk of infection during assisted ventilation, respiratory mechanics, and ventilation strategies for patients with an unprotected airway.

    PubMed

    Wenzel, V; Idris, A H; Dörges, V; Nolan, J P; Parr, M J; Gabrielli, A; Stallinger, A; Lindner, K H; Baskett, P J

    2001-05-01

    The fear of acquiring infectious diseases has resulted in reluctance among healthcare professionals and the lay public to perform mouth-to-mouth ventilation. However, the benefit of basic life support for a patient in cardiopulmonary or respiratory arrest greatly outweighs the risk for secondary infection in the rescuer or the patient. The distribution of ventilation volume between lungs and stomach in the unprotected airway depends on patient variables such as lower oesophageal sphincter pressure, airway resistance and respiratory system compliance, and the technique applied while performing basic or advanced airway support, such as head position, inflation flow rate and time, which determine upper airway pressure. The combination of these variables determines gas distribution between the lungs and the oesophagus and subsequently, the stomach. During bag-valve-mask ventilation of patients in respiratory or cardiac arrest with oxygen supplementation (> or = 40% oxygen), a tidal volume of 6-7 ml kg(-1) ( approximately 500 ml) given over 1-2 s until the chest rises is recommended. For bag-valve-mask ventilation with room-air, a tidal volume of 10 ml kg(-1) (700-1000 ml) in an adult given over 2 s until the chest rises clearly is recommended. During mouth-to-mouth ventilation, a breath over 2 s sufficient to make the chest rise clearly (a tidal volume of approximately 10 ml kg(-1) approximately 700-1000 ml in an adult) is recommended.

  13. Fast or Slow Rescue Ventilations: A Predictive Model of Gastric Inflation.

    PubMed

    Fitz-Clarke, John R

    2018-05-01

    Rescue ventilations are given during respiratory and cardiac arrest. Tidal volume must assure oxygen delivery; however, excessive pressure applied to an unprotected airway can cause gastric inflation, regurgitation, and pulmonary aspiration. The optimal technique provides mouth pressure and breath duration that minimize gastric inflation. It remains unclear if breath delivery should be fast or slow, and how inflation time affects the division of gas flow between the lungs and esophagus. A physiological model was used to predict and compare rates of gastric inflation and to determine ideal ventilation duration. Gas flow equations were based on standard pulmonary physiology. Gastric inflation was assumed to occur whenever mouth pressure exceeded lower esophageal sphincter pressure. Mouth pressure profiles that approximated mouth-to-mouth ventilation and bag-valve-mask ventilation were investigated. Target tidal volumes were set to 0.6 and 1.0 L. Compliance and airway resistance were varied. Rapid breaths shorter than 1 s required high mouth pressures, up to 25 cm H 2 O to achieve the target lung volume, which thus promotes gastric inflation. Slow breaths longer than 1 s permitted lower mouth pressures but increased time over which airway pressure exceeded lower esophageal sphincter pressure. The gastric volume increased with breath durations that exceeded 1 s for both mouth pressure profiles. Breath duration of ∼1.0 s caused the least gastric inflation in most scenarios. Very low esophageal sphincter pressure favored a shift toward 0.5 s. High resistance and low compliance each increased gastric inflation and altered ideal breath times. The model illustrated a general theory of optimal rescue ventilation. Breath duration with an unprotected airway should be 1 s to minimize gastric inflation. Short pressure-driven and long duration-driven gastric inflation regimens provide a unifying explanation for results in past studies. Copyright © 2018 by Daedalus Enterprises.

  14. Collective fluid mechanics of honeybee nest ventilation

    NASA Astrophysics Data System (ADS)

    Gravish, Nick; Combes, Stacey; Wood, Robert J.; Peters, Jacob

    2014-11-01

    Honeybees thermoregulate their brood in the warm summer months by collectively fanning their wings and creating air flow through the nest. During nest ventilation workers flap their wings in close proximity in which wings continuously operate in unsteady oncoming flows (i.e. the wake of neighboring worker bees) and near the ground. The fluid mechanics of this collective aerodynamic phenomena are unstudied and may play an important role in the physiology of colony life. We have performed field and laboratory observations of the nest ventilation wing kinematics and air flow generated by individuals and groups of honeybee workers. Inspired from these field observations we describe here a robotic model system to study collective flapping wing aerodynamics. We microfabricate arrays of 1.4 cm long flapping wings and observe the air flow generated by arrays of two or more fanning robotic wings. We vary phase, frequency, and separation distance among wings and find that net output flow is enhanced when wings operate at the appropriate phase-distance relationship to catch shed vortices from neighboring wings. These results suggest that by varying position within the fanning array honeybee workers may benefit from collective aerodynamic interactions during nest ventilation.

  15. The unique contribution of manual chest compression-vibrations to airflow during physiotherapy in sedated, fully ventilated children.

    PubMed

    Gregson, Rachael K; Shannon, Harriet; Stocks, Janet; Cole, Tim J; Peters, Mark J; Main, Eleanor

    2012-03-01

    This study aimed to quantify the specific effects of manual lung inflations with chest compression-vibrations, commonly used to assist airway clearance in ventilated patients. The hypothesis was that force applied during the compressions made a significant additional contribution to increases in peak expiratory flow and expiratory to inspiratory flow ratio over and above that resulting from accompanying increases in inflation volume. Prospective observational study. Cardiac and general pediatric intensive care. Sedated, fully ventilated children. Customized force-sensing mats and a commercial respiratory monitor recorded force and respiration during physiotherapy. Percentage changes in peak expiratory flow, peak expiratory to inspiratory flow ratios, inflation volume, and peak inflation pressure between baseline and manual inflations with and without compression-vibrations were calculated. Analysis of covariance determined the relative contribution of changes in pressure, volume, and force to influence changes in peak expiratory flow and peak expiratory to inspiratory flow ratio. Data from 105 children were analyzed (median age, 1.3 yrs; range, 1 wk to 15.9 yrs). Force during compressions ranged from 15 to 179 N (median, 46 N). Peak expiratory flow increased on average by 76% during compressions compared with baseline ventilation. Increases in peak expiratory flow were significantly related to increases in inflation volume, peak inflation pressure, and force with peak expiratory flow increasing by, on average, 4% for every 10% increase in inflation volume (p < .001), 5% for every 10% increase in peak inflation pressure (p = .005), and 3% for each 10 N of applied force (p < .001). By contrast, increase in peak expiratory to inspiratory flow ratio was only related to applied force with a 4% increase for each 10 N of force (p < .001). These results provide evidence of the unique contribution of compression forces in increasing peak expiratory flow and peak expiratory to inspiratory flow ratio bias over and above that related to accompanying changes from manual hyperinflations. Force generated during compression-vibrations was the single significant factor in multivariable analysis to explain the increases in expiratory flow bias. Such increases in the expiratory bias provide theoretically optimal physiological conditions for cephalad mucus movement in fully ventilated children.

  16. Automated respiratory cycles selection is highly specific and improves respiratory mechanics analysis.

    PubMed

    Rigo, Vincent; Graas, Estelle; Rigo, Jacques

    2012-07-01

    Selected optimal respiratory cycles should allow calculation of respiratory mechanic parameters focusing on patient-ventilator interaction. New computer software automatically selecting optimal breaths and respiratory mechanics derived from those cycles are evaluated. Retrospective study. University level III neonatal intensive care unit. Ten mins synchronized intermittent mandatory ventilation and assist/control ventilation recordings from ten newborns. The ventilator provided respiratory mechanic data (ventilator respiratory cycles) every 10 secs. Pressure, flow, and volume waves and pressure-volume, pressure-flow, and volume-flow loops were reconstructed from continuous pressure-volume recordings. Visual assessment determined assisted leak-free optimal respiratory cycles (selected respiratory cycles). New software graded the quality of cycles (automated respiratory cycles). Respiratory mechanic values were derived from both sets of optimal cycles. We evaluated quality selection and compared mean values and their variability according to ventilatory mode and respiratory mechanic provenance. To assess discriminating power, all 45 "t" values obtained from interpatient comparisons were compared for each respiratory mechanic parameter. A total of 11,724 breaths are evaluated. Automated respiratory cycle/selected respiratory cycle selections agreement is high: 88% of maximal κ with linear weighting. Specificity and positive predictive values are 0.98 and 0.96, respectively. Averaged values are similar between automated respiratory cycle and ventilator respiratory cycle. C20/C alone is markedly decreased in automated respiratory cycle (1.27 ± 0.37 vs. 1.81 ± 0.67). Tidal volume apparent similarity disappears in assist/control: automated respiratory cycle tidal volume (4.8 ± 1.0 mL/kg) is significantly lower than for ventilator respiratory cycle (5.6 ± 1.8 mL/kg). Coefficients of variation decrease for all automated respiratory cycle parameters in all infants. "t" values from ventilator respiratory cycle data are two to three times higher than ventilator respiratory cycles. Automated selection is highly specific. Automated respiratory cycle reflects most the interaction of both ventilator and patient. Improving discriminating power of ventilator monitoring will likely help in assessing disease status and following trends. Averaged parameters derived from automated respiratory cycles are more precise and could be displayed by ventilators to improve real-time fine tuning of ventilator settings.

  17. In-depth survey report of American Airlines plating facility

    NASA Astrophysics Data System (ADS)

    Mortimer, V. D., Jr.

    1982-12-01

    An in depth survey was conducted at the American Airlines Maintenance and Engineering Center as part of National Institute for Occupational Safety and Health (NIOSH) study evaluating measures to control occupational health hazards associated with the metal plating industry. This American Airlines plating facility, employing approximately 25 workers, is primarily engaged in plating hard chromium, nickel and cadmium on aircraft engine and landing gear parts. Six tanks were studied, including an electroless nickel tank. Area and personal samples for chromium, nickel, cadmium, and cyanide were collected. Ventilation airflow and tank dimensions were measured and data recorded on plating operations. The relationships between air contaminants emitted, local exhaust ventilation flow rate, tank size, and plating activity were evaluated.

  18. Low-flow venovenous CO₂ removal in association with lung protective ventilation strategy in patients who develop severe progressive respiratory acidosis after lung transplantation.

    PubMed

    Ruberto, F; Bergantino, B; Testa, M C; D'Arena, C; Zullino, V; Congi, P; Paglialunga, S G; Diso, D; Venuta, F; Pugliese, F

    2013-09-01

    Primary graft dysfunction (PGD) might occur after lung transplantation. In some severe cases, conventional therapies like ventilatory support, administration of inhaled nitric oxide (iNO), and intravenous prostacyclins are not sufficient to provide an adequate gas exchange. The aim of our study was to evaluate the use of a lung protective ventilation strategy associated with a low-flow venovenous CO2 removal treatment to reduce ventilator-associated injury in patients that develop severe PGD after lung transplantation. From January 2009 to January 2011, 3 patients developed PGD within 24 hours after lung transplantation. In addition to conventional medical treatment, including hemodynamic support, iNO and prostaglandin E1 (PGE1), we initiated a ventilatory protective strategy associated with low-flow venovenous CO2 removal treatment (LFVVECCO2R). Hemodynamic and respiratory parameters were assessed at baseline as well as after 3, 12, 24, and 48 hours. No adverse events were registered. Despite decreased baseline elevated pulmonary positive pressures, application of a protective ventilation strategy with LFVVECCO2R reduced PaCO2 and pulmonary infiltrates as well as increased pH values and PaO2/FiO2 ratios. Every patient showed simultaneous improvement of clinical and hemodynamic conditions. They were weaned from mechanical ventilation and extubated after 24 hours after the use of the low-flow venovenous CO2 removal device. The use of LFVVECCO2R together with a protective lung ventilation strategy during the perioperative period of lung transplantation may be a valid clinical strategy for patients with PGD and severe respiratory acidosis occured despite adequate mechanical ventilation. Copyright © 2013 Elsevier Inc. All rights reserved.

  19. FiO2 delivered by a turbine portable ventilator with an oxygen concentrator in an Austere environment.

    PubMed

    Bordes, Julien; Erwan d'Aranda; Savoie, Pierre-Henry; Montcriol, Ambroise; Goutorbe, Philippe; Kaiser, Eric

    2014-09-01

    Management of critically ill patients in austere environments is a logistic challenge. Availability of oxygen cylinders for the mechanically ventilated patient may be difficult in such a context. A solution is to use a ventilator able to function with an oxygen concentrator. We tested the SeQual Integra™ (SeQual, San Diego, CA) 10-OM oxygen concentrator paired with the Pulmonetic System(®) LTV 1000 ventilator (Pulmonetic Systems, Minneapolis, MN) and evaluated the delivered fraction of inspired oxygen (FiO2) across a range of minute volumes and combinations of ventilator settings. Two LTV 1000 ventilators were tested. The ventilators were attached to a test lung and FiO2 was measured by a gas analyzer. Continuous-flow oxygen was generated by the OC from 0.5 L/min to 10 L/min and injected into the oxygen inlet port of the LTV 1000. Several combinations of ventilator settings were evaluated to determine the factors affecting the delivered FiO2. The LTV 1000 ventilator is a turbine ventilator that is able to deliver high FiO2 when functioning with an oxygen concentrator. However, modifications of the ventilator settings such as increase in minute ventilation affect delivered FiO2 even if oxygen flow is constant on the oxygen concentrator. The ability of an oxygen concentrator to deliver high FiO2 when used with a turbine ventilator makes this method of oxygen delivery a viable alternative to cylinders in austere environments when used with a turbine ventilator. However, FiO2 has to be monitored continuously because delivered FiO2 decreases when minute ventilation is increased. Copyright © 2014 Elsevier Inc. All rights reserved.

  20. ALGORITHMS FOR ESTIMATING RESTING METABOLIC RATE AND ACTIVITY SPECIFIC VENTILATION RATES FOR USE IN COMPLEX EXPOSURE AND INTAKE DOSE MODELS

    EPA Science Inventory

    This work summarizes advancements made that allow for better estimation of resting metabolic rate (RMR) and subsequent estimation of ventilation rates (i.e., total ventilation (VE) and alveolar ventilation (VA)) for individuals of both genders and all ages. ...

  1. Wind Extraction for Natural Ventilation

    NASA Astrophysics Data System (ADS)

    Fagundes, Tadeu; Yaghoobian, Neda; Kumar, Rajan; Ordonez, Juan

    2017-11-01

    Due to the depletion of energy resources and the environmental impact of pollution and unsustainable energy resources, energy consumption has become one of the main concerns in our rapidly growing world. Natural ventilation, a traditional method to remove anthropogenic and solar heat gains, proved to be a cost-effective, alternative method to mechanical ventilation. However, while natural ventilation is simple in theory, its detailed design can be a challenge, particularly for wind-driven ventilation, which its performance highly involves the buildings' form, surrounding topography, turbulent flow characteristics, and climate. One of the main challenges with wind-driven natural ventilation schemes is due to the turbulent and unpredictable nature of the wind around the building that impose complex pressure loads on the structure. In practice, these challenges have resulted in founding the natural ventilation mainly on buoyancy (rather than the wind), as the primary force. This study is the initial step for investigating the physical principals of wind extraction over building walls and investigating strategies to reduce the dependence of the wind extraction on the incoming flow characteristics and the target building form.

  2. An Earth-Based Equivalent Low Stretch Apparatus to Assess Material Flammability for Microgravity and Extraterrestrial Fire-Safety Applications

    NASA Technical Reports Server (NTRS)

    Olson, S. L.; Beeson, H.; Haas, J. P.

    2003-01-01

    The objective of this project is to modify the standard oxygen consumption (cone) calorimeter (described in ASTM E 1354 and NASA STD 6001 Test 2) to provide a reproducible bench-scale test environment that simulates the buoyant or ventilation flow that would be generated by or around a burning surface in a spacecraft or extraterrestrial gravity level. This apparatus will allow us to conduct normal gravity experiments that accurately and quantitatively evaluate a material's flammability characteristics in the real-use environment of spacecraft or extra-terrestrial gravitational acceleration. The Equivalent Low Stretch Apparatus (ELSA) uses an inverted cone geometry with the sample burning in a ceiling fire configuration that provides a reproducible bench-scale test environment that simulates the buoyant or ventilation flow that would be generated by a flame in a spacecraft or extraterrestrial gravity level. Prototype unit testing results are presented in this paper. Ignition delay times and regression rates for PMMA are presented over a range of radiant heat flux levels and equivalent stretch rates which demonstrate the ability of ELSA to simulate key features of microgravity and extraterrestrial fire behavior.

  3. STUDY PROGRAM FOR TURBO-COOLER FOR PRODUCING ENGINE COOLING AIR.

    DTIC Science & Technology

    VANES , STAGNATION POINT, DECELERATION, ACCELERATION, SUPERSONIC DIFFUSERS, TURBINE BLADES , EVAPOTRANSPIRATION, LIQUID COOLED, HEAT TRANSFER, GAS BEARINGS, SEALS...HYPERSONIC AIRCRAFT , COOLING + VENTILATING EQUIPMENT), (*GAS TURBINES , COOLING + VENTILATING EQUIPMENT), HYPERSONIC FLOW, AIR COOLED, AIRCRAFT ... ENGINES , FEASIBILITY STUDIES, PRESSURE, SUPERSONIC CHARACTERISTICS, DESIGN, HEAT EXCHANGERS, COOLING (U) AXIAL FLOW TURBINES , DUCT INLETS, INLET GUIDE

  4. Integrated analysis of numerical weather prediction and computational fluid dynamics for estimating cross-ventilation effects on inhaled air quality inside a factory

    NASA Astrophysics Data System (ADS)

    Murga, Alicia; Sano, Yusuke; Kawamoto, Yoichi; Ito, Kazuhide

    2017-10-01

    Mechanical and passive ventilation strategies directly impact indoor air quality. Passive ventilation has recently become widespread owing to its ability to reduce energy demand in buildings, such as the case of natural or cross ventilation. To understand the effect of natural ventilation on indoor environmental quality, outdoor-indoor flow paths need to be analyzed as functions of urban atmospheric conditions, topology of the built environment, and indoor conditions. Wind-driven natural ventilation (e.g., cross ventilation) can be calculated through the wind pressure coefficient distributions of outdoor wall surfaces and openings of a building, allowing the study of indoor air parameters and airborne contaminant concentrations. Variations in outside parameters will directly impact indoor air quality and residents' health. Numerical modeling can contribute to comprehend these various parameters because it allows full control of boundary conditions and sampling points. In this study, numerical weather prediction modeling was used to calculate wind profiles/distributions at the atmospheric scale, and computational fluid dynamics was used to model detailed urban and indoor flows, which were then integrated into a dynamic downscaling analysis to predict specific urban wind parameters from the atmospheric to built-environment scale. Wind velocity and contaminant concentration distributions inside a factory building were analyzed to assess the quality of the human working environment by using a computer simulated person. The impact of cross ventilation flows and its variations on local average contaminant concentration around a factory worker, and inhaled contaminant dose, were then discussed.

  5. Conservative fluid management prevents age-associated ventilator induced mortality.

    PubMed

    Herbert, Joseph A; Valentine, Michael S; Saravanan, Nivi; Schneck, Matthew B; Pidaparti, Ramana; Fowler, Alpha A; Reynolds, Angela M; Heise, Rebecca L

    2016-08-01

    Approximately 800 thousand patients require mechanical ventilation in the United States annually with an in-hospital mortality rate of over 30%. The majority of patients requiring mechanical ventilation are over the age of 65 and advanced age is known to increase the severity of ventilator-induced lung injury (VILI) and in-hospital mortality rates. However, the mechanisms which predispose aging ventilator patients to increased mortality rates are not fully understood. Ventilation with conservative fluid management decreases mortality rates in acute respiratory distress patients, but to date there has been no investigation of the effect of conservative fluid management on VILI and ventilator associated mortality rates. We hypothesized that age-associated increases in susceptibility and incidence of pulmonary edema strongly promote age-related increases in ventilator associated mortality. 2month old and 20month old male C57BL6 mice were mechanically ventilated with either high tidal volume (HVT) or low tidal volume (LVT) for up to 4h with either liberal or conservative fluid support. During ventilation, lung compliance, total lung capacity, and hysteresis curves were quantified. Following ventilation, bronchoalveolar lavage fluid was analyzed for total protein content and inflammatory cell infiltration. Wet to dry ratios were used to directly measure edema in excised lungs. Lung histology was performed to quantify alveolar barrier damage/destruction. Age matched non-ventilated mice were used as controls. At 4h, both advanced age and HVT ventilation significantly increased markers of inflammation and injury, degraded pulmonary mechanics, and decreased survival rates. Conservative fluid support significantly diminished pulmonary edema and improved pulmonary mechanics by 1h in advanced age HVT subjects. In 4h ventilations, conservative fluid support significantly diminished pulmonary edema, improved lung mechanics, and resulted in significantly lower mortality rates in older subjects. Our study demonstrates that conservative fluid alone can attenuate the age associated increase in ventilator associated mortality. Copyright © 2016 Elsevier Inc. All rights reserved.

  6. Conservative Fluid Management Prevents Age-Associated Ventilator Induced Mortality

    PubMed Central

    Herbert, Joseph A.; Valentine, Michael S.; Saravanan, Nivi; Schneck, Matthew B.; Pidaparti, Ramana; Fowler, Alpha A.; Reynolds, Angela M.; Heise, Rebecca L.

    2017-01-01

    Background Approximately 800 thousand patients require mechanical ventilation in the United States annually with an in-hospital mortality rate of over 30%. The majority of patients requiring mechanical ventilation are over the age of 65 and advanced age is known to increase the severity of ventilator-induced lung injury (VILI) and in-hosptial mortality rates. However, the mechanisms which predispose aging ventilator patients to increased mortality rates are not fully understood. Ventilation with conservative fluid management decreases mortality rates in acute respiratory distress patients, but to date there has been no investigation of the effect of conservative fluid management on VILI and ventilator associated mortality rates. We hypothesized that age-associated increases in susceptibility and incidence of pulmonary edema strongly promote age-related increases in ventilator associated mortality. Methods 2 month old and 20 month old male C57BL6 mice were mechanically ventilated with either high tidal volume (HVT) or low tidal volume (LVT) for up to 4 hours with either liberal or conservative fluid support. During ventilation, lung compliance, total lung capacity, and hysteresis curves were quantified. Following ventilation, bronchoalveolar lavage fluid was analyzed for total protein content and inflammatory cell infiltration. Wet to dry ratios were used to directly measure edema in excised lungs. Lung histology was performed to quantify alveolar barrier damage/destruction. Age matched non-ventilated mice were used as controls. Results At 4hrs, both advanced age and HVT ventilation significantly increased markers of inflammation and injury, degraded pulmonary mechanics, and decreased survival rates. Conservative fluid support significantly diminished pulmonary edema and improved pulmonary mechanics by 1hr in advanced age HVT subjects. In 4hr ventilations, conservative fluid support significantly diminished pulmonary edema, improved lung mechanics, and resulted in significantly lower mortality rates in older subjects. Conclusion Our study demonstrates that conservative fluid alone can attenuate the age associated increase in ventilator associated mortality. PMID:27188767

  7. Miners' Misconceptions of Flow Distribution Within Circuits as a Factor Influencing Underground Mining Accidents.

    NASA Astrophysics Data System (ADS)

    Passaro, Perry David

    Misconceptions can be thought of as naive approaches to problem solving that are perceptually appealing but incorrect and inconsistent with scientific evidence (Piaget, 1929). One type of misconception involves flow distributions within circuits. This concept is important because miners' conceptual errors about flow distribution changes within complex circuits may be in part responsible for fatal mine disasters. Based on the theory that misconceptions of flow distribution changes within circuits were responsible for underground mine disasters involving mine ventilation circuits, a series of studies was undertaken with mining engineering students, professional mining engineers, as well as mine foremen, mine supervisors, mine rescue members, mine maintenance personnel, mining researchers and working miners to identify these conceptual errors and errors in mine ventilation procedures. Results indicate that misconceptions of flow distribution changes within circuits exist in over 70 percent of the subjects sampled. It is assumed that these misconceptions of flow distribution changes within circuits result in errors of judgment when miners are faced with inferring and changing ventilation arrangements when two or more mine sections are connected. Furthermore, it is assumed that these misconceptions are pervasive in the mining industry and may be responsible for at least two mine ventilation disasters. The findings of this study are consistent with Piaget's (1929) model of figurative and operative knowledge. This model states that misconceptions are in part due to a lack of knowledge of dynamic transformations and how to apply content information. Recommendations for future research include the development of an interactive expert system for training miners with ventilation arrangements. Such a system would meet the educational recommendations made by Piaget (1973b) by involving a hands-on approach that allows discovery, interaction, the opportunity to make mistakes and to review the cognitive concepts on which the subject relied during his manipulation of the ventilation system.

  8. A new device for 100 per cent humidification of inspired air

    PubMed Central

    Larsson, Anders; Gustafsson, Ann; Svanborg, Lennart

    2000-01-01

    Introduction: Devices for active humidification of the inspired air in mechanically ventilated patients cause water condensation in the ventilator tubing, which may become contaminated or interfere with the function of the ventilator. The present study describes and tests the performance of a new humidifier, which is designed to eliminate water condensation. Objectives: To test the performance of the new humidifier at different ventilator settings in a lung model, and to compare this new humidifier with a conventional active humidifier in ventilator-treated critically ill patients. Materials and methods: The humidifier (Humid-Heat; Louis Gibeck AB, Upplands Väsby, Sweden) consists of a supply unit with a microprocessor and a water pump, and a humidification device, which is placed between the Y-piece and the endotracheal tube. The humidification device is based on a hygroscopic heat-moisture exchanger (HME), which absorbs the expired heat and moisture and releases it into the inspired gas. External heat and water are then added to the patient side of the HME, so the inspired gas should reach 100% humidity at 37°C (44 mg H2O/l air). The external water is delivered to the humidification device via a pump onto a wick and then evaporated into the inspired air by an electrical heater. The microprocessor controls the water pump and the heater by an algorithm using the minute ventilation (which is fed into the microprocessor) and the airway temperature measured by a sensor mounted in the flex-tube on the patient side of the humidification device. The performance characteristics were tested in a lung model ventilated with a constant flow (inspiratory:expiratory ratio 1:2, rate 12–20 breaths/min and a minute ventilation of 3–25 l/min) or with a decelerating flow (inspiratory:expiratory ratio 1:2, rate 12–15 breaths/min and a minute ventilation of 4.7–16.4 l/min). The device was also tested prospectively and in a randomized order compared with a conventional active humidifier (Fisher & Paykel MR730, Auckland, New Zealand) in eight mechanically ventilated, endotracheally intubated patients in the intensive care unit. The test period with each device was 24 h. The amount of fluid consumed and the amount of water in the water traps were measured. The number of times that the water traps were emptied, changes of machine filters, the suctions and quality of secretions, nebulizations, and the amount of saline instillations and endotracheal tube obstruction were recorded. In order to evaluate increased expiratory resistance due to the device, the airway pressure was measured at the end of a prolonged end-expiratory pause at 1 h of use and at the end of the test, and was compared with the corresponding pressure before the experiment. The body temperature of the patient was measured before and after the test of each device. Results: Both with constant flow and decelerating flow, the Humid-Heat gave an absolute humidity of 41–44 mgH2O/l at 37°C, with the lower level at the highest ventilation. In the patients, both Humid-Heat and the conventional active humidifier (MR730) maintained temperatures, indicating that they provided the intended heat and moisture to the inspired air. With both devices, the body temperature was maintained during the test period. There was no difference in the amount of secretions, the quality of the secretions and the frequency of suctions, saline instillations or nebulizations between the test periods with the two devices. There was no endotracheal tube obstruction, and after 1 h of use and at the end of the test no increased airway resistance was found with either device. When the MR730 was used, however, the water traps needed to be emptied six to 14 (mean eight) times (total amount of fluid in the traps was 100–300 ml) and the machine filters were changed two to six (mean four) times due to an excessive amount of condensed water with flow obstruction. No condensation of water was found in the tubing with the Humid-Heat. The water consumption was 23–65 ml/h (mean 30 ml/h) with the MR730 and 4–8 ml/h (mean 6 ml/h) with the Humid-Heat (P < 0.0008). The same relations were found when the water consumption was corrected for differences in minute ventilation. Discussion: The new humidifier, the Humid-Heat, gave an absolute humidity of 41–44 mg/l at 37°C in the bench tests. The tests in ventilated patients showed that the device was well tolerated and that condensation in the tubing was eliminated. There was no need to empty water traps. The test period was too short to evaluate whether the new device had any other advantages or disadvantages compared with conventional humidifiers. PMID:11056746

  9. Effect of flashlight guidance on manual ventilation performance in cardiopulmonary resuscitation: A randomized controlled simulation study.

    PubMed

    Kim, Ji Hoon; Beom, Jin Ho; You, Je Sung; Cho, Junho; Min, In Kyung; Chung, Hyun Soo

    2018-01-01

    Several auditory-based feedback devices have been developed to improve the quality of ventilation performance during cardiopulmonary resuscitation (CPR), but their effectiveness has not been proven in actual CPR situations. In the present study, we investigated the effectiveness of visual flashlight guidance in maintaining high-quality ventilation performance. We conducted a simulation-based, randomized, parallel trial including 121 senior medical students. All participants were randomized to perform ventilation during 2 minutes of CPR with or without flashlight guidance. For each participant, we measured mean ventilation rate as a primary outcome and ventilation volume, inspiration velocity, and ventilation interval as secondary outcomes using a computerized device system. Mean ventilation rate did not significantly differ between flashlight guidance and control groups (P = 0.159), but participants in the flashlight guidance group exhibited significantly less variation in ventilation rate than participants in the control group (P<0.001). Ventilation interval was also more regular among participants in the flashlight guidance group. Our results demonstrate that flashlight guidance is effective in maintaining a constant ventilation rate and interval. If confirmed by further studies in clinical practice, flashlight guidance could be expected to improve the quality of ventilation performed during CPR.

  10. Efforts and Programs of the Department of Defense Relating to the Prevention, Mitigation, and Treatment of Blast Injuries

    DTIC Science & Technology

    2007-01-01

    Combat Critical Care Engineering: Evaluation of Closed Loop Control of Ventilation and Oxygen Flow During Resuscitation in the Compensatory and...Decompensatory Phases of Hemorrhagic Shock: This effort evaluated closed loop control of ventilation and oxygen flow during resuscitation in the...Cerebral Injury Volume, Cerebral Edema, Cerebral Blood Flow and Reactivity, and Histopathology in a Rat Model of Traumatic Brain Injury and Hemorrhagic

  11. A microprocessor-controlled tracheal insufflation-assisted total liquid ventilation system.

    PubMed

    Parker, James Courtney; Sakla, Adel; Donovan, Francis M; Beam, David; Chekuri, Annu; Al-Khatib, Mohammad; Hamm, Charles R; Eyal, Fabien G

    2009-09-01

    A prototype time cycled, constant volume, closed circuit perfluorocarbon (PFC) total liquid ventilator system is described. The system utilizes microcontroller-driven display and master control boards, gear motor pumps, and three-way solenoid valves to direct flow. A constant tidal volume and functional residual capacity (FRC) are maintained with feedback control using end-expiratory and end-inspiratory stop-flow pressures. The system can also provide a unique continuous perfusion (bias flow, tracheal insufflation) through one lumen of a double-lumen endotracheal catheter to increase washout of dead space liquid. FRC and arterial blood gases were maintained during ventilation with Rimar 101 PFC over 2-3 h in normal piglets and piglets with simulated pulmonary edema induced by instillation of albumin solution. Addition of tracheal insufflation flow significantly improved the blood gases and enhanced clearance of instilled albumin solution during simulated edema.

  12. A method of reconstruction of clinical gas-analyzer signals corrupted by positive-pressure ventilation.

    PubMed

    Farmery, A D; Hahn, C E

    2001-04-01

    The use of sidestream infrared and paramagnetic clinical gas analyzers is widespread in anesthesiology and respiratory medicine. For most clinical applications, these instruments are entirely satisfactory. However, their ability to measure breath-by-breath volumetric gas fluxes, as required for measurement of airway dead space, oxygen uptake, and so on, is usually inferior to that of the mass spectrometer, and this is thought to be due, in part, to their slower response times. We describe how volumetric gas analysis with the Datex Ultima analyzer, although reasonably accurate for spontaneous ventilation, gives very inaccurate results in conditions of positive-pressure ventilation. We show that this problem is a property of the gas sampling system rather than the technique of gas analysis itself. We examine the source of this error and describe how cyclic changes in airway pressure result in variations in the flow rate of the gas within the sampling catheter. This results in the phenomenon of "time distortion," and the resultant gas concentration signal becomes a nonlinear time series. This corrupted signal cannot be aligned or integrated with the measured flow signal. We describe a method to correct for this effect. With the use of this method, measurements required for breath-by-breath gas-exchange models can be made easily and reliably in the clinical setting.

  13. Tolerance of Volume Control Noninvasive Ventilation in Subjects With Amyotrophic Lateral Sclerosis.

    PubMed

    Martínez, Daniel; Sancho, Jesús; Servera, Emilio; Marín, Julio

    2015-12-01

    Noninvasive ventilation (NIV) tolerance has been identified as an independent predictor of survival in amyotrophic lateral sclerosis (ALS). Volume control continuous mandatory ventilation (VC-CMV) NIV has been associated with poor tolerance. The aim of this study was to determine the tolerance of subjects with ALS to VC-CMV NIV. This was a prospective study involving subjects with ALS who were treated with VC-CMV NIV. Respiratory and functional parameters were recorded when the subjects began ventilatory support. NIV tolerance was evaluated after 3 months. Eighty-seven subjects with ALS were included. After 3 months, 80 subjects (92%) remained tolerant of NIV. Tolerant subjects presented greater survival (median 22.0 months, 95% CI 14.78-29.21) than intolerant subjects (median 6.0 months, 95% CI 0.86-11.13) (P = .03). The variables that best predicted NIV tolerance were mechanically assisted cough peak flow (P = .01) and percentage of time spent with SpO2 < 90% at night while on NIV (P = .03) CONCLUSIONS: VC-CMV NIV provides high rates of NIV tolerance in subjects with ALS. Mechanically assisted cough peak flow and percentage of time spent with SpO2 < 90% at night while using NIV are the 2 factors associated with tolerance of VC-CMV NIV in subjects with ALS. Copyright © 2015 by Daedalus Enterprises.

  14. Field study of air change and flow rate in six automobiles.

    PubMed

    Knibbs, L D; de Dear, R J; Atkinson, S E

    2009-08-01

    For many people, a relatively large proportion of daily exposure to a multitude of pollutants may occur inside an automobile. A key determinant of exposure is the amount of outdoor air entering the cabin (i.e. air change or flow rate). We have quantified this parameter in six passenger vehicles ranging in age from 18 years to <1 year, at three vehicle speeds and under four different ventilation settings. Average infiltration into the cabin with all operable air entry pathways closed was between 1 and 33.1 air changes per hour (ACH) at a vehicle speed of 60 km/h, and between 2.6 and 47.3 ACH at 110 km/h, with these results representing the most (2005 Volkswagen Golf) and least air-tight (1989 Mazda 121) vehicles, respectively. Average infiltration into stationary vehicles parked outdoors varied between approximately 0 and 1.4 ACH and was moderately related to wind speed. Measurements were also performed under an air recirculation setting with low fan speed, while airflow rate measurements were conducted under two non-recirculate ventilation settings with low and high fan speeds. The windows were closed in all cases, and over 200 measurements were performed. The results can be applied to estimate pollutant exposure inside vehicles. There is increasing recognition of the often disproportionately large contribution of in-vehicle pollutant exposures to overall measures. This has highlighted the need for accurate and representative quantification of determinant factors to facilitate exposure estimation and mitigation. The ventilation rate in a vehicle cabin is a key parameter affecting the transfer of pollutants from outdoors to the cabin interior, and vice-versa. New data regarding this variable are presented here, and the results indicate substantial variability in outdoor air infiltration into vehicles of differing age. The efficacy of simple measures to reduce outdoor air infiltration into 'leaky' vehicles to increase occupant protection would be a worthwhile avenue of further research.

  15. Quantification of Methane and Ammonia Emissions in a Naturally Ventilated Barn by Using Defined Criteria to Calculate Emission Rates.

    PubMed

    Schmithausen, Alexander J; Schiefler, Inga; Trimborn, Manfred; Gerlach, Katrin; Südekum, Karl-Heinz; Pries, Martin; Büscher, Wolfgang

    2018-05-16

    Extensive experimentation on individual animals in respiration chambers has already been carried out to evaluate the potential of dietary changes and opportunities to mitigate CH₄ emissions from ruminants. Although it is difficult to determine the air exchange rate of open barn spaces, measurements at the herd level should provide similarly reliable and robust results. The primary objective of this study was (1) to define a validity range (data classification criteria (DCC)) for the variables of wind velocity and wind direction during long-term measurements at barn level; and (2) to apply this validity range to a feeding trial in a naturally cross-flow ventilated dairy barn. The application of the DCC permitted quantification of CH₄ and NH₃ emissions during a feeding trial consisting of four periods. Differences between the control group (no supplement) and the experimental group fed a ration supplemented with condensed Acacia mearnsii tannins (CT) became apparent. Notably, CT concentrations of 1% and 3% of ration dry matter did not reduce CH₄ emissions. In contrast, NH₃ emissions decreased 34.5% when 3% CT was supplemented. The data confirm that quantification of trace gases in a naturally ventilated barn at the herd level is possible.

  16. Local exhaust ventilation for the control of welding fumes in the construction industry--a literature review.

    PubMed

    Flynn, Michael R; Susi, Pam

    2012-08-01

    Arc welding is a common unit operation in the construction industry, where frequent changes in location and welding position make it more difficult to control fume exposures than in industries where fixed locations are the norm. Welders may be exposed to a variety of toxic airborne contaminants including manganese (Mn) and hexavalent chromium (CrVI). Local exhaust ventilation (LEV) is a well-known engineering control for welding fumes but has not been adopted widely in the construction industry. This literature review presents data on the performance of a variety of LEV systems for welding fume control from the construction (five references), shipyard (five references), and other industries. The studies indicate that LEV can reduce fume exposures to total particulate, Mn, and CrVI to levels below currently relevant standards. Field studies suggest that 40-50% or more reduction in exposure is possible with portable or fixed LEV systems relative to natural ventilation but that correct positioning of the hood and adequate exhaust flow rates are essential. Successful implementation of extraction guns for gas metal arc welding (GMAW) and flux core arc welding has been demonstrated, indicating that a successful balance between extraction airflow and shielding gas requirements is possible. Work practices are an important part of achieving successful control of fume exposures; in particular, positioning the hood close to the arc, checking exhaust flow rates, and avoiding the plume. Further research is needed on hood size effects for controlling welding fume with portable LEV systems and identifying and overcoming barriers to LEV use in construction.

  17. Comparison between mixed and laminar airflow systems in operating rooms and the influence of human factors: experiences from a Swedish orthopedic center.

    PubMed

    Erichsen Andersson, Annette; Petzold, Max; Bergh, Ingrid; Karlsson, Jón; Eriksson, Bengt I; Nilsson, Kerstin

    2014-06-01

    The importance of laminar airflow systems in operating rooms as protection from surgical site infections has been questioned. The aim of our study was to explore the differences in air contamination rates between displacement ventilation and laminar airflow systems during planned and acute orthopedic implant surgery. A second aim was to compare the influence of the number of people present, the reasons for traffic flow, and the door-opening rates between the 2 systems. Active air sampling and observations were made during 63 orthopedic implant operations. The laminar airflow system resulted in a reduction of 89% in colony forming units in comparison with the displacement system (P < .001). The air samples taken in the preparation rooms showed high levels of bacterial growth (≈ 40 CFU/m(3)). Our study shows that laminar airflow-ventilated operating rooms offer high-quality air during surgery, with very low levels of colony forming units close to the surgical wound. The continuous maintenance of laminar air flow and other technical systems are crucial, because minor failures in complex systems like those in operating rooms can result in a detrimental effect on air quality and jeopardize the safety of patients. The technical ventilation solutions are important, but they do not guarantee clean air, because many other factors, such as the organization of the work and staff behavior, influence air cleanliness. Copyright © 2014 Association for Professionals in Infection Control and Epidemiology, Inc. Published by Mosby, Inc. All rights reserved.

  18. Comparison of ventilator-associated pneumonia (VAP) rates between different ICUs: Implications of a zero VAP rate.

    PubMed

    Sundar, Krishna M; Nielsen, David; Sperry, Paul

    2012-02-01

    Ventilator-associated pneumonia (VAP) is associated with significant morbidity and mortality. Measures to reduce the incidence of VAP have resulted in institutions reporting a zero or near-zero VAP rates. The implications of zero VAP rates are unclear. This study was done to compare outcomes between two intensive care units (ICU) with one of them reporting a zero VAP rate. This study retrospectively compared VAP rates between two ICUs: Utah Valley Regional Medical Center (UVRMC) with 25 ICU beds and American Fork Hospital (AFH) with 9 ICU beds. Both facilities are under the same management and attended by a single group of intensivists. Both ICUs have similar nursing and respiratory staffing patterns. Both ICUs use the same intensive care program for reduction of VAP rates. ICU outcomes between AFH (reporting zero VAP rate) and UVRMC (VAP rate of 2.41/1000 ventilator days) were compared for the years 2007-2008. UVRMC VAP rates during 2007 and 2008 were 2.31/1000 ventilator days and 2.5/1000 ventilator days respectively compared to a zero VAP rate at AFH. The total days of ventilation, mean days of ventilation per patient and mean duration of ICU stay per patient was higher in the UVRMC group as compared to AFH ICU group. There was no significant difference in mean age and APACHE II score between ICU patients at UVRMC and AFH. There was no statistical difference in rates of VAP and mortality between UVRMC and AFH. During comparisons of VAP rate between institutions, a zero VAP rate needs to be considered in the context of overall ventilator days, mean durations of ventilator stay and ICU mortality. Copyright © 2012 Elsevier Inc. All rights reserved.

  19. Association between substandard classroom ventilation rates and students' academic achievement.

    PubMed

    Haverinen-Shaughnessy, U; Moschandreas, D J; Shaughnessy, R J

    2011-04-01

    This study focuses on the relationship between classroom ventilation rates and academic achievement. One hundred elementary schools of two school districts in the southwest United States were included in the study. Ventilation rates were estimated from fifth-grade classrooms (one per school) using CO(2) concentrations measured during occupied school days. In addition, standardized test scores and background data related to students in the classrooms studied were obtained from the districts. Of 100 classrooms, 87 had ventilation rates below recommended guidelines based on ASHRAE Standard 62 as of 2004. There is a linear association between classroom ventilation rates and students' academic achievement within the range of 0.9-7.1 l/s per person. For every unit (1 l/s per person) increase in the ventilation rate within that range, the proportion of students passing standardized test (i.e., scoring satisfactory or above) is expected to increase by 2.9% (95%CI 0.9-4.8%) for math and 2.7% (0.5-4.9%) for reading. The linear relationship observed may level off or change direction with higher ventilation rates, but given the limited number of observations, we were unable to test this hypothesis. A larger sample size is needed for estimating the effect of classroom ventilation rates higher than 7.1 l/s per person on academic achievement. The results of this study suggest that increasing the ventilation rates toward recommended guideline ventilation rates in classrooms should translate into improved academic achievement of students. More studies are needed to fully understand the relationships between ventilation rate, other indoor environmental quality parameters, and their effects on students' health and achievement. Achieving the recommended guidelines and pursuing better understanding of the underlying relationships would ultimately support both sustainable and productive school environments for students and personnel. © 2010 John Wiley & Sons A/S.

  20. Feasibility of the capnogram to monitor ventilation rate during cardiopulmonary resuscitation.

    PubMed

    Aramendi, Elisabete; Elola, Andoni; Alonso, Erik; Irusta, Unai; Daya, Mohamud; Russell, James K; Hubner, Pia; Sterz, Fritz

    2017-01-01

    The rates of chest compressions (CCs) and ventilations are both important metrics to monitor the quality of cardiopulmonary resuscitation (CPR). Capnography permits monitoring ventilation, but the CCs provided during CPR corrupt the capnogram and compromise the accuracy of automatic ventilation detectors. The aim of this study was to evaluate the feasibility of an automatic algorithm based on the capnogram to detect ventilations and provide feedback on ventilation rate during CPR, specifically addressing intervals where CCs are delivered. The dataset used to develop and test the algorithm contained in-hospital and out-of-hospital cardiac arrest episodes. The method relies on adaptive thresholding to detect ventilations in the first derivative of the capnogram. The performance of the detector was reported in terms of sensitivity (SE) and Positive Predictive Value (PPV). The overall performance was reported in terms of the rate error and errors in the hyperventilation alarms. Results were given separately for the intervals with CCs. A total of 83 episodes were considered, resulting in 4880min and 46,740 ventilations (8741 during CCs). The method showed an overall SE/PPV above 99% and 97% respectively, even in intervals with CCs. The error for the ventilation rate was below 1.8min -1 in any group, and >99% of the ventilation alarms were correctly detected. A method to provide accurate feedback on ventilation rate using only the capnogram is proposed. Its accuracy was proven even in intervals where canpography signal was severely corrupted by CCs. This algorithm could be integrated into monitor/defibrillators to provide reliable feedback on ventilation rate during CPR. Copyright © 2016 Elsevier Ireland Ltd. All rights reserved.

  1. Experimental investigation of airborne contaminant transport by a human wake moving in a ventilated aircraft cabin

    NASA Astrophysics Data System (ADS)

    Poussou, Stephane B.

    The air ventilation system in jetliners provides a comfortable and healthy environment for passengers. Unfortunately, the increase in global air traffic has amplified the risks presented by infectious aerosols or noxious material released during flight. Inside the cabin, air typically flows continuously from overhead outlets into sidewall exhausts in a circular pattern that minimizes secondary flow between adjacent seat rows. However, disturbances frequently introduced by individuals walking along an aisle may alter air distribution, and contribute to spreading of contaminants. Numerical simulation of these convoluted transient flow phenomena is difficult and complex, and experimental assessment of contaminant distribution in real cabins often impractical. A fundamental experimental study was undertaken to examine the transport phenomena, to validate computations and to improve air monitoring systems. A finite moving body was modeled in a 10:1 scale simplified aircraft cabin equipped with ventilation, at a Reynolds number (based on body diameter) of the order of 10,000. An experimental facility was designed and constructed to permit measurements of the ventilation and wake velocity fields using particle image velocimetry (PIV). Contaminant migration was imaged using the planar laser induced fluorescence (PLIF) technique. The effect of ventilation was estimated by comparison with a companion baseline study. Results indicate that the evolution of a downwash predominant behind finite bodies of small aspect ratio is profoundly perturbed by the ventilation flow. The reorganization of vortical structures in the near-wake leads to a shorter longitudinal recirculation region. Furthermore, mixing in the wake is modified and contaminant is observed to convect to higher vertical locations corresponding to seated passenger breathing level.

  2. The ventilation problem in schools: literature review

    DOE PAGES

    Fisk, W. J.

    2017-07-06

    Based on a review of literature published in refereed archival journals, ventilation rates in classrooms often fall far short of the minimum ventilation rates specified in standards. We report that there is compelling evidence, from both cross-sectional and intervention studies, of an association of increased student performance with increased ventilation rates. There is evidence that reduced respiratory health effects and reduced student absence are associated with increased ventilation rates. Increasing ventilation rates in schools imposes energy costs and can increase heating, ventilating, and air-conditioning system capital costs. The net annual costs, ranging from a few dollars to about 10 dollarsmore » per person, are less than 0.1% of typical public spending on elementary and secondary education in the United States. Finally, such expenditures seem like a small price to pay given the evidence of health and performance benefits.« less

  3. The ventilation problem in schools: literature review

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Fisk, W. J.

    Based on a review of literature published in refereed archival journals, ventilation rates in classrooms often fall far short of the minimum ventilation rates specified in standards. We report that there is compelling evidence, from both cross-sectional and intervention studies, of an association of increased student performance with increased ventilation rates. There is evidence that reduced respiratory health effects and reduced student absence are associated with increased ventilation rates. Increasing ventilation rates in schools imposes energy costs and can increase heating, ventilating, and air-conditioning system capital costs. The net annual costs, ranging from a few dollars to about 10 dollarsmore » per person, are less than 0.1% of typical public spending on elementary and secondary education in the United States. Finally, such expenditures seem like a small price to pay given the evidence of health and performance benefits.« less

  4. Effect of ventilation rate on air cleanliness and energy consumption in operation rooms at rest.

    PubMed

    Lee, Shih-Tseng; Liang, Ching-Chieh; Chien, Tsung-Yi; Wu, Feng-Jen; Fan, Kuang-Chung; Wan, Gwo-Hwa

    2018-02-27

    The interrelationships between ventilation rate, indoor air quality, and energy consumption in operation rooms at rest are yet to be understood. We investigate the effect of ventilation rate on indoor air quality indices and energy consumption in ORs at rest. The study investigates the air temperature, relative humidity, concentrations of carbon dioxide, particulate matter (PM), and airborne bacteria at different ventilation rates in operation rooms at rest of a medical center. The energy consumption and cost analysis of the heating, ventilating, and air conditioning (HVAC) system in the operation rooms at rest were also evaluated for all ventilation rates. No air-conditioned operation rooms had very highest PM and airborne bacterial concentrations in the operation areas. The bacterial concentration in the operation areas with 6-30 air changes per hour (ACH) was below the suggested level set by the United Kingdom (UK) for an empty operation room. A 70% of reduction in annual energy cost by reducing the ventilation rate from 30 to 6 ACH was found in the operation rooms at rest. Maintenance of operation rooms at ventilation rate of 6 ACH could save considerable amounts of energy and achieve the goal of air cleanliness.

  5. Coal-Packed Methane Biofilter for Mitigation of Green House Gas Emissions from Coal Mine Ventilation Air

    PubMed Central

    Limbri, Hendy; Gunawan, Cindy; Thomas, Torsten; Smith, Andrew; Scott, Jason; Rosche, Bettina

    2014-01-01

    Methane emitted by coal mine ventilation air (MVA) is a significant greenhouse gas. A mitigation strategy is the oxidation of methane to carbon dioxide, which is approximately twenty-one times less effective at global warming than methane on a mass-basis. The low non-combustible methane concentrations at high MVA flow rates call for a catalytic strategy of oxidation. A laboratory-scale coal-packed biofilter was designed and partially removed methane from humidified air at flow rates between 0.2 and 2.4 L min−1 at 30°C with nutrient solution added every three days. Methane oxidation was catalysed by a complex community of naturally-occurring microorganisms, with the most abundant member being identified by 16S rRNA gene sequence as belonging to the methanotrophic genus Methylocystis. Additional inoculation with a laboratory-grown culture of Methylosinus sporium, as investigated in a parallel run, only enhanced methane consumption during the initial 12 weeks. The greatest level of methane removal of 27.2±0.66 g methane m−3 empty bed h−1 was attained for the non-inoculated system, which was equivalent to removing 19.7±2.9% methane from an inlet concentration of 1% v/v at an inlet gas flow rate of 1.6 L min−1 (2.4 min empty bed residence time). These results show that low-cost coal packing holds promising potential as a suitable growth surface and contains methanotrophic microorganisms for the catalytic oxidative removal of methane. PMID:24743729

  6. Comparison of three humidifiers during high-frequency percussive ventilation using the VDR-4® Fail-safe Breathing Circuit Hub.

    PubMed

    Tiffin, Norman H; Short, Kathy A; Jones, Samuel W; Cairns, Bruce A

    2011-01-01

    The VDR-4® high-frequency percussive ventilator (HFPV) has been shown to be beneficial in the management of inhalation injury by improving secretion clearance while maintaining oxygenation and ventilation. Delivery of gas flow during HFPV could lack adequate humidification delivered to the patient because a major portion of the delivered gas flow would bypass the humidifier when using the original VDR-4® ventilator circuit. The authors tested a novel inline vaporizing humidifier and two gas-water interface humidifiers during HFPV using the new VDR-4® Fail-safe Breathing Circuit Hub® to determine whether delivered humidification could be improved. This new humidification system, the Hydrate Omni™, delivers water vapor into the gas flow of the ventilator circuit rather than water droplets as delivered by the gas-water interface humidifiers. Measurements of absolute humidity and gas temperature were made on the three different humidification systems using a test lung model under standard ambient conditions. The authors found that when using the novel inline vaporizer, it provided better humidification when compared with the standard gas-water interface humidifier during HFPV using the new VDR-4® breathing circuit.

  7. Total and regional deposition of inhaled aerosols in supine healthy subjects and subjects with mild-to-moderate COPD

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Darquenne, Chantal; Lamm, Wayne J.; Fine, Janelle M.

    Despite substantial development of sophisticated subject-specific computational models of aerosol transport and deposition in human lungs, experimental validation of predic- tions from these new models is sparse. We collected aerosol retention and exhalation profiles in seven healthy volunteers and six subjects with mild-to-moderate COPD (FEV1 ¼ 50–80%predicted) in the supine posture. Total deposition was measured during continuous breathing of 1 and 2.9 mm-diameter particles (tidal volume of 1 L, flow rate of 0.3 L/s and 0.75 L/s). Bolus inhalations of 1 mm particles were performed to penetration volumes of 200, 500 and 800 mL (flow rate of 0.5 L/s). Aerosolmore » bolus dispersion (H), deposition, and mode shift (MS) were calculated from these data. There was no significant difference in total deposition between healthy subjects and those with COPD. Total deposition increased with increasing particle size and also with increasing flow rate. Similarly, there was no significant difference in aerosol bolus deposition between subject groups. Yet, the rate of increase in dispersion and of decrease in MS with increasing penetration volume was higher in subjects with COPD than in healthy volunteers (H: 0.79870.205 vs. 0.52770.122 mL/mL, p¼ 0.01; MS: - 0.27170.129 vs. - 0.145 70.076 mL/mL, p¼ 0.05) indicating larger ventilation inhomogeneities (based on H) and increased flow sequencing (based on MS) in the COPD than in the healthy group. In conclusion, in the supine posture, deposition appears to lack sensitivity for assessing the effect of lung morphology and/or ventilation distribution alteration induced by mild-to- moderate lung disease on the fate of inhaled aerosols. However, other parameters such as aerosol bolus dispersion and mode shift may be more sensitive parameters for evaluating models of lungs with moderate disease.« less

  8. Outcome-based ventilation: A framework for assessing performance, health, and energy impacts to inform office building ventilation decisions.

    PubMed

    Rackes, A; Ben-David, T; Waring, M S

    2018-07-01

    This article presents an outcome-based ventilation (OBV) framework, which combines competing ventilation impacts into a monetized loss function ($/occ/h) used to inform ventilation rate decisions. The OBV framework, developed for U.S. offices, considers six outcomes of increasing ventilation: profitable outcomes realized from improvements in occupant work performance and sick leave absenteeism; health outcomes from occupant exposure to outdoor fine particles and ozone; and energy outcomes from electricity and natural gas usage. We used the literature to set low, medium, and high reference values for OBV loss function parameters, and evaluated the framework and outcome-based ventilation rates using a simulated U.S. office stock dataset and a case study in New York City. With parameters for all outcomes set at medium values derived from literature-based central estimates, higher ventilation rates' profitable benefits dominated negative health and energy impacts, and the OBV framework suggested ventilation should be ≥45 L/s/occ, much higher than the baseline ~8.5 L/s/occ rate prescribed by ASHRAE 62.1. Only when combining very low parameter estimates for profitable impacts with very high ones for health and energy impacts were all outcomes on the same order. Even then, however, outcome-based ventilation rates were often twice the baseline rate or more. © 2018 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

  9. A multitracer system for multizone ventilation measurement

    NASA Astrophysics Data System (ADS)

    Sherman, Max

    1990-09-01

    Mass transfer due to pressure-driven air flow is one of the most important processes for determining both environmental quality and energy requirements in buildings. Heat, moisture, and contaminants are all transported by air movement between indoors and outdoors as well as between different zones within a building. Measurement of these air flows is critical to understanding the performance of buildings. Virtually all measurements of ventilation are made using the dilution of a tracer gas. The vast majority of such measurements have been made in a single zone, using a single tracer gas. For the past several years LBL has been developing the MultiTracer Measurement System (MTMS) to provide full multizone air flow information in an accurate, real-time manner. MTMS is based on a quadrupole mass spectrometer to provide high-speed concentration analysis of multiple tracer gases in the (low) ppm level that are injected into multiple zones using mass-flow controllers. The measurement and injection system is controlled by a PC and can measure all concentrations in all zones (and adjust the injected tracer flows) within 2 min and can operate unattended for weeks. The resulting injection rate and concentration data can be analyzed to infer the bulk air movement between zones. The system also measures related quantities such as weather and zonal temperature to assist in the data interpretation. Using MTMS, field measurements have been made for the past two years.

  10. Design and Evaluation of a Ventilated Garment for Use in Temperatures up to 200°C

    PubMed Central

    Crockford, G. W.; Hellon, R. F.

    1964-01-01

    The protection of personnel against high air and radiant temperatures is a problem that has been confronting industry for many years now, and for many industrial situations it still has not been solved. The experiments reported here were intended to determine the most suitable form of insulation for a hot entry suit for use primarily in furnace wrecking where mean radiant temperatures of 200°C. are met and where heat-reflecting garments are unsuitable due to the rapid deterioration of the reflecting surface. From a preliminary consideration of the problem it was concluded that a ventilated garment was required and that conventional ventilated garments in which air is induced to flow parallel to the body surfaces (axial ventilation) are basically unsound in design as the air is not utilized for the transfer of heat in the most efficient manner. A new form of ventilation was therefore developed in which air flows out through a permeable suit (radial ventilation). This form of ventilation produces what is called dynamic insulation, and this method of insulation, when compared with two alternative methods on a physical model, was found to be very effective. The model experiments were confirmed by comparative trials of three ventilated suits each using one of three different forms of insulation thought to be suitable for use in heat-protective clothing. Physiological measurements made on the subjects and physical measurement made on the suits confirmed that dynamic insulation is the most suitable insulation for a hot entry suit for furnace wrecking. With the air flows used in these experiments, dynamic insulation had a thermal conductance one-fifth that of conventional static insulation, and sweat losses and oral temperature rises were reduced by one-third and one-half respectively. PMID:14180476

  11. Metabolically Derived Human Ventilation Rates: A Revised Approach Based Upon Oxygen Consumption Rates (Final Report, 2009)

    EPA Science Inventory

    EPA announced the availability of the final report, Metabolically Derived Human Ventilation Rates: A Revised Approach Based Upon Oxygen Consumption Rates. This report provides a revised approach for calculating an individual's ventilation rate directly from their oxygen c...

  12. International Space Station Crew Quarters Ventilation and Acoustic Design Implementation

    NASA Technical Reports Server (NTRS)

    Broyan, James L., Jr.; Cady, Scott M; Welsh, David A.

    2010-01-01

    The International Space Station (ISS) United States Operational Segment has four permanent rack sized ISS Crew Quarters (CQs) providing a private crew member space. The CQs use Node 2 cabin air for ventilation/thermal cooling, as opposed to conditioned ducted air-from the ISS Common Cabin Air Assembly (CCAA) or the ISS fluid cooling loop. Consequently, CQ can only increase the air flow rate to reduce the temperature delta between the cabin and the CQ interior. However, increasing airflow causes increased acoustic noise so efficient airflow distribution is an important design parameter. The CQ utilized a two fan push-pull configuration to ensure fresh air at the crew member's head position and reduce acoustic exposure. The CQ ventilation ducts are conduits to the louder Node 2 cabin aisle way which required significant acoustic mitigation controls. The CQ interior needs to be below noise criteria curve 40 (NC-40). The design implementation of the CQ ventilation system and acoustic mitigation are very inter-related and require consideration of crew comfort balanced with use of interior habitable volume, accommodation of fan failures, and possible crew uses that impact ventilation and acoustic performance. Each CQ required 13% of its total volume and approximately 6% of its total mass to reduce acoustic noise. This paper illustrates the types of model analysis, assumptions, vehicle interactions, and trade-offs required for CQ ventilation and acoustics. Additionally, on-orbit ventilation system performance and initial crew feedback is presented. This approach is applicable to any private enclosed space that the crew will occupy.

  13. A comparative study of behaviors of ventilated supercavities between experimental models with different mounting configurations

    NASA Astrophysics Data System (ADS)

    Lee, Seung-Jae; Kawakami, Ellison; Karn, Ashish; Arndt, Roger E. A.

    2016-08-01

    Small-scale water tunnel experiments of the phenomenon of supercavitation can be carried out broadly using two different kinds of experimental models-in the first model (forward facing model, or FFM), the incoming flow first interacts with the cavitator at front, which is connected to the strut through a ventilation pipe. The second model could have the strut and the ventilation pipe preceding the cavitator (backward facing model, or BFM). This is the continuation of a water tunnel study of the effects of unsteady flows on axisymmetric supercavities. In this study, the unwanted effect of test model configuration on supercavity shape in periodic flows was explored through a comparison of FFM and BFM models. In our experiments, it was found that periodic gust flows have only a minimal effect on the maximum diameter and the cavity length can be shortened above a certain vertical velocity of periodic flows. These findings appear to be robust regardless of the model configuration.

  14. Implementation of a real-time compliance dashboard to help reduce SICU ventilator-associated pneumonia with the ventilator bundle.

    PubMed

    Zaydfudim, Victor; Dossett, Lesly A; Starmer, John M; Arbogast, Patrick G; Feurer, Irene D; Ray, Wayne A; May, Addison K; Pinson, C Wright

    2009-07-01

    Ventilator-associated pneumonia (VAP) causes significant morbidity and mortality in critically ill surgical patients. Recent studies suggest that the success of preventive measures is dependent on compliance with ventilator bundle parameters. Implementation of an electronic dashboard will improve compliance with the bundle parameters and reduce rates of VAP in our surgical intensive care unit (SICU). Time series analysis of VAP rates between January 2005 and July 2008, with dashboard implementation in July 2007. Multidisciplinary SICU at a tertiary-care referral center with a stable case mix during the study period. Patients admitted to the SICU between January 2005 and July 2008. Infection control data were used to establish rates of VAP and total ventilator days. For the time series analysis, VAP rates were calculated as quarterly VAP events per 1000 ventilator days. Ventilator bundle compliance was analyzed after dashboard implementation. Differences between expected and observed VAP rates based on time series analysis were used to estimate the effect of intervention. Average compliance with the ventilator bundle improved from 39% in August 2007 to 89% in July 2008 (P < .001). Rates of VAP decreased from a mean (SD) of 15.2 (7.0) to 9.3 (4.9) events per 1000 ventilator days after introduction of the dashboard (P = .01). Quarterly VAP rates were significantly reduced in the November 2007 through January 2008 and February through April 2008 periods (P < .05). For the August through October 2007 and May through July 2008 quarters, the observed rate reduction was not statistically significant. Implementation of an electronic dashboard improved compliance with ventilator bundle measures and is associated with reduced rates of VAP in our SICU.

  15. Requirements and Sizing Investigation for Constellation Space Suit Portable Life Support System Trace Contaminant Control

    NASA Technical Reports Server (NTRS)

    Paul, Heather L.; Jennings, Mallory A.; Waguespack, Glenn

    2010-01-01

    The Trace Contaminant Control System (TCCS), located within the ventilation loop of the Constellation Space Suit Portable Life Support System (PLSS), is responsible for removing hazardous trace contaminants from the space suit ventilation flow. This paper summarizes the results of a trade study that evaluated if trace contaminant control could be accomplished without a TCCS, relying on suit leakage, ullage loss from the carbon dioxide and humidity control system, and other factors. Trace contaminant generation rates were revisited to verify that values reflect the latest designs for Constellation Space Suit System (CSSS) pressure garment materials and PLSS hardware. Additionally, TCCS sizing calculations were performed and a literature survey was conducted to review the latest developments in trace contaminant technologies.

  16. Acoustic Measurements of an Uninstalled Spacecraft Cabin Ventilation Fan Prototype

    NASA Technical Reports Server (NTRS)

    Koch, L. Danielle; Brown, Clifford A.; Shook, Tony D.; Winkel, James; Kolacz, John S.; Podboy, Devin M.; Loew, Raymond A.; Mirecki, Julius H.

    2012-01-01

    Sound pressure measurements were recorded for a prototype of a spacecraft cabin ventilation fan in a test in the NASA Glenn Acoustical Testing Laboratory. The axial fan is approximately 0.089 m (3.50 in.) in diameter and 0.223 m (9.00 in.) long and has nine rotor blades and eleven stator vanes. At design point of 12,000 rpm, the fan was predicted to produce a flow rate of 0.709 cu m/s (150 cfm) and a total pressure rise of 925 Pa (3.72 in. of water) at 12,000 rpm. While the fan was designed to be part of a ducted atmospheric revitalization system, no attempt was made to throttle the flow or simulate the installed configuration during this test. The fan was operated at six speeds from 6,000 to 13,500 rpm. A 13-microphone traversing array was used to collect sound pressure measurements along two horizontal planes parallel to the flow direction, two vertical planes upstream of the fan inlet and two vertical planes downstream of the fan exhaust. Measurements indicate that sound at blade passing frequency harmonics contribute significantly to the overall audible noise produced by the fan at free delivery conditions.

  17. Computational Study of Ventilation and Disease Spread in Poultry Houses

    NASA Astrophysics Data System (ADS)

    Cimbala, John; Pawar, Sourabh; Wheeler, Eileen; Lindberg, Darla

    2006-11-01

    The air flow in and around poultry houses has been studied numerically with the goal of determining disease spread characteristics and comparing ventilation schemes. A typical manure-belt layer egg production facility is considered. The continuity, momentum, and energy equations are solved for flow both inside and outside poultry houses using the commercial computational fluid dynamics (CFD) code FLUENT. Both simplified two-dimensional and fully three-dimensional geometries are modeled. The spread of virus particles is considered to be analogous to diffusion of a tracer contaminant gas, in this case ammonia. The effect of thermal plumes produced by the hens in the poultry house is also considered. Two ventilation schemes with opposite flow directions are compared. Contours of temperature and ammonia mass fraction for both cases are obtained and compared. The analysis shows that ventilation and air quality characteristics are much better for the case in which the air flow is from bottom to top (enhancing the thermal plume) instead of from top to bottom (fighting the thermal plume) as in most poultry houses. This has implications in air quality control in the event of epidemic outbreaks of avian flu or other infectious diseases.

  18. Wind tunnel and numerical data on the ventilation performance of windcatcher with wing wall.

    PubMed

    Nejat, Payam; Calautit, John Kaiser; Abd Majid, Muhd Zaimi; Hughes, Ben Richard; Zeynali, Iman; Jomehzadeh, Fatemeh

    2016-12-01

    The data presented in this article were the basis for the study reported in the research articles entitled "Evaluation of a two-sided windcatcher integrated with wing wall (as a new design) and comparison with a conventional windcatcher" (P. Nejat, J.K. Calautit, M.Z.A. Majid, B.R. Hughes, I. Zeynali, F. Jomehzadeh, 2016) [1] which presents the effect of wing wall on the air flow distribution under using the windcatchers as a natural ventilation equipment. Here, we detail the wind tunnel testing and numerical set-up used for obtaining the data on ventilation rates and indoor airflow distribution inside a test room with a two-sided windcatcher and wing wall. Three models were integrated with wing wall angled at 30°, 45° and 60° and another windcatcher was a conventional two-sided device. The computer-aided design (CAD) three-dimensional geometries which were produced using Solid Edge modeler are also included in the data article.

  19. Transient states of air parameters after a stoppage and re-start of the main fan / Stany przejściowe parametrów powietrza po postoju i załączeniu wentylatora głównego

    NASA Astrophysics Data System (ADS)

    Wasilewski, Stanisław

    2012-12-01

    A stoppage of the main ventilation fan constitutes a disturbance of ventilation conditions of a deepmine and its effects can cause serious hazards by generating transient states of air and gas flow. Main ventilation fans are the basic deep-mine facilities; therefore, under mining regulations it is only allowed to stop them with the consent and under the conditions specified by the mine maintenance manager. The stoppage of the main ventilation fan may be accompanied by transient air parameters, including the air pressure and flow patterns. There is even the likelihood of reversing the direction of air flow, which, in case of methane mines, can pose a major hazard, particularly in sections of the mine with fire fields or large goaf areas. At the same time, stoppages of deep-mine main ventilation fans create interesting research conditions, which if conducted under the supervision of the monitoring systems, can provide much information about the transient processes of pressure, air and gas flow in underground workings. This article is a discussion of air parameter observations in mine workings made as part of such experiments. It also presents the procedure of the experiments, conducted in three mines. They involved the observation of transient processes of mine air parameters, and most interestingly, the recording of pressure and air and gas flow in the workings of the mine ventilation networks by mine monitoring systems and using specialist recording instruments. In mining practice, both in Poland and elsewhere, software tools and computer modelling methods are used to try and reproduce the conditions prior to and during disasters based on the existing network model and monitoring system data. The use of these tools to simulate the alternatives of combating and liquidation of the gas-fire hazard after its occurrence is an important issue. Measurement data collected during the experiments provides interesting research material for the verification and validation of the software tools used for the simulation of processes occurring in deep-mine ventilation systems.

  20. Air compressor battery duration with mechanical ventilation in a field anesthesia machine.

    PubMed

    Szpisjak, Dale F; Giberman, Anthony A

    2015-05-01

    Compressed air to power field anesthesia machine ventilators may be supplied by air compressor with battery backup. This study determined the battery duration when the compPAC ventilator's air compressor was powered by NiCd battery to ventilate the Vent Aid Training Test Lung modeling high (HC = 0.100 L/cm H2O) and low (LC = 0.020 L/cm H2O) pulmonary compliance. Target tidal volumes (VT) were 500, 750, and 1,000 mL. Respiratory rate = 10 bpm, inspiratory-to-expiratory time ratio = 1:2, and fresh gas flow = 1 L/min air. N = 5 in each group. Control limits were determined from the first 150 minutes of battery power for each run and lower control limit = mean VT - 3SD. Battery depletion occurred when VT was below the lower control limit. Battery duration ranged from 185.8 (±3.2) minutes in the LC-1000 group to 233.3 (±3.6) minutes in the HC-750 group. Battery duration of the LC-1000 group was less than all others (p = 0.027). The differences among the non-LC-1000 groups were not clinically significant. Reprint & Copyright © 2015 Association of Military Surgeons of the U.S.

  1. Potential risk for bacterial contamination in conventional reused ventilator systems and disposable closed ventilator-suction systems.

    PubMed

    Li, Ya-Chi; Lin, Hui-Ling; Liao, Fang-Chun; Wang, Sing-Siang; Chang, Hsiu-Chu; Hsu, Hung-Fu; Chen, Sue-Hsien; Wan, Gwo-Hwa

    2018-01-01

    Few studies have investigated the difference in bacterial contamination between conventional reused ventilator systems and disposable closed ventilator-suction systems. The aim of this study was to investigate the bacterial contamination rates of the reused and disposable ventilator systems, and the association between system disconnection and bacterial contamination of ventilator systems. The enrolled intubated and mechanically ventilated patients used a conventional reused ventilator system and a disposable closed ventilator-suction system, respectively, for a week; specimens were then collected from the ventilator circuit systems to evaluate human and environmental bacterial contamination. The sputum specimens from patients were also analyzed in this study. The detection rate of bacteria in the conventional reused ventilator system was substantially higher than that in the disposable ventilator system. The inspiratory and expiratory limbs of the disposable closed ventilator-suction system had higher bacterial concentrations than the conventional reused ventilator system. The bacterial concentration in the heated humidifier of the reused ventilator system was significantly higher than that in the disposable ventilator system. Positive associations existed among the bacterial concentrations at different locations in the reused and disposable ventilator systems, respectively. The predominant bacteria identified in the reused and disposable ventilator systems included Acinetobacter spp., Bacillus cereus, Elizabethkingia spp., Pseudomonas spp., and Stenotrophomonas (Xan) maltophilia. Both the reused and disposable ventilator systems had high bacterial contamination rates after one week of use. Disconnection of the ventilator systems should be avoided during system operation to decrease the risks of environmental pollution and human exposure, especially for the disposable ventilator system. ClinicalTrials.gov PRS / NCT03359148.

  2. Effect of leak and breathing pattern on the accuracy of tidal volume estimation by commercial home ventilators: a bench study.

    PubMed

    Luján, Manel; Sogo, Ana; Pomares, Xavier; Monsó, Eduard; Sales, Bernat; Blanch, Lluís

    2013-05-01

    New home ventilators are able to provide clinicians data of interest through built-in software. Monitoring of tidal volume (VT) is a key point in the assessment of the efficacy of home mechanical ventilation. To assess the reliability of the VT provided by 5 ventilators in a bench test. Five commercial ventilators from 4 different manufacturers were tested in pressure support mode with the help of a breathing simulator under different conditions of mechanical respiratory pattern, inflation pressure, and intentional leakage. Values provided by the built-in software of each ventilator were compared breath to breath with the VT monitored through an external pneumotachograph. Ten breaths for each condition were compared for every tested situation. All tested ventilators underestimated VT (ranges of -21.7 mL to -83.5 mL, which corresponded to -3.6% to -14.7% of the externally measured VT). A direct relationship between leak and underestimation was found in 4 ventilators, with higher underestimations of the VT when the leakage increased, ranging between -2.27% and -5.42% for each 10 L/min increase in the leakage. A ventilator that included an algorithm that computes the pressure loss through the tube as a function of the flow exiting the ventilator had the minimal effect of leaks on the estimation of VT (0.3%). In 3 ventilators the underestimation was also influenced by mechanical pattern (lower underestimation with restrictive, and higher with obstructive). The inclusion of algorithms that calculate the pressure loss as a function of the flow exiting the ventilator in commercial models may increase the reliability of VT estimation.

  3. [Lung protective ventilation. Ventilatory modes and ventilator parameters].

    PubMed

    Schädler, Dirk; Weiler, Norbert

    2008-06-01

    Mechanical ventilation has a considerable potential for injuring the lung tissue. Therefore, attention has to be paid to the proper choice of ventilatory mode and settings to secure lung-protective ventilation whenever possible. Such ventilator strategy should account for low tidal volume ventilation (6 ml/kg PBW), limited plateau pressure (30 to 35 cm H2O) and positive end-expiratory pressure (PEEP). It is unclear whether pressure controlled or volume controlled ventilation with square flow profile is beneficial. The adjustment of inspiration and expiration time should consider the actual breathing mechanics and anticipate the generation of intrinsic PEEP. Ventilatory modes with the possibility of supporting spontaneous breathing should be used as soon as possible.

  4. Exhaust ventilation in attached garages improves residential indoor air quality.

    PubMed

    Mallach, G; St-Jean, M; MacNeill, M; Aubin, D; Wallace, L; Shin, T; Van Ryswyk, K; Kulka, R; You, H; Fugler, D; Lavigne, E; Wheeler, A J

    2017-03-01

    Previous research has shown that indoor benzene levels in homes with attached garages are higher than homes without attached garages. Exhaust ventilation in attached garages is one possible intervention to reduce these concentrations. To evaluate the effectiveness of this intervention, a randomized crossover study was conducted in 33 Ottawa homes in winter 2014. VOCs including benzene, toluene, ethylbenzene, and xylenes, nitrogen dioxide, carbon monoxide, and air exchange rates were measured over four 48-hour periods when a garage exhaust fan was turned on or off. A blower door test conducted in each garage was used to determine the required exhaust fan flow rate to provide a depressurization of 5 Pa in each garage relative to the home. When corrected for ambient concentrations, the fan decreased geometric mean indoor benzene concentrations from 1.04 to 0.40 μg/m 3 , or by 62% (P<.05). The garage exhaust fan also significantly reduced outdoor-corrected geometric mean indoor concentrations of other pollutants, including toluene (53%), ethylbenzene (47%), m,p-xylene (45%), o-xylene (43%), and carbon monoxide (23%) (P<.05) while having no impact on the home air exchange rate. This study provides evidence that mechanical exhaust ventilation in attached garages can reduce indoor concentrations of pollutants originating from within attached garages. © 2016 Her Majesty the Queen in Right of Canada. Indoor Air published by John Wiley & Sons Ltd.

  5. Effect of sedation with detomidine and butorphanol on pulmonary gas exchange in the horse.

    PubMed

    Nyman, Görel; Marntell, Stina; Edner, Anna; Funkquist, Pia; Morgan, Karin; Hedenstierna, Göran

    2009-05-07

    Sedation with alpha2-agonists in the horse is reported to be accompanied by impairment of arterial oxygenation. The present study was undertaken to investigate pulmonary gas exchange using the Multiple Inert Gas Elimination Technique (MIGET), during sedation with the alpha2-agonist detomidine alone and in combination with the opioid butorphanol. Seven Standardbred trotter horses aged 3-7 years and weighing 380-520 kg, were studied. The protocol consisted of three consecutive measurements; in the unsedated horse, after intravenous administration of detomidine (0.02 mg/kg) and after subsequent butorphanol administration (0.025 mg/kg). Pulmonary function and haemodynamic effects were investigated. The distribution of ventilation-perfusion ratios (VA/Q) was estimated with MIGET. During detomidine sedation, arterial oxygen tension (PaO2) decreased (12.8 +/- 0.7 to 10.8 +/- 1.2 kPa) and arterial carbon dioxide tension (PaCO2) increased (5.9 +/- 0.3 to 6.1 +/- 0.2 kPa) compared to measurements in the unsedated horse. Mismatch between ventilation and perfusion in the lungs was evident, but no increase in intrapulmonary shunt could be detected. Respiratory rate and minute ventilation did not change. Heart rate and cardiac output decreased, while pulmonary and systemic blood pressure and vascular resistance increased. Addition of butorphanol resulted in a significant decrease in ventilation and increase in PaCO2. Alveolar-arterial oxygen content difference P(A-a)O2 remained impaired after butorphanol administration, the VA/Q distribution improved as the decreased ventilation and persistent low blood flow was well matched. Also after subsequent butorphanol no increase in intrapulmonary shunt was evident. The results of the present study suggest that both pulmonary and cardiovascular factors contribute to the impaired pulmonary gas exchange during detomidine and butorphanol sedation in the horse.

  6. Contaminant levels, source strengths, and ventilation rates in California retail stores.

    PubMed

    Chan, W R; Cohn, S; Sidheswaran, M; Sullivan, D P; Fisk, W J

    2015-08-01

    This field study measured ventilation rates and indoor air quality in 21 visits to retail stores in California. Three types of stores, such as grocery, furniture/hardware stores, and apparel, were sampled. Ventilation rates measured using a tracer gas decay method exceeded the minimum requirement of California's Title 24 Standard in all but one store. Concentrations of volatile organic compounds (VOCs), ozone, and carbon dioxide measured indoors and outdoors were analyzed. Even though there was adequate ventilation according to standard, concentrations of formaldehyde and acetaldehyde exceeded the most stringent chronic health guidelines in many of the sampled stores. The whole-building emission rates of VOCs were estimated from the measured ventilation rates and the concentrations measured indoor and outdoor. Estimated formaldehyde emission rates suggest that retail stores would need to ventilate at levels far exceeding the current Title 24 requirement to lower indoor concentrations below California's stringent formaldehyde reference level. Given the high costs of providing ventilation, effective source control is an attractive alternative. Field measurements suggest that California retail stores were well ventilated relative to the minimum ventilation rate requirement specified in the Building Energy Efficiency Standards Title 24. Concentrations of formaldehyde found in retail stores were low relative to levels found in homes but exceeded the most stringent chronic health guideline. Looking ahead, California is mandating zero energy commercial buildings by 2030. To reduce the energy use from building ventilation while maintaining or even lowering formaldehyde in retail stores, effective formaldehyde source control measures are vitally important. Published 2014. This article is a U.S. Government work and is in the public domain in the USA.

  7. Adaptive support ventilation may deliver unwanted respiratory rate-tidal volume combinations in patients with acute lung injury ventilated according to an open lung concept.

    PubMed

    Dongelmans, Dave A; Paulus, Frederique; Veelo, Denise P; Binnekade, Jan M; Vroom, Margreeth B; Schultz, Marcus J

    2011-05-01

    With adaptive support ventilation, respiratory rate and tidal volume (V(T)) are a function of the Otis least work of breathing formula. We hypothesized that adaptive support ventilation in an open lung ventilator strategy would deliver higher V(T)s to patients with acute lung injury. Patients with acute lung injury were ventilated according to a local guideline advising the use of lower V(T) (6-8 ml/kg predicted body weight), high concentrations of positive end-expiratory pressure, and recruitment maneuvers. Ventilation parameters were recorded when the ventilator was switched to adaptive support ventilation, and after recruitment maneuvers. If V(T) increased more than 8 ml/kg predicted body weight, airway pressure was limited to correct for the rise of V(T). Ten patients with a mean (±SD) Pao(2)/Fio(2) of 171 ± 86 mmHg were included. After a switch from pressure-controlled ventilation to adaptive support ventilation, respiratory rate declined (from 31 ± 5 to 21 ± 6 breaths/min; difference = 10 breaths/min, 95% CI 3-17 breaths/min, P = 0.008) and V(T) increased (from 6.5 ± 0.8 to 9.0 ± 1.6 ml/kg predicted body weight; difference = 2.5 ml, 95% CI 0.4-4.6 ml/kg predicted body weight, P = 0.02). Pressure limitation corrected for the rise of V(T), but minute ventilation declined, forcing the user to switch back to pressure-controlled ventilation. Adaptive support ventilation, compared with pressure-controlled ventilation in an open lung strategy setting, delivers a lower respiratory rate-higher V(T) combination. Pressure limitation does correct for the rise of V(T), but leads to a decline in minute ventilation.

  8. A randomised control study comparing the Infant Flow Driver with nasal continuous positive airway pressure in preterm infants.

    PubMed

    Mazzella, M; Bellini, C; Calevo, M G; Campone, F; Massocco, D; Mezzano, P; Zullino, E; Scopesi, F; Arioni, C; Bonacci, W; Serra, G

    2001-09-01

    To compare the effectiveness of the Infant Flow Driver (IFD) with single prong nasal continuous positive airway pressure (nCPAP) in preterm neonates affected by respiratory distress syndrome. Randomised controlled study. Between September 1997 and March 1999, 36 preterm infants who were eligible for CPAP treatment were randomly selected for either nCPAP or IFD and studied prospectively for changes in oxygen requirement and/or respiratory rate. The requirement for mechanical ventilation, complications of treatment, and effects on mid-term outcome were also evaluated. Use of the IFD had a significantly beneficial effect on both oxygen requirement and respiratory rate (p < 0.0001) when compared with nCPAP. Moreover, O(2) requirement and respiratory rate were significantly decreased by four hours (p < 0.001 and p < 0.03 respectively). The probability of remaining supplementary oxygen free over the first 48 hours of treatment was significantly higher in patients treated with the IFD than with nCPAP (p < 0.02). IFD treated patients had a higher success (weaning) rate (94% v 72 %) and shorter duration of treatment (49.3 (31) v 56 (29.7) hours respectively; mean (SD)), although the difference was not significant. IFD appears to be a feasible device for managing respiratory distress syndrome in preterm infants, and benefits may be had with regard to oxygen requirement and respiratory rate when compared with nCPAP. The trend towards reduced requirement for mechanical ventilation, shorter clinical recovery time, and shorter duration of treatment requires further evaluation in a multicentre randomised clinical trial.

  9. Optimal ventilation of the anesthetized pediatric patient.

    PubMed

    Feldman, Jeffrey M

    2015-01-01

    Mechanical ventilation of the pediatric patient is challenging because small changes in delivered volume can be a significant fraction of the intended tidal volume. Anesthesia ventilators have traditionally been poorly suited to delivering small tidal volumes accurately, and pressure-controlled ventilation has become used commonly when caring for pediatric patients. Modern anesthesia ventilators are designed to deliver small volumes accurately to the patient's airway by compensating for the compliance of the breathing system and delivering tidal volume independent of fresh gas flow. These technology advances provide the opportunity to implement a lung-protective ventilation strategy in the operating room based upon control of tidal volume. This review will describe the capabilities of the modern anesthesia ventilator and the current understanding of lung-protective ventilation. An optimal approach to mechanical ventilation for the pediatric patient is described, emphasizing the importance of using bedside monitors to optimize the ventilation strategy for the individual patient.

  10. Potential risk for bacterial contamination in conventional reused ventilator systems and disposable closed ventilator-suction systems

    PubMed Central

    Li, Ya-Chi; Lin, Hui-Ling; Liao, Fang-Chun; Wang, Sing-Siang; Chang, Hsiu-Chu; Hsu, Hung-Fu; Chen, Sue-Hsien

    2018-01-01

    Background Few studies have investigated the difference in bacterial contamination between conventional reused ventilator systems and disposable closed ventilator-suction systems. The aim of this study was to investigate the bacterial contamination rates of the reused and disposable ventilator systems, and the association between system disconnection and bacterial contamination of ventilator systems. Methods The enrolled intubated and mechanically ventilated patients used a conventional reused ventilator system and a disposable closed ventilator-suction system, respectively, for a week; specimens were then collected from the ventilator circuit systems to evaluate human and environmental bacterial contamination. The sputum specimens from patients were also analyzed in this study. Results The detection rate of bacteria in the conventional reused ventilator system was substantially higher than that in the disposable ventilator system. The inspiratory and expiratory limbs of the disposable closed ventilator-suction system had higher bacterial concentrations than the conventional reused ventilator system. The bacterial concentration in the heated humidifier of the reused ventilator system was significantly higher than that in the disposable ventilator system. Positive associations existed among the bacterial concentrations at different locations in the reused and disposable ventilator systems, respectively. The predominant bacteria identified in the reused and disposable ventilator systems included Acinetobacter spp., Bacillus cereus, Elizabethkingia spp., Pseudomonas spp., and Stenotrophomonas (Xan) maltophilia. Conclusions Both the reused and disposable ventilator systems had high bacterial contamination rates after one week of use. Disconnection of the ventilator systems should be avoided during system operation to decrease the risks of environmental pollution and human exposure, especially for the disposable ventilator system. Trial registration ClinicalTrials.gov PRS / NCT03359148 PMID:29547638

  11. Lung volume, breathing pattern and ventilation inhomogeneity in preterm and term infants.

    PubMed

    Latzin, Philipp; Roth, Stefan; Thamrin, Cindy; Hutten, Gerard J; Pramana, Isabelle; Kuehni, Claudia E; Casaulta, Carmen; Nelle, Matthias; Riedel, Thomas; Frey, Urs

    2009-01-01

    Morphological changes in preterm infants with bronchopulmonary dysplasia (BPD) have functional consequences on lung volume, ventilation inhomogeneity and respiratory mechanics. Although some studies have shown lower lung volumes and increased ventilation inhomogeneity in BPD infants, conflicting results exist possibly due to differences in sedation and measurement techniques. We studied 127 infants with BPD, 58 preterm infants without BPD and 239 healthy term-born infants, at a matched post-conceptional age of 44 weeks during quiet natural sleep according to ATS/ERS standards. Lung function parameters measured were functional residual capacity (FRC) and ventilation inhomogeneity by multiple breath washout as well as tidal breathing parameters. Preterm infants with BPD had only marginally lower FRC (21.4 mL/kg) than preterm infants without BPD (23.4 mL/kg) and term-born infants (22.6 mL/kg), though there was no trend with disease severity. They also showed higher respiratory rates and lower ratios of time to peak expiratory flow and expiratory time (t(PTEF)/t(E)) than healthy preterm and term controls. These changes were related to disease severity. No differences were found for ventilation inhomogeneity. Our results suggest that preterm infants with BPD have a high capacity to maintain functional lung volume during natural sleep. The alterations in breathing pattern with disease severity may reflect presence of adaptive mechanisms to cope with the disease process.

  12. Automated ventilator testing.

    PubMed

    Ghaly, J; Smith, A L

    1994-06-01

    A new era has arrived for the Biomedical Engineering Department at the Royal Women's Hospital in Melbourne. We have developed a system to qualitatively test for intermittent or unconfirmed faults, associated with Bear Cub ventilators. Where previous testing has been inadequate, computer logging is now used to interface the RT200 Timeter Calibration Analyser (TCA) to obtain a real time display of data, which can be stored and graphed. Using Quick Basic version 4.5, it was possible to establish communication between the TCA and an IBM compatible computer, such that meaningful displays of machine performance were produced. From the parameters measured it has been possible to obtain data on Peak Pressure, Inspiratory to Expiratory ratio (I:E ratio) Peak Flow and Rate. Monitoring is not limited to these parameters, though these were selected for our particular needs. These parameters are plotted in two ways: 1. Compressed average versus time, up to 24 hours on one screen 2. Raw data, 36 minutes displayed on each screen. The compressed data gives an overview which allows easy identification of intermittent faults. The uncompressed data confirms that the averaged signal is a realistic representation of the situation. One of the major benefits of this type of data analysis, is that ventilator performance may be monitored over a long period of time without requiring the presence of a service technician. It also allows individual ventilator performance to be graphically compared to other ventilators.

  13. Evaluation of a flexible bronchoscope prototype designed for bronchoscopy during mechanical ventilation: a proof-of-concept study.

    PubMed

    Nay, M-A; Auvet, A; Mankikian, J; Herve, V; Dequin, P-F; Guillon, A

    2017-06-01

    Bronchoscopy during mechanical ventilation of patients' lungs significantly affects ventilation because of partial obstruction of the tracheal tube, and may thus be omitted in the most severely ill patients. It has not previously been possible to reduce the external diameter of the bronchoscope without reducing the diameter of the suction channel, thus reducing the suctioning capacity of the device. We believed that a better-designed bronchoscope could improve the safety of bronchoscopy in patients whose lungs were ventilated. We designed a flexible bronchoscope prototype with a drumstick-shaped head consisting of a long, thin proximal portion; a short and large distal portion for camera docking; and a large suction channel throughout the length of the device. The aims of our study were to test the impact of our prototype on mechanical ventilation when inserted into the tracheal tube, and to assess suctioning capacity. We first tested the efficiency of the suction channel, and demonstrated that the suction flow of the prototype was similar to that of conventional adult bronchoscopes. We next evaluated the consequences of bronchoscopy when using the prototype on minute ventilation and intrathoracic pressures during mechanical ventilation: firstly, in vitro using a breathing simulator; and secondly, in vivo using a porcine model of pulmonary ventilation. The insertion of adult bronchoscopes into the tracheal tube immediately impaired the protective ventilation strategy employed, whereas the prototype preserved it. For the first time, we have developed an innovative flexible bronchoscope designed for bronchoscopy during invasive mechanical ventilation, that both preserved the protective ventilation strategy, and enabled efficient suction flow. © 2017 The Association of Anaesthetists of Great Britain and Ireland.

  14. The impact of a ventilator bundle on preventing ventilator-associated pneumonia: a multicenter study.

    PubMed

    Eom, Joong Sik; Lee, Mi-Suk; Chun, Hee-Kyung; Choi, Hee Jung; Jung, Sun-Young; Kim, Yeon-Sook; Yoon, Seon Jin; Kwak, Yee Gyung; Oh, Gang-Bok; Jeon, Min-Hyok; Park, Sun-Young; Koo, Hyun-Sook; Ju, Young-Su; Lee, Jin Seo

    2014-01-01

    For prevention of ventilator-associated pneumonia (VAP), a bundle approach was applied to patients receiving mechanical ventilation in intensive care units. The incidence of VAP and the preventive efficacy of the VAP bundle were investigated. A quasi-experimental study was conducted in adult intensive care units of 6 university hospitals with similar VAP rates. We implemented the VAP bundle between March 2011 and June 2011, then compared the rate of VAP after implementation of the VAP bundle with the rate in the previous 8 months. Our ventilator bundle included head of bed elevation, peptic ulcer disease prophylaxis, deep venous thrombosis prophylaxis, and oral decontamination with chlorhexidine 0.12%. Continuous aspiration of subglottic secretions was an option. Implementation of the VAP bundle reduced the VAP rate from a mean of 4.08 cases per 1,000 ventilator-days to 1.16 cases per 1,000 ventilator-days. The incidence density ratio (rate) was 0.28 (95% confidence interval, 0.275-0.292). Implementing the appropriate VAP bundle significantly decreased the incidence of VAP in patients with mechanical ventilation. Copyright © 2014 Association for Professionals in Infection Control and Epidemiology, Inc. Published by Mosby, Inc. All rights reserved.

  15. An injection and mixing element for delivery and monitoring of inhaled nitric oxide.

    PubMed

    Martin, Andrew R; Jackson, Chris; Fromont, Samuel; Pont, Chloe; Katz, Ira M; Caillobotte, Georges

    2016-08-30

    Inhaled nitric oxide (NO) is a selective pulmonary vasodilator used primarily in the critical care setting for patients concurrently supported by invasive or noninvasive positive pressure ventilation. NO delivery devices interface with ventilator breathing circuits to inject NO in proportion with the flow of air/oxygen through the circuit, in order to maintain a constant, target concentration of inhaled NO. In the present article, a NO injection and mixing element is presented. The device borrows from the design of static elements to promote rapid mixing of injected NO-containing gas with breathing circuit gases. Bench experiments are reported to demonstrate the improved mixing afforded by the injection and mixing element, as compared with conventional breathing circuit adapters, for NO injection into breathing circuits. Computational fluid dynamics simulations are also presented to illustrate mixing patterns and nitrogen dioxide production within the element. Over the range of air flow rates and target NO concentrations investigated, mixing length, defined as the downstream distance required for NO concentration to reach within ±5 % of the target concentration, was as high as 47 cm for the conventional breathing circuit adapters, but did not exceed 7.8 cm for the injection and mixing element. The injection and mixing element has potential to improve ease of use, compatibility and safety of inhaled NO administration with mechanical ventilators and gas delivery devices.

  16. On the Single-Layer Hydraulics Model for Flows and Ventilation over Unban Areas in Stable Stratification

    NASA Astrophysics Data System (ADS)

    Liu, C. H.

    2015-12-01

    Atmospheric stability has substantial effects on the flows and heat/mass transport processes. While extensive studies have been conducted for neutral and unstable stabilities, rather limited studies have been devoted to stable stratification. Major technical reason is the demanding spatio-temporal resolution required to solve the small scales in stratified turbulent flows. Instead of continuous density variation, we use the single-layer hydraulics model (analogous to shallow water equations for global dynamics), to simulate the stratified flows and turbulence structure over hypothetical urban areas. An array of identical ribs in cross flows is used to model an idealized urban surface and the aerodynamic resistance is controlled by the separation among the ribs. Two immiscible fluids (water and air) with a large density difference (three order of magnitude) are used to simulate the stratification. The key assumption is that the density in the (lower) single layer is uniform. As a result, the stratification is measured by the Froude number Fr (= U/(gH)1/2; where U is the flow speed, g the gravitational acceleration and H the single-layer depth). One of the characteristics of single-layer hydraulics model is hydraulic jump which occurs when the flows are slowing down from Fr > 1 (high-speed flows over smoother surfaces) to Fr < 1 (lower-speed flows over rougher surfaces). It is noteworthy that kinetic energy does not conserve across hydraulic jump that, unavoidably, cascades to turbulent kinetic energy (TKE). We thus hypotheses that the elevated TKE could modify the street-level ventilation mechanism in the stratified flows across an abrupt change in surface roughness entering urban areas. Large-eddy simulation and laboratory-scale water channel experiments are sought to improve our understanding of the occurrence of hydraulic jump and the associated street-level ventilation mechanism in the stratified flows over urban areas. Preliminary results, by comparing the dynamics at Fr = 2.4 and Fr = 2.8, demonstrate the notable changes in ventilation performance in the first several rows of ribs of urban areas. Substantial changes in the mean and fluctuating velocities are observed that contribute to the different street-level ventilation mechanism. Detailed results will be reported in the upcoming AGU fall meeting.

  17. Sustained Reduction of Ventilator-Associated Pneumonia Rates Using Real-Time Course Correction With a Ventilator Bundle Compliance Dashboard.

    PubMed

    Talbot, Thomas R; Carr, Devin; Parmley, C Lee; Martin, Barbara J; Gray, Barbara; Ambrose, Anna; Starmer, Jack

    2015-11-01

    The effectiveness of practice bundles on reducing ventilator-associated pneumonia (VAP) has been questioned. To implement a comprehensive program that included a real-time bundle compliance dashboard to improve compliance and reduce ventilator-associated complications. DESIGN Before-and-after quasi-experimental study with interrupted time-series analysis. SETTING Academic medical center. In 2007 a comprehensive institutional ventilator bundle program was developed. To assess bundle compliance and stimulate instant course correction of noncompliant parameters, a real-time computerized dashboard was developed. Program impact in 6 adult intensive care units (ICUs) was assessed. Bundle compliance was noted as an overall cumulative bundle adherence assessment, reflecting the percentage of time all elements were concurrently in compliance for all patients. The VAP rate in all ICUs combined decreased from 19.5 to 9.2 VAPs per 1,000 ventilator-days following program implementation (P<.001). Bundle compliance significantly increased (Z100 score of 23% in August 2007 to 83% in June 2011 [P<.001]). The implementation resulted in a significant monthly decrease in the overall ICU VAP rate of 3.28/1,000 ventilator-days (95% CI, 2.64-3.92/1,000 ventilator-days). Following the intervention, the VAP rate decreased significantly at a rate of 0.20/1,000 ventilator-days per month (95% CI, 0.14-0.30/1,000 ventilator-days per month). Among all adult ICUs combined, improved bundle compliance was moderately correlated with monthly VAP rate reductions (Pearson correlation coefficient, -0.32). A prevention program using a real-time bundle adherence dashboard was associated with significant sustained decreases in VAP rates and an increase in bundle compliance among adult ICU patients.

  18. Late outcomes of a randomized trial of high-frequency oscillation in neonates.

    PubMed

    Zivanovic, Sanja; Peacock, Janet; Alcazar-Paris, Mireia; Lo, Jessica W; Lunt, Alan; Marlow, Neil; Calvert, Sandy; Greenough, Anne

    2014-03-20

    Results from an observational study involving neonates suggested that high-frequency oscillatory ventilation (HFOV), as compared with conventional ventilation, was associated with superior small-airway function at follow-up. Data from randomized trials are needed to confirm this finding. We studied 319 adolescents who had been born before 29 weeks of gestation and had been enrolled in a multicenter, randomized trial that compared HFOV with conventional ventilation immediately after birth. The trial involved 797 neonates, of whom 592 survived to hospital discharge. We compared follow-up data from adolescents who had been randomly assigned to HFOV with follow-up data from those who had been randomly assigned to conventional ventilation, with respect to lung function and respiratory health, health-related quality of life, and functional status, as assessed with the use of questionnaires completed when the participants were 11 to 14 years of age. The primary outcome was forced expiratory flow at 75% of the expired vital capacity (FEF75). The HFOV group had superior results on a test of small-airway function (z score for FEF75, -0.97 with HFOV vs. -1.19 with conventional therapy; adjusted difference, 0.23 [95% confidence interval, 0.02 to 0.45]). There were significant differences in favor of HFOV in several other measures of respiratory function, including forced expiratory volume in 1 second, forced vital capacity, peak expiratory flow, diffusing capacity, and impulse-oscillometric findings. As compared with the conventional-therapy group, the HFOV group had significantly higher ratings from teachers in three of eight school subjects assessed, but there were no other significant differences in functional outcomes. In a randomized trial involving children who had been born extremely prematurely, those who had undergone HFOV, as compared with those who had received conventional ventilation, had superior lung function at 11 to 14 years of age, with no evidence of poorer functional outcomes. (Funded by the National Institute for Health Research Health Technology Assessment Programme and others.).

  19. Clinical assessment of auto-positive end-expiratory pressure by diaphragmatic electrical activity during pressure support and neurally adjusted ventilatory assist.

    PubMed

    Bellani, Giacomo; Coppadoro, Andrea; Patroniti, Nicolò; Turella, Marta; Arrigoni Marocco, Stefano; Grasselli, Giacomo; Mauri, Tommaso; Pesenti, Antonio

    2014-09-01

    Auto-positive end-expiratory pressure (auto-PEEP) may substantially increase the inspiratory effort during assisted mechanical ventilation. Purpose of this study was to assess whether the electrical activity of the diaphragm (EAdi) signal can be reliably used to estimate auto-PEEP in patients undergoing pressure support ventilation and neurally adjusted ventilatory assist (NAVA) and whether NAVA was beneficial in comparison with pressure support ventilation in patients affected by auto-PEEP. In 10 patients with a clinical suspicion of auto-PEEP, the authors simultaneously recorded EAdi, airway, esophageal pressure, and flow during pressure support and NAVA, whereas external PEEP was increased from 2 to 14 cm H2O. Tracings were analyzed to measure apparent "dynamic" auto-PEEP (decrease in esophageal pressure to generate inspiratory flow), auto-EAdi (EAdi value at the onset of inspiratory flow), and IDEAdi (inspiratory delay between the onset of EAdi and the inspiratory flow). The pressure necessary to overcome auto-PEEP, auto-EAdi, and IDEAdi was significantly lower in NAVA as compared with pressure support ventilation, decreased with increase in external PEEP, although the effect of external PEEP was less pronounced in NAVA. Both auto-EAdi and IDEAdi were tightly correlated with auto-PEEP (r = 0.94 and r = 0.75, respectively). In the presence of auto-PEEP at lower external PEEP levels, NAVA was characterized by a characteristic shape of the airway pressure. In patients with auto-PEEP, NAVA, compared with pressure support ventilation, led to a decrease in the pressure necessary to overcome auto-PEEP, which could be reliably monitored by the electrical activity of the diaphragm before inspiratory flow onset (auto-EAdi).

  20. Flow and volume dependence of rat airway resistance during constant flow inflation and deflation.

    PubMed

    Rubini, Alessandro; Carniel, Emanuele Luigi; Parmagnani, Andrea; Natali, Arturo Nicola

    2011-12-01

    The aim of this study was to measure the flow and volume dependence of both the ohmic and the viscoelastic pressure dissipations of the normal rat respiratory system separately during inflation and deflation. The study was conducted in the Respiratory Physiology Laboratory in our institution. Measurements were obtained for Seven albino Wistar rats of both sexes by using the flow interruption method during constant flow inflations and deflations. Measurements included anesthesia induction, tracheostomy and positioning of a tracheal cannula, positive pressure ventilation, constant flow respiratory system inflations and deflations at two different volumes and flows. The ohmic resistance exhibited volume and flow dependence, decreasing with lung volume and increasing with flow rate, during both inflation and deflation. The stress relaxation-related viscoelastic resistance also exhibited volume and flow dependence. It decreased with the flow rate at a constant lung volume during both inflation and deflation, but exhibited a different behavior with the lung volume at a constant flow rate (i.e., increased during inflations and decreased during deflations). Thus, stress relaxation in the rat lungs exhibited a hysteretic behavior. The observed flow and volume dependence of respiratory system resistance may be predicted by an equation derived from a model of the respiratory system that consists of two distinct compartments. The equation agrees well with the experimental data and indicates that the loading time is the critical parameter on which stress relaxation depends, during both lung inflation and deflation.

  1. Measurements of evaporated perfluorocarbon during partial liquid ventilation by a zeolite absorber.

    PubMed

    Proquitté, Hans; Rüdiger, Mario; Wauer, Roland R; Schmalisch, Gerd

    2004-01-01

    During partial liquid ventilation (PLV) the knowledge of the quantity of exhaled perfluorocarbon (PFC) allows a continuous substitution of the PFC loss to achieve a constant PFC level in the lungs. The aim of our in vitro study was to determine the PFC loss in the mixed expired gas by an absorber and to investigate the effect of the evaporated PFC on ventilatory measurements. To simulate the PFC loss during PLV, a heated flask was rinsed with a constant airflow of 4 L min(-1) and PFC was infused by different speeds (5, 10, 20 mL h(-1)). An absorber filled with PFC selective zeolites was connected with the flask to measure the PFC in the gas. The evaporated PFC volume and the PFC concentration were determined from the weight gain of the absorber measured by an electronic scale. The PFC-dependent volume error of the CO2SMO plus neonatal pneumotachograph was measured by manual movements of a syringe with volumes of 10 and 28 mL with a rate of 30 min(-1). Under steady state conditions there was a strong correlation (r2 = 0.999) between the infusion speed of PFC and the calculated PFC flow rate. The PFC flow rate was slightly underestimated by 4.3% (p < 0.01). However, this bias was independent from PFC infusion rate. The evaporated PFC volume was precisely measured with errors < 1%. The volume error of the CO2SMO-Plus pneumotachograph increased with increasing PFC content for both tidal volumes (p < 0.01). However for PFC flow rates up to 20 mL/h the error of the measured tidal volumes was < 5%. PFC selective zeolites can be used to quantify accurately the evaporated PFC volume during PLV. With increasing PFC concentrations in the exhaled air the measurement errors of ventilatory parameters have to be taken into account.

  2. Numerical analysis of natural ventilation system in a studio apartment in Bangladesh

    NASA Astrophysics Data System (ADS)

    Kabir, K. M. Ariful; Hasan, Md. Rakibul; Khan, Md. Abdul Hakim

    2017-07-01

    The study of temperature and air flow for natural ventilation system has been investigated numerically. A finite element model for studio apartment was developed with the aim of achieving detail energy allocation in the real buildings during the transient process in the walls and internal air. A tool of computational fluid dynamics (CFD) is employed to assist the process. In the tropical regions most of the energy is consumed by the heating, cooling and ventilation appliances. Therefore, the optimize ventilation system will be a suitable and valid option for the saving of energy from the household sector to increase cooling performance and ensuring thermal comfort as well. A mathematical exploration is carried out on full scale dwelling and small scale model and indication is given on the relevance of such a comparison. Calculations are carried out with household heat sources for calm and windy period, but without any human. As expected, for windy periods, the wind is the main driving force behind the internal air flow. However, in calm periods for unsteady flow the internal airflow looks like more complexes through observation.

  3. Occupant Interactions and Effectiveness of Natural Ventilation Strategies in Contemporary New Housing in Scotland, UK.

    PubMed

    Sharpe, Tim; Farren, Paul; Howieson, Stirling; Tuohy, Paul; McQuillan, Jonathan

    2015-07-21

    The need to reduce carbon emissions and fuel poverty has led to increased building envelope air tightness, intended to reduce uncontrolled ventilation heat losses. Ventilation strategies in dwellings still allow the use of trickle ventilators in window frames for background ventilation. The extent to which this results in "healthy" Indoor Air Quality (IAQ) in recently constructed dwellings was a concern of regulators in Scotland. This paper describes research to explore this. First a review of literature was conducted, then data on occupant interactions with ventilation provisions (windows, doors, trickle vents) gathered through an interview-based survey of 200 recently constructed dwellings, and measurements made on a sample of 40 of these. The main measured parameter discussed here is CO2 concentration. It was concluded after the literature review that 1000 ppm absolute was a reasonable threshold to use for "adequate" ventilation. The occupant survey found that there was very little occupant interaction with the trickle ventilators e.g., in bedrooms 63% were always closed, 28% always open, and in only 9% of cases occupants intervened to make occasional adjustments. In the measured dwellings average bedroom CO2 levels of 1520 ppm during occupied (night time) hours were observed. Where windows were open the average bedroom CO2 levels were 972 ppm. With windows closed, the combination of "trickle ventilators open plus doors open" gave an average of 1021 ppm. "Trickle ventilators open" gave an average of 1571 ppm. All other combinations gave averages of 1550 to 2000 ppm. Ventilation rates and air change rates were estimated from measured CO2 levels, for all dwellings calculated ventilation rate was less than 8 L/s/p, in 42% of cases calculated air change rate was less than 0.5 ach. It was concluded that trickle ventilation as installed and used is ineffective in meeting desired ventilation rates, evidenced by high CO2 levels reported across the sampled dwellings. Potential implications of the results are discussed.

  4. Code System to Calculate Tornado-Induced Flow Material Transport.

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    ANDRAE, R. W.

    1999-11-18

    Version: 00 TORAC models tornado-induced flows, pressures, and material transport within structures. Its use is directed toward nuclear fuel cycle facilities and their primary release pathway, the ventilation system. However, it is applicable to other structures and can model other airflow pathways within a facility. In a nuclear facility, this network system could include process cells, canyons, laboratory offices, corridors, and offgas systems. TORAC predicts flow through a network system that also includes ventilation system components such as filters, dampers, ducts, and blowers. These ventilation system components are connected to the rooms and corridors of the facility to form amore » complete network for moving air through the structure and, perhaps, maintaining pressure levels in certain areas. The material transport capability in TORAC is very basic and includes convection, depletion, entrainment, and filtration of material.« less

  5. 46 CFR 153.316 - Special cargo pumproom ventilation rate.

    Code of Federal Regulations, 2013 CFR

    2013-10-01

    ... CARGOES SHIPS CARRYING BULK LIQUID, LIQUEFIED GAS, OR COMPRESSED GAS HAZARDOUS MATERIALS Design and Equipment Cargo Handling Space Ventilation § 153.316 Special cargo pumproom ventilation rate. When Table 1...

  6. 46 CFR 153.316 - Special cargo pumproom ventilation rate.

    Code of Federal Regulations, 2012 CFR

    2012-10-01

    ... CARGOES SHIPS CARRYING BULK LIQUID, LIQUEFIED GAS, OR COMPRESSED GAS HAZARDOUS MATERIALS Design and Equipment Cargo Handling Space Ventilation § 153.316 Special cargo pumproom ventilation rate. When Table 1...

  7. 46 CFR 153.316 - Special cargo pumproom ventilation rate.

    Code of Federal Regulations, 2011 CFR

    2011-10-01

    ... CARGOES SHIPS CARRYING BULK LIQUID, LIQUEFIED GAS, OR COMPRESSED GAS HAZARDOUS MATERIALS Design and Equipment Cargo Handling Space Ventilation § 153.316 Special cargo pumproom ventilation rate. When Table 1...

  8. 46 CFR 153.316 - Special cargo pumproom ventilation rate.

    Code of Federal Regulations, 2014 CFR

    2014-10-01

    ... CARGOES SHIPS CARRYING BULK LIQUID, LIQUEFIED GAS, OR COMPRESSED GAS HAZARDOUS MATERIALS Design and Equipment Cargo Handling Space Ventilation § 153.316 Special cargo pumproom ventilation rate. When Table 1...

  9. 46 CFR 153.316 - Special cargo pumproom ventilation rate.

    Code of Federal Regulations, 2010 CFR

    2010-10-01

    ... CARGOES SHIPS CARRYING BULK LIQUID, LIQUEFIED GAS, OR COMPRESSED GAS HAZARDOUS MATERIALS Design and Equipment Cargo Handling Space Ventilation § 153.316 Special cargo pumproom ventilation rate. When Table 1...

  10. Can High-flow Nasal Cannula Reduce the Rate of Endotracheal Intubation in Adult Patients With Acute Respiratory Failure Compared With Conventional Oxygen Therapy and Noninvasive Positive Pressure Ventilation?: A Systematic Review and Meta-analysis.

    PubMed

    Ni, Yue-Nan; Luo, Jian; Yu, He; Liu, Dan; Ni, Zhong; Cheng, Jiangli; Liang, Bin-Miao; Liang, Zong-An

    2017-04-01

    The effects of high-flow nasal cannula (HFNC) on adult patients with acute respiratory failure (ARF) are controversial. We aimed to further determine the effectiveness of HFNC in reducing the rate of endotracheal intubation in adult patients with ARF by comparison to noninvasive positive pressure ventilation (NIPPV) and conventional oxygen therapy (COT). The PubMed, Embase, Medline, and the Cochrane Central Register of Controlled Trials databases, as well as the Information Sciences Institute Web of Science, were searched for all controlled studies that compared HFNC with NIPPV and COT in adult patients with ARF. The primary outcome was the rate of endotracheal intubation; the secondary outcomes were ICU mortality and length of ICU stay. Eighteen trials with a total of 3,881 patients were pooled in our final studies. Except for ICU mortality (I 2  = 67%, χ 2  = 12.21, P = .02) and rate of endotracheal intubation (I 2  = 63%, χ 2  = 13.51, P = .02) between HFNC and NIPPV, no significant heterogeneity was found in outcome measures. Compared with COT, HFNC was associated with a lower rate of endotracheal intubation (z = 2.55, P = .01) while no significant difference was found in the comparison with NIPPV (z = 1.40, P = .16). As for ICU mortality and length of ICU stay, HFNC did not exhibit any advantage over either COT or NIPPV. In patients with ARF, HFNC is a more reliable alternative than NIPPV to reduce the rate of endotracheal intubation than COT. Copyright © 2017 American College of Chest Physicians. Published by Elsevier Inc. All rights reserved.

  11. High-frequency oscillatory ventilation (HFOV) in the treatment of neonatal respiratory disturbances: case reports of two infants.

    PubMed

    Kiszel, J; Seri, I; Machay, T

    1985-01-01

    The technique of high-frequency oscillatory ventilation (HFOV) was successfully used in a preterm infant with severe hyaline membrane disease and in a term neonate presenting with intrauterine pneumonia and associated severe pneumomediastinum. None of the infants could adequately be ventilated by conventional ventilation; both of them deteriorated owing to severe hypoxaemia and hypercapnia. In the preterm infant with HMD a rapid and progressive improvement of oxygenation had been observed immediately after the beginning of HFOV, and he was successfully weaned off the ventilator after 71 hours on HFOV. His recovery was uncomplicated and definitive. In the term neonate presenting with IUP and associated severe PM, an improvement in oxygenation was detected, whereas the retention of paCO2 remained unaltered. On leaving the MAP unchanged but doubling the flow rate, paCO2 and arterial pH also normalised. No sign of PM was seen on the X-ray picture 17.5 hours after the start of HFOV. This patient was weaned off the ventilator after 29 hours on HFOV and his recovery was also uncomplicated. It is believed that recovery of the PM was secondary to the low MAP and to the higher arterial pO2 levels, and that HFOV may also have a direct role in the treatment of preexisting air leaks and perhaps also in their prevention. In our patients HFOV resulted in a definitive recovery, while no improvement had occurred on using conventional ventilation. To determine the exact mechanism of action, the clear cut fields of indications and the possible side effects of HFOV, further investigations are needed.

  12. Hard metal exposures. Part 1: Observed performance of three local exhaust ventilation systems.

    PubMed

    Guffey, S E; Simcox, N; Booth, D W; Hibbard, R; Stebbins, A

    2000-04-01

    Not every ventilation system performs as intended; much can be learned when they do not. The purpose of this study was to compare observed initial performance to expected levels for three saw-reconditioning shop ventilation systems and to characterize the changes in performance of the systems over a one-year period. These three local exhaust ventilation systems were intended to control worker exposures to cobalt, cadmium, and chromium during wet grinding, dry grinding, and welding/brazing activities. Prior to installation the authors provided some design guidance based on Industrial Ventilation, a Manual of Recommended Practice. However, the authors had limited influence on the actual installation and operation and no line authority for the systems. In apparent efforts to cut costs and to respond to other perceived needs, the installed systems deviated from the specifications used in pressure calculations in many important aspects, including adding branch ducts, use of flexible ducts, the choice of fans, and the construction of some hoods. After installation of the three systems, ventilation measurements were taken to determine if the systems met design specifications, and worker exposures were measured to determine effectiveness. The results of the latter will be published as a companion article. The deviations from design and maintenance failures may have adversely affected performance. From the beginning to the end of the study period the distribution of air flow never matched the design specifications for the systems. The observed air flows measured within the first month of installation did not match the predicated design air flows for any of the systems, probably because of the differences between the design and the installed system. Over the first year of operation, hood air flow variability was high due to inadequate cleaning of the sticky process materials which rapidly accumulated in the branch ducts. Poor distribution of air flows among branch ducts frequently produced individual hood air flows that were far below specified design levels even when the total air flow through that system was more than adequate. To experienced practitioners, it is not surprising that deviations from design recommendations and poor maintenance would be associated with poor system performance. Although commonplace, such experiences have not been documented in peer-reviewed publications to date. This publication is a first step in providing that documentation.

  13. Mask Ventilation during Induction of General Anesthesia: Influences of Obstructive Sleep Apnea.

    PubMed

    Sato, Shin; Hasegawa, Makoto; Okuyama, Megumi; Okazaki, Junko; Kitamura, Yuji; Sato, Yumi; Ishikawa, Teruhiko; Sato, Yasunori; Isono, Shiroh

    2017-01-01

    Depending on upper airway patency during anesthesia induction, tidal volume achieved by mask ventilation may vary. In 80 adult patients undergoing general anesthesia, the authors tested a hypothesis that tidal volume during mask ventilation is smaller in patients with sleep-disordered breathing priorly defined as apnea hypopnea index greater than 5 per hour. One-hand mask ventilation with a constant ventilator setting (pressure-controlled ventilation) was started 20 s after injection of rocuronium and maintained for 1 min during anesthesia induction. Mask ventilation efficiency was assessed by the breath number needed to initially exceed 5 ml/kg ideal body weight of expiratory tidal volume (primary outcome) and tidal volumes (secondary outcomes) during initial 15 breaths (UMIN000012494). Tidal volume progressively increased by more than 70% in 1 min and did not differ between sleep-disordered breathing (n = 42) and non-sleep-disordered breathing (n = 38) patients. In post hoc subgroup analyses, the primary outcome breath number (mean [95% CI], 5.7 [4.1 to 7.3] vs. 1.7 [0.2 to 3.2] breath; P = 0.001) and mean tidal volume (6.5 [4.6 to 8.3] vs. 9.6 [7.7 to 11.4] ml/kg ideal body weight; P = 0.032) were significantly smaller in 20 sleep-disordered breathing patients with higher apnea hypopnea index (median [25th to 75th percentile]: 21.7 [17.6 to 31] per hour) than in 20 non-sleep disordered breathing subjects with lower apnea hypopnea index (1.0 [0.3 to 1.5] per hour). Obesity and occurrence of expiratory flow limitation during one-hand mask ventilation independently explained the reduction of efficiency of mask ventilation, while the use of two hands effectively normalized inefficient mask ventilation during one-hand mask ventilation. One-hand mask ventilation is difficult in patients with obesity and severe sleep-disordered breathing particularly when expiratory flow limitation occurs during mask ventilation.

  14. Validating computational predictions of night-time ventilation in Stanford's Y2E2 building

    NASA Astrophysics Data System (ADS)

    Chen, Chen; Lamberti, Giacomo; Gorle, Catherine

    2017-11-01

    Natural ventilation can significantly reduce building energy consumption, but robust design is a challenging task. We previously presented predictions of natural ventilation performance in Stanford's Y2E2 building using two models with different levels of fidelity, embedded in an uncertainty quantification framework to identify the dominant uncertain parameters and predict quantified confidence intervals. The results showed a slightly high cooling rate for the volume-averaged temperature, and the initial thermal mass temperature and window discharge coefficients were found to have an important influence on the results. To further investigate the potential role of these parameters on the observed discrepancies, the current study is performing additional measurements in the Y2E2 building. Wall temperatures are recorded throughout the nightflush using thermocouples; flow rates through windows are measured using hotwires; and spatial variability in the air temperature is explored. The measured wall temperatures are found the be within the range of our model assumptions, and the measured velocities agree reasonably well with our CFD predications. Considerable local variations in the indoor air temperature have been recorded, largely explaining the discrepancies in our earlier validation study. Future work will therefore focus on a local validation of the CFD results with the measurements. Center for Integrated Facility Engineering (CIFE).

  15. Effect of frequency of ventilator circuit changes (3 vs 7 days) on the rate of ventilator-associated pneumonia in PICU.

    PubMed

    Samransamruajkit, Rujipat; Jirapaiboonsuk, Suree; Siritantiwat, Sirirush; Tungsrijitdee, Ornanong; Deerojanawong, Jitladda; Sritippayawan, Suchada; Prapphal, Nuanchan

    2010-03-01

    Ventilator-associated pneumonia (VAP) is associated with significant morbidity and mortality in pediatric intensive care unit (PICU). Our purpose was to evaluate the effects of ventilator circuit change on the rate of VAP in the PICU. A prospective randomized controlled trial was conducted at a university hospital PICU. Children (younger than 18 years) who received mechanical ventilation from December 2006 to November 2007 were randomly assigned to receive ventilator circuit changes every 3 or 7 days. Of 176 patients, 88 were assigned to receive ventilator circuit every 3 days and 88 patients had a change weekly. The rate of VAP was 13.9/1000 ventilator days for the 3-day circuit change (n = 12) vs 11.5/1000 ventilator days (n = 10) for the 7-day circuit change (odds ratio, 0.8; confidence interval, 0.3-1.9; P = .6). There was a trend toward decreased PICU stay and mortality rate in 7-day change group compared to 3-day change group but did not reach statistical significance. Furthermore, switching from a 3-day to a 7-day change policy could save costs up to US $22,000/y. The 7-day ventilator circuit change did not contribute to increased rates of VAP in our PICU. Thus, it may be used as a guide to save workload and supply costs. Crown Copyright 2010. Published by Elsevier Inc. All rights reserved.

  16. Results of the Trace Contaminant Control Needs Evaluation and Sizing Study for Space Suit Life Support Development

    NASA Technical Reports Server (NTRS)

    Paul, Heather L.; Jennings, Mallory A.

    2009-01-01

    The Trace Contaminant Control System (TCCS), located within the ventilation loop of the Portable Life Support System (PLSS) of the Constellation Space Suit Element (CSSE), is responsible for removing hazardous trace contaminants from the space suit ventilation flow. This paper summarizes the results of a trade study that evaluated if trace contaminant control could be accomplished without a TCCS, relying on suit leakage, ullage loss from the carbon dioxide and humidity control system, and other factors. Trace contaminant generation rates were revisited to verify that values reflect the latest designs for CSSE pressure garment materials and PLSS hardware. Additionally, TCCS sizing calculations were performed and a literature survey was conducted to review the latest developments in trace contaminant technologies.

  17. A simple method for isocapnic hyperventilation evaluated in a lung model.

    PubMed

    Hallén, K; Stenqvist, O; Ricksten, S-E; Lindgren, S

    2016-05-01

    Isocapnic hyperventilation (IHV) has the potential to increase the elimination rate of anaesthetic gases and has been shown to shorten time to wake-up and post-operative recovery time after inhalation anaesthesia. In this bench test, we describe a technique to achieve isocapnia during hyperventilation (HV) by adding carbon dioxide (CO2) directly to the breathing circuit of a standard anaesthesia apparatus with standard monitoring equipment. Into a mechanical lung model, carbon dioxide was added to simulate a CO2 production (V(CO2)) of 175, 200 and 225 ml/min. Dead space (V(D)) volume could be set at 44, 92 and 134 ml. From baseline ventilation (BLV), HV was achieved by doubling the minute ventilation and fresh gas flow for each level of V(CO2), and dead space. During HV, CO2 was delivered (D(CO2)) by a precision flow meter via a mixing box to the inspiratory limb of the anaesthesia circuit to achieve isocapnia. During HV, the alveolar ventilation increased by 113 ± 6%. Tidal volume increased by 20 ± 0.1% during IHV irrespective of V(D) and V(CO2) level. D(CO2) varied between 147 ± 8 and 325 ± 13 ml/min. Low V(CO2) and large V(D) demanded a greater D(CO2) administration to achieve isocapnia. The FICO2 level during IHV varied between 2.3% and 3.3%. It is possible to maintain isocapnia during HV by delivering carbon dioxide through a standard anaesthesia circuit equipped with modern monitoring capacities. From alveolar ventilation, CO2 production and dead space, the amount of carbon dioxide that is needed to achieve IHV can be estimated. © 2015 The Acta Anaesthesiologica Scandinavica Foundation. Published by John Wiley & Sons Ltd.

  18. SU-F-J-219: Predicting Ventilation Change Due to Radiation Therapy: Dependency On Pre-RT Ventilation and Effort Correction

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Patton, T; Du, K; Bayouth, J

    Purpose: Ventilation change caused by radiation therapy (RT) can be predicted using four-dimensional computed tomography (4DCT) and image registration. This study tested the dependency of predicted post-RT ventilation on effort correction and pre-RT lung function. Methods: Pre-RT and 3 month post-RT 4DCT images were obtained for 13 patients. The 4DCT images were used to create ventilation maps using a deformable image registration based Jacobian expansion calculation. The post-RT ventilation maps were predicted in four different ways using the dose delivered, pre-RT ventilation, and effort correction. The pre-RT ventilation and effort correction were toggled to determine dependency. The four different predictedmore » ventilation maps were compared to the post-RT ventilation map calculated from image registration to establish the best prediction method. Gamma pass rates were used to compare the different maps with the criteria of 2mm distance-to-agreement and 6% ventilation difference. Paired t-tests of gamma pass rates were used to determine significant differences between the maps. Additional gamma pass rates were calculated using only voxels receiving over 20 Gy. Results: The predicted post-RT ventilation maps were in agreement with the actual post-RT maps in the following percentage of voxels averaged over all subjects: 71% with pre-RT ventilation and effort correction, 69% with no pre-RT ventilation and effort correction, 60% with pre-RT ventilation and no effort correction, and 58% with no pre-RT ventilation and no effort correction. When analyzing only voxels receiving over 20 Gy, the gamma pass rates were respectively 74%, 69%, 65%, and 55%. The prediction including both pre- RT ventilation and effort correction was the only prediction with significant improvement over using no prediction (p<0.02). Conclusion: Post-RT ventilation is best predicted using both pre-RT ventilation and effort correction. This is the only prediction that provided a significant improvement on agreement. Research support from NIH grants CA166119 and CA166703, a gift from Roger Koch, and a Pilot Grant from University of Iowa Carver College of Medicine.« less

  19. Recurrence quantification analysis of heart rate variability and respiratory flow series in patients on weaning trials.

    PubMed

    Arcentales, Andrés; Giraldo, Beatriz F; Caminal, Pere; Benito, Salvador; Voss, Andreas

    2011-01-01

    Autonomic nervous system regulates the behavior of cardiac and respiratory systems. Its assessment during the ventilator weaning can provide information about physio-pathological imbalances. This work proposes a non linear analysis of the complexity of the heart rate variability (HRV) and breathing duration (T(Tot)) applying recurrence plot (RP) and their interaction joint recurrence plot (JRP). A total of 131 patients on weaning trials from mechanical ventilation were analyzed: 92 patients with successful weaning (group S) and 39 patients that failed to maintain spontaneous breathing (group F). The results show that parameters as determinism (DET), average diagonal line length (L), and entropy (ENTR), are statistically significant with RP for T(Tot) series, but not with HRV. When comparing the groups with JRP, all parameters have been relevant. In all cases, mean values of recurrence quantification analysis are higher in the group S than in the group F. The main differences between groups were found on the diagonal and vertical structures of the joint recurrence plot.

  20. Intricate but tight coupling of spiracular activity and abdominal ventilation during locust discontinuous gas exchange cycles.

    PubMed

    Talal, Stav; Gefen, Eran; Ayali, Amir

    2018-03-15

    Discontinuous gas exchange (DGE) is the best studied among insect gas exchange patterns. DGE cycles comprise three phases, which are defined by their spiracular state: closed, flutter and open. However, spiracle status has rarely been monitored directly; rather, it is often assumed based on CO 2 emission traces. In this study, we directly recorded electromyogram (EMG) signals from the closer muscle of the second thoracic spiracle and from abdominal ventilation muscles in a fully intact locust during DGE. Muscular activity was monitored simultaneously with CO 2 emission, under normoxia and under various experimental oxic conditions. Our findings indicate that locust DGE does not correspond well with the commonly described three-phase cycle. We describe unique DGE-related ventilation motor patterns, coupled to spiracular activity. During the open phase, when CO 2 emission rate is highest, the thoracic spiracles do not remain open; rather, they open and close rapidly. This fast spiracle activity coincides with in-phase abdominal ventilation, while alternating with the abdominal spiracle and thus facilitating a unidirectional air flow along the main trachea. A change in the frequency of rhythmic ventilation during the open phase suggests modulation by intra-tracheal CO 2 levels. A second, slow ventilatory movement pattern probably serves to facilitate gas diffusion during spiracle closure. Two flutter-like patterns are described in association with the different types of ventilatory activity. We offer a modified mechanistic model for DGE in actively ventilating insects, incorporating ventilatory behavior and changes in spiracle state. © 2018. Published by The Company of Biologists Ltd.

  1. EVALUATION OF VENTILATION PERFORMANCE FOR INDOOR SPACE

    EPA Science Inventory

    The paper discusses a personal-computer-based application of computational fluid dynamics that can be used to determine the turbulent flow field and time-dependent/steady-state contaminant concentration distributions within isothermal indoor space. (NOTE: Ventilation performance ...

  2. Ventilation of carbon monoxide from a biomass pellet storage tank--a study of the effects of variation of temperature and cross-ventilation on the efficiency of natural ventilation.

    PubMed

    Emhofer, Waltraud; Lichtenegger, Klaus; Haslinger, Walter; Hofbauer, Hermann; Schmutzer-Roseneder, Irene; Aigenbauer, Stefan; Lienhard, Martin

    2015-01-01

    Wood pellets have been reported to emit toxic gaseous emissions during transport and storage. Carbon monoxide (CO) emission, due to the high toxicity of the gas and the possibility of it being present at high levels, is the most imminent threat to be considered before entering a pellet storage facility. For small-scale (<30 tons storage capacity) residential pellet storage facilities, ventilation, preferably natural ventilation utilizing already existing openings, has become the most favored solution to overcome the problem of high CO concentrations. However, there is little knowledge on the ventilation rates that can be reached and thus on the effectiveness of such measures. The aim of the study was to investigate ventilation rates for a specific small-scale pellet storage system depending on characteristic temperature differences. Furthermore, the influence of the implementation of a chimney and the influence of cross-ventilation on the ventilation rates were investigated. The air exchange rates observed in the experiments ranged between close to zero and up to 8 m(3) h(-1), depending largely on the existing temperature differences and the existence of cross-ventilation. The results demonstrate that implementing natural ventilation is a possible measure to enhance safety from CO emissions, but not one without limitations. © The Author 2014. Published by Oxford University Press on behalf of the British Occupational Hygiene Society.

  3. Effect of varying the pressurisation rate during noninvasive pressure support ventilation.

    PubMed

    Prinianakis, G; Delmastro, M; Carlucci, A; Ceriana, P; Nava, S

    2004-02-01

    The aim of the study was to assess the effects of varying the pressurisation rate during noninvasive pressure support ventilation on patients' breathing pattern, inspiratory effort, arterial blood gases, tolerance to ventilation and amount of air leakage. A total of 15 chronic obstructive pulmonary disease patients recovering from an acute episode of hypercapnic acute respiratory failure were studied during four randomised trials with different levels of pressurisation rate. No significant changes were observed in breathing pattern and arterial blood gases between the different runs. The pressure time product of the diaphragm, an estimate of its metabolic consumption, was significantly lower with all pressurisation rates than with spontaneous breathing, but was significantly lowest with the fastest rate. However, air leak, assessed by the ratio between expired and inspired tidal volumes, increased and the patients' tolerance of ventilation, measured using a standardised scale, was significantly poorer with the fastest pressurisation rate. In chronic obstructive pulmonary disease patients recovering from an episode of acute hypercapnic respiratory failure and ventilated with noninvasive pressure support ventilation, different pressurisation rates resulted in different reductions in the pressure time product of the diaphragm; this reduction was greater with the fastest rate, but was accompanied by significant air leaks and poor tolerance.

  4. Software for real-time control of a tidal liquid ventilator.

    PubMed

    Heckman, J L; Hoffman, J; Shaffer, T H; Wolfson, M R

    1999-01-01

    The purpose of this project was to develop and test computer software and control algorithms designed to operate a tidal liquid ventilator. The tests were executed on a 90-MHz Pentium PC with 16 MB RAM and a prototype liquid ventilator. The software was designed using Microsoft Visual C++ (Ver. 5.0) and the Microsoft Foundation Classes. It uses a graphic user interface, is multithreaded, runs in real time, and has a built-in simulator that facilitates user education in liquid-ventilation principles. The operator can use the software to specify ventilation parameters such as the frequency of ventilation, the tidal volume, and the inspiratory-expiratory time ratio. Commands are implemented via control of the pump speed and by setting the position of two two-way solenoid-controlled valves. Data for use in monitoring and control are gathered by analog-to-digital conversion. Control strategies are implemented to maintain lung volumes and airway pressures within desired ranges, according to limits set by the operator. Also, the software allows the operator to define the shape of the flow pulse during inspiration and expiration, and to optimize perfluorochemical liquid transfer while minimizing airway pressures and maintaining the desired tidal volume. The operator can stop flow during inspiration and expiration to measure alveolar pressures. At the end of expiration, the software stores all user commands and 30 ventilation parameters into an Excel spreadsheet for later review and analysis. Use of these software and control algorithms affords user-friendly operation of a tidal liquid ventilator while providing precise control of ventilation parameters.

  5. Assessment of Natural Ventilation System for a Typical Residential House in Poland

    NASA Astrophysics Data System (ADS)

    Antczak-Jarząbska, Romana; Krzaczek, Marek

    2016-09-01

    The paper presents the research results of field measurements campaign of natural ventilation performance and effectiveness in a residential building. The building is located in the microclimate whose parameters differ significantly in relation to a representative weather station. The measurement system recorded climate parameters and the physical variables characterizing the air flow in the rooms within 14 days of the winter season. The measurement results showed that in spite of proper design and construction of the ventilation system, unfavorable microclimatic conditions that differed from the predicted ones caused significant reduction in the efficiency of the ventilation system. Also, during some time periods, external climate conditions caused an opposite air flow direction in the vent inlets and outlets, leading to a significant deterioration of air quality and thermal comfort measured by CO2 concentration and PMV index in a residential area.

  6. Traffic flow in the operating room: an explorative and descriptive study on air quality during orthopedic trauma implant surgery.

    PubMed

    Andersson, Annette Erichsen; Bergh, Ingrid; Karlsson, Jón; Eriksson, Bengt I; Nilsson, Kerstin

    2012-10-01

    Understanding the protective potential of operating room (OR) ventilation under different conditions is crucial to optimizing the surgical environment. This study investigated the air quality, expressed as colony-forming units (CFU)/m(3), during orthopedic trauma surgery in a displacement-ventilated OR; explored how traffic flow and the number of persons present in the OR affects the air contamination rate in the vicinity of surgical wounds; and identified reasons for door openings in the OR. Data collection, consisting of active air sampling and observations, was performed during 30 orthopedic procedures. In 52 of the 91 air samples collected (57%), the CFU/m(3) values exceeded the recommended level of <10 CFU/m(3). In addition, the data showed a strongly positive correlation between the total CFU/m(3) per operation and total traffic flow per operation (r = 0.74; P = .001; n = 24), after controlling for duration of surgery. A weaker, yet still positive correlation between CFU/m(3) and the number of persons present in the OR (r = 0.22; P = .04; n = 82) was also found. Traffic flow, number of persons present, and duration of surgery explained 68% of the variance in total CFU/m(3) (P = .001). Traffic flow has a strong negative impact on the OR environment. The results of this study support interventions aimed at preventing surgical site infections by reducing traffic flow in the OR. Copyright © 2012 Association for Professionals in Infection Control and Epidemiology, Inc. Published by Mosby, Inc. All rights reserved.

  7. Effects of lung protective mechanical ventilation associated with permissive respiratory acidosis on regional extra-pulmonary blood flow in experimental ARDS.

    PubMed

    Hering, Rudolf; Kreyer, Stefan; Putensen, Christian

    2017-10-27

    Lung protective mechanical ventilation with limited peak inspiratory pressure has been shown to affect cardiac output in patients with ARDS. However, little is known about the impact of lung protective mechanical ventilation on regional perfusion, especially when associated with moderate permissive respiratory acidosis. We hypothesized that lung protective mechanical ventilation with limited peak inspiratory pressure and moderate respiratory acidosis results in an increased cardiac output but unequal distribution of blood flow to the different organs of pigs with oleic-acid induced ARDS. Twelve pigs were enrolled, 3 died during instrumentation and induction of lung injury. Thus, 9 animals received pressure controlled mechanical ventilation with a PEEP of 5 cmH 2 O and limited peak inspiratory pressure (17 ± 4 cmH 2 O) versus increased peak inspiratory pressure (23 ± 6 cmH 2 O) in a crossover-randomized design and were analyzed. The sequence of limited versus increased peak inspiratory pressure was randomized using sealed envelopes. Systemic and regional hemodynamics were determined by double indicator dilution technique and colored microspheres, respectively. The paired student t-test and the Wilcoxon test were used to compare normally and not normally distributed data, respectively. Mechanical ventilation with limited inspiratory pressure resulted in moderate hypercapnia and respiratory acidosis (PaCO 2 71 ± 12 vs. 46 ± 9 mmHg, and pH 7.27 ± 0.05 vs. 7.38 ± 0.04, p < 0.001, respectively), increased cardiac output (140 ± 32 vs. 110 ± 22 ml/min/kg, p<0.05) and regional blood flow in the myocardium, brain and spinal cord, adrenal and thyroid glands, the mucosal layers of the esophagus and jejunum, the muscularis layers of the esophagus and duodenum, and the gall and urinary bladders. Perfusion of kidneys, pancreas, spleen, hepatic arterial bed, and the mucosal and muscularis blood flow to the other evaluated intestinal regions remained unchanged. In this porcine model of ARDS mechanical ventilation with limited peak inspiratory pressure resulting in moderate respiratory acidosis was associated with an increase in cardiac output. However, the better systemic blood flow was not uniformly directed to the different organs. This observation may be of clinical interest in patients, e.g. with cardiac, renal and cerebral pathologies.

  8. Exhaled CO2 Parameters as a Tool to Assess Ventilation-Perfusion Mismatching during Neonatal Resuscitation in a Swine Model of Neonatal Asphyxia

    PubMed Central

    Li, Elliott Shang-shun; Cheung, Po-Yin; O'Reilly, Megan; LaBossiere, Joseph; Lee, Tze-Fun; Cowan, Shaun; Bigam, David L.; Schmölzer, Georg Marcus

    2016-01-01

    Background End-tidal CO2 (ETCO2), partial pressure of exhaled CO2 (PECO2), and volume of expired CO2 (VCO2) can be continuously monitored non-invasively to reflect pulmonary ventilation and perfusion status. Although ETCO2 ≥14mmHg has been shown to be associated with return of an adequate heart rate in neonatal resuscitation and quantifying the PECO2 has the potential to serve as an indicator of resuscitation quality, there is little information regarding capnometric measurement of PECO2 and ETCO2 in detecting return of spontaneous circulation (ROSC) and survivability in asphyxiated neonates receiving cardiopulmonary resuscitation (CPR). Methods Seventeen newborn piglets were anesthetized, intubated, instrumented, and exposed to 45-minute normocapnic hypoxia followed by apnea to induce asphyxia. Protocolized resuscitation was initiated when heart rate decreased to 25% of baseline. Respiratory and hemodynamic parameters including ETCO2, PECO2, VCO2, heart rate, cardiac output, and carotid artery flow were continuously measured and analyzed. Results There were no differences in respiratory and hemodynamic parameters between surviving and non-surviving piglets prior to CPR. Surviving piglets had significantly higher ETCO2, PECO2, VCO2, cardiac index, and carotid artery flow values during CPR compared to non-surviving piglets. Conclusion Surviving piglets had significantly better respiratory and hemodynamic parameters during resuscitation compared to non-surviving piglets. In addition to optimizing resuscitation efforts, capnometry can assist by predicting outcomes of newborns requiring chest compressions. PMID:26766424

  9. Economic, Environmental and Health Implications of Enhanced Ventilation in Office Buildings.

    PubMed

    MacNaughton, Piers; Pegues, James; Satish, Usha; Santanam, Suresh; Spengler, John; Allen, Joseph

    2015-11-18

    Current building ventilation standards are based on acceptable minimums. Three decades of research demonstrates the human health benefits of increased ventilation above these minimums. Recent research also shows the benefits on human decision-making performance in office workers, which translates to increased productivity. However, adoption of enhanced ventilation strategies is lagging. We sought to evaluate two of the perceived potential barriers to more widespread adoption-Economic and environmental costs. We estimated the energy consumption and associated per building occupant costs for office buildings in seven U.S. cities, representing different climate zones for three ventilation scenarios (standard practice (20 cfm/person), 30% enhanced ventilation, and 40 cfm/person) and four different heating, ventilation and air conditioning (HVAC) system strategies (Variable Air Volume (VAV) with reheat and a Fan Coil Unit (FCU), both with and without an energy recovery ventilator). We also estimated emissions of greenhouse gases associated with this increased energy usage, and, for comparison, converted this to the equivalent number of vehicles using greenhouse gas equivalencies. Lastly, we paired results from our previous research on cognitive function and ventilation with labor statistics to estimate the economic benefit of increased productivity associated with increasing ventilation rates. Doubling the ventilation rate from the American Society of Heating, Refrigeration and Air-Conditioning Engineers minimum cost less than $40 per person per year in all climate zones investigated. Using an energy recovery ventilation system significantly reduced energy costs, and in some scenarios led to a net savings. At the highest ventilation rate, adding an ERV essentially neutralized the environmental impact of enhanced ventilation (0.03 additional cars on the road per building across all cities). The same change in ventilation improved the performance of workers by 8%, equivalent to a $6500 increase in employee productivity each year. Reduced absenteeism and improved health are also seen with enhanced ventilation. The health benefits associated with enhanced ventilation rates far exceed the per-person energy costs relative to salary costs. Environmental impacts can be mitigated at regional, building, and individual-level scales through the transition to renewable energy sources, adoption of energy efficient systems and ventilation strategies, and promotion of other sustainable policies.

  10. Economic, Environmental and Health Implications of Enhanced Ventilation in Office Buildings

    PubMed Central

    MacNaughton, Piers; Pegues, James; Satish, Usha; Santanam, Suresh; Spengler, John; Allen, Joseph

    2015-01-01

    Introduction: Current building ventilation standards are based on acceptable minimums. Three decades of research demonstrates the human health benefits of increased ventilation above these minimums. Recent research also shows the benefits on human decision-making performance in office workers, which translates to increased productivity. However, adoption of enhanced ventilation strategies is lagging. We sought to evaluate two of the perceived potential barriers to more widespread adoption—Economic and environmental costs. Methods: We estimated the energy consumption and associated per building occupant costs for office buildings in seven U.S. cities, representing different climate zones for three ventilation scenarios (standard practice (20 cfm/person), 30% enhanced ventilation, and 40 cfm/person) and four different heating, ventilation and air conditioning (HVAC) system strategies (Variable Air Volume (VAV) with reheat and a Fan Coil Unit (FCU), both with and without an energy recovery ventilator). We also estimated emissions of greenhouse gases associated with this increased energy usage, and, for comparison, converted this to the equivalent number of vehicles using greenhouse gas equivalencies. Lastly, we paired results from our previous research on cognitive function and ventilation with labor statistics to estimate the economic benefit of increased productivity associated with increasing ventilation rates. Results: Doubling the ventilation rate from the American Society of Heating, Refrigeration and Air-Conditioning Engineers minimum cost less than $40 per person per year in all climate zones investigated. Using an energy recovery ventilation system significantly reduced energy costs, and in some scenarios led to a net savings. At the highest ventilation rate, adding an ERV essentially neutralized the environmental impact of enhanced ventilation (0.03 additional cars on the road per building across all cities). The same change in ventilation improved the performance of workers by 8%, equivalent to a $6500 increase in employee productivity each year. Reduced absenteeism and improved health are also seen with enhanced ventilation. Conclusions: The health benefits associated with enhanced ventilation rates far exceed the per-person energy costs relative to salary costs. Environmental impacts can be mitigated at regional, building, and individual-level scales through the transition to renewable energy sources, adoption of energy efficient systems and ventilation strategies, and promotion of other sustainable policies. PMID:26593933

  11. Comparison of the effectiveness of high flow nasal oxygen cannula vs. standard non-rebreather oxygen face mask in post-extubation intensive care unit patients.

    PubMed

    Brotfain, Evgeni; Zlotnik, Alexander; Schwartz, Andrei; Frenkel, Amit; Koyfman, Leonid; Gruenbaum, Shaun E; Klein, Moti

    2014-11-01

    Optimal oxygen supply is the cornerstone of the management of critically ill patients after extubation, especially in patients at high risk for extubation failure. In recent years, high flow oxygen system devices have offered an appropriate alternative to standard oxygen therapy devices such as conventional face masks and nasal prongs. To assess the clinical effects of high flow nasal cannula (HFNC) compared with standard oxygen face masks in Intensive Care Unit (ICU) patients after extubation. We retrospectively analyzed 67 consecutive ventilated critical care patients in the ICU over a period of 1 year. The patients were allocated to two treatment groups: HFNC (34 patients, group 1) and non-rebreathing oxygen face mask (NRB) (33 patients, group 2). Vital respiratory and hemodynamic parameters were assessed prior to extubation and 6 hours after extubation. The primary clinical outcomes measured were improvement in oxygenation, ventilation-free days, re-intubation, ICU length of stay, and mortality. The two groups demonstrated similar hemodynamic patterns before and after extubation. The respiratory rate was slightly elevated in both groups after extubation with no differences observed between groups. There were no statistically significant clinical differences in PaCO2. However, the use of HFNC resulted in improved PaO2/FiO2 post-extubation (P < 0.05). There were more ventilator-free days in the HFNC group (P< 0.05) and fewer patients required reintubation (1 vs. 6). There were no differences in ICU length of stay or mortality. This study demonstrated better oxygenation for patients treated with HFNC compared with NRB after extubation. HFNC may be more effective than standard oxygen supply devices for oxygenation in the post-extubation period.

  12. Airway Humidification During High-Frequency Percussive Ventilation

    DTIC Science & Technology

    2009-03-01

    Airway Humidification During High-Frequency Percussive Ventilation Patrick F Allan MD, Michael J Hollingsworth CRT, Gordon C Maniere CRT, Anthony K...about the risk of inadequate humidification during high- frequency percussive ventilation (HFPV). METHODS: We studied 5 humidifiers during HFPV with a...50 L/min, and the ConchaTherm Hi-Flow with VDR nebulizer) provided carinal humidification equivalent to the comparator setup, without regard to

  13. Human respiration at rest in rapid compression and at high pressures and gas densities

    NASA Technical Reports Server (NTRS)

    Gelfand, R.; Lambertsen, C. J.; Strauss, R.; Clark, J. M.; Puglia, C. D.

    1983-01-01

    The ventilation (V), end-tidal PCO2 (PACO2), and CO2 elimination rate were determined in men at rest breathing CO2-free gas over the pressure range 1-50 ATA and the gas density range 0.4-25 g/l, during slow and rapid compressions, at stable elevated ambient pressures and during slow decompressions. Progressive increase in pulmonary gas flow resistance due to elevation of ambient pressure and inspired gas density to the He-O2 equivalent of 5000 feet of seawater was found to produce a complex pattern of change in PACO2. It was found that as both ambient pressure and pulmonary gas flow resistance were progressively raised, PACO2 at first increased, went through a maximum, and then declined towards values near the 1 ATA level. It is concluded that this pattern of PACO2 change results from the interaction on ventilation of the increase in pulmonary resistance due to the elevation of gas density with the increase in respiratory drive postulated as due to generalized central nervous system excitation associated with exposure to high hydrostatic pressure. It is suggested that a similar interaction exists between increased gas flow resistance and the increase in respiratory drive related to nitrogen partial pressure and the resulting narcosis.

  14. Effect of substance P on cardiovascular and respiratory function in subjects.

    PubMed

    Fuller, R W; Maxwell, D L; Dixon, C M; McGregor, G P; Barnes, V F; Bloom, S R; Barnes, P J

    1987-04-01

    The effect of substance P (SP), administered both intravenously and by inhalation, has been studied in normal and asthmatic humans. Intravenous infusion of SP (0.2-3.3 pmol X kg-1 X min-1) achieving a plasma concentration of SP between 5 and 25 pM produced vasodilatation (mean +/- SD), maximal increase in skin temperature (0.9 +/- 0.3 degree C) (P less than 0.05), and fall in diastolic blood pressure (8.5 +/- 2.9 mmHg) (P less than 0.05) associated with an increase in heart rate (15 +/- 10 beats/min) (P less than 0.05). All subjects had a fall in Vp30 (airflow at 70% of forced vital capacity measured from total lung capacity after a forced partial expiratory flow maneuver) at low infusion rate (P less than 0.05) and a significant rise at the highest infusion rate (P less than 0.05). Ventilation at rest and when stimulated by transient hypoxia increased (mean increase in resting ventilation 0.73 +/- 0.4 l/min and mean percent increase in transient ventilatory hypoxic response 41 +/- 27%). There was a small nonsignificant increase in plasma norepinephrine but no change in epinephrine or histamine. Inhaled SP, up to 0.7 mumol, caused a small nonsignificant fall in airway function in asthmatic subjects. SP has demonstrable effects on vascular smooth muscle and control of ventilation but at the doses studied had little effect on airway function.

  15. Cardiorespiratory Interactions in Paediatrics: 'It's (almost always) the circulation stupid!'

    PubMed

    Rigby, M L; Rosenthal, M

    2017-03-01

    The interaction of the heart and lungs is probably the most important aspect of life and survival. Fortunately, it is not difficult to understand the fundamentals. The purpose of the lungs and their ventilation is to present oxygen to the circulation via the alveoli and to receive carbon dioxide from the circulation and then expel it. The relations of the heart and lungs and the matching of blood flow to the various organs with ventilation and lung perfusion may be disrupted by a variety of congenital or acquired heart malformations. They include those giving rise to an increased or reduced pulmonary blood flow, elevated pulmonary venous pressure or external physical pressure on the airways or lung parenchyma. Respiratory disorders which compromise cardiac function include states with reduced alveolar ventilation, those with a barrier to ventilation or perfusion, ventilation/perfusion mismatch and pulmonary vascular disease. There is also a fascinating group in which congenital disorders of the heart and lung co-exist to produce very particular modes of abnormal cardiopulmonary interaction. Copyright © 2016. Published by Elsevier Ltd.

  16. A randomised control study comparing the Infant Flow Driver with nasal continuous positive airway pressure in preterm infants

    PubMed Central

    Mazzella, M; Bellini, C; Calevo, M; Campone, F; Massocco, D; Mezzano, P; Zullino, E; Scopesi, F; Arioni, C; Bonacci, W; Serra, G

    2001-01-01

    OBJECTIVE—To compare the effectiveness of the Infant Flow Driver (IFD) with single prong nasal continuous positive airway pressure (nCPAP) in preterm neonates affected by respiratory distress syndrome.
DESIGN—Randomised controlled study.
PATIENTS—Between September 1997 and March 1999, 36 preterm infants who were eligible for CPAP treatment were randomly selected for either nCPAP or IFD and studied prospectively for changes in oxygen requirement and/or respiratory rate. The requirement for mechanical ventilation, complications of treatment, and effects on mid-term outcome were also evaluated.
RESULTS—Use of the IFD had a significantly beneficial effect on both oxygen requirement and respiratory rate (p < 0.0001) when compared with nCPAP. Moreover, O2 requirement and respiratory rate were significantly decreased by four hours (p < 0.001 and p < 0.03 respectively). The probability of remaining supplementary oxygen free over the first 48 hours of treatment was significantly higher in patients treated with the IFD than with nCPAP (p < 0.02). IFD treated patients had a higher success (weaning) rate (94% v 72%) and shorter duration of treatment (49.3 (31) v 56 (29.7) hours respectively; mean (SD)), although the difference was not significant.
CONCLUSIONS—IFD appears to be a feasible device for managing respiratory distress syndrome in preterm infants, and benefits may be had with regard to oxygen requirement and respiratory rate when compared with nCPAP. The trend towards reduced requirement for mechanical ventilation, shorter clinical recovery time, and shorter duration of treatment requires further evaluation in a multicentre randomised clinical trial.

 PMID:11517199

  17. Bacterial burden in the operating room: impact of airflow systems.

    PubMed

    Hirsch, Tobias; Hubert, Helmine; Fischer, Sebastian; Lahmer, Armin; Lehnhardt, Marcus; Steinau, Hans-Ulrich; Steinstraesser, Lars; Seipp, Hans-Martin

    2012-09-01

    Wound infections present one of the most prevalent and frequent complications associated with surgical procedures. This study analyzes the impact of currently used ventilation systems in the operating room to reduce bacterial contamination during surgical procedures. Four ventilation systems (window-based ventilation, supported air nozzle canopy, low-turbulence displacement airflow, and low-turbulence displacement airflow with flow stabilizer) were analyzed. Two hundred seventy-seven surgical procedures in 6 operating rooms of 5 different hospitals were analyzed for this study. Window-based ventilation showed the highest intraoperative contamination (13.3 colony-forming units [CFU]/h) followed by supported air nozzle canopy (6.4 CFU/h; P = .001 vs window-based ventilation) and low-turbulence displacement airflow (3.4 and 0.8 CFU/h; P < .001 vs window-based ventilation and supported air nozzle canopy). The highest protection was provided by the low-turbulence displacement airflow with flow stabilizer (0.7 CFU/h), which showed a highly significant difference compared with the best supported air nozzle canopy theatre (3.9 CFU/h; P < .001). Furthermore, this system showed no increase of contamination in prolonged durations of surgical procedures. This study shows that intraoperative contamination can be significantly reduced by the use of adequate ventilation systems. Copyright © 2012 Association for Professionals in Infection Control and Epidemiology, Inc. Published by Mosby, Inc. All rights reserved.

  18. Estimation of tracheal pressure and imposed expiratory work of breathing by the endotracheal tube, heat and moisture exchanger, and ventilator during mechanical ventilation.

    PubMed

    Uchiyama, Akinori; Yoshida, Takeshi; Yamanaka, Hidenori; Fujino, Yuji

    2013-07-01

    The resistance of the endotracheal tube (ETT), the heat and moisture exchanger (HME), and the ventilator may affect the patient's respiratory status. Although previous studies examined the inspiratory work of breathing (WOB), investigation of WOB in the expiratory phase is rare. We estimated tracheal pressure at the tip of the ETT (Ptrach) and calculated expiratory WOB imposed by the ETT, the HME, and the expiratory valve. We examined imposed expiratory WOB in patients under a continuous mandatory ventilation (CMV) mode and during spontaneous breathing trials (SBTs). We hypothesized that imposed expiratory WOB would increase with heightened ventilatory demand. We measured airway pressure (Paw) and respiratory flow (V). We estimated Ptrach using the equation Ptrach = Paw - K1 × V(K2) - 2.70 × V(L/s)(1.42). K1 and K2 were determined by the inner diameter (ID) of the ETT. Imposed expiratory WOB was calculated from the area of Ptrach above PEEP versus lung volume. We examined imposed expiratory WOB and imposed expiratory resistance in relation to mean expiratory flow. We examined 28 patients under CMV mode, and 29 during SBT. During both CMV and SBT, as mean expiratory flow increased, imposed expiratory WOB increased. The regression curves between mean expiratory flow (x) (L/s) and imposed expiratory WOB (y) (J/L) were y = 1.35x(0.83) (R(2) = 0.79) for 7 mm ID ETT under CMV, y = 1.12x(0.82) (R(2) = 0.73) for 8 mm ID ETT under CMV, y = 1.07x(1.04) (R(2) = 0.85) for 7 mm ID ETT during SBT, and y = 0.84x(0.93) (R(2) = 0.75) for 8 mm ID ETT during SBT. Levels of imposed expiratory WOB were affected by ETT diameter and ventilator mode. The reason for increasing imposed expiratory WOB was an increase in expiratory resistance imposed by the ETT and HME. Under mechanical ventilation, imposed expiratory WOB should be considered in patients with higher minute ventilation.

  19. An experimental study on effects of increased ventilation flow on students' perception of indoor environment in computer classrooms.

    PubMed

    Norbäck, D; Nordström, K

    2008-08-01

    The effects of ventilation in computer classrooms were studied with university students (n = 355) in a blinded study, 31% were women and 3.8% had asthma. Two classrooms had a higher air exchange (4.1-5.2 ac/h); two others had a lower air exchange (2.3-2.6 ac/h). After 1 week, ventilation conditions were shifted. The students reported environmental perceptions during the last hour. Room temperature, RH, CO2, PM10 and ultra-fine particles were measured simultaneously. Mean CO2 was 1185 ppm at lower and 922 ppm at higher air exchange. Mean temperature was 23.2 degrees C at lower and 22.1 degrees C at higher air exchange. After mutual adjustment (temperature, RH, CO2, air exchange), measured temperature was associated with a perception of higher temperature (P < 0.001), lower air movement (P < 0.001), and poorer air quality (P < 0.001). Higher air exchange was associated with a perception of lower temperature (P < 0.001), higher air movement (P = 0.001), and better air quality (P < 0.001). In the longitudinal analysis (n = 83), increased air exchange caused a perception of lower temperature (P = 0.002), higher air movement (P < 0.001), better air quality (P = 0.001), and less odor (P = 0.02). In conclusion, computer classrooms have CO2 levels above 1000 ppm and temperatures above 22 degrees C. Increased ventilation from 7 l/s per person to 10-13 l/s per person can improve thermal comfort and air quality. Computer classrooms are crowded indoor environments with a high thermal load from both students and computer equipment. It is important to control room temperature either by air conditioning, sun shields, or sufficiently high ventilation flow. A high ventilation flow is also crucial to achieving good perceived air quality. Personal ventilation flow should be at least 10 l/s. Possible loss of learning ability due to poor indoor air quality in university buildings deserves more attention.

  20. Energy and cost associated with ventilating office buildings in a tropical climate.

    PubMed

    Rim, Donghyun; Schiavon, Stefano; Nazaroff, William W

    2015-01-01

    Providing sufficient amounts of outdoor air to occupants is a critical building function for supporting occupant health, well-being and productivity. In tropical climates, high ventilation rates require substantial amounts of energy to cool and dehumidify supply air. This study evaluates the energy consumption and associated cost for thermally conditioning outdoor air provided for building ventilation in tropical climates, considering Singapore as an example locale. We investigated the influence on energy consumption and cost of the following factors: outdoor air temperature and humidity, ventilation rate (L/s per person), indoor air temperature and humidity, air conditioning system coefficient of performance (COP), and cost of electricity. Results show that dehumidification of outdoor air accounts for more than 80% of the energy needed for building ventilation in Singapore's tropical climate. Improved system performance and/or a small increase in the indoor temperature set point would permit relatively large ventilation rates (such as 25 L/s per person) at modest or no cost increment. Overall, even in a thermally demanding tropical climate, the energy cost associated with increasing ventilation rate up to 25 L/s per person is less than 1% of the wages of an office worker in an advanced economy like Singapore's. This result implies that the benefits of increasing outdoor air ventilation rate up to 25 L/s per person--which is suggested to provide for productivity increases, lower sick building syndrome symptom prevalence, and reduced sick leave--can be much larger than the incremental cost of ventilation.

  1. Early application of airway pressure release ventilation may reduce the duration of mechanical ventilation in acute respiratory distress syndrome.

    PubMed

    Zhou, Yongfang; Jin, Xiaodong; Lv, Yinxia; Wang, Peng; Yang, Yunqing; Liang, Guopeng; Wang, Bo; Kang, Yan

    2017-11-01

    Experimental animal models of acute respiratory distress syndrome (ARDS) have shown that the updated airway pressure release ventilation (APRV) methodologies may significantly improve oxygenation, maximize lung recruitment, and attenuate lung injury, without circulatory depression. This led us to hypothesize that early application of APRV in patients with ARDS would allow pulmonary function to recover faster and would reduce the duration of mechanical ventilation as compared with low tidal volume lung protective ventilation (LTV). A total of 138 patients with ARDS who received mechanical ventilation for <48 h between May 2015 to October 2016 while in the critical care medicine unit (ICU) of the West China Hospital of Sichuan University were enrolled in the study. Patients were randomly assigned to receive APRV (n = 71) or LTV (n = 67). The settings for APRV were: high airway pressure (P high ) set at the last plateau airway pressure (P plat ), not to exceed 30 cmH 2 O) and low airway pressure ( P low ) set at 5 cmH 2 O; the release phase (T low ) setting adjusted to terminate the peak expiratory flow rate to ≥ 50%; release frequency of 10-14 cycles/min. The settings for LTV were: target tidal volume of 6 mL/kg of predicted body weight; P plat not exceeding 30 cmH 2 O; positive end-expiratory pressure (PEEP) guided by the PEEP-FiO 2 table according to the ARDSnet protocol. The primary outcome was the number of days without mechanical ventilation from enrollment to day 28. The secondary endpoints included oxygenation, P plat , respiratory system compliance, and patient outcomes. Compared with the LTV group, patients in the APRV group had a higher median number of ventilator-free days {19 [interquartile range (IQR) 8-22] vs. 2 (IQR 0-15); P < 0.001}. This finding was independent of the coexisting differences in chronic disease. The APRV group had a shorter stay in the ICU (P = 0.003). The ICU mortality rate was 19.7% in the APRV group versus 34.3% in the LTV group (P = 0.053) and was associated with better oxygenation and respiratory system compliance, lower P plat , and less sedation requirement during the first week following enrollment (P < 0.05, repeated-measures analysis of variance). Compared with LTV, early application of APRV in patients with ARDS improved oxygenation and respiratory system compliance, decreased P plat and reduced the duration of both mechanical ventilation and ICU stay.

  2. Inspiratory High Frequency Airway Oscillation Attenuates Resistive Loaded Dyspnea and Modulates Respiratory Function in Young Healthy Individuals

    PubMed Central

    Morris, Theresa; Sumners, David Paul; Green, David Andrew

    2014-01-01

    Direct chest-wall percussion can reduce breathlessness in Chronic Obstructive Pulmonary Disease and respiratory function may be improved, in health and disease, by respiratory muscle training (RMT). We tested whether high-frequency airway oscillation (HFAO), a novel form of airflow oscillation generation can modulate induced dyspnoea and respiratory strength and/or patterns following 5 weeks of HFAO training (n = 20) compared to a SHAM-RMT (conventional flow-resistive RMT) device (n = 15) in healthy volunteers (13 males; aged 20–36 yrs). HFAO causes oscillations with peak-to-peak amplitude of 1 cm H2O, whereas the SHAM-RMT device was identical but created no pressure oscillation. Respiratory function, dyspnoea and ventilation during 3 minutes of spontaneous resting ventilation, 1 minute of maximal voluntary hyperventilation and 1 minute breathing against a moderate inspiratory resistance, were compared PRE and POST 5-weeks of training (2×30 breaths at 70% peak flow, 5 days a week). Training significantly reduced NRS dyspnoea scores during resistive loaded ventilation, both in the HFAO (p = 0.003) and SHAM-RMT (p = 0.005) groups. Maximum inspiratory static pressure (cm H2O) was significantly increased by HFAO training (vs. PRE; p<0.001). Maximum inspiratory dynamic pressure was increased by training in both the HFAO (vs. PRE; p<0.001) and SHAM-RMT (vs. PRE; p = 0.021) groups. Peak inspiratory flow rate (L.s−1) achieved during the maximum inspiratory dynamic pressure manoeuvre increased significantly POST (vs. PRE; p = 0.001) in the HFAO group only. HFAO reduced inspiratory resistive loading–induced dyspnoea and augments static and dynamic maximal respiratory manoeuvre performance in excess of flow-resistive IMT (SHAM-RMT) in healthy individuals without the respiratory discomfort associated with RMT. PMID:24651392

  3. The effects of opening areas on solar chimney performance

    NASA Astrophysics Data System (ADS)

    Ling, L. S.; Rahman, M. M.; Chu, C. M.; Misaran, M. S. bin; Tamiri, F. M.

    2017-07-01

    To enhance natural ventilation at day time, solar chimney is one of the suitable options for topical country like Malaysia. Solar chimney creates air flow due to stack effect caused by temperature difference between ambient and inside wall. In the solar chimney, solar energy is harvested by the inner wall that cause temperature rise compare to ambient. Therefore, the efficiency of the solar chimney depends on the availability of solar energy as well as the solar intensity. In addition, it is very hard to get good ventilation at night time by using a solar chimney. To overcome this problem one of the suitable valid option is to integrate solar chimney with turbine ventilator. A new type of solar chimney is designed and fluid flow analyzed with the computational fluid dynamics (CFD) software. The aim of CFD and theoretical study are to investigate the effect of opening areas on modified solar chimney performance. The inlet and outlet area of solar chimney are varied from 0.0224m2 to 0.6m2 and 0.1m2 to 0.14m2 respectively based on the changes of inclination angle and gap between inner and outer wall. In the CFD study the constant heat flux is considered as 500W/m2. CFD result shows that there is no significant relation between opening areas and the air flow rate through solar chimney but the ratio between inlet and outlet is significant on flow performance. If the area ratio between inlet and outlet are equal to two or larger, the performance of the solar chimney is better than the solar chimney with ratio lesser than two. The solar chimney performance does not effect if the area ratio between inlet and outlet varies from 1 to 2. This result will be useful for design and verification of actual solar chimney performance.

  4. MR imaging of apparent 3He gas transport in narrow pipes and rodent airways

    NASA Astrophysics Data System (ADS)

    Minard, Kevin R.; Jacob, Richard E.; Laicher, Gernot; Einstein, Daniel R.; Kuprat, Andrew P.; Corley, Richard A.

    2008-10-01

    High sensitivity makes hyperpolarized 3He an attractive signal source for visualizing gas flow with magnetic resonance (MR) imaging. Its rapid Brownian motion, however, can blur observed flow lamina and alter measured diffusion rates when excited nuclei traverse shear-induced velocity gradients during data acquisition. Here, both effects are described analytically, and predicted values for measured transport during laminar flow through a straight, 3.2-mm diameter pipe are validated using two-dimensional (2D) constant-time images of different binary gas mixtures. Results show explicitly how measured transport in narrow conduits is characterized by apparent values that depend on underlying gas dynamics and imaging time. In ventilated rats, this is found to obscure acquired airflow images. Nevertheless, flow splitting at airway branches is still evident and use of 3D vector flow mapping is shown to reveal surprising detail that highlights the correlation between gas dynamics and lung structure.

  5. A program for sustained improvement in preventing ventilator associated pneumonia in an intensive care setting

    PubMed Central

    2012-01-01

    Background Ventilator-associated pneumonia (VAP) is a common infection in the intensive care unit (ICU) and associated with a high mortality. Methods A quasi-experimental study was conducted in a medical-surgical ICU. Multiple interventions to optimize VAP prevention were performed from October 2008 to December 2010. All of these processes, including the Institute for Healthcare Improvement’s (IHI) ventilator bundle plus oral decontamination with chlorhexidine and continuous aspiration of subglottic secretions (CASS), were adopted for patients undergoing mechanical ventilation. Results We evaluated a total of 21,984 patient-days, and a total of 6,052 ventilator-days (ventilator utilization rate of 0.27). We found VAP rates of 1.3 and 2.0 per 1,000 ventilator days respectively in 2009 and 2010, achieving zero incidence of VAP several times during 12 months, whenever VAP bundle compliance was over 90%. Conclusion These results suggest that it is possible to reduce VAP rates to near zero and sustain these rates, but it requires a complex process involving multiple performance measures and interventions that must be permanently monitored. PMID:23020101

  6. A program for sustained improvement in preventing ventilator associated pneumonia in an intensive care setting.

    PubMed

    Caserta, Raquel A; Marra, Alexandre R; Durão, Marcelino S; Silva, Cláudia Vallone; Pavao dos Santos, Oscar Fernando; Neves, Henrique Sutton de Sousa; Edmond, Michael B; Timenetsky, Karina Tavares

    2012-09-29

    Ventilator-associated pneumonia (VAP) is a common infection in the intensive care unit (ICU) and associated with a high mortality. A quasi-experimental study was conducted in a medical-surgical ICU. Multiple interventions to optimize VAP prevention were performed from October 2008 to December 2010. All of these processes, including the Institute for Healthcare Improvement's (IHI) ventilator bundle plus oral decontamination with chlorhexidine and continuous aspiration of subglottic secretions (CASS), were adopted for patients undergoing mechanical ventilation. We evaluated a total of 21,984 patient-days, and a total of 6,052 ventilator-days (ventilator utilization rate of 0.27). We found VAP rates of 1.3 and 2.0 per 1,000 ventilator days respectively in 2009 and 2010, achieving zero incidence of VAP several times during 12 months, whenever VAP bundle compliance was over 90%. These results suggest that it is possible to reduce VAP rates to near zero and sustain these rates, but it requires a complex process involving multiple performance measures and interventions that must be permanently monitored.

  7. Using a Ventilation Controller to Optimize Residential Passive Ventilation For Energy and Indoor Air Quality

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Turner, William; Walker, Iain

    One way to reduce the energy impact of providing residential ventilation is to use passive and hybrid systems. However, these passive and hybrid (sometimes called mixed-mode) systems must still meet chronic and acute health standards for ventilation. This study uses a computer simulation approach to examine the energy and indoor air quality (IAQ) implications of passive and hybrid ventilation systems, in 16 California climate zones. Both uncontrolled and flow controlled passive stacks are assessed. A new hybrid ventilation system is outlined that uses an intelligent ventilation controller to minimise energy use, while ensuring chronic and acute IAQ standards are met.more » ASHRAE Standard 62.2-2010 – the United States standard for residential ventilation - is used as the chronic standard, and exposure limits for PM 2.5, formaldehyde and NO 2 are used as the acute standards.The results show that controlled passive ventilation and hybrid ventilation can be used in homes to provide equivalent IAQ to continuous mechanical ventilation, for less use of energy.« less

  8. 46 CFR 108.437 - Pipe sizes and discharge rates for enclosed ventilation systems for rotating electrical equipment.

    Code of Federal Regulations, 2011 CFR

    2011-10-01

    ... 46 Shipping 4 2011-10-01 2011-10-01 false Pipe sizes and discharge rates for enclosed ventilation... Systems Fixed Carbon Dioxide Fire Extinguishing Systems § 108.437 Pipe sizes and discharge rates for enclosed ventilation systems for rotating electrical equipment. (a) The minimum pipe size for the initial...

  9. Influence of Mechanical Ventilation on the Incidence of Pneumothorax During Infraclavicular Subclavian Vein Catheterization: A Prospective Randomized Noninferiority Trial.

    PubMed

    Kim, Eugene; Kim, Hyun Joo; Hong, Deok Man; Park, Hee-Pyoung; Bahk, Jae-Hyon

    2016-09-01

    It remains unclear whether we have to interrupt mechanical ventilation during infraclavicular subclavian venous catheterization. In practice, the clinicians' choice about lung deflation depends on their own discretion. The purpose of this study was to assess the influence of mechanical ventilation on the incidence of pneumothorax during infraclavicular subclavian venous catheterization. A total of 332 patients, who needed subclavian venous catheterization, were randomly assigned to 1 of the 2 groups: catheterizations were performed with the patients' lungs under mechanical ventilation (ventilation group, n = 165) or without mechanical ventilation (deflation group, n = 167). The incidences of pneumothorax and other complications such as arterial puncture, hemothorax, or catheter misplacements and the success rate of catheterization were compared. The incidences of pneumothorax were 0% (0/165) in the ventilation group and 0.6% (1/167) in the deflation group. The incidence of pneumothorax in the deflation group was 0.6% higher than that in the ventilation group and the 2-sided 90% confidence interval for the difference was (-1.29% to 3.44%). Because the lower bound for the 2-sided 90% confidence interval, -1.29%, was higher than the predefined noninferiority margin of -3%, the inferiority of the ventilation group over the deflation group was rejected at the .05 level of significance. Other complication rates and success rates of catheterization were comparable between 2 groups. The oxygen saturation dropped below 95% in 9 patients in the deflation group, while none in the ventilation group (P = .007). The success and complication rates were similar regardless of mechanical ventilation. During infraclavicular subclavian venous catheterization, interruption of mechanical ventilation does not seem to be necessary for the prevention of pneumothorax.

  10. Minute ventilation of cyclists, car and bus passengers: an experimental study.

    PubMed

    Zuurbier, Moniek; Hoek, Gerard; van den Hazel, Peter; Brunekreef, Bert

    2009-10-27

    Differences in minute ventilation between cyclists, pedestrians and other commuters influence inhaled doses of air pollution. This study estimates minute ventilation of cyclists, car and bus passengers, as part of a study on health effects of commuters' exposure to air pollutants. Thirty-four participants performed a submaximal test on a bicycle ergometer, during which heart rate and minute ventilation were measured simultaneously at increasing cycling intensity. Individual regression equations were calculated between heart rate and the natural log of minute ventilation. Heart rates were recorded during 280 two hour trips by bicycle, bus and car and were calculated into minute ventilation levels using the individual regression coefficients. Minute ventilation during bicycle rides were on average 2.1 times higher than in the car (individual range from 1.3 to 5.3) and 2.0 times higher than in the bus (individual range from 1.3 to 5.1). The ratio of minute ventilation of cycling compared to travelling by bus or car was higher in women than in men. Substantial differences in regression equations were found between individuals. The use of individual regression equations instead of average regression equations resulted in substantially better predictions of individual minute ventilations. The comparability of the gender-specific overall regression equations linking heart rate and minute ventilation with one previous American study, supports that for studies on the group level overall equations can be used. For estimating individual doses, the use of individual regression coefficients provides more precise data. Minute ventilation levels of cyclists are on average two times higher than of bus and car passengers, consistent with the ratio found in one small previous study of young adults. The study illustrates the importance of inclusion of minute ventilation data in comparing air pollution doses between different modes of transport.

  11. Pressure Dynamic Characteristics of Pressure Controlled Ventilation System of a Lung Simulator

    PubMed Central

    Shi, Yan; Ren, Shuai; Cai, Maolin; Xu, Weiqing; Deng, Qiyou

    2014-01-01

    Mechanical ventilation is an important life support treatment of critically ill patients, and air pressure dynamics of human lung affect ventilation treatment effects. In this paper, in order to obtain the influences of seven key parameters of mechanical ventilation system on the pressure dynamics of human lung, firstly, mechanical ventilation system was considered as a pure pneumatic system, and then its mathematical model was set up. Furthermore, to verify the mathematical model, a prototype mechanical ventilation system of a lung simulator was proposed for experimental study. Last, simulation and experimental studies on the air flow dynamic of the mechanical ventilation system were done, and then the pressure dynamic characteristics of the mechanical system were obtained. The study can be referred to in the pulmonary diagnostics, treatment, and design of various medical devices or diagnostic systems. PMID:25197318

  12. Computational fluid dynamics study on the influence of an alternate ventilation configuration on the possible flow path of infectious cough aerosols in a mock airborne infection isolation room.

    PubMed

    Thatiparti, Deepthi Sharan; Ghia, Urmila; Mead, Kenneth R

    2016-01-01

    When infectious epidemics occur, they can be perpetuated within health care settings, potentially resulting in severe health care workforce absenteeism, morbidity, mortality, and economic losses. The ventilation system configuration of an airborne infection isolation room is one factor that can play a role in protecting health care workers from infectious patient bioaerosols. Though commonly associated with airborne infectious diseases, the airborne infection isolation room design can also impact other transmission routes such as short-range airborne as well as fomite and contact transmission routes that are impacted by contagion concentration and recirculation. This article presents a computational fluid dynamics study on the influence of the ventilation configuration on the possible flow path of bioaerosol dispersal behavior in a mock airborne infection isolation room. At first, a mock airborne infection isolation room was modeled that has the room geometry and layout, ventilation parameters, and pressurization corresponding to that of a traditional ceiling-mounted ventilation arrangement observed in existing hospitals. An alternate ventilation configuration was then modeled to retain the linear supply diffuser in the original mock airborne infection isolation room but interchanging the square supply and exhaust locations to place the exhaust closer to the patient source and allow clean air from supply vents to flow in clean-to-dirty flow paths, originating in uncontaminated parts of the room prior to entering the contaminated patient's air space. The modeled alternate airborne infection isolation room ventilation rate was 12 air changes per hour. Two human breathing models were used to simulate a source patient and a receiving health care worker. A patient cough cycle was introduced into the simulation, and the airborne infection dispersal was tracked in time using a multi-phase flow simulation approach. The results from the alternate configuration revealed that the cough aerosols were pulled by the exhaust vent without encountering the health care worker by 0.93 s after patient coughs and the particles were controlled as the aerosols' flow path was uninterrupted by an air particle streamline from patient to the ceiling exhaust venting out cough aerosols. However, not all the aerosols were vented out of the room. The remaining cough aerosols entered the health care worker's breathing zone by 0.98 s. This resulted in one of the critical stages in terms of the health care worker's exposure to airborne virus and presented the opportunity for the health care worker to suffer adverse health effects from the inhalation of cough aerosols. Within 2 s, the cough aerosols reentered and recirculated within the patient and health care worker's surroundings resulting in pockets of old contaminated air. By this time, coalescence losses decreased as the aerosol were no longer in very close proximity and their movement was primarily influenced by the airborne infection isolation room airflow patterns. In the patient and health care worker's area away from the supply, the fresh air supply failed to reach this part of the room to quickly dilute the cough aerosol concentration. The exhaust was also found to have minimal effect upon cough aerosol removal, except for those areas with high exhaust velocities, very close to the exhaust grill. Within 5-20 s after a patient's cough, the aerosols tended to break up to form smaller sized aerosols of less than one micron diameter. They remained airborne and entrained back into the supply air stream, spreading into the entire room. The suspended aerosols resulted in the floating time of more than 21 s in the room due to one cough cycle. The duration of airborne contagion in the room and its prolonged exposure to the health care worker is likely to happen due to successive coughing cycles. Hence, the evaluated alternate airborne infection isolation room is not effective in removing at least 38% particles exposed to health care worker within the first second of a patient's cough.

  13. Does an oral care protocol reduce VAP in patients with a tracheostomy?

    PubMed

    Conley, Patricia; McKinsey, David; Graff, Jason; Ramsey, Anthony R

    2013-07-01

    Several studies have demonstrated that oral care with chlorhexidine gluconate (CHG) 0.12% solution reduces the incidence of ventilator-associated pneumonia (VAP) in mechanically ventilated patients with endotracheal tubes in the ICU. Minimal evidence shows the effectiveness of any oral care protocols in preventing VAP in mechanically ventilated patients with tracheostomies in a step-down or progressive care unit (PCU). To determine the effectiveness of an oral care protocol in reducing the VAP rate in mechanically ventilated patients with tracheostomies in the PCU. A 12-month prospective study was conducted on 75 mechanically ventilated patients who had tracheostomies. The oral care protocol consisted of tooth brushing with toothpaste and applying CHG 0.12% solution every 12 hours. At the conclusion of the study, the VAP rate in the study population was compared with the National Health and Safety Network (NHSN) report for 2009 benchmark of 1.5 per 1,000 ventilator days. After the oral care protocol was implemented in the PCU, the VAP rate was 1.1 per 1,000 ventilator days over 12 months, compared with the NHSN report for 2009 of 1.5 per 1,000 ventilator days. Tooth brushing with toothpaste and applying CHG 0.12% solution may be an effective oral care protocol to reduce the VAP rate in patients in PCUs with tracheostomies who are being mechanically ventilated.

  14. Experimental study of airflow characteristics of stratum ventilation in a multi-occupant room with comparison to mixing ventilation and displacement ventilation.

    PubMed

    Cheng, Y; Lin, Z

    2015-12-01

    The motivation of this study is stimulated by a lack of knowledge about the difference of airflow characteristics between a novel air distribution method [i.e., stratum ventilation (SV)] and conventional air distribution methods [i.e., mixing ventilation (MV) and displacement ventilation (DV)]. Detailed air velocity and temperature measurements were conducted in the occupied zone of a classroom with dimensions of 8.8 m (L) × 6.1 m (W) × 2.4 m (H). Turbulence intensity and power spectrum of velocity fluctuation were calculated using the measured data. Thermal comfort and cooling efficiency were also compared. The results show that in the occupied zone, the airflow characteristics among MV, DV, and SV are different. The turbulent airflow fluctuation is enhanced in this classroom with multiple thermal manikins due to thermal buoyancy and airflow mixing effect. Thermal comfort evaluations indicate that in comparison with MV and DV, a higher supply air temperature should be adopted for SV to achieve general thermal comfort with low draft risk. Comparison of the mean air temperatures in the occupied zone reveals that SV is of highest cooling efficiency, followed by DV and then MV. This study reports the unique profiles of flow, temperature, turbulence intensity, and power spectrum of stratum ventilation, which can have a number of implications for both knowledge and understanding of the flow characteristics in a stratum-ventilated room. With respect to the former, it expounds the fundamental characteristics of this air distribution method; and with respect to the latter, it reveals the mechanism of thermal comfort and energy saving under stratum ventilation. © 2015 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

  15. Healthy Zero Energy Buildings (HZEB) Program - Cross-Sectional Study of Contaminant Levels, Source, Strengths, and Ventilation Rates in Retail Stores

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Chan, Wanyu R.; Sidheswaran, Meera; Cohn, Sebastian

    2014-02-01

    This field study measured ventilation rates and indoor air quality parameters in 21 visits to retail stores in California. The data was collected to guide the development of new, science-based commercial building ventilation rate standards that balance the dual objectives of increasing energy efficiency and maintaining acceptable indoor air quality. Data collection occurred between September 2011 and March 2013. Three types of stores participated in this study: grocery stores, furniture/hardware stores, and apparel stores. Ventilation rates and indoor air contaminant concentrations were measured on a weekday, typically between 9 am and 6 pm. Ventilation rates measured using a tracer gasmore » decay method exceeded the minimum requirement of California’s Title 24 Standard in all but one store. Even though there was adequate ventilation according to Title 24, concentrations of formaldehyde, acetaldehyde, and acrolein exceeded the most stringent chronic health guidelines. Other indoor air contaminants measured included carbon dioxide (CO{sub 2}), carbon monoxide (CO), ozone (O{sub 3}), and particulate matter (PM). Concentrations of CO{sub 2} were kept low by adequate ventilation, and were assumed low also because the sampling occurred on a weekday when retail stores were less busy. CO concentrations were also low. The indoor-outdoor ratios of O{sub 3} showed that the first-order loss rate may vary by store trade types and also by ventilation mode (mechanical versus natural). Analysis of fine and ultrafine PM measurements showed that a substantial portion of the particle mass in grocery stores with cooking-related emissions was in particles less than 0.3 μm. Stores without cooking as an indoor source had PM size distributions that were more similar indoors and outdoors. The whole-building emission rates of volatile organic compounds (VOCs) and PM were estimated from the measured ventilation rates and indoor and outdoor contaminant concentrations. Mass balance models were then used to determine the ventilation rates, filtration strategies, or source reductions needed to maintain indoor contaminant concentrations below reference levels. Several scenarios of potential concern were considered: (i) formaldehyde levels in furniture/hardware stores, (ii) contaminants associated with cooking (e.g., PM, acrolein, and acetaldehyde) in grocery stores, and (iii) outdoor contaminants (e.g., PM and O{sub 3}) impacting stores that use natural ventilation. Estimated formaldehyde emission rates suggest that retail stores would need to ventilate at levels far exceeding the current Title 24 requirement to lower indoor concentrations below California’s stringent formaldehyde reference level. Given the high costs of providing ventilation but only modest chronic health benefit is expected, effective source control is an attractive alternative, as demonstrated by some retail stores in this study. Predictions showed that grocery stores need MERV 13 air filters, instead of MERV 8 filters that are more commonly used, to maintain indoor PM at levels that meet the chronic health standards for PM. Exposure to acrolein is a potential health concern in grocery stores, and should be addressed by increasing the use of kitchen range hoods or improving their contaminant removal efficiency. In stores that rely on natural ventilation, indoor PM can be a health concern if the stores are located in areas with high outdoor PM. This concern may be addressed by switching to mechanical ventilation when the outdoor air quality is poor, while continuing natural ventilation when outdoor air quality is good.« less

  16. Overview of existing regulations for ventilation requirements of enclosed vehicular parking facilities

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Krarti, M.; Ayari, A.M.

    1999-07-01

    This paper provides an overview of the current standards and regulations regarding the ventilation in enclosed parking facilities in the US and other countries. First, the paper discusses the emission rates of motor vehicle pollutants and their health effects. In particular, typical emission rates for different vehicle and fuel types are presented to highlight the effect of various parameters on the ventilation rate requirements for parking garages. In addition, the paper provides a brief description of some of the common ventilation problems reported in the literature for enclosed parking garages.

  17. Factors associated with respiration induced variability in cerebral blood flow velocity.

    PubMed Central

    Coughtrey, H; Rennie, J M; Evans, D H; Cole, T J

    1993-01-01

    A consecutive cohort of 73 very low birthweight infants was studied to determine the presence or absence of beat to beat variability in the velocity of blood flow in the cerebral circulation and its relation with respiration. One minute epochs of information included recordings of cerebral blood flow velocity estimated with Doppler ultrasound, blood pressure, spontaneous respiratory activity, and ventilator cycling. Fourier transformation was used to resolve the frequencies present within the one minute epochs and to classify the cerebral blood flow velocity as showing the presence or absence of any respiratory associated variability. A total of 249 recordings was made on days 1, 2, 3, and 7. Forty seven infants showed respiratory variability in cerebral blood flow velocity on 97 occasions, usually during the first day of life. The infants with respiratory associated variability were of lower gestational age and when the respiratory associated variability was present they were more likely to be ventilated and receiving higher inspired oxygen; these associations were shown to be independent of gestational age. There was no significant independent association with brain injury, cerebral blood flow velocity (cm/s), or blood pressure (mm Hg). The findings suggest that artificial ventilation may entrain normal respiratory associated variability in the cerebral circulation but do not provide evidence that it is harmful. PMID:8466269

  18. Comparison of flow and gas washout characteristics between pressure control and high-frequency percussive ventilation using a test lung

    PubMed Central

    Dutta, Rabijit; Xing, Tao; Swanson, Craig; Heltborg, Jeff; Murdoch, Gordon K

    2018-01-01

    Objective A comparison between flow and gas washout data for high-frequency percussive ventilation (HFPV) and pressure control ventilation (PCV) under similar conditions is currently not available. This bench study aims to compare and describe the flow and gas washout behavior of HFPV and PCV in a newly designed experimental setup and establish a framework for future clinical and animal studies. Approach We studied gas washout behavior using a newly designed experimental setup that is motivated by the multi-breath nitrogen washout measurements. In this procedure, a test lung was filled with nitrogen gas before it was connected to a ventilator. Pressure, volume, and oxygen concentrations were recorded under different compliance and resistance conditions. PCV was compared with two settings of HFPV, namely, HFPV-High and HFPV-Low, to simulate the different variations in its clinical application. In the HFPV-Low mode, the peak pressures and drive pressures of HFPV and PCV are matched, whereas in the HFPV-High mode, the mean airway pressures (MAP) are matched. Main results HFPV-Low mode delivers smaller tidal volume (VT) as compared to PCV under all lung conditions, whereas HFPV-High delivers a larger VT. HFPV-High provides rapid washout as compared to PCV under all lung conditions. HFPV-Low takes a longer time to wash out nitrogen except at a low compliance, where it expedites washout at a smaller VT and MAP compared to PCV washout. Significance Various flow parameters for HFPV and PCV are mathematically defined. A shorter washout time at a small VT in low compliant test lungs for HFPV could be regarded as a hypothesis for lung protective ventilation for animal or human lungs. PMID:29369819

  19. Comparison of flow and gas washout characteristics between pressure control and high-frequency percussive ventilation using a test lung.

    PubMed

    Dutta, Rabijit; Xing, Tao; Swanson, Craig; Heltborg, Jeff; Murdoch, Gordon K

    2018-03-15

    A comparison between flow and gas washout data for high-frequency percussive ventilation (HFPV) and pressure control ventilation (PCV) under similar conditions is currently not available. This bench study aims to compare and describe the flow and gas washout behavior of HFPV and PCV in a newly designed experimental setup and establish a framework for future clinical and animal studies. We studied gas washout behavior using a newly designed experimental setup that is motivated by the multi-breath nitrogen washout measurements. In this procedure, a test lung was filled with nitrogen gas before it was connected to a ventilator. Pressure, volume, and oxygen concentrations were recorded under different compliance and resistance conditions. PCV was compared with two settings of HFPV, namely, HFPV-High and HFPV-Low, to simulate the different variations in its clinical application. In the HFPV-Low mode, the peak pressures and drive pressures of HFPV and PCV are matched, whereas in the HFPV-High mode, the mean airway pressures (MAP) are matched. HFPV-Low mode delivers smaller tidal volume (V T ) as compared to PCV under all lung conditions, whereas HFPV-High delivers a larger V T . HFPV-High provides rapid washout as compared to PCV under all lung conditions. HFPV-Low takes a longer time to wash out nitrogen except at a low compliance, where it expedites washout at a smaller V T and MAP compared to PCV washout. Various flow parameters for HFPV and PCV are mathematically defined. A shorter washout time at a small V T in low compliant test lungs for HFPV could be regarded as a hypothesis for lung protective ventilation for animal or human lungs.

  20. Wind Tunnel Measurement of Turbulent and Advective Scalar Fluxes: A Case Study on Intersection Ventilation

    PubMed Central

    Kukačka, Libor; Nosek, Štĕpán; Kellnerová, Radka; Jurčáková, Klára; Jaňour, Zbyněk

    2012-01-01

    The objective of this study is to determine processes of pollution ventilation in the X-shaped street intersection in an idealized symmetric urban area for the changing approach flow direction. A unique experimental setup for simultaneous wind tunnel measurement of the flow velocity and the tracer gas concentration in a high temporal resolution is assembled. Advective horizontal and vertical scalar fluxes are computed from averaged measured velocity and concentration data within the street intersection. Vertical advective and turbulent scalar fluxes are computed from synchronized velocity and concentration signals measured in the plane above the intersection. All the results are obtained for five approach flow directions. The influence of the approach flow on the advective and turbulent fluxes is determined. The contribution of the advective and turbulent flux to the ventilation is discussed. Wind direction with the best dispersive conditions in the area is found. The quadrant analysis is applied to the synchronized signals of velocity and concentration fluctuation to determine events with the dominant contribution to the momentum flux and turbulent scalar flux. PMID:22649290

  1. Wind tunnel measurement of turbulent and advective scalar fluxes: a case study on intersection ventilation.

    PubMed

    Kukačka, Libor; Nosek, Štĕpán; Kellnerová, Radka; Jurčáková, Klára; Jaňour, Zbyněk

    2012-01-01

    The objective of this study is to determine processes of pollution ventilation in the X-shaped street intersection in an idealized symmetric urban area for the changing approach flow direction. A unique experimental setup for simultaneous wind tunnel measurement of the flow velocity and the tracer gas concentration in a high temporal resolution is assembled. Advective horizontal and vertical scalar fluxes are computed from averaged measured velocity and concentration data within the street intersection. Vertical advective and turbulent scalar fluxes are computed from synchronized velocity and concentration signals measured in the plane above the intersection. All the results are obtained for five approach flow directions. The influence of the approach flow on the advective and turbulent fluxes is determined. The contribution of the advective and turbulent flux to the ventilation is discussed. Wind direction with the best dispersive conditions in the area is found. The quadrant analysis is applied to the synchronized signals of velocity and concentration fluctuation to determine events with the dominant contribution to the momentum flux and turbulent scalar flux.

  2. Energy and Cost Associated with Ventilating Office Buildings in a Tropical Climate

    PubMed Central

    Rim, Donghyun; Schiavon, Stefano; Nazaroff, William W.

    2015-01-01

    Providing sufficient amounts of outdoor air to occupants is a critical building function for supporting occupant health, well-being and productivity. In tropical climates, high ventilation rates require substantial amounts of energy to cool and dehumidify supply air. This study evaluates the energy consumption and associated cost for thermally conditioning outdoor air provided for building ventilation in tropical climates, considering Singapore as an example locale. We investigated the influence on energy consumption and cost of the following factors: outdoor air temperature and humidity, ventilation rate (L/s per person), indoor air temperature and humidity, air conditioning system coefficient of performance (COP), and cost of electricity. Results show that dehumidification of outdoor air accounts for more than 80% of the energy needed for building ventilation in Singapore’s tropical climate. Improved system performance and/or a small increase in the indoor temperature set point would permit relatively large ventilation rates (such as 25 L/s per person) at modest or no cost increment. Overall, even in a thermally demanding tropical climate, the energy cost associated with increasing ventilation rate up to 25 L/s per person is less than 1% of the wages of an office worker in an advanced economy like Singapore’s. This result implies that the benefits of increasing outdoor air ventilation rate up to 25 L/s per person — which is suggested to provide for productivity increases, lower sick building syndrome symptom prevalence, and reduced sick leave — can be much larger than the incremental cost of ventilation. PMID:25822504

  3. Assessment of Factors Related to Auto-PEEP.

    PubMed

    Natalini, Giuseppe; Tuzzo, Daniele; Rosano, Antonio; Testa, Marco; Grazioli, Michele; Pennestrì, Vincenzo; Amodeo, Guido; Marsilia, Paolo F; Tinnirello, Andrea; Berruto, Francesco; Fiorillo, Marialinda; Filippini, Matteo; Peratoner, Alberto; Minelli, Cosetta; Bernardini, Achille

    2016-02-01

    Previous physiological studies have identified factors that are involved in auto-PEEP generation. In our study, we examined how much auto-PEEP is generated from factors that are involved in its development. One hundred eighty-six subjects undergoing controlled mechanical ventilation with persistent expiratory flow at the beginning of each inspiration were enrolled in the study. Volume-controlled continuous mandatory ventilation with PEEP of 0 cm H2O was applied while maintaining the ventilator setting as chosen by the attending physician. End-expiratory and end-inspiratory airway occlusion maneuvers were performed to calculate respiratory mechanics, and tidal flow limitation was assessed by a maneuver of manual compression of the abdomen. The variable with the strongest effect on auto-PEEP was flow limitation, which was associated with an increase of 2.4 cm H2O in auto-PEEP values. Moreover, auto-PEEP values were directly related to resistance of the respiratory system and body mass index and inversely related to expiratory time/time constant. Variables that were associated with the breathing pattern (tidal volume, frequency minute ventilation, and expiratory time) did not show any relationship with auto-PEEP values. The risk of auto-PEEP ≥5 cm H2O was increased by flow limitation (adjusted odds ratio 17; 95% CI: 6-56.2), expiratory time/time constant ratio <1.85 (12.6; 4.7-39.6), respiratory system resistance >15 cm H2O/L s (3; 1.3-6.9), age >65 y (2.8; 1.2-6.5), and body mass index >26 kg/m(2) (2.6; 1.1-6.1). Flow limitation, expiratory time/time constant, resistance of the respiratory system, and obesity are the most important variables that affect auto-PEEP values. Frequency expiratory time, tidal volume, and minute ventilation were not independently associated with auto-PEEP. Therapeutic strategies aimed at reducing auto-PEEP and its adverse effects should be primarily oriented to the variables that mainly affect auto-PEEP values. Copyright © 2016 by Daedalus Enterprises.

  4. Effect of metronome rates on the quality of bag-mask ventilation during metronome-guided 30:2 cardiopulmonary resuscitation: A randomized simulation study.

    PubMed

    Na, Ji Ung; Han, Sang Kuk; Choi, Pil Cho; Shin, Dong Hyuk

    2017-01-01

    Metronome guidance is a feasible and effective feedback technique to improve the quality of cardiopulmonary resuscitation (CPR). The rate of the metronome should be set between 100 to 120 ticks/minute and the speed of ventilation may have crucial effect on the quality of ventilation. We compared three different metronome rates (100, 110, 120 ticks/minute) to investigate its effect on the quality of ventilation during metronome-guided 30:2 CPR. This is a prospective, randomized, crossover observational study using a RespiTrainer○ r . To simulate 30 chest compressions, one investigator counted from 1 to 30 in cadence with the metronome rate (1 count for every 1 tick), and the participant performed 2 consecutive ventilations immediately following the counting of 30. Thirty physicians performed 5 sets of 2 consecutive (total 10) bag-mask ventilations for each metronome rate. Participants were instructed to squeeze the bag over 2 ticks (1.0 to 1.2 seconds depending on the rate of metronome) and deflate the bag over 2 ticks. The sequence of three different metronome rates was randomized. Mean tidal volume significantly decreased as the metronome rate was increased from 110 ticks/minute to 120 ticks/minute (343±84 mL vs. 294±90 mL, P =0.004). Peak airway pressure significantly increased as metronome rate increased from 100 ticks/minute to 110 ticks/minute (18.7 vs. 21.6 mmHg, P =0.006). In metronome-guided 30:2 CPR, a higher metronome rate may adversely affect the quality of bag-mask ventilations. In cases of cardiac arrest where adequate ventilation support is necessary, 100 ticks/minute may be better than 110 or 120 ticks/minute to deliver adequate tidal volume during audio tone guided 30:2 CPR.

  5. Effect of metronome rates on the quality of bag-mask ventilation during metronome-guided 30:2 cardiopulmonary resuscitation: A randomized simulation study

    PubMed Central

    Na, Ji Ung; Han, Sang Kuk; Choi, Pil Cho; Shin, Dong Hyuk

    2017-01-01

    BACKGROUND: Metronome guidance is a feasible and effective feedback technique to improve the quality of cardiopulmonary resuscitation (CPR). The rate of the metronome should be set between 100 to 120 ticks/minute and the speed of ventilation may have crucial effect on the quality of ventilation. We compared three different metronome rates (100, 110, 120 ticks/minute) to investigate its effect on the quality of ventilation during metronome-guided 30:2 CPR. METHODS: This is a prospective, randomized, crossover observational study using a RespiTrainer○r. To simulate 30 chest compressions, one investigator counted from 1 to 30 in cadence with the metronome rate (1 count for every 1 tick), and the participant performed 2 consecutive ventilations immediately following the counting of 30. Thirty physicians performed 5 sets of 2 consecutive (total 10) bag-mask ventilations for each metronome rate. Participants were instructed to squeeze the bag over 2 ticks (1.0 to 1.2 seconds depending on the rate of metronome) and deflate the bag over 2 ticks. The sequence of three different metronome rates was randomized. RESULTS: Mean tidal volume significantly decreased as the metronome rate was increased from 110 ticks/minute to 120 ticks/minute (343±84 mL vs. 294±90 mL, P=0.004). Peak airway pressure significantly increased as metronome rate increased from 100 ticks/minute to 110 ticks/minute (18.7 vs. 21.6 mmHg, P=0.006). CONCLUSION: In metronome-guided 30:2 CPR, a higher metronome rate may adversely affect the quality of bag-mask ventilations. In cases of cardiac arrest where adequate ventilation support is necessary, 100 ticks/minute may be better than 110 or 120 ticks/minute to deliver adequate tidal volume during audio tone guided 30:2 CPR. PMID:28458759

  6. High tidal volume ventilation induces NOS2 and impairs cAMP- dependent air space fluid clearance.

    PubMed

    Frank, James A; Pittet, Jean-Francois; Lee, Hyon; Godzich, Micaela; Matthay, Michael A

    2003-05-01

    Tidal volume reduction during mechanical ventilation reduces mortality in patients with acute lung injury and the acute respiratory distress syndrome. To determine the mechanisms underlying the protective effect of low tidal volume ventilation, we studied the time course and reversibility of ventilator-induced changes in permeability and distal air space edema fluid clearance in a rat model of ventilator-induced lung injury. Anesthetized rats were ventilated with a high tidal volume (30 ml/kg) or with a high tidal volume followed by ventilation with a low tidal volume of 6 ml/kg. Endothelial and epithelial protein permeability were significantly increased after high tidal volume ventilation but returned to baseline levels when tidal volume was reduced. The basal distal air space fluid clearance (AFC) rate decreased by 43% (P < 0.05) after 1 h of high tidal volume but returned to the preventilation rate 2 h after tidal volume was reduced. Not all of the effects of high tidal volume ventilation were reversible. The cAMP-dependent AFC rate after 1 h of 30 ml/kg ventilation was significantly reduced and was not restored when tidal volume was reduced. High tidal volume ventilation also increased lung inducible nitric oxide synthase (NOS2) expression and air space total nitrite at 3 h. Inhibition of NOS2 activity preserved cAMP-dependent AFC. Because air space edema fluid inactivates surfactant and reduces ventilated lung volume, the reduction of cAMP-dependent AFC by reactive nitrogen species may be an important mechanism of clinical ventilator-associated lung injury.

  7. Study on the Enhanced Operation of Self-Ventilation-Based Coupling System for Domestic Wastewater Treatment

    NASA Astrophysics Data System (ADS)

    Kong, Lingwei; Wang, Lu; Zhang, Yi; Mei, Rongwu; Zhang, Yu

    2018-06-01

    In this study, a new coupling system of biological filter bed and subsurface-flow constructed wetland based on the self-ventilation network was proposed, and the comparative pollutant removal efficiency at low and high influent concentration of the pilot coupling system with different substrates configurations were investigated. The study found that: The comparison system (b) had better removal rates than that of the original system (a), and the removal rate when treating low influent concentration was 74.10%, 94.14%, 73.57% and 69.53%, while in high influent concentration case was 81.30%, 90.28%, 88.57% and 75.36% for CODCr , NH4+ -N, TN and TP, respectively. The removal of the above main water indexes of the comparison system (b) promoted by 11.00%, 11.55%, 2.69% and 8.09% respectively in low influent concentration case and 4.20%, 9.20%, 7.66% and 13.61% respectively in high influent concentration case when comparing to the original system (a), which showed that the optimized configuration of various kinds of substrates was significant and was more beneficial to the degradation and removal of pollutants. The adsorption and interception function of substrates in the constructed wetland was the main way of phosphorus removal. The function of self-ventilation ensured the amount of DO in the coupling system, making the phosphorus removal was less affected comparing to structure of traditional wetland.

  8. Effect of sedation with detomidine and butorphanol on pulmonary gas exchange in the horse

    PubMed Central

    Nyman, Görel; Marntell, Stina; Edner, Anna; Funkquist, Pia; Morgan, Karin; Hedenstierna, Göran

    2009-01-01

    Background Sedation with α2-agonists in the horse is reported to be accompanied by impairment of arterial oxygenation. The present study was undertaken to investigate pulmonary gas exchange using the Multiple Inert Gas Elimination Technique (MIGET), during sedation with the α2-agonist detomidine alone and in combination with the opioid butorphanol. Methods Seven Standardbred trotter horses aged 3–7 years and weighing 380–520 kg, were studied. The protocol consisted of three consecutive measurements; in the unsedated horse, after intravenous administration of detomidine (0.02 mg/kg) and after subsequent butorphanol administration (0.025 mg/kg). Pulmonary function and haemodynamic effects were investigated. The distribution of ventilation-perfusion ratios (VA/Q) was estimated with MIGET. Results During detomidine sedation, arterial oxygen tension (PaO2) decreased (12.8 ± 0.7 to 10.8 ± 1.2 kPa) and arterial carbon dioxide tension (PaCO2) increased (5.9 ± 0.3 to 6.1 ± 0.2 kPa) compared to measurements in the unsedated horse. Mismatch between ventilation and perfusion in the lungs was evident, but no increase in intrapulmonary shunt could be detected. Respiratory rate and minute ventilation did not change. Heart rate and cardiac output decreased, while pulmonary and systemic blood pressure and vascular resistance increased. Addition of butorphanol resulted in a significant decrease in ventilation and increase in PaCO2. Alveolar-arterial oxygen content difference P(A-a)O2 remained impaired after butorphanol administration, the VA/Q distribution improved as the decreased ventilation and persistent low blood flow was well matched. Also after subsequent butorphanol no increase in intrapulmonary shunt was evident. Conclusion The results of the present study suggest that both pulmonary and cardiovascular factors contribute to the impaired pulmonary gas exchange during detomidine and butorphanol sedation in the horse. PMID:19422714

  9. Use of ventilator associated pneumonia bundle and statistical process control chart to decrease VAP rate in Syria.

    PubMed

    Alsadat, Reem; Al-Bardan, Hussam; Mazloum, Mona N; Shamah, Asem A; Eltayeb, Mohamed F E; Marie, Ali; Dakkak, Abdulrahman; Naes, Ola; Esber, Faten; Betelmal, Ibrahim; Kherallah, Mazen

    2012-10-01

    Implementation of ventilator associated pneumonia (VAP) bundle as a performance improvement project in the critical care units for all mechanically ventilated patients aiming to decrease the VAP rates. VAP bundle was implemented in 4 teaching hospitals after educational sessions and compliance rates along with VAP rates were monitored using statistical process control charts. VAP bundle compliance rates were steadily increasing from 33 to 80% in hospital 1, from 33 to 86% in hospital 2 and from 83 to 100% in hospital 3 during the study period. The VAP bundle was not applied in hospital 4 therefore no data was available. A target level of 95% was reached only in hospital 3. This correlated with a decrease in VAP rates from 30 to 6.4 per 1000 ventilator days in hospital 1, from 12 to 4.9 per 1000 ventilator days in hospital 3, whereas VAP rate failed to decrease in hospital 2 (despite better compliance) and it remained high around 33 per 1000 ventilator days in hospital 4 where VAP bundle was not implemented. VAP bundle has performed differently in different hospitals in our study. Prevention of VAP requires a multidimensional strategy that includes strict infection control interventions, VAP bundle implementation, process and outcome surveillance and education.

  10. Metronome improves compression and ventilation rates during CPR on a manikin in a randomized trial.

    PubMed

    Kern, Karl B; Stickney, Ronald E; Gallison, Leanne; Smith, Robert E

    2010-02-01

    We hypothesized that a unique tock and voice metronome could prevent both suboptimal chest compression rates and hyperventilation. A prospective, randomized, parallel design study involving 34 pairs of paid firefighter/emergency medical technicians (EMTs) performing two-rescuer CPR using a Laerdal SkillReporter Resusci Anne manikin with and without metronome guidance was performed. Each CPR session consisted of 2 min of 30:2 CPR with an unsecured airway, then 4 min of CPR with a secured airway (continuous compressions at 100 min(-1) with 8-10 ventilations/min), repeated after the rescuers switched roles. The metronome provided "tock" prompts for compressions, transition prompts between compressions and ventilations, and a spoken "ventilate" prompt. During CPR with a bag/valve/mask the target compression rate of 90-110 min(-1) was achieved in 5/34 CPR sessions (15%) for the control group and 34/34 sessions (100%) for the metronome group (p<0.001). An excessive ventilation rate was not observed in either the metronome or control group during CPR with a bag/valve/mask. During CPR with a bag/endotracheal tube, the target of both a compression rate of 90-110 min(-1) and a ventilation rate of 8-11 min(-1) was achieved in 3/34 CPR sessions (9%) for the control group and 33/34 sessions (97%) for the metronome group (p<0.001). Metronome use with the secured airway scenario significantly decreased the incidence of over-ventilation (11/34 EMT pairs vs. 0/34 EMT pairs; p<0.001). A unique combination tock and voice prompting metronome was effective at directing correct chest compression and ventilation rates both before and after intubation. Copyright 2009 Elsevier Ireland Ltd. All rights reserved.

  11. Surfactant and pulmonary blood flow distributions following treatment of premature lambs with natural surfactant.

    PubMed Central

    Jobe, A; Ikegami, M; Jacobs, H; Jones, S

    1984-01-01

    Prematurely delivered lambs were treated with radiolabeled natural surfactant by either tracheal instillation at birth and before the onset of mechanical ventilation, or after 23 +/- 1 (+/- SE) min of mechanical ventilation. Right ventricular blood flow distributions, left ventricular outputs, and left-to-right ductal shunts were measured with radiolabeled microspheres. After sacrifice, the lungs of lambs receiving surfactant at birth inflated uniformly with constant distending pressure while the lungs of lambs treated after a period of ventilation had aerated, partially aerated, and atelectatic areas. All lungs were divided into pieces which were weighed and catalogued as to location. The amount of radiolabeled surfactant and microsphere-associated radioactivity in each piece of lung was quantified. Surfactant was relatively homogenously distributed to pieces of lung from lambs that were treated with surfactant at birth; 48% of lung pieces received amounts of surfactant within +/- 25% of the mean value. Surfactant was preferentially recovered from the aerated pieces of lungs of lambs treated after a period of mechanical ventilation, and the distribution of surfactant to these lungs was very nonhomogeneous. Right ventricular blood flow distributions to the lungs were quite homogeneous in both groups of lambs. However, in 8 of 12 lambs, pulmonary blood flow was preferentially directed away from those pieces of lung that received relatively large amounts of surfactant and toward pieces of lung that received less surfactant. This acute redirection of pulmonary blood flow distribution may result from the local changes in compliances within the lung following surfactant instillation. PMID:6546766

  12. Effect Of Ventilation On Chronic Health Risks In Schools And Offices

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Parthasarathy, Srinandini; Fisk, William J.; McKone, Thomas E.

    This study provides a risk assessment for chronic health risks from inhalation exposure to indoor air pollutants in offices and schools with a focus how ventilation impacts exposures to, and risks from, volatile organic compounds (VOCs) and particulate matter (PM2.5). We estimate how much health risks could change with varying ventilation rates under two scenarios: (i) halving the measured ventilation rates and (ii) doubling the measured ventilation rates. For the hazard characterization we draw upon prior papers that identified pollutants potentially affecting health with indoor air concentrations responsive to changes in ventilation rates. For exposure assessment we determine representative concentrationsmore » of pollutants using data available in current literature and model changes in exposures with changes in ventilation rates. As a metric of disease burden, we use disability adjusted life years (DALYs) to address both cancer and non-cancer effects. We also compare exposures to guidelines published by regulatory agencies to assess chronic health risks. Chronic health risks are driven primarily by particulate matter exposure, with an estimated baseline disease burden of 150 DALYs per 100,000 people in offices and 140 DALYs per 100,000 people in schools. Study results show that PM2.5-related DALYs are not very sensitive to changes in ventilation rates. Filtration is more effective at controlling PM2.5 concentrations and health effects. Non-cancer health effects contribute only a small fraction of the overall chronic health burden of populations in offices and schools (<1 DALY per 100,000 people). Cancer health effects dominate the disease burden in schools (3 DALYs per 100,000) and offices (5 DALYs per 100,000), with formaldehyde being the primary risk driver. In spite of large uncertainties in toxicological data and dose-response modeling, our results support the finding that ventilation rate changes do not have significant impacts on estimated chronic disease burdens. Median estimates of DALYs are approximately doubled when the ventilation rates are halved and there is little reduction in health risks associated with doubling ventilation rates, but the very low baseline disease burden from the indoor exposures we considered makes this unremarkable. In exploring the full range of exposure concentrations, to find the fraction exceeding the Office of Environmental Health and Hazard Assessment’s (OEHHAs) chronic reference exposure levels (cRELs) and United States Environmental Protection Agency’s (USEPA) chronic reference dose (RfD) we found only minor shifts in exposure safety margins when ventilation was doubled or halved. We combined our exposure estimates with cancer potency factors published by OEHHA and USEPA to determine that the annual excess cancer risk per capita are below 1 in a million under all ventilation rate scenarios for individual pollutants. The results indicate that chronic health risks (cancer and non-cancer) associated with VOC and PM2.5 exposure in offices and schools are low and thus the chronic disease burden or health benefits of ventilation changes are likely to be well below both the level of detection by health surveillance studies and the level of regulatory thresholds.« less

  13. Summary of human responses to ventilation.

    PubMed

    Seppänen, O A; Fisk, W J

    2004-01-01

    It is known that ventilation is necessary to remove indoor-generated pollutants from indoor air or dilute their concentration to acceptable levels. But as the limit values of all pollutants are not known the exact determination of required ventilation rates based on pollutant concentrations is seldom possible. The selection of ventilation rates has to be based also on epidemiological research, laboratory and field experiments and experience. The existing literature indicates that ventilation has a significant impact on several important human outcomes including: (1) communicable respiratory illnesses; (2) sick building syndrome symptoms; (3) task performance and productivity, and (4) perceived air quality (PAQ) among occupants or sensory panels (5) respiratory allergies and asthma. In many studies, prevalence of sick building syndrome symptoms has also been associated with characteristics of HVAC-systems. Often the prevalence of SBS symptoms is higher in air-conditioned buildings than in naturally ventilated buildings. The evidence suggests that better hygiene, commissioning, operation and maintenance of air handling systems may be particularly important for reducing the negative effects of HVAC systems. Ventilation may also have harmful effects on indoor air quality and climate if not properly designed, installed, maintained and operated. Ventilation may bring indoors harmful substances or deteriorate indoor environment. Ventilation interacts also with the building envelope and may deteriorate the structures of the building. Ventilation changes the pressure differences across the structures of building and may cause or prevent infiltration of pollutants from structures or adjacent spaces. Ventilation is also in many cases used to control the thermal environment or humidity in buildings. The paper summarises the current knowledge on positive and negative effects of ventilation on health and other human responses. The focus is on office-type working environment and residential buildings. The review shows that ventilation has various positive impacts on health and productivity of building occupants. Ventilation reduces the prevalence of airborne infectious diseases and thus the number of sick leave days. In office environment a ventilation rate up to 20-25 L/s per person seem to decrease the prevalence of SBS-symptoms. Air conditioning systems may increase the prevalence of SBS-symptoms relative to natural ventilation if not clean. In residential buildings the air change rate in cold climates should not be below app. 0.5 ach. Ventilation systems may cause pressure differences over the building envelope and bring harmful pollutants indoors.

  14. Inadequate ventilation for nosocomial tuberculosis prevention in public hospitals in Central Thailand.

    PubMed

    Jiamjarasrangsi, W; Bualert, S; Chongthaleong, A; Chaindamporn, A; Udomsantisuk, N; Euasamarnjit, W

    2009-04-01

    Forty-two community and general hospitals in central Thailand. To examine the adequacy of indoor ventilation for nosocomial tuberculosis (TB) prevention in public hospitals in central Thailand. A cross-sectional survey was conducted among 323 patient care and ancillary areas in the target hospitals. Data on indoor ventilation rate were collected by the tracer gas method and reported as air changes per hour (ACH). The adequacy of the measured ventilation rates were then determined by comparison with the international recommended standard values. Indoor ventilation rates were inadequate in almost half of the studied areas (144/323, 44.6%). The inadequacy was particularly serious in the emergency rooms (ERs) and radiological areas, where 73.8% (31/42 each) of the rooms had ACH below the recommended standards. Detailed analysis showed that most of the rooms with natural ventilation had air exchange rates that exceeded the recommended standards, while the opposite was the case for rooms with air-conditioning, particularly the window or wall-mount type. Indoor ventilation in high-risk nosocomial TB areas in public hospitals in Thailand was inadequate due to the installation of air-conditioning systems in modern buildings.

  15. Effectiveness and predictors of failure of noninvasive mechanical ventilation in acute respiratory failure.

    PubMed

    Martín-González, F; González-Robledo, J; Sánchez-Hernández, F; Moreno-García, M N; Barreda-Mellado, I

    2016-01-01

    To assess the effectiveness and identify predictors of failure of noninvasive ventilation. A retrospective, longitudinal descriptive study was made. Adult patients with acute respiratory failure. A total of 410 consecutive patients with noninvasive ventilation treated in an Intensive Care Unit of a tertiary university hospital from 2006 to 2011. Noninvasive ventilation. Demographic variables and clinical and laboratory test parameters at the start and two hours after the start of noninvasive ventilation. Evolution during admission to the Unit and until hospital discharge. The failure rate was 50%, with an overall mortality rate of 33%. A total of 156 patients had hypoxemic respiratory failure, 87 postextubation respiratory failure, 78 exacerbation of chronic obstructive pulmonary disease, 61 hypercapnic respiratory failure without chronic obstructive pulmonary disease, and 28 had acute pulmonary edema. The failure rates were 74%, 54%, 27%, 31% and 21%, respectively. The etiology of respiratory failure, serum bilirubin at the start, APACHEII score, radiological findings, the need for sedation to tolerate noninvasive ventilation, changes in level of consciousness, PaO2/FIO2 ratio, respiratory rate and heart rate from the start and two hours after the start of noninvasive ventilation were independently associated to failure. The effectiveness of noninvasive ventilation varies according to the etiology of respiratory failure. Its use in hypoxemic respiratory failure and postextubation respiratory failure should be assessed individually. Predictors of failure could be useful to prevent delayed intubation. Copyright © 2015 Elsevier España, S.L.U. and SEMICYUC. All rights reserved.

  16. Fluid mechanical dispersion of airborne pollutants inside urban street canyons subjecting to multi-component ventilation and unstable thermal stratifications.

    PubMed

    Mei, Shuo-Jun; Liu, Cheng-Wei; Liu, Di; Zhao, Fu-Yun; Wang, Han-Qing; Li, Xiao-Hong

    2016-09-15

    The pedestrian level pollutant transport in street canyons with multiple aspect ratios (H/W) is numerically investigated in the present work, regarding of various unstable thermal stratification scenarios and plain surrounding. Non-isothermal turbulent wind flow, temperature field and pollutant spread within and above the street canyons are solved by the realizable k-ε turbulence model along with the enhanced wall treatment. One-vortex flow regime is observed for shallow canyons with H/W=0.5, whereas multi-vortex flow regime is observed for deep canyons with H/W=2.0. Both one-vortex and multi-vortex regimes could be observed for the street canyons with H/W=1.0, where the secondary vortex could be initiated by the flow separation and intensified by unstable thermal stratification. Air exchange rate (AER) and pollutant retention time are adopted to respectively evaluate the street canyon ventilation and pollutant removal performance. A second-order polynomial functional relationship is established between AER and Richardson number (Ri). Similar functional relationship could be established between retention time and Ri, and it is only valid for canyons with one-vortex flow regime. In addition, retention time could be prolonged abruptly for canyons with multi-vortex flow regime. Very weak secondary vortex is presented at the ground level of deep canyons with mild stratification, where pollutants are highly accumulated. However, with the decrease of Ri, pollutant concentration adjacent to the ground reduces accordingly. Present research could be applied to guide the urban design and city planning for enhancing pedestrian environment. Copyright © 2016 Elsevier B.V. All rights reserved.

  17. Characterization of Ventilatory Modes in Dragonfly Nymph

    NASA Astrophysics Data System (ADS)

    Roh, Chris; Saxton-Fox, Theresa; Gharib, Morteza

    2013-11-01

    A dragonfly nymph's highly modified hindgut has multiple ventilatory modes: hyperventilation (i.e. jet propulsion), gulping ventilation (extended expiratory phase) and normal ventilation. Each mode involves dynamic manipulation of the exit diameter and pressure. To study the different fluid dynamics associated with the three modes, Anisopteran larvae of the family Aeshnidae were tethered onto a rod for flow visualization. The result showed distinct flow structures. The hyperventilation showed a highly turbulent and powerful jet that occurred at high frequency. The gulping ventilation produced a single vortex at a moderate frequency. The normal ventilation showed two distinct vortices, a low-Reynolds number vortex, followed by a high-Reynolds number vortex. Furthermore, a correlation of the formation of the vortices with the movement of the sternum showed that the dragonfly is actively controlling the timing and the speed of the vortices to have them at equal distance from the jet exit at the onset of inspiration. This behavior prevents inspiration of the oxygen deficient expirated water, resulting in the maximization of the oxygen intake. Supported by NSF GRFP.

  18. Generator stator core vent duct spacer posts

    DOEpatents

    Griffith, John Wesley; Tong, Wei

    2003-06-24

    Generator stator cores are constructed by stacking many layers of magnetic laminations. Ventilation ducts may be inserted between these layers by inserting spacers into the core stack. The ventilation ducts allow for the passage of cooling gas through the core during operation. The spacers or spacer posts are positioned between groups of the magnetic laminations to define the ventilation ducts. The spacer posts are secured with longitudinal axes thereof substantially parallel to the core axis. With this structure, core tightness can be assured while maximizing ventilation duct cross section for gas flow and minimizing magnetic loss in the spacers.

  19. Determining the ventilation and aerosol deposition rates from routine indoor-air measurements.

    PubMed

    Halios, Christos H; Helmis, Costas G; Deligianni, Katerina; Vratolis, Sterios; Eleftheriadis, Konstantinos

    2014-01-01

    Measurement of air exchange rate provides critical information in energy and indoor-air quality studies. Continuous measurement of ventilation rates is a rather costly exercise and requires specific instrumentation. In this work, an alternative methodology is proposed and tested, where the air exchange rate is calculated by utilizing indoor and outdoor routine measurements of a common pollutant such as SO2, whereas the uncertainties induced in the calculations are analytically determined. The application of this methodology is demonstrated, for three residential microenvironments in Athens, Greece, and the results are also compared against ventilation rates calculated from differential pressure measurements. The calculated time resolved ventilation rates were applied to the mass balance equation to estimate the particle loss rate which was found to agree with literature values at an average of 0.50 h(-1). The proposed method was further evaluated by applying a mass balance numerical model for the calculation of the indoor aerosol number concentrations, using the previously calculated ventilation rate, the outdoor measured number concentrations and the particle loss rates as input values. The model results for the indoors' concentrations were found to be compared well with the experimentally measured values.

  20. Performance of Leak Compensation in All-Age ICU Ventilators During Volume-Targeted Neonatal Ventilation: A Lung Model Study.

    PubMed

    Itagaki, Taiga; Bennett, Desmond J; Chenelle, Christopher T; Fisher, Daniel F; Kacmarek, Robert M

    2017-01-01

    Volume-targeted ventilation is increasingly used in low birthweight infants because of the potential for reducing volutrauma and avoiding hypocapnea. However, it is not known what level of air leak is acceptable during neonatal volume-targeted ventilation when leak compensation is activated concurrently. Four ICU ventilators (Servo-i, PB980, V500, and Avea) were compared in available invasive volume-targeted ventilation modes (pressure control continuous spontaneous ventilation [PC-CSV] and pressure control continuous mandatory ventilation [PC-CMV]). The Servo-i and PB980 were tested with (+) and without (-) their proximal flow sensor. The V500 and Avea were tested with their proximal flow sensor as indicated by their manufacturers. An ASL 5000 lung model was used to simulate 4 neonatal scenarios (body weight 0.5, 1, 2, and 4 kg). The ASL 5000 was ventilated via an endotracheal tube with 3 different leaks. Two minutes of data were collected after each change in leak level, and the asynchrony index was calculated. Tidal volume (V T ) before and after the change in leak was assessed. The differences in delivered V T between before and after the change in leak were within ±5% in all scenarios with the PB980 (-/+) and V500. With the Servo-i (-/+), baseline V T was ≥10% greater than set V T during PC-CSV, and delivered V T markedly changed with leak. The Avea demonstrated persistent high V T in all leak scenarios. Across all ventilators, the median asynchrony index was 1% (interquartile range 0-27%) in PC-CSV and 1.8% (0-45%) in PC-CMV. The median asynchrony index was significantly higher in the Servo-i (-/+) than in the PB980 (-/+) and V500 in 1 and 2 kg scenarios during PC-CSV and PC-CMV. The PB980 and V500 were the only ventilators to acclimate to all leak scenarios and achieve targeted V T . Further clinical investigation is needed to validate the use of leak compensation during neonatal volume-targeted ventilation. Copyright © 2017 by Daedalus Enterprises.

  1. High resolution pCO2 monitoring reveals ventilation of Bunker Cave (NW Germany) and its impact on speleothem growth

    NASA Astrophysics Data System (ADS)

    Riechelmann, Sylvia; Breitenbach, Sebastian F. M.; Schröder-Ritzrau, Andrea; Immenhauser, Adrian

    2017-04-01

    Understanding the environmental processes that influence geochemical proxies archived in speleothems depends critically on detailed cave monitoring. Cave air pCO2 is one of the most important factors controlling speleothem growth. The pCO2 concentration of cave air depends on (i) the productivity of its source(s), (ii) CO2-transport dynamics through the epikarst and (iii) cave ventilation processes. We monitored the pCO2 concentration ca. 100 m from the lower entrance of the Bunker-Emst-Cave system (NW Germany) with a CORA CO2-logger at a two-hourly resolution between April 2012 and February 2014. Near-atmospheric minimum pCO2 concentrations of 408 ppm are observed in winter, while higher values up to 811 ppm are recorded in summer. Higher summer concentrations are due to increased plant and soil microbial activity, resulting in elevated CO2 in the soil, which is transferred to the cave with infiltrating water. Generally, the front passages of Bunker Cave are well ventilated. Besides the seasonal pattern, pCO2 concentrations vary at diurnal scale. Correlations of pCO2 with the temperature difference between surface and cave air are positive during summer and negative in winter, with no clear pattern for spring and autumn months. Thus, Bunker Cave ventilation is driven by temperature and density differences between cave and surface air, with two entrances at different elevations allowing dynamic ventilation. During summer, relatively cooler cave air flows from the upper to the lower entrance, while in winter this pattern is reversed due to ascending warm cave air. The situation is further complicated by preferential south/southwestern winds that point directly on the cave entrances. Thus, cave ventilation is frequently disturbed, especially during periods of higher wind speed. Modern ventilation systematics only developed when the two cave entrances were artificially opened (1863 and 1926). Before that, ventilation was restricted and cave pCO2 concentrations were presumably higher under natural conditions. Thus, the present-day ventilation system of Bunker Cave is not a direct analogue for natural ventilation conditions. pCO2 concentrations are relatively low compared to other caves, and because the difference between summer and winter pCO2 is relatively low (max. 400 ppm), a significant effect on seasonal speleothem growth rate is unlikely. In case of Bunker Cave, it is rather a combination of the availability of water, and thus of calcium and carbonate ions and pCO2 concentrations that allow higher carbonate precipitation during winter than summer. Holocene speleothems from Bunker Cave display relatively slow growth rates. We suggest that - with absence of major entrances to the cave system during the Holocene - ventilation was minimal and pCO2 concentrations significantly higher, making winterly water supply the governing factor regulating speleothem growth. Thus, stalagmites from Bunker Cave are likely to record a climatic signal biased towards the winter season.

  2. Tracheostomy and invasive mechanical ventilation in amyotrophic lateral sclerosis: decision-making factors and survival analysis.

    PubMed

    Kimura, Fumiharu

    2016-04-28

    Invasive and/or non-invasive mechanical ventilation are most important options of respiratory management in amyotrophic lateral sclerosis. We evaluated the frequency, clinical characteristics, decision-making factors about ventilation and survival analysis of 190 people with amyotrophic lateral sclerosis patients from 1990 until 2013. Thirty-one percentage of patients underwent tracheostomy invasive ventilation with the rate increasing more than the past 20 years. The ratio of tracheostomy invasive ventilation in patients >65 years old was significantly increased after 2000 (25%) as compared to before (10%). After 2010, the standard use of non-invasive ventilation showed a tendency to reduce the frequency of tracheostomy invasive ventilation. Mechanical ventilation prolonged median survival (75 months in tracheostomy invasive ventilation, 43 months in non-invasive ventilation vs natural course, 32 months). The life-extending effects by tracheostomy invasive ventilation were longer in younger patients ≤65 years old at the time of ventilation support than in older patients. Presence of partners and care at home were associated with better survival. Following factors related to the decision to perform tracheostomy invasive ventilation: patients ≤65 years old: greater use of non-invasive ventilation: presence of a spouse: faster tracheostomy: higher progression rate; and preserved motor functions. No patients who underwent tracheostomy invasive ventilation died from a decision to withdraw mechanical ventilation. The present study provides factors related to decision-making process and survival after tracheostomy and help clinicians and family members to expand the knowledge about ventilation.

  3. Respiration of resting honeybees

    PubMed Central

    Kovac, Helmut; Stabentheiner, Anton; Hetz, Stefan K.; Petz, Markus; Crailsheim, Karl

    2011-01-01

    The relation between the respiratory activity of resting honeybees and ambient temperature (Ta) was investigated in the range of 5–40 °C. Bees were kept in a temperature controlled flow through respirometer chamber where their locomotor and endothermic activity, as well as abdominal ventilatory movements was recorded by infrared thermography. Surprisingly, true resting bees were often weakly endothermic (thorax surface up to 2.8 °C warmer than abdomen) at a Ta of 14–30 °C. Above 33 °C many bees cooled their body via evaporation from their mouthparts. A novel mathematical model allows description of the relationship of resting (standard) metabolic rate and temperature across the entire functional temperature range of bees. In chill coma (<11 °C) bees were ectothermic and CO2 release was mostly continuous. CO2 release rate (nl s−1) decreased from 9.3 at 9.7 °C to 5.4 at 5 °C. At a Ta of >11 °C CO2 was released discontinuously. In the bees’ active temperature range mean CO2 production rate (nl s−1) increased sigmoidally (10.6 at 14.1 °C, 24.1 at 26.5 °C, and 55.2 at 38.1 °C), coming to a halt towards the upper lethal temperature. This was primarily accomplished by an exponential increase in gas exchange frequency (0.54 and 3.1 breaths min−1 at 14.1 and 38.1 °C) but not in released CO2 volume per respiratory cycle (1487 and 1083 nl cycle−1 at 14.1 and 38.1 °C). Emission of CO2 bursts was mostly (98%) accompanied by abdominal ventilation movements even in small CO2 bursts. Larger bursts coincided with a longer duration of active ventilation. An increased amount of CO2 expelled per unit time of ventilation indicates a higher efficiency of ventilation at high ambient temperatures. PMID:17707395

  4. Transthoracic impedance used to evaluate performance of cardiopulmonary resuscitation during out of hospital cardiac arrest.

    PubMed

    Stecher, Frederik S; Olsen, Jan-Aage; Stickney, Ronald E; Wik, Lars

    2008-12-01

    There is a need to measure cardiopulmonary resuscitation (CPR) in order to document whether ambulance personnel follow CPR guidelines. Our goal was to do this using defibrillator technology based on changes in transthoracic impedance (TTI) produced by chest compressions and ventilations. 122 incidents of out-of-hospital cardiac arrest between May 2003 and February 2004 were analysed based on data recorded from defibrillators in Oslo EMS. New software was used to analyze chest compressions and ventilations based on changes in thoracic impedance between the defibrillator pads, as well as ECG and other event data. In total, 25+/-14% (varying from 76% to 3%) of the time chest compressions were not performed on patients without spontaneous circulation (NFR=No Flow Ratio). When adjusting for time spent on analysis of ECG, pulse check and defibrillation, NFR was 20+/-13% (varying from 70% to 3%). Mean compressions delivered per minute was 87+/-16 and the compression rate during active compressions was 117+/-9min(-1). Individual variation was 31-117min(-1) (mean) and 95-144min(-1) (active periods). A mean of 14+/-3ventilations/min was recorded, varying from 8 to 26min(-1). Compared with the rest of the episode, the first 5min had a significantly higher proportion of time without chest compressions; 30+/-17% (p<0.001) and significantly lower mean compression and ventilation rates; 80+/-19min(-1) and 12+/-4min(-1), respectively (p<0.001 in both cases). Core CPR values can be measured from TTI signals by using a standard defibrillator and new software. NFR was 25% (20% adjusted) with great rescuer variability.

  5. Effects of salinity and hypoxia-induced hyperventilation on oxygen consumption and cost of osmoregulation in the estuarine red drum (Sciaenops ocellatus).

    PubMed

    Ern, Rasmus; Esbaugh, Andrew J

    2018-04-23

    Understanding the physiological responses of fishes to salinity changes and aquatic hypoxia is essential for the conservation of marine species. Salinity changes affect the osmotic gradient across the gill epithelium, while hypoxia increases gill ventilation and the flow of water over the gills. Both processes affect the diffusive movement of ions and water across the gill epithelium, and the rate of active ion transport required for maintaining osmotic homeostasis. Consequently, salinity and hypoxia may affect the energetic cost of osmoregulation, and consequently the energy available for other physiological functions such as migration, growth, and reproduction. Historically, studies have assessed the costs of osmoregulation and ventilation in fishes via standard metabolic rate (SMR); however, few studies have used a multi-stressor approach that fully accounts for the osmorespiratory compromise. Here, we determined the combined effects of salinity and hypoxia on SMR, routine metabolic rate (RMR), and plasma ion concentrations in red drum (Sciaenops ocellatus) acclimated to salinities ranging from freshwater to hypersalinity. Surprisingly, there was no significant change in any parameter as a consequence of salinity or hypoxia, including the relatively extreme scenario of combined hypersalinity and hypoxia exposure. We conclude that changes in the osmotic gradient across the gill epithelium and the flow of water over the gills have a negligible effect on the whole animal energy budget of S. ocellatus, suggesting that the cost of osmoregulation is a minor component of basal metabolism regardless of oxygenation status. Copyright © 2018 Elsevier Inc. All rights reserved.

  6. Determination of respiratory gas flow by electrical impedance tomography in an animal model of mechanical ventilation

    PubMed Central

    2014-01-01

    Background A recent method determines regional gas flow of the lung by electrical impedance tomography (EIT). The aim of this study is to show the applicability of this method in a porcine model of mechanical ventilation in healthy and diseased lungs. Our primary hypothesis is that global gas flow measured by EIT can be correlated with spirometry. Our secondary hypothesis is that regional analysis of respiratory gas flow delivers physiologically meaningful results. Methods In two sets of experiments n = 7 healthy pigs and n = 6 pigs before and after induction of lavage lung injury were investigated. EIT of the lung and spirometry were registered synchronously during ongoing mechanical ventilation. In-vivo aeration of the lung was analysed in four regions-of-interest (ROI) by EIT: 1) global, 2) ventral (non-dependent), 3) middle and 4) dorsal (dependent) ROI. Respiratory gas flow was calculated by the first derivative of the regional aeration curve. Four phases of the respiratory cycle were discriminated. They delivered peak and late inspiratory and expiratory gas flow (PIF, LIF, PEF, LEF) characterizing early or late inspiration or expiration. Results Linear regression analysis of EIT and spirometry in healthy pigs revealed a very good correlation measuring peak flow and a good correlation detecting late flow. PIFEIT = 0.702 · PIFspiro + 117.4, r2 = 0.809; PEFEIT = 0.690 · PEFspiro-124.2, r2 = 0.760; LIFEIT = 0.909 · LIFspiro + 27.32, r2 = 0.572 and LEFEIT = 0.858 · LEFspiro-10.94, r2 = 0.647. EIT derived absolute gas flow was generally smaller than data from spirometry. Regional gas flow was distributed heterogeneously during different phases of the respiratory cycle. But, the regional distribution of gas flow stayed stable during different ventilator settings. Moderate lung injury changed the regional pattern of gas flow. Conclusions We conclude that the presented method is able to determine global respiratory gas flow of the lung in different phases of the respiratory cycle. Additionally, it delivers meaningful insight into regional pulmonary characteristics, i.e. the regional ability of the lung to take up and to release air. PMID:24779960

  7. Ischemic-Anoxia of the Central Nervous System: Iron Dependent Oxidative Injury during Reperfusion.

    DTIC Science & Technology

    1986-10-15

    much deeper tissue acidosis and augmented injury is seen in contrast to complete ischemic-anoxia. 4 8. The delocalized iron catalyzes the production of...of deep metabolic acidosis (HCO5 at about 10 meq/L). OCCM maintained good oxygenation, ventilation and acid base balance. The blood gas differences to...lactic acidosis which occurs in the brain under the influence of such low flow rates. 4 3. Siesjo’s study of the pH dependence of lipid peroxidation in

  8. A Stress Test to Evaluate the Physical Capacity of Performing L-1 Anti-G Straining Maneuvers

    DTIC Science & Technology

    1988-09-01

    similar to hyperventilation can produce respiratory alkalosis . Therefore, lowered G-tolerance may result from decreased cerebral blood flow. Another...VE), respiratory exchange ratio (RER) and heart rate (HR were determined for steady state exercise at AGSM duty cyclEs cf 2,, 215%, 33% and 50%. Thesc...V02, ml/kg/min), minute pulmonary ventilation (VE, I/min) and respiratory exchange ratio (RER, VCO 2/VO2) at each duty cycle of discontinuous AGSM

  9. Information flow to assess cardiorespiratory interactions in patients on weaning trials.

    PubMed

    Vallverdú, M; Tibaduisa, O; Clariá, F; Hoyer, D; Giraldo, B; Benito, S; Caminal, P

    2006-01-01

    Nonlinear processes of the autonomic nervous system (ANS) can produce breath-to-breath variability in the pattern of breathing. In order to provide assess to these nonlinear processes, nonlinear statistical dependencies between heart rate variability and respiratory pattern variability are analyzed. In this way, auto-mutual information and cross-mutual information concepts are applied. This information flow analysis is presented as a short-term non linear analysis method to investigate the information flow interactions in patients on weaning trials. 78 patients from mechanical ventilation were studied: Group A of 28 patients that failed to maintain spontaneous breathing and were reconnected; Group B of 50 patients with successful trials. The results show lower complexity with an increase of information flow in group A than in group B. Furthermore, a more (weakly) coupled nonlinear oscillator behavior is observed in the series of group A than in B.

  10. Vortex dynamics and scalar transport in the wake of a bluff body driven through a steady recirculating flow

    NASA Astrophysics Data System (ADS)

    Poussou, Stephane B.; Plesniak, Michael W.

    2012-09-01

    The air ventilation system in wide-body aircraft cabins provides passengers with a healthy breathing environment. In recent years, the increase in global air traffic has amplified contamination risks by airborne flu-like diseases and terrorist threats involving the onboard release of noxious materials. In particular, passengers moving through a ventilated cabin may transport infectious pathogens in their wake. This paper presents an experimental investigation of the wake produced by a bluff body driven through a steady recirculating flow. Data were obtained in a water facility using particle image velocimetry and planar laser induced fluorescence. Ventilation attenuated the downward convection of counter-rotating vortices produced near the free-end corners of the body and decoupled the downwash mechanism from forward entrainment, creating stagnant contaminant regions.

  11. Investigations of High Pressure Acoustic Waves in Resonators with Seal-like Features

    NASA Technical Reports Server (NTRS)

    Daniels, Christopher; Steinetz, Bruce; Finkbeiner, Joshua

    2003-01-01

    A conical resonator (having a dissonant acoustic design) was tested in four configurations: (1) baseline resonator with closed ends and no blockage, (2) closed resonator with internal blockage, (3) ventilated resonator with no blockage, and (4) ventilated resonator with an applied pressure differential. These tests were conducted to investigate the effects of blockage and ventilation holes on dynamic pressurization. Additionally, the investigation was to determine the ability of acoustic pressurization to impede flow through the resonator. In each of the configurations studied, the entire resonator was oscillated at the gas resonant frequency while dynamic pressure, static pressure, and temperature of the fluid were measured. In the final configuration, flow through the resonator was recorded for three oscillation conditions. Ambient condition air was used as the working fluid.

  12. Low-flow CO₂ removal integrated into a renal-replacement circuit can reduce acidosis and decrease vasopressor requirements.

    PubMed

    Forster, Christian; Schriewer, Jens; John, Stefan; Eckardt, Kai-Uwe; Willam, Carsten

    2013-07-24

    Lung-protective ventilation in patients with ARDS and multiorgan failure, including renal failure, is often paralleled with a combined respiratory and metabolic acidosis. We assessed the effectiveness of a hollow-fiber gas exchanger integrated into a conventional renal-replacement circuit on CO₂ removal, acidosis, and hemodynamics. In ten ventilated critically ill patients with ARDS and AKI undergoing renal- and respiratory-replacement therapy, effects of low-flow CO₂ removal on respiratory acidosis compensation were tested by using a hollow-fiber gas exchanger added to the renal-replacement circuit. This was an observational study on safety, CO₂-removal capacity, effects on pH, ventilator settings, and hemodynamics. CO₂ elimination in the low-flow circuit was safe and was well tolerated by all patients. After 4 hours of treatment, a mean reduction of 17.3 mm Hg (-28.1%) pCO₂ was observed, in line with an increase in pH. In hemodynamically instable patients, low-flow CO₂ elimination was paralleled by hemodynamic improvement, with an average reduction of vasopressors of 65% in five of six catecholamine-dependent patients during the first 24 hours. Because no further catheters are needed, besides those for renal replacement, the implementation of a hollow-fiber gas exchanger in a renal circuit could be an attractive therapeutic tool with only a little additional trauma for patients with mild to moderate ARDS undergoing invasive ventilation with concomitant respiratory acidosis, as long as no severe oxygenation defects indicate ECMO therapy.

  13. Termite mounds harness diurnal temperature oscillations for ventilation

    PubMed Central

    King, Hunter; Ocko, Samuel; Mahadevan, L.

    2015-01-01

    Many species of millimetric fungus-harvesting termites collectively build uninhabited, massive mound structures enclosing a network of broad tunnels that protrude from the ground meters above their subterranean nests. It is widely accepted that the purpose of these mounds is to give the colony a controlled microclimate in which to raise fungus and brood by managing heat, humidity, and respiratory gas exchange. Although different hypotheses such as steady and fluctuating external wind and internal metabolic heating have been proposed for ventilating the mound, the absence of direct in situ measurement of internal air flows has precluded a definitive mechanism for this critical physiological function. By measuring diurnal variations in flow through the surface conduits of the mounds of the species Odontotermes obesus, we show that a simple combination of geometry, heterogeneous thermal mass, and porosity allows the mounds to use diurnal ambient temperature oscillations for ventilation. In particular, the thin outer flutelike conduits heat up rapidly during the day relative to the deeper chimneys, pushing air up the flutes and down the chimney in a closed convection cell, with the converse situation at night. These cyclic flows in the mound flush out CO2 from the nest and ventilate the colony, in an unusual example of deriving useful work from thermal oscillations. PMID:26316023

  14. Termite mounds harness diurnal temperature oscillations for ventilation.

    PubMed

    King, Hunter; Ocko, Samuel; Mahadevan, L

    2015-09-15

    Many species of millimetric fungus-harvesting termites collectively build uninhabited, massive mound structures enclosing a network of broad tunnels that protrude from the ground meters above their subterranean nests. It is widely accepted that the purpose of these mounds is to give the colony a controlled microclimate in which to raise fungus and brood by managing heat, humidity, and respiratory gas exchange. Although different hypotheses such as steady and fluctuating external wind and internal metabolic heating have been proposed for ventilating the mound, the absence of direct in situ measurement of internal air flows has precluded a definitive mechanism for this critical physiological function. By measuring diurnal variations in flow through the surface conduits of the mounds of the species Odontotermes obesus, we show that a simple combination of geometry, heterogeneous thermal mass, and porosity allows the mounds to use diurnal ambient temperature oscillations for ventilation. In particular, the thin outer flutelike conduits heat up rapidly during the day relative to the deeper chimneys, pushing air up the flutes and down the chimney in a closed convection cell, with the converse situation at night. These cyclic flows in the mound flush out CO2 from the nest and ventilate the colony, in an unusual example of deriving useful work from thermal oscillations.

  15. Flow and contaminant transport in an airliner cabin induced by a moving body: Model experiments and CFD predictions

    NASA Astrophysics Data System (ADS)

    Poussou, Stephane B.; Mazumdar, Sagnik; Plesniak, Michael W.; Sojka, Paul E.; Chen, Qingyan

    2010-08-01

    The effects of a moving human body on flow and contaminant transport inside an aircraft cabin were investigated. Experiments were performed in a one-tenth scale, water-based model. The flow field and contaminant transport were measured using the Particle Image Velocimetry (PIV) and Planar Laser-Induced Fluorescence (PLIF) techniques, respectively. Measurements were obtained with (ventilation case) and without (baseline case) the cabin environmental control system (ECS). The PIV measurements show strong intermittency in the instantaneous near-wake flow. A symmetric downwash flow was observed along the vertical centerline of the moving body in the baseline case. The evolution of this flow pattern is profoundly perturbed by the flow from the ECS. Furthermore, a contaminant originating from the moving body is observed to convect to higher vertical locations in the presence of ventilation. These experimental data were used to validate a Computational Fluid Dynamic (CFD) model. The CFD model can effectively capture the characteristic flow features and contaminant transport observed in the small-scale model.

  16. Numerical Study of the Transition Between Reentrant Jet and Twin Vortex Flow Regimes in Ventilated Cavitation

    NASA Astrophysics Data System (ADS)

    Adama Maiga, Mahamadou; Coutier-Delgosha, Olivier; Bois, Gérard

    2018-06-01

    Contrary to natural cavitation, ventilated cavitation is controllable and is not harmful. It is particularly used to reduce the drag of the hydraulic vehicles. The ventilated cavitation is characterized by various gas regimes. The mechanisms of ventilated cavitation are investigated in the present work with CFD based on a 2D solver. The attention is especially focused on the transition between the reentrant jet and twin vortex regimes. The results confirm that the product of ventilated cavitation number and Froude number is lower than 1 (σ c Fr < 1) in the twin vortex regime, while it is higher than 1 (σ c Fr > 1) in the reentrant jet regime, as reported in the literature. Further analysis shows that ventilated cavitation is significantly influenced by the natural cavitation number.

  17. Mechanical Ventilation: State of the Art.

    PubMed

    Pham, Tài; Brochard, Laurent J; Slutsky, Arthur S

    2017-09-01

    Mechanical ventilation is the most used short-term life support technique worldwide and is applied daily for a diverse spectrum of indications, from scheduled surgical procedures to acute organ failure. This state-of-the-art review provides an update on the basic physiology of respiratory mechanics, the working principles, and the main ventilatory settings, as well as the potential complications of mechanical ventilation. Specific ventilatory approaches in particular situations such as acute respiratory distress syndrome and chronic obstructive pulmonary disease are detailed along with protective ventilation in patients with normal lungs. We also highlight recent data on patient-ventilator dyssynchrony, humidified high-flow oxygen through nasal cannula, extracorporeal life support, and the weaning phase. Finally, we discuss the future of mechanical ventilation, addressing avenues for improvement. Copyright © 2017 Mayo Foundation for Medical Education and Research. Published by Elsevier Inc. All rights reserved.

  18. Computational Fluid Dynamics (CFD) investigation onto passenger car disk brake design

    NASA Astrophysics Data System (ADS)

    Munisamy, Kannan M.; Kanasan Moorthy, Shangkari K.

    2013-06-01

    The aim of this study is to investigate the flow and heat transfer in ventilated disc brakes using Computational Fluid Dynamics (CFD). NACA Series blade is designed for ventilated disc brake and the cooling characteristic is compared to the baseline design. The ventilated disc brakes are simulated using commercial CFD software FLUENTTM using simulation configuration that was obtained from experiment data. The NACA Series blade design shows improvements in Nusselt number compared to baseline design.

  19. Regional distribution of ventilation assessed by Kr-81m scintigraphy employing temporal Fourier transform

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Slosman, D.; Susskind, H.; Cinotti, L.

    1986-01-01

    Temporal Fourier analysis was applied to Kr-81m ventilation scintigraphy to determine the amplitude (AMP1) and phase (PHA1) of the first harmonic of a single composite respiratory cycle and to compare regional patterns in subjects with obstructive pulmonary disease (COPD) and nonobstructed subjects. Six nonobstructed subjects, three subjects with small airway disease, six subjects with COPD, and one subject with restrictive disease were investigated. The mean value of the functional PHA1 image (PHA1m) correlated negatively with 1-second forced expiratory volume (FEV1) (r = -0.801, P less than .001), with %FEV1/FVC (r = -0.636, P less than .01) and maximum midexpiratory flowmore » rate (FEF25-75%) (r = -0.723, P less than .002), and correlated positively with residual volume (r = 0.640, P less than .01). PHA1m values for the six subjects with COPD were significantly higher (t = 2.359, P less than .05) than for the ten nonobstructed subjects. Display of phase and amplitude functional images permits a visual evaluation of the regional distribution of ventilation to be made. Regional abnormalities of air flow were detected in obstructed subjects, and the presence of airway obstruction could be predicted. Dynamic ventilation imaging, therefore, appears to be a potentially useful noninvasive technique to assess lung impairment on a localized level.« less

  20. Ventilation requirements for control of occupancy odor and tobacco smoke odor: laboratory studies. Final report

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Cain, W.S.; Isseroff, R.; Leaderer, B.P.

    1981-04-01

    Experiments on occupancy odor addressed the question of why required ventilation rate per occupant increased progressively with increases in the number of persons in a space. In order to investigate ventilation requirements under approximately ideal conditions, we constructed an aluminum-lined environmental chamber with excellent control over environmental conditions and a ventilation system that provided rapid and uniform mixing of air. Psychophysical experiments on occupancy odor explored 47 different combinations of occupancy density, temperature and humidity, and ventilation rate. The experiments collected judgements both from visitors, who smelled air from the chamber only once every few minutes, and from occupants, whomore » remained in the chamber for an hour at a time. The judgements of visitors revealed that occupancy odor increased only gradually over time and rarely reached very high or objectionable levels. Judgements of occupants also revealed rather minor dissatisfaction. Only during combinations of high temperature and humidity did objectionability become more than a minor issue to either group. Experiments on cigarette smoking explored rates of 4, 8, and 16 cigarettes per hour under various environmental conditions and with ventilation rates as high as 68 cfm (34 L.s/sup -1/) per occupant. As soon as occupants lit cigarettes in the chamber, the odor level increased dramatically. At ventilation rates far greater than necessary to control occupancy odor, the odor from cigarette smoking remained quite intense. In general, the odor proved impossible to control adequately even with a ventilation rate of 68 cfm (34 L.s/sup -1/) per occupant (4 occupants) and even when only one occupant smoked at a time. As in the case of occupancy odor, a combination of high temperature and humidity exacerbated the odor problem.« less

  1. The Competition Between a Localised and Distributed Source of Buoyancy

    NASA Astrophysics Data System (ADS)

    Partridge, Jamie; Linden, Paul

    2012-11-01

    We propose a new mathematical model to study the competition between localised and distributed sources of buoyancy within a naturally ventilated filling box. The main controlling parameters in this configuration are the buoyancy fluxes of the distributed and local source, specifically their ratio Ψ. The steady state dynamics of the flow are heavily dependent on this parameter. For large Ψ, where the distributed source dominates, we find the space becomes well mixed as expected if driven by an distributed source alone. Conversely, for small Ψ we find the space reaches a stable two layer stratification. This is analogous to the classical case of a purely local source but here the lower layer is buoyant compared to the ambient, due to the constant flux of buoyancy emanating from the distributed source. The ventilation flow rate, buoyancy of the layers and also the location of the interface height, which separates the two layer stratification, are obtainable from the model. To validate the theoretical model, small scale laboratory experiments were carried out. Water was used as the working medium with buoyancy being driven directly by temperature differences. Theoretical results were compared with experimental data and overall good agreement was found. A CASE award project with Arup.

  2. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Singer, Brett C.; Pass, Rebecca Zarin; Delp, William W.

    Combustion pollutant concentrations were measured during the scripted operation of natural gas cooking burners in nine homes. Boiling and simmering activities were conducted on the stovetop and in the oven with and without range hood exhaust ventilation or air mixing via a forced air system. Time-resolved concentrations of carbon dioxide (CO 2), nitric oxide (NO), nitrogen oxides (NO x), nitrogen dioxide (NO 2), particles with diameters of 6 nm or larger (PN), carbon monoxide (CO), and fine particulate matter (PM 2.5) were measured in the kitchen and bedroom area of each home. Four of the nine homes had kitchen 1more » h NO 2 exceed the national ambient air quality standard (100 ppb). In all homes, the highest 1 h integrated PN exceeded 2 × 10 5 cm -3-h, and the highest 4 h PN exceeded 3 × 10 5 cm -3-hr in the kitchen. Range hood performance varied widely, but one with a large capture volume and a measured flow of 108 L/s reduced concentrations 80–95%. Increased awareness of the need to ventilate when cooking, along with building standards for minimum range hood flow rates and volume, could substantially reduce exposures to NO 2 and ultrafine particles in homes.« less

  3. Comparison of the OxyMask and Venturi mask in the delivery of supplemental oxygen: Pilot study in oxygen-dependent patients

    PubMed Central

    Beecroft, Jaime M; Hanly, Patrick J

    2006-01-01

    BACKGROUND: The OxyMask (Southmedic Inc, Canada) is a new face mask for oxygen delivery that uses a small ‘diffuser’ to concentrate and direct oxygen toward the mouth and nose. The authors hypothesized that this unique design would enable the OxyMask to deliver oxygen more efficiently than a Venturi mask (Hudson RCI, USA) in patients with chronic hypoxemia. METHODS: Oxygen-dependent patients with chronic, stable respiratory disease were recruited to compare the OxyMask and Venturi mask in a randomized, single-blind, cross-over design. Baseline blood oxygen saturation (SaO2) was established breathing room air, followed in a random order by supplemental oxygen through the OxyMask or Venturi mask. Oxygen delivery was titrated to maintain SaO2 4% to 5% and 8% to 9% above baseline for two separate 30 min periods of stable breathing. Oxygen flow rate, partial pressure of inspired and expired oxygen (PO2) and carbon dioxide (PCO2), minute ventilation, heart rate, nasal and oral breathing, SaO2 and transcutaneous PCO2 were collected continuously. The study was repeated following alterations to the OxyMask design, which improved clearance of carbon dioxide. RESULTS: Thirteen patients, aged 28 to 79 years, were studied initially using the original OxyMask. Oxygen flow rate was lower, inspired PO2 was higher and expired PO2 was lower while using the OxyMask. Minute ventilation and inspired and expired PCO2 were significantly higher while using the OxyMask, whereas transcutaneous PCO2, heart rate and the ratio of nasal to oral breathing did not change significantly throughout the study. Following modification of the OxyMask, 13 additional patients, aged 18 to 79 years, were studied using the same protocol. The modified OxyMask provided a higher inspired PO2 at a lower flow rate, without evidence of carbon dioxide retention. CONCLUSIONS: Oxygen is delivered safely and more efficiently by the OxyMask than by the Venturi mask in stable oxygen-dependent patients. PMID:16896425

  4. Capnography and chest wall impedance algorithms for ventilation detection during cardiopulmonary resuscitation

    PubMed Central

    Edelson, Dana P.; Eilevstjønn, Joar; Weidman, Elizabeth K.; Retzer, Elizabeth; Vanden Hoek, Terry L.; Abella, Benjamin S.

    2009-01-01

    Objective Hyperventilation is both common and detrimental during cardiopulmonary resuscitation (CPR). Chest wall impedance algorithms have been developed to detect ventilations during CPR. However, impedance signals are challenged by noise artifact from multiple sources, including chest compressions. Capnography has been proposed as an alternate method to measure ventilations. We sought to assess and compare the adequacy of these two approaches. Methods Continuous chest wall impedance and capnography were recorded during consecutive in-hospital cardiac arrests. Algorithms utilizing each of these data sources were compared to a manually determined “gold standard” reference ventilation rate. In addition, a combination algorithm, which utilized the highest of the impedance or capnography values in any given minute, was similarly evaluated. Results Data were collected from 37 cardiac arrests, yielding 438 min of data with continuous chest compressions and concurrent recording of impedance and capnography. The manually calculated mean ventilation rate was 13.3±4.3/min. In comparison, the defibrillator’s impedance-based algorithm yielded an average rate of 11.3±4.4/min (p=0.0001) while the capnography rate was 11.7±3.7/min (p=0.0009). There was no significant difference in sensitivity and positive predictive value between the two methods. The combination algorithm rate was 12.4±3.5/min (p=0.02), which yielded the highest fraction of minutes with respiratory rates within 2/min of the reference. The impedance signal was uninterpretable 19.5% of the time, compared with 9.7% for capnography. However, the signals were only simultaneously non-interpretable 0.8% of the time. Conclusions Both the impedance and capnography-based algorithms underestimated the ventilation rate. Reliable ventilation rate determination may require a novel combination of multiple algorithms during resuscitation. PMID:20036047

  5. The clinical practice of high-flow nasal cannula oxygen therapy in adults: A Japanese cross-sectional multicenter survey.

    PubMed

    Ito, Jiro; Nagata, Kazuma; Sato, Susumu; Shiraki, Akira; Nishimura, Naoki; Izumi, Shinyu; Tachikawa, Ryo; Morimoto, Takeshi; Tomii, Keisuke

    2018-05-01

    High-flow nasal cannula oxygen therapy (HFNC) is widely used mainly in the acute care setting, but limited data are available on real-world practice in adults. The objective of this study was to describe HFNC practices in Japanese adults. A retrospective cross-sectional multicenter survey of adult patients receiving HFNC from January through March 2015 was conducted in 33 participating hospitals in Japan. We obtained information on 321 patients (median age, 76; 218 men, 103 women; median estimated PaO 2 /F I O 2, 178 mm Hg) from 22 hospitals. Do-not-intubate status was determined in 37.4% of patients. Prior to HFNC, 57.9% of patients received conventional oxygen therapy; 25.9%, noninvasive ventilation; and 15.0%, invasive mechanical ventilation. The common indications for HFNC were acute hypoxemic respiratory failure (ARF) (65.4%), postoperative respiratory support (15.9%), and post-extubation respiratory support (11.2%). The underlying etiology of ARF included interstitial lung disease, pneumonia, and cardiogenic pulmonary edema. HFNC was administered mostly in intensive care units or intermittent care units (60.7%) and general wards (36.1%). Median duration of HFNC was 4 days; median total flow rate, 40 L/min; and median F I O 2 , 50%. HFNC significantly improved PaO 2 , PaCO 2 , SpO 2 and respiratory rate from baseline. Two-thirds of patients finally survived to be discharged or transferred. We documented patient demographics, clinical indications, and settings of HFNC use in the real world. We also demonstrated positive effects of HFNC on respiratory parameters. Further studies are urgently needed regarding the efficacy and safety of HFNC in populations outside of previous clinical trials. Copyright © 2018 The Japanese Respiratory Society. Published by Elsevier B.V. All rights reserved.

  6. High-Frequency Percussive Ventilation: Pneumotachograph Validation and Tidal Volume Analysis

    DTIC Science & Technology

    2010-06-01

    protocol, preliminary experience has shown that the flow sensor is amenable to near-automated “plug-and- play ” adaptability, permitting clinicians the...400. 6. Velmahos GC, Chan LS, Tatevossian R, Cornwell EE 3rd, Dough - erty WR, Escudero J, Demetriades D. High-frequency percussive ventilation

  7. Healthy Building Design for the Commercial, Industrial, and Institutional Marketplace.

    ERIC Educational Resources Information Center

    Turner, William A.

    Building design and construction that helps deliver both superior air quality and occupant thermal comfort, while minimizing energy consumption, are examined. The paper explores an integrated building systems approach that combines the principles of "directed air flow control" and "demand controlled ventilation" where ventilation is effectively…

  8. Comparison of coarse coal dust sampling techniques in a laboratory-simulated longwall section.

    PubMed

    Patts, Justin R; Barone, Teresa L

    2017-05-01

    Airborne coal dust generated during mining can deposit and accumulate on mine surfaces, presenting a dust explosion hazard. When assessing dust hazard mitigation strategies for airborne dust reduction, sampling is done in high-velocity ventilation air, which is used to purge the mining face and gallery tunnel. In this environment, the sampler inlet velocity should be matched to the air stream velocity (isokinetic sampling) to prevent oversampling of coarse dust at low sampler-to-air velocity ratios. Low velocity ratios are often encountered when using low flow rate, personal sampling pumps commonly used in underground mines. In this study, with a goal of employing mine-ready equipment, a personal sampler was adapted for area sampling of coarse coal dust in high-velocity ventilation air. This was done by adapting an isokinetic nozzle to the inlet of an Institute of Occupational Medicine (Edinburgh, Scotland) sampling cassette (IOM). Collected dust masses were compared for the modified IOM isokinetic sampler (IOM-MOD), the IOM without the isokinetic nozzle, and a conventional dust sampling cassette without the cyclone on the inlet. All samplers were operated at a flow rate typical of personal sampling pumps: 2 L/min. To ensure differences between collected masses that could be attributed to sampler design and were not influenced by artifacts from dust concentration gradients, relatively uniform and repeatable dust concentrations were demonstrated in the sampling zone of the National Institute for Occupational Safety and Health experimental mine gallery. Consistent with isokinetic theory, greater differences between isokinetic and non-isokinetic sampled masses were found for larger dust volume-size distributions and higher ventilation air velocities. Since isokinetic sampling is conventionally used to determine total dust concentration, and isokinetic sampling made a difference in collected masses, the results suggest when sampling for coarse coal dust the IOM-MOD may improve airborne coarse dust assessments over "off-the-shelf" sampling cassettes.

  9. Comparison of coarse coal dust sampling techniques in a laboratory-simulated longwall section

    PubMed Central

    Patts, Justin R.; Barone, Teresa L.

    2017-01-01

    Airborne coal dust generated during mining can deposit and accumulate on mine surfaces, presenting a dust explosion hazard. When assessing dust hazard mitigation strategies for airborne dust reduction, sampling is done in high-velocity ventilation air, which is used to purge the mining face and gallery tunnel. In this environment, the sampler inlet velocity should be matched to the air stream velocity (isokinetic sampling) to prevent oversampling of coarse dust at low sampler-to-air velocity ratios. Low velocity ratios are often encountered when using low flow rate, personal sampling pumps commonly used in underground mines. In this study, with a goal of employing mine-ready equipment, a personal sampler was adapted for area sampling of coarse coal dust in high-velocity ventilation air. This was done by adapting an isokinetic nozzle to the inlet of an Institute of Occupational Medicine (Edinburgh, Scotland) sampling cassette (IOM). Collected dust masses were compared for the modified IOM isokinetic sampler (IOM-MOD), the IOM without the isokinetic nozzle, and a conventional dust sampling cassette without the cyclone on the inlet. All samplers were operated at a flow rate typical of personal sampling pumps: 2 L/min. To ensure differences between collected masses that could be attributed to sampler design and were not influenced by artifacts from dust concentration gradients, relatively uniform and repeatable dust concentrations were demonstrated in the sampling zone of the National Institute for Occupational Safety and Health experimental mine gallery. Consistent with isokinetic theory, greater differences between isokinetic and non-isokinetic sampled masses were found for larger dust volume-size distributions and higher ventilation air velocities. Since isokinetic sampling is conventionally used to determine total dust concentration, and isokinetic sampling made a difference in collected masses, the results suggest when sampling for coarse coal dust the IOM-MOD may improve airborne coarse dust assessments over “off-the-shelf” sampling cassettes. PMID:27792474

  10. Nasal high-flow therapy reduces work of breathing compared with oxygen during sleep in COPD and smoking controls: a prospective observational study

    PubMed Central

    Kirkness, Jason P.; Grote, Ludger; Fricke, Kathrin; Schwartz, Alan R.; Smith, Philip; Schneider, Hartmut

    2017-01-01

    Patients with chronic obstructive pulmonary disease (COPD) endure excessive resistive and elastic loads leading to chronic respiratory failure. Oxygen supplementation corrects hypoxemia but is not expected to reduce mechanical loads. Nasal high-flow (NHF) therapy supports breathing by reducing dead space, but it is unclear how it affects mechanical loads of patients with COPD. The objective of this study was to compare the effects of low-flow oxygen and NHF therapy on ventilation and work of breathing (WOB) in patients with COPD and controls during sleep. Patients with COPD (n = 12) and controls (n = 6) were recruited and submitted to polysomnography to measure sleep parameters and ventilation in response to administration of oxygen and NHF. A subset of six patients also had an esophageal catheter inserted for the purpose of measuring WOB. Patients with COPD had similar minute ventilation (V̇e) but lower tidal volumes than matched controls. With oxygen, SaO2was increased and V̇e was reduced in both controls and patients with COPD, but there was an increase in transcutaneous CO2 levels. NHF produced a greater reduction in V̇e and was associated with a reduction in CO2 levels. Although NHF halved WOB, oxygen produced only a minor reduction in this parameter. We conclude that oxygen produced little change in WOB, which was associated with CO2 elevations. On the other hand, NHF produced a large reduction in V̇e and WOB with a concomitant decrease in CO2 levels. Our data indicate that NHF improves alveolar ventilation during sleep compared with oxygen and room air in patients with COPD and therefore can decrease their cost of breathing. NEW & NOTEWORTHY Nasal high-flow (NHF) therapy can support ventilation in patients with chronic obstructive pulmonary disease during sleep by decreasing the work of breathing and improving CO2 levels. On the other hand, oxygen supplementation corrects hypoxemia, but it produces only a minimal reduction in work of breathing and is associated with increased CO2 levels. Therefore, NHF can be a useful method to assist ventilation in patients with increased respiratory mechanical loads. PMID:27815367

  11. Management and outcome of mechanically ventilated neurologic patients.

    PubMed

    Pelosi, Paolo; Ferguson, Niall D; Frutos-Vivar, Fernando; Anzueto, Antonio; Putensen, Christian; Raymondos, Konstantinos; Apezteguia, Carlos; Desmery, Pablo; Hurtado, Javier; Abroug, Fekri; Elizalde, José; Tomicic, Vinko; Cakar, Nahit; Gonzalez, Marco; Arabi, Yaseen; Moreno, Rui; Esteban, Andres

    2011-06-01

    To describe and compare characteristics, ventilatory practices, and associated outcomes among mechanically ventilated patients with different types of brain injury and between neurologic and nonneurologic patients. Secondary analysis of a prospective, observational, and multicenter study on mechanical ventilation. Three hundred forty-nine intensive care units from 23 countries. We included 552 mechanically ventilated neurologic patients (362 patients with stroke and 190 patients with brain trauma). For comparison we used a control group of 4,030 mixed patients who were ventilated for nonneurologic reasons. None. We collected demographics, ventilatory settings, organ failures, and complications arising during ventilation and outcomes. Multivariate logistic regression analysis was performed with intensive care unit mortality as the dependent variable. At admission, a Glasgow Coma Scale score ≤8 was observed in 68% of the stroke, 77% of the brain trauma, and 29% of the nonneurologic patients. Modes of ventilation and use of a lung-protective strategy within the first week of mechanical ventilation were similar between groups. In comparison with nonneurologic patients, patients with neurologic disease developed fewer complications over the course of mechanical ventilation with the exception of a higher rate of ventilator-associated pneumonia in the brain trauma cohort. Neurologic patients showed higher rates of tracheotomy and longer duration of mechanical ventilation. Mortality in the intensive care unit was significantly (p < .001) higher in patients with stroke (45%) than in brain trauma (29%) and nonneurologic disease (30%). Factors associated with mortality were: stroke (in comparison to brain trauma), Glasgow Coma Scale score on day 1, and severity at admission in the intensive care unit. In our study, one of every five mechanically ventilated patients received this therapy as a result of a neurologic disease. This cohort of patients showed a higher mortality rate than nonneurologic patients despite a lower incidence of extracerebral organ dysfunction.

  12. Ventilation rate in adults with a tracheal tube during cardiopulmonary resuscitation: A systematic review.

    PubMed

    Vissers, Gino; Soar, Jasmeet; Monsieurs, Koenraad G

    2017-10-01

    The optimal ventilation rate during cardiopulmonary resuscitation (CPR) with a tracheal tube is unknown. We evaluated whether in adults with cardiac arrest and a secure airway (tracheal tube), a ventilation rate of 10min -1 , compared to any other rate during CPR, improves outcomes. A systematic review up to 14 July 2016. We included both adult human and animal studies. A GRADE (Grades of Recommendation, Assessment, Development and Evaluation) approach was used to evaluate the quality of evidence for each outcome. We identified one human observational study with 67 patients and ten animal studies (234 pigs and 30 dogs). All studies carried a high risk of bias. All studies evaluated for return of spontaneous circulation (ROSC). Studies showed no improvement in ROSC with a ventilation rate of 10 min-1 compared to any other rate. The evidence for longer-term outcomes such as survival to discharge and survival with favourable neurological outcome was very limited. A ventilation rate recommendation of 10 min-1 during adult CPR with a tracheal tube and no pauses for chest compression is a very weak recommendation based on very low quality evidence. Copyright © 2017 Elsevier B.V. All rights reserved.

  13. Positive Rates and Factors Associated with Abnormal Lung Function of Greenhouse Workers in China: A Cross-Sectional Study.

    PubMed

    Zhu, Xiaojun; Gao, Panjun; Gu, Yishuo; Xiao, Pei; Liu, Mengxuan; Chen, Juan; Cen, Yacai; Ma, Wenjun; Li, Tao

    2017-08-24

    Since the number of greenhouse workers are increasing in China, this observational cross-sectional study was designed to evaluate lung function and discuss the potential risk factors, to provide evidence in the surveillance of greenhouse workers' health. 678 greenhouse workers in Gansu Province, China were enrolled. A questionnaire which included demographic and occupational information was used. Vital capacity (VC), forced vital capacity (FVC), forced expiratory volume in 1 s (FEV₁), and FEV₁:FVC ratios (FEV₁/FVC), maximal expiratory flow after 50% of the FVC has not been exhaled (MEF 50 ), maximal expiratory flow after 25% of the FVC has not been exhaled (MEF 25 ) and maximal mid-expiratory flow curve (MMEF) were measured as lung function indicators. The mean values and standard deviations (SDs) of VC% predicted, FVC% predicted, FEV₁% predicted and FEV₁/FVC ratio were 106.07 ± 13.36, 107.60 ± 13.95, 97.19 ± 14.80 and 89.76 ± 10.78 respectively. The positive rates of above four and abnormal lung ventilation function were 2.9%, 2.8%, 11.2%, 4.6% and 6.5% respectively. Gender, age, BMI and number of greenhouses owned were influence factors of lung ventilation function ( p < 0.05). The mean values and SDs of MEF 50 % predicted, MEF 25 % predicted and MMEF% predicted were 69.63 ± 24.95, 54.04 ± 24.94 and 66.81 ± 24.53. The positive rates of above three and abnormal small airway function were 45.0%, 72.1%, 47.2% and 49.4% respectively. Age, education and number of greenhouses owned were influence factors for small airway function ( p < 0.05). Working in a greenhouse might influence lung function of the workers. Small airway function indicators could be used as priority indicators for the surveillance of greenhouse workers' health.

  14. Hypoxic pulmonary vasoconstriction in reptiles: a comparative study of four species with different lung structures and pulmonary blood pressures.

    PubMed

    Skovgaard, Nini; Abe, Augusto S; Andrade, Denis V; Wang, Tobias

    2005-11-01

    Low O2 levels in the lungs of birds and mammals cause constriction of the pulmonary vasculature that elevates resistance to pulmonary blood flow and increases pulmonary blood pressure. This hypoxic pulmonary vasoconstriction (HPV) diverts pulmonary blood flow from poorly ventilated and hypoxic areas of the lung to more well-ventilated parts and is considered important for the local matching of ventilation to blood perfusion. In the present study, the effects of acute hypoxia on pulmonary and systemic blood flows and pressures were measured in four species of anesthetized reptiles with diverse lung structures and heart morphologies: varanid lizards (Varanus exanthematicus), caimans (Caiman latirostris), rattlesnakes (Crotalus durissus), and tegu lizards (Tupinambis merianae). As previously shown in turtles, hypoxia causes a reversible constriction of the pulmonary vasculature in varanids and caimans, decreasing pulmonary vascular conductance by 37 and 31%, respectively. These three species possess complex multicameral lungs, and it is likely that HPV would aid to secure ventilation-perfusion homogeneity. There was no HPV in rattlesnakes, which have structurally simple lungs where local ventilation-perfusion inhomogeneities are less likely to occur. However, tegu lizards, which also have simple unicameral lungs, did exhibit HPV, decreasing pulmonary vascular conductance by 32%, albeit at a lower threshold than varanids and caimans (6.2 kPa oxygen in inspired air vs. 8.2 and 13.9 kPa, respectively). Although these observations suggest that HPV is more pronounced in species with complex lungs and functionally divided hearts, it is also clear that other components are involved.

  15. T-piece resuscitator versus self-inflating bag for preterm resuscitation: an institutional experience.

    PubMed

    Jayaram, Archana; Sima, Adam; Barker, Gail; Thacker, Leroy R

    2013-07-01

    Manual ventilation in the delivery room is provided with devices such as self-inflating bags (SIBs), flow-inflating bags, and T-piece resuscitators. To compare the effect of type of manual ventilation device on overall response to resuscitation among preterm neonates born at < 35 weeks gestation. Retrospective data were collected in 2 time periods. Primary outcome was overall response to resuscitation, as measured by Apgar score. Secondary outcomes were incidence of air leaks, need for chest compressions/epinephrine, need for intubation, and surfactant use. We identified 294 resuscitations requiring ventilation. SIB was used for 135 neonates, and T-piece was used for 159 neonates. There was no significant difference between the 1-min and 5-min Apgar scores between SIB and T-piece (P = .77 and P = .11, respectively), nor were there significant differences in secondary outcomes. The rate of rise of Apgar score was higher, by 0.47, with T-piece, compared to SIB (95% CI 0.08-0.87, P = .02). Although some manikin studies favor T-piece for providing reliable and consistent pressures, our experience did not indicate significant differences in effectiveness of resuscitation between the T-piece and SIB in preterm resuscitations.

  16. Distribution of pulmonary ventilation and perfusion during short periods of weightlessness

    NASA Technical Reports Server (NTRS)

    Michels, D. B.; West, J. B.

    1978-01-01

    Airborne experiments were conducted on four trained normal male subjects (28-40 yr) to study pulmonary function during short periods (22-27 sec) of zero gravity obtained by flying a jet aircraft through appropriate parabolic trajectories. The cabin was always pressurized to a sea-level altitude. The discussion is limited to pulmonary ventilation and perfusion. The results clearly demonstrate that gravity is the major factor causing nonuniformity in the topographical distribution of pulmonary ventilation. More importantly, the results suggest that virtually all the topographical nonuniformity of ventilation, blood flow, and lung volume observed under 1-G conditions are eliminated during short periods of zero gravity.

  17. A Mathematical Model for the Middle Ear Ventilation

    NASA Astrophysics Data System (ADS)

    Molnárka, G.; Miletics, E. M.; Fücsek, M.

    2008-09-01

    The otitis media is one of the mostly existing illness for the children, therefore investigation of the human middle ear ventilation is an actual problem. In earlier investigations both experimental and theoretical approach one can find in ([l]-[3]). Here we give a new mathematical and computer model to simulate this ventilation process. This model able to describe the diffusion and flow processes simultaneously, therefore it gives more precise results than earlier models did. The article contains the mathematical model and some results of the simulation.

  18. Continuous on-line measurements of respiratory system, lung and chest wall mechanics during mechanic ventilation.

    PubMed

    Kárason, S; Søndergaard, S; Lundin, S; Stenqvist, O

    2001-08-01

    We present a concept of on-line, manoeuvre-free monitoring of respiratory mechanics during dynamic conditions, displaying calculated alveolar pressure/volume curves continuously and separating lung and chest wall mechanics. Prospective observational study. Intensive care unit of a university hospital. Ten ventilator-treated patients with acute lung injury. Different positive end-expiratory pressure (PEEP) and tidal volumes, low flow inflation. Previously validated methods were used to present a single-value dynostatic compliance for the whole breath and a dynostatic volume-dependent initial, middle and final compliance within the breath. A high individual variation of respiratory mechanics was observed. Reproducibility of repeated measurements was satisfactory (coefficients of variations for dynostatic volume-dependent compliance: < or =9.2% for total respiratory system, < or =18% for lung). Volume-dependent compliance showed a statistically significant pattern of successively decreasing compliance from the initial segment through the middle and final parts within each breath at all respiratory settings. This pattern became more prominent with increasing PEEP and tidal volume, indicating a greater distension of alveoli. No lower inflection point (LIP) was seen in patients with respiratory rate 20/min and PEEP at 4 cmH2O. A trial with low flow inflation in four of the patients showed formation of a LIP in three of them and an upper inflection in one. The monitoring concept revealed a constant pattern of successively decreasing compliance within each breath, which became more prominent with increasing PEEP and tidal volume. The monitoring concept offers a simple and reliable method of monitoring respiratory mechanics during ongoing ventilator treatment.

  19. Prevention of Ventilator-Associated Pneumonia: The Multimodal Approach of the Spanish ICU “Pneumonia Zero” Program*

    PubMed Central

    Palomar-Martínez, Mercedes; Sánchez-García, Miguel; Martínez-Alonso, Montserrat; Álvarez-Rodríguez, Joaquín; Lorente, Leonardo; Arias-Rivera, Susana; García, Rosa; Gordo, Federico; Añón, José M.; Jam-Gatell, Rosa; Vázquez-Calatayud, Mónica; Agra, Yolanda

    2018-01-01

    Objectives: The “Pneumonia Zero” project is a nationwide multimodal intervention based on the simultaneous implementation of a comprehensive evidence-based bundle measures to prevent ventilator-associated pneumonia in critically ill patients admitted to the ICU. Design: Prospective, interventional, and multicenter study. Setting: A total of 181 ICUs throughout Spain. Patients: All patients admitted for more than 24 hours to the participating ICUs between April 1, 2011, and December 31, 2012. Intervention: Ten ventilator-associated pneumonia prevention measures were implemented (seven were mandatory and three highly recommended). The database of the National ICU-Acquired Infections Surveillance Study (Estudio Nacional de Vigilancia de Infecciones Nosocomiales [ENVIN]) was used for data collection. Ventilator-associated pneumonia rate was expressed as incidence density per 1,000 ventilator days. Ventilator-associated pneumonia rates from the incorporation of the ICUs to the project, every 3 months, were compared with data of the ENVIN registry (April–June 2010) as the baseline period. Ventilator-associated pneumonia rates were adjusted by characteristics of the hospital, including size, type (public or private), and teaching (postgraduate) or university-affiliated (undergraduate) status. Measurements and Main Results: The 181 participating ICUs accounted for 75% of all ICUs in Spain. In a total of 171,237 ICU admissions, an artificial airway was present on 505,802 days (50.0% of days of stay in the ICU). A total of 3,474 ventilator-associated pneumonia episodes were diagnosed in 3,186 patients. The adjusted ventilator-associated pneumonia incidence density rate decreased from 9.83 (95% CI, 8.42–11.48) per 1,000 ventilator days in the baseline period to 4.34 (95% CI, 3.22–5.84) after 19–21 months of participation. Conclusions: Implementation of the bundle measures included in the “Pneumonia Zero” project resulted in a significant reduction of more than 50% of the incidence of ventilator-associated pneumonia in Spanish ICUs. This reduction was sustained 21 months after implementation. PMID:29023261

  20. NanoClusters Enhance Drug Delivery in Mechanical Ventilation

    NASA Astrophysics Data System (ADS)

    Pornputtapitak, Warangkana

    The overall goal of this thesis was to develop a dry powder delivery system for patients on mechanical ventilation. The studies were divided into two parts: the formulation development and the device design. The pulmonary system is an attractive route for drug delivery since the lungs have a large accessible surface area for treatment or drug absorption. For ventilated patients, inhaled drugs have to successfully navigate ventilator tubing and an endotracheal tube. Agglomerates of drug nanoparticles (also known as 'NanoClusters') are fine dry powder aerosols that were hypothesized to enable drug delivery through ventilator circuits. This Thesis systematically investigated formulations of NanoClusters and their aerosol performance in a conventional inhaler and a device designed for use during mechanical ventilation. These engineered powders of budesonide (NC-Bud) were delivered via a MonodoseRTM inhaler or a novel device through commercial endotracheal tubes, and analyzed by cascade impaction. NC-Bud had a higher efficiency of aerosol delivery compared to micronized stock budesonide. The delivery efficiency was independent of ventilator parameters such as inspiration patterns, inspiration volumes, and inspiration flow rates. A novel device designed to fit directly to the ventilator and endotracheal tubing connections and the MonodoseRTM inhaler showed the same efficiency of drug delivery. The new device combined with NanoCluster formulation technology, therefore, allowed convenient and efficient drug delivery through endotracheal tubes. Furthermore, itraconazole (ITZ), a triazole antifungal agent, was formulated as a NanoCluster powder via milling (top-down process) or precipitation (bottom-up process) without using any excipients. ITZ NanoClusters prepared by wet milling showed better aerosol performance compared to micronized stock ITZ and ITZ NanoClusters prepared by precipitation. ITZ NanoClusters prepared by precipitation methods also showed an amorphous state while milled ITZ NanoClusters maintained the crystalline character. Overall, NanoClusters prepared by various processes represent a potential engineered drug particle approach for inhalation therapy since they provide effective aerosol properties and stability due to the crystalline state of the drug powders. Future work will continue to explore formulation and delivery performance in vitro and in vivo..

  1. Randomized clinical trial of extended use of a hydrophobic condenser humidifier: 1 vs. 7 days.

    PubMed

    Thomachot, Laurent; Leone, Marc; Razzouk, Karim; Antonini, François; Vialet, Renaud; Martin, Claude

    2002-01-01

    To determine whether extended use (7 days) would affect the efficiency on heat and water preservation of a hydrophobic condenser humidifier as well as the rate of ventilation-acquired pneumonia, compared with 1 day of use. Prospective, controlled, randomized, not blinded, clinical study. Twelve-bed intensive care unit of a university hospital. One hundred and fifty-five consecutive patients undergoing mechanical ventilation for > or = 48 hrs. After randomization, patients were allocated to one of the two following groups: a) heat and moisture exchangers (HMEs) changed every 24 hrs; b) HMEs changed only once a week. Devices in both groups could be changed at the discretion of the staff when signs of occlusion or increased resistance were identified. Efficient airway humidification and heating were assessed by clinical variables (numbers of tracheal suctionings and instillations required, peak and mean airway pressures). The frequency rates of bronchial colonization and ventilation-acquired pneumonia were evaluated by using clinical and microbiological criteria. Endotracheal tube occlusion, ventilatory support variables, duration of mechanical ventilation, length of intensive care, acquired multiorgan dysfunction, and mortality rates also were recorded. The two groups were similar at the time of randomization. Endotracheal tube occlusion never occurred. In the targeted population (patients ventilated for > or = 7 days), the frequency rate of ventilation-acquired pneumonia was 24% in the HME 1-day group and 17% in the HME 7-day group (p > .05, not significant). Ventilation-acquired pneumonia rates per 1000 ventilatory support days were 16.4/1000 in the HME 1-day group and 12.4/1000 in the HME 7-day group (p > .05, not significant). No statistically significant differences were found between the two groups for duration of mechanical ventilation, intensive care unit length of stay, acquired organ system derangements, and mortality rate. There was indirect evidence of very little, if any, change in HME resistance. Changing the studied hydrophobic HME after 7 days did not affect efficiency, increase resistance, or altered bacterial colonization. The frequency rate of ventilation-acquired pneumonia was also unchanged. Use of HMEs for > 24 hrs and up to 7 days is safe.

  2. A high sensitivity fiber optic macro-bend based gas flow rate transducer for low flow rates: Theory, working principle, and static calibration

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Schena, Emiliano; Saccomandi, Paola; Silvestri, Sergio

    2013-02-15

    A novel fiber optic macro-bend based gas flowmeter for low flow rates is presented. Theoretical analysis of the sensor working principle, design, and static calibration were performed. The measuring system consists of: an optical fiber, a light emitting diode (LED), a Quadrant position sensitive Detector (QD), and an analog electronic circuit for signal processing. The fiber tip undergoes a deflection in the flow, acting like a cantilever. The consequent displacement of light spot center is monitored by the QD generating four unbalanced photocurrents which are function of fiber tip position. The analog electronic circuit processes the photocurrents providing voltage signalmore » proportional to light spot position. A circular target was placed on the fiber in order to increase the sensing surface. Sensor, tested in the measurement range up to 10 l min{sup -1}, shows a discrimination threshold of 2 l min{sup -1}, extremely low fluid dynamic resistance (0.17 Pa min l{sup -1}), and high sensitivity, also at low flow rates (i.e., 33 mV min l{sup -1} up to 4 l min{sup -1} and 98 mV min l{sup -1} from 4 l min{sup -1} up to 10 l min{sup -1}). Experimental results agree with the theoretical predictions. The high sensitivity, along with the reduced dimension and negligible pressure drop, makes the proposed transducer suitable for medical applications in neonatal ventilation.« less

  3. Design and Development of a Regenerative Blower for EVA Suit Ventilation

    NASA Technical Reports Server (NTRS)

    Izenson, Michael G.; Chen, Weibo; Hill, Roger W.; Phillips, Scott D.; Paul, Heather L.

    2011-01-01

    Ventilation subsystems in future space suits require a dedicated ventilation fan. The unique requirements for the ventilation fan - including stringent safety requirements and the ability to increase output to operate in buddy mode - combine to make a regenerative blower an attractive choice. This paper describes progress in the design, development, and testing of a regenerative blower designed to meet requirements for ventilation subsystems in future space suits. We have developed analysis methods for the blower s complex, internal flows and identified impeller geometries that enable significant improvements in blower efficiency. We verified these predictions by test, measuring aerodynamic efficiencies of 45% at operating conditions that correspond to the ventilation fan s design point. We have developed a compact motor/controller to drive the blower efficiently at low rotating speed (4500 rpm). Finally, we have assembled a low-pressure oxygen test loop to demonstrate the blower s reliability under prototypical conditions.

  4. Effects of vascular flow and PEEP in a multiple hit model of lung injury in isolated perfused rabbit lungs.

    PubMed

    Piacentini, Enrique; López-Aguilar, Josefina; García-Martín, Carolina; Villagrá, Ana; Saenz-Valiente, Alicia; Murias, Gastón; Fernández-Segoviano, Pilar; Hotchkiss, John R; Blanch, Lluis

    2008-07-01

    High vascular flow aggravates lung damage in animal models of ventilator-induced lung injury. Positive end-expiratory pressure (PEEP) can attenuate ventilator-induced lung injury, but its continued effectiveness in the setting of antecedent lung injury is unclear. The objective of the present study was to evaluate whether the application of PEEP diminishes lung injury induced by concurrent high vascular flow and high alveolar pressures in normal lungs and in a preinjury lung model. Two series of experiments were performed. Fifteen sets of isolated rabbit lungs were randomized into three groups (n = 5): low vascular flow/low PEEP; high vascular flow/low PEEP, and high vascular flow/high PEEP. Subsequently, the same protocol was applied in an additional 15 sets of isolated rabbit lungs in which oleic acid was added to the vascular perfusate to produce mild to moderate lung injury. All lungs were ventilated with peak airway pressure of 30 cm H2O for 30 minutes. Outcome measures included frequency of gross structural failure, pulmonary hemorrhage, edema formation, changes in static compliance, pulmonary vascular resistance, and pulmonary ultrafiltration coefficient. In the context of high vascular flow, application of a moderate level of PEEP reduced pulmonary rupture, edema formation, and lung hemorrhage. The protective effects of PEEP were not observed in lungs concurrently injured with oleic acid. Under these experimental conditions, PEEP attenuates lung injury in the setting of high vascular flow. The protective effect of PEEP is lost in a two-hit model of lung injury.

  5. Ventilator-associated pneumonia rates in critical care units in 3 Arabian Gulf countries: A 6-year surveillance study.

    PubMed

    El-Saed, Aiman; Al-Jardani, Amina; Althaqafi, Abdulhakeem; Alansari, Huda; Alsalman, Jameela; Al Maskari, Zaina; El Gammal, Ayman; Al Nasser, Wafa; Al-Abri, Seif S; Balkhy, Hanan H

    2016-07-01

    Data estimating the rates of ventilator-associated pneumonia (VAP) in critical patients in Gulf Cooperation Council (GCC) countries are very limited. The aim of this study was to estimate VAP rates in GCC hospitals and to compare rates with published reports of the U.S. National Healthcare Safety Network (NHSN) and International Nosocomial Infection Control Consortium (INICC). VAP rates and ventilator utilization between 2008 and 2013 were calculated from aggregate VAP surveillance data using NHSN methodology pooled from 6 hospitals in 3 GCC countries: Saudi Arabia, Oman, and Bahrain. The standardized infection ratios of VAP in GCC hospitals were compared with published reports of the NHSN and INICC. A total of 368 VAP events were diagnosed during a 6-year period covering 76,749 ventilator days and 134,994 patient days. The overall VAP rate was 4.8 per 1,000 ventilator days (95% confidence interval, 4.3-5.3), with an overall ventilator utilization of 0.57. The VAP rates showed a wide variability between different types of intensive care units (ICUs) and were decreasing over time. After adjusting for the differences in ICU type, the risk of VAP in GCC hospitals was 217% higher than NHSN hospitals and 69% lower than INICC hospitals. The risk of VAP in ICU patients in GCC countries is higher than pooled U.S. VAP rates but lower than pooled rates from developing countries participating in the INICC. Copyright © 2016 Association for Professionals in Infection Control and Epidemiology, Inc. Published by Elsevier Inc. All rights reserved.

  6. Indoor thermal environment, air exchange rates, and carbon dioxide concentrations before and after energy retro fits in Finnish and Lithuanian multi-family buildings.

    PubMed

    Leivo, Virpi; Prasauskas, Tadas; Du, Liuliu; Turunen, Mari; Kiviste, Mihkel; Aaltonen, Anu; Martuzevicius, Dainius; Haverinen-Shaughnessy, Ulla

    2018-04-15

    Impacts of energy retrofits on indoor thermal environment, i.e. temperature (T) and relative humidity (RH), as well as ventilation rates and carbon dioxide (CO 2 ) concentrations, were assessed in 46 Finnish and 20 Lithuanian multi-family buildings, including 39 retrofitted case buildings in Finland and 15 in Lithuania (the remaining buildings were control buildings with no retrofits). In the Finnish buildings, high indoor T along with low RH levels was commonly observed both before and after the retrofits. Ventilation rates (l/s per person) were higher after the retrofits in buildings with mechanical exhaust ventilation than the corresponding values before the retrofits. Measured CO 2 levels were low in vast majority of buildings. In Lithuania, average indoor T levels were low before the retrofits and there was a significant increase in the average T after the retrofits. In addition, average ventilation rate was lower and CO 2 levels were higher after the retrofits in the case buildings (N=15), both in apartments with natural and mixed ventilation. Based on the results, assessment of thermal conditions and ventilation rates after energy retrofits is crucial for optimal indoor environmental quality and energy use. Copyright © 2017 Elsevier B.V. All rights reserved.

  7. Effects of Classroom Ventilation Rate and Temperature on Students' Test Scores.

    PubMed

    Haverinen-Shaughnessy, Ulla; Shaughnessy, Richard J

    2015-01-01

    Using a multilevel approach, we estimated the effects of classroom ventilation rate and temperature on academic achievement. The analysis is based on measurement data from a 70 elementary school district (140 fifth grade classrooms) from Southwestern United States, and student level data (N = 3109) on socioeconomic variables and standardized test scores. There was a statistically significant association between ventilation rates and mathematics scores, and it was stronger when the six classrooms with high ventilation rates that were indicated as outliers were filtered (> 7.1 l/s per person). The association remained significant when prior year test scores were included in the model, resulting in less unexplained variability. Students' mean mathematics scores (average 2286 points) were increased by up to eleven points (0.5%) per each liter per second per person increase in ventilation rate within the range of 0.9-7.1 l/s per person (estimated effect size 74 points). There was an additional increase of 12-13 points per each 1°C decrease in temperature within the observed range of 20-25°C (estimated effect size 67 points). Effects of similar magnitude but higher variability were observed for reading and science scores. In conclusion, maintaining adequate ventilation and thermal comfort in classrooms could significantly improve academic achievement of students.

  8. Effects of Classroom Ventilation Rate and Temperature on Students’ Test Scores

    PubMed Central

    2015-01-01

    Using a multilevel approach, we estimated the effects of classroom ventilation rate and temperature on academic achievement. The analysis is based on measurement data from a 70 elementary school district (140 fifth grade classrooms) from Southwestern United States, and student level data (N = 3109) on socioeconomic variables and standardized test scores. There was a statistically significant association between ventilation rates and mathematics scores, and it was stronger when the six classrooms with high ventilation rates that were indicated as outliers were filtered (> 7.1 l/s per person). The association remained significant when prior year test scores were included in the model, resulting in less unexplained variability. Students’ mean mathematics scores (average 2286 points) were increased by up to eleven points (0.5%) per each liter per second per person increase in ventilation rate within the range of 0.9–7.1 l/s per person (estimated effect size 74 points). There was an additional increase of 12–13 points per each 1°C decrease in temperature within the observed range of 20–25°C (estimated effect size 67 points). Effects of similar magnitude but higher variability were observed for reading and science scores. In conclusion, maintaining adequate ventilation and thermal comfort in classrooms could significantly improve academic achievement of students. PMID:26317643

  9. Assessment of ventilation and indoor air pollutants in nursery and elementary schools in France.

    PubMed

    Canha, N; Mandin, C; Ramalho, O; Wyart, G; Ribéron, J; Dassonville, C; Hänninen, O; Almeida, S M; Derbez, M

    2016-06-01

    The aim of this study was to characterize the relationship between Indoor Air Quality (IAQ) and ventilation in French classrooms. Various parameters were measured over one school week, including volatile organic compounds, aldehydes, particulate matter (PM2.5 mass concentration and number concentration), carbon dioxide (CO2 ), air temperature, and relative humidity in 51 classrooms at 17 schools. The ventilation was characterized by several indicators, such as the air exchange rate, ventilation rate (VR), and air stuffiness index (ICONE), that are linked to indoor CO2 concentration. The influences of the season (heating or non-heating), type of school (nursery or elementary), and ventilation on the IAQ were studied. Based on the minimum value of 4.2 l/s per person required by the French legislation for mechanically ventilated classrooms, 91% of the classrooms had insufficient ventilation. The VR was significantly higher in mechanically ventilated classrooms compared with naturally ventilated rooms. The correlations between IAQ and ventilation vary according to the location of the primary source of each pollutant (outdoor vs. indoor), and for an indoor source, whether it is associated with occupant activity or continuous emission. © 2015 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

  10. Design of evaporative-cooling roof for decreasing air temperatures in buildings in the humid tropics

    NASA Astrophysics Data System (ADS)

    Kindangen, Jefrey I.; Umboh, Markus K.

    2017-03-01

    This subject points to assess the benefits of the evaporative-cooling roof, particularly for buildings with corrugated zinc roofs. In Manado, many buildings have roofed with corrugated zinc sheets; because this material is truly practical, easy and economical application. In general, to achieve thermal comfort in buildings in a humid tropical climate, people applying cross ventilation to cool the air in the room and avoid overheating. Cross ventilation is a very popular path to achieve thermal comfort; yet, at that place are other techniques that allow reducing the problem of excessive high temperature in the room in the constructions. This study emphasizes applications of the evaporative-cooling roof. Spraying water on the surface of the ceiling has been executed on the test cell and the reuse of water after being sprayed and cooled once more by applying a heat exchanger. Initial results indicate a reliable design and successfully meet the target as an effective evaporative-cooling roof technique. Application of water spraying automatic and cooling water installations can work optimally and can be an optimal model for the cooling roof as one of the green technologies. The role of heat exchangers can lower the temperature of the water from spraying the surface of the ceiling, which has become a hot, down an average of 0.77° C. The mass flow rate of the cooling water is approximately 1.106 kg/h and the rate of heat flow is around 515 Watt, depend on the site.

  11. [Heat and moisture exchangers for conditioning of inspired air of intubated patients in intensive care. The humidification properties of passive air exchangers under clinical conditions].

    PubMed

    Rathgeber, J; Züchner, K; Kietzmann, D; Weyland, W

    1995-04-01

    Heat and moisture exchangers (HME) are used as artificial noses for intubated patients to prevent tracheo-bronchial or pulmonary damage resulting from dry and cold inspired gases. HME are mounted directly on the tracheal tube, where they collect a large fraction of the heat and moisture of the expired air, adding this to the subsequent inspired breath. The effective performance depends on the water-retention capacity of the HME: the amount of water added to the inspired gas cannot exceed the stored water uptake of the previous breath. This study evaluates the efficiency of four different HME under laboratory and clinical conditions using a new moisture-measuring device. METHODS. In a first step, the absolute efficiency of four different HME (DAR Hygrobac, Gibeck Humid-Vent 2P, Pall BB 22-15 T, and Pall BB 100) was evaluated using a lung model simulating physiological heat and humidity conditions of the upper airways. The model was ventilated with tidal volumes of 500, 1,000, and 1,500 ml and different flow rates. The water content of the ventilated air was determined between tracheal tube and HME using a new high-resolution humidity meter and compared with the absolute water loss of the exhaled air at the gas outlet of a Siemens Servo C ventilator measured with a dew-point hygrometer. Secondly, the moisturizing efficiency was evaluated under clinical conditions in an intensive care unit with 25 intubated patients. Maintaining the ventilatory conditions for each patient, the HME were randomly changed. The humidity data were determined as described above and compared with the laboratory findings. RESULTS AND DISCUSSION. The water content at the respirator outlet is inversely equivalent to the humidity of the inspired gases and represents the water loss from the respiratory tract if the patient is ventilated with dry gases. Moisture retention and heating capacity decreased with higher volumes and higher flow rates. These data are simple to obtain without affecting the patient and can easily be interpreted. It was demonstrated that, compared to physiological conditions, the DAR Hygrobac and Gibeck Humid Vent 2P-HME coated with hygroscopic salts-were able to maintain sufficient inspiratory humidity and heat. The Pall-HME, solely a condensation humidifier, did not meet the physiological requirements.

  12. Momentum rate probe for use with two-phase flows

    NASA Astrophysics Data System (ADS)

    Bush, S. G.; Bennett, J. B.; Sojka, P. E.; Panchagnula, M. V.; Plesniak, M. W.

    1996-05-01

    An instrument for measuring the momentum rate of two-phase flows is described, and design and construction details are provided. The device utilizes a conelike body to turn the flow from the axial to the radial direction. The force resulting from the change in momentum rate of the turning flow is measured using a strain-gage-instrumented cantilevered beam. The instrument is applicable to a wide range of flows including nuclear reactor coolant streams, refrigerants in heating-ventilating air-conditioning equipment, impingement cooling of small scale electronic hardware (computer chips are one example), supercritical fuel injection (in Diesel engines, for instance), and consumer product sprays (such as hair-care product sprays produced using effervescent atomizers). The latter application is discussed here. Features of the instrument include sensitivity to a wide range of forces and the ability to damp oscillations of the deflection cone. Instrument sensitivity allows measurement of momentum rates considerably lower (below 0.01 N) than those that could be obtained using previous devices. This feature is a direct result of our use of precision strain gages, capable of sensing strains below 20 μm/m, and the damping of oscillations which can overwhelm the force measurements. Oscillation damping results from a viscous fluid damper whose resistance is easily varied by changing fluids. Data used to calibrate the instrument are presented to demonstrate the effectiveness of the technique. As an example of the instrument's utility, momentum rate data obtained using it will be valuable in efforts to explain entrainment of surrounding air into effervescent atomizer-produced sprays and also to model the effervescent atomization process.

  13. Convective heat transfer from circular cylinders located within perforated cylindrical shrouds

    NASA Technical Reports Server (NTRS)

    Daryabeigi, K.; Ash, R. L.

    1986-01-01

    The influence of perforated cylindrical shrouds on the convective heat transfer to circular cylinders in transverse flow has been studied experimentally. Geometries studied were similar to those used in industrial platinum resistance thermometers. The influence of Reynolds number, ventilation factor (ratio of the open area to the total surface area of shroud), radius ratio (ratio of shroud's inside radius to bare cylinder's radius), and shroud orientation with respect to flow were studied. The experiments showed that perforated shrouds with ventilation factors in the range 0.1 to 0.4 and radius ratios in the range 1.1 to 2.1 could enhance the convective heat transfer to bare cylinders up to 50%. The maximum enhancement occurred for a radius ratio of 1.4 and ventilation factors between 0.2 and 0.3. It was found that shroud orientation influenced the heat transfer, with maximum heat transfer generally occurring when the shroud's holes were centered on either side of the stagnation line. However, the hole orientation effect is of second order compared to the influence of ventilation factor and radius ratio.

  14. Increased dead space in face mask continuous positive airway pressure in neonates.

    PubMed

    Hishikawa, Kenji; Fujinaga, Hideshi; Ito, Yushi

    2017-01-01

    Continuous positive airway pressure (CPAP) by face mask is commonly performed in newborn resuscitation. We evaluated the effect of face mask CPAP on system dead space. Face mask CPAP increases dead space. A CPAP model study. We estimated the volume of the inner space of the mask. We devised a face mask CPAP model, in which the outlet of the mask was covered with plastic; and three modified face mask CPAP models, in which holes were drilled near to the cushion of the covered face mask to alter the air exit. We passed a continuous flow of 21% oxygen through each model and we controlled the inner pressure to 5 cmH 2 O by adjusting the flow-relief valve. To evaluate the ventilation in the inner space of each model, we measured the oxygen concentration rise time, that is, the time needed for the oxygen concentration of each model to reach 35% after the oxygen concentration of the continuous flow was raised from 21% to 40%. The volume of inner space of the face mask was 38.3 ml. Oxygen concentration rise time in the face mask CPAP model was significantly longer at various continuous flow rates and points of the inner space of the face mask compared with that of the modified face mask CPAP model. Our study indicates that face mask CPAP leads to an increase in dead space and a decrease in ventilation efficiency under certain circumstances. Pediatr Pulmonol. 2017;52:107-111. © 2016 Wiley Periodicals, Inc. © 2016 Wiley Periodicals, Inc.

  15. SponTaneous Respiration using IntraVEnous anaesthesia and Hi-flow nasal oxygen (STRIVE Hi) maintains oxygenation and airway patency during management of the obstructed airway: an observational study.

    PubMed

    Booth, A W G; Vidhani, K; Lee, P K; Thomsett, C-M

    2017-03-01

    High-flow nasal oxygen (HFNO) has been shown to benefit oxygenation, ventilation and upper airway patency in a range of clinical scenarios, however its use in spontaneously breathing patients during general anaesthesia has not been described. Spontaneous respiration using i.v. anaesthesia is the primary technique used at our institution for tubeless airway surgery. We hypothesized that the addition of HFNO would increase our margin of safety, particularly during management of an obstructed airway. A retrospective observational study was conducted using a SponTaneous Respiration using IntraVEnous anaesthesia and High-flow nasal oxygen (STRIVE Hi) technique to manage 30 adult patients undergoing elective laryngotracheal surgery. Twenty-six patients (87%) presented with significant airway and/or respiratory compromise (16 were stridulous, 10 were dyspnoeic). No episodes of apnoea or complete airway obstruction occurred during the induction of anaesthesia using STRIVE Hi. The median [IQR (range)] lowest oxygen saturation during the induction period was 100 [99–100 (97–100)] %. The median [IQR (range)] overall duration of spontaneous ventilation was 44 [40–49.5 (18–100)] min. The median [IQR (range)] end-tidal carbon dioxide (ETCO2) level at the end of the spontaneous ventilation period was 6.8 [6.4–7.1 (4.8–8.9)] kPa. The mean rate of increase in ETCO2 was 0.03 kPa min−1. STRIVE Hi succeeded in preserving adequate oxygen saturation, end-tidal carbon dioxide and airway patency. We suggest that the upper and lower airway benefits attributed to HFNO, are ideally suited to a spontaneous respiration induction, increasing its margin of safety. STRIVE Hi is a modern alternative to the traditional inhalation induction. © The Author 2017. Published by Oxford University Press on behalf of the British Journal of Anaesthesia.

  16. Night ventilation control strategies in office buildings

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Wang, Zhaojun; Yi, Lingli; Gao, Fusheng

    2009-10-15

    In moderate climates night ventilation is an effective and energy-efficient approach to improve the indoor thermal environment for office buildings during the summer months, especially for heavyweight construction. However, is night ventilation a suitable strategy for office buildings with lightweight construction located in cold climates? In order to answer this question, the whole energy-consumption analysis software EnergyPlus was used to simulate the indoor thermal environment and energy consumption in typical office buildings with night mechanical ventilation in three cities in northern China. The summer outdoor climate data was analyzed, and three typical design days were chosen. The most important factorsmore » influencing night ventilation performance such as ventilation rates, ventilation duration, building mass and climatic conditions were evaluated. When night ventilation operation time is closer to active cooling time, the efficiency of night ventilation is higher. With night ventilation rate of 10 ach, the mean radiant temperature of the indoor surface decreased by up to 3.9 C. The longer the duration of operation, the more efficient the night ventilation strategy becomes. The control strategies for three locations are given in the paper. Based on the optimized strategies, the operation consumption and fees are calculated. The results show that more energy is saved in office buildings cooled by a night ventilation system in northern China than ones that do not employ this strategy. (author)« less

  17. Effect of prone positioning on cannula function and impaired oxygenation during extracorporeal circulation.

    PubMed

    Masuda, Yoshiki; Tatsumi, Hiroomi; Imaizumi, Hitoshi; Gotoh, Kyoko; Yoshida, Shinichiro; Chihara, Shinya; Takahashi, Kanako; Yamakage, Michiaki

    2014-03-01

    Prone ventilation is an effective method for improving oxygenation in patients with acute respiratory failure. However, in extracorporeal circulation, there is a risk of cannula-related complications when changing the position. In this study, we investigated cannula-related complications when changing position for prone ventilation and the effect of prone ventilation on impaired oxygenation in patients who underwent extracorporeal membrane oxygenation (ECMO). The study subjects were patients who underwent prone ventilation during ECMO in the period from 2004 to 2011. Indication for prone ventilation was the presence of dorsal infiltration shown by lung computed tomography. Factors investigated were cannula insertion site, dislodgement or obstruction of the cannula, malfunction of vascular access and unplanned dislodgement of the catheters when changing position. Mean arterial pressure, PaO2/FiO2, PEEP level, blood flow and rotation speed of the pump were also determined before and after position change. Five patients were selected as study subjects. The mean duration of prone positioning was 15.3 ± 0.5 h. Strict management during position changes prevented cannula-related complications in the patients who underwent extracorporeal circulation. There were no significant changes in mean arterial pressure, PEEP level, blood flow and rotation speed of the pump when changing position. Low PaO2/FiO2 prior to prone ventilation was significantly increased after supine to prone and then prone to supine position. Prone positioning to improve impaired oxygenation is a safe procedure and not a contraindication in patients receiving extracorporeal circulation.

  18. Metabolically-Derived Human Ventilation Rates: A Revised Approach Based Upon Oxygen Consumption Rates (External Review Draft)

    EPA Science Inventory

    EPA has released a draft report entitled, Metabolically-Derived Human Ventilation Rates: A Revised Approach Based Upon Oxygen Consumption Rates, for independent external peer review and public comment. NCEA published the Exposure Factors Handbook in 1997. This comprehens...

  19. Carbon dioxide rebreathing during non-invasive ventilation delivered by helmet: a bench study.

    PubMed

    Mojoli, Francesco; Iotti, Giorgio A; Gerletti, Maddalena; Lucarini, Carlo; Braschi, Antonio

    2008-08-01

    To define how to monitor and limit CO(2) rebreathing during helmet ventilation. Physical model study. Laboratory in a university teaching hospital. We applied pressure-control ventilation to a helmet mounted on a physical model. In series 1 we increased CO(2) production (V'CO(2)) from 100 to 550 ml/min and compared mean inhaled CO(2) (iCO(2),mean) with end-inspiratory CO(2) at airway opening (eiCO(2)), end-tidal CO(2) at Y-piece (yCO(2)) and mean CO(2) inside the helmet (hCO(2)). In series 2 we observed, at constant V'CO(2), effects on CO(2) rebreathing of inspiratory pressure, respiratory mechanics, the inflation of cushions inside the helmet and the addition of a flow-by. In series 1, iCO(2),mean linearly related to V'CO(2). The best estimate of CO(2) rebreathing was provided by hCO(2): differences between iCO(2),mean and hCO(2), yCO(2) and eiCO(2) were 0.0+/-0.1, 0.4+/-0.2 and -1.3+/-0.5%. In series 2, hCO(2) inversely related to the total ventilation (MVtotal) delivered to the helmet-patient unit. The increase in inspiratory pressure significantly increased MVtotal and lowered hCO(2). The low lung compliance halved the patient:helmet ventilation ratio but led to minor changes in MVtotal and hCO(2). Cushion inflation, although it decreased the helmet's internal volume by 33%, did not affect rebreathing. A 8-l/min flow-by effectively decreased hCO(2). During helmet ventilation, rebreathing can be assessed by measuring hCO(2) or yCO(2), but not eiCO(2). It is directly related to V'CO(2), inversely related to MVtotal and can be lowered by increasing inspiratory pressure or adding a flow-by.

  20. Method for calculating internal radiation and ventilation with the ADINAT heat-flow code

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Butkovich, T.R.; Montan, D.N.

    1980-04-01

    One objective of the spent fuel test in Climax Stock granite (SFTC) is to correctly model the thermal transport, and the changes in the stress field and accompanying displacements from the application of the thermal loads. We have chosen the ADINA and ADINAT finite element codes to do these calculations. ADINAT is a heat transfer code compatible to the ADINA displacement and stress analysis code. The heat flow problem encountered at SFTC requires a code with conduction, radiation, and ventilation capabilities, which the present version of ADINAT does not have. We have devised a method for calculating internal radiation andmore » ventilation with the ADINAT code. This method effectively reproduces the results from the TRUMP multi-dimensional finite difference code, which correctly models radiative heat transport between drift surfaces, conductive and convective thermal transport to and through air in the drifts, and mass flow of air in the drifts. The temperature histories for each node in the finite element mesh calculated with ADINAT using this method can be used directly in the ADINA thermal-mechanical calculation.« less

  1. Assessment and monitoring of flow limitation and other parameters from flow/volume loops.

    PubMed

    Dueck, R

    2000-01-01

    Flow/volume (F/V) spirometry is routinely used for assessing the type and severity of lung disease. Forced vital capacity (FVC) and timed vital capacity (FEV1) provide the best estimates of airflow obstruction in patients with asthma, chronic obstructive pulmonary disease (COPD) and emphysema. Computerized spirometers are now available for early home recognition of asthma exacerbation in high risk patients with severe persistent disease, and for recognition of either infection or rejection in lung transplant patients. Patients with severe COPD may exhibit expiratory flow limitation (EFL) on tidal volume (VT) expiratory F/V (VTF/V) curves, either with or without applying negative expiratory pressure (NEP). EFL results in dynamic hyperinflation and persistently raised alveolar pressure or intrinsic PEEP (PEEPi). Hyperinflation and raised PEEPi greatly enhance dyspnea with exertion through the added work of the threshold load needed to overcome raised pleural pressure. Esophageal (pleural) pressure monitoring may be added to VTF/V loops for assessing the severity of PEEPi: 1) to optimize assisted ventilation by mask or via endotracheal tube with high inspiratory flow rates to lower I:E ratio, and 2) to assess the efficacy of either pressure support ventilation (PSV) or low level extrinsic PEEP in reducing the threshold load of PEEPi. Intraoperative tidal volume F/V loops can also be used to document the efficacy of emphysema lung volume reduction surgery (LVRS) via disappearance of EFL. Finally, the mechanism of ventilatory constraint can be identified with the use of exercise tidal volume F/V loops referenced to maximum F/V loops and static lung volumes. Patients with severe COPD show inspiratory F/V loops approaching 95% of total lung capacity, and flow limitation over the entire expiratory F/V curve during light levels of exercise. Surprisingly, patients with a history of congestive heart failure may lower lung volume towards residual volume during exercise, thereby reducing airway diameter and inducing expiratory flow limitation.

  2. A gas flow indicator for portable life support systems

    NASA Technical Reports Server (NTRS)

    Bass, R. L., III; Schroeder, E. C.

    1975-01-01

    A three-part program was conducted to develop a gas flow indicator (GFI) to monitor ventilation flow in a portable life support system. The first program phase identified concepts which could potentially meet the GFI requirements. In the second phase, a working breadboard GFI, based on the concept of a pressure sensing diaphragm-aneroid assembly connected to a venturi, was constructed and tested. Extensive testing of the breadboard GFI indicated that the design would meet all NASA requirements including eliminating problems experienced with the ventilation flow sensor used in the Apollo program. In the third program phase, an optimized GFI was designed by utilizing test data obtained on the breadboard unit. A prototype unit was constructed using prototype materials and fabrication techniques, and performance tests indicated that the prototype GFI met or exceeded all requirements.

  3. 30 CFR 57.22212 - Air flow (I-C, II-A, and V-A mines).

    Code of Federal Regulations, 2010 CFR

    2010-07-01

    ... Standards for Methane in Metal and Nonmetal Mines Ventilation § 57.22212 Air flow (I-C, II-A, and V-A mines). Air flow across each working face shall be sufficient to carry away any accumulation of methane, smoke...

  4. High flow nasal cannula oxygen versus noninvasive ventilation in adult acute respiratory failure: a systematic review of randomized-controlled trials.

    PubMed

    Beng Leong, Lim; Wei Ming, Ng; Wei Feng, Lee

    2018-06-19

    We reviewed the use of noninvasive ventilation (NIV) versus high flow nasal cannula (HFNC) oxygen in adult acute respiratory failure (ARF). We searched major databases and included randomized trials comparing at least NIV with HFNC or NIV+HFNC with NIV in ARF. Primary outcomes included intubation/re-intubation rates. Secondary outcomes were ICU mortality and morbidities. Five trials were included; three compared HFNC with NIV, one compared HFNC, NIV and oxygen whereas one compared HFNC+NIV with NIV. Patients had hypoxaemic ARF (PaO2/FiO2≤300 mmHg). Heterogeneity prevented result pooling. Three and two studies had superiority and noninferiority design, respectively. Patients were postcardiothoracic surgery, mixed medical/surgical patients and those with pneumonia. Two trials were conducted after extubation, two before intubation and one during intubation. Three trials reported intubation/re-intubation rates as the primary outcomes. The other two trials reported the lowest peripheral capillary oxygen saturation readings during bronchoscopy or intubation. In the former three trials, the odds ratio for intubation/re-intubation rates between HFNC versus the NIV group ranged from 0.80 (95% confidence interval: 0.54-1.19) to 1.65 (95% confidence interval: 0.96-2.84). In the latter two trials, only one reported a difference in the lowest peripheral capillary oxygen saturation between NIV+HFNC versus the NIV group during intubation [100% (interquartile range: 95-100) vs. 96% (interquartile range: 92-99); P=0.029]. The secondary outcomes included differences in ICU mortality and patient tolerability, favouring HFNC, were conflicting, but highlighted future research directions. These include patients with hypercapneic ARF, more severe hypoxaemia (PaO2/FiO2≤200 mmHg), a superiority design, an oxygen arm and patient-centred outcomes.

  5. The simulated air flow pattern around a moving animal transport vehicle as the basis for a prospective biosecurity risk assessment.

    PubMed

    Seedorf, Jens; Schmidt, Ralf-Gunther

    2017-08-01

    Research that investigates bioaerosol emissions from animal transport vehicles (ATVs) and their importance in the spread of harmful airborne agents while the ATVs travel on roads is limited. To investigate the dynamical behaviour of theoretically released particles from a moving ATV, the open-source computational fluid dynamics (CFD) software OpenFOAM was used to calculate the external and internal air flow fields with passive and forced ventilated openings of a common ATV moving at a speed of 80 km/h. In addition to a computed flow rate of approximately 40,000 m 3 /h crossing the interior of the ATV, the visualization of the trajectories has demonstrated distinct patterns of the spatial distribution of potentially released bioaerosols in the vicinity of the ATV. Although the front openings show the highest air flow to the outside, the recirculations of air masses between the interior of the ATV and the atmosphere also occur, which complicate the emission and the dispersion characterizations. To specify the future emission rates of ATVs, a database of bioaerosol concentrations within the ATV is necessary in conjunction with high-performance computing resources to simulate the potential dispersion of bioaerosols in the environment.

  6. ETS levels in hospitality environments satisfying ASHRAE standard 62-1989: "ventilation for acceptable indoor air quality"

    NASA Astrophysics Data System (ADS)

    Moschandreas, D. J.; Vuilleumier, K. L.

    Prior to this study, indoor air constituent levels and ventilation rates of hospitality environments had not been measured simultaneously. This investigation measured indoor Environmental Tobacco Smoke-related (ETS-related) constituent levels in two restaurants, a billiard hall and a casino. The objective of this study was to characterize ETS-related constituent levels inside hospitality environments when the ventilation rates satisfy the requirements of the ASHRAE 62-1989 Ventilation Standard. The ventilation rate of each selected hospitality environment was measured and adjusted. The study advanced only if the requirements of the ASHRAE 62-1989 Ventilation Standard - the pertinent standard of the American Society of Heating, Refrigeration and Air Conditioning Engineers - were satisfied. The supply rates of outdoor air and occupant density were measured intermittently to assure that the ventilation rate of each facility satisfied the standard under occupied conditions. Six ETS-related constituents were measured: respirable suspended particulate (RSP) matter, fluorescent particulate matter (FPM, an estimate of the ETS particle concentrations), ultraviolet particulate matter (UVPM, a second estimate of the ETS particle concentrations), solanesol, nicotine and 3-ethenylpyridine (3-EP). ETS-related constituent levels in smoking sections, non-smoking sections and outdoors were sampled daily for eight consecutive days at each hospitality environment. This study found that the difference between the concentrations of ETS-related constituents in indoor smoking and non-smoking sections was statistically significant. Differences between indoor non-smoking sections and outdoor ETS-related constituent levels were identified but were not statistically significant. Similarly, differences between weekday and weekend evenings were identified but were not statistically significant. The difference between indoor smoking sections and outdoors was statistically significant. Most importantly, ETS-related constituent concentrations measured indoors did not exceed existing occupational standards. It was concluded that if the measured ventilation rates of the sampled facilities satisfied the ASHRAE 62-1989 Ventilation Standard requirements, the corresponding ETS-related constituents were measured at concentrations below known harmful levels as specified by the American Conference of Governmental Industrial Hygiene (ACGIH).

  7. First experiences with methods to measure ammonia emissions from naturally ventilated cattle buildings in the U.K.

    NASA Astrophysics Data System (ADS)

    Demmers, T. G. M.; Burgess, L. R.; Short, J. L.; Phillips, V. R.; Clark, J. A.; Wathes, C. M.

    A method has been developed to measure the emission rate of ammonia from naturally ventilated U.K. livestock buildings. The method is based on measurements of ammonia concentration and estimates of the ventilation rate of the building by continuous release of carbon monoxide tracer within the building. The tracer concentration is measured at nine positions in openings around the perimeter of the building, as well as around a ring sampling line. Two criteria were evaluated to decide whether, at any given time, a given opening in the building acted as an air inlet or as an air outlet. Carbon monoxide concentration difference across an opening was found to be a better criterion than the temperature difference across the opening. Ammonia concentrations were measured continuously at the sampling points using a chemiluminescence analyser. The method was applied to a straw-bedded beef unit and to a slurry-based dairy unit. Both buildings were of space-boarded construction. Ventilation rates estimated by the ring line sample were consistently higher than by the perimeter samples. During calm weather, the ventilation estimates by both samples were similar (10-20 air changes h -1). However, during windy conditions (>5 m s -1) the ventilation rate was overestimated by the ring line sample (average 100 air changes h -1) compared to the perimeter samples (average 50 air changes h -1). The difference was caused by incomplete mixing of the tracer within the building. The ventilation rate estimated from the perimeter samples was used for the calculation of the emission rate. Preliminary estimates of the ammonia emission factor were 6.0 kg NH 3 (500 kg live-weight) -1 (190 d) -1 for the slurry-based dairy unit and 3.7 for the straw-bedded beef unit.

  8. Evaluation of a Mapleson D CPAP system for weaning of mechanical ventilation in pediatric patients.

    PubMed

    Palomero-Rodríguez, Miguel Angel; de Arteaga, Héctor Chozas; Báez, Yolanda Laporta; de Vicente Sánchez, Jesús; Carretero, Pascual Sanabria; Conde, Pilar Sánchez; Pérez Ferrer, Antonio

    2016-01-01

    Over the last years, we have used a flow-inflating bag circuit with a nasotracheal or nasopharyngeal tube as an interface to deliver effective CPAP support in infants ("Mapleson D CPAP system"). The primary goal of this study was to assess the usefulness of the "Mapleson D CPAP system" for weaning of mechanical ventilation (MV) in infants who received MV over 24 h. All infants who received MV for more than 24 h in the last year were enrolled in the study. Demographic data included age, gender, weight, and admission diagnosis. Heart rate, respiratory rate, blood pressure, and oxygen saturation were measured during MV, 2 h after the nasotracheal Mapleson D CPAP system and 2 h after extubation. Patients were classified into two groups: patients MV more than 48 h, and patients with MV fewer than 48 h. P < 0.05 was considered statistically significant. A total of 50 children were enrolled in the study, with a median age was 34 ± 45 months (range, 1-59 months) and median weight was 11.98 ± 9.31 kg (range, 1-48 kg). Median duration of MV was 480 h (range, 2-570). There were no significant differences in PaO2, PaCO2, and pH among MV, 2 h after the nasotracheal Mapleson D CPAP system and 2 h after extubation and spontaneous ventilation with the nasopharyngeal Mapleson D CPAP system or with nasal prongs. The overall extubation failure rate was 26% (n = 13). Weight and age were significantly associated with extubation failure (P < 0.05). The Mapleson D CPAP system, in our opinion, is a useful and safe alternative to more complex and expensive noninvasive CPAP and BiPAP weaning from MV in infants.

  9. Temperature and velocity conditions of air flow in vertical channel of hinged ventilated facade of a multistory building.

    NASA Astrophysics Data System (ADS)

    Statsenko, Elena; Ostrovaia, Anastasia; Pigurin, Andrey

    2018-03-01

    This article considers the influence of the building's tallness and the presence of mounting grooved lines on the parameters of heat transfer in the gap of a hinged ventilated facade. A numerical description of the processes occurring in a heat-gravitational flow is given. The average velocity and temperature of the heat-gravitational flow of a structure with open and sealed rusts are determined with unchanged geometric parameters of the gap. The dependence of the parameters influencing the thermomechanical characteristics of the enclosing structure is derived depending on the internal parameters of the system. Physical modeling of real multistory structures is performed by projecting actual parameters onto a reduced laboratory model (scaling).

  10. Relationship of structure and function of the avian respiratory system to disease susceptibility.

    PubMed

    Fedde, M R

    1998-08-01

    The avian respiratory system exchanges oxygen and carbon dioxide between the gas and the blood utilizing a relatively small, rigid, flow-through lung, and a system of air sacs that act as bellows to move the gas through the lung. Gas movement through the paleopulmonic parabronchi, the main gas exchanging bronchi, in the lung is in the same direction during both inspiration and expiration, i.e., from the mediodorsal secondary bronchi to the medioventral secondary bronchi. During inspiration, acceleration of the gas at the segmentum accelerans of the primary bronchus increases gas velocity so it does not enter the medioventral secondary bronchi. During expiration, airway resistance is increased in he intrapulmonary primary bronchus because of dynamic compression causing gas to enter the mediodorsal secondary bronchi. Reduction in air flow velocity may decrease the efficiency of this aerodynamic valving and thereby decrease the efficiency of gas exchange. The convective gas flow in the avian parabronchus is orientated at a 90 degree angle with respect to the parabronchial blood flow; hence, the cross-current designation of this gas exchanger. With this design, the partial pressure of oxygen in the blood leaving the parabronchus can be higher than that in the gas exiting this structure, giving the avian lung a high gas exchange efficacy. The relationship of the partial pressure of oxygen in the moist inspired gas to that in the blood leaving the lung is dependent on he rate of ventilation. A low ventilation rate may produce a ow oxygen partial pressure in part of the parabronchus, thereby inducing hypoxic vasoconstriction in the pulmonary arterioles supplying this region. Inhaled foreign particles are removed by nasal mucociliary action, by escalator in the trachea, primary bronchi, and secondary bronchi. Small particles that enter parabronchi appear to be phagocytized by the epithelial cells in eh atria and infundibulum. These particles can e transported to interstitial macrophages but the disposition of the particles from this site is unknown. The predominant site of respiratory infections in the caudal air sacs, compared to other parts of the respiratory system, can be explained by the gas flow pathway and the mechanisms present in the parabronchi for particle removal.

  11. Validation of a new mixing chamber system for breath-by-breath indirect calorimetry.

    PubMed

    Kim, Do-Yeon; Robergs, Robert Andrew

    2012-02-01

    Limited validation research exists for applications of breath-by-breath systems of expired gas analysis indirect calorimetry (EGAIC) during exercise. We developed improved hardware and software for breath-by-breath indirect calorimetry (NEW) and validated this system as well as a commercial system (COM) against 2 methods: (i) mechanical ventilation with known calibration gas, and (ii) human subjects testing for 5 min each at rest and cycle ergometer exercise at 100 and 175 W. Mechanical calibration consisted of medical grade and certified calibration gas ((4.95% CO(2), 12.01% O(2), balance N(2)), room air (20.95% O(2), 0.03% CO(2), balance N(2)), and 100% nitrogen), and an air flow turbine calibrated with a 3-L calibration syringe. Ventilation was mimicked manually using complete 3-L calibration syringe manouvers at a rate of 10·min(-1) from a Douglas bag reservoir of calibration gas. The testing of human subjects was completed in a counterbalanced sequence based on 5 repeated tests of all conditions for a single subject. Rest periods of 5 and 10 min followed the 100 and 175 W conditions, respectively. COM and NEW had similar accuracy when tested with known ventilation and gas fractions. However, during human subjects testing COM significantly under-measured carbon dioxide gas fractions, over-measured oxygen gas fractions and minute ventilation, and resulted in errors to each of oxygen uptake, carbon dioxide output, and respiratory exchange ratio. These discrepant findings reveal that controlled ventilation and gas fractions are insufficient to validate breath-by-breath, and perhaps even time-averaged, systems of EGAIC. The errors of the COM system reveal the need for concern over the validity of commercial systems of EGAIC.

  12. [Intraoperative monitoring in artificial respiration of premature and newborn infants. I. Monitoring of respiratory parameters and alveolar ventilation].

    PubMed

    Lenz, G; Heipertz, W; Leidig, E; Madee, S

    1986-06-01

    Monitoring of ventilation serves to ensure adequate alveolar ventilation and arterial oxygenation, and to avoid pulmonary damage due to mechanical ventilation. Basic clinical monitoring, i.e., inspection, auscultation (including precordial or oesophageal stethoscope) and monitoring of heart rate and blood pressure, is mandatory. Mechanical ventilation is monitored by ventilation pressures (peak pressure, plateau pressure and endexpiratory pressure), ventilation volumes (measured at the in/expiratory valve of the respirator and by hot-wire anemometry at the tube connector), ventilation rate, and inspiratory oxygen concentration (FiO2). Alveolar ventilation should be continuously and indirectly recorded by capnometry (pECO2) and by measurement of transcutaneous pCO2 (tcpCO2), whereas oxygenation is determined via measurement of transcutaneous pO2 (tcpO2). Invasive monitoring of gas exchange is essential in prolonged or intrathoracic interventions as well as in neonates with cardiopulmonary problems. paCO2 may be estimated by capillary or venous blood gas analysis; arterial blood gas analysis is required for exact determination of paCO2 as well as arteriocutaneous pCO2 (atcDCO2) and arterio-end-expiratory (aEDCO2) gradients.

  13. Effect of outside air ventilation rate on VOC concentrations and emissions in a call center

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Hodgson, A.T.; Faulkner, D.; Sullivan, D.P.

    2002-01-01

    A study of the relationship between outside air ventilation rate and concentrations of VOCs generated indoors was conducted in a call center. Ventilation rates were manipulated in the building's four air handling units (AHUs). Concentrations of VOCs in the AHU returns were measured on 7 days during a 13-week period. Indoor minus outdoor concentrations and emission factors were calculated. The emission factor data was subjected to principal component analysis to identify groups of co-varying compounds based on source type. One vector represented emissions of solvents from cleaning products. Another vector identified occupant sources. Direct relationships between ventilation rate and concentrationsmore » were not observed for most of the abundant VOCs. This result emphasizes the importance of source control measures for limiting VOC concentrations in buildings.« less

  14. Employing a Modified Diffuser Momentum Model to Simulate Ventilation of the Orion CEV (DRAFT)

    NASA Technical Reports Server (NTRS)

    Straus, John; Ball, Tyler; OHara, William; Barido, Richard

    2011-01-01

    Computational Fluid Dynamics (CFD) is used to model the flow field in the Orion CEV cabin. The CFD model employs a momentum model used to account for the effect of supply grilles on the supply flow. The momentum model is modified to account for non-uniform velocity profiles at the approach of the supply grille. The modified momentum model is validated against a detailed vane-resolved model before inclusion into the Orion CEV cabin model. Results for this comparison, as well as that of a single ventilation configuration are presented.

  15. Stratification of welding fumes and grinding particles in a large factory hall equipped with displacement ventilation.

    PubMed

    Niemelä, R; Koskela, H; Engström, K

    2001-08-01

    The purpose of the study was to investigate the performance of displacement ventilation in a large factory hall where large components of stainless steel for paper, pulp and chemical industries were manufactured. The performance of displacement ventilation was evaluated in terms of concentration distributions of welding fumes and grinding particles, flow field of the supply air and temperature distributions. Large differences in vertical stratification patterns between hexavalent chromium (Cr(VI)) and other particulate contaminants were observed. The concentration of Cr(VI) was notably lower in the zone of occupancy than in the upper part of the factory hall, whereas the concentrations of total airborne particles and trivalent chromium (Cr(III)) were higher in the occupied zone than in the upper zone. The stratification of Cr(VI) had the same tendency as the air temperature stratification caused by the displacement flow field.

  16. THE EFFECT OF VENTILATION ON EMISSION RATES OF WOOD FINISHING MATERIALS

    EPA Science Inventory

    The rate of emission of organic compounds from building materials varies according to: type of material, material loading (area of material/volume of room), compound emitted, temperature, humidity, and ventilation rate. For some compounds and materials (e.g., formaldehyde from pa...

  17. Computational fluid dynamics study on the influence of an alternate ventilation configuration on the possible flow path of infectious cough aerosols in a mock airborne infection isolation room

    PubMed Central

    THATIPARTI, DEEPTHI SHARAN; GHIA, URMILA; MEAD, KENNETH R.

    2017-01-01

    When infectious epidemics occur, they can be perpetuated within health care settings, potentially resulting in severe health care workforce absenteeism, morbidity, mortality, and economic losses. The ventilation system configuration of an airborne infection isolation room is one factor that can play a role in protecting health care workers from infectious patient bioaerosols. Though commonly associated with airborne infectious diseases, the airborne infection isolation room design can also impact other transmission routes such as short-range airborne as well as fomite and contact transmission routes that are impacted by contagion concentration and recirculation. This article presents a computational fluid dynamics study on the influence of the ventilation configuration on the possible flow path of bioaerosol dispersal behavior in a mock airborne infection isolation room. At first, a mock airborne infection isolation room was modeled that has the room geometry and layout, ventilation parameters, and pressurization corresponding to that of a traditional ceiling-mounted ventilation arrangement observed in existing hospitals. An alternate ventilation configuration was then modeled to retain the linear supply diffuser in the original mock airborne infection isolation room but interchanging the square supply and exhaust locations to place the exhaust closer to the patient source and allow clean air from supply vents to flow in clean-to-dirty flow paths, originating in uncontaminated parts of the room prior to entering the contaminated patient’s air space. The modeled alternate airborne infection isolation room ventilation rate was 12 air changes per hour. Two human breathing models were used to simulate a source patient and a receiving health care worker. A patient cough cycle was introduced into the simulation, and the airborne infection dispersal was tracked in time using a multi-phase flow simulation approach. The results from the alternate configuration revealed that the cough aerosols were pulled by the exhaust vent without encountering the health care worker by 0.93 s after patient coughs and the particles were controlled as the aerosols’ flow path was uninterrupted by an air particle streamline from patient to the ceiling exhaust venting out cough aerosols. However, not all the aerosols were vented out of the room. The remaining cough aerosols entered the health care worker’s breathing zone by 0.98 s. This resulted in one of the critical stages in terms of the health care worker’s exposure to airborne virus and presented the opportunity for the health care worker to suffer adverse health effects from the inhalation of cough aerosols. Within 2 s, the cough aerosols reentered and recirculated within the patient and health care worker’s surroundings resulting in pockets of old contaminated air. By this time, coalescence losses decreased as the aerosol were no longer in very close proximity and their movement was primarily influenced by the airborne infection isolation room airflow patterns. In the patient and health care worker’s area away from the supply, the fresh air supply failed to reach this part of the room to quickly dilute the cough aerosol concentration. The exhaust was also found to have minimal effect upon cough aerosol removal, except for those areas with high exhaust velocities, very close to the exhaust grill. Within 5–20 s after a patient’s cough, the aerosols tended to break up to form smaller sized aerosols of less than one micron diameter. They remained airborne and entrained back into the supply air stream, spreading into the entire room. The suspended aerosols resulted in the floating time of more than 21 s in the room due to one cough cycle. The duration of airborne contagion in the room and its prolonged exposure to the health care worker is likely to happen due to successive coughing cycles. Hence, the evaluated alternate airborne infection isolation room is not effective in removing at least 38% particles exposed to health care worker within the first second of a patient’s cough. PMID:28736744

  18. [The effect of the ventilation rate on air particle and air microbe concentration in operating rooms with conventional ventilation. 1. Measurement without surgical activity].

    PubMed

    Kruppa, B; Rüden, H

    1993-05-01

    The question was if a reduction of airborne particles and bacteria in conventionally (turbulently), ventilated operating theatres in comparison to Laminar-Airflow (LAF) operating theatres does occur at high air-exchange-rates. Within the framework of energy consumption measures the influence of air-exchange-rates on airborne particle and bacteria concentrations was determined in two identical operating theatres with conventional ventilation (wall diffusor panel) at the air-exchange-rates 7.5, 10, 15 and 20/h without surgical activity. This was established by means of the statistical procedure of analysis of variance. Especially for the comparison of the air-exchange-rates 7.5 and 15/h statistical differences were found for airborne particle concentrations in supply and ambient air. Concerning airborne bacteria concentrations no differences were found among the various air-exchange-rates. Explanation of variance is quite high for non-viable particles (supply air: 37%, ambient air: 81%) but negligible for viable particles (bacteria) with values below 15%.

  19. Epidemiology of Noninvasive Ventilation in Pediatric Cardiac ICUs.

    PubMed

    Romans, Ryan A; Schwartz, Steven M; Costello, John M; Chanani, Nikhil K; Prodhan, Parthak; Gazit, Avihu Z; Smith, Andrew H; Cooper, David S; Alten, Jeffrey; Mistry, Kshitij P; Zhang, Wenying; Donohue, Janet E; Gaies, Michael

    2017-10-01

    To describe the epidemiology of noninvasive ventilation therapy for patients admitted to pediatric cardiac ICUs and to assess practice variation across hospitals. Retrospective cohort study using prospectively collected clinical registry data. Pediatric Cardiac Critical Care Consortium clinical registry. Patients admitted to cardiac ICUs at PC4 hospitals. None. We analyzed all cardiac ICU encounters that included any respiratory support from October 2013 to December 2015. Noninvasive ventilation therapy included high flow nasal cannula and positive airway pressure support. We compared patient and, when relevant, perioperative characteristics of those receiving noninvasive ventilation to all others. Subgroup analysis was performed on neonates and infants undergoing major cardiovascular surgery. To examine duration of respiratory support, we created a casemix-adjustment model and calculated adjusted mean durations of total respiratory support (mechanical ventilation + noninvasive ventilation), mechanical ventilation, and noninvasive ventilation. We compared adjusted duration of support across hospitals. The cohort included 8,940 encounters from 15 hospitals: 3,950 (44%) received noninvasive ventilation and 72% were neonates and infants. Medical encounters were more likely to include noninvasive ventilation than surgical. In surgical neonates and infants, 2,032 (55%) received postoperative noninvasive ventilation. Neonates, extracardiac anomalies, single ventricle, procedure complexity, preoperative respiratory support, mechanical ventilation duration, and postoperative disease severity were associated with noninvasive ventilation therapy (p < 0.001 for all). Across hospitals, noninvasive ventilation use ranged from 32% to 65%, and adjusted mean noninvasive ventilation duration ranged from 1 to 4 days (3-d observed mean). Duration of total adjusted respiratory support was more strongly correlated with duration of mechanical ventilation compared with noninvasive ventilation (Pearson r = 0.93 vs 0.71, respectively). Noninvasive ventilation use is common in cardiac ICUs, especially in patients admitted for medical conditions, infants, and those undergoing high complexity surgery. We observed wide variation in noninvasive ventilation use across hospitals, though the primary driver of total respiratory support time seems to be duration of mechanical ventilation.

  20. Status asthmaticus in children: a one-year study.

    PubMed Central

    McKenzie, S A; Edmunds, A T; Godfrey, S

    1979-01-01

    50 children were admitted on 72 occasions during one year in status asthmaticus, defined as wheezing not relieved by two doses of bronchodilator 4 hours apart. At least one-third of children were hypercapnoeic on admission. They were managed with either oral prednisolone and nebulised salbutamol. Those with peak expiratory flow rates of greater than 25% expected for height were satisfactorily managed on the oral regimen. One child needed assisted ventilation. 73% were fit for discharge within 4 days and more than half of them needed a change in maintanance treatment. PMID:389173

  1. 30 CFR 57.22206 - Main ventilation failure (I-A, II-A, III, and V-A mines).

    Code of Federal Regulations, 2010 CFR

    2010-07-01

    ... AND NONMETAL MINES Safety Standards for Methane in Metal and Nonmetal Mines Ventilation § 57.22206..., tests for methane shall be conducted in affected active workings until normal air flow has resumed. (b... less than 1.0 percent methane. Persons other than examiners shall not reenter a Subcategory II-A mine...

  2. The effect of fresh gas flow rate and type of anesthesia machine on time to reach target sevoflurane concentration.

    PubMed

    Shin, Hye Won; Yu, Hae Na; Bae, Go Eun; Huh, Hyub; Park, Ji Yong; Kim, Ji Young

    2017-01-19

    Anesthesia machines have been developed by the application of new technology for rapid and easier control of anesthetic concentration. In this study, we used a test lung to investigate whether the time taken to reach the target sevoflurane concentration varies with the rate of fresh gas flow (FGF) and type of anesthesia machine (AM). We measured the times taken to reach the target sevoflurane concentration (2 minimum alveolar concentration = 4%) at variable rates of FGF (0.5, 1, or 3 L/min) and different types of AM (Primus ® , Perseus ® , and Zeus ® [Zeus ® -F; Zeus ® fresh gas mode, Zeus ® -A; Zeus ® auto-mode]). Concomitant ventilation was supplied using 100% O 2. The AMs were connected to a test lung. A sevoflurane vaporizer setting of 6% was used in Primus ® , Perseus ® , and Zeus ® -F; a target end-tidal setting of 4% was used in Zeus ® -A (from a vaporizer setting of 0%). The time taken to reach the target concentration was measured in every group. When the same AM was used (Primus ® , Perseus ® , or Zeus ® -F), the times to target concentration shortened as the FGF rate increased (P < 0.05). Conversely, when the same FGF rate was used, but with different AMs, the time to target concentration was shortest in Perseus ® , followed by Primus ® , and finally by Zeus ® -F (P < 0.05). With regards to both modes of Zeus ® , at FGF rates of 0.5 and 1 L/min, the time to target concentration was shorter in Zeus ® -A than in Zeus ® -F; however, the time was longer in Zeus ® -A than in Zeus ® -F at FGF rate of 3 L/min (P < 0.05). Shorter times taken to reach the target concentration were associated with high FGF rates, smaller internal volume of the AM, proximity of the fresh gas inlets to patients, absence of a decoupling system, and use of blower-driven ventilators in AM.

  3. Airway humidification during high-frequency percussive ventilation.

    PubMed

    Allan, Patrick F; Hollingsworth, Michael J; Maniere, Gordon C; Rakofsky, Anthony K; Chung, Kevin K; Naworol, Gregory A; Ward, John A; Perello, Michelle; Morris, Michael J

    2009-03-01

    We were concerned about the risk of inadequate humidification during high-frequency percussive ventilation (HFPV). We studied 5 humidifiers during HFPV with a lung model, at bias gas flows of 10 L/min, 30 L/min, and 50 L/min, and compared the results to those from a comparator ventilator/humidifier setup and to the minimum temperature (30 degrees C) and humidity (30 mg/L) [corrected] recommended by the American Association for Respiratory Care, at both regular room temperature and a high ambient temperature. Temperature was measured at the humidifier outflow point and at the artificial carina. Humidity was measured at the artificial carina. Of the 7 HFPV/humidifier combinations, 2 (the MR850 at a bias flow of 50 L/min, and the ConchaTherm Hi-Flow with VDR nebulizer) provided a carinal temperature equivalent to the comparator setup at room temperature, whereas one HFPV/humidifier combination (the ConchaTherm Hi-Flow with modified programming, at bias flows of 30 L/min and 50 L/min) provided a higher carinal temperature. At high ambient temperature, all of the setups delivered lower carinal temperature than the comparator setup. Only 2 setups (the ConchaTherm with modified programming at a bias flow of 50 L/min, and the ConchaTherm Hi-Flow with VDR nebulizer) provided carinal humidification equivalent to the comparator setup, without regard to ambient temperature; the other humidifiers were less effective. The ConchaTherm with modified programming, and the ConchaTherm with the VDR nebulizer provided the most consistent humidification. HFPV's distinctive gas-flow mechanism may impair gas heating and humidification, so all humidification systems should be tested with HFPV prior to clinical use.

  4. Oxygen consumption of a pneumatically controlled ventilator in a field anesthesia machine.

    PubMed

    Szpisjak, Dale F; Javernick, Elizabeth N; Kyle, Richard R; Austin, Paul N

    2008-12-01

    Field anesthesia machines (FAM) have been developed for remote locations where reliable supplies of compressed medical gases or electricity may be absent. In place of electricity, pneumatically controlled ventilators use compressed gas to power timing circuitry and actuate valves. We sought to determine the total O(2) consumption and ventilator gas consumption (drive gas [DG] plus pneumatic control [PC] gas) of a FAM's pneumatically controlled ventilator in mechanical models of high (HC) and low (LC) total thoracic compliance. The amount of total O(2) consumed by the Magellan-2200 (Oceanic Medical Products, Atchison, KS) FAM with pneumatically controlled ventilator was calculated using the ideal gas law and the measured mass of O(2) consumed from E cylinders. DG to the bellows canister assembly was measured with the Wright Respirometer Mk 8 (Ferraris Respiratory Europe, Hertford, UK). PC gas consumption was calculated by subtracting DG and fresh gas flow (FGF) from the total O(2) consumed from the E cylinder. The delivered tidal volume (V(T)) was measured with a pneumotach (Hans Rudolph, KS City, MO). Three different V(T) were tested (500, 750, and 1000 mL) with two lung models (HC and LC) using the Vent Aid Training Test Lung (MI Instruments, Grand Rapids, MI). Respiratory variables included an I:E of 1:2, FGF of 1 L/min, and respiratory rate of 10 breaths/min. Total O(2) consumption was directly proportional to V(T) and inversely proportional to compliance. The smallest total O(2) consumption rate (including FGF) was 9.3 +/- 0.4 L/min in the HC-500 model and the largest was 15.9 +/- 0.5 L/min in the LC-1000 model (P < 0.001). The mean PC circuitry consumption was 3.9 +/- 0.24 L/min or 390 mL +/- 24 mL/breath. To prepare for loss of central DG supply, patient safety will be improved by estimating cylinder duration for low total thoracic compliance. Using data from the smaller compliance and greatest V(T) model (LC-1000), a full O(2) E cylinder would be depleted in <42 min, whereas a full H cylinder would last approximately 433 min.

  5. Control aspects of the human cardiovascular-respiratory system under a nonconstant workload.

    PubMed

    Calderon, Pio Gabrielle B; Habib, Mustafa; Kappel, Franz; de Los Reyes, Aurelio A

    2017-07-01

    The human cardiovascular system (CVS) and respiratory system (RS) work together in order to supply oxygen (O 2 ) and other substrates needed for metabolism and to remove carbon dioxide (CO 2 ). Global and local control mechanisms act on the CVS in order to adjust blood flow to the different parts of the body. This, in turn, affects the RS since the amount of O 2 and CO 2 transported, respectively to and away from the tissues depends on the cardiac output and blood flow in both the systemic and pulmonary circuits of the CVS. Local metabolic control is influenced by local concentrations of blood gases affecting systemic resistance, resulting to vasoconstriction/vasodilation. Thus, the exchange of blood gases demands a tight coordination between blood flow and ventilation of the lungs. In this work, a model of the cardiovascular-respiratory system (CVRS) is considered to obtain an optimal control for time-dependent ergometric workloads by using the Euler-Lagrange formulation of the optimal control problem. The essential controls in the CVRS model are variations in the heart rate and alveolar ventilation through which the central nervous system restricts the arterial partial pressure of CO 2 ( [Formula: see text] ) close to 40  mmHg. Further, penalization terms in the cost functional are included to match the metabolic need for O 2 and the metabolic production of CO 2 with O 2 - and CO 2 -transport by blood. Copyright © 2017. Published by Elsevier Inc.

  6. Noninvasive versus conventional ventilation to treat hypercapnic encephalopathy in chronic obstructive pulmonary disease.

    PubMed

    Scala, Raffaele; Nava, Stefano; Conti, Giorgio; Antonelli, Massimo; Naldi, Mario; Archinucci, Ivano; Coniglio, Giovanni; Hill, Nicholas S

    2007-12-01

    We recently reported a high success rate using noninvasive positive pressure ventilation (NPPV) to treat COPD exacerbations with hypercapnic encephalopathy. This study compared the hospital outcomes of NPPV vs. conventional mechanical ventilation (CMV) in COPD exacerbations with moderate to severe hypercapnic encephalopathy, defined by a Kelly score of 3 or higher. A 3-year prospective matched case-control study in a respiratory semi-intensive care unit (RSICU) and intensive care unit (ICU). From 103 consecutive patients the study included 20 undergoing NPPV and 20 CMV, matched for age, simplified acute physiology score II, and baseline arterial blood gases. ABG significantly improved in both groups after 2 h. The rate of complications was lower in the NPPV group than in the CMV group due to fewer cases of nosocomial pneumonia and sepsis. In-hospital mortality, 1-year mortality, and tracheostomy rates were similar in the two groups. Fewer patients remained on ventilation after 30 days in NPPV group. The NPPV group showed a shorter duration of ventilation. In COPD exacerbations with moderate to severe hypercapnic encephalopathy, the use of NPPV performed by an experienced team compared to CMV leads to similar short and long-term survivals with a reduced nosocomial infection rate and duration of ventilation.

  7. Ventilation rates in recently constructed U.S. school classrooms.

    PubMed

    Batterman, S; Su, F-C; Wald, A; Watkins, F; Godwin, C; Thun, G

    2017-09-01

    Low ventilation rates (VRs) in schools have been associated with absenteeism, poorer academic performance, and teacher dissatisfaction. We measured VRs in 37 recently constructed or renovated and mechanically ventilated U.S. schools, including LEED and EnergyStar-certified buildings, using CO 2 and the steady-state, build-up, decay, and transient mass balance methods. The transient mass balance method better matched conditions (specifically, changes in occupancy) and minimized biases seen in the other methods. During the school day, air change rates (ACRs) averaged 2.0±1.3 hour -1 , and only 22% of classrooms met recommended minimum ventilation rates. HVAC systems were shut off at the school day close, and ACRs dropped to 0.21±0.19 hour -1 . VRs did not differ by building type, although cost-cutting and comfort measures resulted in low VRs and potentially impaired IAQ. VRs were lower in schools that used unit ventilators or radiant heating, in smaller schools and in larger classrooms. The steady-state, build-up, and decay methods had significant limitations and biases, showing the need to confirm that these methods are appropriate. Findings highlight the need to increase VRs and to ensure that energy saving and comfort measures do not compromise ventilation and IAQ. © 2017 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

  8. Carbon Dioxide Detection and Indoor Air Quality Control.

    PubMed

    Bonino, Steve

    2016-04-01

    When building ventilation is reduced, energy is saved because it is not necessary to heat or cool as much outside air. Reduced ventilation can result in higher levels of carbon dioxide, which may cause building occupants to experience symptoms. Heating or cooling for ventilation air can be enhanced by a DCV system, which can save energy while providing a comfortable environment. Carbon dioxide concentrations within a building are often used to indicate whether adequate fresh air is being supplied to the building. These DCV systems use carbon dioxide sensors in each space or in the return air and adjust the ventilation based on carbon dioxide concentration; the higher the concentration, the more people occupy the space relative to the ventilation rate. With a carbon dioxide sensor DCV system, the fresh air ventilation rate varies based on the number ofpeople in the space, saving energy while maintaining a safe and comfortable environment.

  9. Effects of types of ventilation system on indoor particle concentrations in residential buildings.

    PubMed

    Park, J S; Jee, N-Y; Jeong, J-W

    2014-12-01

    The objective of this study was to quantify the influence of ventilation systems on indoor particle concentrations in residential buildings. Fifteen occupied, single-family apartments were selected from three sites. The three sites have three different ventilation systems: unbalanced mechanical ventilation, balanced mechanical ventilation, and natural ventilation. Field measurements were conducted between April and June 2012, when outdoor air temperatures were comfortable. Number concentrations of particles, PM2.5 and CO2 , were continuously measured both outdoors and indoors. In the apartments with natural ventilation, I/O ratios of particle number concentrations ranged from 0.56 to 0.72 for submicron particles, and from 0.25 to 0.60 for particles larger than 1.0 μm. The daily average indoor particle concentration decreased to 50% below the outdoor level for submicron particles and 25% below the outdoor level for fine particles, when the apartments were mechanically ventilated. The two mechanical ventilation systems reduced the I/O ratios by 26% for submicron particles and 65% for fine particles compared with the natural ventilation. These results showed that mechanical ventilation can reduce exposure to outdoor particles in residential buildings. Results of this study confirm that mechanical ventilation with filtration can significantly reduce indoor particle levels compared with natural ventilation. The I/O ratios of particles substantially varied at the naturally ventilated apartments because of the influence of variable window opening conditions and unsteadiness of wind flow on the penetration of outdoor air particles. For better prediction of the exposure to outdoor particles in naturally ventilated residential buildings, it is important to understand the penetration of outdoor particles with variable window opening conditions. © 2014 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

  10. The effect of ventilation on indoor exposure to semivolatile organic compounds.

    PubMed

    Liu, C; Zhang, Y; Benning, J L; Little, J C

    2015-06-01

    A mechanistic model was developed to examine how natural ventilation influences residential indoor exposure to semivolatile organic compounds (SVOCs) via inhalation, dermal sorption, and dust ingestion. The effect of ventilation on indoor particle mass concentration and mass transfer at source/sink surfaces, and the enhancing effect of particles on mass transfer at source/sink surfaces are included. When air exchange rate increases from 0.6/h to 1.8/h, the steady-state SVOC (gas-phase plus particle phase with log KOA varying from 9 to 13) concentration in the idealized model decreases by about 60%. In contrast, for the same change in ventilation, the simulated indoor formaldehyde (representing volatile organic compounds) gas-phase concentration decreases by about 70%. The effect of ventilation on exposure via each pathway has a relatively insignificant association with the KOA of the SVOCs: a change of KOA from 10(9) to 10(13) results in a change of only 2-30%. Sensitivity analysis identifies the deposition rate of PM2.5 as a primary factor influencing the relationship between ventilation and exposure for SVOCs with log KOA = 13. The relationship between ventilation rate and air speed near surfaces needs to be further substantiated. © 2014 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

  11. Determination of eustachius tube ventilation functioning among benign type chronic suppurative otitis media and non-otitis media subjects using sonotubometry

    NASA Astrophysics Data System (ADS)

    Ikhwan, M.; Hafil, A. F.; Bramanthyo, B.

    2017-08-01

    The Eustachian tube (ET) is responsible for the ventilation, protection, and cleaning of the middle ear. ET dysfunction plays an important role in the pathogenesis of otitis media cases, and thus the treatment and prognosis of these cases is extremely dependent on adequate ET function, which can ultimately affect the success rate of middle ear reconstruction practices. Data research on the ET’s ventilation function is needed to ensure the success of therapy and surgery treatments in the case of Chronic Suppurative Otitis Media (CSOM) patients. This study aims to investigate ET ventilation functioning in benign type CSOM and non-otitis media subjects and to develop another modality to measure ET ventilation functioning in patients with intact and perforated tympanic membranes. A comparative cross-sectional study of 36 benign type CSOM patients and 80 non-otitis media subjects will be conducted using sonotubometry and the rated parameter measurements of ET opening frequency, amplitude and ET opening duration. Malfunctioning ventilation of the ET is more common among benign type CSOM subjects (47%) than among non-otitis media subjects (18.75%). There is a significant difference (p = 0.002) between the ET ventilation functioning of benign type CSOM subjects and non-otitis media subjects—benign type CSOM subjects have rates of malfunctioning ET ventilation that are 3.88 times higher than those of non-otitis media subjects. Patients with benign type CSOM are more likely to experience malfunctioning ET ventilation than are non-otitis media subjects.

  12. Evidence of inadequate ventilation in portable classrooms: results of a pilot study in Los Angeles County.

    PubMed

    Shendell, D G; Winer, A M; Weker, R; Colome, S D

    2004-06-01

    The prevalence of prefabricated, portable classrooms (portables) for United States public schools has increased; in California, approximately one of three students learn inside portables. Limited research has been conducted on indoor air and environmental quality in American schools, and almost none in portables. Available reports and conference proceedings suggest problems from insufficient ventilation due to poor design, operation, and/or maintenance of heating, ventilation and air conditioning (HVAC) systems; most portables have one mechanical, wall-mounted HVAC system. A pilot assessment was conducted in Los Angeles County, including measurements of integrated ventilation rates based on a perfluorocarbon tracer gas technique and continuous monitoring of temperature (T) and relative humidity (RH). Measured ventilation rates were low [mean school day integrated average 0.8 per hour (range: 0.1-2.9 per hour)]. Compared with relevant standards, results suggested adequate ventilation and associated conditioning of indoor air for occupant comfort were not always provided to these classrooms. Future school studies should include integrated and continuous measurements of T, RH, and ventilation with appropriate tracer gas methods, and other airflow measures. Adequate ventilation has the potential to mitigate concentrations of chemical pollutants, particles, carbon dioxide, and odors in portable and traditional classrooms, which should lead to a reduction in reported health outcomes, e.g., symptoms of 'sick building syndrome', allergies, asthma. Investigations of school indoor air and environmental quality should include continuous temperature and relative humidity data with inexpensive instrumentation as indicators of thermal comfort, and techniques to measure ventilation rates.

  13. The effect of perfluorocarbon vapour on the measurement of respiratory tidal volume during partial liquid ventilation.

    PubMed

    Davies, M W; Dunster, K R

    2000-08-01

    During partial liquid ventilation perfluorocarbon vapour is present in the exhaled gases. The volumes of these gases are measured by pneumotachometers. Error in measuring tidal volumes will give erroneous measurement of lung compliance during partial liquid ventilation. We aim to compare measured tidal volumes with and without perfluorocarbon vapour using tidal volumes suitable for use in neonates. Tidal volumes were produced with a 100 ml calibration syringe from 20 to 100 ml and with a calibrated Harvard rodent ventilator from 2.5 to 20 ml. Control tidal volumes were drawn from a humidifier chamber containing water vapour and the PFC tidal volumes were drawn from a humidifier chamber containing water and perfluorocarbon (FC-77) vapour. Tidal volumes were measured by a fixed orifice, target, differential pressure flowmeter (VenTrak) or a hot-wire anenometer (Bear Cub) placed between the calibration syringe or ventilator and the humidifier chamber. All tidal volumes measured with perfluorocarbon vapour were increased compared with control (ANOVA p < 0.001 and post t-test p < 0.0001). Measured tidal volume increased from 7 to 16% with the fixed orifice type flow-meter, and from 35 to 41% with the hot-wire type. In conclusion, perfluorocarbon vapour flowing through pneumotachometers gives falsely high tidal volume measurements. Calculation of lung compliance must take into account the effect of perfluorocarbon vapour on the measurement of tidal volume.

  14. Effect of the laryngeal tube on the no-flow-time in a simulated two rescuer basic life support setting with inexperienced users.

    PubMed

    Schröder, J; Bucher, M; Meyer, O

    2016-09-01

    Intubation with a laryngeal tube (LT) is a recommended alternative to endotracheal intubation during advanced life support (ALS). LT insertion is easy; therefore, it may also be an alternative to bag-mask ventilation (BMV) for untrained personnel performing basic life support (BLS). Data from manikin studies support the influence of LT on no-flow-time (NFT) during ALS. We performed a prospective, randomized manikin study using a two-rescuer model to compare the effects of ventilation using a LT and BMV on NFT during BLS. Participants were trained in BMV and were inexperienced in the use of a LT. There was no significant difference in total NFT with the use of a LT and BMV (LT: mean 83.1 ± 37.3 s; BMV: mean 78.7 ± 24.5 s; p = 0.313), but we found significant differences in the progression of the scenario: in the BLS-scenario, the proportion of time spent performing chest compressions was higher when BMV was used compared to when a LT was used. The quality of chest compressions and the ventilation rate did not differ significantly between the two groups. The mean tidal volume and mean minute volume were significantly larger with the use of a LT compared with the use of BMV. In conclusion, in a two-rescuer BLS scenario, NFT is longer with the use of a LT (without prior training) than with the use of BMV (with prior training). The probable reasons for this result are higher tidal volumes with the use of a LT leading to longer interruptions without chest compressions.

  15. Simulation of late inspiratory rise in airway pressure during pressure support ventilation.

    PubMed

    Yu, Chun-Hsiang; Su, Po-Lan; Lin, Wei-Chieh; Lin, Sheng-Hsiang; Chen, Chang-Wen

    2015-02-01

    Late inspiratory rise in airway pressure (LIRAP, Paw/ΔT) caused by inspiratory muscle relaxation or expiratory muscle contraction is frequently seen during pressure support ventilation (PSV), although the modulating factors are unknown. We investigated the effects of respiratory mechanics (normal, obstructive, restrictive, or mixed), inspiratory effort (-2, -8, or -15 cm H2O), flow cycle criteria (5-40% peak inspiratory flow), and duration of inspiratory muscle relaxation (0.18-0.3 s) on LIRAP during PSV using a lung simulator and 4 types of ventilators. LIRAP occurred with all lung models when inspiratory effort was medium to high and duration of inspiratory muscle relaxation was short. The normal lung model was associated with the fastest LIRAP, whereas the obstructive lung model was associated with the slowest. Unless lung mechanics were normal or mixed, LIRAP was unlikely to occur when inspiratory effort was low. Different ventilators were also associated with differences in LIRAP speed. Except for within the restrictive lung model, changes in flow cycle level did not abolish LIRAP if inspiratory effort was medium to high. Increased duration of inspiratory relaxation also led to the elimination of LIRAP. Simulation of expiratory muscle contraction revealed that LIRAP occurred only when expiratory muscle contraction occurred sometime after the beginning of inspiration. Our simulation study reveals that both respiratory resistance and compliance may affect LIRAP. Except for under restrictive lung conditions, LIRAP is unlikely to be abolished by simply lowering flow cycle criteria when inspiratory effort is strong and relaxation time is rapid. LIRAP may be caused by expiratory muscle contraction when it occurs during inspiration. Copyright © 2015 by Daedalus Enterprises.

  16. DIRECT MEASUREMENT TECHNIQUE FOR THE DETERMINING VENTILATION RATE IN THE DEPOSIT FEEDING CLAM, MACOMA NASUTA (BIVALVIA, TELLINACEAE)

    EPA Science Inventory

    An exposure chamber, the "clambox", was developed to measure ventilation rate, sediment processing rate, and efficiency of pollutant uptake byMacoma nasuta Conrad, a surface surface deposit-feeding clams. Clams, collected from Yaquina, Bay, Oregon, USA, were cemented into a hole ...

  17. Challenges on non-invasive ventilation to treat acute respiratory failure in the elderly.

    PubMed

    Scala, Raffaele

    2016-11-15

    Acute respiratory failure is a frequent complication in elderly patients especially if suffering from chronic cardio-pulmonary diseases. Non-invasive mechanical ventilation constitutes a successful therapeutic tool in the elderly as, like in younger patients, it is able to prevent endotracheal intubation in a wide range of acute conditions; moreover, this ventilator technique is largely applied in the elderly in whom invasive mechanical ventilation is considered not appropriated. Furthermore, the integration of new technological devices, ethical issues and environment of treatment are still largely debated in the treatment of acute respiratory failure in the elderly.This review aims at reporting and critically analyzing the peculiarities in the management of acute respiratory failure in elderly people, the role of noninvasive mechanical ventilation, the potential advantages of applying alternative or integrated therapeutic tools (i.e. high-flow nasal cannula oxygen therapy, non-invasive and invasive cough assist devices and low-flow carbon-dioxide extracorporeal systems), drawbacks in physician's communication and "end of life" decisions. As several areas of this topic are not supported by evidence-based data, this report takes in account also "real-life" data as well as author's experience.The choice of the setting and of the timing of non-invasive mechanical ventilation in elderly people with advanced cardiopulmonary disease should be carefully evaluated together with the chance of using integrated or alternative supportive devices. Last but not least, economic and ethical issues may often challenges the behavior of the physicians towards elderly people who are hospitalized for acute respiratory failure at the end stage of their cardiopulmonary and neoplastic diseases.

  18. Noninvasive Ventilation During Immediate Postoperative Period in Cardiac Surgery Patients: Systematic Review and Meta-Analysis

    PubMed Central

    Pieczkoski, Suzimara Monteiro; Margarites, Ane Glauce Freitas; Sbruzzi, Graciele

    2017-01-01

    Objective To verify the effectiveness of noninvasive ventilation compared to conventional physiotherapy or oxygen therapy in the mortality rate and prevention of pulmonary complications in patients during the immediate postoperative period of cardiac surgery. Methods Systematic review and meta-analysis recorded in the International Prospective Register of Ongoing Systematic Reviews (number CRD42016036441). The research included the following databases: MEDLINE, Cochrane Central, PEDro, LILACS and manual search of the references of studies published until March 2016. The review included randomized controlled trials with patients during the immediate postoperative period of cardiac surgery, which compared the use of noninvasive ventilation, BiLevel modes, continuous positive airway pressure, intermittent positive pressure breathing and positive pressure ventilation with conventional physiotherapy or oxygen therapy, and assessed the mortality rate, occurrence of pulmonary complications (atelectasis, pneumonia, acute respiratory failure, hypoxemia), reintubation rate, ventilation time, time spent in the intensive care unit (ICU), length of hospital stay and partial pressure of oxygen. Results Among the 479 selected articles, ten were included in the systematic review (n=1050 patients) and six in the meta-analysis. The use of noninvasive ventilation did not significantly reduce the risk for atelectasis (RR: 0.60; CI95% 0.28-1.28); pneumonia (RR: 0.20; CI95% 0.04-1.16), reintubation rate (RR: 0.51; CI95%: 0.15-1.66), and time spent in the ICU (-0.04 days; CI95%: -0.13; 0.05). Conclusion Prophylactic noninvasive ventilation did not significantly reduce the occurrence of pulmonary complications such as atelectasis, pneumonia, reintubation rate and time spent in the ICU. The use is still unproven and new randomized controlled trials should be carried out. PMID:28977203

  19. Hood entry coefficients of compound exhaust hoods.

    PubMed

    Figueroa, Crescente E

    2011-12-01

    A traditional method for assessing the flow rate in ventilation systems is based on multiple readings of velocity or velocity pressure (VP) (usually 10 or 20 points) taken in ductwork sections located away from fittings (> seven × diameters of straight duct). This study seeks to eliminate the need for a multiple-point evaluation and replace it with a simplified method that requires only a single measurement of hood static pressure (SP(h)) taken at a more accessible location (< three × diameters of straight duct from the hood entry). The SP(h) method is widely used for the assessment of flow rate in simple hoods. However, industrial applications quite often use compound hoods that are regularly of the slot/plenum type. For these hoods, a "compound coefficient of entry" has not been published, which makes the use of the hood static pressure method unfeasible. This study proposes a model for the computation of a "compound coefficient of entry" and validates the use of this model to assess flow rate in two systems of well-defined geometry (multi-slotted/plenum and single-slotted/tapered or "fish-tail" types). When using a conservative value of the slot loss factor (1.78), the proposed model yielded an estimate of the volumetric flow rate within 10% of that provided by a more comprehensive method of assessment. The simplicity of the hood static pressure method makes it very desirable, even in the upper range of experimental error found in this study.

  20. Estimated flows of gases and carbon within CEEF ecosystem composed of human, crops and goats

    NASA Astrophysics Data System (ADS)

    Tako, Y.; Komatsubara, O.; Honda, G.; Arai, R.; Nitta, K.

    The Closed Ecology Experiment Facilities (CEEF) can be used as a test bed for Controlled Ecological Life Support Systems (CELSS), because technologies developed for the CEEF system facilitate self-sufficient material circulation necessary for long term missions such as Lunar and Mars exploration. In the experiment conducted under closed condition in FY2003, rice and soybeans were cultivated sequentially in two chambers and a chamber, each having a cultivation area of 30 m2 and floor area of 43 m2, inside the Plantation Module with artificial lighting of the CEEF. In the chamber having a cultivation area of 60 m2 and floor area of 65 m2, inside the Plantation Module with natural and artificial lighting, peanuts and safflowers were also cultivated. Stable transplant (or seeding) and harvest of each crop were maintained during a month. Flows of CO2, O2 and carbon to and from the crops were analyzed during the stable cultivation period. Simulated works and stay in the CEEF lasting five days were conducted two times under ventilating condition in FY2003. Gas exchange of human was estimated using heart rate data collected during the experiments and correlation between gas exchange rate and heart rate. Gas exchange rate and carbon balance of female goats were determined using an open-flow measurement system with a gastight chamber. From these results, flows of gases and carbon in the CEEF were discussed.

  1. Noise exposure is increased with neonatal helmet CPAP in comparison with conventional nasal CPAP.

    PubMed

    Trevisanuto, D; Camiletti, L; Doglioni, N; Cavallin, F; Udilano, A; Zanardo, V

    2011-01-01

    in adults, noninvasive ventilation via a helmet is associated with significantly greater noise than nasal and facial masks. We hypothesized that noise exposure could be increased with neonatal helmet continuous positive airway pressure (CPAP) in comparison with conventional nasal CPAP (nCPAP). Our primary objective was to compare the noise intensity produced by a neonatal helmet CPAP and a conventional nCPAP system. Furthermore, we aimed to evaluate the effect of the gas flow rate and the presence of the humidifier and the filter on noise levels during neonatal helmet CPAP treatment. in this bench study, noise intensity was measured in the following settings: helmet CPAP, nCPAP, incubator and the neonatal intensive care unit. In helmet CPAP, noise measurements were performed at different gas flow rates (8, 10 and 12 l/min), while in nCPAP, the flow rate was 8 l/min. For both CPAP systems, the level of pressure was maintained constant at 5 cmH(2) O. during neonatal helmet CPAP, the median (interquartile range) noise levels were significantly higher than those during nCPAP: 70.0 dB (69.9-70.4) vs. 62.7 dB (62.5-63.0); P<0.001. In the helmet CPAP, the noise intensities changed with increasing flow rate and with the presence of a humidifier or a filter. noise intensities generated by the neonatal helmet CPAP were significantly higher than those registered while using a conventional nCPAP system. In the helmet, the noise intensity depends on the gas flow rate, and the presence of a humidifier and a filter in the system. 2010 The Acta Anaesthesiologica Scandinavica Foundation.

  2. Experimental and CFD evidence of multiple solutions in a naturally ventilated building.

    PubMed

    Heiselberg, P; Li, Y; Andersen, A; Bjerre, M; Chen, Z

    2004-02-01

    This paper considers the existence of multiple solutions to natural ventilation of a simple one-zone building, driven by combined thermal and opposing wind forces. The present analysis is an extension of an earlier analytical study of natural ventilation in a fully mixed building, and includes the effect of thermal stratification. Both computational and experimental investigations were carried out in parallel with an analytical investigation. When flow is dominated by thermal buoyancy, it was found experimentally that there is thermal stratification. When the flow is wind-dominated, the room is fully mixed. Results from all three methods have shown that the hysteresis phenomena exist. Under certain conditions, two different stable steady-state solutions are found to exist by all three methods for the same set of parameters. As shown by both the computational fluid dynamics (CFD) and experimental results, one of the solutions can shift to another when there is a sufficient perturbation. These results have probably provided the strongest evidence so far for the conclusion that multiple states exist in natural ventilation of simple buildings. Different initial conditions in the CFD simulations led to different solutions, suggesting that caution must be taken when adopting the commonly used 'zero initialization'.

  3. Gas-phase organics in environmental tobacco smoke. 1. Effects of smoking rate, ventilation, and furnishing level on emission factors.

    PubMed

    Singer, Brett C; Hodgson, Alfred T; Guevarra, Karla S; Hawley, Elisabeth L; Nazaroff, William W

    2002-03-01

    We measured the emissions of 26 gas-phase organic compounds in environmental tobacco smoke (ETS) using a model room that simulates realistic conditions in residences and offices. Exposure-relevant emission factors (EREFs), which include the effects of sorption and re-emission over a 24-h period, were calculated by mass balance from measured compound concentrations and chamber ventilation rates in a 50-m3 room constructed and furnished with typical materials. Experiments were conducted at three smoking rates (5, 10, and 20 cigarettes day(-1)), three ventilation rates (0.3, 0.6, and 2 h(-1)), and three furnishing levels (wallboard with aluminum flooring, wallboard with carpet, and full furnishings). Smoking rate did not affect EREFs, suggesting that sorption was linearly related to gas-phase concentration. Furnishing level and ventilation rate in the model room had little effect on EREFs of several ETS compounds including 1,3-butadiene, acrolein, acrylonitrile, benzene, toluene, and styrene. However, sorptive losses at low ventilation with full furnishings reduced EREFs for the ETS tracers nicotine and 3-ethenylpyridine by as much as 90 and 65% as compared to high ventilation, wallboard/aluminum experiments. Likewise, sorptive losses were 40-70% for phenol, cresols, naphthalene, and methylnaphthalenes. Sorption persisted for many compounds; for example, almost all of the sorbed nicotine and most of the sorbed cresol remained sorbed 3 days after smoking. EREFs can be used in models and with ETS tracer-based methods to refine and improve estimates of exposures to ETS constituents.

  4. Multiscale image-based modeling and simulation of gas flow and particle transport in the human lungs

    PubMed Central

    Tawhai, Merryn H; Hoffman, Eric A

    2013-01-01

    Improved understanding of structure and function relationships in the human lungs in individuals and sub-populations is fundamentally important to the future of pulmonary medicine. Image-based measures of the lungs can provide sensitive indicators of localized features, however to provide a better prediction of lung response to disease, treatment and environment, it is desirable to integrate quantifiable regional features from imaging with associated value-added high-level modeling. With this objective in mind, recent advances in computational fluid dynamics (CFD) of the bronchial airways - from a single bifurcation symmetric model to a multiscale image-based subject-specific lung model - will be reviewed. The interaction of CFD models with local parenchymal tissue expansion - assessed by image registration - allows new understanding of the interplay between environment, hot spots where inhaled aerosols could accumulate, and inflammation. To bridge ventilation function with image-derived central airway structure in CFD, an airway geometrical modeling method that spans from the model ‘entrance’ to the terminal bronchioles will be introduced. Finally, the effects of turbulent flows and CFD turbulence models on aerosol transport and deposition will be discussed. CFD simulation of airflow and particle transport in the human lung has been pursued by a number of research groups, whose interest has been in studying flow physics and airways resistance, improving drug delivery, or investigating which populations are most susceptible to inhaled pollutants. The three most important factors that need to be considered in airway CFD studies are lung structure, regional lung function, and flow characteristics. Their correct treatment is important because the transport of therapeutic or pollutant particles is dependent on the characteristics of the flow by which they are transported; and the airflow in the lungs is dependent on the geometry of the airways and how ventilation is distributed to the peripheral tissue. The human airway structure spans more than 20 generations, beginning with the extra-thoracic airways (oral or nasal cavity, and through the pharynx and larynx to the trachea), then the conducting airways, the respiratory airways, and to the alveoli. The airways in individuals and sub-populations (by gender, age, ethnicity, and normal vs. diseased states) may exhibit different dimensions, branching patterns and angles, and thickness and rigidity. At the local level, one would like to capture detailed flow characteristics, e.g. local velocity profiles, shear stress, and pressure, for prediction of particle transport in an airway (lung structure) model that is specific to the geometry of an individual, to understand how inter-subject variation in airway geometry (normal or pathological) influences the transport and deposition of particles. In a systems biology – or multiscale modeling – approach, these local flow characteristics can be further integrated with epithelial cell models for the study of mechanotransduction. At the global (organ) level, one would like to match regional ventilation (lung function) that is specific to the individual, thus ensuring that the flow that transports inhaled particles is appropriately distributed throughout the lung model. Computational models that do not account for realistic distribution of ventilation are not capable of predicting realistic particle distribution or targeted drug deposition. Furthermore, the flow in the human lung can be transitional or turbulent in the upper and proximal airways, and becomes laminar in the distal airways. The flows in the laminar, transitional and turbulent regimes have different temporal and spatial scales. Therefore, modeling airway structure and predicting gas flow and particle transport at both local and global levels require image-guided multiscale modeling strategies. In this article, we will review the aforementioned three key aspects of CFD studies of the human lungs: airway structure (conducting airways), lung function (regional ventilation and boundary conditions), and flow characteristics (modeling of turbulent flow and its effect on particle transport). For modeling airway structure, we will focus on the conducting airways, and review both symmetric vs. asymmetric airway models, idealized vs. CT-based airway models, and multiscale subject-specific airway models. Imposition of physiological subject-specific boundary conditions (BCs) in CFD is essential to match regional ventilation in individuals, which is also critical in studying preferential deposition of inhaled aerosols in sub-populations, e.g. normals vs. asthmatics that may exhibit different ventilation patterns. Subject-specific regional ventilation defines flow distributions and characteristics in airway segments and bifurcations, which subsequently determines the transport and deposition of aerosols in the entire lungs. Turbulence models are needed to capture the transient and turbulent nature of the gas flow in the human lungs. Thus, the advantages and disadvantages of different turbulence models as well as their effects on particle transport will be discussed. The ultimate goal of the development is to identify sensitive structural and functional variables in sub-populations of normal and diseased lungs for potential clinical applications. PMID:23843310

  5. Building America Case Study: Ventilation System Effectiveness and Tested Indoor Air Quality Impacts, Tyler, Texas

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    ?Ventilation system effectiveness testing was conducted at two unoccupied, single-family, detached lab homes at the University of Texas - Tyler. Five ventilation system tests were conducted with various whole-building ventilation systems. Multizone fan pressurization testing characterized building and zone enclosure leakage. PFT testing showed multizone air change rates and interzonal airflow filtration. Indoor air recirculation by a central air distribution system can help improve the exhaust ventilation system by way of air mixing and filtration. In contrast, the supply and balanced ventilation systems showed that there is a significant benefit to drawing outside air from a known outside location, andmore » filtering and distributing that air. Compared to the Exhaust systems, the CFIS and ERV systems showed better ventilation air distribution and lower concentrations of particulates, formaldehyde and other VOCs. System improvement percentages were estimated based on four System Factor Categories: Balance, Distribution, Outside Air Source, and Recirculation Filtration. Recommended System Factors could be applied to reduce ventilation fan airflow rates relative to ASHRAE Standard 62.2 to save energy and reduce moisture control risk in humid climates. HVAC energy savings were predicted to be 8-10%, or $50-$75/year. Cumulative particle counts for six particle sizes, and formaldehyde and other Top 20 VOC concentrations were measured in multiple zones. The testing showed that single-point exhaust ventilation was inferior as a whole-house ventilation strategy.« less

  6. Measurement of changes in respiratory mechanics during partial liquid ventilation using jet pulses.

    PubMed

    Schmalisch, Gerd; Schmidt, Mario; Proquitté, Hans; Foitzik, Bertram; Rüdiger, Mario; Wauer, Roland R

    2003-05-01

    To compare the changes in respiratory mechanics within the breathing cycle in healthy lungs between gas ventilation and partial liquid ventilation using a special forced-oscillation technique. Prospective animal trial. Animal laboratory in a university setting. A total of 12 newborn piglets (age, <12 hrs; mean weight, 725 g). After intubation and instrumentation, lung mechanics of the anesthetized piglets were measured by forced-oscillation technique at the end of inspiration and the end of expiration. The measurements were performed during gas ventilation and 80 mins after instillation of 30 mL/kg perfluorocarbon PF 5080. Brief flow pulses (width, 10 msec; peak flow, 16 L/min) were generated by a jet generator to measure the end-inspiratory and the end-expiratory respiratory input impedance in the frequency range of 4-32 Hz. The mechanical variables resistance, inertance, and compliance were determined by model fitting, using the method of least squares. At least in the lower frequency range, respiratory mechanics could be described adequately by an RIC single-compartment model in all piglets. During gas ventilation, the respiratory variables resistance and inertance did not differ significantly between end-inspiratory and end-expiratory measurements (mean [sd]: 4.2 [0.7] vs. 4.1 [0.6] kPa x L(-1) x sec, 30.0 [3.2] vs. 30.7 [3.1] Pa x L(-1) x sec2, respectively), whereas compliance decreased during inspiration from 14.8 (2.0) to 10.2 (2.4) mL x kPa(-1) x kg(-1) due to a slight lung overdistension. During partial liquid ventilation, the end-inspiratory respiratory mechanics was not different from the end-inspiratory respiratory mechanics measured during gas ventilation. However, in contrast to gas ventilation during partial liquid ventilation, compliance rose from 8.2 (1.0) to 13.0 (3.0) mL x kPa(-1) x kg(-1) during inspiration. During expiration, when perfluorocarbon came into the upper airways, both resistance and inertance increased considerably (mean with 95% confidence interval) by 34.3% (23.1%-45.8%) and 104.1% (96.0%-112.1%), respectively. The changes in the respiratory mechanics within the breathing cycle are considerably higher during partial liquid ventilation compared with gas ventilation. This dependence of lung mechanics from the pulmonary gas volume hampers the comparability of dynamic measurements during partial liquid ventilation, and the magnitude of these changes cannot be detected by conventional respiratory-mechanical analysis using time-averaged variables.

  7. Employing a Modified Diffuser Momentum Model to Simulate Ventilation of the Orion CEV

    NASA Technical Reports Server (NTRS)

    Straus, John; Lewis, John F.

    2011-01-01

    The Ansys CFX CFD modeling tool was used to support the design efforts of the ventilation system for the Orion CEV. CFD modeling was used to establish the flow field within the cabin for several supply configurations. A mesh and turbulence model sensitivity study was performed before the design studies. Results were post-processed for comparison with performance requirements. Most configurations employed straight vaned diffusers to direct and throw the flow. To manage the size of the models, the diffuser vanes were not resolved. Instead, a momentum model was employed to account for the effect of the diffusers. The momentum model was tested against a separate, vane-resolved side study. Results are presented for a single diffuser configuration for a low supply flow case.

  8. Implementation of clinical practice guidelines for ventilator-associated pneumonia: a multicenter prospective study.

    PubMed

    Sinuff, Tasnim; Muscedere, John; Cook, Deborah J; Dodek, Peter M; Anderson, William; Keenan, Sean P; Wood, Gordon; Tan, Richard; Haupt, Marilyn T; Miletin, Michael; Bouali, Redouane; Jiang, Xuran; Day, Andrew G; Overvelde, Janet; Heyland, Daren K

    2013-01-01

    Ventilator-associated pneumonia is an important cause of morbidity and mortality in critically ill patients. Evidence-based clinical practice guidelines for the prevention, diagnosis, and treatment of ventilator-associated pneumonia may improve outcomes, but optimal methods to ensure implementation of guidelines in the intensive care unit are unclear. Hence, we determined the effect of educational sessions augmented with reminders, and led by local opinion leaders, as strategies to implement evidence-based ventilator-associated pneumonia guidelines on guideline concordance and ventilator-associated pneumonia rates. Two-year prospective, multicenter, time-series study conducted between June 2007 and December 2009. Eleven ICUs (ten in Canada, one in the United States); five academic and six community ICUs. At each site, 30 adult patients mechanically ventilated >48 hrs were enrolled during four data collection periods (baseline, 6, 15, and 24 months). Guideline recommendations for the prevention, diagnosis, and treatment of ventilator-associated pneumonia were implemented using a multifaceted intervention (education, reminders, local opinion leaders, and implementation teams) directed toward the entire multidisciplinary ICU team. Clinician exposure to the intervention was assessed at 6, 15, and 24 months after the introduction of this intervention. The main outcome measure was aggregate concordance with the 14 ventilator-associated pneumonia guideline recommendations. One thousand three hundred twenty patients were enrolled (330 in each study period). Clinician exposure to the multifaceted intervention was high and increased during the study: 86.7%, 93.3%, 95.8%, (p < .001), as did aggregate concordance (mean [SD]): 50.7% (6.1), 54.4% (7.1), 56.2% (5.9), 58.7% (6.7) (p = .007). Over the study period, ventilator-associated pneumonia rates decreased (events/330 patients): 47 (14.2%), 34 (10.3%), 38 (11.5%), 29 (8.8%) (p = .03). A 2-yr multifaceted intervention to enhance ventilator-associated pneumonia guideline uptake was associated with a significant increase in guideline concordance and a reduction in ventilator-associated pneumonia rates.

  9. Non-invasive ventilation with intelligent volume-assured pressure support versus pressure-controlled ventilation: effects on the respiratory event rate and sleep quality in COPD with chronic hypercapnia.

    PubMed

    Nilius, Georg; Katamadze, Nato; Domanski, Ulrike; Schroeder, Maik; Franke, Karl-Josef

    2017-01-01

    COPD patients who develop chronic hypercapnic respiratory failure have a poor prognosis. Treatment of choice, especially the best form of ventilation, is not well known. This study compared the effects of pressure-controlled (spontaneous timed [ST]) non-invasive ventilation (NIV) and NIV with intelligent volume-assured pressure support (IVAPS) in chronic hypercapnic COPD patients regarding the effects on alveolar ventilation, adverse patient/ventilator interactions and sleep quality. This prospective, single-center, crossover study randomized patients to one night of NIV using ST then one night with the IVAPS function activated, or vice versa. Patients were monitored using polysomnography (PSG) and transcutaneous carbon dioxide pressure (PtcCO 2 ) measurement. Patients rated their subjective experience (total score, 0-45; lower scores indicate better acceptability). Fourteen patients were included (4 females, age 59.4±8.9 years). The total number of respiratory events was low, and similar under pressure-controlled (5.4±6.7) and IVAPS (8.3±10.2) conditions ( P =0.064). There were also no clinically relevant differences in PtcCO 2 between pressure-controlled and IVAPS NIV (52.9±6.2 versus 49.1±6.4 mmHg). Respiratory rate was lower under IVAPS overall; between-group differences reached statistical significance during wakefulness and non-rapid eye movement sleep. Ventilation pressures were 2.6 cmH 2 O higher under IVAPS versus pressure-controlled ventilation, resulting in a 20.1 mL increase in breathing volume. Sleep efficiency was slightly higher under pressure-controlled ventilation versus IVAPS. Respiratory arousals were uncommon (24.4/h [pressure-controlled] versus 25.4/h [IVAPS]). Overall patient assessment scores were similar, although there was a trend toward less discomfort during IVAPS. Our results show that IVAPS NIV allows application of higher nocturnal ventilation pressures versus ST without affecting sleep quality or inducing ventilation- associated events.

  10. VENTILATION RESEARCH: A REVIEW OF RECENT INDOOR AIR QUALITY LITERATURE

    EPA Science Inventory

    The report gives results of a literature review, conducted to survey and summarize recent and ongoing engineering research into building ventilation, air exchange rate, pollutant distribution and dispersion, and other effects of heating, ventilation, and air-conditioning (HVAC) s...

  11. Ventilation Rates and Airflow Pathways in Patient Rooms: A Case Study of Bioaerosol Containment and Removal.

    PubMed

    Mousavi, Ehsan S; Grosskopf, Kevin R

    2015-11-01

    Most studies on the transmission of infectious airborne disease have focused on patient room air changes per hour (ACH) and how ACH provides pathogen dilution and removal. The logical but mostly unproven premise is that greater air change rates reduce the concentration of infectious particles and thus, the probability of airborne disease transmission. Recently, a growing body of research suggests pathways between pathogenic source (patient) and control (exhaust) may be the dominant environmental factor. While increases in airborne disease transmission have been associated with ventilation rates below 2 ACH, comparatively less data are available to quantify the benefits of higher air change rates in clinical spaces. As a result, a series of tests were conducted in an actual hospital to observe the containment and removal of respirable aerosols (0.5-10 µm) with respect to ventilation rate and directional airflow in a general patient room, and, an airborne infectious isolation room. Higher ventilation rates were not found to be proportionately effective in reducing aerosol concentrations. Specifically, increasing mechanical ventilation from 2.5 to 5.5 ACH reduced aerosol concentrations only 30% on average. However, particle concentrations were more than 40% higher in pathways between the source and exhaust as was the suspension and migration of larger particles (3-10 µm) throughout the patient room(s). Computational analyses were used to validate the experimental results, and, to further quantify the effect of ventilation rate on exhaust and deposition removal in patient rooms as well as other particle transport phenomena. © The Author 2015. Published by Oxford University Press on behalf of the British Occupational Hygiene Society.

  12. Infiltration as Ventilation: Weather-Induced Dilution

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Sherman, Max H.; Turner, William J.N.; Walker, Iain S.

    The purpose of outdoor air ventilation is to dilute or remove indoor contaminants to which occupants are exposed. It can be provided by mechanical or natural means. In most homes, especially older homes, weather-driven infiltration provides the dominant fraction of the total ventilation. As we seek to provide good indoor air quality at minimum energy cost, it is important to neither over-ventilate nor under-ventilate. Thus, it becomes critically important to evaluate correctly the contribution infiltration makes to the total outdoor air ventilation rate. Because weather-driven infiltration is dependent on building air leakage and weather-induced pressure differences, a given amount ofmore » air leakage will provide different amounts of infiltration. Varying rates of infiltration will provide different levels of contaminant dilution and hence effective ventilation. This paper derives these interactions and then calculates the impact of weather-driven infiltration for different climates. A new “N-factor” is introduced to provide a convenient method for calculating the ventilation contribution of infiltration for over 1,000 locations across North America. The results of this work could be used in indoor air quality standards (specifically ASHRAE 62.2) to account for the contribution of weather-driven infiltration towards the dilution of indoor pollutants.« less

  13. Carbon dioxide narcosis due to inappropriate oxygen delivery: a case report.

    PubMed

    Herren, Thomas; Achermann, Eva; Hegi, Thomas; Reber, Adrian; Stäubli, Max

    2017-07-28

    Oxygen delivery to patients with chronic obstructive pulmonary disease may be challenging because of their potential hypoxic ventilatory drive. However, some oxygen delivery systems such as non-rebreathing face masks with an oxygen reservoir bag require high oxygen flow for adequate oxygenation and to avoid carbon dioxide rebreathing. A 72-year-old Caucasian man with severe chronic obstructive pulmonary disease was admitted to the emergency department because of worsening dyspnea and an oxygen saturation of 81% measured by pulse oximetry. Oxygen was administered using a non-rebreathing mask with an oxygen reservoir bag attached. For fear of removing the hypoxic stimulus to respiration the oxygen flow was inappropriately limited to 4L/minute. The patient developed carbon dioxide narcosis and had to be intubated and mechanically ventilated. Non-rebreathing masks with oxygen reservoir bags must be fed with an oxygen flow exceeding the patient's minute ventilation (>6-10 L/minute.). If not, the amount of oxygen delivered will be too small to effectively increase the arterial oxygen saturation. Moreover, the risk of carbon dioxide rebreathing dramatically increases if the flow of oxygen to a non-rebreathing mask is lower than the minute ventilation, especially in patients with chronic obstructive pulmonary disease and low tidal volumes. Non-rebreathing masks (with oxygen reservoir bags) must be used cautiously by experienced medical staff and with an appropriately high oxygen flow of 10-15 L/minute. Nevertheless, arterial blood gases must be analyzed regularly for early detection of a rise in partial pressure of carbon dioxide in arterial blood in patients with chronic obstructive pulmonary disease and a hypoxic ventilatory drive. These patients are more safely managed using a nasal cannula with an oxygen flow of 1-2L/minute or a simple face mask with an oxygen flow of 5L/minute.

  14. Mask ventilation with two different face masks in the delivery room for preterm infants: a randomized controlled trial.

    PubMed

    Cheung, D; Mian, Q; Cheung, P-Y; O'Reilly, M; Aziz, K; van Os, S; Pichler, G; Schmölzer, G M

    2015-07-01

    If an infant fails to initiate spontaneous breathing after birth, international guidelines recommend a positive pressure ventilation (PPV). However, PPV by face mask is frequently inadequate because of leak between the face and mask. Despite a variety of available face masks, none have been prospectively compared in a randomized fashion. We aimed to evaluate and compare leak between two commercially available round face masks (Fisher & Paykel (F&P) and Laerdal) in preterm infants <33 weeks gestational age in the delivery room. Infants born at the Royal Alexandra Hospital from April to September 2013 at <33 weeks gestational age who received mask PPV in the delivery room routinely had a flow sensor placed between the mask and T-piece resuscitator. Infants were randomly assigned to receive PPV with either a F&P or Laerdal face mask. All resuscitators were trained in the use of both face masks. We compared mask leak, airway pressures, tidal volume and ventilation rate between the two groups. Fifty-six preterm infants (n=28 in each group) were enrolled; mean±s.d. gestational age 28±3 weeks; birth weight 1210±448 g; and 30 (52%) were male. Apgar scores at 1 and 5 min were 5±3 and 7±2, respectively. Infants randomized to the F&P face mask and Laerdal face mask had similar mask leak (30 (25-38) versus 35 (24-46)%, median (interquartile range), respectively, P=0.40) and tidal volume (7.1 (4.9-8.9) versus 6.6 (5.2-8.9) ml kg(-1), P=0.69) during PPV. There were no significant differences in ventilation rate, inflation time or airway pressures between groups. The use of either face mask during PPV in the delivery room yields similar mask leak in preterm infants <33 weeks gestational age.

  15. Dissolution of biogenic ooze over basement edifices in the equatorial Pacific with implications for hydrothermal ventilation of the oceanic crust

    USGS Publications Warehouse

    Bekins, B.A.; Spivack, A.J.; Davis, E.E.; Mayer, L.A.

    2007-01-01

    Recent observations indicate that curious closed depressions in carbonate sediments overlying basement edifices are widespread in the equatorial Pacific. A possible mechanism for their creation is dissolution by fluids exiting basement vents from off-axis hydrothermal flow. Quantitative analysis based on the retrograde solubility of calcium carbonate and cooling of basement fluids during ascent provides an estimate for the dissolution capacity of the venting fluids. Comparison of the dissolution capacity and fluid flux with typical equatorial Pacific carbonate mass accumulation rates shows that this mechanism is feasible. By maintaining sediment-free basement outcrops, the process may promote widespread circulation of relatively unaltered seawater in the basement in an area where average sediment thicknesses are 300-500 m. The enhanced ventilation can explain several previously puzzling observations in this region, including anomalously low heat flux, relatively unaltered seawater in the basement, and aerobic and nitrate-reducing microbial activity at the base of the sediments. ?? 2007 The Geological Society of America.

  16. Effect of socioemotional stress on the quality of cardiopulmonary resuscitation during advanced life support in a randomized manikin study.

    PubMed

    Bjørshol, Conrad Arnfinn; Myklebust, Helge; Nilsen, Kjetil Lønne; Hoff, Thomas; Bjørkli, Cato; Illguth, Eirik; Søreide, Eldar; Sunde, Kjetil

    2011-02-01

    The aim of this study was to evaluate whether socioemotional stress affects the quality of cardiopulmonary resuscitation during advanced life support in a simulated manikin model. A randomized crossover trial with advanced life support performed in two different conditions, with and without exposure to socioemotional stress. The study was conducted at the Stavanger Acute Medicine Foundation for Education and Research simulation center, Stavanger, Norway. Paramedic teams, each consisting of two paramedics and one assistant, employed at Stavanger University Hospital, Stavanger, Norway. A total of 19 paramedic teams performed advanced life support twice in a randomized fashion, one control condition without socioemotional stress and one experimental condition with exposure to socioemotional stress. The socioemotional stress consisted of an upset friend of the simulated patient who was a physician, spoke a foreign language, was unfamiliar with current Norwegian resuscitation guidelines, supplied irrelevant clinical information, and repeatedly made doubts about the paramedics' resuscitation efforts. Aural distractions were supplied by television and cell telephone. The primary outcome was the quality of cardiopulmonary resuscitation: chest compression depth, chest compression rate, time without chest compressions (no-flow ratio), and ventilation rate after endotracheal intubation. As a secondary outcome, the socioemotional stress impact was evaluated through the paramedics' subjective workload, frustration, and feeling of realism. There were no significant differences in chest compression depth (39 vs. 38 mm, p = .214), compression rate (113 vs. 116 min⁻¹, p = .065), no-flow ratio (0.15 vs. 0.15, p = .618), or ventilation rate (8.2 vs. 7.7 min⁻¹, p = .120) between the two conditions. There was a significant increase in the subjective workload, frustration, and feeling of realism when the paramedics were exposed to socioemotional stress. In this advanced life support manikin study, the presence of socioemotional stress increased the subjective workload, frustration, and feeling of realism, without affecting the quality of cardiopulmonary resuscitation.

  17. CFD Simulations to Improve Ventilation in Low-Income Housing

    NASA Astrophysics Data System (ADS)

    Ho, Rosemond; Gorle, Catherine

    2017-11-01

    Quality of housing plays an important role in public health. In Dhaka, Bangladesh, the leading causes of death include tuberculosis, lower respiratory infections, and chronic obstructive pulmonary disease, so improving home ventilation could potentially mitigate these negative health effects. The goal of this project is to use computational fluid dynamics (CFD) to predict the relative effectiveness of different ventilation strategies for Dhaka homes. A Reynolds-averaged Navier-Stokes CFD model of a standard Dhaka home with apertures of different sizes and locations was developed to predict air exchange rates. Our initial focus is on simulating ventilation driven by buoyancy-alone conditions, which is often considered the limiting case in natural ventilation design. We explore the relationship between ventilation rate and aperture area to determine the most promising configurations for optimal ventilation solutions. Future research will include the modeling of wind-driven conditions, and extensive uncertainty quantification studies to investigate the effect of variability in the layout of homes and neighborhoods, and in local wind and temperature conditions. The ultimate objective is to formulate robust design recommendations that can reduce risks of respiratory illness in low-income housing.

  18. Nonstationary heat and mass transfer in the multilayer building construction with ventilation channels

    NASA Astrophysics Data System (ADS)

    Kharkov, N. S.

    2017-11-01

    Results of numerical modeling of the coupled nonstationary heat and mass transfer problem under conditions of a convective flow in facade system of a three-layer concrete panel for two different constructions (with ventilation channels and without) are presented. The positive effect of ventilation channels on the energy and humidity regime over a period of 12 months is shown. Used new method of replacement a solid zone (requiring specification of porosity and material structure, what complicates process of convergence of the solution) on quasi-solid in form of a multicomponent mixture (with restrictions on convection and mass fractions).

  19. The influence of opening windows and doors on the natural ventilation rate of a residential building

    EPA Science Inventory

    An analysis of air exchange rates due to intentional window and door openings in a research test house located in a residential environment is presented. These data inform the development of ventilation rate control strategies as building envelopes are tightened to improve the e...

  20. Define and Quantify the Physics of Air Flow, Pressure Drop and Aerosol Collection in Nuclear Grade HEPA Filters

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Moore, Murray E.

    Objective: Develop a set of peer-review and verified analytical methods to adjust HEPA filter performance to different flow rates, temperatures and altitudes. Experimental testing will measure HEPA filter flow rate, pressure drop and efficiency to verify the analytical approach. Nuclear facilities utilize HEPA (High Efficiency Particulate Air) filters to purify air flow for workspace ventilation. However, the ASME AG-1 technical standard (Code on Nuclear Air and Gas Treatment) does not adequately describe air flow measurement units for HEPA filter systems. Specifically, the AG-1 standard does not differentiate between volumetric air flow in ACFM (actual cubic feet per minute)compared to massmore » flow measured in SCFM (standard cubic feet per minute). More importantly, the AG-1 standard has an overall deficiency for using HEPA filter devices at different air flow rates, temperatures, and altitudes. Technical Approach: The collection efficiency and pressure drops of 18 different HEPA filters will be measured over a range of flow rates, temperatures and altitudes. The experimental results will be compared to analytical scoping calculations. Three manufacturers have allocated six HEPA filters each for this effort. The 18 filters will be tested at two different flow rates, two different temperatures and two different altitudes. The 36 total tests will be conducted at two different facilities: the ATI Test facilities (Baltimore MD) and the Los Alamos National Laboratory (Los Alamos NM). The Radiation Protection RP-SVS group at Los Alamos has an aerosol wind tunnel that was originally designed to evaluate small air samplers. In 2010, modifications were started to convert the wind tunnel for HEPA filter testing. (Extensive changes were necessary for the required aerosol generators, HEPA test fixtures, temperature control devices and measurement capabilities.) To this date, none of these modification activities have been funded through a specific DOE or NNSA program. This is expected to require six months of time, after receipt of funding. Benefits: US DOE facilities that use HEPA filters will benefit from access to the new operational measurement methods. Uncertainty and guesswork will be removed from HEPA filter operations.« less

  1. Effect of Postextubation High-Flow Nasal Cannula vs Noninvasive Ventilation on Reintubation and Postextubation Respiratory Failure in High-Risk Patients: A Randomized Clinical Trial.

    PubMed

    Hernández, Gonzalo; Vaquero, Concepción; Colinas, Laura; Cuena, Rafael; González, Paloma; Canabal, Alfonso; Sanchez, Susana; Rodriguez, Maria Luisa; Villasclaras, Ana; Fernández, Rafael

    2016-10-18

    High-flow conditioned oxygen therapy delivered through nasal cannulae and noninvasive mechanical ventilation (NIV) may reduce the need for reintubation. Among the advantages of high-flow oxygen therapy are comfort, availability, lower costs, and additional physiopathological mechanisms. To test if high-flow conditioned oxygen therapy is noninferior to NIV for preventing postextubation respiratory failure and reintubation in patients at high risk of reintubation. Multicenter randomized clinical trial in 3 intensive care units in Spain (September 2012-October 2014) including critically ill patients ready for planned extubation with at least 1 of the following high-risk factors for reintubation: older than 65 years; Acute Physiology and Chronic Health Evaluation II score higher than 12 points on extubation day; body mass index higher than 30; inadequate secretions management; difficult or prolonged weaning; more than 1 comorbidity; heart failure as primary indication for mechanical ventilation; moderate to severe chronic obstructive pulmonary disease; airway patency problems; or prolonged mechanical ventilation. Patients were randomized to undergo either high-flow conditioned oxygen therapy or NIV for 24 hours after extubation. Primary outcomes were reintubation and postextubation respiratory failure within 72 hours. Noninferiority margin was 10 percentage points. Secondary outcomes included respiratory infection, sepsis, and multiple organ failure, length of stay and mortality; adverse events; and time to reintubation. Of 604 patients (mean age, 65 [SD, 16] years; 388 [64%] men), 314 received NIV and 290 high-flow oxygen. Sixty-six patients (22.8%) in the high-flow group vs 60 (19.1%) in the NIV group were reintubation (absolute difference, -3.7%; 95% CI, -9.1% to ∞); 78 patients (26.9%) in the high-flow group vs 125 (39.8%) in the NIV group experienced postextubation respiratory failure (risk difference, 12.9%; 95% CI, 6.6% to ∞) [corrected]. Median time to reintubation did not significantly differ: 26.5 hours (IQR, 14-39 hours) in the high-flow group vs 21.5 hours (IQR, 10-47 hours) in the NIV group (absolute difference, -5 hours; 95% CI, -34 to 24 hours). Median postrandomization ICU length of stay was lower in the high-flow group, 3 days (IQR, 2-7) vs 4 days (IQR, 2-9; P=.048). Other secondary outcomes were similar in the 2 groups. Adverse effects requiring withdrawal of the therapy were observed in none of patients in the high-flow group vs 42.9% patients in the NIV group (P < .001). Among high-risk adults who have undergone extubation, high-flow conditioned oxygen therapy was not inferior to NIV for preventing reintubation and postextubation respiratory failure. High-flow conditioned oxygen therapy may offer advantages for these patients. clinicaltrials.gov Identifier: NCT01191489.

  2. Optimal Chest Compression Rate and Compression to Ventilation Ratio in Delivery Room Resuscitation: Evidence from Newborn Piglets and Neonatal Manikins

    PubMed Central

    Solevåg, Anne Lee; Schmölzer, Georg M.

    2017-01-01

    Cardiopulmonary resuscitation (CPR) duration until return of spontaneous circulation (ROSC) influences survival and neurologic outcomes after delivery room (DR) CPR. High quality chest compressions (CC) improve cerebral and myocardial perfusion. Improved myocardial perfusion increases the likelihood of a faster ROSC. Thus, optimizing CC quality may improve outcomes both by preserving cerebral blood flow during CPR and by reducing the recovery time. CC quality is determined by rate, CC to ventilation (C:V) ratio, and applied force, which are influenced by the CC provider. Thus, provider performance should be taken into account. Neonatal resuscitation guidelines recommend a 3:1 C:V ratio. CCs should be delivered at a rate of 90/min synchronized with ventilations at a rate of 30/min to achieve a total of 120 events/min. Despite a lack of scientific evidence supporting this, the investigation of alternative CC interventions in human neonates is ethically challenging. Also, the infrequent occurrence of extensive CPR measures in the DR make randomized controlled trials difficult to perform. Thus, many biomechanical aspects of CC have been investigated in animal and manikin models. Despite mathematical and physiological rationales that higher rates and uninterrupted CC improve CPR hemodynamics, studies indicate that provider fatigue is more pronounced when CC are performed continuously compared to when a pause is inserted after every third CC as currently recommended. A higher rate (e.g., 120/min) is also more fatiguing, which affects CC quality. In post-transitional piglets with asphyxia-induced cardiac arrest, there was no benefit of performing continuous CC at a rate of 90/min. Not only rate but duty cycle, i.e., the duration of CC/total cycle time, is a known determinant of CC effectiveness. However, duty cycle cannot be controlled with manual CC. Mechanical/automated CC in neonatal CPR has not been explored, and feedback systems are under-investigated in this population. Evidence indicates that providers perform CC at rates both higher and lower than recommended. Video recording of DR CRP has been increasingly applied and observational studies of what is actually done in relation to outcomes could be useful. Different CC rates and ratios should also be investigated under controlled experimental conditions in animals during perinatal transition. PMID:28168185

  3. Comparison of Ventilation and Cardiac Compressions When Utilizing the Impact Model 730 Automatic Transport Ventilator Versus a Conventional Bag Valve With a Facemask in a Model of Adult Cadiopulmonary Arrest

    DTIC Science & Technology

    2005-10-25

    Chest compression , Ventilation 2P251107 256Page 2 1. Introduction During the initial stages of cardiopulmonary resuscitation ( CPR ), ventilation is...a metronome to facilitate chest compression timing. Twenty-eight nurses alternated performing 4 minutes of CPR using the BVM or Impact 730 to deliver... Chest compression and ventilation rates during cardiopulmonary resuscitation : the effects of audible tone guidance. Academic Emergency Medicine

  4. The impact of hospital-wide use of a tapered-cuff endotracheal tube on the incidence of ventilator-associated pneumonia.

    PubMed

    Bowton, David L; Hite, R Duncan; Martin, R Shayn; Sherertz, Robert

    2013-10-01

    Aspiration of colonized oropharyngeal secretions is a major factor in the pathogenesis of ventilator-associated pneumonia (VAP). A tapered-cuff endotracheal tube (ETT) has been demonstrated to reduce aspiration around the cuff. Whether these properties are efficacious in reducing VAP is not known. This 2-period, investigator-initiated observational study was designed to assess the efficacy of a tapered-cuff ETT to reduce the VAP rate. All intubated, mechanically ventilated patients over the age of 18 were included. During the baseline period a standard, barrel-shaped-cuff ETT (Mallinckrodt Hi-Lo) was used. All ETTs throughout the hospital were then replaced with a tapered-cuff ETT (TaperGuard). The primary outcome variable was the incidence of VAP per 1,000 ventilator days. We included 2,849 subjects, encompassing 15,250 ventilator days. The mean ± SD monthly VAP rate was 3.29 ± 1.79/1,000 ventilator days in the standard-cuff group and 2.77 ± 2.00/1,000 ventilator days in the tapered-cuff group (P = .65). While adherence to the VAP prevention bundle was high throughout the study, bundle adherence was significantly higher during the standard-cuff period (96.5 ± 2.7%) than in the tapered-cuff period (90.3 ± 3.5%, P = .01). In the setting of a VAP rate very near the average of ICUs in the United States, and where there was high adherence to a VAP prevention bundle, the use of a tapered-cuff ETT was not associated with a reduction in the VAP rate.

  5. Cerebral glucose deficiency versus oxygen deficiency in neonatal encephalopathy.

    PubMed

    Rudolph, A M

    2018-04-24

    Hypoxic-ischemic encephalopathy (HIE) in newborn infants is generally considered to result from decreased arterial oxygen content or cerebral blood flow. Cerebral injury similar to that of HIE has been noted with hypoglycemia. Studies in fetal lambs have shown that ventilation with 3% oxygen did not change cerebral blood flow, but ventilation with 100% oxygen resulted in marked reduction in cerebral blood flow, glucose delivery and glucose consumption. Blood glucose concentration falls markedly after birth; this, associated with the fall in cerebral blood flow, greatly reduces glucose supply to the brain. In preterm infants, blood glucose levels tend to be very low. Also persistent patency of the ductus arteriosus may reduce cerebral flow in diastole, thus exaggerating the decrease in glucose supply. I propose that glycopenic-ischemic encephalopathy is a more appropriate term for the cerebral insult. We should consider more aggressive management of the low blood glucose concentrations in the neonate, and particularly in preterm infants. Administration of high levels of oxygen in inspired air should be avoided to reduce the enhancement of cerebral vasoconstriction and decreased flow that normally occurs after birth.

  6. Effects of respiratory muscle work on respiratory and locomotor blood flow during exercise.

    PubMed

    Dominelli, Paolo B; Archiza, Bruno; Ramsook, Andrew H; Mitchell, Reid A; Peters, Carli M; Molgat-Seon, Yannick; Henderson, William R; Koehle, Michael S; Boushel, Robert; Sheel, A William

    2017-11-01

    What is the central question of this study? Does manipulation of the work of breathing during high-intensity exercise alter respiratory and locomotor muscle blood flow? What is the main finding and its importance? We found that when the work of breathing was reduced during exercise, respiratory muscle blood flow decreased, while locomotor muscle blood flow increased. Conversely, when the work of breathing was increased, respiratory muscle blood flow increased, while locomotor muscle blood flow decreased. Our findings support the theory of a competitive relationship between locomotor and respiratory muscles during intense exercise. Manipulation of the work of breathing (WOB) during near-maximal exercise influences leg blood flow, but the effects on respiratory muscle blood flow are equivocal. We sought to assess leg and respiratory muscle blood flow simultaneously during intense exercise while manipulating WOB. Our hypotheses were as follows: (i) increasing the WOB would increase respiratory muscle blood flow and decrease leg blood flow; and (ii) decreasing the WOB would decrease respiratory muscle blood flow and increase leg blood flow. Eight healthy subjects (n = 5 men, n = 3 women) performed a maximal cycle test (day 1) and a series of constant-load exercise trials at 90% of peak work rate (day 2). On day 2, WOB was assessed with oesophageal balloon catheters and was increased (via resistors), decreased (via proportional assist ventilation) or unchanged (control) during the trials. Blood flow was assessed using near-infrared spectroscopy optodes placed over quadriceps and the sternocleidomastoid muscles, coupled with a venous Indocyanine Green dye injection. Changes in WOB were significantly and positively related to changes in respiratory muscle blood flow (r = 0.73), whereby increasing the WOB increased blood flow. Conversely, changes in WOB were significantly and inversely related to changes in locomotor blood flow (r = 0.57), whereby decreasing the WOB increased locomotor blood flow. Oxygen uptake was not different during the control and resistor trials (3.8 ± 0.9 versus 3.7 ± 0.8 l min -1 , P > 0.05), but was lower on the proportional assist ventilator trial (3.4 ± 0.7 l min -1 , P < 0.05) compared with control. Our findings support the concept that respiratory muscle work significantly influences the distribution of blood flow to both respiratory and locomotor muscles. © 2017 The Authors. Experimental Physiology © 2017 The Physiological Society.

  7. Effect of one-lung ventilation on end-tidal carbon dioxide during cardiopulmonary resuscitation in a pig model of cardiac arrest.

    PubMed

    Ryu, Dong Hyun; Jung, Yong Hun; Jeung, Kyung Woon; Lee, Byung Kook; Jeong, Young Won; Yun, Jong Geun; Lee, Dong Hun; Lee, Sung Min; Heo, Tag; Min, Yong Il

    2018-01-01

    Unrecognized endobronchial intubation frequently occurs after emergency intubation. However, no study has evaluated the effect of one-lung ventilation on end-tidal carbon dioxide (ETCO2) during cardiopulmonary resuscitation (CPR). We compared the hemodynamic parameters, blood gases, and ETCO2 during one-lung ventilation with those during conventional two-lung ventilation in a pig model of CPR, to determine the effect of the former on ETCO2. A randomized crossover study was conducted in 12 pigs intubated with double-lumen endobronchial tube to achieve lung separation. During CPR, the animals underwent three 5-min ventilation trials based on a randomized crossover design: left-lung, right-lung, or two-lung ventilation. Arterial blood gases were measured at the end of each ventilation trial. Ventilation was provided using the same tidal volume throughout the ventilation trials. Comparison using generalized linear mixed model revealed no significant group effects with respect to aortic pressure, coronary perfusion pressure, and carotid blood flow; however, significant group effect in terms of ETCO2 was found (P < 0.001). In the post hoc analyses, ETCO2 was lower during the right-lung ventilation than during the two-lung (P = 0.006) or left-lung ventilation (P < 0.001). However, no difference in ETCO2 was detected between the left-lung and two-lung ventilations. The partial pressure of arterial carbon dioxide (PaCO2), partial pressure of arterial oxygen (PaO2), and oxygen saturation (SaO2) differed among the three types of ventilation (P = 0.003, P = 0.001, and P = 0.001, respectively). The post hoc analyses revealed a higher PaCO2, lower PaO2, and lower SaO2 during right-lung ventilation than during two-lung or left-lung ventilation. However, the levels of these blood gases did not differ between the left-lung and two-lung ventilations. In a pig model of CPR, ETCO2 was significantly lower during right-lung ventilation than during two-lung ventilation. However, interestingly, ETCO2 during left-lung ventilation was comparable to that during two-lung ventilation.

  8. Concomitant administration of nitrous oxide and remifentanil reduces oral tissue blood flow without decreasing blood pressure during sevoflurane anesthesia in rabbits.

    PubMed

    Kasahara, Masataka; Ichinohe, Tatsuya; Okamoto, Sota; Okada, Reina; Kanbe, Hiroaki; Matsuura, Nobuyuki

    2015-06-01

    To determine whether continuous administration of nitrous oxide and remifentanil—either alone or together—alters blood flow in oral tissues during sevoflurane anesthesia. Eight male tracheotomized Japanese white rabbits were anesthetized with sevoflurane under mechanical ventilation. Heart rate (HR), systolic blood pressure (SBP), diastolic blood pressure (DBP), mean arterial pressure (MAP), common carotid arterial blood flow (CCBF), tongue mucosal blood flow (TMBF), mandibular bone marrow blood flow (BBF), masseter muscle blood flow (MBF), upper alveolar tissue blood flow (UBF), and lower alveolar tissue blood flow (LBF) were recorded in the absence of all test agents and after administration of the test agents (50 % nitrous oxide, 0.4 μg/kg/min remifentanil, and their combination) for 20 min. Nitrous oxide increased SBP, DBP, MAP, CCBF, BBF, MBF, UBF, and LBF relative to baseline values but did not affect HR or TMBF. Remifentanil decreased all hemodynamic variables except DBP. Combined administration of nitrous oxide and remifentanil recovered SBP, DBP, MAP, and CCBF to baseline levels, but HR and oral tissue blood flow remained lower than control values. Our findings suggest that concomitant administration of nitrous oxide and remifentanil reduces blood flow in oral tissues without decreasing blood pressure during sevoflurane anesthesia in rabbits.

  9. Formaldehyde and acetaldehyde exposure mitigation in US residences: In-home measurements of ventilation control and source control

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Hult, Erin L.; Willem, Henry; Price, Phillip N.

    2014-10-01

    Measurements were taken in new US residences to assess the extent to which ventilation and source control can mitigate formaldehyde exposure. Increasing ventilation consistently lowered indoor formaldehyde concentrations. However, at a reference air exchange rate of 0.35 h -1, increasing ventilation was up to 60% less effective than would be predicted if the emission rate were constant. This is consistent with formaldehyde emission rates decreasing as air concentrations increase, as observed in chamber studies. In contrast, measurements suggest acetaldehyde emission was independent of ventilation rate. To evaluate the effectiveness of source control, formaldehyde concentrations were measured in Leadership in Energymore » and Environmental Design (LEED) certified/Indoor airPLUS homes constructed with materials certified to have low emission rates of volatile organic compounds (VOC). At a reference air exchange rate of 0.35 h -1, and adjusting for home age, temperature and relative humidity, formaldehyde concentrations in homes built with low-VOC materials were 42% lower on average than in reference new homes with conventional building materials. Without adjustment, concentrations were 27% lower in the low-VOC homes. The mean and standard deviation of formaldehyde concentration were 33 μg m-3 and 22 μg m -3 for low-VOC homes and 45 μg m -3 and 30 μg m -3 for conventional.« less

  10. Prototype Systems for Measuring Outdoor Air Intake Rates in Rooftop Air Handlers

    DOE Office of Scientific and Technical Information (OSTI.GOV)

    Fisk, William J.; Chan, Wanyu R.; Hotchi, Toshifumi

    2015-01-01

    The widespread absence of systems for real-time measurement and feedback control, of minimum outdoor air intake rates in HVAC systems contributes to the poor control of ventilation rates in commercial buildings. Ventilation rates affect building energy consumption and influence occupant health. The project designed fabricated and tested four prototypes of systems for measuring rates of outdoor air intake into roof top air handlers. All prototypes met the ±20% accuracy target at low wind speeds, with all prototypes accurate within approximately ±10% after application of calibration equations. One prototype met the accuracy target without a calibration. With two of four prototypemore » measurement systems, there was no evidence that wind speed or direction affected accuracy; however, winds speeds were generally below usually 3.5 m s -1 (12.6 km h -1) and further testing is desirable. The airflow resistance of the prototypes was generally less than 35 Pa at maximum RTU air flow rates. A pressure drop of this magnitude will increase fan energy consumption by approximately 4%. The project did not have resources necessary to estimate costs of mass produced systems. The retail cost of components and materials used to construct prototypes ranged from approximately $1,200 to $1,700. The test data indicate that the basic designs developed in this project, particularly the designs of two of the prototypes, have considerable merit. Further design refinement, testing, and cost analysis would be necessary to fully assess commercial potential. The designs and test results will be communicated to the HVAC manufacturing community.« less

  11. Effect of tubing deposition, breathing pattern, and temperature on aerosol mass distribution measured by cascade impactor.

    PubMed

    Gurses, Burak K; Smaldone, Gerald C

    2003-01-01

    Aerosols produced by nebulizers are often characterized on the bench using cascade impactors. We studied the effects of connecting tubing, breathing pattern, and temperature on mass-weighted aerodynamic particle size aerosol distributions (APSD) measured by cascade impaction. Our experimental setup consisted of a piston ventilator, low-flow (1.0 L/min) cascade impactor, two commercially available nebulizers that produced large and small particles, and two "T"-shaped tubes called "Tconnector(cascade)" and "Tconnector(nebulizer)" placed above the impactor and the nebulizer, respectively. Radiolabeled normal saline was nebulized using an airtank at 50 PSIG; APSD, mass balance, and Tconnector(cascade) deposition were measured with a gamma camera and radioisotope calibrator. Flow through the circuit was defined by the air tank (standing cloud, 10 L/min) with or without a piston pump, which superimposed a sinusoidal flow on the flow from the air tank (tidal volume and frequency of breathing). Experiments were performed at room temperature and in a cooled environment. With increasing tidal volume and frequency, smaller particles entered the cascade impactor (decreasing MMAD; e.g., Misty-Neb, 4.2 +/- 0.9 microm at lowest ventilation and 2.7 +/- 0.1 microm at highest, p = 0.042). These effects were reduced in magnitude for the nebulizer that produced smaller particles (AeroTech II, MMAD 1.8 +/- 0.1 to 1.3 +/- 0.1 microm; p = 0.0044). Deposition on Tconnector(cascade) increased with ventilation but was independent of cascade impactor flow. Imaging of the Tconnector(cascade) revealed a pattern of deposition unaffected by cascade impactor flow. These measurements suggest that changes in MMAD with ventilation were not artifacts of tubing deposition in the Tconnector(cascade). At lower temperatures, APSD distributions were more polydisperse. Our data suggest that, during patient inhalation, changes in particle distribution occur that are related to conditions in the tubing and may reduce the diameters of particles entering the patient. This effect is more significant for nebulizers producing large particles. Changes in ambient temperature did not affect these observations.

  12. Hypopnea consequent to reduced pulmonary blood flow in the dog.

    PubMed

    Stremel, R W; Whipp, B J; Casaburi, R; Huntsman, D J; Wasserman, K

    1979-06-01

    The ventilatory responses to diminished pulmonary blood flow (Qc), as a result of partial cardiopulmonary bypass (PCB), were studied in chloralose-urethan-anesthetized dogs. Qc was reduced by diverting vena caval blood through a membrane gas exchanger and returning it to the ascending aorta. PCB flows of 400--1,600 ml/min were utilized for durations of 2--3 min. Decreasing Qc, while maintaining systemic arterial blood gases and perfusion, results in a significant (P less than 0.05) decrease in expiratory ventilation (VE) (15.9%) and alveolar ventilation (VA) (31.0%). The ventilatory decreases demonstrated for this intact group persist after bilateral cervical vagotomy (Vx), carotid body and carotid sinus denervation (Cx), and combined Vx and Cx. The changes in VE and VA were significantly (P less than 0.001) correlated with VCO2 changes, r = 0.80 and r = 0.93, respectively. These ventilatory changes were associated with an overall average decrease in left ventricular PCO2 of 2.1 Torr; this decrease was significant (P less than 0.05) only in the intact and Cx groups. Decreasing pulmonary blood flow results in a decrease in ventilation that may be CO2 related; however, the exact mechanism remains obscure but must have a component that is independent of vagally mediated cardiac and pulmonary afferents and peripheral baroreceptor and chemoreceptor afferents.

  13. DOE Office of Scientific and Technical Information (OSTI.GOV)

    Scofield, C.M.; Des Champs, N.H.

    This article examines a design concept for classroom air conditioning systems that guarantees minimum ventilation rates are met. The topics of the article include new ventilation requirements, design concept, outside air induction diffuser, low-velocity ducts and plenums, the relationship of humidity to school absenteeism rates, retrofit applications, and saving energy.

  14. Quantitative relationship of sick building syndrome symptoms with ventilation rates

    EPA Science Inventory

    Data from published studies were combined and analyzed to develop best-fit equations and curves quantifying the change in sick building syndrome (SBS) symptom prevalence in office workers with ventilation rate. For each study, slopes were calculated, representing the fractional...

  15. Hepatic Blood Perfusion Estimated by Dynamic Contrast-Enhanced Computed Tomography in Pigs Limitations of the Slope Method

    PubMed Central

    Winterdahl, Michael; Sørensen, Michael; Keiding, Susanne; Mortensen, Frank V.; Alstrup, Aage K. O.; Hansen, Søren B.; Munk, Ole L.

    2012-01-01

    Objective To determine whether dynamic contrast-enhanced computed tomography (DCE-CT) and the slope method can provide absolute measures of hepatic blood perfusion from hepatic artery (HA) and portal vein (PV) at experimentally varied blood flow rates. Materials and Methods Ten anesthetized 40-kg pigs underwent DCE-CT during periods of normocapnia (normal flow), hypocapnia (decreased flow), and hypercapnia (increased flow), which was induced by adjusting the ventilation. Reference blood flows in HA and PV were measured continuously by surgically-placed ultrasound transit-time flowmeters. For each capnic condition, the DCE-CT estimated absolute hepatic blood perfusion from HA and PV were calculated using the slope method and compared with flowmeter based absolute measurements of hepatic perfusions and relative errors were analyzed. Results The relative errors (mean±SEM) of the DCE-CT based perfusion estimates were −21±23% for HA and 81±31% for PV (normocapnia), 9±23% for HA and 92±42% for PV (hypocapnia), and 64±28% for HA and −2±20% for PV (hypercapnia). The mean relative errors for HA were not significantly different from zero during hypo- and normocapnia, and the DCE-CT slope method could detect relative changes in HA perfusion between scans. Infusion of contrast agent led to significantly increased hepatic blood perfusion, which biased the PV perfusion estimates. Conclusions Using the DCE-CT slope method, HA perfusion estimates were accurate at low and normal flow rates whereas PV perfusion estimates were inaccurate and imprecise. At high flow rate, both HA perfusion estimates were significantly biased. PMID:22836307

  16. Carbon dioxide clearance in rabbits during expiratory phase intratracheal pulmonary ventilation.

    PubMed

    Meyappan, Raju T; Raszynski, Andre; Bohorquez, Jorge; Totapally, Balagangadhar R; Koul, Pulin B; Norozian, Faraz M; Valcourt, Karl; Torbati, Dan

    2007-01-01

    The purpose of this study was to compare the efficacy of CO2 removal during conventional mechanical ventilation (CMV) with and without expiratory phase intratracheal pulmonary ventilation (expiratory ITPV or Exp-ITPV); and to compare CO2 clearance during Exp-ITPV, in pressure-controlled ventilation (PCV) and in volume-controlled ventilation (VCV) modes. Seven anesthetized rabbits were tracheotomized and intubated using a 4 mm endotracheal tube. Venous and arterial lines were established. The rabbits were paralyzed, mechanically ventilated, and ventilation parameters were adjusted to achieve baseline arterial hypercapnia. Animals were then ventilated during 30-minute trials of CMV and Exp-ITPV, in both PCV and VCV modes. A custom-built, microprocessor-controlled solenoid valve was used to limit ITPV gas flow to the expiratory phase. Proximal and carinal airway pressures and hemodynamic variables were continuously recorded, and arterial blood gases were analyzed at the end of each trial. Exp-ITPV, as compared with CMV, reduced arterial PCO2 by 12% and 21% in PCV and VCV modes, respectively (p < 0.02 and p < 0.001; one-sided paired t test), without significant changes in other cardiorespiratory variables. In conclusion, Exp-ITPV is more effective than CMV in clearing CO2 through a small endotracheal tube. Exp-ITPV is also more effective in VCV mode than PCV mode.

  17. Transnasal Humidified Rapid-Insufflation Ventilatory Exchange (THRIVE): a physiological method of increasing apnoea time in patients with difficult airways.

    PubMed

    Patel, A; Nouraei, S A R

    2015-03-01

    Emergency and difficult tracheal intubations are hazardous undertakings where successive laryngoscopy-hypoxaemia-re-oxygenation cycles can escalate to airway loss and the 'can't intubate, can't ventilate' scenario. Between 2013 and 2014, we extended the apnoea times of 25 patients with difficult airways who were undergoing general anaesthesia for hypopharyngeal or laryngotracheal surgery. This was achieved through continuous delivery of transnasal high-flow humidified oxygen, initially to provide pre-oxygenation, and continuing as post-oxygenation during intravenous induction of anaesthesia and neuromuscular blockade until a definitive airway was secured. Apnoea time commenced at administration of neuromuscular blockade and ended with commencement of jet ventilation, positive-pressure ventilation or recommencement of spontaneous ventilation. During this time, upper airway patency was maintained with jaw-thrust. Transnasal Humidified Rapid-Insufflation Ventilatory Exchange (THRIVE) was used in 15 males and 10 females. Mean (SD [range]) age at treatment was 49 (15 [25-81]) years. The median (IQR [range]) Mallampati grade was 3 (2-3 [2-4]) and direct laryngoscopy grade was 3 (3-3 [2-4]). There were 12 obese patients and nine patients were stridulous. The median (IQR [range]) apnoea time was 14 (9-19 [5-65]) min. No patient experienced arterial desaturation < 90%. Mean (SD [range]) post-apnoea end-tidal (and in four patients, arterial) carbon dioxide level was 7.8 (2.4 [4.9-15.3]) kPa. The rate of increase in end-tidal carbon dioxide was 0.15 kPa.min(-1) . We conclude that THRIVE combines the benefits of 'classical' apnoeic oxygenation with continuous positive airway pressure and gaseous exchange through flow-dependent deadspace flushing. It has the potential to transform the practice of anaesthesia by changing the nature of securing a definitive airway in emergency and difficult intubations from a pressured stop-start process to a smooth and unhurried undertaking. © 2014 The Authors Anaesthesia published by John Wiley & Sons Ltd on behalf of Association of Anaesthetists of Great Britain and Ireland.

  18. Numerical investigation of pulmonary drug delivery under mechanical ventilation conditions

    NASA Astrophysics Data System (ADS)

    Banerjee, Arindam; van Rhein, Timothy

    2012-11-01

    The effects of mechanical ventilation waveform on fluid flow and particle deposition were studied in a computer model of the human airways. The frequency with which aerosolized drugs are delivered to mechanically ventilated patients demonstrates the importance of understanding the effects of ventilation parameters. This study focuses specifically on the effects of mechanical ventilation waveforms using a computer model of the airways of patient undergoing mechanical ventilation treatment from the endotracheal tube to generation G7. Waveforms were modeled as those commonly used by commercial mechanical ventilators. Turbulence was modeled with LES. User defined particle force models were used to model the drag force with the Cunningham correction factor, the Saffman lift force, and Brownian motion force. The endotracheal tube (ETT) was found to be an important geometric feature, causing a fluid jet towards the right main bronchus, increased turbulence, and a recirculation zone in the right main bronchus. In addition to the enhanced deposition seen at the carinas of the airway bifurcations, enhanced deposition was also seen in the right main bronchus due to impaction and turbulent dispersion resulting from the fluid structures created by the ETT. Authors acknowledge financial support through University of Missouri Research Board Award.

  19. Transpulmonary hypothermia: a novel method of rapid brain cooling through augmented heat extraction from the lungs.

    PubMed

    Kumar, Matthew M; Goldberg, Andrew D; Kashiouris, Markos; Keenan, Lawrence R; Rabinstein, Alejandro A; Afessa, Bekele; Johnson, Larry D; Atkinson, John L D; Nayagam, Vedha

    2014-10-01

    Delay in instituting neuroprotective measures after cardiac arrest increases death and decreases neuronal recovery. Current hypothermia methods are slow, ineffective, unreliable, or highly invasive. We report the feasibility of rapid hypothermia induction in swine through augmented heat extraction from the lungs. Twenty-four domestic crossbred pigs (weight, 50-55kg) were ventilated with room air. Intraparenchymal brain temperature and core temperatures from pulmonary artery, lower esophagus, bladder, rectum, nasopharynx, and tympanum were recorded. In eight animals, ventilation was switched to cooled helium-oxygen mixture (heliox) and perfluorocarbon (PFC) aerosol and continued for 90min or until target brain temperature of 32°C was reached. Eight animals received body-surface cooling with water-circulating blankets; eight control animals continued to be ventilated with room air. Brain and core temperatures declined rapidly with cooled heliox-PFC ventilation. The brain reached target temperature within the study period (mean [SD], 66 [7.6]min) in only the transpulmonary cooling group. Cardiopulmonary functions and poststudy histopathological examination of the lungs were normal. Transpulmonary cooling is novel, rapid, minimally invasive, and an effective technique to induce therapeutic hypothermia. High thermal conductivity of helium and vaporization of PFC produces rapid cooling of alveolar gases. The thinness and large surface area of alveolar membrane facilitate rapid cooling of the pulmonary circulation. Because of differences in thermogenesis, blood flow, insulation, and exposure to the external environment, the brain cools at a different rate than other organs. Transpulmonary hypothermia was significantly faster than body surface cooling in reaching target brain temperature. Copyright © 2014 Elsevier Ireland Ltd. All rights reserved.

  20. Influence of chest compression artefact on capnogram-based ventilation detection during out-of-hospital cardiopulmonary resuscitation.

    PubMed

    Leturiondo, Mikel; Ruiz de Gauna, Sofía; Ruiz, Jesus M; Julio Gutiérrez, J; Leturiondo, Luis A; González-Otero, Digna M; Russell, James K; Zive, Dana; Daya, Mohamud

    2018-03-01

    Capnography has been proposed as a method for monitoring the ventilation rate during cardiopulmonary resuscitation (CPR). A high incidence (above 70%) of capnograms distorted by chest compression induced oscillations has been previously reported in out-of-hospital (OOH) CPR. The aim of the study was to better characterize the chest compression artefact and to evaluate its influence on the performance of a capnogram-based ventilation detector during OOH CPR. Data from the MRx monitor-defibrillator were extracted from OOH cardiac arrest episodes. For each episode, presence of chest compression artefact was annotated in the capnogram. Concurrent compression depth and transthoracic impedance signals were used to identify chest compressions and to annotate ventilations, respectively. We designed a capnogram-based ventilation detection algorithm and tested its performance with clean and distorted episodes. Data were collected from 232 episodes comprising 52 654 ventilations, with a mean (±SD) of 227 (±118) per episode. Overall, 42% of the capnograms were distorted. Presence of chest compression artefact degraded algorithm performance in terms of ventilation detection, estimation of ventilation rate, and the ability to detect hyperventilation. Capnogram-based ventilation detection during CPR using our algorithm was compromised by the presence of chest compression artefact. In particular, artefact spanning from the plateau to the baseline strongly degraded ventilation detection, and caused a high number of false hyperventilation alarms. Further research is needed to reduce the impact of chest compression artefact on capnographic ventilation monitoring. Copyright © 2017 Elsevier B.V. All rights reserved.

Top